30
R.L.Paul, a P. Halsted, a J.Ballinger, a P.J. Blower, a P.K. Marsden, a A. Trivett A, c D. Lloyd, c M.J. O’Doherty b K. Wood, d D.J.Honess, d M.I. Saunders, e a Guy’s, King’s & St Thomas’ School of Medicine, London, UK b Guy's and St Thomas' NHS Trust, London, UK c University of Kent, Canterbury, UK d Gray Cancer Institute Northwood e University College, London, UK Production of Cu-64 and Synthesis and Biological Evaluation of 64 CuATSM and second generation analogues

Transcript of Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. ·...

Page 1: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

R.L.Paul,a P. Halsted,a J.Ballinger,a P.J. Blower,a P.K. Marsden,a A.

Trivett A,c D. Lloyd,c M.J. O’Dohertyb

K. Wood,d D.J.Honess,d M.I. Saunders,e

aGuy’s, King’s & St Thomas’ School of Medicine, London, UKbGuy's and St Thomas' NHS Trust, London, UK

cUniversity of Kent, Canterbury, UKdGray Cancer Institute Northwood

eUniversity College, London, UK

Production of Cu-64 and Synthesis and

Biological Evaluation of 64CuATSM and

second generation analogues

Page 2: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

COPPER RADIONUCLIDESCOPPER RADIONUCLIDES

Isotope Half Life Radiation Source

60Cu 20 min +

(93%), EC (7%) cyclotron

61Cu 3.3 hours +

(62%), EC (38%) cyclotron

62Cu 9.74 mins +

(98%), EC (2%) generator/cyclotron

64Cu 12.7 hours +

(18%), EC (41%), - (37%)

reactor/cyclotron

66Cu 5.1 mins -

(100%) reactor/cyclotron

67Cu 62 hours -

(100%) (52%) reactor/cyclotron

Page 3: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

ObjectivesObjectives

�� Production of CuProduction of Cu--6464

6464Ni(Ni(p,np,n))6464CuCu

T1/2 12.7 hT1/2 12.7 h

+ (18%), - (37%),

EC (41%)

� Synthesis of hypoxia-selective agents based on [64Cu(ATSM)]

NN N

HS

N

SH

H3C CH3

NNH3C CH3

H H

ATSM

Page 4: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Production of CuProduction of Cu--64 : 3 main steps64 : 3 main steps

�� Electroplating of NiElectroplating of Ni--6464

�� Irradiation of the TargetIrradiation of the Target

�� Purification of CuPurification of Cu--64 64

�� tracer productiontracer production

Page 5: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

11.5 11.5 MeVMeV proton beamproton beam

Page 6: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

The ChemistryThe Chemistry

Requires separation of CuRequires separation of Cu--64 from other isotopes 64 from other isotopes after target irradiationafter target irradiation

99.699.60.330.330.0020.0020.020.020.040.04Content (%)Content (%)

NiNi--6464NiNi--6262NiNi--6161NiNi--6060NiNi--5858IsotopeIsotope

<10<10<10<1020020010103030<10<10800800<10<10<10<10<1<1ContentContent

((ppmppm))

ZnZnCuCuFeFeMnMnCrCrTiTiCaCaSiSiMgMgBBElementElement

Page 7: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

64CuCl2 after Purification

0

20000

40000

60000

80000

100000

120000

140000

160000

Energy (keV)

no. C

ounts

(4 m

in)

Cu-64, 511 KeV

Page 8: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

CuATSMCuATSM

and analogueand analogue

NN N

S

N

S

Cu

H3C CH3

NNH3C CH3

H H

NN N

S

N

S

Cu

CH3

NNH3C CH3

H H

CuATSM(hypoxia)

E1/2 -0.59 V

CuPTSM(flow imaging agent)

E1/2 -0.51 V

Page 9: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

ATSE: uptake at different oxygen levels in vitro

70

0

10

20

30

40

50

60

70

0 10 20 30 40 50 60

Time (min)

