Pre-clinical Evaluation of the New Xoft-AxxentTM ... · – TG-56 Code of Practice for...
Transcript of Pre-clinical Evaluation of the New Xoft-AxxentTM ... · – TG-56 Code of Practice for...
Pre-clinical Evaluation of the New Xoft-AxxentTM Electronic Brachytherapy System
PrePre--clinical Evaluation of the New clinical Evaluation of the New XoftXoft--AxxentAxxentTMTM Electronic Electronic Brachytherapy SystemBrachytherapy System
J Turian, D Bernard, Z Hu, A Dickler, J ChuJ Turian, D Bernard, Z Hu, A Dickler, J ChuRush University Medical Center, Chicago, IL 60612Rush University Medical Center, Chicago, IL 60612
WEWE--CC--224C224C--0808
9/5/2006 WE-C-224C-08 2
OutlineOutlineOutline
Introduction
System description
Testing methodology
Tests performed/Results
Conclusions
IntroductionIntroduction
System descriptionSystem description
Testing methodologyTesting methodology
Tests performed/ResultsTests performed/Results
ConclusionsConclusions
9/5/2006 WE-C-224C-08 3
IntroductionIntroductionIntroduction
Why EBT– Dosimetric characteristic similar to existing radionuclides
– Ability to switch the source on/off
– Adjustable radiation output
– Less regulatory requirements
– Less shielding
– Elimination of radioactive waste
Why EBTWhy EBT–– Dosimetric characteristic similar to existing Dosimetric characteristic similar to existing radionuclidesradionuclides
–– Ability to switch the source on/offAbility to switch the source on/off
–– Adjustable radiation outputAdjustable radiation output
–– Less regulatory requirementsLess regulatory requirements
–– Less shieldingLess shielding
–– Elimination of radioactive wasteElimination of radioactive waste
9/5/2006 WE-C-224C-08 4
ObjectivesObjectivesObjectives
Investigate the Xoft-Axxent EB system in a clinical setting using an existing framework
Provide feedback on the system’s performance and suggest improvements
Evaluate and improve a proposed set of QA procedures
Investigate the Investigate the XoftXoft--AxxentAxxent EB system in a clinical EB system in a clinical setting using an existing frameworksetting using an existing framework
Provide feedback on the systemProvide feedback on the system’’s performance s performance and suggest improvementsand suggest improvements
Evaluate and improve a proposed set of QA Evaluate and improve a proposed set of QA proceduresprocedures
9/5/2006 WE-C-224C-08 5
System DescriptionSystem DescriptionSystem Description
Cooling connections
HV connection
Micro-miniature X-ray source
9/5/2006 WE-C-224C-08 6
System description: X-ray sourceSystem description: XSystem description: X--ray sourceray source
Miniature X-ray source inserted into a “flexible” cooling catheter
High vacuum x-ray tube technology50 kV operating potentialOutput: ~18 Gy/min @1cm in tissueWater cooled
X-Ray Tube HV Cable
X-Ray Source Tip Detail
9/5/2006 WE-C-224C-08 7
System Description: X-ray SourceSystem Description: XSystem Description: X--ray Sourceray Source
9/5/2006 WE-C-224C-08 8
System Description: ApplicatorSystem Description: ApplicatorSystem Description: Applicator
Multiple BalloonShapes and Sizes
HDR X-ray SourceLumen
Balloon Inflation Valve
Multi-Lumen Shaft
Drainage Port Valve
Drainage Holes
9/5/2006 WE-C-224C-08 9
Testing MethodologyTesting MethodologyTesting Methodology
AAPM Task Group Reports:– TG-43,TG-43U Brachytherapy Dosimetry Formalism
– TG-56 Code of Practice for Brachytherapy Physics
– TG-59 High Dose Rate Tx Delivery
System log files examination
Measurements using equipment:– Provided by Xoft
– General therapy equipment available in the clinic
AAPM Task Group Reports:AAPM Task Group Reports:–– TGTG--43,TG43,TG--43U Brachytherapy 43U Brachytherapy DosimetryDosimetry FormalismFormalism
–– TGTG--56 Code of Practice for Brachytherapy Physics56 Code of Practice for Brachytherapy Physics
–– TGTG--59 High Dose Rate 59 High Dose Rate TxTx DeliveryDelivery
System log files examination System log files examination
