Technical specification FLUXUS G601 TSFLUXUS_G601V2-3US_Lus, 2019-05-01 Features • Precise bidirectional and highly dynamic flow measurement with the non-invasive clamp-on technology • High precision at fast and slow flow rates, high temperature and zero point stability • Portable, easy-to-use flow transmitter with 2 flow channels, multiple inputs/outputs, an integrated data logger with a serial interface • Water and dust-tight (NEMA 4); resistant against oil, many liquids and dirt • Li-Ion battery provides up to 25 hours of measurement opera- tion • Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results • User-friendly design • Transducers available for a wide range of inner pipe diameters and fluid temperatures • Probe for wall thickness measurement available • Robust, water-tight (NEMA 4) transport case with comprehen- sive accessories • QuickFix for fast mounting of the flow transmitter in difficult conditions • Including measurement of liquids Applications Designed for industrial use in harsh environments, in gas pro- cessing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications: • Measurement on natural gas pipelines and in natural gas stor- age installations • Measurement of synthesized gas and injection gas • Measurement for the gas supply industry • Supervision of permanently installed meters, service and main- tenance FLUXUS G601 Measurement equipment in transport case Portable ultrasonic flow measurement of gas and liquids Portable instrument for non-invasive, quick ultrasonic flow measurement with clamp-on technology for all types of piping

Transcript of Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty...

Page 1: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification

FLUXUS G601

TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Features

• Precise bidirectional and highly dynamic flow measurement with the non-invasive clamp-on technology

• High precision at fast and slow flow rates, high temperature and zero point stability

• Portable, easy-to-use flow transmitter with 2 flow channels, multiple inputs/outputs, an integrated data logger with a serial interface

• Water and dust-tight (NEMA 4); resistant against oil, many liquids and dirt

• Li-Ion battery provides up to 25 hours of measurement opera-tion

• Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results

• User-friendly design

• Transducers available for a wide range of inner pipe diameters and fluid temperatures

• Probe for wall thickness measurement available

• Robust, water-tight (NEMA 4) transport case with comprehen-sive accessories

• QuickFix for fast mounting of the flow transmitter in difficult conditions

• Including measurement of liquids

Applications

Designed for industrial use in harsh environments, in gas pro-cessing and natural gas extraction, chemical industry and in thepetroleum industry. Practical applications:

• Measurement on natural gas pipelines and in natural gas stor-age installations

• Measurement of synthesized gas and injection gas

• Measurement for the gas supply industry

• Supervision of permanently installed meters, service and main-tenance

FLUXUS G601

Measurement equipment in transport case

Portable ultrasonic flow measurement of gas and liquids

Portable instrument for non-invasive, quick ultrasonic flow measurement with clamp-on technology for all types of piping

Page 2: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus2

Table of contents

Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Calculation of volumetric flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Number of sound paths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Typical measurement setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Standard volumetric flow rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Standard scope of supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Example for the equipment of a transport case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Transducer selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Transducer order code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Transducer mounting fixture. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Coupling materials for transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Damping material (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Damping mats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Damping coat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Connection systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Clamp-on temperature probe (optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Fixation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Wall thickness measurement (optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Page 3: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

3TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Function

Measurement principleThe transducers are mounted on the pipe which is completely filled with the fluid. The ultrasonic signals are emitted alter-nately by a transducer and received by the other. The physical quantities are determined from the transit times of the ultra-sonic signals.

As the fluid where the ultrasound propagates is flowing, the transit time of the ultrasonic signal in flow direction is shorterthan the one against the flow direction.

The transit time difference ∆t is measured and allows the flowmeter to determine the average flow velocity along the prop-agation path of the ultrasonic signals. A flow profile correction is then performed in order to obtain the area averaged flowvelocity, which is proportional to the volumetric flow rate.

The integrated microprocessors control the entire measuring cycle. The received ultrasonic signals are checked for mea-surement usability and evaluated for their reliability. Noise signals are eliminated.

