34
QUOLIS 5000 DENTAL UNIT INSTALLATION INSTRUCTIONS IMPORTANT This manual provides installation instructions for the BELMONT QUOLIS5000 UNIT. The instructions contained in this booklet should be thoroughly read and understood before installing the chair and unit. After the installation has been completed, keep this manual in a safe place and refer to it for future maintenance. R

Transcript of QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual,...

Page 1: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

QUOLIS 5000 DENTAL UNIT

INSTALLATION INSTRUCTIONS

IMPORTANT

This manual provides installation instructions for the BELMONT QUOLIS5000 UNIT. The instructions contained in this booklet should be thoroughly read and understood before installing the chair and unit.After the installation has been completed, keep this manual in a safe place and refer to it for future maintenance.

R

Page 2: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve
Page 3: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

Intended Use of the Product This product is an active therapeutic device intended to administer or exchange energy of electric, air

and water for the exclusive use for diagnoses, treatments and relative procedures of dentistry, and its

characteristic is not in a potentially hazardous way between such energy and human body, taking

account of the nature, the density and site of application of the energy.

The product must be operated or handled by the qualified dentists or by dental staffs under the

supervision of the dentist.

Such dentists or dental staffs should instruct and/or assist the patients to approach to and leave from

the product.

Patients should not be allowed to operate or handle the product unless he/she is so instructed.

The product is supplied together with the handpieces like electric micromotor, air turbine and/or

motor, scaler and so on.

Environmental Requirements Ambient Temperature Operating 41F to 104F Storage 14F to 122F

Humidity 10 % - 80%

Atmospheric Pressure 8.7psi – 15.4psi

Classification of Eqipment a. Type of shock protection : Class I Equipment

b. Degree of shock protection : Type B Applied part

c. Degree of protection against water ingress : Ordinary Equipment (All Products)

d. Mode of operation : Continuous Operation

e. Flammable Gasses : Not suitable for use in the presence of a flammable anesthetic mixture with air,

oxygen, or nitrous oxide.

f. Earth : Protective earth (ground)

g. Hospital Grade Plug : Grounding reliability can be achieved only when the equipment is connected to an

equivalent receptacle marked hospital only or hospital grade.

CAUTION

Ground reliability can only be achieved when the equipment is connected to an equivalent receptacle marked HOSPITAL only or HOSPITAL GRADE.

Page 4: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

TABLE OF CONTENTS Page 1 . Overview and Major Components ----------------------------------------- 1 1-1. Vac Pac System 1-2. PMU System 2.DimensionsandSpecifications ----------------------------------------------- 3 2-1. Vac Pac System 2-2. PMU System 3 . Introduction -------------------------------------------------------------------- 5 3-1. Precautions 3-2. Necessary tools 4 . Unpacking of Units ------------------------------------------------------------ 5 5 . Swing Arm ---------------------------------------------------------------------- 7 5-1. How to Install the Swing Arm 6 . Junction Box -------------------------------------------------------------------- 7 6-1. How to Install the Junction Box 7 . Balance Arm for the Doctor’ s Control ----------------------------------- 8 7-1. How to Install the Balance Arm 7-2. How to Plumb the Doctor’ s Control 7-3. How to Plumb the Water Bottle 7-4. How to Install the Foot Control 8 . Monitor Mount (Optional) -------------------------------------------------- 11 8-1. How to Install the Monitor Arm 9 . Light Pole (Optional) --------------------------------------------------------- 12 9-1. How to Install the Light Pole (Optional) 10 . Vac Pac (Optional) ------------------------------------------------------------- 12 10-1. How to Install the Vac Pac 10-2. How to Plumb the J-Box Utility 11 . PMU (Optional) --------------------------------------------------------------- 14 11-1. How to Install the PMU 11-2. How to Install the Cuspidor Bowl 11-3. How to Plumb the J-Box Utility 11-4. How to install accessories 12 . Final Adjustment --------------------------------------------------------------- 17 12-1. Swing Arm 12-2. How to Adjust the Swing Arm Friction 12-3. How to Adjust the Rotation Tension of Assistant Arm 12-4. How to adjust the water flow volume of Cup fill and Bowl Rins13.ConnectthePowerCableandConfirmsfunctionlity ------------------ 18 14 . Option --------------------------------------------------------------------------- 18 14-1. How to Install the Tray Holder 15 . Flow Diagram ------------------------------------------------------------------- 19 15-1. Doctor’ s Table Section 15-2. Swing Arm Section 15-3. Vac Pac Section 15-4. PMU Section 15-5. Junction Box Section 15-6. Overall Unit Flow Diagram(Vac Pac System) 15-7. Overall Unit Flow Diagram(PMU System) 16 . Electrical Diagram ----------------------------------------------------------- 25 16-1. Chair Section 16-2. Doctor’s Table Section 16-3. PMU Section 17 . Location of Product Labels ------------------------------------------------- 28 18 . Floor Template ---------------------------------------------------------------- 29

Page 5: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

14

15

16

24

17 1918

21

2223

20

1

2

34

78

9

10

11

12

13

56

1. Overview and Major Components 1-1. Vac Pac System

(1) Control Head

(2) Main Control Panel

(3) Digital Display

(4) Handle

(5) Handpiece Holders

(6) Dr’s Syringe

(7) Handpiece Hose

(8) Foot Control

(9) Balance Arm

(10) Dr Swing Arm]

(11) Water Bottle

(12) Junction Box

(13) Dental Light Swing Arm

(14) Light Pole

(15) Assistant Instrument Holder

(16) HVE

(17) SE

(18) Assistant’s Syringe

(19) Assistant Touch Pad

(20) Solids Collector

(21) Assistant Swing Arm

(22) Assistant Holder Arm

(23) Vacuum Pack Housing

(24) Dental Light

(*Note) Quolis 5000 series does not include syringe tips. Syringe tips manufactured by DCI are

compatible with this device.

- 1 -

Page 6: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

14

15

16

24

17

19

18

21

22

2320

1

2

34

7

8

9

10

11

1213

5 6

25 2627

28

1. Overview and Major Components 1-2. PMU System

(1) Control Head

(2) Main Control Panel

(3) Digital Display

(4) Handle

(5) Handpiece Holders

(6) Dr’s Syringe (*)

(7) Handpiece Hose

(8) Foot Control

(9) Balance Arm

(10) Dr Swing Arm

(11) Water Bottle

(12) Junction Box

(13) Dental Light Swing Arm

(14) Light Pole

(15) Assistant Instrument Holder

(16) HVE

(17) SE

(18) Assistant’s Syringe (*)

(19) Assistant Touch Pad

(20) Solids Collector

(21) Assistant Swing Arm

(22) Assistant Holder Arm

(23) Assistant Section

(24) Dental Light

(25) Cuspidor Bowl

(26) Cup filler switch and Bowl Rinse switch

(27) Cup filler nozzle and Bowl Rinse nozzle

(28) Q.D.Section

(*Note) Quolis 5000 series does not include syringe tips. Syringe tips manufactured by DCI are

compatible with this device.