% uptake

0% 0.1% 0.5% 5% 10% 20%

A

70

0

10

20

30

40

50

60

70

0 10 20 30 40 50 60

Time (min)

% uptake

0% 0.1% 0.5% 5% 10% 20%

A

0

10

20

30

40

50

60

70

80

90

100

1 10 100 1000 10000 100000 1000000

Oxygen conc (PPM)

Cu-ATSM Cu-ATSE Cu-ASSM

ATSE

ATSE

ATSM 1 h

ATSE uptake switches on at higher O2 levels (10x less hypoxic) than ATSM: more relevant to radiobiological hypoxia

NN N

S

N

S

Cu

H3C CH3

NN

H H

CH3H3C

Page 10: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

ATSE vs. ATSM: ATSE vs. ATSM:

biodistributionbiodistribution

ATSM and ATSE

BALB/c mice, EMT6 tumour, 20 min

0

5

10

15

20

25

30

35

blo

od

lun

g

live

r

sp

lee

n

kid

ne

y

he

art

bra

in

bo

ne

tum

ou

r

gu

t

mu

sc

le

tissue

%ID

/g ATSM

ATSE

ATSE uptake higher in tumour, lower in liver and kidney

NN N

S

N

S

Cu

H3C CH3

NN

H H

NN N

S

N

S

Cu

H3C CH3

NNH3C CH3

H H

CH3H3C

ATSE ATSM

Page 11: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Mechanism of Hypoxia SelectivityMechanism of Hypoxia Selectivity

Me

NH

N

Me

H

N

S

N N

S

N

MeMe

Cu

Me

NH

N

Me

H

N

S

N N

S

N

MeMe

Cu

Me

NH

N

Me

H

N

S

N N

S

N

MeMe

Cu

+ e-

Cu(II)ATSM

-

hypoxic

O2fast fast

Me

NH

N

Me

H

N

S

N N

S

N

MeMe

H H

“Cu” +slow

O2

Page 12: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

NNN

S

N

SN NR3

R4

R3

R4

R1 R2

Cu

R1 R2 R3 R4

GTS H H H H

GTSM H H CH3 H

PTS CH3 H H H

PTSM CH3 H CH3 H

PTSM2 CH3 H CH3 CH3

PTSE CH3 H C2H5 H

PTSP CH3 H C6H5 H

ATS CH3 CH3 H H

ATSM CH3 CH3 CH3 H

CTS C2H5 CH3 H H

CTSM C2H5 CH3 CH3 H

DTS C2H5 C2H5 H H

DTSM C2H5 C2H5 CH3 H

Biological Evaluation: Screening of ‘new’Bis(thiosemi-carbazones)

Page 13: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Screening of ‘new’ Bis(thiosemi-carbazones)

NNN

S

N

SN NR3

R4

R3

R4

R1 R2

Cu

R1 R2 R3 R4

PTS CH 3 H H HPTSM CH 3 H CH 3 HPTSM2 CH 3 H CH 3 CH 3

PTSE CH 3 H C2H5 H

ATS CH 3 CH 3 H HATSM CH 3 CH 3 CH 3 HCTSCTSM C2H5 CH 3 CH 3 HDTS C2H5 C2H5 H HDTSM C2H5 C2H5 CH 3 H

Page 14: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Screening of ‘new’ Bis(thiosemi-carbazones)

NNN

S

N

SN NR3

R4

R3

R4

R1 R2

Cu

R1 R2 R3 R4

PTSE CH 3 H C2H5 H

ATS CH 3 CH 3 H HATSM CH 3 CH 3 CH 3 H

DTS C2H5 C2H5 H HDTSM C2H5 C2H5 CH 3 H

Page 15: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Screening of ‘new’ Bis(thiosemi-carbazones)

0

1

2

3

4

5

6

7

8

9

10

0 10 20 30 40 50 60 70

20.90%

0.00%

Time (mins)

0

1

2

3

4

5

6

7

8

0 10 20 30 40 50 60 70

20.90%

0.50%

0.00%

ATS

ATSM

NNN

S

N

SN N

Cu

NNN

S

N

SN N

Cu

Page 16: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic gadolinium enhanced MRI in a rat tumourmodel