Calculation of volumetric flow rate = kRe · A · ka ·

where

Path of the ultrasonic signal in the flowing fluid

Transit time difference ∆t

- volumetric flow rate

kRe - fluid mechanics calibration factor

A - cross-sectional pipe area

ka - acoustical calibration factor

∆t - transit time difference

tγ - average of transit times in the fluid

∆tt0

V· t2 t-----------

Page 4: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus4

Number of sound pathsThe number of sound paths is the number of transits of the ultrasonic signal through the fluid in the pipe. Depending on thenumber of sound paths, the following methods of installation exist:

• reflect arrangementThe number of sound paths is even. The transducers are mounted on the same side of the pipe. Correct positioning of the transducers is easier.

• diagonal arrangementThe number of sound paths is odd. The transducers are mounted on opposite sides of the pipe.

• direct modeDiagonal arrangement with 1 sound path. This should be used in the case of a high signal attenuation by the fluid, pipe or coatings.

The preferred method of installation depends on the application. While increasing the number of sound paths increasesthe accuracy of the measurement, signal attenuation increases as well. The optimum number of sound paths for the pa-rameters of the application will be determined automatically by the transmitter.

As the transducers can be mounted with the transducer mounting fixture in reflect arrangement or diagonal arrangement,the number of sound paths can be adjusted optimally for the application.

a - transducer distance

Typical measurement setup

Reflect arrangement, number of sound paths: 2

Diagonal arrangement, number of sound paths: 3

Direct mode, number of sound paths: 1 Direct mode, number of sound paths: 1, negative transducer distance

Example of a reflect arrangement with connection of the inputs to an external process pressure and process temperature measurement for standard volumetric flow rate calculation

a

a

a > 0 a < 0

transducers

damping mat

transmitter

inputs

temperature probee.g., external pressure sensor

Page 5: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

5TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Standard volumetric flow rateThe standard volumetric flow rate can be selected as physical quantity to be measured. It will be calculated internally by:

N = · · ·

where

The operational pressure p and the operational temperature T of the fluid will be entered directly as fixed values into thetransmitter.

or:

If inputs are installed (optional), pressure and temperature can be measured by the customer and fed in the transmitter.

The gas compressibility coefficient K of the gas is entered in the transmitter:

• as fixed value or

• as approximation according to e.g., AGA8 or GERG

N - standard volumetric flow rate

- operating volumetric flow rate

pN - standard pressure (absolute value)

p - operating pressure (absolute value)

TN - standard temperature in K

T - operating temperature in K

K compressibility coefficient of the gas: ratio of the compressibility factors of the gas at operating conditions andat standard conditions Z/ZN

V· V· ppN-----

TN

T----- 1

K---

Page 6: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus6

Transmitter

Technical data

FLUXUS G601

design portablemeasurementmeasurement principle

transit time difference correlation principle

flow velocity ft/s 0.03 to 115 ft/s, depending on pipe diameterrepeatability 0.15 % of reading ±0.02 ft/sfluid all acoustically conductive gases,

e.g., nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propanetemperature com-pensation

corresponding to the recommendations in ANSI/ASME MFC-5.1-2011

measurement uncertainty (volumetric flow rate)measurement uncer-tainty of measuring system1

±0.3 % of reading ±0.02 ft/sincludes calibration certificate traceable to NISTcalibration facility ISO 17025 accredited

measurement uncer-tainty at the measu-ring point

±1 to 3 % of reading ±0.02 ft/s, contact FLEXIM for an application specific uncertainty evaluation

transmitterpower supply • 100 to 230 V/50 to 60 Hz (power supply unit)

• 10.5 to 15 V DC (socket at transmitter)

• integrated batteryintegrated battery Li-Ion, 7.2 V/6.2 Ah• operating time h • > 14 h (without outputs, inputs and backlight)

• > 25 h (1 measuring channel, ambient temperature > 50 °F, without outputs, inputs and backlight)power consumption W < 6 (with outputs, inputs and backlight), charging: 18number of measuring channels

