Quality Assurance in Diagnostic Medical Sonography

HHHoldorf

Page 2: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Quality Assurance -The “routine” periodic evaluation of the

US system to guarantee optimal image quality

Requirements1. Assessment of system components2. Repairs3. Preventive maintenance4. Record keeping

Done for medical and legal matters

Page 3: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE

Goals -proper equipment operation -detect gradual changes -minimize downtime -reduce number of repeat scans

Page 4: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Methods -test under known conditions -constant instrument settings *Results based on Transducer Type Receiver Gain Power Position of transducer

Depth*Make sure that when testing equip. that the

derived conclusion is system variation and not change in procedure

-use phantom with measurable characteristics

Page 5: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCEPerformed by the Sonographer -It is the duty of the sonographer to

perform quality assurance -Although the manufacturer’s service

engineer may be involved, the ultimate responsibility for QA always rests with the sonographer

-Various aspects of all instrumentation should be routinely inspected to ensure consistency of its performance

-It is important to validate the reliability of the images produced and the measurements made with each transducer on each US system.

Page 6: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCEAIUM 100mm Test Object -fluid filled tank containing strategically

located stainless steel pins or plastic strings -speed of sound in the AIUM object is

identical to that of soft tissue -Evaluates accuracy and performance

characteristics of a system. -BUT, the AIUM test object does not have

the attenuation properties of soft tissue, so a grayscale cannot be evaluated (BISTABLE)

Page 7: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

AIUM 100mm Test object

Page 8: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Placing the transducer on the top, sides and on

the oblique side of the object provides a variety of orientations between the sound beam and the pins.

1. Axial resolution= evaluated when the pins in the test object are parallel to the sound beam’s main axis

2. Lateral resolution= evaluated when the pins in the test object are perpendicular to the sound beam

3. Electronic caliper accuracy= evaluated by comparing the distances between reflections on the display with the actual distances in the test object

4. Dead zone= evaluated by scanning the pins located at the top of the test object, very close to the transducer.

Page 9: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Tissue Equivalent Phantom -have ultrasonic features similar to soft

tissue. -more suitable to evaluate the characteristics

of modern systems (grayscale, multi-focus, phased array transducers)

-strategically located pins, structures that mimic cysts & solid masses embedded in the phantom

-similar to soft tissue 1. Speed of sound2. Attenuation3. Scattering characteristics4. Echogenicity

Page 10: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

Tissue Equivalent Phantom

Page 11: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

Tissue Equivalent Phantom

Page 12: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Doppler Phantom

-flow phantoms are the devices of choice for evaluating Doppler systems

-used to assess the accuracy of pulsed, continuous wave, power and color flow systems.

1. vibrating string

2. moving belt phantoms

3. flow phantom

Page 13: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

Doppler Phantom

Page 14: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Modern phantoms include a circulation pump which propels a fluid

through vessels embedded in a tissue equivalent phantom Used most often a suspension mimicking blood.

(problems: air bubbles and changing consistency over time)

Tests for effective:

1. Penetration of Doppler beam

2. Ability to discriminate between different flow directions

3. Accuracy of measured flow speed

Page 15: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Slice Thickness Phantom -evaluation of slice thickness is also called

elevational thickness -imaging plane is thicker than either the beam

width or the SPL. -phantom’s medium mimics soft tissue. -thicker slices diminish spatial resolution and

reduce the ability to visualize small, low contrast lesions

-When the US beam is overly thick, cystic structures may appear filled in “ARTIFACTS”

Page 16: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

Slice Thickness Phantom

Page 17: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Performance MeasuresSensitivity is the ability of a system to display low

level echoes with a tissue equivalent phantom.1. Minimum sensitivity (image deeper) Pick a deep pin, with the TGC set flat, increasing

the gain from its minimum value to the point when an echo is displayed on the CRT determines the minimum sensitivity.

The same rod in the test object should always be imaged for continuity.

Assesses the detection of low level echoes in the far field

Page 18: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE2. Normal sensitivity Settings are those which all the pins, solid

masses, and cystic structures in the test phantom are accurately displayed.

Output power, TGC, and amplification are adjusted to establish normal sensitivity

Found at higher gain than the minimum sensitivity All subsequent quality assurance and performance

measurements are made at the normal sensitivity.

(because we have to see all the structures to make measurements)

Page 19: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE3. Maximum sensitivity (image shallower) Evaluated with the output power and amplification

of the system set to the maximum practical levels With these settings, a tissue equivalent phantom

is imaged, and the depth of tissue-like texture is determined.

