08/22/08 1 Performance Testing and Dosimetry in Computer Tomography Asim Gilani Radiation Protection Unit Medical Physics Department Dubai Hospital

description

Performance Testing of CT (Computer Tomography) using ACR (American College of Radiology) Phantom

Transcript of Qaulity Control of CT

Page 1: Qaulity Control of CT

08/22/08 1

Performance Testing and Dosimetry in Computer

Tomography

Asim Gilani

Radiation Protection Unit

Medical Physics Department

Dubai Hospital

Page 2: Qaulity Control of CT

08/22/08 2

Annual Performance Evaluation

•Alignment light accuracy

•Alignment of table to gantry

•Table/gantry tilt

•Slice localization from localization image (SPR)

•Table incrementation accuracy

•Slice thickness

•Image quality

– High contrast (spatial) resolution

– Low contrast resolution

– Image uniformity

– Noise

– Artefact evaluation

•CT number accuracy and constancy

Page 3: Qaulity Control of CT

08/22/08 3

Annual Performance Evaluation

•Display devices– Video display

– Hard-copy display

•Dosimetry-Computed Tomography dosimetry index (CTDI)

-Patient radiation dose for representative examinations

•Safety evaluation– Visual inspection

– Audible/visual signals

– Posting requirements

– Scattered radiation measurements

Page 4: Qaulity Control of CT

08/22/08 4

ACR Performance Phantom

Page 5: Qaulity Control of CT

08/22/08 5

Positioning Accuracy

CT # Accuracy

Slice Width Accuracy

Low Contrast Resolution

High Contrast (Spatial) Resolution

Image Uniformity and Noise

Image Artifacts

Distance Measurement Accuracy

The ACR CT phantom was designed to examine a broad range of scanner parameters

Page 6: Qaulity Control of CT

08/22/08 6

4 3 2 1

20 c

m

4 cm

Head Foot

AlignmentAlignment

CT # CT #

Slice widthSlice width

Low Low contrast contrast

resolutionresolution

Uniformity Uniformity & noise & noise

Distance Distance accuracy accuracy

& SSP& SSP

High High contrast contrast

resolutionresolution

Page 7: Qaulity Control of CT

08/22/08 7

•Align Module 1 to lasers

•Scan with Hi Resolution Chest protocol

Test 1: Phantom and Scanner Test 1: Phantom and Scanner AlignmentAlignment

Page 8: Qaulity Control of CT

08/22/08 8

Page 9: Qaulity Control of CT

08/22/08 9

ACR Performance Phantom

Page 10: Qaulity Control of CT

08/22/08 10

Air ≈ -1000 HU

Water ≈ 0 HU

Polyethylene ≈ -97 HU

“Bone” ≈ +910 HU

Acrylic ≈ +120 HU

MODULE 1

Page 11: Qaulity Control of CT

08/22/08 11

MODULE 1

Page 12: Qaulity Control of CT

08/22/08 12

High-resolution chest technique

Must see all four BBs

Longer wire must have same number of lines above and below (±1)

Wires are 0.5 mm apart in z-direction

WW = 1000WL = 0

Page 13: Qaulity Control of CT

08/22/08 13

•Prescribe a scan at center of Module 4 i.e., 120 mm superior

•Scan with Hi Resolution Chest protocol

Test 2: Table Distance Travel Test 2: Table Distance Travel TestTest

Page 14: Qaulity Control of CT

08/22/08 14

12

5 9

4

6

7

8

10

Bar patterns:lp/cm MODULE 4

Page 15: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 16: Qaulity Control of CT

08/22/08 16

Test 3: CT Number Accuracy and Test 3: CT Number Accuracy and Slice ThicknessSlice Thickness

•Align Module 1 to lasers

•Scan with Adult Abdomen protocol & all kVp’s available

•Record all HU and slice thickness

•Also scan with 3, 5, 7 mm & hi resolution chest slice thickness

•Record water HU and slice thickness for each scan.

Page 17: Qaulity Control of CT

08/22/08 17

Adult abdomen technique

CT number measured in 5 materials

CT number of water measured at several scan widths and kVps

WW = 400WL = 0

“Bone”

Water

Poly

Acrylic Air

Page 18: Qaulity Control of CT

08/22/08 18

Adult abdomen technique

Scan width measured at several scan widths

Page 19: Qaulity Control of CT

08/22/08

DA

TA

FO

RM

DA

TA

FO

RM

Page 20: Qaulity Control of CT

08/22/08 20

Test 4: Low Contrast ResolutionTest 4: Low Contrast Resolution

• Scan Module 2 with Abdomen and Head protocols

• Record the smallest rod seen

Page 21: Qaulity Control of CT

08/22/08 21

25 mm

6 mm

5 mm

4 mm

3 mm

2 mm

Low contrast = 6 HU ± 0.5 HU

MODULE 2

Page 22: Qaulity Control of CT

08/22/08 22

Adult abdomen and head techniques

Record diameter of the smallest set of rods for which all 4 rods can be seen

ROI check of absolute contrast using large rod WW = 100

WL = 100

Page 23: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 24: Qaulity Control of CT

08/22/08 24

Test 5: Uniformity, Noise, Test 5: Uniformity, Noise, Artifacts, and Distance AccuracyArtifacts, and Distance Accuracy

• Scan Module 3 with Abdomen protocol• Record HU & SD in center• Record HU at 12, 3, 6, 9 o’clock• Examine for artifacts and record• Check distance accuracy (optional)

Page 25: Qaulity Control of CT

08/22/08 25

MODULE 3

100 mm

Page 26: Qaulity Control of CT

08/22/08 26

Adult abdomen and head techniques

Measure uniformity and noise

Assess for artifacts

Measure distance accuracy(optional)

WW = 100WL = 0

Page 27: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 28: Qaulity Control of CT

08/22/08 28

• Scan Module 4: – Hi Resolution Chest protocol– Abdomen protocol

• Record limiting lp/mm

Test 6: High Contrast Test 6: High Contrast ResolutionResolution

Page 29: Qaulity Control of CT

08/22/08 29

12

5 9

4

6

7

8

10

Bar patterns:lp/cm MODULE 4

Page 30: Qaulity Control of CT

08/22/08 30

Adult abdomen, adult head and high-resolution chest techniques

Record the first highest frequency bar pattern for which the bars and spaces merge

WW = 100WL 1100

Page 31: Qaulity Control of CT

08/22/08 31

WW = 400WL = 0

CT ART?

At soft-tissue window settings these streaks are visible and not of concern

Page 32: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 33: Qaulity Control of CT

Questions

Page 34: Qaulity Control of CT

08/22/08 34

• ACR Manual

ReferencesReferences

Page 35: Qaulity Control of CT

08/22/08 35

Thank You