Poster No. M1010 Tecan Platform for LC-MS Bioanalysis of ... · Cartridge-based IC was evaluated...

1
We would like to thank Dr. Lee Hoang from PhyNexus Inc., Ms. Adrienne Chow and Mr. Bob Clark from Tecan US for their help in Tecan setup for cartridge-based immunocapture using PhyTip™ columns. RESULTS (Continued) Integration of Automated Bead- and Cartridge-Based Immunocapture Workflows on a Generic Tecan Platform for LC-MS Bioanalysis of Protein Therapeutics and Biomarkers Naiyu Zheng, Huidong Gu, Kristin Taylor, Yan J. Zhang, Jianing Zeng Analytical and Bioanalytical Operations, Bristol-Myers Squibb Company, Princeton, NJ 08543 PURPOSE Recently, LC–MS technology has been increasingly used for the bioanalysis of protein therapeutics and biomarkers to complement ligand-binding assays (LBA) in drug discovery and development. Regardless of the direct mass spectrometric detection of intact proteins or indirect measurement of their tryptic surrogate peptides, immunocapture (IC) has become the sample preparation technique of choice to achieve the best sample cleanup for ultrasensitive detection of protein analytes by LC-MS. Current IC workflows include manual procedure (e.g. using magnetic beads), automated bead-based automated IC (e.g. Thermo Fisher Scientific’s KingFisher™ platform) and cartridge-based automated IC (e.g. AssayMap Cartridges with Agilent Bravo, MSIA tips on Thermo Scientific’s Versette platform, or PhyTip® columns on TECAN platform). Naiyu Zheng, PhD Rt. 206 & Province Line Road Princeton, NJ 08540 Phone: 609-252-5494 E-mail: [email protected] 2017 AAPS National Biotechnology Conference, May 1 - 3, 2017. San Diego, CA. METHODS RESULTS (Continued) New immunocapture (IC) strategy has been successfully developed by integrating and automating both bead- and cartridge-based IC on existing and generic Tecan freedom EVO® platforms for LC-MS bioanalysis of protein therapeutics and biomarkers. This strategy allows the IC enrichment of protein analytes from diverse biological sample sizes and applications with high flexibility and high throughput without additional capital investment. ACKNOWLEDGEMENTS Integration of Immunocapture (IC) Workflow on a Generic Tecan Freedom EVO® Platform Many commercially available robotic liquid handlers (RLH) specialized for IC are highly vendor-specific. These IC workflows usually require significant capital investment for new RLH and the use of costly cartridges (tips) from the same vendor. In addition, they generally can do only one IC workflow: either bead-based or cartridge-based IC. The purpose of this study is to develop an integrated IC strategy for both bead- and cartridge- based IC on the existing Tecan freedom EVO® platforms for LC-MS bioanalysis of protein therapeutics and biomarkers for diverse applications without additional capital investment. ASSAY 1: Bead-Based Immunocapture (IC) Used for the Quantitation of Two Modified Protein Biomarkers, Protein I and Protein II, in Monkey Blood Lysate To improve LC-MS/MS assay sensitivity, IC of the analytes from 3 mL blood lysate was required. Due to large sample volume, IC in 96-well format is impossible. Automated sample enrichment using bead-based IC in 4-mL sample tubes on Tecan is desirable. Assay 2: Tip-Based IC Used for the Quantitation of An Antibody Drug Conjugate (ADC), Protein III, in Rat Serum by LC-MS/MS LEAP 4X Ultra Pump (uHPLC) & HTC PAL autosampler A Fluidx® XSD-48 automated capping/decapping system was used for automated capping and decapping during IC and subsequent washes. Dynabeads® MyOne™ streptavidin T1 beads with the use of a biotinylated antibody were evaluated. IC of protein analytes was achieved by incubation of 3 mL of monkey blood lysate with the immobilized biotinylated antibody on magnetic beads in 4-mL automation friendly Greiner Bio-One tubes (Cat No: 127261). After IC of the protein analytes by shaking at room temperature for 2 h, followed by centrifugation, the unwanted biological