Pulsed Voltammetric Techniques

10
1 of 10 Pulsed Voltammetric Techniques Typical Set up for Polarography Pulsed Voltammetric Methods

description

Pulsed Voltammetric Methods. Pulsed Voltammetric Techniques. Typical Set up for Polarography. Pulsed Voltammetric Methods. DC Polarography with Dropping Mercury Electrode (DME). Pulsed Voltammetric Methods. Sample-and-Hold or Tast (sampled-current) Voltammetry. D E. E. t. t. - PowerPoint PPT Presentation

Transcript of Pulsed Voltammetric Techniques

Page 1: Pulsed Voltammetric Techniques

1 of 10

Pulsed Voltammetric Techniques

Typical Set up for Polarography

Pulsed Voltammetric Methods

Page 2: Pulsed Voltammetric Techniques

2 of 10

DC Polarography with Dropping Mercury Electrode (DME)

Pulsed Voltammetric Methods

Page 3: Pulsed Voltammetric Techniques

3 of 10

Sample-and-Hold or Tast (sampled-current) Voltammetry

Pulsed Voltammetric Methods

E

t

E

6/13/2*2/1708)( mCnDi OOd

3/13/2 )(00567.0)( EEmCi PZCid

Current is sampled late in the drop life, so faradaic current is minimized versus near onset.

LOD ~ 10-5 M

Page 4: Pulsed Voltammetric Techniques

4 of 10Normal Pulse (NPV or NPP) vs. Tast Voltammetry

Pulsed Voltammetric Methods

Current sampled and held.

2/12/1

*2/1

)'()(

OO

NPVdCnFADi

' If > 5RsCd (~0.05-15 ms), icharging is almost insignificant. LOD ~ 10-5 -10-6M

Page 5: Pulsed Voltammetric Techniques

5 of 10

Page 6: Pulsed Voltammetric Techniques

6 of 10

DPV

NPV

Differential Pulse (DPV) vs. Normal Pulse VoltammetryPulsed Voltammetric Methods

Small ifaradaic

when away from E0’

Large ifaradaic when near E0’

LOD ~ 10-6 -10-7 M

Page 7: Pulsed Voltammetric Techniques

7 of 10

NPV

DPV

OsteryoungSquareWaveVoltammetry

Comparison of Excitation Waveforms for Pulsed VoltammetriesPulsed Voltammetric Methods

Page 8: Pulsed Voltammetric Techniques

8 of 10Osteryoung Square Wave Voltammetry Pulsed Voltammetric Methods

LOD ~ 10-8 M

Page 9: Pulsed Voltammetric Techniques

9 of 10Osteryoung Square Wave Voltammetry

Pulsed Voltammetric Methods

Page 10: Pulsed Voltammetric Techniques

10 of 10

Stripping Analysis Coupled to DPV - DPASV

ip = 0.138 Qmetal/tpulse-width

~50-nm thick Hg film on Glassy Carbon (vitreous) electrode by electrodeposition from acidic Hg2+

DPVDeposit M0

Limit of Detection of ~0.005 ppm

Pulsed Voltammetric Methods