1C Series
C SE
RIESC S E R I E S
Digital pressure calibrator
! Ranges up to 1000 bar/15,000 psi
! Accuracy all ranges 0.025% FS
! Self contained in a carrying case with detachable lid
! 25 units of pressure
! In-built pressure generation
! Data storage with RS232 interface
2C Series
C S E R I E SDigital pressure calibrator
The Pressurements C Series is a microprocessor based pressurecalibrator with built-in pressure generation.
This self-contained instrument includes fine pressure control, deviceenergising, output measurement and data storage capabilities. Therugged design incorporates a carrying case and detachable lid forworkshop or field use. Ergonomic layout ensures that test leads canbe connected without compromising visibility of the large, dualparameter display.
A single piece elastomeric keypad, which forms the user interface,combines a high quality tactile feel with exceptional service life. ThemA step and ramp outputs and built-in continuity tester extend thecapabilities to include the commissioning and maintenance of controlloops.
A highly accurate and easy to use calibrator is only part of the solutionfor improving overall data quality and working efficiency. The C series,with data-storage and RS232 will communicate through Terminal TM orHyper TerminalTM to assist the production of spread sheets andcertificates to reduce calibration times and eliminate data recordingerrors.
Improved Performance
The C series combines practical design with state of the artperformance.
Accuracy
0.025% FS for all pressures
Ranges
From –1 to 700 bar
Integral pneumatic pressure source
0 to 20 bar
Vacuum source
0 to – 800 mbar (Optional)
Integral hydraulic pressure source
0 to 700 bar
Measure
Pressure, mA, V, switch state(open/closed) & ambient temp.
Source
Pressure, mA step, mA ramp & mA value
Energising supplies
10V and 24Vdc
Simplified Operation
The key to the simple operation of the C Series is theTask Menu. Specific operating modes such as P-I, switch test and leaktest are configured at the touch of a button by simple menu selection.This innovative approach ensures correct set up of the calibrator forthe job at hand and leaves the operator free to concentrate on hiswork.
The operating system can work in several languages and in 25different pressure units, including a special user defined unit. Processfeatures such as max/min, filter and % span facilitate themeasurement of fluctuating signals and flow parameters.
3C Series
Set up key to determinedefault working parameters
Automatic zero correction forany input or output parameter
Input/output parameter selectionkeys with dedicated task key forinstant access to task menu
Documenting keys to store and recallscreen snapshots, multi-channel datalog and calibration results
Multi-lingual software
Input and output connectors
Tactile elastomeric keypad
Function keys used inresponse to display prompts
5C Series
Standard models available
AIR OPERATED MODEL
Model Range
CA020P* -1 to 20 bar/-14.5 to 300 psi
CA070 0 to 70 bar/0 to 1000 psi
CA200 0 to 200 bar/0 to 3000 psi
CA350 0 to 350 bar/0 to 5000 psi
*Includes handpump. All other gas operated models require an externalpressure source.
OIL OPERATED MODELS
Model Range
C020 0 to 20 bar/0 to 300 psi
C200 0 to 200 bar/0 to 3000 psi
C350 0 to 350 bar/0 to 5000 psi
C700 0 to 700 bar/0 to 10000 psi
C1000P 0 to 1000 bar/0 to 15000 psi
WATER OPERATED MODELS
Model Range
CW020 0 to 20 bar/0 to 300 psi
CW200 0 to 200 bar/0 to 3000 psi
CW350 0 to 350 bar/0 to 5000 psi
CW700 0 to 700 bar/0 to 10000 psi
CW1000P 0 to 1000 bar/0 to 15000 psi
Other pressure ranges available
Options Available• Dual Pressure ranges available for hydraulic models.
• Built in handpump on all hydraulic models allowing large volumes to be generated quickly. Standard on models C & CW1000P.
• Built in vacuum handpump (0 to -800 mbar), air operated models only.
