PRESSURE REGULATORS, FLOW METERS, QUICK COUPLING PROBES AND ACCESSORIES
TABLE OF CONTENTS
3
13485QUALITYASSURANCE
ISO0029
INTRODUCTION P.004
- CEODEUX MEDITEC solutions P.004
- Quality first... P.004
- The main medical gases P.005
PRODUCT RANGE OVERVIEW P.006
PRESSURE REGULATORS | STANDARD RESPIRATORY THERAPIES P.008
- ALPINOX medical oxygen regulator P.008
- MINIOX medical oxygen regulator P.010
PRESSURE REGULATORS | SPECIAL THERAPIES (ASSISTED) P.012
- Series BSC 280/380 P.012
- Series BSC 290/390 P.014
- Series DI 135 P.016
IN-HOSPITAL THERAPY EQUIPMENT P.018
- M730 | Alpi-flow Flowmeter with calibrated orifices P.018
- Alpinox Column Flowmeter P.019
- ALPI-VAC | vaccum regulator P.020
- Alpi-vent | suction units P.021
ACCESORIES P.022
- Respiratoty therapy accessories P.022
- Suction accessories P.023
All Rotarex regulators are produced in Europe in accordance with international standards (ISO; CGA....) and are guaran-teed to provide safe and reliable performance in operation. All locations are ISO 13485.
CEODEUX MEDITEC SOLUTIONS
QUALITY FIRST...
4 INTRODUCTION
ALL RIGHTS OF CHANGE RESERVED
Patient therapy and safety is our main focus in the design, installation, commissioning and operation of medical gas equipment
QUANTITY OF SUPPLYRegulators are designed in such a way sufficient medical gas capacity can be ensured in order to provide the required pressures and flows of gases for the intended number of patients to be treated at any time.
IDENTITY OF SUPPLYAll our products are designed, produced and controlled for a specific medical gas. All points to which the user or technical staff can connect are gas specified. There is no risk of misconnection.
QUALITY OF SUPPLYOur regulators maintain quality of supplied medical gases thanks to appro-priate selection of material, components cleanliness and internal filtration equipment.Products designs ensure safe supply by minimizing ignition risks, by avoid-ing patient poisoning risks and by guaranteeing very good performances for accurate therapies.
CEODEUX MEDITEC products are designed, validated, produced and con-trolled during their whole lifetime according to a very stringent quality system which exceeds requirements of ISO 13485 standard.
DESIGNAll our products comply with local and international standards. Some reference standards:
Product design is done through a risk assessment process and is combining proven technologies and innovation to reach the best performance levels.
VALIDATIONProduct validation process guarantees that products are developed and produced to withstand the complete lifecycle (installation, commission-ing, normal service and maintenance)
PRODUCTIONAll products are produced in a controlled environment in order to maintain cleanlinessAn individual traceability is ensured with individual serial numberProducts are individually controlled and data are recorded for a perfect product management.All products are cleaned according to ISO 15001
Products design and validation
CGA E-7 Medical gas pressure regulators Flowmeters and orifice flow selectors
ISO 10524-1 Pressure regulators for use with medical gases - Part 1: Pressure regulators and pressure regulators with flow-metering devices
ISO 15002 Flowmeters for use with medical gases
Materials selection and cleanliness ISO 15001:2011 Anesthetic and respiratory equipment. - Compatibility with oxygen
5
THE MAIN MEDICAL GASES
5INTRODUCTION
ALL RIGHTS OF CHANGE RESERVED
Symbol Name Application Source Type
O2 OxygenOxygen is one of the most extensively used gases for respiratory therapy and life-support and is additionally used in anesthetic procedures.
- Compressed gas cylinder - Liquid tank - Concentrator
Air AirMedical air is mainly used in respiratory therapy as a power source for patient ventilators, and for blending with oxygen. It is also used as the driving gas for nebulized drugs and chemotherapy agents.
- Compressor - Compressed gas cylinder
N2O Nitrous OxideNitrous oxide is used for anesthetic and analgesic purposes, being mixed with air, oxygen, and nebulized agents.
- Liquefied gas cylinder
CO2 Carbon DioxideCarbon dioxide is used less commonly now as a respiratory stimulant, and for insufflation during surgery.
- Liquefied gas cylinder
Air Surgical AirSurgical air (of medical air quality) is also used, at a higher pressure, to power a variety of surgical tools and other devices such as tourniquets.
- Compressor - Compressed gas cylinder
N2 NitrogenAs an alternative, nitrogen can be used for the same purpose as surgical air.
