Maintenance of analyzer Sysmex-CA 1500 Presented by Saima fazal B.S (3 rd yr )

Transcript of Ppt

Page 1: Ppt

Maintenance of analyzer

Sysmex-CA 1500Presented by Saima fazalB.S (3rd yr )

Page 2: Ppt

INTRODUCTION

• The Sysmex CA-1500 System is a fully-automated blood coagulation analyzer

• It can quickly analyze large quantities of samples with a high-degree of accuracy

• The analyzed data can be retained as stored data, displayed, or printed

ABOUT INSTRUMENT

Page 3: Ppt
Page 4: Ppt

FeaturesFlexible communication options for new

and current users of CA systems.

• Up to 120 tests per hour

• Authentic random access analysis.

• Complete quality control software

• Detection for clotting, chromogenic and immunoturbidimetric assays

Page 5: Ppt

MaintenanceWhy maintenance is necessary?

To insure the instrument in optimal operating condition it requires periodic maintenance which includes :

Daily maintenance Weekly maintenance As-needed maintenance

Page 6: Ppt

Daily maintenance clean the probes Discard the used reaction tubes Discard waste Remove condensation from the reagent tray Check and discard trap chamber fluid Run controls

Page 7: Ppt

Weekly maintenance

Prime the hydraulic line with rinse solution Clean the instrument

As-needed maintenance

Adjust the pressure

Page 8: Ppt

Daily maintenance

Cleaning the probes:

• From the main menu screen ,press rinse probe key, the rinse probe screen will appear

• Set CA clean 1 to both the reagent holder A2 and A3

• Probe cleaning will start

Page 9: Ppt

Discarding the used reaction tubes

• The used reaction tubes are automatically thrown in the trash box

• At completion of every 250 tests, or at least once every 24 hours, discard the used reaction tubes from the trash box and clean it with tap water

Page 10: Ppt
Page 11: Ppt

Discarding waste

• At completion of the day analysis, discard the waste fluid that has collected in the waste tank.

• If tank has becomes full it will alarm a sound and a confirmation screen will appear

• Open the waste tank and discard the waste

Page 12: Ppt
Page 13: Ppt

Removing condensation from tray

• After completion of the day’s analysis or at least once every 24 hours, check to see if condensation has formed on the reagent tray.

• Before cleaning the reagent trays, turn the power OFF

• Failure to do so can result in electric shock

Page 14: Ppt
Page 15: Ppt

Weekly maintenance

Priming the hydraulic line with rinse solutionSTEPS:•Check that there is sufficient rinse solution in the rinse tank and waste tank is empty

•From the main menu screen select special menu and then select maintain, the maintenance sub menu will appear.

Maintenance System

Test

Return

Rinse &

prepare

Pressure

adjust

Calib.

lamp

Syringe

cycles

Program

Ver. up

Format

FD

Return

Maintenance sub menu

Page 16: Ppt

• Select the Rinse and prepare key, the prime and Deprime execution screen will appear.

• Then select execute key, priming will be start.

Page 17: Ppt
Page 18: Ppt

Cleaning the instrument

Why we clean the instrument?

To ensure the instrument will be maintained in a good condition, clean it once a week.

Which includes: • Cleaning instrument from exterior• Cleaning instrument from interior

Page 19: Ppt

As needed maintenance

Adjusting the pressure

• The pressure are adjusted 2.2 kg/cm and 0.1 kg/cm

• The pressure are constantly monitored by pressure sensor

• If an abnormality is detected, an error message will appear

• It any pressure related error message is displayed, check the

tubing connections for leaks.

• It nothing is found, adjust the pressure

Page 20: Ppt
Page 21: Ppt

Displaying the pressure adjustment

•From main menu select special key menu•Then from there select maintenance key, maintenance screen will appear.

•Press adjust pressure