PS2016 Series High Precision Isolated Data Collector PS2016V /PS2016I PS2016V-485 /PS2016I-485 User Manual Rev: D Beijing Smacq technologies Co., Ltd. www.smacq.com www.smacq.cn

Transcript of PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device...

Page 1: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

PS2016 Series High Precision Isolated Data Collector

PS2016V /PS2016I

PS2016V-485 /PS2016I-485

User Manual

Rev: D

Beijing Smacq technologies Co., Ltd.

www.smacq.com

www.smacq.cn

Page 2: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

2

Content

1. PRODUCT OVERVIEW ............................................................................................................... 3

1.1. GENERAL INTRODUCTION ............................................................................................................... 3

1.2. PRODUCT FEATURES ...................................................................................................................... 3

1.3. PRODUCT SPECIFICATIONS ............................................................................................................... 4

2. PRODUCT APPEARANCE AND SIGNAL CONNECTION ................................................................. 5

2.1. PRODUCT APPEARANCE .................................................................................................................. 5

2.2. SIGNAL CONNECTION ..................................................................................................................... 6

3. INSTALLATION AND TESTING .................................................................................................... 7

3.1. DRIVER INSTALLATION .................................................................................................................... 7

3.2. SOFTWARE INSTALLATION................................................................................................................ 8

3.3. VOLTAGE SIGNAL TEST .................................................................................................................... 8

3.4. TEMPERATURE SIGNAL TEST ........................................................................................................... 11

3.5. DATA FILE DETAIL ......................................................................................................................... 14

4. PROGRAMMING GUIDE ......................................................................................................... 15

4.1. OVERVIEW ................................................................................................................................. 15

4.2. GENERAL COMMANDS ................................................................................................................. 15

4.3. SETTING COMMANDS ................................................................................................................... 15

4.4. MEASUREMENT COMMANDS ......................................................................................................... 16

5. ORDERING INFORMATION ..................................................................................................... 17

6. SERVICE AND WARRANTY ...................................................................................................... 18

7. REVISION HISTORY ................................................................................................................. 19

Page 3: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

3

1. Product Overview

1.1. General introduction

Smacq PS-2000 Series data collector is a compact, fully isolated multi-channel testing device

with up to four and a half readings to ensure the accuracy of test data. It is ideal for design

verification, automatic test and data acquisition, aging monitoring and many other test and

measurement applications.

Only one PS-2000 series data collector is needed to start multi-channel test tasks, scaling up

to 500 channels. It can help you start test tasks at very low cost and complete the task with very

high efficiency.

1.2. Product features

up to 4 ½ digit reading resolution,and 0.015% DC voltage accuracy

Max 200V DC voltage range

DC voltage measurement 16 channel full differential input

Only 1 module is needed to start multi-channel testing tasks, with 500-channel

scalability.

Optical isolation between measurement input and computer port

Flexible configuration system. The switch unit can collaborate with third-party test

equipment to set up test system.

SCPI instruction, Visual Studio, LabVIEW and MATLAB development support, convenient

for independent development of automated test system.

USB or RS-485 interface optional

Page 4: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

4

1.3. Product specifications

The following specification parameters, if not specified otherwise, were tested after running the

device for 30 minutes under the temperature of 25 degree Celsius and the relative humidity of

40%.

Diagram 1、Product specifications

Model PS2016V PS2016V-485 PS2016I PS2016I-485

Parameters Voltage, Temperature(Thermocouple) Current

Resolution 4½ - Digit

Basic accuracy

±(% reading + digits)

Voltage:0.02%+4

0.05%+4 Temperature

(Thermocouple)

K ±2℃

J ±2℃

R ±3℃

S ±3℃

E ±2℃

T ±2℃

N ±2℃

Range

Voltage:20V,200V

100mA Temperature

(Thermocouple)

K -50~1200℃

J -50~910℃

R 0~1750℃

S 0~1750℃

E -50~670℃

T -100~400℃

N -50~1300℃

Resolution

20V range:1mV

200V range:10mV

Temperature:0.1℃

10uA

Input resistance 1MΩ 20Ω

Sampling rate 2.5Sa/s

Page 5: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

5

Channel 16

Channel type 2 line differential input

Channel isolation >10GΩ

Relay pull time 6ms

Relay release time 4ms

Relay life expectancy Mechanical 100,000,000 cycles;Electrical 500,000 cycles (full load)

Connection method Bundled screw terminals

Control method SCPI instruction

Communication

interface USB RS-485 USB RS-485

PSU(Built-in

self-healing fuse) USB power 5VDC USB power 5VDC

Power consumption 5VDC,<150mA

Dimensions Refer to chapter 2.1 dimensions diagram

Weight Without screw terminals:500g

With screw terminals:525g

Working

environment

Temperature: 0°C to +50°C

Relative humidity: 5% to 80% (No condensation)

Storage

requirements

Temperature: –20°C to 80°C

Relative humidity: 5% to 90% (No condensation)

Calibration cycle 1 year

2. Product Appearance and Signal Connection

2.1. Product appearance

PS-2000 series DAQ device utilizes metal shield case. It uses plug-in screw terminal connectors to

connect signals. The USB communication interface uses Type-B USB connector. Figure 1 shows its

appearance. The dimensions inside the figure are millimeter.