2

damping s 0 to 100 (adjustable)measuring cycle Hz 100 to 1000 (1 channel)response time s 1 (1 channel), option: 0.07housing material PA, TPE, AutoTex, stainless steeldegree of protection NEMA 4dimensions in see dimensional drawingweight lb 4.6fixation QuickFix pipe mounting fixtureambient temperature °F 14 to 140display 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanishmeasuring functionsphysical quantities operating volumetric flow rate, standard volumetric flow rate, mass flow rate, flow velocitytotalizer volume, masscalculation functions average, difference, sumdiagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit timescommunication interfacesservice interfaces • RS232

• USB (with adapter)process interfaces • Modbus RTU (optional)accessoriesserial data kit• сable RS232• adapter RS232 - USBsoftware • FluxDiagReader: download of measured values and parameters, graphical presentation

• FluxDiag (optional): download of measurement data, graphical presentation, report generation

• FluxSubstanceLoader: upload of fluid data setsadapter • output adapter (if number of outputs > 4)

• input adapter (if number of inputs > 2)transport case dimensions: 19.7 x 15.7 x 7.5 indata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured values1 with aperture calibration of the transducers

For the technical data in the flow measurement of liquids mode see Technical specification TSFLUXUS_F601V*-*.

Page 7: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

7TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Dimensions

outputsThe outputs are galvanically isolated from the transmitter.

number see standard scope of supply, max. on request• current outputrange mA 0/4 to 20accuracy 0.1 % of reading ±15 μAactive output Rext < 750 Ω (Uint = 24 V DC)passive output Uext = 4 to 16 V, depending on Rext (Rext < 500 Ω at 16 V)• frequency outputrange kHz 0 to 5open collector 24 V/4 mA• binary outputoptorelay 26 V/100 mAbinary output as alarm output• functions limit, change of flow direction or errorbinary output as pulse output• functions mainly for totalizing• pulse value units 0.01 to 1000• pulse width ms 1 to 1000inputs

The inputs are galvanically isolated from the transmitter.number see standard scope of supply, max. 4• temperature inputtype Pt100/Pt1000connection 4-wirerange °F -238 to +1040resolution K 0.01accuracy ±0.01 % of reading ±0.03 K• current inputaccuracy 0.1 % of reading ±10 μApassive input Rint = 50 Ω, Pint < 0.3 W• range mA -20 to +20• voltage inputrange V 0 to 1accuracy 0.1 % of reading ±1 mVinternal resistance Rint = 1 MΩ

*60*

FLUXUS G601

1 with aperture calibration of the transducers

For the technical data in the flow measurement of liquids mode see Technical specification TSFLUXUS_F601V*-*.

CHANNEL BCHANNEL A

BATTERY

1 20

4 5 67

SNAPQ- Q+

MUX

NEXT

9DISP

3DISP

–MODEFAST

.LIGHT

3x OFF

3x QOFF

8QON

ENTER ENTER

BRKON

C

8.9

2.3

28.

39

in inch

Page 8: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus8

Standard scope of supply

Adapters

G601 Standard G601 Extended Standard G601 Multi-functional G601 CA-Energy

application flow measurement of gas flow measurement of com-pressed air, industrial gases and liquids

2 independent measuring channelscalculation of standard volu-metric flow rate

calculation of standard volumetric flow rate, with optional use of current measured pressure and temperature valuesincluding energy calculator for BTU and heat measurements

simultaneous monitoring of 2 energy flows

liquids: integrated thermal en-ergy computer for monitoring of energy flows

outputspassive current output 2 2 2 2 2 2 2 2binary output 2 1 1 1 2 2 2 1frequency output - - 1 - 1 0 - -Modbus - x - x - x - xinputstemperature input - - - - 1 1 2 2passive current input - - 2 2 2 2 2 2voltage input - - - - 1 1 - -accessoriestransport case x x x x x x x xpower supply unit, mains cable

x x x x x x x x

battery x x x x x x x xoutput adapter - - - - x x - -input adapter - - 2 2 2 2 2 2adapter for voltage and current inputs