matrix were transferred to the waste, and the remaining beads in the tubes were transferred into a 96-well plate on top of a MagnaBot 96 magnetic separation stand. After two PBST washes, followed by one wash with 25 mM NH 4 OAc with 0.05% Tween-2, the samples were used for on-bead digestion. Quantitation of Proteins I and II was achieved by LC-MS/MS analysis of two representative surrogate peptides. For each processed sample, a 50 µL volume of the digest was injected into an a Waters CORTECS™ UPLC C18+ column (2.1 x 100 mm, 1.6 µm) and analyzed on a Triple Quad 5500 system using MRM detection for Peptide A, Peptide B and SIL-Peptide A (IS). Assay 2: Cartridge-Based IC: LC-MS Bioanalysis of an Antibody Drug Conjugate (ADC), Protein III in Rat Serum Assay 1: LLOQ of 0.125 nM and 0.250 nM in monkey blood lysate were achieved for Protein I and Protein II Assay 1: In Addition to Cost-Effectiveness, Bead-based Immunocapture Using a Tecan Allows the Sensitivity Improvement from Being Able to Process More Sample Volume In Each Sample With High Throughput Assay 2: Automated Cartridge-Based Immunocapture (IC) Workflow on Tecan Platform Assay 2: Comparable LC-MS Responses at LLOQ Concentration Level Achieved from Immunocapture Using PhyTip™ Columns vs. AssayMap™ Cartridges Assay 2: Comparison of Standard Curves for Protein III Obtained from Immunocapture Using Two Platforms (PhyTip™ Column on Tecan vs AssayMap™ Cartridges on Bravo) CONCLUSIONS AssayMap Cartridges Using Agilent’s Bravo™ Platform MSIA Tips Using Thermo Scientific’s Versette™ Platform Magnetic Beads Using Thermo Scientific’s KingFisher™ Platform TECAN Freedom EVO Dynabeads® Streptavidin On bead digestion with trypsin Time (Min) Rel. Abundance Peptide A Peptide B LC-MS/MS Peptide A Magnetic stand Capture antibody biotinylated Protein I Protein II Immobilized Antibody Buffer wash Buffer wash Interference Proteins blood lysate Immuno- Capture (IC) Bead-Based Immunocapture On TECAN Platform Enrich Protein I, and II Peptide A IC-LC-MS/MS for the bioanalysis of Protein III (ADC) in rat plasma was previously developed and validated using an Agilent Bravo™ platform. Evaluation of new IC workflow using PhyTip® columns on an existing TECAN has several advantages, such as cost saving in capital investment of new IC platform, less expensive with the tips, more flexibility and capability with TECAN than using standalone IC platforms. Capture analyte Elute PhyTip™ columns resin Time (Min) Rel. Abundance LC-MS/MS Trypsin Digestion Serum Samples Buffer Wash Immobilize antibody Enriched Protein III Buffer Wash Aspirate Dispense Protein III (ADC) Interference Proteins PhyTip™Streptavidin Resin Capture antibody biotinylated Tip-Based Immunocapture On TECAN Platform Assay 1: Bead-Based IC: LC-MS Bioanalysis of Two Modified Protein Biomarkers, Protein I and Protein II that Derived from an Endogenous Protein (MW: 90 KDa) in Monkey Blood Lysate. PhyTip® streptavidin columns (PhyNeXus, Inc) (20 µL resin bed size on 200-µL Tecan tips) with the use of another biotinylated antibody were evaluated. Cartridge-based IC was evaluated for LC-MS bioanalysis of an antibody drug conjugate (ADC), Protein III in rat serum (15 µL serum + 145 µL PBS). The IC was performed by aspiration and dispensing of the serum samples through PhyTip® columns for 16 cycles (4.2 µL/sec, 10 s delay), followed by a wash prior to acid elution and trypsin digestion. Quantitation of Protein III was achieved by LC-MS/MS analysis of a surrogate peptide. For each processed sample, a 40 µL volume of the digest was injected into an a Waters Acquity™ UPLC HSS T3 column (2.1 x 50 mm, 1.8 µm) and analyzed on a Triple Quad 6500 system with MRM detection for Peptide C and the IS (SIL-Peptide C), respectively. RESULTS Assay 1: Automation of Bead-Based IC Workflow on TECAN Platform Magnetic stand Blood lysate Samples 3 mL/Tube Add capture antibody on beads Shake at RT for 2 h (IC) Centrifugation Decap Discard top