Maximum visualization depth is used to assess sensitivity, and should not differ from one routine evaluation to the next

***Sensitivity is also assessed when the sonographer adjusts the system controls to change echo brightness from barely visible to full brightness

Page 20: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Dead Zone

-results from the time that it takes for the system to switch from the transmit to the receive mode

-region close to the transducer where images are inaccurate

-extends from the transducer to the shallowest depth from which meaningful reflections appear

-information within the dead zone is unreliable and may not be used in the diagnostic setting

-the dead zone is assessed with the most shallow series of pins in a test object

Page 21: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

Dead Zone

Page 22: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Dead Zone

-an acoustic standoff, or “gel pack” positioned between the transducer and the patient allows accurate imaging of important superficial structures (50 cc bag of IV fluid)

-an increasingly deeper dead zone may indicate a cracked crystal, detached backing material, or a longer pulse duration

Page 23: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Registration Accuracy

-ability of the system to place reflections in proper positions while imaging from different orientations

Range Accuracy (vertical depth calibration)

-system’s accuracy in placing reflectors at correct depths located parallel to the sound beam

-if differences appear the error may be caused by

1. System malfunction

2. The speed of sound in the phantom is different than 1,540 m/s

Page 24: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Horizontal calibration

-system’s ability to place echoes in their correct position perpendicular to the sound beam

Focal zone (surrounds focus)

-focus of phased array transducers, must be carefully evaluated

-lateral resolution is excellent in the focal zone

-systems with dynamic receive focusing should produce narrow reflections over a wide range of depths

Page 25: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Axial Resolution

-smallest distance at which two pins parallel to the sound beam are displayed as two distinct echoes.

-evaluated by scanning a set of closely spaced pins within the phantom

-uses pins parallel to the sound beam Lateral Resolution

-minimum distance that two side-by-side rods are displayed as two distinct images

-look to see if pins are perpendicular to sound beam

Page 26: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCECompensation Operation or Uniformity

-Using a tissue equivalent phantom and scanning from the top, the echoes are displayed with TIME GAIN COMPENSATION (TGC) off and then with the TGC on.

-With the TGC off, the echoes should be displayed with reduced amplitude as depth increases

-With the TGC on, all of the echoes should have the same amplitude, regardless of depth

Page 27: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

QUALITY ASSURANCE Mock Cysts and Tumors -Using the tissue equivalent phantom to check

dimensions of cysts -Also, note texture and fill-in Display and Grayscale Dynamic Range -Adjusting the system’s output power and

amplification should produce changes in the grayscale display.

-Important to compare the relationship between the image on the system’s screen with the output of all other display devices, such as remote viewing stations

Page 28: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS GOLD STANDARD: a perfect technique that we deem 100% accurate to

which our ultrasound results are compared.

Since the majority of sound energy produced by the transducer remains in the body, it is important to know output of the ultrasound system.

Page 29: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Hydrophone

-small needle with a piezoelectric crystal at its end

- a wire connects the PZT crystal to an oscilloscope and the needle is placed in the ultrasound beam

-displayed are acoustic signals received by the crystal

-since the hydrophone is small, the acoustic pressure is measured at specific locations within the sound beam

-can quantitate amplitude, period, pulse duration and pulse repetition period

Page 30: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

Hydrophone

Page 31: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTSRadiation Force -transducer’s sound beam creates a

very small, but measureable force on any target that it strikes.

-if the target is a balance or a float, the measured force relates to the power in the beam

-when the sound beam is entirely absorbed or reflected by the target, the target acts as an extremely sensitive miniature postal scale

-measures SATA or SPTA intensity

Page 32: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS

Acoustic-Optics

-Based on the interaction of sound and light

-A shadowing system, called Schlieren, uses this principle to allow us to view the shape of a sound beam in a medium

-Quantitation of the PRP, amplitude, period, and pulse duration

Page 33: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Three devices measure the output of

ultrasound transducers by absorption, the conversion of sound energy into heat:

1. Calorimeter -measures the total power of the entire

sound beam through the process of absorption

-the sound beam is directed into the calorimeter where the sound energy is converted into heat (absorption)

-the sound beam’s total power is calculated by measuring the temperature rise and the time of heating

Page 34: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Thermocouple

-tiny electronic thermometer

-a dab of absorbing material is placed on the thermocouple

-the thermocouple is inserted into the sound beam and the temperature is measured by absorbing material

-the temperature rise is related to the power of the sound beam at the particular location where the device is located

Page 35: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTSLiquid Crystals

-certain crystals change color based on their temperature

-tank full of crystals

-When a sound beam strikes a crystal, the sound energy is absorbed

-The change in temperature causes a change in their color, providing insight into the shape and strength of the sound beam

Page 36: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS RISK-BENEFIT RELATIONSHIP

1. Benefits must outweigh the risks of the exam

2. Obstetrics and fetal medicine is a common application of diagnostic sonography. There is no confirmation of harm resulting from its use

3. Extremely high ultrasound intensities damage biologic tissue

4. Low intensity ultrasound has no known Bioeffects

5. Under controlled circumstances, Bioeffects are beneficial

Page 37: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS AIUM’S AND FDA’S Bioeffects intensity limit:

SPTA (Spatial Peak/Temporal Average)

Highest output intensities with pulsed Doppler

Lowest output intensities with gray scale imaging

100 mW/cm2 unfocused1,000mW/cm2 focused

Page 38: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Focused beams are less likely to cause temperature elevation in tissues

Unfocused beams are more likely to cause temperature elevation in tissues

**this occurs because a narrow beam heats only a small region of tissue and the heat is rapidly transferred to and dissipated by adjacent tissues that were not heated by the US beam.