- - - - 3 3 2 2

QuickFix pipe mounting fixture for transmitter

x x x x x x x x

serial data kit x x x x x x x xmeasuring tape x x x x x x x xwall thickness probe - - - - x x x xuser manual,Quick start guide

x x x x x x x x

connector board at the up-per side of the transmitter

*601

output adapter

input adapter

adapter for voltage and current inputs

transducersmeasuring channel A

outputs inputs

transducersmeasuring channel B

power supply unit/battery charging unitRS232

Page 9: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

9TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Example for the equipment of a transport case

G601

software

power supply unit,mains cable

measuring tape

transmittertransducers

user manual,Quick start guide

coupling compound

wall thickness probe (optional)

QuickFix pipe mounting fixture

temperature probes (optional)

serial data kit

transducer mounting fixture

Page 10: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus10

Transducers

Transducer selection

Step 1a

Select a Lamb wave transducer:

Step 1b

If the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:

Step 2

inner pipe diameter d dependent on the flow velocity v of the fluid in the pipe

The transducers are selected from the characteristics (see next page). Lamb wave transducers are selected from the leftcolumn, shear wave transducers from the right column.

Lamb wave transducers: If the values d and v are not in the range, the diagonal arrangement with 1 sound path may beused, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the values are still not in therange, shear waves transducers regarding the pipe wall thickness have to be selected in step 1b.

transducer order code

GLF 0.59 1.3 1.4

GLG 0.43 0.94

GLH 0.31 0.63

GLK 0.2 0.39

GLM 0.1 to 0.2

GLP 0.05 to 0.12

GLQ 0.02 to 0.05

0.2 0.4 0.6 0.8 1 1.2 1.4pipe wall thickness [in]

transducer order code

GSG 0.43

GSK 0.2

GSM 0.1

GSP 0.05

GSQ 0.02

0.2 0.4 0.6 0.8 1 1.2 1.4pipe wall thickness [in]

recommended possible

Page 11: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

11TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Lamb wave transducer1 shear wave transducer1

GLF

GLG GSG

GLH

GLK GSK

GLM GSM

GLP GSP

GLQ GSQ

1 inner pipe diameter and max. flow velocity for a typical application with natural gas, nitrogen, oxygen in reflect arrangement with 2 sound paths (Lamb wave transducers)/1 sound path (shear wave transducers)

Page 12: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus12

Step 3

min. fluid pressure

Example

Step 4

for the characters 4 to 11 of the transducer order code (ambient temperature, explosion protection, connection system, ex-tension cable) see page 13

Step 5

for the technical data of the selected transducer see page 14 et seqq.

Lamb wave transducer shear wave transducer

transducer or-der code

fluid pressure1 [psi] transducer or-der code

fluid pressure1 [psi]metal pipe plastic pipe metal pipe plastic pipemin. min. extended min. min. min. extended min.

GLF 218 145 15GLG 218 145 15 GSG 435 290 15GLH 218 145 15GLK 218 (d > 4.7 in)

145 (d < 4.7 in)145 (d > 4.7 in)44 (d < 4.7 in)

15 GSK 435 290 15

GLM 145 (d > 2.4 in)73 (d < 2.4 in)

44 (d < 2.4 in) 15 GSM 435 290 15

GLP 145 (d > 1.4 in)73 (d < 1.4 in)

44 (d < 1.4 in) 15 GSP 435 290 15

GLQ 145 (d > 0.59 in)73 (d < 0.59 in)

44 (d < 0.59 in) 15 GSQ 435 290 15

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air

d = inner pipe diameter

step

1 pipe wall thickness in 0.56 0.34 1.5selected transducer GLG or GLH GLH or GLK GS

2 inner pipe diameter in 22.9 3.8 5.6max. flow velocity ft/s 49 98 98selected transducer GLG GLK GSK