- Compressed gas cylinder
6
ALL RIGHTS OF CHANGE RESERVED
ALPINOX
ALPINOX MEDICAL OXYGEN P. 08
REGULATOR WITH FLOW SELECTORALPINOX MEDICAL OXYGEN P. 08
REGULATOR WITH QUICK CONNECTORALPINOX PIN INDEX MEDICAL P. 08
OXYGEN REGULATOR WITH FLOW SELECTOR
MINIOX
MINIOX QUICK CONNECTOR P. 10 MINIOX COLUMN FLOWMETER P. 10
PRESSURE REGULATORS | SPECIAL THERAPIES (ASSISTED)
BSC 280/380 P. 12
SINGLE STAGE CARTRIDGE REGULATORBSC 290/390 P. 14
SINGLE STAGE CARTRIDGE REGULATOR WITH HOSE BARB
DI 235 P. 16
NITRIC OXIDE APPLICATIONS DOUBLE STAGE BELLOW REGULATOR
PRESSURE REGULATORS | STANDARD RESPIRATORY THERAPIES
PRODUCT RANGE OVERVIEW
7
ALL RIGHTS OF CHANGE RESERVED
IN HOSPITAL THERAPY EQUIPMENT
FLOW METERING DEVICES
ALPI-FLOW FLOWMETER WITH CALIBRATED ORIFICES P. 18 COLUMN FLOWMETER P. 19
ACCESORIES
RESPIRATOTY THERAPY ACCESSORIES
REMOVABLE HOSE BARB P. 22 HUMIDIFIER P. 22 ALPI-CLICK P. 22
MEDICAL PROBES
MEDICAL CONNECTORS (SINGLE AND DOUBLE)
SUCTION EQUIPMENT
ALPI-VAC VACCUM REGULATOR P. 20 ALPI-VENT SUCTION UNITS P. 21
SUCTION ACCESSORIES
SMALL COLLECTION JAR P. 23 LARGE COLLECTION CONTAINERS P. 23 OUTLET HOSE CONNECTION P. 23
PRODUCT RANGE OVERVIEW
8
ALL RIGHTS OF CHANGE RESERVED
KEY FEATURES
- Single stage piston regulator ensuring high life cycle - Preset integrated pressure relief valve - Flowmeter with stop at 0 - Easy to clean - High precision fl ow from 0.25 lpm to 25 lpm - Continuous fl ow from 0 to maximal fl ow - Integrated safety relief device protected and none accessible
- Ergonomic handwheel and modern design. Red dot design award
- Easily cleanable - Rubber protection on the gauge - Plastic protection on the regulator
- Ergonomic plastic hand-wheel on the inlet connection - Position 25 lpm intended for the nebulization - 20μm Filter inside the seat regulator - Design free of fl uorinated polymers whose combustion products are toxic
- Individual number for traceability
OPTIONS
- Lateral marking
- Cylinder regulator for hospital and home care oxygen therapy
- A new generation of medical regulator with high fl ow precision and continuous fl ow between settings.
- In compliance with EN ISO 10524-1
ALPINOX MEDICAL OXYGEN REGULATOR
ALPINOX
STANDARD ALPINOX
ALPINOX WITH QUICK CONNECTOR
ALPINOX PIN INDEX
9
ALL RIGHTS OF CHANGE RESERVED
SPECIFICATIONS
Gas Medical oxygen, Air
Max Inlet pressure
200 bar
Gauge 0-315 bar, 0-4500 psi
Filter 20 micron, sintered bronze
Flow range From 0-2Lpm to 0-25Lpm
Body material white coated brass
O-ring material EPDM
Seat material Nylon
Regulatory status Complies with medical devices directive 93/42/EECand EN ISO 10524
CE marking CE0029
ALPINOX
PRODUCT CONFIGURATOR
Gas Working Pressure Cylinder Connection Outlet Pressure Flow Range Pressure outlet Flow outlet
M730 O2 200 NF 4.5 2L N HB6
O2 O2 200 bar 200 AFNOR NF 3.5 bar 3.5 0-2Lpm 2L None N Hose Barb Ø6mm - Fixed
HB6
Air AIR 2000 psi 2000 DIN D 4.5 bar 4.5 0-6Lpm 6L AFNOR NF Hose Barb Ø6mm - Removable 9/16”
9/16
N2O/O2 N2O 3000 psi 3000 BS B 0-15Lpm 15L DIN DIN 0-15Lpm M12
UNI U 0-25Lpm 25L BS BS 0-25Lpm 1/4
Pin-Index P SS SS
10
ALL RIGHTS OF CHANGE RESERVED
KEY FEATURES
Safety first - Smart choice of materials in conformance with ISO 15001 and the most restrictive additional requirements
- Design free of fluorinated polymers - Fitted with a safety relief valve in case of over pressure
- Even in case of failure, no high pressure exposure
High performances - Better outlet pressure and flow stability - Inverted seat regulator for a longer lifetime - Design resistant to pressure shocks & adiabatic compression
Lower operating costs - Compact and lightweight design for an easy handling and storage
- Suitable for all types of medical gases - Precise flow and improved stability for better gas consumption management
REGULATORY
- ISO 10524-1 - CGA E-7
MINIOX MEDICAL OXYGEN REGULATOR
- Medical pressure regulator for hospital and homecare respiratory therapy.
- Used to reduce cylinder pressure to a lower pressure suitable for medical equipment and/or for gas delivery direct to patient.