Page 39: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Dosimetry

-Science of identifying and measuring the characteristics of an ultrasound beam that are relevant to its potential for producing biological effects.

Bioeffects may be conducted in two broad areas:

1. In Vivo- living

2. In Vitro- nonliving

Page 40: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS 1. In Vivo

-performed within the living body of a plant or an animal.

e.g. Studying the effects of US exposure on the lung tissue of a laboratory rat.

***Difficult to study in vivo because of attenuation***

2. In Vitro

-performed outside the living body in an artificial environment.

e.g. A computer model estimating the temperature elevation of tissues during exposure to US.

Page 41: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS

AIUM Statement on In Vitro Bioeffects

1. In vitro Bioeffects research is important

2. In vitro Bioeffects are real even though they may not apply to the clinical setting

3. In vitro Bioeffects which claim direct clinical significance should be viewed with caution (without in vivo validation)

Page 42: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS STUDY TECHNIQUES:

1. Mechanistic Approach - begins as a proposal that a specific

mechanism has the potential to produce Bioeffects

-based on that proposal, a theoretical analysis is performed to estimate the scope of the Bioeffects at various exposure levels.

-searches for a cause and effect relationship

Page 43: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS EMPERICAL APPROACH

-based on acquisition and review of information from patients or animals exposed to US

-studying charts of patients who have been exposed to US

-seeks a relationship between the exposure to US and the effects of that exposure

-searches for a relationship between an exposure and a response

***strongest conclusions when conclusions to both approaches agree.

Page 44: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS MECHANISMS OF BIOEFFECTS

1. Thermal=proposes that Bioeffects result from tissue temperature elevation

-as sound travels in the body, energy is converted into heat.

-bone is an absorber. There fore temperature elevation at a tissue-bone interface is more likely.

-temperature elevation in fetal soft tissue is considered of potentially greater harm than adults. Thus, fetal soft tissues adjacent to bone are of great concern

Page 45: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTSBody core temperature is regulated at 37

degrees C. Life processes do not function normally at other temperatures.

Any exam that causes an elevation in temperature of less than 2 degrees C may be used without reservation

Any exam that causes a temperature elevation to greater than 41 degrees C is considered potentially harmful to a fetus.

Page 46: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Thermal index (TI):useful predictor of max. temperature increase

under most clinically relevant conditions.

3 forms:

1. TIS=assumes that sound is traveling in soft tissue

2. TIB=assumes that bone is at or near the focus of the sound beam

3. TIC=assumes that the cranial bone is in the sound beam’s near field

Page 47: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Empirical findings:1. Serious tissue damage occurs from

prolonged elevation of body temperature

2. Tissue heating is related to the output characteristics of the transducer and the properties of the tissue

3. A 2 degree to 4 degree in testicular temperature can cause infertility

4. A combination of temp and exposure time determine the likelihood of harmful Bioeffects

Page 48: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS

5. No confirmed Bioeffects have been reported for temp elevations of up to 2 degrees above normal for exposures of less than 50 hrs.

6. Max. heating is related to the beam’s SPTA intensity

7. Fetal tissues appear less tolerant of tissue heating than adults.

8. A greater amount of acoustic energy is absorbed by bone than by soft tissue.

Page 49: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS Mechanistic findings:

1. Theoretical models appear to correlate with experimental data even though:

-the ultrasound beam is quite complex

-diagnostic equipment is diverse

-tissue characteristics are different

***strong argument***

Page 50: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS2. Cavitation=interaction of sound waves

with microscopic, stabilized, gas bubbles in the tissues.

-Also known as gaseous nuclei= NOT CONTRAST AGENTS. NATURAL GASEOUS NUCLEI THAT OCCURS NATURALLY IN THE BODY.

- Excitation takes on the form of shrinking and expanding of the bubble. Potential of near total energy absorption where the nuclei exist may lead to thermal injury

- MI relates to cavitation

Page 51: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS1. Stable Cavitation

-occurs at lower MI levels

-the gaseous nuclei tend to oscillate or expand and contract.

-bubbles that are a few millimeters in diameter might double in size-bubble does not burst

-the bubbles intercept and absorb much of the acoustic energy

-EFFECTS: the fluids surrounding the cells undergo micro streaming and the cells are exposed to shear stresses

Page 52: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

BIOEFFECTS2. Transient Cavitation (inertial or normal) -at higher MI levels, transient cavitation

occurs -bubble bursting -EFFECTS: produces highly localized, violent

effects, including:1. Colossal temperatures2. Shock waves -The destructive effects of transient cavitation

are not considered clinically important since they are highly localizes and affect few cells.

-The pressure threshold for transient is only 10% higher than that required for stable cavitation

Page 53: Quality Assurance in Diagnostic Medical Sonography HHHoldorf.

Conclusion

The American College of Radiology (ACR) strongly recommends that the training for “appropriately trained sonographers or service engineers” conducting routine QC be provided by a qualified medical physicist.

If unable to acquire training by a qualified medical physicist, training can be achieved through the ultrasound equipment manufacturer or through an appropriate course.