Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual...
Transcript of Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual...
![Page 1: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/1.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 1 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM Seegene
Prepared by QA Committee
Issued by: Laboratory Manager Revision Date: 4/26/2021
Approved by Laboratory Director:
Microbiologist-in-Chief
Next Review Date:
Uncontrolled When Printed
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY " in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
TABLE OF CONTENTS
Introduction ..................................................................................................................................... 3
Collection and Transport & Storage ................................................................................................ 3
Procedure ......................................................................................................................................... 6
General Precautions: ........................................................................................................................ 6
Pre Board Preparation: ..................................................................................................................... 7
Microlab STARlet ............................................................................................................................ 8
CFX96TM
Amplification and Detection .......................................................................................... 20
PCR Analysis ................................................................................................................................. 21
Interpretation ................................................................................................................................. 28
Seegene Report Interpretation ........................................................................................................ 28
General Seegene Reporting Process ............................................................................................... 29
Seegene Reporting Process for Repeat Samples ............................................................................ 31
Reporting ....................................................................................................................................... 32
PRELIMINARY RESULTS (for UHN/MSH/STAT patients only): ....................................... 32
FINAL RESULTS ........................................................................................................................ 33
Cleaning ......................................................................................................................................... 36
Quality Control .............................................................................................................................. 37
![Page 2: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/2.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 2 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Troubleshooting ............................................................................................................................. 38
Related Documents ........................................................................................................................ 38
References: .................................................................................................................................... 38
Record of Edited Revisions ........................................................................................................... 39
![Page 3: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/3.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 3 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Introduction
The 2019–20 coronavirus outbreak is an ongoing public health emergency of international concern
involving coronavirus disease 2019. It is caused by SARS-CoV-2, first identified in Wuhan, Hubei,
China.
SARS-CoV-2 is closely related to the original SARS-CoV. It is thought to have a zoonotic origin.
Genetic analysis has revealed that the coronavirus genetically clusters with the genus
Betacoronavirus, in lineage B of the subgenus Sarbecovirus together with two bat-derived strains. It
is 96% identical at the whole genome level to other bat coronavirus samples (BatCov RaTG13). In
February 2020, Chinese researchers found that there is only one amino acid difference in certain
genome sequences between the viruses found in pangolins and those from human patients, implying
that pangolins may have been an intermediate host.
Allplex™ 2019-nCoV Assay (100T) is a qualitative in vitro test for multiple detection of 3 types of
gene (RdRp gene, Envelope gene, and Nucleocapsid gene).
Collection and Transport & Storage
Note: Nasopharyngeal (NP) swabs, throat swabs, saliva, sputum, Bronchial Alveolar Lavages
(BALs), EDTA plasma/serum, stool and breast milk are verified sample types for this assay.
I. Respiratory samples
Respiratory samples such as Sputum, Throat swabs, Nasopharyngeal (NP) swabs
and Bronchial Alveolar Lavages (BAL) are tested.
Sputum and BAL can be stored at 4°C for up to 72 hours after collection in sterile
container.
NP and Throat swabs can be stored 4°C for up to 72 hours after collection in
Universal Transport Media (UTM™) or equivalent.
If the specimen is not going to be tested within 72 hours of collection, then it
should be stored at -70°C or lower.
II. Blood samples
Specimen collected in serum separator tube/red top tube or EDTA tubes may be
tested
Follow the tube manufacture’s processing instructions for serum and plasma
collection tubes. Gravity separation is not sufficient for specimen preparation.
![Page 4: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/4.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 4 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Separated serum /EDTA plasma can be stored in 4°C for up to 72 hours after
collection. If the specimen is not going to be tested within 72 hours of collection,
then it should be stored at -70°C or lower.
III. Placenta/ conception swabs
Placenta/conception swabs can be stored at 4°C for up to 72 hours after collection
If the specimen is not going to be tested within 72 hours of collection, then it
should be stored at -70°C or lower.