supernatant on TECAN Transfer beads to 96-well plate LC-MS/MS On bead digestion at 50°C for 30 min Wash with PBST Automated decapper TECAN Freedom EVO Thaw Centrifugation Immunocapture (2h) Trypsin Digestion (0.5h) Blood Lysate 3 mL Waste lysate TECAN Freedom EVO Washed with buffer three times Assay 1: Representative Chromatograms for the Modified Protein Biomarker, Protein I and Protein II Blank 0 200 400 600 5.11 LLOQ at 0.250 nM of Protein I 0 250 500 750 5.02 SIL-Peptide B (IS) 4.0 4.5 5.0 5.5 6.0 Time, min 0.0 2.0e4 4.0e4 4.82 0 100 200 300 LLOQ at 0.125 nM of Protein II 0 200 400 600 800 4.83 4.60 SIL-Peptide B (IS) 4.0 4.5 5.0 5.5 6.0 0.0 2.0e4 4.0e4 4.82 5.11 Intensity, cps Blank Protein I Protein II "Quadratic" Regression ("1 / x" weighting) (r = 0.9995) 2.0 4.0 6.0 8.0 10.0 12.0 Protein II Conc (nM) 0.00 0.20 0.40 0.60 Protein II (0.125 – 12.5 nM) "Quadratic" Regression ("1 / x" weighting) (r = 0.9986) 2.0 4.0 6.0 8.0 10.0 12.0 0.00 0.10 0.20 0.30 0.40 Analyte Area / IS Area Protein I (0.250-12.5 nM) Protein I Conc (nM) Analyte Area / IS Area One of the approaches to improve LC-MS/MS assay sensitivity for protein quantitation is to enrich protein analytes from a large sample volume using IC. However, when the volume is larger than the capacity of the 96-well format, it could be very challenging. By using a Tecan Freedom EVO platform with a Fluidx® XSD-48 automated capping/ decapping system, sample preparation throughput was increased from 32 samples per batch performed manually to 96 samples per batch when performed on Tecan. Platforms TECAN Freedom EVO Manual Sample Racks 2 Fluidx® XSD-48 Racks 4 DynaMag™-5 Magnet Racks Sample Size Per Process 96 Samples 32 Samples Time for 96 Samples 1 day (< 4 hours) 3 days (3 separate batches are required) Productivity Improvement 3-fold N/A Comment Existing TECAN platform; Capable to handle all liquid transfer as well. N/A PhyTip® columns PhyTip® columns are compatible with Tecan Freedom EVO Step 1 Load PhyTip™ columns: 200-µL PhyTip™ columns packed with 20-µL of Streptavidin resin Step 2 Equilibration: 200 µL PBS (pH 7.4, 10 mM), 4 cycles. Step 3 Load antibody: 100 µL of 100 µg/mL, 8 cycles. Step 4 Wash: 200 µL PBS (pH 7.4, 10 mM), 2 cycles. Step 5 Capture from Serum: 160 µL (15 µL serum + 145 µL PBS), 16 cycles. Step 6 Wash: 200 µL PBS (pH 7.4, 10 mM), 2 cycles. Step 7 Elute: 40 µL 30% ACN with 0.1% formic acid, 4 cycles. Step 8 Neutralize: 10 µL of 1 M NH 4 HCO 3 Reduction/alkylation Trypsin Digestion. PhyTip™ Columns on Tecan AssayMap™ Cartridges on Bravo IS Max 6621 0.5 ug/mL Protein III (LLOQ) Max 4.9e5 IS Max 6041 Max 4.5e5 0.5 ug/mL Protein III (LLOQ) Assay 2: Immunocapture Using PhyTip™ on Tecan Freedom EVO Is More Cost-Effective than Agilent’s Bravo When New Capital Investment is not Required PhyTip™ on Tecan on Tecan (Only test 4 points) AssayMap™ on Bravo Tecan (0.5 -4 ug/mL) "Linear" Regression (1/X 2 weighting): y = 0.0295 x -0.0027 (r = 0.9946) 1.0 2.0 3.0 4.0 Protein III (µg/mL) 0.000 0.025 0.050 0.075 0.100 0.125 Analyte Area / IS Area "Linear" Regression (1/X 2 weighting): y = 0.0142x+0.0000312 (r=0.9983) 0 50 100 150 0.0 0.4 0.8 1.2 1.6 2.0 Protein III (µg/mL) Platforms TECAN Freedom EVO Agilent’s Bravo Tips or Beads PhyTip™ Column AssayMap™ Tip Typical Volume Size for Tips 200 µL 50 µL Typical Packed Resin Bed on Tips 20 µL 5 µL Maximal capacity for antibody loading 138 µg IgG 100 µg IgG Time for One Plate 45 min > 1 h Capital Investment Existing platforms $100K Comment PhyTip™ columns are also compatible with Tecan, Hamilton and other platforms; Capable to handle other liquid transfer as well. May need other liquid handlers for other liquid handling, such as sample dilution. Protein I Protein II Poster No. M1010 Preliminary results indicated that the standard curve generated from the 4 lowest concentration levels using PhyTip™ columns were comparable with those obtained from the previously validated method using AssayMap™ cartridges.

