Probe Characterization

30
Probe Characterization Amber Precision Instruments, Inc. (API) Jul. 23, 2014

description

Probe Characterization. Amber Precision Instruments, Inc. (API). Our EMI Probes. EMI Probes: Up to 6 GHz Up to 20 GHz Up to 40 GHz. Optional EMI Probes; Choose: Size Frequency range Field Component. * EMI Hx 2 mm. Characterization Structure up to 15 GHz. - PowerPoint PPT Presentation

Transcript of Probe Characterization

Page 1: Probe Characterization

Probe Characterization

Amber Precision Instruments, Inc. (API)

Jul. 23, 2014

Page 2: Probe Characterization

2

Our EMI Probes

EMI Probes:- Up to 6 GHz- Up to 20 GHz- Up to 40 GHz

Optional EMI Probes;Choose:- Size- Frequency range- Field Component

* EMI Hx 2 mm

Page 3: Probe Characterization

3

Characterization Structureup to 15 GHz

50 Ohms Microstrip Line (MSL)

Page 4: Probe Characterization

4

Characterization Structureup to 40 GHz

50 Ohms Air-Dielectric* Coplanar Waveguide (CPW)

* To avoid the effect of non-TEM waves on probe characterization at higher freqs (>15 GHz).

Page 5: Probe Characterization

5

Probe Characterization Setup

Page 6: Probe Characterization

6

Typical EMI Probe S21

1 2 3 4 5 6 7 8 9 10x 10

9

-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

* EMI Hx 2 mm: up to 10 GHz

Page 7: Probe Characterization

7

What Are the Specs?

108

109

1010-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

108

109

1010-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

20 dB/dec Line

Log freq: Low freq with 20 dB/dec slope

* EMI Hx 2 mm

Page 8: Probe Characterization

8

What Are the Specs?

High freq: How high the probe measures?

0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2x 10

10

-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

* EMI Hx 2 mm

Page 9: Probe Characterization

9

What Are the Specs?

108

109

1010

-100

-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Hx Field, = 0

Unwanted Field, = 90

Unwanted field: Decoupling of unwanted components

* EMI Hx 2 mm

Page 10: Probe Characterization

10

What Are the Specs?

108

109

1010-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

107

108

109

1010-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Low freq: How low the probe measures in a “practical” setup?

20 dB/dec slope60 dB dynamic range

4 MHz

Page 11: Probe Characterization

11

High Freq Probe

109

1010

-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Hx Field, = 0

Unwanted Field, = 90

* EMI Hx 1 mm L: up to 20 GHz

Page 12: Probe Characterization

12

Probe Factor

* EMI Hx 2 mm

108

109

1010

20

30

40

50

Frequency [Hz]

|PF|

[dB

(A/m

)/V]

Measured Probe FactorTheoretical Open-Circuit Probe Factor

Probe factor: Provide probe factor SmartScan imports probe factor SmartScan measures and calculates system factor

Page 13: Probe Characterization

13

Other Methods ofData Presentation

• How other people characterize their probes?

• How the characterization data is presented?

• Is their data presentation fair?

Page 14: Probe Characterization

14

Example* of 10 kHz (?) to 6 GHz Probe Data Presentation

Let’s analyze the data!Extract

* Aprel: http://media.wix.com/ugd/36d46a_46d2902bfdd24c25b401582e6a488370.pdf

Page 15: Probe Characterization

15

Is Freq Scale OK?

Why the freq axis of extracted data don’t

match?!

Let’s see!

Let’s analyze the data!

Page 16: Probe Characterization

16

Is Freq Scale OK?

0.99999 GHz 1.3 GHz 1.3 GHz 1.3 GHz 1.1 GHz

Why all ticks are equal?!

Well, not a big deal. Maybe just a typo!

Page 17: Probe Characterization

17

Lower Frequencies

Why all ticks are equal?!

Well, not a big deal. Maybe just a typo!