IV. Stool samples
Stool can be stored at 4°C for up to 72 hours after collection in sterile container.
If the specimen is not going to be tested within 72 hours of collection, then it
should be stored at -70°C or lower.
V. Saliva samples
To maintain optimum viability, the specimen should be stored and transported at
2-8°C. If transport to the laboratory will be delayed, specimen should be frozen at
-70°C or below and shipped on dry ice.
Acceptable for testing up to 48hrs at room temperature
VI. Other specimen types (e.g. urine, tissue, miscellaneous samples, etc.) need to be
approved by the Microbiologists
Materials, Equipment and Facilities:
Clean Room: Biosafety Cabinet (MIBCT3), freezer (MIFTG)
Specimen Preparation area: Biosafety Cabinet –2.5 level
Microlab STARlet
CFX96TM
IVD
Nimbus-96 Deep Well Micro Plate
1.5ml micro tube
Optical Flat 8-Cap strips
Hard-Shell 96 Plate, White shell, White wells
High volume Tips (1000 µL)
Standard Volume Tips (300 µL)
Sample and reagent racks 1.5 ml RNase-free microcentrifuge tubes
Variable volume Rainin pipettes: 1 to 20 µL, 10 to 200 µL, 100 to 1000 µL
![Page 5: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/5.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 5 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Reagents:
AllplexTM
2019-nCoV Assay
STARMag 96×4 Cartridge Kit
i. STARMag 96×4 Universal Cartridge Kit
ii. STARMag 96×4 C-Type Cartridge Kit
iii. STARMag 96×4 L-Type Cartridge Kit
< µL > Original version C version L Version A
PK 10 Not required Not required Not required
IC 10 10 10
Lysis Buffer 150 150 350
Sample 200 200 200
Beads 10 10 Included in LB
Binding buffer 586 586 Included in LB
Washing Buffer 1 490 490 350
Washing Buffer 2 490 490 350
Washing Buffer 3 500 500 Not required
Elution Buffer 100 100 100
External controls: Positive external controls (positive for all E/N/R genes) and Negative
external controls (negative for all E/N/R genes) to be run for each new lot/shipment if QC
fails
![Page 6: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/6.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 6 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Procedure
General Precautions:
There must be separate PCR work areas:
Clean room
Specimen preparation room-2.5 level lab
Powder-free Gloves should only be in use in PCR areas.
Change gloves frequently and keep tubes closed whenever possible.
Prepare Working 1% sodium hypochloride daily.
Specimen Preparation Supplies and equipment must be dedicated to Specimen Prep Area and
not used for other activities and never used in Clean Room.
Change lab coats and gloves between work areas.
Use only Aerosol Resistant Tips (ART)
Use only sterile RNase-free, DNAse-free microtubes
Thaw components completely at room temperature.
Clean PCR work areas (Clean Room and Specimen Preparation Area) bench tops and
equipment after each shift.
Do not pool reagents from different lots or from different tubes of the same lot.
Do not use the product after its expiry date.
Sample Preparation:
1. Sputum:
Viscous sputum sample should be treated with Sputolysin prior to testing at a 1:2
dilution.
2. Stool:
Pipette 500 µL PCR grade water or Universal Transport Media (UTM) into 1.5mL
tube
Take a swab from Cepheid VRE kit to swab the stool
Insert swab into the stool specimen and roll the swab over the specimen to be fully
coated.
Note: make sure that swab is brown but not including a large amount of fecal
material
Vortex the stool resuspension tube at the highest setting for 15 seconds. Ensure full
vortex is performed during mixing.
Centrifuge stool resuspension Tube at a minimum of 2,000 x g (4643 rpm) or 13,200
rcf for a minimum of 30 seconds.
![Page 7: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/7.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 7 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Pipette 300 µL of supernatant into another new 1.5mL tube for subsequent testing
3. Tissue
If PCR is performed from tissue in formalin, first take the tissue out of formalin and
place it in UTM before grinding.
