QUALITY ASSURANCE IN THE METALS LABORATORY
FACTORS YIELDING HIGH QUALITY
Control all parameters Ruggedize system Regular maintenance
SAMPLE INTRODUCTION SYSTEM
CONTROLLING UPTAKE RATE%
EFFI
CIE
NC
Y
0
10
20
30
40
50
60
0.03 0.11 0.27 0.5 0.61 1.4 2.1
UPTAKE RATE (ML/MIN)
Transport Efficiency vs. Uptake
CONTROLLING UPTAKE RATE
Worn pump tubing?
Worn pump rollers?
Improperly adjusted pump tension?
Clogged nebulizer? Kinked sample capillary?
Real-Time Sample Flow Measurement
How it works: Thermal Flow Measurement
Flow Rate is related to (TempA – TempB)
Heater
Temp. Sensor B
sample
Temp. Sensor A
CONTROLLING UPTAKE RATE
Features of TruFlo:
Adjustable damping
Settable alarm limits
Digital display
Recordable graph
Range: 0 to 4mL/min
CONTROLLING UPTAKE RATE
CONTROLLING TEMPERATURE
Standard glass cyclonic Twister spray chamber; e.g. KT-1013
Quartz cyclonic Twister spray chamber; e.g. KT-1013Q
High purity PFA cyclonic spray chamber; e.g. KT-1013P
Low volume Twinnabar spray chamber; e.g. KT-1013T
Glass, quartz and PFA spray chambers are interchangeable
IsoMist Configurations:
CONTROLLING TEMPERATURE
0.91
1.11.21.31.41.51.61.71.81.9
22.12.2
10C 15C 20C 25C 30C 35C
ISOMIST TEMPERATURE
Nor
mal
ised
em
issi
on s
igna
l
Al 396.153As 193.696Ca 317.933Cd 214.440Cr 267.716Cu 327.393Fe 238.204Mg 285.213Mg 280.271Mn 257.610P 178.221Pb 220.353Se 196.026Si 251.611Ti 334.940Zn 202.548Zn 206.200Zn 213.857
Intensity vs. Temperature (1ml/min uptake)
Higher accuracy
Higher productivity
Greater reproducibility
Better experiments
CONTROLLING TEMPERATURE
CONTROLLING TEMPERATUREAv
erag
e In
tens
ity (c
ps)
Temperature (˚C)
Ir 212 Pt 214 Ru 240 Au 242 Pd 340 Rh 343
110000
90000
70000
50000
3000045403530252015105
ENHANCED SENSITIVITY FOR PRECIOUS METALS
CONTROLLING TEMPERATURE
Pt 214
Pt 265
Pt Gravimetric ValuePerc
ent P
latin
um (%
)
Temperature (˚C)
16.0
15.9
15.8
15.7
15.6
4540353025201515.5
IMPROVED ACCURACY FOR PRECIOUS METALS
CONTROLLING INTERFERENCES
ICP-MS
Ionization
Space-charge
Suppression
ICP-OES
EIE
Suppression
Noise
Internal Standardization:
Trident Internal Standard Kit
Modular construction Zero dead-volume
connections
Built-in mixing chamber
SeaSpray nebulizer
Twister spray chamber Argon Humidifier
Ceramic outer tube
DEALING WITH HIGH TDS
IMPROVED STABILITYFully Ceramic D-Torch – Less Maintenance & Improved Stability with High TDS
Glass Expansion October 2013 Newsletter, “How to Achieve High Accuracy with Difficult Samples.”
ENHANCE DAY-TO-DAY REPRODUCIBILTY
Re-usable inline filter.
Nebulizer cleaning tool.
Black flushing adapter.
Real-Time sample monitor.
QA Kit: P/N KT-1136
QC SUMMARY
Control sample uptake
Control spray chamber temperature
Control interferences
Control down-time