Transcript of Poster No. M1010 Tecan Platform for LC-MS Bioanalysis of ... · Cartridge-based IC was evaluated...

Page 1: Poster No. M1010 Tecan Platform for LC-MS Bioanalysis of ... · Cartridge-based IC was evaluated for LC-MS bioanalysis of an antibody drug conjugate (ADC), Protein III in rat serum

We would like to thank Dr. Lee Hoang from PhyNexus Inc., Ms. Adrienne Chow and Mr. Bob Clark from Tecan US for their help in Tecan setup for cartridge-based immunocapture using PhyTip™ columns.

RESULTS (Continued)

Integration of Automated Bead- and Cartridge-Based Immunocapture Workflows on a GenericTecan Platform for LC-MS Bioanalysis of Protein Therapeutics and Biomarkers

Naiyu Zheng, Huidong Gu, Kristin Taylor, Yan J. Zhang, Jianing ZengAnalytical and Bioanalytical Operations, Bristol-Myers Squibb Company, Princeton, NJ 08543

PURPOSE Recently, LC–MS technology has been increasingly used for the bioanalysis of protein therapeutics and biomarkers to complement

ligand-binding assays (LBA) in drug discovery and development. Regardless of the direct mass spectrometric detection of intact proteins or indirect measurement of their tryptic surrogate peptides,

immunocapture (IC) has become the sample preparation technique of choice to achieve the best sample cleanup for ultrasensitive detection of protein analytes by LC-MS.

Current IC workflows include manual procedure (e.g. using magnetic beads), automated bead-based automated IC (e.g. Thermo Fisher Scientific’s KingFisher™ platform) and cartridge-based automated IC (e.g. AssayMap Cartridges with Agilent Bravo, MSIA tips on Thermo Scientific’s Versette platform, or PhyTip® columns on TECAN platform).

Naiyu Zheng, PhDRt. 206 & Province Line Road

Princeton, NJ 08540Phone: 609-252-5494

E-mail: [email protected]

2017 AAPS National Biotechnology Conference, May 1 - 3, 2017. San Diego, CA.

METHODS RESULTS (Continued)

New immunocapture (IC) strategy has been successfully developed by integrating and automating both bead- and cartridge-based IC on existing and generic Tecan freedom EVO® platforms for LC-MS bioanalysis of protein therapeutics and biomarkers.