3 min. fluid pressure psi 290 218 580selected transducer GLG GLK GSK

Page 13: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

13TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Transducer order code

1, 2 3 4 5, 6 7, 8 9 to 11 no. of character

tra

ns

du

ce

r

tra

ns

du

ce

r fr

equ

ency

- am

bie

nt

tem

pe

ratu

re

exp

losi

on

pro

tec

tio

n

co

nn

ec

tio

n s

yste

m

- ex

ten

sio

n c

ab

le

/ op

tio

n

de

sc

rip

tio

n

GS set of ultrasonic flow transducers for gas measurement, shear waveGL set of ultrasonic flow transducers for gas measurement, Lamb wave

F 0.15 MHzG 0.2 MHzH 0.3 MHzK 0.5 MHzM 1 MHzP 2 MHzQ 4 MHz

N normal temperature rangeE extended temperature range

NN not explosion proofNL with Lemo connector

XXX 0 m: without extension cable> 0 m: with extension cable

LC long transducer cable

Page 14: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus14

Technical data

Shear wave transducers (nonEx, NL)

order code GSG-NNNNL/** GSK-NNNNL/** GSM-NNNNL/** GSP-NNNNL/** GSQ-NNNNL/**technical type G(DL)G1NZ7 G(DL)K1NZ7 G(DL)M1NZ7 G(DL)P1NZ7 G(DL)Q1NZ7transducer frequency MHz 0.2 0.5 1 2 4fluid pressure1

min. extended psi metal pipe: 290min. psi metal pipe: 435, plastic pipe: 15inner pipe diameter d2

min. extended in 7.1 2.4 1.2 0.59 0.28min. recommended in 8.7 3.1 1.6 0.79 0.39max. recommended in 35.4 11.8 5.9 2 0.87max. extended in 43.3 14.2 7.1 2.4 1.2pipe wall thicknessmin. in 0.43 0.2 0.1 0.05 0.02materialhousing PEEK with stainless steel cap 304 stainless steel 304contact surface PEEK PEEKdegree of protection NEMA 6transducer cabletype 1699length ft 16 13 9length (***-*****/LC) ft 29dimensionslength l in 5.1 4.98 2.36 1.67width b in 2.01 2.01 1.18 0.71height h in 2.64 2.66 1.32 0.85dimensional drawing

weight (without cable)

lb 1 0.79 0.08 0.03

pipe surface temperaturemin. °F -40max. °F +266ambient temperaturemin. °F -40max. °F +266temperature com-pensation

x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other fluids on requestinner pipe diameter max. recommended/max. extended: in reflect arrangement and for a flow velocity of 49 ft/s

l

hb

l

hb

l

hb

l

hb

Page 15: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

15TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Shear wave transducers (nonEx, NL, extended temperature range)

order code GSM-ENNNL/** GSP-ENNNL/** GSQ-ENNNL/**technical type G(DL)M1EZ7 G(DL)P1EZ7 G(DL)Q1EZ7transducer frequency MHz 1 2 4fluid pressure1

min. extended psi metal pipe: 290min. psi metal pipe: 435, plastic pipe: 15inner pipe diameter d2

min. extended in 1.2 0.59 0.28min. recommended in 1.6 0.79 0.39max. recommended in 5.9 2 0.87max. extended in 7.1 2.4 1.2pipe wall thicknessmin. in 0.1 0.05 0.02materialhousing stainless steel 304contact surface Sintimiddegree of protection NEMA 4transducer cabletype 1699length ft 13 9length (***-*****/LC) ft 29dimensionslength l in 2.36 1.67width b in 1.18 0.71height h in 1.32 0.85dimensional drawing

weight (without cable)

lb 0.09 0.02

pipe surface temperaturemin. °F -22max. °F +392ambient temperaturemin. °F -22max. °F +392temperature com-pensation

x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other fluids on requestinner pipe diameter max. recommended/max. extended: in reflect arrangement and for a flow velocity of 49 ft/s

l

hb

l

hb

Page 16: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus16

Lamb wave transducers

Lamb wave transducers (nonEx, NL)