MINIOX
MINIOX WITH QUICK CONNECTOR MINIOX WITH COLUMN FLOWMETER
11
ALL RIGHTS OF CHANGE RESERVED
SPECIFICATIONS
Gas O2 / Air / N2O
Max. Inlet Pressure 200 bar
Gauge 0-315 bar
Temperaturerange
-20°C to +60°C
Body material Chrome plated brass
Diaphragm EPDM
O-rings EPDM
Seat PA 6.6
Filter 20 μm sinteredbronze
Outlet pressure 3.5 bar/4.5 bar
Max . Flow rate 40 Lpm
Weight 0,640 kg
Inlet Connection According to national standards
Outlet Connection According to national standards
PRODUCT DESCRIPTION
COLUMN FLOWMETER0-18 Lpm Outlet according to national standards (Available versions: single outlet & double outlet)
HIGH PRESSURE GAUGE 0-315 bar Cylinder content visual indication
INTEGRATED RELIEF VALVE Protect regulator against overpressure risks
CYLINDER VALVE CONNECTIONAccording to national standards
PRESSURE REGULATORDiaphragm technology
1
2
3
4
5
MINIOX
3
1 2
5
4
Gas Inlet pressure Outlet pressure Inlet connection Outlet dimensions Flow rate
M743 O2 200 4,5 DIN G1/4 15Oxygen O2 200 bar 200 4,5 bar 4,5 G¼ ” Female O2 G¼ ” G¼ 0-15 Lpm
Air AIR 2000 psi 2000 3,5 bar 3,5 NF 29-650 NF M12x125 M12 0-25 Lpm
DIN 477-1 DIN 9⁄16” UNF 9⁄16
BS 341-3 BS
NEN 3268 NEN
UNI 11144 UNI
CGA V1 CGA
Pin Index PI
PRODUCT CONFIGURATOR
Gas Inlet pressure Outlet pressure Outlet types Inlet connection Outlet dimensions
M743QC O2 200 4,5 SQC DIN DINOxygen O2 200 bar 200 4,5 bar 4,5 Single quick connector SQC G¼ ” Female O2 DIN 13260 DIN
3,5 bar 3,5 Dual quick connector DQC NF 29-650 NF AFNOR NF S90-116 AF
DIN 477-1 DIN SS8752430 SS
BS 341-3 BS
NEN 3268 NEN
UNI 11144 UNI
CGA V1 CGA
Pin Index PI
Flowmeter version
Quick connector version
ALL RIGHTS OF CHANGE RESERVED
12
ALL RIGHTS OF CHANGE RESERVED
Right inlet Left inlet
SERIES BSC 280/380 | SINGLE STAGE CARTRIDGE REGULATOR
INCLUDED FUNCTIONS
- Inlet port - Outlet port - Pressure regulator - Safety relief valve - High pressure measurement - Low pressure measurement
REGULATORY
- Medical certifi ed according to MDD 93/42/EEC amended 2007/47/EC
- Product complies with EN ISO 10524-1
- Enables gas supply to therapy machines
- Especially designed for accurate NO mixtures therapies
- Can be fi tted with pressure sensor for proper therapies parameters management
- 6 ports for multiple possible confi gurations and appropriate product integration thanks to compactness
SAFETY FIRST - Fitted with a safety relief valve in case of over pressure which can be collected
- Dedicated HP and LP pressure gauges for a permanent indication - Compatible materials with Oxygen and medical applications - No gas contamination due to a metal-metal sealing
HIGH PERFORMANCE - Proven Cartridge Seat Technology - Better outlet pressure and fl ow stability - Smart choice of material in conformance with EN ISO 15001
LOWER OPERATING COSTS - Compact and lightweight design for an easy handling and storage - Longer product life thanks to less impingement on the diaphragm - Easy maintenance operations thanks to Cartridge Technology
KEY ADVANTAGES
Pressure gauges
Outlet (LP)
Safety valve
Inlet (HP)
Plugged Plugged
HIGH PRESSURE REGULATOR
ALL RIGHTS OF CHANGE RESERVED
13
SPECIFICATIONS
ALL RIGHTS OF CHANGE RESERVED
Body Chrome plated brass
Seat seal Polyamide
O-ring EPDM
Diaphragm Hastelloy®
Inlet Connection Cylinder connection
Outlet Connection Hose Barb or Medical connector
Leak rate 10-8 mbar l/s He
Weight ± 1.4 kg
Inlet pressure 200 / 300 bar2900 / 4350 psi
Outlet pressure 4 / 8 bar58 / 116 psi
Nominal Flow 7 / 25 Nm3/h (N2)
Temperature range
-40°C to +60°C
PRODUCT CONFIGURATOR
Material Inlet Pressure Port Configuration Outlet pressure Inlet Connection Outlet connection Pressure Sensor Gas
BSC I 280 R 5 DIN NIST B NO
Stainless steel
I 200 bar 280 Right inlet R 5 bar 5 DIN477-1 DIN G¼” F G14F without pressure sensor
A NO (max 1000ppm)/N2
