Page 1: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

Activities in KRISS

Wan-Cho Cho

2017 CCAUV

CCAUV/17-57

Page 2: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

1

General

Organization change

Acoustics

Vibration

Ultrasound

Page 3: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

2

General

CMC updates (approved at April 2017)

Peer review

Acoustics Whole previously existing items are re-approved

Ultrasound The items of ultrasonic power are newly added

Vibration Whole previously existing items of linear vibration are re-approved The items of angular vibrations are newly added

Acoustics The items for free-field microphone sensitivity for the frequency range of 1 kHz -31.5 kHz

are newly added based on the results of CCAUV.A-K4

Ultrasound The items of ultrasonic power are updated and the uncertainties are updated

Vibration The frequency ranges of items of linear vibration are extended to 0.5 Hz The items of angular vibration are updated and the uncertainties are updated

Page 4: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

1 Sound in Air

- Diffuse-field sensitivity - Optical measurement method

Page 5: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

4

Introduction

Types of sound field

Pressure field Uniform pressure, only diaphragm exposed to pressure Small coupler, cross-section of plane wave duct High S/N ratio & high stability

Free field No reverberant field, SPL distribution satisfy the inverse square law Direct field + effect of sensor body (infinite rod) Representing outdoor condition

Diffuse field Uniform pressure, plane wave incidence with same

probability for every direction Representing indoor condition

Pressure field

Free field

Radom incidence (Diffuse) field

Images from http://www,gras.dk

Page 6: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

5

Introduction

Diffuse Sound Field

General definition Sound field having equal probability of energy flow in all direction and the energy density is uniform in

a volume (Jacobsen, DTU Tech. rep., 1979; Nelisse & Nicolas, JASA, 1997) Summation of infinite-equally spaced plane wave sources General characteristics High reverberation time Uniform distribution of sound pressure Spatial correlation is given by sinc function

Sound field in real life

Page 7: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

6

Introduction

Previous Works

Standard on the Diffuse Field Sensitivity Calibration No standard issued for calibrating the diffuse field sensitivity Pressure sensitivity: IEC 61094-2 (Primary), IEC 61094-5 (Comparison) Free-field sensitivity: IEC 61094-3 (Primary), IEC TS 61094-7 (Correction), IEC 61094-8 (Comparison) No ‘Calibration and Measurement Capabilities’ registered by NMI No international comparison has been conducted

Research Works on the Diffuse Field Sensitivity Calibration Reciprocity method Basic concept & formulation: Diestel, JASA, 1961 Measurement method with a scaled reverberation chamber: Barrera-Figueroa et al., JASA, 2008 Improved result using diffuser: Milhomem et al., Internoise 2016 Random Incidence method Measurement method result: Barrera-Figueroa et al., JASA, 2007

Page 8: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

7

Random Incidence Sensitivity Method

Measurement System

System Configuration Rod: Hollowed pipe, 1 m length, same outer diameter with mic, hanged by fishing line Rotator: Motor controlled with high precision (0.1 degree resolution) Measured with 5 degree step Sound source: Compression driver + Horn Data acquisition system: B&K PULSE + LAN-XI 3060 (FFT Analyzer + Generator) Distance between source and test mic.: 4 m

Test mic.

Compression driver

Rotator

Monitor microphone

DAQ (B&K LAN-XI 3060)

Gen. CH2 CH1 PC

Controller

Page 9: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

8

Random Incidence Sensitivity Method

Measurement System

System Configuration Rod: Hollowed pipe, 1 m length, same outer diameter with mic, hanged by fishing line Rotator: Motor controlled with high precision (0.1 degree resolution) Measured with 5 degree step Sound source: Compression driver + Horn Data acquisition system: B&K PULSE + LAN-XI 3060 (FFT Analyzer + Generator) Distance between source and test mic.: 4 m

Page 10: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

9

Random Incidence Sensitivity Method

Signal Processing

Gating Process Various reflection & scattering effects are included Reflection removal should be applied Applying low-pass filtering for stabilizing impulse response (Kwon & Cho, JASA, 2013) Fluctuation in 2 kHz – 5 kHz is reduced

Frequency response at frontal direction (4160.1792662)

Steady state

response Impulse response (Filtered)

iFFT

Filtered FRF LPF

Direct path IR (Filtered)

Time gating

Direct path FRF (Filtered)

FFT Inverse LPF

Direct path response

0 0.5 1 1.5 2 2.5

x 104

0

0.5

1

1.5

2

2.5

3

3.5

Frequency (Hz)

Freq

uenc

y re

spon

se

Before gatingAfter gating

Averaged measured directivity with 3 microphones

103

104

0

1

2

3

4

5

6

7

8

9

Frequency (Hz)

Dire

ctiv

ity in

dex

(dB

)

LS1, w/ gatingLS1, w/o gatingLS2, w/ gatingLS2, w/o gating

Page 11: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

10

Reciprocity Method

Measurement System

Scaled Mini Reverberation Chamber Non-paralleled walls of aluminum plates Volume: 2.8 m3 Transmission loss of door: 30~45 dB Reverberation time: MDF plate wall: 0.38 ~ 0.60 s Steel plate wall: 0.52 ~ 0.84 s

1000 10000

0.3

0.4

0.5

0.6

0.7

0.8

0.9 MDF plate wall Steel plate wall

Rebe

rver

atio

n Ti

me

(s)

Frequency (Hz)

Page 12: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

11

Reciprocity Method

Measurement System

DAQ (B&K PULSE 3560-C)

Trans. - Receiver Amp. Insert voltage switch

(B&K 5998)

PC

Output Reference

Amplifier with 20 dB Gain (B&K NEXUS)

Output Ref.