order code GLF-NNNNL GLG-NNNNL GLH-NNNNL GLK-NNNNL GLM-NNNNL GLP-NNNNL GLQ-NNNNL technical type G(RT)F1NC3 G(RT)G1NC3 G(RT)H1NC3 G(RT)K1NC3 G(RT)M1NC3 G(RT)P1NC3 G(RT)Q1NC3transducer frequency MHz 0.15 0.2 0.3 0.5 1 2 4fluid pressure1

min. extended psi metal pipe: 145 metal pipe:145 (d > 4.7 in)44 (d < 4.7 in)

metal pipe:44 (d < 2.4 in)

metal pipe:44 (d < 1.4 in)

metal pipe:44 (d < 0.59 in)

min. psi metal pipe: 218plastic pipe: 15

metal pipe:218 (d > 4.7 in)145 (d < 4.7 in)plastic pipe: 15

metal pipe:145 (d > 2.4 in)73 (d < 2.4 in)plastic pipe: 15

metal pipe:145 (d > 1.4 in)73 (d < 1.4 in)plastic pipe: 15

metal pipe:145 (d > 0.59 in)73 (d < 0.59 in)plastic pipe: 15

inner pipe diameter d2

min. extended in 8.7 7.1 4.3 2.4 1.2 0.59 0.28min. recommended in 10.6 8.7 5.5 3.1 1.6 0.79 0.39max. recommended in 47.2 35.4 23.6 11.8 5.9 2 0.87max. extended in 63 55.1 39.4 14.2 7.1 2.4 1.2pipe wall thicknessmin. in 0.59 0.43 0.31 0.2 0.1 0.05 0.02max. in 1.3 0.94 0.63 0.39 0.2 0.12 0.05max. extended 1.4 - - - - - -materialhousing PPSU with

stainless steel cap 316Ti

PPSU with stainless steel cap 304

contact surface PPSUdegree of protection NEMA 4transducer cabletype 1699length ft 16 13 9length (***-*****/LC) ft 29dimensionslength l in 6.42 5.06 2.91 1.65width b in 2.13 2.01 1.26 0.87height h in 3.59 2.66 1.59 1dimensional drawing

weight (without cable)

lb 2.1 1 0.17 0.04

pipe surface temperaturemin. °F -40max. °F +302 +338ambient temperaturemin. °F -40max. °F +302 +338temperature com-pensation

x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other fluids on requestinner pipe diameter max. recommended: in reflect arrangement (diagonal arrangement) and for a flow velocity of 49 ft/s (98 ft/s)inner pipe diameter max. extended: in reflect arrangement (diagonal arrangement) and for a flow velocity of 39 ft/s (82 ft/s)

l

h

l

hb

lh

bl

hb

Page 17: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

17TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Transducer mounting fixture

Order code

1, 2 3 4 5 6 7 to 9 no. of charactertr

ansd

uce

r m

ou

nti

ng

fi

xtu

re

tra

ns

du

ce

r

- mea

su

rem

ent

arr

an

gem

en

t

siz

e

- fix

ati

on

ou

ter

pip

e d

iam

ete

r

de

scr

ipti

on

VP portable VariofixTB tension belts

A all transducersD reflect arrangement or diagonal arrangement/direct modeR reflect arrangement

S smallM medium

C chainsG tension beltsN without fixation

055 0.39 to 21.7 in150 2 to 59.1 in210 2 to 82.7 in

portable Variofix VP and chains (optional)

material: stainless steel 304, 301, 303

dimensions: 16.3 x 3.7 x 2.99 in

chain length: 6 ft

portable Variofix VP and tension belts

tension belts TB

material: steel, powder coated and textile tension belt

length: 16/22 ft

ambient temperature: max. 140 °F

outer pipe diameter:max. 59.1/82.7 in

(optional)

Page 18: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus18

Coupling materials for transducers

Technical data

normal temperature range(4th character of transducer order code = N)

extended temperature range(4th character of transducer order code = E)

< 212 °F < 338 °F < 302 °F < 392 °Fcoupling compound type N

coupling compound type E

coupling compound type E

coupling compoundtype E or H

type ambient temperature

°Fcoupling compound type N -22 to +266

coupling compound type E -22 to +392

coupling compound type H -22 to +482

Page 19: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

19TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Damping material (optional)Damping material will be used for the gas measurement to reduce acoustic noise influences on the measurement.

