Quantitative Sialic Acid Analysis - Ludger › docs › application-note › ludger-APN... ·...

3
Why is sialic acid analysis important? Sialic acids are negavely charged monosaccharides found on the non-reducing termini of glycans. They are important for the stability and 3D conformaon of glycoproteins and are involved in many biological interacons. Sialic acids oſten have a pivotal funconal impact: for example sialylaon of the N-glycans on IgG increases an-inflammatory acvity; O-acetylated sialic acids can change ligand interacons and affect degradaon (O-acetylated sialic acids including Neu5,9Ac 2 are found on EPO); and the presence of sialic acids also increases the serum half-life of glycoproteins by prevenng uptake by the asialoglycoprotein-receptor located on liver cells. A diverse range of sialic acids are found in nature, but the two major sialic acids found on N-glycans and O-glycans in biopharmaceucals are N-acetyl-neuraminic acid (Neu5Ac, or NANA) and N-glycolyl-neuraminic acid (Neu5Gc, or NAGA). Humans cannot synthesise Neu5Gc and its presence on a drug can lead to immune reacons such as chronic inflammaon. An-Neu5Gc anbodies have been detected in normal human sera, and can neutralise any Neu5Gc containing biopharmaceucal, thus lowering the drug’s efficacy. It is important to be aware that the choice of cell line can greatly influence the type of sialic acids present on a biopharmaceucal, for instance a large proporon of the sialic acids on mouse IgG are oſten Neu5Gc. It is therefore imperave, for drug safety and efficacy, to monitor both the level and types of sialic acids during all stages of the product life cycle as well as QC batch to batch consistency. Hence, sialic acid analysis is a regulatory requirement laid out in the ICH Q6B guidelines for characterisaon of biopharmaceucals. O OH O OH HO N OH OH AcO H O Quantave Sialic Acid Analysis Royle L, Hendel J, Kozak RP Ludger Ltd, Culham Science Centre, Oxfordshire, UK Application Note #APN002 How to perform quantave sialic acid analysis using Ludger kits and standards: Release of Sialic Acids Sialic acids are released by mild acid hydrolysis from the glycoprotein samples; typically from triplicate 50 µg amounts, although larger amounts may be required for samples with very low levels of sialylaon e.g. MAbs. We recommend a number of process controls are also taken through the process: sample buffer or water as negave controls; fetuin glycoprotein (GCP-FET-50U) and/or a quantave glycopepde (BQ-GPEP-A2G2S2-10U) as posive controls. Samples and process controls are incubated with 2M acec acid at 80°C for 2 hours. DMB Labelling To enable sensive detecon the released sialic acids are then fluorescently labelled with 1,2-diamino-4,5-methylenedioxybenzene. 2HCl (DMB) – see figure 1. When the sialic acid ring opens it exposes a ketone group next to the carboxylic acid, the DMB binds across these two adjacent keto groups producing a fluorophore (other monosaccharides do not have these adjacent keto groups, so they are not fluorescently labelled). The acid hydrolysed samples and process controls plus Ludger sialic acid standards: Neu5Ac (CM-NEUAC-01), Neu5Gc (CM-NEUGC-01), sialic acid reference panel (CM-SRP-01) and Neu5,9Ac 2 (CM-NEU5,9,AC2-01) if using, are incubated with DMB at 50°C for 3 hours. LC Analysis The Neu5Ac and Neu5Gc standards are diluted to produce a standard curve for quantaon of these two sialic acids. The sialic acid reference panel (CM-SRP-01) and Neu5,9Ac 2 are used for sialic acid idenficaon. Samples and controls are analysed by either HPLC using a LudgerSep-R1 column LS-R1-4.6x150 (table 1), or by UHPLC using a LudgerSep-uR2 column LS-UR2-2.1x100 (table 2) for faster run mes. Typical profiles for the sialic acid reference panel are shown in figure1 (A and B). What informaon is obtained by using Ludger Kits and Standards? Quantaon of sialic acids (Neu5Ac & Neu5Gc) as nmol/mg protein Relave quantaon of the O-acetylated Neu5,9Ac 2 N-glycolyl neuraminic acid Neu5Gc O OH O OH HO N OH OH HO H O HO 5-N-acetyl-9-O-acetyl neuraminic acid Neu5,9Ac 2 N-acetyl neuraminic acid Neu5Ac

Transcript of Quantitative Sialic Acid Analysis - Ludger › docs › application-note › ludger-APN... ·...

Page 1: Quantitative Sialic Acid Analysis - Ludger › docs › application-note › ludger-APN... · 2019-02-08 · How to perform quantitative sialic acid analysis using Ludger kits and

Why is sialic acid analysis important?