Page 6: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

6

2.2. Signal connection

The screw terminal connector in front of the DAQ device is for connecting the measured

signals.

On the back of the DAQ device, there are two groups of connectors, one USB interface for

connecting with computers directly; the other RS485 interface for connection with computers or

embedded processing system like PLC.

Figure 2 and Diagram 2 show its pin distribution and the corresponding signal in detail.

Diagram 2、Screw terminal connector corresponding signals

Name Notes

ch+ ch0+ to ch15+ is the signal positive input of channel 0 to 15

ch- ch0- to ch15- is the signal negative input of channel 0 to 15

5VDC 5VDC power input of RS-485 interface

GND Power ground of RS-485 interface

A+ Positive input of RS-485 communication signal

B- Negative input of RS-485 communication signal

Attention:For models with USB interface, there is no need of extra 5VDC power supply.

Figure 2、Screw terminal connector pin detail

Page 7: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

7

3. Installation and Testing

3.1. Driver installation

When PS2016 DAQ device is connected to computers running Microsoft Windows for the first

time, it will require installing the driver before the computer can recognize PS2016 correctly.

Open the folder of "\PS Series DAQ/driver" on the Resources CD, double click the file

"SETUP.EXE", then click "Install", and there will be a dialog of success after the driver is installed.

After correctly installing the driver, Windows Device Manager will recognize PS2016 as a virtual

serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port

number for the device as shown in parentheses in Figure 3.

Figure 3、The Device Manager with correctly recognized PS2016

Page 8: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

8

3.2. Software installation

Open the folder "\PS Series DAQ\PS2016 - DAQ Lite Setup\Volume" and run the file "setup.exe".

Set the installation directory for the software, click "Next" all the way till it's installed. Usually you

can keep the default installation directory. Start the software after the installation. Figure 4

shows the software interface.

3.3. Voltage signal test

Connect a voltage signal of a known voltage to the PS2016 data collector, as shown in Figure

5.

Figure 4、PS2016-DAQ Lite software interface

Page 9: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

9

Make corresponding settings on PS2016-DAQ Lite, as shown in Figure 6, with detailed notes

below.

1:Select the correct serial port connection.

2:Set the sampling cycle in seconds.

3:Set the range. Can be set as 20V or 200V range when measuring voltage signals.

4:Set the path of data storage.

5:Choose the channel to run the DAQ device in.

Figure 5、Voltage signal test connection

Attention:When measuring voltage signals, if the thermocouple range is

selected by mistake, the PS2016 DAQ device may be damaged after it runs.

Page 10: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

10

Within the data acquisition interface after finishing the settings, click "START" button to

begin the data acquisition and real-time data storage, as showed in Figure 7, with detailed notes

below.

1: Start/Stop button

2: Real time value display.

3: Trend graph display.

4: Corresponding channel legend display.

In the trend graph display, the ordinate scale value is automatically scaled by default. You

can cancel the "automatic Y scale adjustment" by right-clicking the pop-up menu on the ordinate

scale, so as to achieve fixed vertical coordinates. After fixing the ordinate, you can double-click

the ordinate number to modify the ordinate scale value.

The same method can be used to modify the coordinate scale value.

Figure 6、Voltage testing settings

Page 11: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

11

3.4. Temperature signal test

When using PS2016 DAQ device to measure temperature, the thermocouple needs to be

connected to the signal as shown in Figure 8 below.

When using K type thermocouple, the maximum measuring range is 1100℃.

When using other types of thermocouples, the maximum measuring range is the temperature

range corresponding to 45.108mV on its indexing table.

Figure 7、Voltage test acquisition interface

The thermocouple probe must be insulated during the connection. The exposed metal

on the thermocouple probe cannot be connected directly to any voltage signal.

Otherwise, the PS2016 DAQ device may be damaged.

Page 12: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

12

You will need to set the range as "thermocouple" in settings, and choose the thermocouple

type you use, as shown in Figure 9, with detailed notes below:

1:Choose correct serial port number

2:Set sampling cycle, measured in seconds.

3:Set range. Choose thermocouple range when measuring temperature and choose

corresponding thermocouple type.

4:Set data storage path.

5:Choose the channel to acquire.

Figure 8、Thermocouple signal test connection

Page 13: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

13

Within the data acquisition interface after finishing the settings, click "START" button to

begin the data acquisition and real-time data storage, as showed in Figure 10, with detailed notes

below.

1: Start/Stop button

2: Real time value display.

3: Trend graph display.