Damping matsDamping mats will be used for the gas measurement to reduce acoustic noise influences on the measurement.

Transducer damping mats will be installed below the transducers.

Pipe damping mats will be installed at reflection points, e.g., flange, weld.

Selection of damping mats

Quantity for pipe damping mat - type A

(depending on the outer pipe diameter)

Length of pipe damping mat - type B

(length l depending on transducer frequency and outer pipe diameter)

damping coat

damping mat

-22 0 +176 +392 fluid temperature [°F]

type description outer pipe dia-meter

dimensionsl x b x h

transducer frequency technical type

ambient tempe-rature

remark

in in F G H K M P Q °F

transducer damping mat

D for temporary installation (multip-le use), fixed with coupling com-pound

< 3.1 17.72 x 4.53 x 0.02 - - - - x x x D20S3 -13 to +140≥ 3.1 35.43 x 9.06 x 0.02 - - - x x - - D20S2

35.43 x 9.06 x 0.05 x x x - - - - D50S2

pipe damping mat

A for temporary installation (multip-le use), fixed with coupling com-pound

< 11.8 11.81 x 4.53 x 0.02 x x x x x x x A20S4 -13 to +140 for quantity see ta-ble below

B self-adhesive ≥ 11.8 l x 3.94 x 0.04 x x x x x x - B35R2 -31 to +122 l - see table below

outer pipe diameter D transducer frequency

in F, G, H K, M, P, Q3.9 12 67.9 24 1211.8 32 16

outer pipe diameter D transducer frequency

F, G, H K, M, Pin ft ft11.8 39 1919.7 104 5239.4 413 206

b

lD

transducer damping mat pipe damping mat

reflect arrangementdiagonal arrangement/direct mode

D = outer pipe diameter

reflection point

Page 20: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus20

Damping coatFor high temperatures it is recommended to apply the damping coat onto the pipe.

Technical data

Dimensioning

material multipolymeric matrix/inorganic ceramic coatingpacking drum gal 1properties heat resistant, inert

transducer number of packing drums

outer pipe diameter

≤15.7 ≤23.6 ≤31.5in

F 1 2 2G 1 1 2H 1 1 1K 1 - -M 1 - -P 1 - -Q 1 - -

Page 21: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

21TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Connection systems

Cable

Cable length

connection system NL

direct connection/connection with extension cable transducerstechnical type*****Z7*****C3

transducer cable

type 1699weight lb/ft 0.06ambient temperature °F -67 to +392cable jacketmaterial PTFEouter diameter in 0.11thickness in 0.01color brownshield xsheathmaterial stainless steel 304outer diameter in 0.31

extension cable

type 1750 2551standard length ft 16

32-

max. length ft 32 see table belowweight lb/ft 0.08 0.06ambient temperature °F < 144 -13 to +176cable jacketmaterial PE TPE-Oouter diameter in 0.24 0.31thickness in 0.02color black blackshield x xsheathmaterial stainless steel 304 -outer diameter in 0.35 -remark optional

transducer frequency

F, G, H, K M, P Q S

connection system NLtransducerstechnical type

x y l x y l x y l x y l

*D***Z71

*R***C31ft 6 9 ≤ 82 6 6 ≤ 82 6 3 ≤ 82 3 3 ≤ 65

option LC:*L***Z71

*T***C31

ft 6 22 ≤ 82 22 6 ≤ 82 26 3 ≤ 82 - - -

1 l > 82 to 328 ft on request

x, y = transducer cable length

l = max. length of extension cable

tra

nsm

itte

r

l x y

Page 22: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus22

Clamp-on temperature probe (optional)