- 2 -

Page 7: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

2. Dimensions and Specifications 2-1. Vac Pac System 2-1-1. Dimensions - Inch / (mm) - Tolerance (± 10%)

2-1-2. Specifications Power Consumption ------------------------------- AC120V 1A Frequency ------------------------------------------- 60 Hz Fuse value ------------------------------------------- AC125V 10A Control Voltage ------------------------------------- DC5V Dr’s control Net Weight --------------------------- 55 lbs (25 kg) Swing Arm Net Weight ---------------------------- 55 lbs (25 kg) (Without Dental Light) Vacuum Pack Net Weight ------------------------- 44 lbs (20 kg) Junction Net Weight ------------------------------- 16 lbs (7 kg) Doctor Table Maximum Load -------------------- 4.40 lbs (2 kg) Operating Pressure --------------------------------- Water 30-40 psi (0.2-0.28 Mpa), Air 75 psi (0.5 MPa)

- 3 -

Page 8: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

2. Dimensions and Specifications 2-2. PMU System 2-2-1. Dimensions - Inch / (mm) - Tolerance (± 10%)

2-2-2. Specifications Power Consumption ------------------------------- AC120V 1A Frequency ------------------------------------------- 60 Hz Fuse value ------------------------------------------- AC125V 10A Control Voltage ------------------------------------- DC5V Dr’s control Net Weight --------------------------- 55 lbs (25 kg) Swing Arm Net Weight ---------------------------- 55 lbs (25 kg) (Without Dental Light) PMU Net Weight ----------------------------------- 66 lbs (30 kg) Junction Net Weight ------------------------------- 16 lbs (7 kg) Doctor Table Maximum Load -------------------- 4.40 lbs (2 kg) Operating Pressure --------------------------------- Water 30-40 psi (0.2-0.28 Mpa), Air 75 psi (0.5 MPa)

- 4 -

Page 9: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

3. Introduction 3-1. Precautions for installation

• Do not connect to power supply other than 120V/60Hz.

• Ground the unit properly.

• Attach the junction box firmly to the floor.

• When electric micromotor (option) or electric scaler (option) is installed, please refer to the manual

for the each device.

• Do not drop or hit the chair.

• When the installation process has been completed, verify that all the mechanical and electrical

functions are working properly and that there is no evidence of oil, water or air leakage.

• Attach the dental chair to the floor before mounting the unit.

• Refer to the installation manual of Quolis 5000 dental chair to know how to install the dental chair.

• Refer to Floor Template for connection of water, air, power, drain hose and vacuum hose.

3-2. Necessary Tools

• Phillips screw drivers set #1, #2.

• Slotted screw driver

• Hex key wenches set (metric, 2.5mm – 8mm)

• Adjustable wrench(over 24mm)

• Long nose pliers

• Wire cutter

• Socket set 17mm, 7mm, 1/4”

• Level

4. Unpacking of units Each carton number corresponds to a component

Check all the components and parts

damage after unpacking.

Opening cartons with a sharp object may cause damage to parts

CAUTION

Box No. Module

3 Doctor’s control

4 Swing arm (s)

5 Vac pacPMU

No. 3No.5

No.4

6

No.6

- 5 -

Page 10: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

Box No.3 (Doctor’s control)

(1) Doctor's control assembly 1 set

(2) Junction box assembly 1 set • Junction box with cover 1 set • Duct hose 800mm 1 pc. • Utility package (air)

(3) Balance arm end cap 1pc.

(4) Stainless tray and non-slip pad 1 set

(5) Foot controller 1 set

(6) Syringe parts 1 set

(7) Operating manual, installation manual 1 set

(8) Miscellaneous parts for installation as follow • Sleeve for 1/4” OD tubing 14 pcs. • Sleeve for 1/8” OD tubing 12 pcs. • Barb fitting for 1/4” OD tubing 3 pcs. • Barb fitting for 1/8” OD tubing 3 pc. • Hex plug 2 pcs. • Tubing 1/8” OD brown 500mm 1 pc. • Tubing 1/4 Green 500mm 1 pc. • Panhead Philips screw M4 x 10mm 2 pcs • Round head Philips wood screw M5 x 25mm 3 pcs.

Box No.4 (Swing arms)

(9) Swing arm Assembly 1 set

(10) Water bottle 1 pc

(11) Lower tubing cover 1 pc.

(12) Quick disconnect adaptor for water bottle 1pc.

(13) Swing arm locking pin for dental light 1 pc. (For the swing-

mounted light only)

(14) Miscellaneous Parts for installation as follow. • Socket screw M10 x 40mm (for mounting) 4 pcs. • Socket screw M10 x 20mm (for level) 4 pcs. • Pan head Phillips screw M5 x 16mm 4 pcs. • Flat head screw M10×35mm 1 pc. • 1/4 sleeve 2 pcs. • 1/8 sleeve 1 pc.

Box No.5 (Vac Pac)

(15) Vac Pac Assembly 1 set

(16) Syringe parts 1 set

(17) Miscellaneous Parts for installation as follow. • Socket screw M10 x 40mm (for mounting) 4 pcs. • 1/4 sleeve 2 pcs. • Barb fitting for 1/4 OD tubing 2 pcs.

Box No.6 (PMU)

(18) PMU Assembly 1 set

(19) Junction Box Assembly 1 set

(20) Syringe parts 1 set

(21) Cuspidor Bowl Section

• Cup filler nozzle 1pc.

• Bowl Rinse nozzle 1pc.

• Basket Strainer & Cap 1pc.

(22) Miscellaneous Parts for installation as follow. • Socket screw M10 x 40mm (for mounting) 4 pcs. • 1/4 sleeve 8 pcs. • Barb fitting for 1/4 OD tubing 5 pcs.

(2)

(3), (4), (5) (6)

(16)

(20) (21)

(10) , (11)

(12) , (13)

(19)

- 6 -

Page 11: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

5. Swing Arm 5-1. How to Install the Swing Arm The swing arm assembly shall be mounted at the front of the chair under the seat. 1. Raise the seat frame of the chair as shown.

2. Insert M10 x 35mm flat head screw in the threaded hole on the top side of the plate, center . Leave 1" from the top of screw to plate surface.

3. Raise the seat.

4. Attach swing arm assembly to chair by aligning M10 x 35mm screw with recess in front of the seat and slide screw into slot.

5. Locate 4 - M10 x 20mm leveling set screws and insert into bottom side of swing arm plate.

6. Adjust the level of the swing arm plate with the four socket set screws.

7. Tighten the 4 - M10 x 40 mm socket head screws. Lift the swing arm if needed.

6. Junction Box 6-1. How to Install the Junction Box Refer to Floor Template for connection of water, air, power, drain hose and vacuum hose.

1. Raise the chair up

2. Turn off the power of the chair

3. Remove two slotted M5 x 10mm screws from each side of the pump cover and remove pump cover.

4. Place the junction box frame in front of the chair. (The plan of the plumbing position needs to be done in advance. See the flow diagram)

5. Attach the junction box frame to the floor with three M5 x 25mm round head philips wood screws.

Failure to lift seat frame as shown can result in damage to paint finish of swing arm in subsequent steps.

CAUTION Seat Frame

Flat Head Screw

Flange cover

Swing arm plate

4 X Socket head screws (M10 x 40mm) for securing arm assembly to the chair

4 X Socket head screws(M10 x 20mm) for level adjustment

Main SwitchScrews for J.box cover Screws for pump cover

Screws x 3

- 7 -

Page 12: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

7. Balance Arm for the Doctor’s Control 7-1. How to Install the Balance Arm

1. Lower the chair.

2. Insert the balance arm into the top of the swing arm.

3. Slide the balance arm cover off balance arm and route tubing and

cables down into the swing arm. Make sure that the hoses are not

twisted.

4. Remove the seat flange cover from chair by removing four M5 x

15mm screws and sub-link cover by

removing two M5 x 15mm screws on the back of the chair

cantilever lift arm.

5. Route umbilical tubing & wires from swing arm into chair. Tubing &

wires should be routed over the top of the pivot bar in the front of

the cantilever arm.

6. Feed umbilical down between 2nd pivot bar and into plastic retaining

loop.