Measurements using equipment:Measurements using equipment:–– Provided by Provided by XoftXoft
–– General therapy equipment available in the clinic General therapy equipment available in the clinic
9/5/2006 WE-C-224C-08 10
Tests PerformedTests PerformedTests Performed
Controller’s GUI functionality
Beam stability
Output stability
Timer accuracy / End effects
Source transit time
Positional accuracy
Treatment planning & delivery
Safety characteristics
Some general design constraints
ControllerController’’s GUI functionalitys GUI functionality
Beam stabilityBeam stability
Output stability Output stability
Timer accuracy / End effectsTimer accuracy / End effects
Source transit timeSource transit time
Positional accuracyPositional accuracy
Treatment planning & delivery Treatment planning & delivery
Safety characteristicsSafety characteristics
Some general design constraintsSome general design constraints
9/5/2006 WE-C-224C-08 11
Controller FunctionalityController FunctionalityController Functionality
Hope you had a chance to see the unit (Xoft booth)
Setup and delivery:– Enter patient data
– Import dwell time file
– Perform source calibration
– Treatment delivery
– Treatment interrupt recovery
– Treatment log files
Status indicator lights
Emergency off
Hope you had a chance to see the unit (Hope you had a chance to see the unit (XoftXoft booth)booth)
Setup and delivery:Setup and delivery:–– Enter patient dataEnter patient data
–– Import dwell time fileImport dwell time file
–– Perform source calibrationPerform source calibration
–– Treatment deliveryTreatment delivery
–– Treatment interrupt recoveryTreatment interrupt recovery
–– Treatment log filesTreatment log files
Status indicator lightsStatus indicator lights
Emergency offEmergency off
9/5/2006 WE-C-224C-08 12
Beam Parameters StabilityBeam Parameters StabilityBeam Parameters Stability
Generator kVp, mA– Direct measurement was
impractical (use the log files)
Beam current (µA) –from log files
Startup dose– Magnitude
– Beam quality
– How to account for it
Generator kVp, Generator kVp, mAmA–– Direct measurement was Direct measurement was
impractical (use the log impractical (use the log files)files)
Beam current (Beam current (µµAA) ) ––from log filesfrom log files
Startup doseStartup dose–– MagnitudeMagnitude
–– Beam qualityBeam quality
–– How to account for itHow to account for it
20s DwT exposure; kVp, filament current, beam current vs time
0.001
0.01
0.1
1
10
100
0:000:000:000:000:000:000:000:000:000:020:060:100:140:180:22
time(m:ss)
VOLTAGE(kV) FILAMENT(A)Beam(mA)
20s DwT exposure; kVp, filament current, beam current vs time
0.001
0.01
0.1
1
10
100
0:000:000:000:000:000:000:000:000:000:020:060:100:140:180:22
time(m:ss)
VOLTAGE(kV) FILAMENT(A)Beam(mA)
9/5/2006 WE-C-224C-08 13
Beam Parameters StabilityBeam Parameters StabilityBeam Parameters Stability
voltagevoltage 0.26%0.26% 0.17%0.17% 0.23%0.23% 0.20%0.20%
currentcurrent 0.62%0.62% 0.43%0.43% 0.42%0.42% 0.39%0.39%
Single tube over different days (1σ)
Different tube on same day (1σ)
SNSN 27232723 26892689 27582758 29042904voltagevoltage 0.19%0.19% 0.17%0.17% 0.04%0.04% 0.05%0.05%
currentcurrent 0.18%0.18% 0.39%0.39% 0.26%0.26% 0.15%0.15%
9/5/2006 WE-C-224C-08 14
Output Constancy of Four SourcesOutput Constancy of Four SourcesOutput Constancy of Four Sources
Standard Imaging Well chamber100 sec collection time
tube Calibration current (nA)
Corrected collection
charge (nC) 2689 106.3 1.415 2723 100.5 1.446 2758 104.5 1.451 2904 85.6 1.403
tube Calibration current (nA)
Corrected collection
charge (nC) 2689 106.3 1.415 2723 100.5 1.446 2758 104.5 1.451 2904 85.6 1.403
σ=1.6%
9/5/2006 WE-C-224C-08 15
Time Adjustment Factor FTime Adjustment Factor Time Adjustment Factor FF
elecSK
k
PINSF××
= 06.1
I - measured current corrected for temp and pressure
Pelec – electrometer correction factor
1.06 – balloon correction factor
Sk – treatment planning AKS (nominal value)
NSK – AKS ion chamber coefficient NSK=R·C