Technical data

PT13N

design clamp-onwith connector

type Pt1000connection 4-wiremeasuring range °F -40 to +392accuracy T ±(0.27 °F + 2 . 10-3 . (|T [°F]| -

32 °F))class A

accuracy ∆T(2x Pt matched according to EN 1434-1)

≤ 0.03 °F (at 50 °F)

housing 360 brass alloydegree of protection NEMA 4dimensionslength l in 0.79width b in 0.59height h in 0.49dimensional drawing

weight lb 0.437 (without connector)accessoriesthermal conductivity foil 482 °F

x

PT13F

design clamp-onshort response time,with connector

type Pt1000connection 4-wiremeasuring range °F -58 to +482accuracy T ±(0.27 °F + 2 . 10-3 . (|T [°F]| -

32 °F))class A

accuracy ∆T(2x Pt matched according to EN 1434-1)

≤ 0.1 K (3 K < ∆T < 6 K), more corresponding to EN 1434-1

response time s 8housing PEEK, stainless steel 304, cop-

perdegree of protection NEMA 4dimensionslength l in 0.55width b in 1.18height h in 1.06dimensional drawing

weight lb 0.7 (without connector)accessoriesthermal conductivity paste 392 °F

x

thermal conductivity foil 482 °F

x

plastic protection plate, insulation foam

x

Connection system

Connection

Cable

direct connection/connection with extension cable

temperature probe extension cable connector

pin

red black 2

red green 6

white white 1

white red 7

temperature probe extension cable

type 4 x 24 AWG 4 x 18 AWGstandard length ft 20 -max. length ft - 656cable jacket PTFE LS PVC

extension cable

h

lb

Connection system

Connection

Cable

direct connection/connection with extension cable

temperature probe extension cable connector

pin

red black 2

red/blue green 6

white/blue white 1

white red 7

temperature probe extension cable

type 4 x 0.25 mm² black 4 x 18 AWGstandard length ft 9 -max. length ft - 656cable jacket PTFE LS PVC

extension cable

h

bl

Page 23: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

Technical specification FLUXUS G601

23TSFLUXUS_G601V2-3US_Lus, 2019-05-01

Fixation

tension strap PT13N

material: stainless steel 301, 410

ball chain PT13F

material: stainless steel 316L

length: 3 ft

Page 24: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

FLUXUS G601 Technical specification

2019-05-01, TSFLUXUS_G601V2-3US_Lus24

Wall thickness measurement (optional)The pipe wall thickness is an important pipe parameter which has to be determined exactly for a good measurement. How-ever, the pipe wall thickness often is unknown.

The wall thickness probe can be connected to the transmitter instead of the flow transducers and the wall thickness mea-surement mode is activated automatically.

Acoustic coupling compound is applied to the wall thickness probe which then is placed firmly on the pipe. The wall thick-ness is displayed and can be stored directly in the transmitter.

Technical data

Cable

DWR1NZ7

measuring range1 in 0.04 to 9.8resolution in 0.0004accuracy 1 % ±0.004 influid temperature °F -4 to +392,

short-time peak max. 932сabletype 2616length ft 41 The measuring range depends on the attenuation of the ultrasonic signal in the pipe. For strongly attenuating plastics (e.g., PFA, PTFE, PP) the measuring

range is smaller.

2616

ambient temperature °F <392cable jacketmaterial FEPouter diameter in 0.2color blackshield x

DWR1NZ7

Page 25: Portable ultrasonic flow measurement of gas and liquids · 2019-05-06 · measurement uncertainty (volumetric flow rate) measurement uncer-tainty of measuring system1 ±0.3 % of reading

2019-05-01, TSFLUXUS_G601V2-3US_Lus

FLEXIM AMERICAS CorporationEdgewood, NY 11717

USA

Tel.:(631) 492-2300Fax:(631) 492-2117

internet: www.flexim.come-mail: [email protected]

1-888-852-7473

Subject to change without notification. Errors excepted.FLUXUS is a registered trademark of FLEXIM GmbH.

Copyright (©) FLEXIM GmbH 2019