If there is significant amount of formalin, dilute out further by placing in second
UTM vial before grinding.
a portion of the ground up sample gets placed into a freezer box inside MIRT16 fridge
the remaining ungrounded sample goes into MIRT17 fridge
Scan request form into donor folder
4. Saliva
Dilute 1:1 with saline (ex. 300 µL saliva + 300 µL saline)
Pre Board Preparation:
1. Make sure the specimen ID label is just underneath the rim, straight up and no fold made
2. Make sure there is at least 300 µL in the 1.5ml conical tube
3. Place samples into 65°C for 30 minutes
4. After 30 minutes, take the samples out of the incubator and bring samples to BSC
5. Load the tubes onto Seegene sample carriers and CAREFULLY uncap the tubes
Note: Please avoid contamination between tubes. Change gloves if needed
6. Make sure specimen IDs on the tubes are fully visible for automatic barcode scanner to read.
(If they are well attached on the tubes, all barcodes will be read.)
Note: Leave the sample carries in the BSC till ready to load onto Seegene
![Page 8: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/8.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 8 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Microlab STARlet
1. Click Maintenance button
2. Select proper maintenance (daily or Weekly) and click Start button
3. Following the steps on the screen to finish the necessary maintenance process.
4. Close the maintenance screen
5. Double click SLnCoV icon on the desk top
6. The following message window pops up, click OK button
Start button
![Page 9: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/9.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 9 of 40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
7. Click LAUNCHER RUN
8. Select One Step for 2019-nCoV
Note the following procedure changes when using the C or L Kit
C-Kit procedure L-Kit procedure A-Kit procedure
1. There will be no PK
2. Before use, you should aliquot
1. There will be no PK in version
L
![Page 10: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/10.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 10 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
1.8ml magnetic beads from the
big bottle given into four empty
2ml tubes. Make sure to vortex
the magnetic beads bottle VERY
thoroughly WELL to remove any
aggregation before the dispense.
If a partial and new kit are
loaded, 2 tubes are required
If only a new kit is loaded, 1 tube
is required to be loaded.
3. Start a protocol
4. Scan the extraction cartridge
barcode as it (see a picture
below) prompts after starting the
protocol.
5. Remaining process is the same
2. Magnetic beads are in Lysis
buffer so make sure to mix
them well before the use by
pipetting up and down in the
liquid (NOT above the liquid
due to bubbles generated) by
1000 µL pipette or the lysis
buffer tube itself can be
separated. Before opening the
plastic cap, take the lysis buffer
tube ifself out and gently invert
it to mix them well.
3. Start a protocol
4. Scan the extraction cartridge
barcode as it(see a picture
below) prompts after starting
the protocol.
Remaining process is the same
9. Prepare Waste Bag-Make sure the tip waste is in position.
If not, place it correctly.
Click Next
10. Prepare 1000 µL Tip
Place the 1000 µL filter tips to the required position according to the screen display
There are 7 positions for the 1000 µL tips
Click Next
![Page 11: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/11.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 11 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
11. Prepare 300 µL Tip
Place the 300 µL filter tips to the required position according to the screen display (4
positions)
There are 4 positions for the 300 µL tips
Click Next
12. Prepare Small Tube Rack Carrier
Place the small tube rack carrier at Tack No. 24
After confirming PCR information, place the PCR reagent in the small tube rack 1 or
1,2
Click Next
![Page 12: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/12.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 12 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
13. Prepare IC
Place RP-VIC on appropriate position in the small tube rack 1
Click Next
14. Prepare PK
Prepare PK if required
Prepare the PK:
Add 2600 µL of Proteinase Buffer Universal PB in to Universal Proteinase K to
lyophilize the powder and let the bottle stand still for 10 minutes before use
Note: Instrument needs a little more than what the system calculates, please top it up
(i.e. if the system say 1170 µL, top it up to 1200 µL; if it says 270 µL, top it up to
300 µL)
Type the total sample number in the Sample Qty window, the necessary volume of
PK is calculated by the system.