4: Corresponding channel legend display.

In the trend graph display, the ordinate scale value is automatically scaled by default. You

can cancel the "automatic Y scale adjustment" by right-clicking the pop-up menu on the ordinate

scale, so as to achieve fixed vertical coordinates. After fixing the ordinate, you can double-click

the ordinate number to modify the ordinate scale value.

The same method can be used to modify the coordinate scale value.

Figure 9、Temperature test setting

Page 14: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

14

3.5. Data file detail

All data files are saved in .csv format, readable by Microsoft Excel, as shown in Figure 11 below.

The data file consists of beginning time, cycle, measuring type and all original data for this

acquisition.

Figure 10、Temperature test acquisition interface

Figure 11、Data file

Page 15: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

15

4. Programming Guide

4.1. Overview

When you need to embed PS2016 into your own measuring and control system, you can

control PS2016 via the SCPI instructions detailed in this chapter.

USB interface

The USB interface on PS2016 DAQ device is virtual serial port. It has a default baud of 9600,

8 bit data bits, 1 stop bit, no hardware handshake and its command uses linefeed "\n" as the end

character, i.e. 0x0A.

RS-485 interface

If your PS2016 DAQ device is equipped with RS-485 interface, it has a default baud of 9600,

8 bit data bits, 1 stop bit, no hardware handshake and its command starts from 1-byte address,

with linefeed "\n" as the end character, i.e. 0x0A. When the address byte is set to 0x00, it

represents the broadcast.

4.2. General commands

*IDN?

Get device information.

*RST

Reset device default settings. This command will disconnect all channels but will not

change the baud setting.

4.3. Setting commands

BPS:9600

Setting serial communication baud as 9600.

BPS:115200

Setting serial communication baud as 9600.

RNG:[n]

Page 16: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

16

Set DAQ device range. N can be set as 2, 3, or 9。

When set as 2, it represents the range of 20V; when set as 3, it represents the range of

200V; when set as 9, it represents measuring thermocouple.

ADDR:[n]

Set the address of RS-485 interface. This command is only for models with RS-485

interface. During the setting, there can be only one PS2016 connected to RS-485 bus. Then

the device sends a broadcast address to set its address.

4.4. Measurement commands

CH[n]:V?

Read measured voltage from specified channel with return value measured in V. The value

of n ranges from 0 to 15, corresponding to channel 0 to channel 15.

For example:USB send CH0:V?\n

USB answer 0.01000\n

RS-485 send \00 CH0:V?\n

RS-485 answer \00 0.01000\n

CH[n]:T[TYPE]?

Read measured temperature from specified channel with return value measured in ℃. The

value of n ranges from 0 to 15, corresponding to channel 0 to channel 15.

TYPE can be set as K, J, R, S, B, E, T or N, corresponding to respective type of thermocouple.

For example:USB send CH0:TK?\n

USB answer 25.5678\n

RS-485 send \00 CH0:TK?\n

RS-485 answer \00 25.5678\n

Page 17: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

17

5. Ordering information

Device

Model Notes

PS2016V 16 channel differential voltage input, Thermocouple measurement with cold

end compensation, USB interface

PS2016I 16 channel current input, USB interface

PS2016V-485 16 channel differential voltage input, Thermocouple measurement with cold

end compensation, RS485 interface

PS2016I-485 16 channel current input, RS485 interface

Optional accessories

Model Notes

USB-A-B USB connector,1.5 meters,USB-A type to USB-B type

TB8-3.81 Screw terminal connector,8 digits,with spacing of 3.81mm

Optional accessories

Model Notes

SDIN 35mmDIN guide mounting bracket

STHK-1.5 K type thermocouple,1.5 meters

STHK-3 K type thermocouple,3 meters

STHK-5 K type thermocouple,5 meters

Page 18: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

18

6. Service and Warranty

All Smacq products can be serviced free of charge under normal use within their warranty. Refer

to the warranty notes inside their package for details.

Unless otherwise provided in this manual or warranty statement, Smacq does not provide any

other warranties, either express or implied.

If you need technical support or more service details, or if you have any questions using this

product or reading this document, feel free to contact us.

Tel:010-52482802

Email:[email protected]

Website:www.smacq.com / www.smacq.cn

Page 19: PS2016 Series High Precision Isolated Data Collector DAQ UserManual en.pdf · serial port device under the name of "USB-SERIAL CH341A". The OS will assign a serial port number for

19

7. Revision history

Date Version Notes

2016.8.30 Rev: A Initial release

2016.11.14 Rev: B

1、 Added independent range setting for every channel in

software PS2016-DAQ Lite.

2、 K-type thermocouple supports measuring up to 1100℃.

2017.3.30 Rev: C

Change RS-485 connector pin detail to correspond to the

changes in products with RS-485 interface manufactured after

April 2017.

2017.4.20 Rev: D Update the accuracy of thermocouples measurements.