CT-to-TPS-to-verification summary of simulation process
Vibha Chaswal, Ph.D.
Know the co-ordinate system TPS uses
Patient co-ordinate system (not used in Pinnacle)
CT co-ordinate system
Image data on acquisition on CT
• Image request form: Patient positioned using lap lasers (ct bb’s on lap lasers)
• Immobilization devices marked• Orientation and technique (protocol) selected
in CT• Topogram acquisition• CT scan acquisition• Dicom images send to TPS and archive
TPS processes• Image data imported• Scanning information verified ( patient position,
orientation)• CT bb’s are localized w.r.t. CT room lasers co-ordinate
system• ROIs are contoured and isocenter is defined w.r.t machine
co-ordinate system• Shifts from ct-bb’s to isocenter are given w.r.t treatment
table (up/down, L/R, in/out)• Segmentation, TP (beams, blocks, dose) and DRRs
generated (set-up and field)• Transfer TP to record and verify (LEX), DRRs to machine
(coherence)
Verification processes• Set patient to CT-bb’s• Get shifts to isocenter from DRR’s, make shifts• Check SSD• Take portal image using field DRRs or Set-up
DRRs (orthogonal and oblique DRRs)• Compare DRR with portal image and verify shift
(orthogonal gives sup-inf and left-right info oblique gives A-P)
• Mark lasers on patient if everything matches, these are treatment isocenter localization marks
• Bite-block – physicist verifies the physical positioning of bite-block before CT scan, radiocam localization no ct bb’s
• Cone-beam – Cone beam field with CT image and segmented structures sent to machine, MVCT image compared with the help of bony land-marks and contours