7. Feed tubing between PCB box and pump and out to the front of the chair.

8. Slide in the balance arm cover and into the channels along the balance arm and the

balance arm end cap with two M4 x 10mm screws.

9. Adjust M6 x 5mm balance arm friction

adjustment screw (M6 x 5mm).

Balance arm cover

Flange cover Sub link cover

Main Link Axis

Sub Link Axis

Screws for flange cover

Route tubing and wires over the top of the pivot bar in front of the chair

Feed tubing between PCB box and Pump

Pump

PCB box

Tubing is routed throughthe plastic retainer loop and out thoroug the front of the chair

Balance arm friction adjustment screw

End cap

Screws

- 8 -

Page 13: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

7-2. How to Plumb the Doctor’s Control

1. Cut the duct hose to desired length and insert it between retainers inside of the pump cover and the junction box.

CAUTION

If Vac Pac or PMU is installed with Dr. table, run tubings for Vac Pac or PMU at the same time.

AIR

WATER(Optional)

Utility Package

Duct Hose

1/4 Sleeve

1/4 Sleeve

1/8 Sleeve

1/8 Sleeve

Earth(Yellow/Green)

14V/24V Wire(Red/Black/Blue/Purple)

1/8 Tubing (Orange)

1/8 Tubing (Brown)

1/8 Tubing (Blue)

Groundings

3D view of air utility part

Master Air (Orange 1/8”)Pirot Air (Brown 1/8”)

1/4 Tubing (Yellow)

2. Attach the manual on/off valve to the air supply pipe. Purge air through valve to clean debris. Attach automatic valve assembly for air to manual valve. Do the same thingfor the water assembly ordered with "city water" option.

If PMU is not purchased, Utility package for water becomes an option.The pressure gauge for air has higher scale range.

3. As below drawing shows, connect tubings (2,7,8 & 10), connect 14V/24V wire and connect the wire for protective earth (ground).

- 9 -

Page 14: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

4. Connect wires in J Box as follows.

(1) Dr.Table and (3) Dr.Table only

(4) PMU only

(2) Dr.Table and Vac Pac

Wiring in J Box

PMU Line

9P

2P

2P

3P

Char Line

Dr.Table Line

Vac Pac Line 9P

2P3P

Chair LineDr.Table Line

Do not connect

Do not connect

Do not connect

9P

2P3P

Chair Line

Dr.Table Line

PMU LineDo not connect

Do not connect

9P

2P2P

2P

3PChair Line

7-3. How to Plumb the Water Bottle

1. Connect the tubing labeled 3, 7 and 11 to the manifold inside the junction box.

Utility package for water is an option. Make sure that City/Bottle

switch is equipped on the Swing Arm.

If City/Bottle switch is not equipped, do not connect green 1/4 tube (11). This will cause water leak.

CAUTION

WATER

1/8 Sleeve1/4 Sleeve

1/4 Sleeve

1/8 Tubing (Brown)1/4 Tubing (Blue)

1/4 Tubing (Green)(Option)

City/Bottle switch

Swing arm

- 10 -

Page 15: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

7-4. How to install the Foot Control

Number Color (Size) Function

6 Red (1/4” OD) Drive

10 Yellow (1/4” OD) In

5 Yellow (1/8” OD) Chip (Air for spray, not used)

9 Gray (1/8” OD) Water (Pilot air for water for spray, not used)

1. Insert foot control tubing through channel in chair base

and feed through chairs into J.box.

2. Connect the tubing labeled 5, 6, 9 and 10 to the tubing

from the doctor table.

Note : Although the 1/8” OD gray tubing and yellow 1/8”

OD tubing are not used, connect to prevent air leaks.

3. Attach the foot control tubing to the chair base with a

nylon clamp (see photo).

Drive

Water Pilot Chip

IN

Foot control tbings

1/4 Sleeve

1/4 Tubing (Red)1/8 Tubing (Yellow)1/8 Tubing (Gray)

1/4 Tubing (Yellow)

INDRIVEChipWater Pilot

1/8 Sleeve

8. Monitor Mount(Optional) 8-1. How to Install the Monitor Arm

1. Install the monitor arm to the swing arm. See 5-1 How to install swing arm.

2. Feed the cables thought the swing arm and the chair, and bring the end into the junction box.

3. Install the monitor mount to the monitor arm.

4. Connect the cables inside Junction box.

Note : LCD monitors are not supplied from Belmont. The mounting holes on back of monitor should be

designed for attachment of 75 or 1000 mm VESA mgt pattern.

- 11 -

Page 16: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

9. Light Pole (Optional) 9-1. How to Install the Light Pole

1. For installation of the dental light, see the installation and

operation manual of Quolis dental light.

2. Position the swing arm to either side of the chair and lock it in

place with the dental light locking pin.

When you combine Dr. Table and PMU with a water bottle.

Set Lock Pin on A position to avoid the interference of a bottle

and a swing arm.

Locking Pin

< A Position >

CAUTION

Do not remove the cardboard paper put on the swing arm untilthe seat cushion has been attached to the seat frame with screws,otherwise the surface of the light pole will be scratched off whenswinging the arm.

10. Vac Pac(Optional) 10-1. How to Install the Vac Pac

1. Remove four screws M5 x 15mm and the flange cover of the chair.

2. Insert two M10 x 40mm socket head cap screws into the top

threaded hole on the rear side of the chair. Screws should

extend approximately 1" from chair.

3. Hook the assistant arm mounting bracket underneath the two

socket head cap screws.Tilt the bracket slightly and lift it.

Place the lower edge of the mounting bracket on the

projections of the flange.

4. Tighten lower M10 x 40mm socket head cap screws through

swing arm mounting bracket. Tighten all four screws.

5. Route the umbilical tubing with unit umbilical down through

chair, between the main link and sub link Axis, and bring

the end to the junction box. Tie the tubing with the white

nylon clamp.

Make sure that hoses are not kinked

Socket head cap screw

Projections of

approx. 1”

Screw for flange cover

the flange

Tighten

Projection of the flangeMain Link Axis Sub Link Axis

Nylon Clamp

- 12 -

Page 17: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

10-2. How to Plumb the JBOX-Utility (Vac Pac)

1. Cut the duct hose to desired length and insert it

between retainers inside of the pump cover and

the junction box.

2. Attach the manual on/off valve to the air supply

pipe.Purge air through valve to clean debris.

Attach automatic valve assembly for air to

manual valve. Do the same thing for the water

assembly ordered with "city water" option.

3. Connect 5/8 Vacuum hose.

4. As below drawing shows, connect

tubings and wires. If connection end is

plugged, use supplied 1/4 fitting.

5. Connect wires under a pump cover.

AIR

WATER(Optional)

Utility Package

Duct Hose

Vacuum

1/4 Sleeve

1/4 Tube Blue

1/4 Sleeve Yellow

Vac Pac Line Chair Line

- 13 -

Page 18: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

11. PMU(Optional) 11-1. How to Install the PMU

1. Remove four screws M5 x 15mm and the flange cover of the

chair.

2. Insert two M10 x 40mm socket head cap screws into the top

threaded hole on the rear side of the chair. Screws should

extend approximately 1" from chair.

3. Hook the assistant arm mounting bracket underneath the two

socket head cap screws.

Tilt the bracket slightly and lift it. Place the lower edge of the

mounting bracket on the projections of the flange.

4. Tighten lower M10 x 40mm socket head cap screws through

swing arm mounting bracket. Tighten all four screws.

5. Route the umbilical tubing with unit umbilical down through

chair, between the main link and sub link, and bring the end

to the junction box. Tie the tubing with the white nylon

clamp.

