Quality Testing — Ethanol in Hand Sanitizers · 2020-04-28 · Gas Chromatography (GC) Quality...

2
Gas Chromatography (GC) Quality Testing — Ethanol in Hand Sanitizers S-GC-F26 Ethanol World Health Organization (WHO) recommends the practice of good hand hygiene as one of the basic protective measure against COVID-19. This includes washing hands with soap and water or clean your hands with alcohol-based hand sanitizer. The U.S. Centers for Disease Control (CDC) recommends to use alcohol-based hand sanitizer that contain at least 60% ethanol or 70% isopropanol. Shimadzu GC is most suited for measurement of alcohol content in alcohol-based hand sanitizer. Nexis GC-2030 Column SH-Rt™-Q-BOND 30 m, 0.53 mm ID, df 20 μm (P/N: 221-75765-30) Inj. volume 3.0 uL (split ratio 11:1) Inj. temp. 250 °C Oven temp. 200 °C isothermal Detector TCD, 260°C The National Standard “Hygienic standard for alcohol disinfectants” (GB/T 26373-2010) specifies the amount of ethanol used in hand sanitizer to be 70-80%. GC-FID can provide a quick, easy, and accurate determination as illustrated. 1.935 2.215 2.723 2.527 min. 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 Column SH-Rtx™-624 30 m, 0.32 mm ID, df 1.8 μm (P/N: 221-75864-30) Inj. volume 1.0 μL (split ratio 20:1) Inj. temp. 280 °C Oven temp. Detector FID, 260°C 40 °C (hold 12 min) to 240°C @ 10°C/min (hold 10 min) Chinese Pharmacopoeia (CIP 2010 ver. 2) specifies the impurities in ethanol as such: methanol 0.02%; (acetaldehyde + acetal) 0.001% ; benzene 0.0002% ; other impurities 0.03% Measurement of Ethanol Content Maximum Allowable Impurities in Ethanol 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 0 2500 5000 7500 10000 12500 uV 1 2 3 4 5 6 FID1 Acetaldehyde Rs = 1.66 CIP Requirement: Rs > 1.5 for peak 1 & 2 Methanol Ethanol methane 2.063 water methanol ethanol acetone

Transcript of Quality Testing — Ethanol in Hand Sanitizers · 2020-04-28 · Gas Chromatography (GC) Quality...

Gas Chromatography (GC)

Quality Testing —Ethanol in Hand Sanitizers

S-GC-F26

EthanolWorld Health Organization (WHO) recommends the practice of good hand hygiene as one of the basic protective measure against COVID-19. This includes washing hands with soap and water or clean your hands with alcohol-based hand sanitizer.

The U.S. Centers for Disease Control (CDC) recommends to use alcohol-based hand sanitizer that contain at least 60%ethanol or 70% isopropanol. Shimadzu GC is most suited for measurement of alcohol content in alcohol-based hand sanitizer.

Nexis GC-2030

ColumnSH-Rt™-Q-BOND30 m, 0.53 mm ID,df 20 µm (P/N: 221-75765-30)

Inj. volume3.0 uL (split ratio 11:1)

Inj. temp.250 °C

Oven temp.200 °C isothermal

DetectorTCD, 260°C

The National Standard “Hygienic standard for alcoholdisinfectants” (GB/T 26373-2010) specifies theamount of ethanol used in hand sanitizer to be70-80%. GC-FID can provide a quick, easy, andaccurate determination as illustrated.

1.935 2.215 2.723

2.527

min. 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4

ColumnSH-Rtx™-62430 m, 0.32 mm ID,df 1.8 µm (P/N: 221-75864-30)

Inj. volume1.0 µL (split ratio 20:1)

Inj. temp.280 °C

Oven temp.

DetectorFID, 260°C

40 °C (hold 12 min) to240°C @ 10°C/min(hold 10 min)

Chinese Pharmacopoeia (CIP 2010 ver. 2) specifies theimpurities in ethanol as such: methanol ≤ 0.02%;(acetaldehyde + acetal) ≤ 0.001% ; benzene≤ 0.0002% ; other impurities ≤ 0.03%

Measurement of Ethanol Content Maximum Allowable Impurities in Ethanol

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 min

0

2500

5000

7500

10000

12500uV

12

3

4

5 6

FID1

AcetaldehydeRs = 1.66

CIP Requirement: Rs > 1.5 forpeak 1 & 2

Methanol Ethanol

methane

2.063 water

methanol

ethanol

acetone

Shimadzu Europa GmbH www.shimadzu.eu

For Research Use Only. Not for use in diagnostic procedures. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its affiliates, whether or not they are used with trademark symbol “TM” or “®”. Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.

The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.

Shimadzu (Asia Pacific) Pte Ltd., April 2020

An advanced interface enables intuitive operation with clear graphics.Shimadzu's latest tool-free maintenance technology makes dailymaintenance easy.

Nexis GC-2030Designed with the Analyst in Mind

Achieves the world’s highest sensitivity on the all of the detectors,such as FID and BID. The advanced flow controller (AFC) enhancesreliability with excellent repeatability.

World’s Highest Sensitivity and Reproducibility

Nexis GC-2030 can be customized to meet a customers' specific requirements and needs. Options and functions to use hydrogen carrier gas safely in high-speed analysis maximize analysis productivity.

Exceptional Expandability and Productivity

N E W P R O D U C T

I N N O V A T I O N

A W A R D 2 0 1 7

227-36322-01-20 to 240 °C

ID df 20m

0.10 mm 0.50 µm

Temp. Range 30m 60m 75m

– – –

20 µm

ID

0.32 mm

df 30 m

227-36327-01

Shimadzu GC column (SH-Rtx™-624)

Shimadzu GC column (SH-Rt™-Q-BOND)

227-36259-01-20 to 240 °C0.18 mm

0.25 mm

0.32 mm

0.53 mm

1.00 µm

1.40 µm

1.80 µm

3.00 µm

-20 to 240 °C

-20 to 240 °C

-20 to 240 °C 221-75865-75

– – –

10 µm0.53 mm 227-36302-01

Temp. Range

to 190 °C

to 190 °C

227-75865-60

227-36215-01

227-75864-60

227-75865-30

227-75863-30

227-75864-30

* Maximum temperatures listed are for shorter length columns. Longer columns may have a different maximum temperature.