This strategy allows the IC enrichment of protein analytes from diverse biological sample sizes and applications with high flexibility and high throughput without additional capital investment.

ACKNOWLEDGEMENTS

Integration of Immunocapture (IC) Workflow on a Generic Tecan Freedom EVO® Platform

Many commercially available robotic liquid handlers (RLH) specialized for IC are highly vendor-specific. These IC workflows usually require significant capital investment for new RLH and the use of costly cartridges (tips) from the same vendor. In addition, they generally can do only one IC workflow: either bead-based or cartridge-based IC.

The purpose of this study is to develop an integrated IC strategy for both bead- and cartridge-based IC on the existing Tecan freedom EVO® platforms for LC-MS bioanalysis of protein therapeutics and biomarkers for diverse applications without additional capital investment.

ASSAY 1: Bead-Based Immunocapture (IC) Used for the Quantitation of Two Modified Protein Biomarkers, Protein I and Protein II, in Monkey Blood Lysate

To improve LC-MS/MS assay sensitivity, IC of the analytes from 3 mL blood lysate was required. Due to large sample volume, IC in 96-well format is impossible. Automated sample enrichment using bead-based IC in 4-mL sample

tubes on Tecan is desirable.

Assay 2: Tip-Based IC Used for the Quantitation of An Antibody Drug Conjugate (ADC), Protein III, in Rat Serum by LC-MS/MS

LEAP 4X Ultra Pump (uHPLC)& HTC PAL autosampler

A Fluidx® XSD-48 automated capping/decapping system was used for automated capping and decapping during IC and subsequent washes.

Dynabeads® MyOne™ streptavidin T1 beads with the use of a biotinylated antibody were evaluated.

IC of protein analytes was achieved by incubation of 3 mL of monkey blood lysate with the immobilized biotinylated antibody on magnetic beads in 4-mL automation friendly Greiner Bio-One tubes (Cat No: 127261).

After IC of the protein analytes by shaking at room temperature for 2 h, followed by centrifugation, the unwanted biological matrix were transferred to the waste, and the remaining beads in the tubes were transferred into a 96-well plate on top of a MagnaBot 96 magnetic separation stand. After two PBST washes, followed by one wash with 25 mM NH4OAc with 0.05% Tween-2, the samples were used for on-bead digestion.

Quantitation of Proteins I and II was achieved by LC-MS/MS analysis of two representative surrogate peptides. For each processed sample, a 50 µL volume of the digest was injected into an a Waters CORTECS™ UPLC C18+ column (2.1 x 100 mm, 1.6 µm) and analyzed on a Triple Quad 5500 system using MRM detection for Peptide A, Peptide B and SIL-Peptide A (IS).

Assay 2: Cartridge-Based IC: LC-MS Bioanalysis of an Antibody Drug Conjugate (ADC), Protein III in Rat Serum

Assay 1: LLOQ of 0.125 nM and 0.250 nM in monkey blood lysate were achieved for Protein I and Protein II

Assay 1: In Addition to Cost-Effectiveness, Bead-based Immunocapture Using a Tecan Allows the Sensitivity Improvement from Being Able to Process More Sample Volume In Each Sample With High Throughput

Assay 2: Automated Cartridge-Based Immunocapture (IC) Workflow on Tecan Platform

Assay 2: Comparable LC-MS Responses at LLOQ Concentration Level Achieved from Immunocapture Using PhyTip™ Columns vs. AssayMap™ Cartridges

Assay 2: Comparison of Standard Curves for Protein III Obtained from Immunocapture Using Two Platforms (PhyTip™ Column on Tecan vs AssayMap™ Cartridges on Bravo)

CONCLUSIONS

AssayMap Cartridges Using Agilent’s Bravo™ Platform

MSIA Tips Using Thermo Scientific’s Versette™ Platform

Magnetic Beads Using Thermo Scientific’s KingFisher™ Platform

TECAN Freedom EVO

Dynabeads® Streptavidin

On bead digestionwith trypsin

Time (Min)