Make sure that hoses are not kinked

Socket head cap screw

Projections of

approx. 1”

Tighten

Main Link Axis Sub Link Axis

Nylon Clamp

Screw for flange cover

the flange

Projection of the flange

- 14 -

Page 19: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

11-2. How to Install the Cuspidor Bowl.

1. Take the main cover off by unscrewing

three screws.

2. Remove a cap bolt.

3. Insert Cuspidor Bowl into Rotation Axis.

Do not forget to insert an attached washer.

4. As below drawing shows, connect drain hose, tubings and wires.

(1) We supply long hoses. As needed, cut them accordingly.

(2) To avoid water leak, insert barb fittings to the end.

(Use a connector, when an optional Cuspidor Drain Kit in installed)

(3) Connect the drain hoses by a hose clamp.

(4) Connect tubings by matching colors and numbers.

(5) Connect wirings by using a 4P wire connector.

5. When connections have been completed, put the main cover and cap bolt back.

CAUTION

Please avoid the external force on Drain Hose, tubings and Wires.

This will cause the malfunction or damage to the equipment.

Cuspidor Bowl

Washer

Main Cover

Body

Cap Bolt

hose clamp

Connect tubings

Connect

Cuspidor Drain Valve (Option)

Connect Wirings

1/4 Sleeve

Wire4P Connector

1/4 Tube Blue1/4 Tube Green

- 15 -

Page 20: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

11-3. How to Plimb the JBOX-Utility

1. Cut the duct hose to desired length and insert it between retainers inside of the pump cover and the junction box.

2. Attach the manual on/off valve to the air supply

pipe.Purge air through valve to clean debris. Attach automatic valve assembly for air to manual valve. Do the same thing for the water assembly ordered with "city water" option.

3. Connect Vacuum Hose (5/8) and Drain Hose (3/4) Vacuum hose have uneven surface. We supply long hoses. As needed, cut them accordingly.

4. As below drawing shows, connect tubings and

wires. If connection end is plugged, use

supplied 1/4 fitting.Connect Dr. Table side

wirings to the wirings on PMU side.

5. Connect wires in J Box as follows.

AIR

WATER(Optional)

Utility Package

Duct Hose

Drain

Vacuum

1/4 Sleeve

1/8 Sleeve

1/4 Sleeve

1/8 Sleeve

Earth(Yellow/Green)

14V/24V Wire(Red/Black/Blue/Purple)

1/4 Tube Green

1/8 Tube Brown1/8 Tube Orange

1/4Tube Blue

1/8 Tube Brown

1/4 Tube Yellow

Groundings3D view of air utility part

Pirot Air(Brown 1/8”)

Option for PMU onlyDo not Connect

Connect to Dr.Table Line

(1) Dr.Table and PMU

(3) Dr.Table only

(4) PMU only

(2) Dr.Table and Vac Pac

Wiring in J Box

PMU Line

9P

2P

2P

3P

Chair Line

Dr.Table Line

Vac Pac Line 9P

2P3P

Chair LineDr.Table Line

Do not connect

Do not connect

Do not connect

9P

2P3P

Chair Line

Dr.Table Line

PMU LineDo not connect

Do not connect

9P

2P2P

2P

3PChair Line

- 16 -

Page 21: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

11-4. Installation of Accessories.

1. Install, Nozzles, Cap & Basket Strainer.Cap

Basket Strainer

Nozzles

12. Final Adjustments 12-1. Swing Arm

1. Attach the lower tubing cover to the bottom of the swing arm mounting bracket with four M5 x 16mm.

2. Replace the chair pump cover and the junction box cover and fasten with M5 x 10mm screw and

tighten firmly with a slotted screw driver.

CAUTION

Fasten the pump cover and junction box cover firmly with a screw driver with screws provided and tighten

12-2. How to Adjust the Swing Arm Friction

The friction of the swing arm for Doctor's control,

dental light and monitor must be adjusted independently.

1. Adjust the M6 x 5mm socket set screw beneath the

swing arm.

12-3. How to Adjust the Rotation Tension of Assistant Arm.

1. Adjusting screw for vertical angle is located here.

To access adjusting screw, pull the rubber cover off.

12-4. How to adjust the water flow volume of Cupfiller and

Bowl Rinse.

1. Adjust the water flow by pinch valves.

Set screw

Set screw

Green : Bowl Rinse

Pinch Valves

Blue : Cupfiller

- 17 -

Page 22: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

13. Connect the Power Cable and Confirms functionality. 1. Connect the power cables(120V), electrical cables

and the connectors to the electrical receptacle inside

the junction box for chair and unit.

2. Adjust the regulator pressures. Air : 75 psi. Water (optional) : 30-40psi

3. When all of the installation work are done, turn the main switch on.

Confirms that each part functions correctly.

Then put the cover on.

14. Option 14-1. How to Install the Tray Holder

The tray holder can be mounted at either side of the Doctor's control.

1. Fit the tray holder to the tray arm.

Install the tray arm to beneath the Doctor's control with two

screws M6 x 25mm.

Ground reliability can only be achieved when the

equipment is connected to an equivalent receptacle

marked HOSPITAL only or HOSPITAL GRADE.

CAUTION

ONOFF

Two holes on the back

This can be installed on the other side.

- 18 -

Page 23: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

15. Flow Diagram 15-1. Doctor’s control section

- 19 -

Page 24: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

15-2. Swing Arm Section

15-3. Vac Pac Section

Air regulatorAir regulator

City / bottle

Option Bottle water City Water (Option)

City / bottle

- 20 -

Page 25: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

15-4. PMU Section

Syringe

Ejector

Vacuum

FLOW CONTROL FOR SYRINGE

Air

Water Outlet Flow Control

WATER

AIR

Pinch Valve

Cuspider Bowl

Drain Hose 3/4

Pinch Valve

Vacuum Hose 5/8

Orange 1/8 : Master Air IN(Option)Brown 1/8 : Master Air OUT(Option)Green 1/4 : City Water

Yellow 1/4 : AirBlue 1/4 : Bottled Water

ManifoldSolenoid Valve for Bowl Frash

Solenoid Valve for Water Water

Solid Collector

Bowl Flash Nozzle

Cupfiller Nozzle

Air

Water

- 21 -

Page 26: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

15-5. Junction Box Section

AIR

(Drain For PMU only)

VACUUM

WATER (Optional)

Yellow 1/4”OD

Green 1/4”OD

Blue 1/4”OD

Orange 1/8”OD

Blue 1/8”OD

Brown 1/8”OD

- 22 -

Page 27: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

15-6. Overall Unit (Vac Pac System)

AIR

(Dra

info

r P

MU

on

ly)

VA

CU

UM

WA

TE

R

(Op

tio

na

l)

- 23 -

Page 28: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

AIR

(Dra

info

r P

MU

on

ly)

VA

CU

UM

WA

TE

R

15-7. Overall Unit (PMU System)

- 24 -

Page 29: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

16. Electric Diagram

16-1. Chair Section

12

34

56

78

9

84736251

21

21

SV1

SV2

SV4

SV3

1 2 3 4 5 6

12

SV6

98

76

54

32

1

1 2W

hite

Yello

w

XA9P

ANFU

07A0

11

Trun

k Line

B fo

r Sole

noid

Valve

VL2P

Lock

Red

Brow

n

1051 6 2 7 3 8 4 9

CN9

5566

-10A

(MO

LEX)

43

12

VL2P

Fuse

F 10

ABlac

k

Gre

en/Y

ellow

Black

Whit

e

Whit

eBl

ack

LN

CN18

IL-5

P-S3

EN2

(JAE

)