Insert voltage junction

Transmitter (B&K ZE0796)

Pre. Amp. with 20 dB gain (B&K 2673-W-001)

Small reverberation chamber (2.8 m3)

Voltage ratio measurement Same to the free-field reciprocity measurement system - Insert voltage method - Isolating the channels of transmitter and receiver to suppress cross-talk - 20 dB preamp. gain to increase S/N ratio

Acquisition process FRF with Random noise

Page 13: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

0 100 200 300 400 500 600 700 800 900 1000-1.5

-1

-0.5

0

0.5

1

1.5x 10

-5

Time (ms)

Impu

lse

resp

onse

12

Reciprocity Method

Signal Processing

Contribution of direct wave For the face to face configuration, contribution of

frontal direction become larger than other directions To apply the removal process Freq. step smaller than (1/RT) Hz is required

2 2.5 3 3.5 4 4.5 5-1

-0.5

0

0.5

1

1.5x 10

-5

Time (ms)

Impu

lse

resp

onse

Direct wave

Averaging process Frequency band average: 1 Hz 1/3 octave band Spatial averaging: 18 separated points (no face to face configuration)

103

104

10-1

100

101

102

Frequency (Hz)

Vol

tage

ratio

103

104

10-1

100

101

102

Frequency (Hz)

Vol

tage

ratio

103

104

10-1

100

101

102

Frequency (Hz)V

olta

ge ra

tio

Raw data Band averaged data Spatially averaged data

Page 14: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

13

Measurement Result

Results comparison

Comparison with the sensitivities in other types of sound fields Diffuse-field sensitivity estimated by the different methods are placed

in between the pressure and free-field sensitivity

1000 10000

-40

-38

-36

-34

-32

-30

-28

Sens

itivity

(dB

re. 1

V/Pa

)

Frequency (Hz)

Pressure sensitiviy Free-field sensitivity Random incidence sensitivity Diffuse field sensitivity with FFT

1000 10000

-40

-38

-36

-34

-32

-30

-28

Sens

itivity

(dB

re. 1

V/Pa

)

Frequency (Hz)

Pressure sensitiviy Free-field sensitivity Random incidence sensitivity Diffuse field sensitivity with FFT

4180.2341431 4180.2208287

Page 15: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

14

Optical method for the future primary standard

Measurement system with gated photon correlation method

System setup Feasibility test with sine-wave tube Installation of free-field measurement system Collection of measurement data and the detailed investigation are on-going

This works was conducted by collaborating with Dr. T. Koukoulas

a former principal research scientist at NPL

Page 16: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

2 Ultrasound

- High ultrasonic power measurement

Page 17: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

16

Ultrasound

Contact point Name: Dr. Yong Tae Kim

(Head, Center for medical convergence metrology) Email: [email protected]

Recent activities Ultrasound power Development absorbing target available to RFB method up to 300 W High power measurement system construction Fabrication of high power transducers (1 MHz, 3 MHz) Pilot study for comparison with NMIJ, NIS (2 MHz – 15 MHz) Report is submitted to SCI journal. HIFU Transducer using meta-acoustic lens HIFU phantom research for histotripsy

Future Plan (2018 - ) Pilot study high power measurement comparison in APMP APMP.U-K1 RMO KC (2018)

Page 18: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

17

High ultrasonic power measurement

High power target

Requirement Echo Reduction: 50 dB (sufficient) Single layer

Characteristics of the developed target Specific Insertion Loss: 15 dB/cm 3 cm-thickness IL = 45 dB

Page 19: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

18

High ultrasonic power measurement

System configuration

300 W US-Power

Generating Part

Measuring Part

Power transducer

Power amplifiers

Step attenuator

1 MHz Power TR.

3 MHz Power TR.

1 kW Power Amp. 500 W Power Amp.

250 W Power Amp.

100 W Power Amp. Thermal voltage converters

1 V TVC 3 V TVC 5 V TVC 10 V TVC 20 V TVC 50 V TVC

Precision Balance Absorbing target Alignment stage

50 g / 0.001 mg

100 g / 0.01 mg

Automation SW

ER: 50 dB / IL: 15 dB/cm

ER: 50 dB / IL: 30 dB/cm

H. Power 50 Ohm termination

Cooling circulator

Page 20: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

19

High ultrasonic power measurement

High power measurement and water temperature problem

Water temperature continues to rise after measurement. Difficult to maintain water temperature: (21.5 ± 2.0 ) ℃ New water tank with Cooling circulator Water temperature can be controlled

Page 21: PowerPoint 프레젠테이션 - BIPM · Measurement system with gated photon correlation method System setup Feasibility test with sine -wave tube Installation of free -field measurement

KRISS

20

Summary

KRISS AUV Re-organization Role, activities and research area are not changed

Acoustics Diffuse-field sensitivity Optical method for the future primary standard

On-going Research Works

Ultrasound High power measurement method Related topics to HIFU

Vibration Research on the national seismic monitoring system to make it traceable to the international

metrology standard