Product Information PI · PAINT COLOR:WHITE(Munsell N9.5) (4-M10) 140 145 90 444 10 155...
Transcript of Product Information PI · PAINT COLOR:WHITE(Munsell N9.5) (4-M10) 140 145 90 444 10 155...
The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Canon Electron Tubes & Devices Co., Ltd. (CETD) for any infringements of patents or other rights of the third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of CETD or others.
The information contained herein may be changed without prior notice. It is therefore, advisable to contact to CETD before processing with the design of equipment incorporating this product.
Product Information PI
XRC-4552X
Rotating Anode X-ray Tube Assembly Compact X-ray tube assembly for CT scanner featuring
extremely high cooling performance.
Liquid metal lubricated bearings (LM bearings) is applied in the rotation system.
All-metal extra-heavy anode disc is constructed with specially processed rhenium-tungsten-faced molybdenum alloy target.
The anode heat content is 2840 kJ (4000 kHU) and the maximum anode heat dissipation rate is 10.2 kW.
General Data IEC Classification (IEC60601-1:2005+A1:2012) .................. Class Ⅰ ME EQUIPMENT Electrical:
Circuit: High Voltage Generator ............................................ Constant Potential High-Voltage Generator Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage ................................................................................................... 135 kV Nominal Focal Spot Value:
Large Focus .................................................................................................................... 1.4 × 1.4 Small Focus .................................................................................................................... 0.9 × 0.7
Nominal Anode Input Power (at 4s): Large Focus ........................................................................................................................ 48 kW Small Focus ........................................................................................................................ 30 kW
Nominal CT Anode Input Power: Large Focus ........................................................................................................................ 48 kW Small Focus ........................................................................................................................ 26 kW
Nominal CT Scan Power Index (CTSPI): Large Focus ........................................................................................................................ 44 kW Small Focus ........................................................................................................................ 22 kW
XRC-4552X
- 2 -
Motor Ratings:
Stator: XS-AY
1st
Start-up 2nd
Start-up Stand-by
Boost-up from
Stand-by Scanning
Driven Frequency [Hz] 50 120 100 120 120 Input Power [W] 1500 500 350 950 500 Voltage [V] 130 168 120 200 168 Current [A] 15 7.5 6.0 9.0 7.5 Min. Speed Up [s] 5 15 6 9 - Notes 1) To be obtained with 3-phase starter ST-7018 or equivalent. 2) These data are indicating standard values. 3) Anode keeps continuous rotation regardless X-ray output.
Anode Speed ..................................................................................................... Minimum 6360 min-1 Stator:
Type .......................................................................................................................... Three-phase Each Winding Impedance .................................................................................................... 3.8 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ (Measured by DC 500 V)
Heat Exchanger Input Power ................................................................................................... 360 W Heat Exchanger Input Voltage ...................................................................... Single-phase AC 200 V Heat Exchanger Source Frequency ..................................................................................... 50/60 Hz Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 °C Mode of Operation ........................................................................................................... Intermittent Envelope Current (135kV / 100mA) .......................................................................... (Approx) 10 mA Envelope Voltage (135kV / 100mA) ............................................................................................ 0 kV
Mechanical:
Dimensions ................................................................................................ See Dimensional Outline. Overall Length .................................................................................................................. 493 mm Maximum Diameter .......................................................................................................... 213 mm
Target: Anode Angle ................................................................................................................. 7 degrees Diameter .......................................................................................................................... 140 mm Construction ..................................................................................................... Rhenium-tungsten
Permanent Filtration ................................................................. 1.1 mm Al / 75 kV IEC60522:1999 Radiation Protection:
X-ray Leakage ............................................................................................ Maximum 0.87 mGy/h Leakage Technique Factor ................................................................................. 135 kV, 29.6 mA
X-ray Coverage: Longitudinal Direction of Tube Axis ...................................................................... Anode side 3.3°
Cathode side 21° Perpendicular Direction of Tube Axis .................................................................................... ± 27°
Weight: Tube Housing Unit .................................................................................................. Approx. 41 kg Heat Exchanger ...................................................................................................... Approx. 26 kg
High Voltage Receptacle ......................... To Meet Requirements of IEC60526 Corrigendum1:2010 Cooling Method ........................................................................................................ Heat Exchanger Position During Operation ............................................................... ± 30° with Respect to Tube Axis Position During Shipping ................................................................................. Anode Facing Upward G Proof in Gantry Rotation ........................................................................... Maximum 12 × 9.8 m/s2 Housing Model Number ......................................................................................................... XH-168
XRC-4552X
- 3 -
Absolute Maximum and Minimum Ratings (At any time, these values must not be exceeded.)