Note: - Oil operated model shown
6C Series
Accuracy
0.025% (non-linearity, hysteresis, repeatability)
Temperature Effects
±0.004% reading/0C averaged over – 100C to 400C w.r.t. 200C
Overpressure
Safe to 2 x FS (except –1 to 20 bar range safe to 35 bar max, 700 bar range safeto 1000 bar max)
* Temperature coefficient ±0.0075% reading/°C w.r.t. 20°C.
* Temperature coefficient ±0.0075% reading/°C w.r.t. 20°C.
Pressure units
25 scale units plus one user defined.
mA step
Continuous cycle at 10 sec intervals.
mA ramp
Continuous cycle with configurable end values and 60 second travel time.
Data log
Multi-parameter with internal memory for 10,000 values.Variable sample period or log on key press.
Leak Test
Check for pressure leaks prior to calibration or during routine maintenance.
Snapshot
Paperless notepad. Stores up to 20 complete displays.
Computer interface
RS 232.
Process functions
Tare, max/min, filter, flow, % span.
Language
English, French, German, Italian, Portuguese and Spanish.
Power management
Auto power OFF, auto backlight OFF, battery low indicator and status on key press
Panel
60 x 60 mm graphic LCD with backlight.
Readout
±99999 capability, 2 readings per second.
SPECIFICATIONS
ELECTRICAL INPUTS
Input Range Accuracy Resolution Remarks
Voltage* ±50Vd.c. ±0.05%Rdg. ±0.004%F.S. 100µV max. Autoranging, >10MΩCurrent* ±55mA ±0.05%Rdg. ±0.004%F.S. 0.001mA 10O, 50V max.Temperature -10° to 40°C ±1°C 1°C Local ambientSwitch Open/Closed 5mA whetting
ELECTRICAL OUTPUTS
Function mA Output
4 to 20mA linear 4 8 12 16 200 to 20mA linear 0 5 10 15 204 to 20mA flow 4 5 8 13 200 to 20mA flow 0 1.25 5 11.25 204 to 20mA valve 3.8 4 4.2 12 19 20 21
DISPLAY
ENVIRONMENTAL
Temperature
Operating –10 to 500CCalibrated –10 to 400C
Humidity
0-90% non-condensing
Conformity
EN61010, EN50081-1, EN50082-1, CE marked
Including lid: 47 x 32 x20cms (LxDxH)Weight: 12kg
1/8” ¼” 3/8” ½” BSP or NPT female.
Air:- Hand pump or external source (models above 20 bar)Hydraulic:- Ram screwOptional:- Fitted hydraulic handpump Fitted vacuum handpump (0 to –800 mbar)
Rechargeable Ni-cad battery pack, supplied with a universal input charger which also allowsthe instrument to be used while charging.
Other productsand services
Deadweight Testers
A full of primary standards for pressure from –1 to 4000 bar,
also absolute and differential pressure.
Comparison Pumps
Used to check pressure measuring instruments against master test
gauges or digital indicators.
Calibration
Calibrations traceable to the National Physical Laboratory (UK),
National Institute of Science and Technology (US) as standard.
Optional NAMAS certification available.
Training
The “Pressure School” will train delegates in all aspects of
pressure measurement and calibration with courses tailored for
specific requirements.
C S E R I E SStandard Specification
Output Range Accuracy Resolution Remarks
Voltage* 10Vd.c. ±0.1% Max. load 10mA24Vd.c. ±5% Max. load 26mA
Current* 0 to 24mA ±0.05%Rdg. ±0.01%F.S. 0.001mA
SPECIAL FEATURES
Due to Pressurements’ process of continuous improvement, specifications are subject to change without notice.
Represented by
Pressurements LimitedUnit 22, Apex Business Centre,Boscombe Road, Dunstable, Beds LU5 4SBTel: +44 (0) 1582 471535Fax: +44(0) 1582 601185
DIMENSIONS
PRESSURE CONNECTIONS
PRESSURE GENERATION
POWER SUPPLY