54

32

1

FLAS

H

CN17

IL-7

P-S3

EN2

(JAE

)

76

54

32

1

RY-2

005

(1.2

)

(3.4

)

(2.5

)

(3.4

)

3

11V9V

Trun

k Line

for C

apac

itor

ANFU

12A0

Brow

n

Yello

w

Ora

nge

Red

Gra

y

Purp

le

Blue

Gre

en

CN1-

2(1-

1)IL

-10P

-S3E

N2(J

AE)

109

87

65

344

221

CN13

IL-3

P-S3

EN2

(JAE

)

CN14

IL-3

P-S3

EN2

(JAE

)

CN7

IL-3

P-S3

EN2

(JAE

)

CN6

IL-3

P-S3

EN2

(JAE

)

CN5

IL-3

P-S3

EN2

(JAE

)

CN12

IL-3

P-S3

EN2

(JAE

)321

CN15

IL-2

P-S3

EN2

(JAE

)211

CN2-

2(2-

1)IL

-6P-

S3EN

2(J

AE)

CN1-

1(1-

2)IL

-10P

-S3E

N2(J

AE)

CN16

B03B

-XAS

K-1

CN9

B09B

-XAS

K-1

CN10

IL-1

0P-S

3EN2

(JAE

)

1 2 3 4 5 6

21

1 2 31 2 3

VL8P

56

109

87

65

344

221

1 2 3 1 2 3

Pote

ntio

met

er fo

r Bac

kres

tBr

own

Red

Ora

nge

109

87

65

43

21

Cont

rol P

.C.B

.(AG

FS13

*0)

XX-S

TD20

07/0

0TI

Relay

P.C

.B.

120V

Yello

w

Blac

k do

t ind

icate

s m

ale

pin

side

[Rem

arks

]Co

nnec

tor S

ymbo

l

Cap

Plug

21

Mot

orUp Do

wn

Raise

Recli

ning

Pote

ntio

met

er fo

r Sea

tBr

own

Red

Ora

nge

TSM

-CBC

-5LP

-M1-

E

Tran

sform

erTK

-120

V-2H

Red

Whi

te/R

ed

Brown

SW2

SW3

LIM

NOL

Lim

itST

ORE

EWS

VDA

S20

E53

M

321

1 2 33211 2 3

Capa

citor

45uF

1 5 2 6 3 7 4 8

4

231 6 4 5

2 3 1 2

32

1

54

32

1

CN6

5566

-08A

(MO

LEX)

CN4

B02P

-NV

CN3

B3P-

VH

CN1

B2P-

VH

CN2

B10B

-XAS

K-1N

-ACN

8B0

3B-X

ASK-

1N-A

CN7

B04B

-XAS

K-1N

-A

CN10

5566

-06A

(MO

LEX)

CN5

B04P

-NV

109

87

65

344

221 Brown

544

221

Yello

w

Yello

wAN

FT58

A0Tr

unk L

Ine

for M

otor

/ Ca

pacit

orVL

6P

ANFT

57A0

Tran

sform

er S

ectio

n(12

0V)

ANET

56A0

Mot

or S

ectio

n (1

15V)

ANFT

56A0

ANDR

50A0

ANDR

36A0

Trun

k Line

for R

elay

Sole

noid

Val

ve S

ectio

n(1

20V)

1 2 3 4

Chai

r Dow

n Pr

ohib

ition

Chai

r Ope

ratio

n Pr

ohib

ition

CN4-

1(4-

2)IL

-4P-

S3EN

2(J

AE)

CN8

B04B

-XAS

K-1

12

34

CN3

B05B

-XAS

K-1

SW4

MO

DESW

131

(Pro

porti

onal

Sol

enoi

d Va

lve)

67

89

12

12

34

5

1

Whit

e

Red

10

NMF2

P

N.O.

COM

1 3 2

CN11

5556

-04A

Mai

n Sw

itch

L2

N4

24

Up Down

Raise

Recli

ning

CN2-

1(2-

2)IL

-6P-

S3EN

2(J

AE)

1 2 3 4 5 6

CN19

B05B

-XAS

K-1

3

1 2 3 4

DNCS

42A0

CN4-

1(4-

2)IL

-4P-

S3EN

2(J

AE)

N.O.

COM

Unus

ed

12

CN9

IL-2

P-S3

EN2

(JAE

)

109

87

65

344

221

1112

13

Arm

rest

Mem

bran

e SW

- L(A

GFS

14A0

)

109

87

65

344

221

1112

13

109

87

65

34 4

221

1112

13

109

87

65

34 4

221

1112

1310

98

76

53

4422

111

1213

Dist

ribut

ion

P.C.

B.(A

GFS

12A0

)

(Dist

ribut

ion

P.C.

B. in

the

right

sid

e ar

mre

st)

XA9P

XA2P

Touc

hpad

Swi

tch(A

GFS

16A0

)

To S

eria

l Com

mun

icatio

n Li

ne fo

rDo

ctor

Tabl

e

D Li

ght P

CB.

4

Rota

tion

Lock

EWS

VDA

S20

E53

DNCS

42A0

ANFT

71A0

Mai

n Po

wer S

witch

Sec

tion

ANFT

64A0

Trun

k Li

ne fo

r Den

tal L

ight

Pow

er S

ourc

e

ANFT

65A0

Trun

k Line

ffor

Cha

ir Po

wer S

ourc

e

ANFT

63A0

Earth

ANFT

55B0

XA13

P

XA13

PXA

13P

XA13

PXA

13P

ANFT

60A0

ANFT

61A0

Trun

k Line

B fo

r Mem

bran

e SW

ANFT

62A0

Trun

k Line

C fo

r Mem

bran

e SW

ANFT

67A0

Seria

l Com

mun

icatio

n Li

ne

ANFT

68A0

D Li

ght C

omm

unica

tion

Line

ANFT

69A0

Touc

hpad

SW

Trun

k Line

A fo

r

Safe

ty Sw

itch

for U

nder

Pum

p Co

ver

(AG

FS15

A0)

3

122

34 4

56

78

910

XA10

P

ANFT

76A0

Assis

tant

Sid

e

Green

22

44

ANFT

70A0

Powe

r Cab

le Se

ctionSa

fety

Switc

h fo

r Und

er P

ump

Cove

r

Brow

nRe

d

Brow

nRe

d

Chai

r Dow

n Pr

ohib

ition

Chai

r Ope

ratio

n Pr

ohib

ition

MotorUpDown

RecliningRaise

Up/Down Slow SlideBackrest Slow SlideLock Cancel

RedOrangeYellowGreenBluePurpleGray

Black

RedOrangeYellowGreenBluePurpleGray

Pink

WhiteBlack

White/Red

Light Blue

AR

UpDown

RecliningRaise

P1P2LP

COM

5VLock CancelSafetySafety LED.COM

BrownRed

Seria

l Tra

nsm

issio

n

Seria

l Rec

eptio

n

Seria

l Tra

nsm

issio

n

Seria

l Rec

eptio

n

RedBrown

RedBrown

OrangeYellowGreenBluePurpleGray

Black

Touc

hpad

SW

Trun

k Line

B fo

r

AR

UpDown

RecliningRaise

P1P2LP

COM

RedBrown

OrangeYellowGreenBluePurpleGrayBlack

RedBrown

OrangeYellowGreenBluePurpleGray

Black

AR

UpDown

RecliningRaise

P1P2LP

COM

Touc

hpad

SW

Trun

k Line

C fo

r

BrownRedOrangeYellowGreenBluePurpleGrayWhiteBlackPinkLight BlueWhite/Red

BrownRedOrangeYellowGreenBluePurpleGrayWhiteBlackPinkLight BlueWhite/Red

BrownRedOrangeYellowGreenBluePurpleGrayWhiteBlackPinkLight BlueWhite/Red

AR

UpDown

RecliningRaise

P1P2LP

5V

Lock CancelSafety

Safety LED.