C – ion chamber calibration factor for I-125 (6711)
R – ratio of Xoft/6711 (for 50 kVp: R=1.939)
I I -- measured current corrected for temp and pressuremeasured current corrected for temp and pressure
PPelecelec –– electrometer correction factorelectrometer correction factor
1.06 1.06 –– balloon correction factorballoon correction factor
SSkk –– treatment planning AKS (nominal value)treatment planning AKS (nominal value)
NNSKSK –– AKS ion chamber coefficient NAKS ion chamber coefficient NSKSK=R=R··CC
C C –– ion chamber calibration factor for Iion chamber calibration factor for I--125 (6711)125 (6711)
R R –– ratio of Xoft/6711 (for 50 kVp: R=1.939)ratio of Xoft/6711 (for 50 kVp: R=1.939)
9/5/2006 WE-C-224C-08 16
Output Constancy versus TimeOutput Constancy versus TimeOutput Constancy versus Time
Standard Imaging Well chamber100 sec collection time
tube Calibration current (nA)
Corrected collection
charge (nC) 1/30 102.1 1.463 1/31 100.5 1.446 2/1 101.5 1.464 2/2 102.4 1.445
tube Calibration current (nA)
Corrected collection
charge (nC) 1/30 102.1 1.463 1/31 100.5 1.446 2/1 101.5 1.464 2/2 102.4 1.445
σ=0.8% σ=0.7%
9/5/2006 WE-C-224C-08 17
End EffectsEnd EffectsEnd Effects
Materials– QA phantom provided by Xoft
– PTW TN34013 (0.0053cm3) soft X-ray ion chamber
– CNMC electrometer
Methods– Measure the charge collected for
different exposure times
Including start up
During stable operation
MaterialsMaterials–– QA phantom provided by QA phantom provided by XoftXoft
– PTW TN34013 (0.0053cm3) soft Xsoft X--ray ion chamberray ion chamber
–– CNMC electrometerCNMC electrometer
MethodsMethods–– Measure the charge collected for Measure the charge collected for
different exposure timesdifferent exposure times
Including start upIncluding start up
During stable operationDuring stable operation
9/5/2006 WE-C-224C-08 18
Timer Error and LinearityTimer Error and LinearityTimer Error and Linearity
Obtained from:– 20,40,80,160,320 sec
exposures
– charge collected using the PTW chamber
– Regression results
Timer accuracy 2.0 sec
Timer linearity 0.32%
2s may be clinically significant and should be taken into account
Obtained from:– 20,40,80,160,320 sec
exposures
– charge collected using the PTW chamber
– Regression results
Timer accuracy 2.0 sec
Timer linearity 0.32%
2s may be clinically 2s may be clinically significant and should be significant and should be taken into accounttaken into account
Timer Linearity / Accuracy
y = 0.9968x - 1.9923R2 = 1
0.0050.00
100.00150.00200.00250.00300.00350.00400.00450.00
0.00 100.00 200.00 300.00 400.00 500.00
Xoft Dwell Time (s)M
easu
red
Dw
ell T
ime
(s)
Timer Linearity / Accuracy
y = 0.9968x - 1.9923R2 = 1
0.0050.00
100.00150.00200.00250.00300.00350.00400.00450.00
0.00 100.00 200.00 300.00 400.00 500.00
Xoft Dwell Time (s)M
easu
red
Dw
ell T
ime
(s)
9/5/2006 WE-C-224C-08 19
Start Up DoseStart Up DoseStart Up Dose
Source Ramp-Up Sequence
0 4 8 12 16 20 24 28 32
Time (seconds)
High Voltage (kV)
Filament Current (A)
Beam Current (uA)20 kV
50 kV
1.1 A1.5 A
300 uA
100 uA
Source Ramp-Up Sequence
0 4 8 12 16 20 24 28 32
Time (seconds)
High Voltage (kV)
Filament Current (A)
Beam Current (uA)20 kV
50 kV
1.1 A1.5 A
300 uA
100 uA
Up to 10% of the prescribed dose
Beam quality is reduced (37.5 for 50 kV beam)
Corrections should be applied for the first dwell position
(courtesy of Xoft)
Up to 10% of the prescribed dose
Beam quality is reduced (37.5 for 50 kV beam)
Corrections should be applied for the first dwell position
(courtesy of Xoft)
Dose During Source Turn-On Sequence(Axxent Source in Water)
0.01
0.10
1.00
10.00
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0
Radius (cm)
Del
iver
ed D
ose
(Gy)
Turn-On3.0 sec at 50 kV5.7 sec at 37.5 kV
Dose During Source Turn-On Sequence(Axxent Source in Water)
0.01
0.10
1.00
10.00
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0
Radius (cm)
Del
iver
ed D
ose
(Gy)
Turn-On3.0 sec at 50 kV5.7 sec at 37.5 kV
9/5/2006 WE-C-224C-08 20