![Page 13: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/13.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 13 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Pipette required volume of PK into the 1.5 ml sterile microtube according to the
number of the samples and place it to the designated position in the small tube rack
1
Click Next
Note: Prepare PK before running the instrument. The exact volume will be indicated
on the Work List
15. Prepare Magnetic Bead
Re-suspend the bead tube with hand tapping and then vortex for more than 1
minute
1st Bead position:
Place the Bead tube on the designated position in the small tube rack 1.
2nd
Bead position:
If the total number of prepared samples exceeds the remaining samples covered by
previous cartridge, add a new tube of magnetic beads (see the picture below)
![Page 14: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/14.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 14 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Click Next
Note: Must pair extraction cartridge and bead location. (#1 to #1, #2 to #2)
16. Prepare Extraction Reagent
Add absolute ethanol to the Wash Buffer cartridge before using for the first time.
Refer to the Tub Label of WB2.
Universal Kit & C-Type Kit: add 48mL of ethanol into WB2
L-Type Kit: add 38mL of ethanol into WB2
A Kit: add 15mL of ethanol into WB1 and 48mL of ethanol into WB2
Place the extraction reagent cartridge at Track No.18
After completing work list, scan the reagent barcode at the Reagent Barcode Page
and load the required extraction reagent to the designated position (see picture
below)
Click Next
![Page 15: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/15.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 15 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
17. Prepare 96 DWP
Place the 96DWP on the designated position.
When using barcoded 96 DWP, the barcode of DWP should be placed to right side
of plate.
Click Next
18. Prepare PCR Plate
Place the Plate Bio-Rad White on the designated position.
After confirming PCR information, place the Plate Bio-Rad White in the plate 1
and 2.
When the samples exceed 96 well in plate, it automatically turns over to plate 2.
Note: Normally the sample number is less or equal to 96, only need to load PCR
plate 1
When using barcoded PCR plate, the barcode of plate should be placed to right side
of plate
Click Next
![Page 16: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/16.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 16 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
19. Prepare Sample
Prepare each sample on tubes, referring to sample type and tube size
Then place them into appropriate position
Each tube must contain more than a guided volume of samples (Refer to Sample
preparation guide)
Insert the stop barcode tube into the following location after loading all the
samples.
Sample Preparation Guide
Minimum Volume (µL)
Tube Specimen
1.5mL 300
12 mm 800
16 mm 2000
20. Prepare Sample Carrier
Place the 32 sample carriers at Track No.13, 14 and 15
If Sample Unload is turned “ON”, after sample dispense, sample carriers are
automatically unloaded
Click Worklist
![Page 17: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/17.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 17 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Read specimen IDs:
I. Automatically read specimen IDs :
Click Barcode button on the top right hand side (no need to type
quantity of the samples)
II. Manually scan specimen IDs:
Manually scan specimen IDs by using hand barcode provided (need to
type the quantity of samples)
Note: In case you let the system read the sample IDs automatically, if ID labels are
not well placed, there is a high chance it will not be read. Therefore, have
worksheet with all IDs that you loaded on the system ready so any missed
specimen IDs you can manually scan IDs accordingly
21. Scan extraction reagent barcode. When there are not enough reagents left in the 1st cartridge,
there will be a pop-up window asking you to place the second set of reagent cartridge. Click
OK and then scan the barcode from the second reagent cartridge.
22. Click Next
![Page 18: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/18.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 18 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
23. Scan PCR reagent barcode. When there are not enough reagents left in the 1st PCR reagents,
there will be a pop-up window asking you to place the 2nd
set of PCR reagents. Click OK and
then scan the barcode from the second one.
24. Click Next
25. Place each PCR reagent accordingly. Make sure to completely thaw, vortex (3~5 seconds) and
quick spin them down. If there are not enough PCR reagents left from the 1st kit, you need to
place new PCR reagents in the 2nd
row as you scanned the barcode.