Rel. A

bund

ancePeptide A Peptide B

LC-MS/MS

Peptide A

Magnetic stand

Capture antibodybiotinylated

Protein I

Protein II

ImmobilizedAntibody

Buffer wash

Buffer wash

Interference Proteins

blood lysate

Immuno-Capture (IC)

Bead-Based ImmunocaptureOn TECAN Platform

Enrich Protein I, and II

Peptide A

IC-LC-MS/MS for the bioanalysis of Protein III (ADC) in rat plasma was previously developed and validated using an Agilent Bravo™ platform. Evaluation of new IC workflow using PhyTip® columns on an existing TECAN has several advantages, such as cost saving in capital investment of new IC platform, less expensive with the tips, more flexibility and capability with TECAN than usingstandalone IC platforms.

Capture analyte Elute

PhyTip™ columns

resin

Time (Min)Rel. A

bund

ance

LC-MS/MS

TrypsinDigestion

Serum Samples

BufferWash

Immobilizeantibody

Enriched Protein III

BufferWash

Aspi

rate

Disp

ense

Protein III (ADC)

Interference Proteins

PhyTip™Streptavidin ResinCapture antibodybiotinylated

Tip-Based ImmunocaptureOn TECAN Platform

Assay 1: Bead-Based IC: LC-MS Bioanalysis of Two Modified Protein Biomarkers, Protein I and Protein II that Derived from an Endogenous Protein (MW: 90 KDa) in Monkey Blood Lysate.

PhyTip® streptavidin columns (PhyNeXus, Inc) (20 µL resin bed size on 200-µL Tecan tips) with the use of another biotinylated antibody were evaluated.

Cartridge-based IC was evaluated for LC-MS bioanalysis of an antibody drug conjugate (ADC), Protein III in rat serum (15 µL serum + 145 µL PBS).

The IC was performed by aspiration and dispensing of the serum samples through PhyTip® columns for 16 cycles (4.2 µL/sec, 10 s delay), followed by a wash prior to acid elution and trypsin digestion.

Quantitation of Protein III was achieved by LC-MS/MS analysis of a surrogate peptide. For each processed sample, a 40 µL volume of the digest was injected into an a Waters Acquity™ UPLC HSS T3 column (2.1 x 50 mm, 1.8 µm) and analyzed on a Triple Quad 6500 system with MRM detection for Peptide C and the IS (SIL-Peptide C), respectively.

RESULTSAssay 1: Automation of Bead-Based IC Workflow on TECAN Platform

Magnetic stand

Blood lysateSamples3 mL/Tube

Add capture antibodyon beadsShake at RT for 2 h (IC)

Centrifugation

Decap

Discard top supernatanton TECAN

Transfer beads to 96-well plate

LC-MS/MS

On bead digestion at 50°C for 30 min

Wash with PBST

Automated decapper TECAN Freedom EVO

ThawCentrifugation Immunocapture (2h)

Trypsin Digestion (0.5h)

Blood Lysate 3 mL

Waste lysate

TECAN Freedom EVO

Washed with buffer three times

Assay 1: Representative Chromatograms for the Modified Protein Biomarker, Protein I and Protein II

Blank

0200400600 5.11

LLOQ at 0.250 nMof Protein I

0250500750 5.02

SIL-Peptide B (IS)

4.0 4.5 5.0 5.5 6.0Time, min

0.0

2.0e4

4.0e44.82

0100200300

LLOQat 0.125 nM of Protein II

0200400600800 4.83

4.60

SIL-Peptide B (IS)

4.0 4.5 5.0 5.5 6.00.0

2.0e4

4.0e4 4.82

5.11

Intensity, cps

Blank

Protein I Protein II

"Quadratic" Regression ("1 / x" weighting) (r = 0.9995)

2.0 4.0 6.0 8.0 10.0 12.0

Protein II Conc (nM)

0.00

0.20

0.40

0.60Protein II(0.125 – 12.5 nM)

"Quadratic" Regression ("1 / x" weighting) (r = 0.9986)

2.0 4.0 6.0 8.0 10.0 12.00.00

0.10

0.20

0.30

0.40

Analyte Area / IS Area

Protein I(0.250-12.5 nM)

Protein I Conc (nM)

Analyte Area / IS Area

One of the approaches to improve LC-MS/MS assay sensitivity for protein quantitation is to enrich protein analytes from a large sample volume using IC. However, when the volume is larger than the capacity of the 96-well format, it could be very challenging.