NO
Left inlet L ISO5145 ISO ¼”NPT F N14F with HP pressure sensors
B
CGA V1 CGA with LP pressure sensor
C
with HP & LP pressure sensors
D
HIGH PRESSURE REGULATOR
ALL RIGHTS OF CHANGE RESERVED
14
ALL RIGHTS OF CHANGE RESERVED
Right inlet Left inlet
SERIES BSC 290/390 | HP CYLINDER PRESSURE REGULATOR
INCLUDED FUNCTIONS
- Inlet port - Outlet port - Pressure regulator - Safety relief valve - High pressure measurement - Low pressure measurement
REGULATORY
- Medical certified according to MDD 93/42/EEC amended 2007/47/EC
- Product complies with EN ISO 10524-1
- Enables gas supply to therapy machines
- Can be fitted with pressure sensor for proper therapies parameters management
- 6 ports bodies for multiple possible configurations and appropriate product integration
SAFETY FIRST - Fitted with a safety relief valve in case of over pressure which can be collected
- Dedicated HP and LP pressure gauges for a permanent indication - Compatible materials with Oxygen and medical applications - No gas contamination due to a metal-metal sealing
HIGH PERFORMANCE - Proven Cartridge Seat Technology - Better outlet pressure and flow stability - Smart choice of material in conformance with EN ISO 15001
LOWER OPERATING COSTS - Compact and lightweight design for an easy handling and storage - Longer product life thanks to less impingement on the diaphragm - Easy maintenance operations thanks to Cartridge Technology
KEY ADVANTAGES
Pressure gauges
Outlet (LP)
Safety valve
Inlet (HP)
Plugged Plugged
HIGH PRESSURE REGULATOR
ALL RIGHTS OF CHANGE RESERVED
15
SPECIFICATIONS
ALL RIGHTS OF CHANGE RESERVED
Body Chrome plated brass
Seat seal Polyamide
O-ring EPDM
Diaphragm Hastelloy®
Inlet Connection Cylinder connection
Outlet Connection Hose barb or medical connector
Leak rate 10-8 mbar l/s He
Weight ± 1.4 kg
Inlet pressure 200 / 300 bar2900 / 4350 psi
Outlet pressure 4 / 8 bar58 / 116 psi
Nominal Flow 7 / 25 Nm3/h (N2)
Temperature range
-40°C to +60°C
PRODUCT CONFIGURATOR
Material Inlet Pressure Port Configuration Outlet pressure Inlet Connection Outlet connection Pressure sensor Gas Type
BSC L 290 R 4 NF NIST B O2
Chrome plated brass
L 200 bar 2900 psi
290 Right inlet R 4 bar 58 psi
4 G¼” Female O2 Hose Barb 6 mm
HB without pressure sensor
A O2 O2
300 bar 4350 psi
390 Left inlet L 8 bar 116 psi
8 NF 29-650 NF AFNOR Medical Connector
NF with HP pressure sensors
B Air AIR
10 bar 145 psi
10 DIN 477-1 DIN DIN Medical Connector
DIN with LP pressure sensor
C N2O N2O
BS 341-3 BS NIST Medical Connector
NIST with HP & LP pressure sensors
D CO2 CO2
NEN 3268 NEN N2 N2
UNI 11144 UNI
CGA V1 CGA
HIGH PRESSURE REGULATOR
ALL RIGHTS OF CHANGE RESERVED
16 PRESSURE REGULATORS | SPECIAL THERAPIES (ASSISTED)
APPLICATIONS
- Designed for cylinder regulator applications. - Ideally suited for special NO/N2 mixtures (N2 with 1000 ppm NO) applications requiring a very stable outlet pressure together with a very sensitive set up of this outlet pressure.
KEY FEATURES
- The DI 235 regulator is based on the S 20 proven bellow technology.
- Accurate pressure control for reliable service and guarantees a stable outlet pressure due to the combination of the diaphragm and bellow technology.
- Compact and lightweight design. - Can also be equipped with a needle or shut off valve at the outlet.
Compact and lightweight design
1 inlet / 2 outlets Inlet/outlet pressure gauges 2 safety relief valves
Special requirements on request
- For NO/N2 mixtures - Diaphragm/bellow dual stage
- Purity up to 6.0 - Max inlet pressure: 300 bar (4350 psi)
- Preset outlet pressure: 2,5/5 bar (36/72 psi)
SERIES DI 235 | NITRIC OXIDE DOUBLE STAGE BELLOW REGULATOR
ALL RIGHTS OF CHANGE RESERVED
0
14,5
29
43,5
58
72,5
87
101,5
116
0,0
1,0
2,0
3,0
4,0
5,0
6,0
17
SPECIFICATIONS
PRESSURE REGULATORS | SPECIAL THERAPIES (ASSISTED)