Sialic acids are negatively charged monosaccharides found on the non-reducing termini of glycans. They are important for the stability and 3D conformation of glycoproteins and are involved in many biological interactions. Sialic acids often have a pivotal functional impact: for example sialylation of the N-glycans on IgG increases anti-inflammatory activity; O-acetylated sialic acids can change ligand interactions and affect degradation (O-acetylated sialic acids including Neu5,9Ac2 are found on EPO); and the presence of sialic acids also increases the serum half-life of glycoproteins by preventing uptake by the asialoglycoprotein-receptor located on liver cells.

A diverse range of sialic acids are found in nature, but the two major sialic acids found on N-glycans and O-glycans in biopharmaceuticals are N-acetyl-neuraminic acid (Neu5Ac, or NANA) and N-glycolyl-neuraminic acid (Neu5Gc, or NAGA). Humans cannot synthesise Neu5Gc and its presence on a drug can lead to immune reactions such as chronic inflammation. Anti-Neu5Gc antibodies have been detected in normal human sera, and can neutralise any Neu5Gc containing biopharmaceutical, thus lowering the drug’s efficacy. It is important to be aware that the choice of cell line can greatly influence the type of sialic acids present on a biopharmaceutical, for instance a large proportion of the sialic acids on mouse IgG are often Neu5Gc.

It is therefore imperative, for drug safety and efficacy, to monitor both the level and types of sialic acids during all stages of the product life cycle as well as QC batch to batch consistency. Hence, sialic acid analysis is a regulatory requirement laid out in the ICH Q6B guidelines for characterisation of biopharmaceuticals.

O

OHO

OH

HON

OHOH

AcO

HO

Quantitative Sialic Acid AnalysisRoyle L, Hendel J, Kozak RP Ludger Ltd, Culham Science Centre, Oxfordshire, UK

Application Note#APN002

How to perform quantitative sialic acid analysis using Ludger kits and standards:

Release of Sialic AcidsSialic acids are released by mild acid hydrolysis from the glycoprotein samples; typically from triplicate 50 µg amounts, although larger amounts may be required for samples with very low levels of sialylation e.g. MAbs. We recommend a number of process controls are also taken through the process: sample buffer or water as negative controls; fetuin glycoprotein (GCP-FET-50U) and/or a quantitative glycopeptide (BQ-GPEP-A2G2S2-10U) as positive controls. Samples and process controls are incubated with 2M acetic acid at 80°C for 2 hours.

DMB LabellingTo enable sensitive detection the released sialic acids are then fluorescently labelled with 1,2-diamino-4,5-methylenedioxybenzene. 2HCl (DMB) – see figure 1. When the sialic acid ring opens it exposes a ketone group next to the carboxylic acid, the DMB binds across these two adjacent keto groups producing a fluorophore (other monosaccharides do not have these adjacent keto groups, so they are not fluorescently labelled). The acid hydrolysed samples and process controls plus Ludger sialic acid standards: Neu5Ac (CM-NEUAC-01), Neu5Gc (CM-NEUGC-01), sialic acid reference panel (CM-SRP-01) and Neu5,9Ac2 (CM-NEU5,9,AC2-01) if using, are incubated with DMB at 50°C for 3 hours.

LC AnalysisThe Neu5Ac and Neu5Gc standards are diluted to produce a standard curve for quantitation of these two sialic acids. The sialic acid reference panel (CM-SRP-01) and Neu5,9Ac2 are used for sialic acid identification. Samples and controls are analysed by either HPLC using a LudgerSep-R1 column LS-R1-4.6x150 (table 1), or by UHPLC using a LudgerSep-uR2 column LS-UR2-2.1x100 (table 2) for faster run times. Typical profiles for the sialic acid reference panel are shown in figure1 (A and B).

What information is obtained by using Ludger Kits and Standards?

• Quantitation of sialic acids (Neu5Ac & Neu5Gc) as nmol/mg protein• Relative quantitation of the O-acetylated Neu5,9Ac2

N-glycolylneuraminic acid

Neu5Gc

O

OHO

OH

HON

OHOH

HO

HO

HO

Neu5Gc

5-N-acetyl-9-O-acetyl neuraminic acid

Neu5,9Ac2

N-acetyl neuraminic acid

Neu5Ac

Page 2: Quantitative Sialic Acid Analysis - Ludger › docs › application-note › ludger-APN... · 2019-02-08 · How to perform quantitative sialic acid analysis using Ludger kits and

Minutes0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00

Fluo

resc

ence

12

3

4

5

6

7

*

*0.00 1.25 2.50 3.75 5.00 6.25 7.50

Minutes

Fluo

resc

ence

12

3

4

5

6

7

**

BA

LC Data

Time (min) Flow mL/min %A %B

0 0.5 100 0

19 0.5 100 0

19.5 0.5 10 90

23.5 0.5 10 90

24 0.5 100 0

30 0.5 100 0

Time (min) Flow mL/min %A %B

0 0.25 100 0

7 0.25 100 0

7.5 0.25 10 90

8 0.25 10 90

8.5 0.25 100 0

15 0.25 100 0

Table 1. 30 min running method for HPLC analysis using a LudgerSep-R1 column (4.6 x 150 mm, 3 µm particles) LS-R1-4.6x150. Injection volume = 25 µL.