Source Transit TimeSource Transit TimeSource Transit Time
Catheter length: 25 cmSource travel distance: 7.5 cm– HDR sources:
Nucletron-12 cm for 0.25 cm step size; 24 cm for 0.5 cm step sizeVarisource-5 cm and 10 cm at 0.25 and 0.5 cm step, respectively
Maximum source speed: 0.83 cm/sFor the 7 cm balloon the transit time is 8.4 s (0.5cm step size) This relatively large transit time should be investigated and accounted for in the treatment planning processFor a 4cm diameter balloon there is a calculated 0.5% +/-1.4% dose uncertainty due to transit time
Catheter length: 25 cmCatheter length: 25 cmSource travel distance: 7.5 cmSource travel distance: 7.5 cm–– HDR sources:HDR sources:
NucletronNucletron--12 cm for 0.25 cm step size; 24 cm for 0.5 cm step size12 cm for 0.25 cm step size; 24 cm for 0.5 cm step sizeVarisourceVarisource--5 cm and 10 cm at 0.25 and 0.5 cm step, respectively5 cm and 10 cm at 0.25 and 0.5 cm step, respectively
Maximum source speed: 0.83 cm/sMaximum source speed: 0.83 cm/sFor the 7 cm balloon the transit time is 8.4 s (0.5cm step For the 7 cm balloon the transit time is 8.4 s (0.5cm step size) size) This relatively large transit time should be investigated and This relatively large transit time should be investigated and accounted for in the treatment planning processaccounted for in the treatment planning processFor a 4cm diameter balloon there is a calculated 0.5% +/For a 4cm diameter balloon there is a calculated 0.5% +/--1.4% dose uncertainty due to transit time 1.4% dose uncertainty due to transit time
9/5/2006 WE-C-224C-08 21
Source Positional AccuracySource Positional AccuracySource Positional Accuracy
Determined using RC film – Precision better than 1 mm over span of source travel
(7.5 cm)
Step size accuracy (pull back mechanism)– External markers
– ≤0.5mm
All sources manufactured to identical lengths
A better source positioning QA device will be useful
Determined using RC film Determined using RC film –– Precision better than 1 mm over span of source travel Precision better than 1 mm over span of source travel
(7.5 cm)(7.5 cm)
Step size accuracy (pull back mechanism)Step size accuracy (pull back mechanism)–– External markersExternal markers
–– ≤≤0.5mm 0.5mm
All sources manufactured to identical lengthsAll sources manufactured to identical lengths
A better source positioning QA device will be A better source positioning QA device will be useful useful
9/5/2006 WE-C-224C-08 22
“Flexibility” of the Source Catheter““FlexibilityFlexibility”” of the Source Catheterof the Source Catheter
Maximum bend in the catheter– 15 deg flex
– Source out of the nest
– Fault
Comparison– HDR source wires can negotiate a curvature with a minimum 2 cm
radius
Clinical Implications– Proper balloon placement
– Ensure straight approach to balloon’s X-ray source lumen
Maximum bend in the catheterMaximum bend in the catheter–– 15 deg flex15 deg flex
–– Source out of the nestSource out of the nest
–– FaultFault
ComparisonComparison–– HDR source wires can negotiate a curvature with a minimum 2 cm HDR source wires can negotiate a curvature with a minimum 2 cm
radiusradius
Clinical ImplicationsClinical Implications–– Proper balloon placement Proper balloon placement
–– Ensure straight approach to balloonEnsure straight approach to balloon’’s Xs X--ray source lumenray source lumen
9/5/2006 WE-C-224C-08 23
Treatment PlanningTreatment PlanningTreatment Planning
Source parameters specified using the TG43 framework and modeled using Plato TPS
Xoft versus Ir-192 dose distributions
Xoft surface to target ratios (STR) 2.1 to 3.5 versus 1.8 to 2.8 for Ir-192
Source parameters specified using the TG43 Source parameters specified using the TG43 framework and modeled using Plato TPS framework and modeled using Plato TPS
XoftXoft versus Irversus Ir--192 dose distributions 192 dose distributions
XoftXoft surface to target ratios (STR) 2.1 to 3.5 surface to target ratios (STR) 2.1 to 3.5 versus 1.8 to 2.8 for Irversus 1.8 to 2.8 for Ir--192192