PRODUCT CONFIGURATIONS
Body Material Outlet Pressure End Connections Gauges
D I 235 5 G 1
Stainless steel I 2,5 bar36 psi
2,5 In: 16 x 1.336 Out: G 3⁄8
G Without 0
5 bar72 psi
5 ¼ NPT - ¼ NPT N With 1
Female ports 16 x 1.336 (inlet) - G 3⁄8 (outlet)or ¼ NPT (inlet/outlet)
Seat seal PTFE (1st stage) EPDM (2nd stage)
O-ring FPM
Diaphragm Hastelloy
Bellow AISI 316L (SS version)
Weight ± 1,4 kg± 3.0 lbs
Leak rate 10-8 mbar ℓ/s He
Temperature range - 20°C to + 60°C- 4°F to + 140°F
Gauges High and low pressure (M10 x 1 or ¼ NPT)
Max inlet pressure 300 bar4350 psi
Outlet pressure 2,5/5 bar 36/72 psi
Nominal Flow 1,8/3 Nm3/h (N2)
FLOW CURVES
OU
TLET
PR
ESSU
RE-
bar
an
d p
si
PRESSUREbar
PRESSUREpsi
Blue line P1 = 300 barRed line P1 = 150 barGreen line P1 = 75 bar
FLOW CURVE DI 235 REGULATORSET: 5 bar / 72.5 psi
2.5 bar / 36.25 psi
Nm3/h 1 2 3 4 5 6
Nlpm 17 33 50 66 83 100
Slpm 18 35 53 70 88 105
SCFM 0.6 1.2 1.9 2.5 3.1 3.7
FLOW RATE in Nitrogen (N2)
360°
TO BE MOUNTED ON A
RAIL WITH FLEX HOSE
(ACC. TO EN ISO 5359)
18
ALL RIGHTS OF CHANGE RESERVED
M730 | ALPI-FLOW FLOWMETER WITH CALIBRATED ORIFICES
KEY FEATURES
- Continuous fl ow between settings - Stop at 0 - Inlet pressure : 4 bar 0/+1 bar - Flow precision acc. EN ISO 15002 - Flow range: 0-15 lpm (12 positions)
NEW GENERATION ALPINOX FLOWMETER
MEDICAL FLOWMETER
- High fl ow precision from 0.1 lpm to 25 lpm - Continuous fl ow from 0 to maximal fl ow - Ergonomic and modern design - Easily cleanable - Plastic protection on the fl owmeter - Position 25 lpm intended for the nebulization - Swiveling design 360° for optimal readability - Design free of fl uorinated polymers whose combustion products are toxic
- Flow-meter for hospital and ambulance oxygen therapy
- New generation of medical fl ow-meter with high fl ow precision and continuous fl ow between settings. The fl ow-meter is in compliance with the EN ISO 15002
TO BE CONNECTED ON CEODEUX MEDITEC TERMINAL UNITS
IN-HOSPITAL THERAPY EQUIPMENT
SPECIFICATIONS
Inlet Connection Medical connector acc. to national standards
Flow Outlet M12x1.25 M.-9⁄16”UNF M. -G¼ ” M.
Dimensions 136x33x82mm
Weight 0.230 Kg
Inlet pressure 6 bar max
Temperature range
-20°C to +60°C
Flow Range 0-15 Lpm
Accuracy ± 10 % read value or ± 0.5 Lpm if greater
Body Chrome plated
O-ring EPDM
Measuring tube Polycarbonate
PRODUCT CONFIGURATOR
Gas Inlet Pressure Flow range Inlet Connection Flow outlet
M730-FM O2 3.5 2L NF 9/16
O2 O2 3.5 bar 3.5 0-2Lpm 2L AFNOR NF Hose Barb Ø6mm - Removable 9⁄16”
9/16
Air AIR 4.5 bar 4.5 0-6Lpm 6L DIN DIN Hose Barb Ø6mm - Removable
NF
N2O N2O 0-15Lpm 15L BS BS M12x125 M12
0-25Lpm 25L SS SS Hose Barb Ø6mm - Removable G¼ ”
1/4
TO BE MOUNTED ON A
RAIL WITH FLEX HOSE
(ACC. TO EN ISO 5359)
TO BE CONNECTED ON CEODEUX MEDITEC TERMINAL UNITS
19
ALL RIGHTS OF CHANGE RESERVED
IN-HOSPITAL THERAPY EQUIPMENT
KEY FEATURES
- Swiveling design 360° for an optimal orientation - Gas specifi c marking and color - Available in twin units versions for two independent suppliers from one single source.
REGULATORY
- Medical CE certifi ed according to MDD 93/42/EEC amended 2007/47/EC
- Product complies with EN ISO 15002
ALPINOX COLUMN FLOWMETER
- Medical Column Flowmeter for Oxygen and others medical gases (Air, N2, CO2, N2O)
- Combination of a precise needle valve and an accurate measuring device for a proper dosage of medical gases
SPECIFICATIONS
Inlet Connection Medical connector acc. to national standards
Flow Outlet M12x1.25 M.-9⁄16”UNF M. -G¼ ” M.
Dimensions 136x33x82mm
Weight 0.230 Kg
Inlet pressure 6 bar max
Temperature range
-20°C to +60°C
Flow Range 0-15 Lpm
Accuracy ± 10 % read value or ± 0.5 Lpm if greater
Body Chrome plated
O-ring EPDM
Measuring tube Polycarbonate
Removable Hose barb
M12x1.25 F. -9⁄16”UNF F.-G¼ ” F.