Maximum X-ray Tube Voltage ................................................................................................ 135 kV Between Anode (or Cathode) and Ground ............................................................................. 67.5 kV Minimum X-ray Tube Voltage ................................................................................................... 80 kV Maximum X-ray Tube Current:
Large Focus ...................................................................................................................... 400 mA Small Focus ...................................................................................................................... 300 mA
Maximum Filament Current (25 kHz Maximum): Large Focus .......................................................................................................................... 5.2 A Small Focus .......................................................................................................................... 5.2 A
Filament Voltage (25 kHz Maximum): Large Focus (At Maximum Filament Current 5.2 A) ................................................... 9.7 ~ 13.2 V Small Focus (At Maximum Filament Current 5.2 A) ................................................. 10.5 ~ 14.4 V
Continuous Anode Input Power ......................................................................... 4.0 kW (5.63 kHU/s) Thermal Characteristics:
Maximum Anode Heat Content ..................................................................... 2840 kJ (4000 kHU) Maximum Anode Heat Dissipation .............................................................. 10.2 kW (14.4 kHU/s) Operating Anode Heat Dissipation ................................................................ 5.7 kW (8.03 kHU/s) X-ray Tube Assembly Heat Content .............................................................. 3750 kJ (5280 kHU) Nominal Continuous Input Power:
With Heat Exchanger ............................................................................... 4.0 kW (5.63 kHU/s) Ambient Temperature 40 Oil Temperature 80
Peak Power Input Ratings: Do not exceed these value in any anode heat content.
Exposure Time [sec] Peak Power [kW]
Small Focal Spot Large Focal Spot
4 10 20 30 40
100 300
30 29 27 26 25 15
0
48 43.5 40 37 36 20
0
XRC-4552X
- 4 -
Environmental Limits
Operating Limits: Temperature ................................................................................................................ 18 ~ 45
Humidity ........................................................................................................................ 30 ~ 80 % (No Condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa Altitude ................................................................................................................................ 2000 m
Shipping and Storage Limits: Temperature .............................................................................................................. -20 ~ 75
Humidity ........................................................................................................................ 20 ~ 90 % (No Condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
XRC-4552X
- 5 -
0
10
20
30
40
50
60
70
80
90
100
0 3 6 9 12 15 18 21 24 27 30
HE
AT
ST
OR
AG
E [
%]
TIME [min]
COOLING
HEATING
10.2kW
5.7kW
Anode Heating / Cooling Curve
Notes
1. Heat storage of the anode is calculated based on the anode thermal characteristics.
This determines the X-ray input conditions for subsequent X-ray exposure. For specifying
conditions, contact CETD.
2. To avoid the over-load, the OLP (overload protection) program should be used to restrict the
X-ray input conditions. The X-ray tube assembly should not be used in systems which do not
incorporate the OLP program.
When absolutely impossible to use a system which incorporate the OLP program, contact
CETD for operating procedures.
