How to calibrate an RTD or Platinum Resistance Thermometer ...
Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance...
Transcript of Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance...
Image sheet
Precision Resistance DecadesPrecision RTD Simulators
M-622, M-612M-602, M-602A
Basic technical data
Resistance decade M-602
Precision RTD simulator M-612
Precision resistance decade M-622
Resistance range from 0.1 Ohm to 10 MOhmAccuracy 0.02 %Temperature coefficient < 25ppm/ CMax. load 2 W (120 V or 0.6 A)Simulation of RTD (Pt, Ni) temperature sensorsAccuracy +/- 0.2 C2, 3, 4 - wire connectionPower line adapter (Internal accumulator optionally)Interface RS-232 (IEEE 488 optionaly)
o
o
Resistance range from 16.0000 Ohm to 10 kOhmAccuracy 0.005 %Temperature coefficient < 1ppm/ CSimulation of RTD (Pt, Ni) temperature sensorsSimulation of cable resistanceAccuracy +/- 0.02 C2, 3, 4 - wire connectionPower line adapter (Internal accumulator optionally)Interface RS-232 (IEEE 488 optionaly)
o
o
Resistance range from 1.000 00 Ohm to 1.200 000 MOhmAccuracy 0.005 %Temperature coefficient < 1ppm/ COperating voltage 120VSimulation of RTD (Pt, Ni) temperature sensorsAccuracy +/- 0.02 C2, 3, 4 - wire connectionInternal accumulator / power line adapterInterface RS-232 (IEEE 488 optionaly)
o
o
Multifunction display shows:
Output terminals
Numerical keypadwith cursor keys
Housing holder
* status of the decade * preset values * SETUP and Calibration menu* calibration data * local or remote control
* 2W, 4W resistance* Simulation RTD sensors ( Pt, Ni ) * Open/Short functions
GND terminalprotection earth Power on/off button
Front panel ergonomy
Series of M6x2 Precision resistance decades/RTD simulators offers decades of electric resistance up 0,1 Ohm to 10 MOhm.
Basic application fields are calibration laboratories, manufacturers of electric resistance meters regulators and process meters.
Front panel is divided into logical areas :
* output terminals field * numerical keypad with cursor keys * display with contact and visibility from all directions
LED indicated4W function
LED indicated2W function
Rear panel with interface
Power supply connector
Power line adapter
Series of M6x2 Precision resistance decades/RTD simulators offers decades of electric resistance up 0,1 Ohm to 10 MOhm.
Basic application fields are calibration laboratories, manufacturers of electric resistance meters, regulators and process meters.
RS-232 interface
GPIB interface
2 W Resistance calibration
Simple connection and one touch buttons simplifytwo-terminal calibration of UUTs: * Connect UUT input to the precision decade 2W output terminals * Set R-function and requested value * Press ENTER
LED indicated2W function
1. Select function R2. Select 4W / 2W 3. Set value 0 Ohm
4. Set requested value using numerical keypad5. Press ENTER
Decade display R mode.
4 W Resistance calibration
Simple connection and one touch buttons simplify four-terminal calibration of UUTs: * Connect UUT input to the precision decade 4W output terminals * Set R-function and requested value * Press ENTER
LED indicated4W function
1. Select function R2. Select 4W / 2W 3. Set value 0-10 kOhm
4. Set requested value using numerical keypad5. Press ENTER
Decade display R mode.
2 W / 4W Temperature calibration
Simple connection and one touch buttons simplify two/four-terminal calibration of UUTs: * Connect UUT input to the precision decade output terminals * Set RTD simulation and requested value * Press ENTER
LED indicated2W function
1. Select function and temperature scale Pt(68),Pt(90),Pt(US), Ni2. Set value of Ro(Pt,Ni) 3. Set T.unit (°C), (°F)
4. Set requested value of temperature using numerical keypad5. Press ENTER
Decade display RTD mode.
Heat meters calibration
LED indicated2W function
1. Select function and temperature scale Pt(68),Pt(90),Pt(US), Ni2. Set value of Ro(Pt,Ni) 3. Set T.unit (°C), (°F)
4. Set requested value of temperature using numerical keypad5. Press ENTER
Simple connection and one touch buttons simplify two/four-terminal calibration of heat meters: * Connect UUT input 1 to the RTD simulator output terminals * Connect UUT input 2 to the RTD simulator output terminals * Set RTD simulation and requested value T1, T2 * Press ENTER
1
2
1
2
Input 1Input 2
Decade display RTD mode.
Requested accuracy
1 mOhm1 mOhm1 mOhm
.
.
.20 Ohm
Standard(terminals)
R00 (4W)R01 (4W)R02 (4W)
.
.
.R35 (R2W/4W)
Nominal value
2.0 Ohm3.9 Ohm7.8 Ohm
.
.
.602 kOhm
Decade recalibration
Simple procedure of recalibration is accesible using decade CALIBRATION mode: - correct password must be entered befor calibration
Decade display password mode.
Decade display calibration mode.
- recalibration procedure consist of measuring basic resistance values and entering their actually measured data
- in following table nominal values of calibration points and requested recalibration accuracy are described
Remote control & Automatic calibration
Precision decades full automatic operatio using remote control mode. Following intrefaces are available for connection the calibrator to PC:
- GPIB ( IEEE-488 ) interface. Connector is located on the rear panel Nationa instruments GPIB card is requested to be installed in the computer.
- RS-232 serial line. connector is located on the rear panel.
- USB interface using RS-232-USB converter
- by user created application SW, using control commands SCPI
- using Meatest CALIBER application SW
Decades can be applied in systems for automatic and semiautomatic calibration ofUUTs:
CALIBER application software for automatic and semiautomatic calibration of meters of electric quantities.
CALIBER ProcedureCalibrationcertificate
Standard unit
Unit under test
CALIBER application software is based on instrument cards. Basic instrument card are delivered with decades,however new card be created by calibration laboratory.
- output from the CALIBER is list of calibration points with measured and evaluated results including uncertainty of calibration
- calibration procedures can be easily created by calibration laboratory
Ordering information - options
Resistance decade M-602 M602-Vxxxx - basic version (10W - 300kW)
M602A-Vxxxx - extende d version (0.1W - 10MW) Bus M602-V1xxx - RS232 M602-V2xxx - IEEE488
Addit ional functions M602-Vx0xx - none M602-Vx1xx - Short/Open function
Housing M602-Vxx0x - table version M602-Vxx1x - module 19“, 3HE M602-Vxx2x - table version with holde r
Power supplying M602-Vxxx0 - power line adaptor M602-Vxxx1 - battery + power line adapt or Example of order:
M602A-V2110 - resistance decade 0.1W - 10MW, IEEE488, Short/Open, 19“ rack, without battery
Precision RTD simulator M-612
Bus M612-V1xxx - RS232 M612-V2xxx - GPIB
Addit ional functions M612-Vx0xx - none M612-Vx1xx - Short/Open function
Housing M612-Vxx0x - table version M612-Vxx1x - module 19“, 3HE
Bus M622-V1xxx - RS232 M622-V2xxx - IEEE488
Addit ional functions M622-Vx0xx - none M622-Vx1xx - Short/Open function
Housing M622-Vxx0x - table version M622-Vxx1x - module 19“, 3HE
Precision resistance decade M-622