Humidifi ers M12x1.25 F. -9⁄16”UNF F.-G¼ ” F.200mL / 300mL / 500mL
OPTIONS
PRODUCT CONFIGURATOR
Gas Inlet Pressure Flow range Inlet Connection Flow outlet
CFM O2 3.5 15L NF 9/16
O2 O2 3.5 bar 3.5 0-15Lpm 15L AFNOR NF Hose Barb Ø6mm - Removable 9⁄16”
9/16
Air AIR 4.5 bar 4.5 DIN DIN Hose Barb Ø6mm - Removable
NF
BS BS M12x125 M12
SS SS Hose Barb Ø6mm - Removable G¼ ”
1/4
ALL RIGHTS OF CHANGE RESERVED
- Medical vaccum regulator (VAC)
- Regulator suitable for all medical suction applications in hospitals and nursing homes
- Insure safe, quick and reliable use thanks to performant technology and easy manipulation
ALPI-VAC | VACCUM REGULATOR
SPECIFICATIONS
Sizes (LxWxH) 103x92x170 mm
Weight ALPI-VAC 250: 0.42 Kg ALPI-VAC 600/1000: 0.37 Kg
Maximum suction rate
ALPI-VAC 250: 50 L/min. ±5 L/min. at -220 mbar ALPI-VAC 600: 72 L/min. ±5 L/min. at -550 mbar ALPI-VAC 1000: 115 L/min. ±5 L/min. at -950 mbar
Maximum adjustable depression
ALPI-VAC 250: -220 mbar ±20 mbarALPI-VAC 600: -550 mbar ±20 mbarALPI-VAC 1000: -950 mbar
Valve protecting against excessive
negative pressures
ALPI-VAC 250/600: preset with anti-clogging system ALPI-VAC 1000: N/A
Maximum supply depression
-950 mbar
Suction adjusting system
With membrane
Standard vacuum supply connection
ISO G. ¼” M
Outlet connection ISO G. ½” M
I/0 switch Quick push button type
SUNCTION EQUIPMENT20
KEY FEATURES
The ALPI-VAC continuous suction vacuum regulator is particularly suitable for all medical suction applications in hospitals and nursing homes.
The device is made of a strong techno-polymer body, with a quick I/0 switch-button, a suction adjustment knob and a control vacuum gauge with three possible end-of-scale choices: -250 mbar, -600 mbar and -1000 mbar. A protective silicone cover prevents gauge damages caused by possible shocks received during transport or use.
The ALPI-VAC vacuum regulators are also designed, through a threaded connection, for connecting directly or indirectly to the normal collection containers for suctioned fluids.
The extreme rationality and simplicity, combined with the sophisticated technical execution, allow both operators and patients to appreciate the safety and reliability of this medical device.
ALL RIGHTS OF CHANGE RESERVED
21SUNCTION EQUIPMENT
ALPI-VENT | SUCTION UNITS
KEY FEATURES
The ALPI-VENT/500 and ALPI-VENT/1000 vacuum generating system units are particularly suitable in all hospital applications in which it is necessary to get values of vacuum up to -850 mbar and where a central facility of vacuum generation for suction is missing, often for technical choice.
As it is well known, the physical principle called “Venturi” allows to get depression from a fluid under pressure, which is obtained, in case of these devices, using medical compressed air coming from the centralized distribution plant.
Such devices are composed essentially of a brass chrome-plated body, a needle valve with vacuum adjustment knob and a vacuum gauge. The version ALPI-VENT/1000 fits also a quick I/0 button switch and a large gauge, while in the version ALPI-VENT/500 a protective silicone cover on vacuum gauge avoids damages caused by possible impacts during transport or use.
- Vaccum generating system unit (VAC)
- Up to 850 mbar - Suitable for all hospital applications where a central facility of vacuum generation for suction is missing
SPECIFICATIONS
Sizes (without collection jar)
(LxWxH)
ALPI-VENT/500: 88x70x92 mm ALPI-VENT/1000: 140x80x140 mm
Weight (without collection jar)
ALPI-VENT/500: 0.80 Kg ALPI-VENT/1000: 0.95 Kg
Max. suction range
ALPI-VENT/500: 25 L/min. ±2 L/min. ALPI-VENT/1000: 27 L/min. ±2 L/min.
Vacuum gauge scale
ALPI-VENT/500: 0 ÷ -1000 mbar ALPI-VENT/1000: 0 ÷ -1 bar
Suction range ALPI-VENT/500: -800 mbar ±25 mbar ALPI-VENT/1000: -850 mbar ±25 mbar
Standard supply pressure
2.7 ÷ 5.5 bar
Adjustment of suction level
needle valve
Air consumption at the max.
suction value
47 L/min. ±2 L/min.
Inlet connection ISO G. ¼” M.
Outlet connection
to the patient
ISO G. ½” M.
ALPI-VENT/500
ALPI-VENT/1000
22
ALL RIGHTS OF CHANGE RESERVED
ADVANTAGES
- Swiveling design 360° for optimal use. - Design free of fl uorinated polymers whose combustion products are toxic.
- Used to connect or disconnect two specifi c medical devices for anesthesia, ventilation, oxygen therapy or various medical therapies.
- Designed and approved in accordance with the following standards: NF S 90-116 , DIN 13260-2 , SS 8752430 , EN737-6 .
22
RESPIRATORY THERAPY ACCESSORIES
REMOVABLE HOSE BARB
HUMIDIFIER
ALPI-CLICK MEDICAL PROBES MEDICAL CONNECTORS (SINGLE AND DOUBLE)
ACCESORIES
- Devices used to connect or disconnect specifi c medical devices for anesthesia, ventilation, oxygen therapy or various medical therapies.
- Designed according to EN 13544-2 for ø 6mm tube.