26. Click Next
27. PCR plate information tells you how many 8 strip caps you will need at the end. Click Next
![Page 19: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/19.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 19 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
28. Follow the following check list to make sure everything is ok
1000 uL Tip
300 uL Tip
Sample
Carrier
Extraction
Reagent
![Page 20: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/20.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 20 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
29. Click Run
30. At the last pop up screen, be sure that you placed everything in the correct position and mark
the check box as you read it line by line once you followed.
31. Do not leave as you will need to indicate tip positions. In 2~3 minutes, there will be pop up
screen for you to indicate tips location and make them exactly identical to where you placed.
Click and drag onto to the gray-out 1000 µL tips spot to make sure it turns to brown
indicating that all 1000 µL tips are present, click OK
Click and drag onto to the gray-out 300 µL tips spot to make sure it turns to brown
indicating that all 300 µL tips are present, click OK 32. Once extraction is completed, pop-up window saying Method Completed, click OK button
Note: This will generate the plrn file to the Biorad Thermocycler
CFX96TM
Amplification and Detection
1. On the desktop chose plrn (starlet-PC)
2. Choose the run you want to start by double clicking on it Choose open lid and load plate
into thermocycler
3. Choose open lid and load plate into thermocycler
4. Choose close lid
5. Go to plate tab
6. Go to edit select and check the sample IDs and make sure it is correct run
![Page 21: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/21.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 21 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
7. Click OK
8. Choose next
9. Start Run
10. Name the run and save to current folder
PCR Analysis
1. Check that the signal at the beginning of cycling is OK
2. Check that the curves are sigmoidal
3. In the top menu bar, Click Tools
4. Select Seegene export
![Page 22: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/22.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 22 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
5. Select COVID-19 RAW DATA
6. Click OK
7. Click “X” at the top right corner of the window to close file
Click NO to the following Message Window “Do you want to save the changes…:,
Do not save changes, store raw data as is
![Page 23: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/23.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 23 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
8. Click “X” at the top right corner of the window to close the Bio-Rad CFX program
9. On the desktop, double click 2019-nCOV Viewer
10. The following message window pops up, click “OK”
11. Open Raw Data
a. Click File->Open
b. Click Desktop
c. Select COVID-19 RAW DATA folder
d. Click Open
![Page 24: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/24.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 24 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Open current run by date and time
Open Nimbus-yes
Desktop-plrn(STARLET-PC)-open your run (need to scroll right)
Select all samples (well info)
12. Select Quantstep4, Click Open
13. Select current run by date and time and click Open
![Page 25: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/25.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 25 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
14. Nimbus/STARlet Setting window pops up, click Open
15. Click Desktop, select plrn(STARLET-PC) – Shortcut, Click Open
16. Select your run (need to scroll right), click Open
![Page 26: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/26.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 26 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
17. Create the results in PDF file
a. Select all samples by checking well info
b. Click Print Icon
c. Type the name of the operator
d. Click PDF
e. Click Desktop button
f. Select COVID PDF RESULTS
g. Click Open
b
d
c
a
e f
g
![Page 27: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/27.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 27 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
h. Type the file name including the date and operator’s name
i. Click Save
j. Must wait for the complete message window pops up
k. Click OK
l. On the desktop, double click the folder of COVID PDF RESULTS
m. Open your PDF result and print
Auto-Transfer of Results
1. In the 2019-nCOV Viewer uncheck any sample numbers that have to be exported
manually i.e.; BGI repeats that are negative, invalid non-NP, non-throat and non-
sputum samples, offline repeats
2. Create a CSV file by using File > Export CSV
3. Save file under the shortcut to the Seegene (eg. 1_Seegene1 shortcut on the desktop)
using the same name as the PDF file
4. Close 2019 nCoV-Viewer
5. Open bioftp program on the desktop
6. Drag CSV file from the Seegene folder from the left window to the right window
![Page 28: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/28.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 28 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Interpretation
When a successful run is completed (positive and negative controls yield expected corresponding results)
use the table below for sample interpretation.
