Alan Neal Charge Scientist Pathlab Waikato My Opinion & approach Quality Assurance in the 21 st Century

Transcript of Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification –...

Page 1: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Alan NealCharge ScientistPathlab Waikato

My Opinion & approach

Quality Assurance in the21st Century

Page 2: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

…. in 2019?• few laboratories today actually use quality goals

when managing the quality of their testing processes

• Laboratories continue with statistical QC procedures that are selected without consideration of the quality needed for the test and sometimes without consideration of the performance observed for the method……..

•Way forward – Risk management

www.westgard.com

Page 3: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

ISO 15189 - 2012

Specific criteria - no further clarification

Page 4: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Title

Page 5: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •
Page 6: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

https://www.elitecme.com/resource-center/laboratory/calibration-verification-webinar/

Page 7: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Calibration:

Assumed that all haematology analysers have a linear response over measurable range, this characteristic enables the use of single point calibration.

Calibration performed:

• Analyser set up• As directed by manufacture• Failure of calibration verification

In a previously stable analyser, Calibration is the last option in the resolution of poor performance, all other options must be excluded prior to this

procedure.

Page 8: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Calibration Verification:

Calibration verification frequency 6 monthly*, also:

• Direction of manufacture(s)• Direction of laboratory QMS• Review of reagent Lot to Lot variance• Major maintenance• Replacement of critical parts• Complete reagent change• QC trouble shooting

Page 9: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

*Calibration Verification – Automated cell counters:

If:

• follow manufactures instructions for operation• two level of controls material tested each day• Control materials within acceptable limits

Then this fulfils the calibration verification criteria

Also• Three or more levels of calibration material &• Includes Low, Mid & High values &• Performed at least six monthly

Then this would fulfil the calibration verification requirements

Calibration material* – contains a ‘known’ amount of analyte*Proficiency or control materials with know values are examples of calibration material

Page 10: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

*Calibration Verification – Automated cell counters:

Therefore if using IQC & / or EQA scheme and this has:

• Three or more levels of calibration material &• Includes Low, Mid & High values &• Information over 6 month period &• Good performance

Then this would fulfil the calibration verification requirements

Page 11: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Accuracy & Precision via IQC review:

• Only perform reproducibility check if high CVI• Risk of this approach – Group think / Bias!

Page 12: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Different approach but same outcome:

Page 13: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Pathlab Approach – Calibration / Precision:

Primary Analyser Secondary Analyser(s)- ‘Truth’ - Clone of Primary (xM xB values)- Calibration verification annual* - IQC’s compared to primary analyser- All EQA’s analysed - Monthly Cal validation (Intra-analyser comparison)

- (Review Web based IQC) - (Review Web based IQC)

Identical set up to primary analyser – xB, xM (targets and limits).

*Unless concern Re analyser performance – EQA, IQC poor performance? Of any value

Page 14: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Consideration(s):– Calibration / Calibration verification processes:

Matrix effect

Calibration material artificial (Alligator, fixed etc. & doesn’t behave the same as native

blood.

Page 15: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •

Summary….• Customise QA procedures for

individual assays based on their quality requirement, performance & risk

• Less (effort) for more (reward)

• Minimise false rejections in order to maximise response to real problems

• Spend time & effort on tasks that need it & not on those that don’t !!

Page 16: Quality Assurance in the 21st Century · • QC trouble shooting *Calibration Verification – Automated cell counters: If: • follow manufactures instructions for operation •