COM

AR

UpDown

RecliningRaise

P1P2LP

5V

Lock CancelSafety

Safety LED.

COM

Arm

rest

Mem

bran

e SW

- R

Yello

wYe

llow

Yello

w

Yello

wW

hite

Red

Yello

wYe

llow

Gra

y

Gra

y

Trun

k Line

A fo

r Mem

bran

e SW

BlueBlue

Light Blue

Blue

WhiteWhite

BlackBlack

White

Blac

kBl

ack

Whit

e

Whit

eYe

llow

Whit

eYe

llow

Whit

eYe

llow

Trun

k Line

A fo

r Sole

noid

Valve

MotorUp

Down

RecliningRaise

COM Com

Up/Down Slow Slide

Backrest Slow Slide

COM

COM

Lock Cancel

SV8

SV7

3 2

VL4P4 1

Yello

wW

hite

Yello

wW

hite

Gre

enBl

ue

Yello

wO

rang

e

Up/D

own

Slow

Slid

e

Back

rest

Slow

Slid

e

Back

rest

Slow

Up/D

own

Slow

1 2

XA2P

Ther

mist

or

NTC

Ther

mist

or

Red

Brow

nBl

ack

Blac

kTh

erm

istor

for S

olen

oid

Valve

sCh

air u

pper

fram

e

Earth

Lin

e Base

Whit

e

- 25 -

Page 30: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

16-2. Dr.Table Section

12

1 2 3 4

1 2

HP

Hol

der S

W P

CB. F

ourt

h Li

ne

4 3 2 1

21G

ray

Purp

le

14142424

(2.3

)

(1)

CN36

B03B

-XA

SK-1

N-A

321

Tim

erSp

ray

F(Fu

nctio

n)ST

ORE

LEDM

AIN

(Gre

en)

LOCK

(Ora

nge)

Ligh

t Pac

k(G

reen

)

Spra

y A

ir(Bl

ue)

Operation Switch

-+SE

LECT

Ligh

t Pac

kD

enta

l Lig

ht

Chair Operation Switch

LPP2P1AR

Rais

eD

own

UP

Trun

k Li

ne fo

r Dis

play

AN

FT93

XX

NO

COM

Min

i Air

Switc

h(F/

C)

AN

FT97

XX

AN

FT95

XX

AN

FT96

XX

AN

FT83

XX

AN

FT80

XX

AN

FT79

XX

AN

FT87

XX

AN

FT86

A0

AN

FT82

XX

AN

FT85

XX

Tran

sfor

mer

Sec

tion

AN

FT92

XX

Trun

k Li

ne fo

r Den

tal L

ight

Pow

er S

ourc

e (J

BOX)

AN

FT81

XX

21

32

13

21

3

Pow

er C

able

(UL)

JBO

X

AN

FT84

XX

Seria

l Com

mun

icat

ion

Line

AN

FT78

XX

Trun

k Li

ne fo

r HP

Hol

der

AN

FT77

XX

AGFS

17XX

AGFS

20XX

Air

SW

Fuse

F 10

A

120V

14V 24V

Tran

sfor

mer

FK-

023

Junc

tion

Box

Sect

ion

AC IN

1 2211 221

21

1 2 3

12

12

34

21

21

34

56

71

2

87654321 4321

8 7 6 5 4 3 2 1 87654321

1 2 3 4 5 6 7 8 1 2 3 4 1 21 221 1 221

1 2 3 4

43211 2

CN31

IL-2

P-S3

EN2

(JA

E)

21

CN3

IL-3

P-S3

EN2

(JA

E)

To S

eria

l Com

mun

icat

ion

Line

(ASS

)Se

rial T

rans

mis

sion

(1.2

)

Seria

l Rec

eptio

n(3

.4)

CN29

236-

747(

Gre

en)

(WAG

O)

CN26

IL-3

P-S3

EN2

(JA

E)PE

RIO

END

O

CN16

IL-4

P-S3

EN2

(JA

E)

Wat

er

Doc

tor M

embr

ane

SW(A

GFS

18XX

)

12

34

56

78

910

12

344

665

Uni

t P.C

.B.

(AG

FS22

XX)

CN21

IL-4

P-S3

EN2

(JA

E)

POW

ER

CN2

IL-2

P-S3

EN2

(JA

E)CN4

IL-2

P-S3

EN2

(JA

E)

CN6

IL-2

P-S3

EN2

(JA

E)

CN7

IL-3

P-S3

EN2

(JA

E)

CN10

IL-2

P-S3

EN2

(JA

E)

CN12

IL-2

P-S3

EN2

(JA

E)

CN17

IL-2

P-S3

EN2

(JA

E)

Air

SW

CN24

IL-4

P-S3

EN2

(JA

E)

CN25

IL-4

P-S3

EN2

(JA

E)

LP1

LP2

LP3

LP4

CN27

236

-401

(Gra

y)W

AGO

12

34

56

78

910

12

344

665

CN14

IL-2

P-S3

EN2

(JA

E)

Air

Red

Blac

kRe

dBl

ack

CN30

IL-2

P-S3

EN2

(JA

E)

Dis

play

(AG

FS19

XX)

12

34

56

78

910

CN5

IL-1

0P-S

3EN

2(J

AE)

CN18

IL-2

P-S3

CN2

(JA

E)W

hite

Blac

k

HP1

SV

HP2

SV

HP3

SV

HP4

SV

CN8

IL-8

P-S3

EN2

(JA

E)

HP1HP2

HP4

12V

0V0VHP3

CN13

IL-8

P-S3

EN2

(JA

E)

CN1

IL-3

P-S3

EN2

(JA

E)

CN36

236

-401

(Gra

y)W

AGO

CN37

236

-401

(Gra

y)W

AGO

CN38

236

-401

(Gra

y)W

AGO

CN23

IL-4

P-S3

EN2

(JA

E)

CN11

IL-8

P-S3

EN2

(JA

E)

FLA

SH

CN28

IL-7

P-S3

EN2

(JA

E)

CN19

IL-2

P-S3

EN2

(JA

E)

CN9

IL-4

P-S3

EN2

(JA

E)

GN

DCN

22IL

-2P-

S3EN

2(J

AE)

HP4

HP3

HP2

HP1

CN32

IL-2

P-S3

EN2

(JA

E)CN

33IL

-2P-

S3EN

2(J

AE)

CN34

IL-2

P-S3

EN2

(JA

E)CN

35IL

-2P-

S3EN

2(J

AE)

CN20

IL-1

0P-S

3EN

2(J

AE)

CN15

IL-6

P-S3

EN2

(JA

E)

XA-1

3PXA

-6P

AN

FT98

XX

Trun

k Li

ne A

for D

r Mem

bran

e SW

AN

FT99

XX

HP3HP2HP1

HP4

---

+++

-+

8

Sole

noid

Val

ve fo

r Pilo

t Air

Sole

noid

Val

ve fo

r Pilo

tA

ND

T08X

X

-+

-+

25

Sole

noid

Val

ve fo

r Spr

ayA

ND

T07X

X

1 2

12

1 2 3 4

12

12

1 21 21 2H

P H

olde

r SW

PCB

.