What happened to lower freqs?!

Is it extrapolated or really measured?

What is the measurementinstrument setup (IF BW,

span, no. of points, etc.) to measure down to 10 kHz?

Is it a “practical” setup?

Page 18: Probe Characterization

18

Plot It Differently

104

105

106

107

108

109

-120

-100

-80

-60

-40

Frequency [Hz]

|S21

| [dB

]

What if we plot it in a different way

(log freq)?

104

105

106

107

108

109

-120

-100

-80

-60

-40

Frequency [Hz]

|S21

| [dB

] 20 dB/dec LineThe probe is NOT purely inductiveat lower freqs(as claimed).

Is it probe issue?Or, measurement

issue?Or, simply not

extrapolated the right way?

Note: Simple extrapolation (even if it is done the right

way) does NOT mean probe can measure down to that

freq!

Page 19: Probe Characterization

19

How about 10 kHz to20 GHz (?) Probe?

Extract

Let’s analyze the data!

Page 20: Probe Characterization

20

Flat Response (?)

Let’s analyze the data!

The data is from -30 dB to -60 dB. Why the scale

is from 0 to -180 dB?!

Well, it looks better* this way! :-)

*“flat response acrossthe frequency range”

Page 21: Probe Characterization

21

Plot It Differently

0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2x 10

10

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

What if we plot it in a different way

(tight scale)?

Page 22: Probe Characterization

22

Plot It Differently

What if we plot it in a different way

(log freq)?

108

109

1010

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

108

109

1010

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

20 dB/dec Line

The lower freqs seemsto be fine for this prob.The deviation might be

just data extractionerror.

Page 23: Probe Characterization

23

108

109

1010

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

20 dB/dec Line

Higher Frequencies

How about the higher freqs?

There is about 20 dB notch @

13 GHz!

A deep notch is usuallya BIG red flag for

“unwanted field” coupling!

Page 24: Probe Characterization

24

Higher Frequencies

The question is:How to make sure this probemeasures the “desired field” up to 20 GHz (as claimed)?

The “unwanted field” decouplingdata of probe need to be provided.

This probe is suspected to work <10 GHz,

unless the rejection info is available!

Page 25: Probe Characterization

25

10 kHz to 40 GHz Probe

Let’s analyze the data!

Extract

Page 26: Probe Characterization

26

Flat Response (?)

Let’s analyze the data!

Same question:The data is from -35 dB to -55 dB. Why the scale

is from 0 to -120 dB?!

Page 27: Probe Characterization

27

Plot It DifferentlySingle Probe Solution (?)

What if we plot it in a different way(tight scale and

log freq)?

1010.2

1010.3

1010.4

1010.5

1010.6

-50

-45

-40

-35

Frequency [Hz]

|S21

| [dB

]

This info is NOT sufficient to evaluate the probe up

to 40 GHz!

The “unwanted field” decouplingdata of probe need to be provided.

What does happen to lower freqs?Does this probe have sufficient

sensitivity to measure down to 10 kHz?!

Page 28: Probe Characterization

28

Another Example* of Data Presentation

* Detectus AB: http://www.detectus.se/products_nearfieldprobes.html

Again, the data is from -30 dB to -60 dB. Why the scale is

from 0 to -60 dB?!

Without unwanted field data, there is no sufficient info to

evaluate the probe.

Log freq scale is required forlow freq evaluation.

Page 29: Probe Characterization

29

What to Look for inProbe Characterization Data?

• Unwanted field rejection data need to be provided to determine higher usable freq.

• Is lower usable freq achievable in a “practical” setup?

• Is the probe purely inductive/capacitive at lower freqs?

• Is the characterization structure (MSL, CPW, etc.) suitable for freq of interest?

• What is the measurement setup/condition?

• Is the scale proportional to data range?

Page 30: Probe Characterization

30

Questions?

Contact us: [email protected]

www.amberpi.com