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Gas Oxygen, Air, 50% N20 / 50% 02 Materials Brass chrome plated, EPDM, Nylon CE marking CE 0029
Inner volume CH/200: 210 mlTR/200: 120 mlMAK/300: 285 mlMAK/500: 355 ml
Sizes (LxWxH) CH/200: 55x67x200 mmTR/200: 60x75x190 mmMAK/300: 70x81x175 mmMAK/500: 70x81x207 mm
Weight CH/200: 0.11 KgTR/200: 0.13 KgMAK/300: 0.15 KgMAK/500: 0.16 Kg
Applicable pressure and fl ow
Pressure: 500 kPaFlow: 10 L/min.
Inlet connection ¼ ” ISO 3253 F3⁄8” ISO 3253 FM 12x1.25 F9⁄16” UNF F
Outlet connection Hose connector Ø 6÷9 mm
Thread M12x125G¼ ”9⁄16” UNF
Materials Brass chrome plated CE marking CE 0029
- Used to connect or disconnect specifi c medical devices for anesthesia, ventilation, oxygen therapy or various medical therapies.
- Designed and approved in accordance with the EN ISO 9170-1
ALL RIGHTS OF CHANGE RESERVED
23ACCESORIES
SUCTION ACCESSORIES
ADVANTAGES
- Connector for hose Ø 7.5 mm, ISO G. 1/2” F. inlet thread.
- Vacuum regulator outlet hose connection
- Use to connect ALPI-VAC (Vaccum regulator) and its equipment MAK (collection container)
SMALL COLLECTION JAR
LARGE COLLECTION CONTAINERS
OUTLET HOSE CONNECTION
ADVANTAGES
- The two models have capacities of 300 ml and 500 ml respectively in standard execution, or fi tting a special lid for housing an antibacterial fi lter.
- The jar and the lid are made in polycarbonate, with fi ttings in brass, with ergonomic screw nut in high-strength polycarbonate. All parts can be sterilized in an autoclave (121 °C - 15 min.)
- Small collection jar are used for suction of small volume of fl uids and have been designed and manufactured for applications of “high fl ow and high vacuum”
- Can also be use as a safety container to protect the equipment and the plant upstream of the section source
- Matched with ALPI-VAC vacuum regulators
ADVANTAGES
- These devices are available with a capacity of 1000 ml with screw cap, of 2000 ml with screw and pressure lid and 4000 ml with pressure lid.
- The jar and the lid are made in polycarbonate, with fi ttings in brass, with ergonomic screw nut in high-strength polycarbonate. All parts can be sterilized in an autoclave (121 °C - 15 min.).
- Large collection container for suctioned liquid for applications of “high fl ow” and “high vaccum”
- Can also be use as a safety container to protect the equipment and the plant upstream of the section source
- Matched with ALPI-VAC vacuum regulators
Inner volume MAK/300/300 a.b: 300 mlMAK/500/500 a.b: 500 ml
Sizes (LxWxH) MAK/300: 70x98x180 mmMAK/300 a.b: 70x98x210 mmMAK/500: 70x98x215 mmMAK/500 a.b: 70x98x245 mm
Weight MAK/300: 0.20 KgMAK/300 a.b: 0.24 KgMAK/500: 0.21 KgMAK/500 a.b: 0.25 Kg
Supply depression max -950 mbar - 5 min
Overfl ow valve Floating valve
Inlet connection ISO G ½ ” F.
Outlet connection Hose connector Ø 8.0 ÷ 9.2 mm
Inner volume MAK/1000: 1000 mlMAK/2000 with screw lid: 2000 mlMAK/2000 with pressure lid: 2000 mlMAK/4000: 4000 ml
Sizes (LxWxH) MAK/1000: 110x110x243 mmMAK/2000 with screw lid/ with pressure lid: 140x140x298 mmMAK/4000: 175x175x310 mm
Weight MAK/1000: 0.38 KgMAK/2000 with screw lid: 0.49 KgMAK/2000 with pressure lid: 0.49 KgMAK/4000: 0.99 Kg
Supply depr. max -950 mbar- 5 min
Overfl ow valve Floating valve
Inlet connection Hose connector Ø 8.0÷9.2 mm or ISO G 1/2” F (for MAK/1000 and MAK/2000 screw lid only)
Outlet connection Hose connector Ø 8.0÷9.2 mm
Antibacteria system also available for both model (see specifi cation table: a.b)
NOTES
24
ALL RIGHTS OF CHANGE RESERVED
NOTES
25
ALL RIGHTS OF CHANGE RESERVED
A FULL LINE OF GAS CONTROL SOLUTIONS
ROTAREX is helping engineers worldwide to get better gas results: from ultra high purity production and medical care facilities to industrial and LPG applications, as well as alternative energy vehicles, fire suppression, diving, aerospace, cryogenics, laboratory, petro-chemical and welding. ROTAREX applies over 90 years of know-how and experience to custom design, develop and manufacture the high performance valves, regulators and fittings to suit your needs, all in one hand. Discover the difference ROTAREX can make in your world.
COMPLETE SOLUTIONS FROM SOURCE TO PROCESS.