L N

Blac

k

VL4P

VL2P

Pow

er S

ourc

e fo

r Den

tal L

ight

AGD

R39X

X

Brow

n

Red

XA2PXA

2P

Gre

enBl

ue

Ora

nge

Yello

w

Brow

nRe

d

Stan

dard

Spe

c.

Syrin

ge T

urbi

ne x

3

Brow

n

AN

FU04

XX

Mai

n A

ir SW

Sec

tion

AGFL

34XX

AN

FU05

XX

Eart

h (D

r Tab

le)

CN1

PS-L

A-1

4P

CN1

12

1314

Brow

nRe

dO

rang

eG

reen

Yello

w

Blue

Gra

yPu

rple

Whi

te

Dynamic Output 1Dynamic Output 2Dynamic Output 3Dynamic LED 1Dynamic LED 2Dynamic LED 3

Red

BlackWhiteGray

BrownRed

OrangeYellowGreen

Blue

PurpleBlue

GreenYellow

Orange

Brown

Trun

k lin

e B

for 2

4V.1

2V

Powe

r Sou

rce

Purple

Blue

PurpleBlue

Purple

Blue

Black

Red

Mai

n G

auge

Cla

mp

Red

Blue

Red

Blac

k

Purp

leBl

ue

Trun

k lin

e A

for 2

4V.1

2V

Powe

r Sou

rce

Fuse

6A

Fuse

2A

Trun

k lin

e (B

lue)

for F

use

Hold

er

Trun

k lin

e (R

ed) f

or F

use

Hold

er

Gre

en/Y

ello

w

Trun

k lin

e (B

lack

) for

Fus

e Ho

lder

Chai

r Pow

er S

uppl

y Se

ctio

n

Chai

r Pow

er S

uppl

y Bl

ack

Blac

k

Blac

k

Whi

te

Whi

te

Whi

teG

reen

Blac

k

Blac

k

Gra

y

Whi

teW

hite

Whi

te

Gre

en/Y

ello

w

Earth

Purp

le

Purp

leO

rang

e

Ora

nge

Red Blac

kBl

uePu

rple

J-Bo

x G

auge

Cla

mp

Sect

ion

Both

con

nect

ors

to b

e co

nnec

ted,

2P

& 3

P)

Red

Blac

kRe

dBl

ack

Red

Blac

kRe

dBl

ack

HP

Hol

der S

W P

CB.

HP

Hol

der S

W P

CB.

Seria

l Tra

nsm

issi

on

Seria

l Rec

eptio

n

Brow

nBr

own

Brow

nBr

own

Brow

n

BrownRed

OrangeYellowGreen

Blue

WhiteGray

Purple

0V5VVFD-RS

Low

erSp

ray

Wat

er(B

lue)

Trun

k Li

ne B

for D

r Mem

bran

e SW

VFD-R/WVFD-E

VFD-1

VFD-3VFD-2

VFD-0

AN

FU69

XXTr

unk

line

(Bla

ck) f

or F

use

Hold

er

HP2

HP1

HP3

Input AInput BInput CInput DInput EInput FInput GInput H

Display LED.Display LED.

XA3P

1 32

Ora

nge

Yello

w

To S

eria

l Com

mun

icat

ion

Line

(CH

)Co

nnec

t 2P

conn

ecto

r onl

y

With

PM

U

With

out P

MU

12

34

PurpleBlueBlackRed

To te

rmin

al o

n as

sist

ant s

ide

1A03

XDXX

Trun

k lin

e C

for 2

4V.1

2V

Powe

r Sou

rce

Conn

ect t

runk

line

C fo

r 24V

.12V

fr

om c

uspi

dor

With

PM

U

With

out P

MU

Conn

ect t

runk

line

A a

nd B

Trun

k lin

e(Pu

rple

) for

Dr P

CB.

Powe

r Sou

rce

Trun

k lin

e(Bl

ue) f

or D

r PCB

. Po

wer S

ourc

e

Rota

tion

Dire

ctio

nFo

rwor

d Ro

tatio

n(G

reen

)

Reve

rse

Rota

tion(

Ora

nge)

B1A

1A

2B3

A3

B6A

6B7

A7

A4

A5

B2B4

B5

- 26 -

Page 31: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

16-3. PMU Section

21

21

2121

1 21 2211 2 211 2

321

4321

1 2 3654321

1 2 1 21 2 3

1 2 3 44321

12

21

1 2 3 4 5 6

43

21

1 2

1G01

6Q**

Gau

ge C

lam

p (C

uspi

dor)

Chai

r Con

trol P

CB.

(AG

FS13

**)

12

34

56

78

998

76

54

32

1

XA9P

1A03

YA**

1

CN1-

2(1-

1)IL

-10P

-S3E

N2(J

AE)

109

87

65

34

21

XA9P

(AG

FS16

**)

ANFT

69**

12

34

56

78

9

XA9P

1A03

Y9**

Touc

hpad

Swi

tch

1G01

64**

Earth

(3m

)

1A03

Y8**

1A03

Y0**

1A03

XX**

1A03

XW**

1A03

XV**1G

0163

**

CB S

witc

h Se

ctio

n1G

0163

**N.

O.

N.O

.CO

MCO

M

(Top

View

)

(Sid

e Vi

ew)

XA C

onne

ctor

43

21

IL C

onne

ctor

CN16

B02B

-XAS

K-1

Cent

ral V

acuu

m S

W2

1

(JAE

)IL

-2P-

S3EN

2CN

6-B

(Cha

ir O

pera

tion

Proh

ibitio

n)

LOCK

CN17

-BB0

2B-X

ASK-

1

CN17

-AB0

2B-X

ASK-

1

CN18

B05B

-XAS

K-1

54

32

1

XA2P

XA2P

XA6P

Norm

ally

#3 s

ide(

4800

bps)

(#1

side

2400

bps)

Tran

smiss

ion

Spee

d

CN14LED6

Red

(Ful

l cap

acity

of c

upfill

er)

LED5

Ora

nge(

CN3

- dat

a tra

nsm

issio

n)

LED4

Ora

nge(

CN4

- dat

a tra

nsm

issio

n)

LED2

Gre

en(C

upLE

D-De

tect

a c

ap)

LED1

Gre

en(P

ower

LED

)

(JAE

)IL

-2P-

S3EN

2CN

6-A

LOCK

CN10

IL-2

P-S3

EN2

(JAE

)CU

R-HO

LD

CN11

IL-2

P-S3

EN2

(JAE

)SI

G-C

UR

(JAE

)IL

-2P-

S3EN

2CN

12Ex

tra

CN4

(JAE

)IL

-2P-

S3EN

2

(JAE

)IL

-2P-

S3EN

2CN

3

TXD

(JAE

)IL

-4P-

S3EN

2CN

2

RXD

CN7

(JAE

)IL

-3P-

S3EN

2

(JAE

)IL

-3P-

S3EN

2CN

8

FULL

Sen

sor

CUP

Sens

or

(JAE

)IL

-6P-

S3EN

2CN

9

CUP

SW

CUP

SV

BOW

L SV

(WAG

O)

236-

747

CN12

1

Assis

tant

PCB

.(A

GDJ

73**

)

POW

ER

1A03

XZ**

1A03

Y1**

1A03

Y7**

Trun

k Li

ne fo

r 14V

Pow

er S

ouce

1A03

XY**

3

12

34

56

7

FLAS

H

CN19

B07B

-XAS

K-1

Notc

hed

for #

1 Pi

n sid

e

12

34

5 .

. .