Honess DJ1, Wood KA1,2 , Maxwell RJ3, Wilson I1, Paul RL4, O’Doherty MJ4, Marsden PK4, Blower PJ4, Sanghera B5, Wong W5, Saunders MI2

1 University of Oxford Gray Cancer Institute, Northwood, Middlesex, UK2 University College Hospital, London, UK 3 Northern Institute for Cancer Research, University of Newcastle, Newcastle, UK4 Guy’s, King’s, St Thomas’ School of Medicine, London, UK5 Paul Strickland Scanner Centre, Mount Vernon Hospital, Northwood, Middlesex, UK

Page 17: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Background

• Regional hypoxia within human tumours is a significant

cause of treatment failure.

• Non-invasive identification of hypoxic regions could improve

outcome by use of intensity modulated radiotherapy (IMRT).

• A reliable non-invasive hypoxia marker is not yet available.

• 64CuATSM is a PET tracer being

evaluated as a hypoxia marker.

• Uptake of the tracer in the first few minutes may

discriminate between hypoxia and normoxia (Lewis et al; J Nucl

Med 1999;40:177-183).

Me

NH

N

Me

H

N

S

N N

S

N

MeMe

Cu

Cu(II)ATSM

64Cu t 1/2 = 12.7h

Page 18: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Aim

To assess the degree of influence of blood flow on marker uptake

into a rat tumour, using:

1. Gadolinium-enhanced dynamic MRI to monitor blood flow

into the tumour immediately before….

2. PET scanning to monitor tumour 64CuATSM uptake in the

same slice.

Tumour methods

• P22 tumours s/c in left flank of BDXI rats, n = 8

• Rats anaesthetised

• Rats placed in same jig for MR and PET scanning

Page 19: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

MR methods:• 4.7 T Varian with 6 cm RF coil

• 100 gradient echo images, one every 6 sec

• (TR 60 ms; TE 2.5 ms; 1.0 mm slice thickness)

• After 30 sec, Gd-DTPA given by i/v infusion for 5 sec

• AUC of the uptake curve for first 90 sec (AUC90)

calculated; this is a robust measure of blood flow in

this tumour

PET methods:• Concorde MicroPET Focus 220

• 10-60 MBq 64CuATSM given i/v; scanning for

60 min

• Images reconstructed in 5 minute segments

for viewing tracer uptake and 10 minute

segments for calculation of mean standard

uptake values (SUVs) for the central tumour

slice (0.8 mm thick)

Page 20: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

CTI CTI microPETmicroPET Studies Studies

with with 6464CuCu--ATSMATSM

Non tumour bearing rat

10.09 MBq 64Cu-ATSM

Coronal view (head at top)

Uptake seen in liver, kidneys and bowel

Page 21: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

P22 tumour bearing rat: 4.2 P22 tumour bearing rat: 4.2 MBqMBq 6464CuCu--ATSMATSM

Axial ViewThrough tumour

Coronal view Through tumour

Sagittal view, liver attop

Page 22: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

MR results:Images of a typical tumour are shown below.

Before Gd-DTPA At 90sec (60 sec after Gd-DTPA)

tumour

Tumour Gd-DTPA uptake appears to be greater at the periphery, with less well-perfused central areas.

tumour

Page 23: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

PET results:Transaxial PET images of the same slice after 20.3 MBq

64CuATSM.

Mean activity for 0 - 5 min after 64CuATSM

Mean activity for 55 -60 min after 64CuATSM

At early times 64CuATSM uptake is greater at the periphery than the centre, with a very similar pattern to Gd-DTPA distribution.

