DARE Page 1 Non-Invasive Induction Link Model for Implantable Biomedical Microsystems: Pacemaker to Monitor Arrhythmic Patients in Body Area Networks Prepared by: Anum Tauqir

description

Non-Invasive Induction Link Model for Implantable Biomedical Microsystems: Pacemaker to Monitor Arrhythmic Patients in Body Area Networks. Prepared by: Anum Tauqir. Outline. Background Problem Statement Motivation Mathematical Model Equivalent Circuits Equations Simulation Results - PowerPoint PPT Presentation

Transcript of Prepared by: Anum Tauqir

Page 1: Prepared by: Anum Tauqir

DARE Page 1

Non-Invasive Induction Link Model for Implantable Biomedical Microsystems:

Pacemaker to MonitorArrhythmic Patients in Body Area

Networks

Prepared by:Anum Tauqir

Page 2: Prepared by: Anum Tauqir

DARE Page 2

Outline Background

Problem Statement

Motivation

Mathematical Model

Equivalent Circuits

Equations

Simulation Results

Conclusion

Page 3: Prepared by: Anum Tauqir

DARE Page 3

Background

Page 4: Prepared by: Anum Tauqir

DARE Page 4

Medical Implants aim to: replace missing body parts or

deliver medication, monitor body functions, or provide support to

organs and tissues.

Most widely implanted device: Pacemaker

monitor patients with heart related issues

Most commonly occurring is arrhythmia

Page 5: Prepared by: Anum Tauqir

DARE Page 5

an abnormal heart rhythm, due to changes in the

conduction of electrical impulses through the heart.

Pacemakers: use low-energy electrical pulses to overcome this abnormality.

They create forced rhythms according to natural human heart

beats, to let the heart to function in a normal manner.

consists of a small battery, a generator and wires attached to the

sensor to be inserted into the patients heart.

Arrhythmia

Page 6: Prepared by: Anum Tauqir

DARE Page 6

Working of Pacemaker

Page 7: Prepared by: Anum Tauqir

DARE Page 7

Problem Statement

Page 8: Prepared by: Anum Tauqir

DARE Page 8

To cater for arrhythmia: generated pulses carry sensed information regarding different

events occurring inside the heart to the doctor

processing and transmission of data,

create a strain on the battery of a pacemaker to consume huge

amount of power that ultimately;

depletes the sensor and hence becomes unable to further carry any

informational data.

Page 9: Prepared by: Anum Tauqir

DARE Page 9

Motivation

Page 10: Prepared by: Anum Tauqir

DARE Page 10

Induction technique is presented to: recharge the sensors battery, implanted inside a pacemaker to

avoid early depletion

Technique focuses on enhancing:

voltage gain

link efficiency

Two equivalent circuits: Series tuned primary circuit

Series tuned primary and parallel tuned secondary circuit

Page 11: Prepared by: Anum Tauqir

DARE Page 11

Mathematical Model

Page 12: Prepared by: Anum Tauqir

DARE Page 12

Induction Link

Primary Circuit powered by a voltage source

Secondary Circuit Source generates magnetic flux in order to induce power at

secondary side, implanted inside human body. Interface

skin acts as an interface or a barrier between the two circuits.

Page 13: Prepared by: Anum Tauqir

DARE Page 13

Induction Link Parameters Coupling Co-efficient (k)

degree of coupling between the two circuits.enhances the link efficiency In WBANs for body tissues safety:

k < 0.45

Voltage Gain (Vout / Vin) ratio that, indicates an increase in the voltage at the output side in

relative to the voltage applied at primary side Link Efficiency (η)

ability of transferring power from primary side to secondary side in an efficient manner.

Page 14: Prepared by: Anum Tauqir

DARE Page 14

Equivalent Circuits

Page 15: Prepared by: Anum Tauqir

DARE Page 15

Series Tuned Primary Circuit (STPC)

Page 16: Prepared by: Anum Tauqir

DARE Page 16

a capacitor is connected in series at primary side. as, only a small amount of voltage induces because of a low

coupling factor of 0.45 so, a series tuned circuit is used in order to:

induce sufficient amount of voltage to the secondary coil

Page 17: Prepared by: Anum Tauqir

DARE Page 17

Series Tuned Primary and Parallel Tuned Secondary Circuit (STPPTSC)

Page 18: Prepared by: Anum Tauqir

DARE Page 18

capacitor C2p is connected in parallel at secondary side as, the sensors implanted inside a human body operate under low

frequencies. parallel capacitor let the circuit to act as a low pass filter

which, allows low frequencies to pass through and, blocking the higher frequencies thereby,

preventing damages to body tissues

Page 19: Prepared by: Anum Tauqir

DARE Page 19

Model ParametersParameter Value

Operating frequency f = 13.56 MHz

Primary coil L1 = 5.48 μH

Secondary coil L2 = 1 μH

Parasitic resistance of the transmitter coil

RL1 2.12 ≃ Ω

Parasitic resistance of the receiver coil

RL2 1.63 ≃ Ω

Load resistance Rload = 320 Ω

Page 20: Prepared by: Anum Tauqir

DARE Page 20

Equations

Page 21: Prepared by: Anum Tauqir

DARE Page 21

Voltage Gain

For STPC

For STPPTSC

where,

Page 22: Prepared by: Anum Tauqir

DARE Page 22

Link Efficiency

For STPC

For STPPTSC

where,

where,

Page 23: Prepared by: Anum Tauqir

DARE Page 23

Simulation Results

Page 24: Prepared by: Anum Tauqir

DARE Page 24

Voltage Gain of Series Tuned Primary Circuit

Page 25: Prepared by: Anum Tauqir

DARE Page 25

Link Efficiency of Series Tuned Primary Circuit

Page 26: Prepared by: Anum Tauqir

DARE Page 26

Voltage Gain of Series Tuned Primary and Parallel Tuned Secondary Circuit

Page 27: Prepared by: Anum Tauqir

DARE Page 27

Link Efficiency of Series Tuned Primary and Parallel Tuned Secondary Circuit

Page 28: Prepared by: Anum Tauqir

DARE Page 28

Conclusion