12

34

CN20

Norm

ally

# 3

side

(#1

side

for w

ritin

g of

dat

a)

Data

inpu

tXA

3P

J Bo

x Se

ctio

n

12

34

56

78

IL-8

P-S3

EN2

CN5

(JAE

)

VL4P

VL4P

XA2P

XA2P

Cusp

idor

Sec

tion

Cupf

iller S

witc

hBo

wl R

inse

Swi

tch

XA4P

Gre

en/Y

ello

w

Blac

k do

t ind

icate

rs m

ale

pin

side

Cap

Plug

21

XA2P

1G01

7J**

1G01

7K**

XA2P

Opt

ion

32

1

WRI

TEO

UT1

23

2009

0717

BDS-

5000

-PMU

UpDown

ARP1P2LP

COM

RaiseReclining

BrownRed

GreemBluePurpleGrayBlack

OrangeYellow

BrownRed

GreemBluePurpleGrayBlack

OrangeYellow

BrownRed

GreemBluePurpleGrayBlack

OrangeYellow

Trun

k Li

ne A

for

Touc

hpad

SW

Trun

k Li

ne B

for

To

uchp

ad S

W (f

or P

MU)

Trun

k Li

ne C

for

Touc

hpad

SW

(for

PM

U)

Sole

noid

Val

ve fo

r Bow

l Rin

seVD

W21

-6G

-2-M

5

Sole

noid

Val

ve fo

r Cup

fille

rVD

W21

-6G

-2-M

5

Sole

noid

Val

ve fo

r Bow

l Rin

seV-

114-

6G-M

5-X7

0

Red

Blac

kR

edBl

ack

Brow

nR

edO

rang

eYe

llow

Red

Blac

k

Cup

fille

r Sol

enoi

d Va

lve

Sect

ion

Bow

l Rin

se S

olen

oid

Valv

e Se

ctio

n

Bowl

Rin

se S

olen

oid

Valve

Se

ctio

n

Red

Blac

k

Trun

k Li

ne fo

r Sol

enoi

d Va

lve

-+

-+

-+

[Rem

arks

]Co

nnec

tor S

ymbo

l X

A .

. .

XA C

onne

ctor

VL

. .

. VL

Con

nect

or

Brown

RedOrange

EXTCOM

ARLPP2P1

BOWLCUP

(Cha

ir O

pera

tion

Proh

ibitio

n)

TXDSCLKRXDBUSYRESETCNVSSGND

PurpleBlue

18V PowerSignal(Red)

Signal(Green)Signal(Blue)

Signal(Orange)

TXD

Safe

ty S

W fo

r Dr

.Tab

le A

rm

Safe

ty S

W fo

r Dr

.Tab

le A

rm

TXD

Trun

k Li

ne(A

ss) A

for

Seria

l Com

mun

icatio

n Li

ne

Trun

k Li

ne(A

ss) B

for

Seria

l Com

mun

icatio

n Li

neBr

own

Red

Ora

nge

Yello

wBr

own

Red

Brow

nR

edTo

Cha

ir Se

rial

Com

mun

icat

ion

Line

(CH

)

To D

r.Tab

le S

eria

l C

omm

unic

atio

n Li

ne(U

nit)

Trun

k Li

ne C

for C

B Sw

itch

Ora

nge

Blue

Brow

nPu

rple

Red

Ora

nge

Blue

Brow

nPu

rple

Red

Red Red

Blac

k

BrownOrange

RedRed

Trun

k Li

ne B

for C

B Sw

itch

Trun

k Li

ne A

for C

B Sw

itch

Trun

k Li

ne C

for 2

4V/1

4V P

ower

So

uce

BrownRed

BrownRed

CB S

witc

h Se

ctio

n

Red

Blac

kBl

uePu

rple

Conn

ect t

o Tr

unk

Line

B fo

r 24

V/14

V Po

wer S

ouce

Conn

ect t

o Tr

unk

Line

A fo

r 24

V/14

V Po

wer S

ouce

(Sid

e Vi

ew)

Pin

Num

ber A

rrang

emen

tLE

D3 O

rang

e(CN

2 - d

ata

trans

miss

ion)

5 .

. .

5 .

. .

- 27 -

Page 32: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

17. Location of the Product Labels

17-3. Product label (SERIAL No.) is

located side of the assistant

swing arm.

17-2. Product label (SERIAL No.) is

located side of the doctor swing arm.

17-4. Product label (SERIAL No.) is

located side of the PMU swing arm.

ET

L L

IST

ED

US

CONF

ORM

S TO

UL S

TD 73

CERT

IFIE

D TO

CAN/

CSA

STD

C22.

2NO

. 68

2007

Model QUOLIS5000 UNITSerial No.Rated VoltageRated CurrentClassification

ManufacturedProduct of Japan

ZZ535521

AW********* ****1V ~ 60Hz

1A

GS1-128

(**)**************(**)*********

2-1-1.Higashishinsaibashi,Chuo-ku,Osaka,Japan

101 BELMONT DRIVE, SOMERSET, NJ 08873BELMONT EQUIPMENT, A Division of Takara Belmont USA Inc.

10 2007** ****

Dental Delivery Unit

2007

ModelQUOLIS5000 UNIT

Serial No.

Rated Voltage

Rated Current

Classification

Manufactured

Product of Japan

ZZ535521

AW*********

****120V  ~ 60Hz

1 A

GS1-128

(**)**************(**)*********

2-1-1.Higashishinsaibashi,Chuo-ku,Osaka,Japan

Takara Belmont 

101 BELMONT DRIVE, SOMERSET, NJ 08873

BELMONT EQUIPMENT, A Division of Takara Belmont USA Inc.

10 2007

** ****

RECOGNIZED COMPONENT

PART1: GENERALREQUIREMENTS

MEDICAL ELECTRICALEQUIPMENT

CONFORMS TO UL STD 60601-13058278

Takara Belmont Corp.

17-1. Product label & ETL label are

located under the doctor table.

- 28 -

Page 33: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

18. Floor Template

18. Floor Template

8

5-1/

2

3/8

1-1/

2

5-3/

4

5-3/

82

1

2

TRANSFORMER

4-3/8

5-3/8

1-3/8

4

3-3/4

GRAVITY DRAIN(1-1/2 NORMAL PIPE)FOR PMU

GRAVITY DRAIN(1-1/2 NORMAL PIPE)

FINISHED FLOOR

TELECOMMUNICATIONS(1-1/2 NORMAL PIPE)

TELECOMMUNICATIONS(1-1/2 NORMAL PIPE)

ELECTRICAL CONDUITFOR QUAD BOX(1/2 O.D.)

ELECTRICAL CONDUITFOR QUAD BOX(1/2 O.D.)

AIR(1/2’ THREADED PIPE)

AIR(1/2’ THREADED PIPE)

WATER (For option)(1/2 THREADED PIPE)

WATER(1/2 THREADED PIPE)

CENTRAL VACUUM(5/8 O.D. TUBE)

CENTRAL VACUUM(5/8 O.D. TUBE)

-28-

- 29 -

Page 34: QUOLIS 5000 DENTAL UNIT...(5) Foot controller 1 set (6) Syringe parts 1 set (7) Operating manual, installation manual 1 set (8) Miscellaneous parts for installation as follow • Sleeve

NOTE

BELMONT EQUIPMENT, Division of Takara Belmont USA, Inc.101 Belmont Drive Somerset, New Jersey 08873 U.S.A. TEL.:(732) 469-5000 / (800) 223-1192 Fax.:(732)526-6322 / (800) 280-7504

TAKARA CO, CANADA LTD.2076 S. Sheridan Way, Mississauga, Ont., L5J2M4, Can. TEL.:(905) 822-2755 Fax.:(905)822-6203

R