Table 2. 15 min running method for UHPLC analysis using a LudgerSep-uR2 column (2.1 x 100 mm, 1.9 µm particles) LS-UR2-2.1x100. Injection volume = 5 µL.

LC Conditions Solvent A = acetonitrile:methanol:water 9:7:84Solvent B = acetonitrileFluorescence: Excitation ƛ: 373 nm; Emission ƛ: 448 nmColumn temp = 30°C; Sample temp = 10°C

Figure 1: Chromatogram of DMB Labelled Sialic Acid Reference Panel (CM-SRP-01) run on the LudgerSep-R1 HPLC column (A) or LudgerSep-uR2 UHPLC column (B)Peaks: 1 = Neu5Gc; 2 = Neu5Ac; 3 = Neu5,7Ac2; 4 = Neu5Gc,9Ac; 5 = Neu5,8Ac2; 6 = Neu5,9Ac2; 7 = Neu5,x,xAc3 (where x is an unknown acetyl position); * = Reagent.

Results

The peak areas of Neu5Ac and Neu5Gc in samples are compared to standard curves to provide quantitative data on the amounts of the sialic acids in your samples and in the process control. Figure 2 shows data for sialic acids released from mouse and human IgGs. The mouse IgG has high levels of the non-human sialic acid Neu5Gc, and in this case the human IgG has 4 times more sialic acids than the mouse IgG. Figure 3 shows data for the sialic acids in EPO. The absolute amounts of the Neu5Ac and Neu5Gc can be determined by comparison to the standard curve. However, EPO also contains the O-acetylated sialic acid Neu5,9Ac2. Under aqueous conditions the acetate on the 9 position can migrate to the 8 position, so it is difficult to have a fully quantitative standard for Neu5,9Ac2. The sialic acids are therefore identified by comparison to the retention times of the standards, and the relative proportions of the Neu5Ac, Neu5Gc and Neu5,9Ac2 are determined from their peak areas.

Figure 2: Identification and quantitation of Neu5Ac and Nue5Gc in mouse and human IgG.Analysis on LudgerSep-R1 HPLC column.

Neu5Gc

Neu5Ac

Fluo

resc

ence

Mouse IgG

Human IgG Human IgG nmol/mg protein

Sialic Acid Average %CV

Neu5Gc 0.00 -

Neu5Ac 8.90 2.81

Mouse IgG nmol/mg protein

Sialic Acid Average %CV

Neu5Gc 1.65 6.06

Neu5Ac 0.53 5.41

Minutes8.00 10.00 12.00

Neu5Ac

Page 3: Quantitative Sialic Acid Analysis - Ludger › docs › application-note › ludger-APN... · 2019-02-08 · How to perform quantitative sialic acid analysis using Ludger kits and

Figure 3: Identification and quantitation of sialic acids in EPO. Analysis on LudgerSep-R1 HPLC column.

© 2017 Ludger Ltd - Document #APN002 - Version 1.0 www.ludger.com

Ludger Ltd, Oxfordshire OX14 3EB UK

Product/Ordering Information

Sialic acid Release, and LabellingLT-KDMB-A1 LudgerTag DMB sialic acid release and labelling kit

Positive ControlsGCP-FET-50U-X4 Fetuin Glycoprotein Standard BQ-GPEP-A2G2S2-10U LudgerBioQuant GPEP A2G2S2 glycopeptide standard

Sialic Acid StandardsCM-NEUAC-01 N-acetylneuraminic acid (Neu5Ac or NANA) quantitative standardCM-NEUGC-01 N-glycolylneuraminic acid (Neu5Gc or NGNA) quantitative standardCM-NEU5,9,AC2-01 5-N-acetyl-9-O-acetyl neuraminic acid standard (Neu5,9Ac2)CM-SRP-01 Sialic acid reference panel

LC Analysis:LS-R1-4.6x150 LudgerSep R1 HPLC ColumnLS-UR2-2.1x100 LudgerSep uR2 UHPLC Column

Neu5GcNeu5Ac

Neu5

,7Ac

2

Neu5

,Gc,

9Ac

Neu5,9Ac2

Fluo

resc

ence

* Reagent

*

Sialic acid Reference

Panel

EPO

Neu

5Gc Neu

5Ac

Minutes6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00

Neu

5,9A

c 2

Relative % Area

Sialic Acid Average %CV

Neu5Gc 0.99 0.24

Neu5,Ac 93.91 0.02

Neu59Ac2 5.09 0.39

nmol/mg protein

Sialic Acid Average %CV

Neu5Gc 3.61 5.11

Neu5Ac 270.32 5.31

Neu5Gc & Neu5Ac quantified in nmol/mg protein by reference to standard curves

Relative proportions of Neu5Gc, Neu5Ac & Neu5,9Ac2 calculated from % peak areas.Ne

u5,8

,Ac 2

Neux

x,Ac

3

LT-KDMB-A1