3. Before you make the OLP program, contact CETD for the detail conditions.
XRC-4552X
- 6 -
0
500
1000
1500
2000
2500
3000
3500
0 30 60 90 120 150 180
HE
AT
ST
OR
AG
E [
kJ]
TIME [min]
3750
COOLING
HEATING
4.0kW
X-ray Tube Assembly Heating / Cooling Curve
XRC-4552X
- 7 -
0
100
200
300
4.0 4.2 4.4 4.6 4.8 5.0 5.2
Ep=120kV
Ep=135kVEp=100kV
Ep=80kV
TU
BE
CU
RR
EN
T [
mA
]
FILAMENT CURRENT [A]
0
100
200
300
400
4.0 4.2 4.4 4.6 4.8 5.0 5.2
Ep=120kV
Ep=135kV
Ep=100kV
Ep=80kV
420
TU
BE
CU
RR
EN
T [
mA
]
FILAMENT CURRENT [A]
Emission Characteristics
Large Focus
Ep: Tube Voltage
For Reference Only
Small Focus
Ep: Tube Voltage
For Reference Only
XRC-4552X
- 8 -
5
6
7
8
9
10
11
12
13
3.0 3.5 4.0 4.5 5.0
FIL
AM
EN
T V
OL
TA
GE
[V
]
FILAMENT CURRENT [A]
5.2
SF
LF
Filament Characteristics
SF : Small Focus LF : Large Focus
For Reference Only
XRC-4552X
- 9 -
30°20
6.3
130
57.2
259
.5
35°
58.3
130
(20
)(20
)
55.2±1
234
259
φ92
φ213
100
φ177
108
130
112
.21126
8.8
115.382
67.6
174
.421
4
φ165A
B
PRESSURE RELIEF VALVE (1.96MPa)(NORMALLY CLOSED)
No Painting.Oil hose length 2m.
CABLE A
-1.0mm≦A≦1.0mm
-1.0mm≦B≦1.0mm
ANODE
OIL INLET
FOCAL SPOT
OIL OUTLET
L:LARGE FOCUSS:SMALL FOCUSC:COMMON
CATHODE
CABLE B
CEN
TR
AL R
AY A
ND
REFER
EN
CE A
XIS
CENTRAL OF FOCAL SPOT
FOCAL SPOT POSITION
7°
4-M6 15 DEEP
(EARTH TERMINAL OF INSSRT TUBE)
Dimensional Outline of Tube Housing Unit Unit: mm
XRC-4552X
- 10 -
Position of Screw Hole for Fixing to Gantry unit : mm
XRC-4552X
- 11 -
UVWEGME
: U PHASE WINDING OF THE STATOR COIL: V PHASE WINDING OF TNE STATOR COIL: W PHASE WINDING OF THE STATOR COIL: EARTH TERMINAL OF HOUSING ASS.: GETTER TERMINAL : EARTH TERMINAL OF INSERT TUBE METAL ENVELOPE
CONNECTOR (ALDEN H-403)
TUBE HOUSING UNIT
CONNECTOR (ALDEN H-407)
WU V
4
4
1
1
2
2
CABLE A 1
WH
ITH
32
E GG
63 5
5
7
7
8
8
ME
9
9
4 65 7
31 2 4 5 6 87TERMINAL TBS01
9
8 9WHITH BLACK
CABLE B
BLA
CK
OR
AN
GE
GR
EEN
/YELLO
W
YELLO
W
PU
RP
LE
WH
ITE
TEMPERATURE RELAY (80 3 C)(NORMALLY CLOSED)
Terminal Connections of Tube Housing Unit
XRC-4552X
- 12 -
SWITCH BUTTON
VIEW FROM A-A'
A A' LED (YELLOW)
PAINT COLOR:WHITE(Munsell N9.5)
(4-M
10)
140
145 90
444
10
90155
465
305
2000
G F
OR
CE
GANTRYROTATINGDIRECTION
12
4 4
26
4
CABLE C
AIR
FLO
W
Dimensional Outline of Heat Exchanger Unit: mm
XRC-4552X
- 13 -
Terminal Connections of Heat Exchanger
250 [V] 10 [A]
Marketing Engineering Group, Sales Department 1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan Tel: +81-287-26-6666 Fax: +81-287-26-6060 https://etd.canon
PE-XRC-4552X 2018-11