After ~ 15 -20 min this pattern disperses and uptake is distributed across the tumour

Page 24: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Correlation of quantitative MR and PET

data - early:

There is good correlation between MRI measurement of blood flow

(AUC90) and early 64CuATSM uptake (mean SUV 0-10 min):

AUC90 = 6856 mmol.sec for this tumour

Mean SUV 0 -10 min = 1.31 for the same tumour

The correlation coefficient R for all 8 tumours = 0.75(p = 0.04)

Page 25: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Correlation of quantitative MR and PET

data - later:But there is no correlation between MRI measurement of blood flow

(AUC90) and later 64CuATSM uptake (mean SUV 50-60 min):

AUC90 = 6856 mmol.sec for this tumour

Mean SUV 50-60 min = 3.49 for the same tumour

The correlation coefficient R for all 8 tumours = 0.44(p = 0.27)

Page 26: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Conclusions:

• Early 64CuATSM uptake appears to be dominated by perfusion.

• Later, by 50-60 min, this quantitative correlation is lost.

• Hence later tracer retention in this model is determined by other

factors, possibly hypoxia.

Further investigation is in progress using a vascular damaging agent to modify

tumour hypoxia and pimonidazole, an immunohistochemical hypoxia marker.

• The data indicate that clinical PET imaging of 64CuATSM should

not be carried out immediately after tracer administration, but

time allowed for factors other than perfusion to determine tracer

retention.

Page 27: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

AcknowledgementsAcknowledgements

Guys’, Kings’, St Thomas’School of Medicine

Mr Phil Halsted

Prof Phil BlowerProf Phil Blower

Dr Paul Marsden

Dr Jim Ballinger

EPSRC for funding

The University of KentThe University of Kent

Ms Amanda Ms Amanda TrivettTrivett

Dr Dan LloydDr Dan Lloyd

The Gray Cancer InstituteThe Gray Cancer Institute

Dr Dr DavinaDavina HonessHoness

Dr Katie WoodDr Katie Wood

Prof Michelle SaundersProf Michelle Saunders

Page 28: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

Purification of 64-CuCl2 by Column Chromatography

0

10

20

30

40

50

60

70

80

1 2 3 4 5 6 7 8 9 10

Fraction no. (~ 2 ml)

Actinity (M

Bq)

Page 29: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

ICP MS AnalysisICP MS Analysis

0.5M0.5M

6M6M

8M8M

[HCl][HCl]

1010

99

88

77

66

55

44

33

22

11

FractionFraction

from from

Column Column

(2ml)(2ml)

--0.00200.00200.00040.00040.01300.013025.84325.843

0.00070.0007--0.00010.00010.00460.004610.09810.098

0.00480.0048--0.00010.00010.00620.00626.6076.607

0.00550.0055--0.00010.00010.00690.006912.75212.752

0.01830.0183--0.00020.00020.00770.007715.56415.564

61.324961.32490.00050.00050.01400.0140880.026880.026

606.4071606.40710.02000.02000.55490.55499777.949777.94

428.6611428.66110.01380.01380.36470.36477820.097820.09

--0.00170.0017--0.00020.00020.00730.00736.7586.758

--0.00120.0012--0.00000.00000.00340.00343.5063.506

NiNi--6464

((ppmppm))

NiNi--6161

((ppmppm))

NiNi--6060

((ppmppm))

ZnZn--6464

((ppmppm))

Page 30: Production of Cu-64 and Synthesis and Biological Evaluation of 64 … · 2019. 4. 26. · Preliminary evaluation of the effects of blood flow on PET detection of 64Cu-ATSM by dynamic

CTI CTI microPETmicroPET StudiesStudies

�� P22 P22 carcinosarcomacarcinosarcoma is a tumour with clinically relevant is a tumour with clinically relevant hypoxia levels.hypoxia levels.

�� CuATSMCuATSM is clearly taken up in the tumours and is clearly taken up in the tumours and distribution dependent on blood flow to an extent.distribution dependent on blood flow to an extent.

�� Uptake is nonUptake is non--uniform and there are regions of hypoxia.uniform and there are regions of hypoxia.

�� Tumour blood flow is being monitored by GdTumour blood flow is being monitored by Gd--enhanced enhanced MRI before the 64CuMRI before the 64Cu--ATSM PET scanning and ATSM PET scanning and comparison with comparison with immunohistochemicalimmunohistochemical pimonidazolepimonidazolestaining.staining.