IrIr--192 18 192 18 posnposnXoftXoft 1 1 posnposn XoftXoft 5 5 posnposn
9/5/2006 WE-C-224C-08 24
Safety Characteristics Safety Characteristics Safety Characteristics
Treatment pause button:
– worked
– treatment resumption possible
Treatment interruption
(displace source from nest):
– treatment resumption possible
Radiation Safety
– FlexiShield adequate
– Distortion of the balloon by the shield was not investigated
Treatment pause button:Treatment pause button:
–– workedworked
–– treatment resumption treatment resumption possiblepossible
Treatment interruption Treatment interruption
(displace source from nest):(displace source from nest):
–– treatment resumption treatment resumption possiblepossible
Radiation SafetyRadiation Safety
–– FlexiShieldFlexiShield adequateadequate
–– Distortion of the balloon by Distortion of the balloon by the shield was not the shield was not investigated investigated
Emergency stop button:
– worked
– treatment log file recovered
Sudden power failure (unplugged unit):
– treatment log file recovered
Treatment stop button:
– worked
– treatment log file recovered
Emergency stop button:Emergency stop button:
–– workedworked
–– treatment log file recoveredtreatment log file recovered
Sudden power failure Sudden power failure (unplugged unit):(unplugged unit):
–– treatment log file recoveredtreatment log file recovered
Treatment stop button:Treatment stop button:
–– workedworked
–– treatment log file recoveredtreatment log file recovered
9/5/2006 WE-C-224C-08 25
Possible Design DeficienciesPossible Design DeficienciesPossible Design Deficiencies
Tube source is easily displaced from the nest on positioning arm
Source attachment clip is fragile
Source travel might be too limited (7.5 cm)
Physics mode will be helpful
Pressure and temperature gauge as part of the system
Interlock failure: X-ray generation with no water flow in cooling tube
– not reproducible
Tube source is easily displaced from the nest on Tube source is easily displaced from the nest on positioning armpositioning arm
Source attachment clip is fragileSource attachment clip is fragile
Source travel might be too limited (7.5 cm)Source travel might be too limited (7.5 cm)
Physics mode will be helpfulPhysics mode will be helpful
Pressure and temperature gauge as part of the systemPressure and temperature gauge as part of the system
Interlock failure: XInterlock failure: X--ray generation with no water flow in ray generation with no water flow in cooling tubecooling tube
–– not reproduciblenot reproducible
9/5/2006 WE-C-224C-08 26
ConclusionsConclusionsConclusions
Very stable outputSource position accuracy < 1.0 mmTimer accuracy and linearity adequate“Turn on” dose needs to be quantified and documentedSource transit effect on delivered dose is ≤1.0% STR’s range is “similar” to Ir-192 Safety interlocks functionalPatient exposure history was always recoverableSource movement should be extendedSource attachment to nest could be redesignedThe accuracy of the dose distribution was not investigated
Very stable outputVery stable outputSource position accuracy < 1.0 mmSource position accuracy < 1.0 mmTimer accuracy and linearity adequateTimer accuracy and linearity adequate““Turn onTurn on”” dose needs to be quantified and documenteddose needs to be quantified and documentedSource transit effect on delivered dose is Source transit effect on delivered dose is ≤≤1.0% 1.0% STRSTR’’ss range is range is ““similarsimilar”” to Irto Ir--192 192 Safety interlocks functionalSafety interlocks functionalPatient exposure history was always recoverablePatient exposure history was always recoverableSource movement should be extendedSource movement should be extendedSource attachment to nest could be redesignedSource attachment to nest could be redesignedThe accuracy of the dose distribution was not investigatedThe accuracy of the dose distribution was not investigated