MEDITECCYLINDER VALVES EQUIPMENT FIRETEC AUTOMOTIVE LPG/SRG
AUSTRALIA ⁄ BRASIL ⁄ CHINA ⁄ FRANCE ⁄ INDIA ⁄ ITALY ⁄ JAPAN ⁄ LUXEMBOURG ⁄ MIDDLE EAST ⁄ RUSSIA ⁄ SINGAPORE ⁄ SPAIN ⁄ THAILAND ⁄ USA
ULTRA HIGH PURITY VALVES MEDICAL VALVES & EQUIPMENT INDUSTRIAL CYLINDER VALVES REFRIGERANT CYLINDER VALVES
PRESSURE REGULATORS SUPPLY & SWITCH OVER BOARDS LINE VALVES FITTINGS & ADAPTORS
FIXED INSTALLATION FIRE SYSTEMS
OBJECT FIRE SUPPRESSION SYSTEMS
AUTOMOTIVE VALVES & REGULATORS WATER CARBONATION
LPG TANK VALVES & REGULATORS
LPG CYLINDER VALVES & REGULATORS
DIGITAL MEASUREMENT SYSTEMS PLASTIC INJECTION MOULDING
www.rotarex.comwww.rotarex.comwww.rotarex.com
WORLDWIDE HEADQUARTERS
Ceodeux Meditec Customer Service
24 rue de Diekirch, B.P. 19
7505 Lintgen, Luxembourg
Tel.: +352 32 78 32 1
Fax: +352 32 78 32 854
E-mail: [email protected]
© 2012 - ROTAREX S.A.
CEODEUX MEDITEC - CEODEUX S.A.
MPR-EN-1115
REGIONAL / COUNTRY HEAD QUARTERS
NORTH AMERICA
USARotarex North America
221 Westec Drive
Westmoreland Technology Park I
Mt. Pleasant, Pennsylvania 15666 USA
Tel.: +1 724-696-43 40
Fax: +1 724-696-43 64
E-mail: [email protected]
SOUTH AMERICA
BRASILRotarex Brazil Ltda
Cond. Ind. Portal da Anhanguera Estr. Municipal
Gov. Mário Covas, S/N
13279-411 Bairro Macuco -Valinhos
São Paulo Brazil
Tel.: +55 19 3518 0800
Fax: +55 19 3869-1503
E-mail: [email protected]
EUROPE
FRANCE, BENELUX, GERMANY24, rue de Diekirch, BP 19
L-7505 Lintgen, Luxembourg
Tel.: +352 32 78 32-1
Fax: +352 32 78 32 317
E-mail: [email protected]
ITALYRotarex Italia S.r.l.
46 Via Giacomo Matteotti
I-25080, Ciliverghe di Mazzano (BS) Italy
Tel.: +39 030 212 05 50
Fax: +39 030 212 23 62
E-mail: [email protected]
SPAINRotarex Spain
C/Alcala, 4° Oficina 404
28014 Madrid
Tel.: +34 (91) 570 89 05
Mobile: + 34 (650) 908 856
E-mail: [email protected]
RUSSIARotarex Rus
Tverskaya street, 20/1 bldg. 1 of 506
125009, Moscow Russia
Tel.: +7 985 125 7776
Skype: rotarexrus
E-mail: [email protected]
POLANDRotarex Polska
13 Gröbli
PL 49-300 Brzeg, Poland
Tel.: +48 77 416 40 16
Fax: +48 77 416 20 99
E-mail: [email protected]
ASIA
MIDDLE EASTRotarex Middle East
Jebel Ali Free Zone - Lob 17, Office 415
P.O. Box 261749
Dubai, U.A.E.
Tel.: +971 (0) 4 887 6701
Fax: +971 (0) 4 887 6702
E-mail: [email protected]
CHINARotarex Star
60 Yuan Zhong Road
Shanghai Nanhui Industrial Zone
201300, Shanghai China
Tel.: +86-21 5800 4000
Fax: +86-21 5800 3226
E-mail: [email protected]
SINGAPORERotarex Fareast Pte Ltd
10 Ubi Crescent
Ubi Techpark, Lobby C, #06-55
408564 Singapore
Tel.: +65 64 72 37 27
Fax: +65 64 72 45 28
E-mail: [email protected]
JAPANRotarex Japan Ltd
6F, Yawaragi Bldg., 1-27-11 Taito
Taito-ku, Tokyo
110-0016 JAPAN
Tel.: +81 3 5817 8108
Mob.: +81 80 3755 2539
Fax: +81 3 5817 8168
E-mail: [email protected]
INDIARotarex ENGG. PVT. LTD.
C-105, Mahindra Apartments
Vikaspuri, New Delhi 110018
mob.: +91-9873467666
E-mail: [email protected]
THAILANDRotarex (Thailand) Co Ltd.
297, Wanglee Tower
3rd floor, Unit C, Surawong Road
Suriyawong, Bangrak,
Bangkok 10500 Thailand
Tel.: +66 2635 71 79
E-mail: [email protected]
SOUTH KOREARotarex Luz Tech Co.
305-500, 547 Yongsan-dong
Yuseong-gu
Taejon, Korea
Tel.: +82 42 825 8911
Fax: +82 42 825 8913
E-mail: [email protected]
OCEANIA
AUSTRALIARotarex Australia/New Zealand
Hornsby Heights,
Sydney, NSW, 2077
Tel.: +61 477 477 481
E-mail: [email protected]