By using a Tecan Freedom EVO platform with a Fluidx® XSD-48 automated capping/ decapping system, sample preparation throughput was increased from 32 samples per batch performed manually to 96 samples per batch when performed on Tecan.

Platforms TECAN Freedom EVO Manual

Sample Racks

2 Fluidx® XSD-48 Racks 4 DynaMag™-5 Magnet Racks

Sample Size Per Process 96 Samples 32 Samples

Time for 96 Samples 1 day (< 4 hours) 3 days (3 separate batches are required)

Productivity Improvement 3-fold N/A

Comment Existing TECAN platform; Capable to handle all liquid transfer as well. N/A

PhyTip® columns PhyTip® columns are compatible with Tecan Freedom EVO

Step 1 Load PhyTip™ columns: 200-µL PhyTip™ columns packed with 20-µL of Streptavidin resin

Step 2 Equilibration: 200 µL PBS (pH 7.4, 10 mM), 4 cycles.

Step 3 Load antibody: 100 µL of 100 µg/mL, 8 cycles.

Step 4 Wash: 200 µL PBS (pH 7.4, 10 mM), 2 cycles.

Step 5 Capture from Serum: 160 µL (15 µL serum + 145 µL PBS), 16 cycles.

Step 6 Wash: 200 µL PBS (pH 7.4, 10 mM), 2 cycles.

Step 7 Elute: 40 µL 30% ACN with 0.1% formic acid, 4 cycles.

Step 8 Neutralize: 10 µL of 1 M NH4HCO3 Reduction/alkylation Trypsin Digestion.

PhyTip™ Columns on Tecan AssayMap™ Cartridges on Bravo

IS

Max 66210.5 ug/mL

Protein III (LLOQ)

Max 4.9e5

IS

Max 6041

Max 4.5e5

0.5 ug/mLProtein III (LLOQ)

Assay 2: Immunocapture Using PhyTip™ on Tecan Freedom EVO Is More Cost-Effective than Agilent’s Bravo When New Capital Investment is not Required

PhyTip™ on Tecan on Tecan (Only test 4 points) AssayMap™ on Bravo

Tecan (0.5 -4 ug/mL)

"Linear" Regression (1/X2 weighting): y = 0.0295 x -0.0027 (r = 0.9946)

1.0 2.0 3.0 4.0Protein III (µg/mL)

0.000

0.025

0.050

0.075

0.100

0.125

Analyte Area / IS Area

"Linear" Regression (1/X2 weighting): y = 0.0142x+0.0000312 (r=0.9983)

0 50 100 1500.0

0.4

0.8

1.2

1.6

2.0

Protein III (µg/mL)

Platforms TECAN Freedom EVO Agilent’s Bravo

Tips or Beads

PhyTip™ Column AssayMap™ Tip

Typical Volume Size for Tips 200 µL 50 µL

Typical Packed Resin Bed on Tips 20 µL 5 µL

Maximal capacity for antibody loading 138 µg IgG 100 µg IgG

Time for One Plate 45 min > 1 h

Capital Investment Existing platforms $100K

CommentPhyTip™ columns are also compatible with Tecan,

Hamilton and other platforms; Capable to handle other liquid transfer as well.

May need other liquid handlers for other liquid handling, such as sample dilution.

Protein I Protein II

Poster No.

M1010

Preliminary results indicated that the standard curve generated from the 4 lowest concentration levels using PhyTip™ columns were comparable with those obtained from the previously validated method using AssayMap™ cartridges.