Seegene Report Interpretation
Negative Internal Control readings in HEX (IC) may occur in strongly positive FAM, Cal Red 610
and Quasar 670 channels results. High signal for E/N/RdRP gene in the sample often deplete the
amplification/detection materials (eg. nucleotides), leading to reduced or absent Internal Control
signals.
E-gene RdRP-gene N-gene IC Interpretation
- - - + COVID-19 Virus NOT Detected
+ + + +/- COVID-19 Virus Detected
- + + +/- COVID-19 Virus Detected
+ + - +/- COVID-19 Virus Detected
+ - + +/- COVID-19 Virus Detected
- + - +/- See “General Seegene Reporting Process”
+ - - +/- See “General Seegene Reporting Process”
- - + +/- See “General Seegene Reporting Process”
- - - - INVALID
See “General Seegene Reporting
Process”Repeat in Neat for all non-NP,
non-throat, non-sputum samples
![Page 29: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/29.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 29 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
General Seegene Reporting Process
![Page 30: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/30.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 30 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
![Page 31: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/31.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 31 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Seegene Reporting Process for Repeat Samples
![Page 32: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/32.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 32 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Reporting
Follow “General Seegene Reporting Process”.
PRELIMINARY RESULTS (for UHN/MSH/STAT patients only):
PRELIMINARY POSTIVE RESULTS:
Result in the isolate window as COVID-19 virus
In the Isolate comment, choose \DPCR & \ind
Status as PRELIMINARY
Inform all positive results according to COVID RESULT COMMUNICATION
Example:
COVID-19 virus
DETECTED by real-time PCR.
* * * * * * * * * * * * * * * * * *
(Low level detection)
Sample is being retested. Further report to follow.
Testing performed using the Seegene Allplex 2019-nCoV Assay.
NOTE: The Seegene Allplex 2019-nCoV Assay has been approved
by Health Canada for Emergency Use Access (EUA) and has
been verified by the University Health Network/Sinai
Health Microbiology Laboratory.
![Page 33: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/33.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 33 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
FINAL RESULTS
NEGATIVE RESULTS:
Result in the test window using COVID-19 Virus Not Detected test comment
Choose from the keypad: }-NCS (Seegene)
Status as FINAL
When Seegene .csv file for a run is exported to LIS, all negative samples should auto-
finalize.
Example:
COVID-19 virus NOT detected by real-time PCR.
* * * * * * * * * * * * * * * * * * *
Testing performed using the Seegene Allplex 2019-nCoV Assay.
NOTE: The Seegene Allplex 2019-nCoV Assay has been approved
by Health Canada for Emergency Use Access (EUA) and has
been verified by the University Health Network/Sinai
Health Microbiology Laboratory.\
POSITIVE RESULTS:
Result in the isolate window as COVID-19 virus
In the Isolate comment, choose \CVS+ (Seegene)
Status as FINAL
Inform all positive results according to COVID RESULT COMMUNICATION
Example:
COVID-19 virus
DETECTED by real-time PCR
* * * * * * * * * * * * * * * * * *
Testing performed using the Seegene Allplex 2019-nCoV Assay.
NOTE: The Seegene Allplex 2019-nCoV Assay has been approved
by Health Canada for Emergency Use Access (EUA) and has
been verified by the University Health Network/Sinai
Health Microbiology Laboratory
![Page 34: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/34.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 34 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
INVALID RESULTS:
Repeat in Neat for all non-NP, non-throat and non-sputum samples
Result in the test window using COVID-19 virus PCR test unable to be completed test
comment
Choose from the keypad: }INCS (Seegene)
Status as FINAL
Inform all invalid results to the ward /clinic
Example:
COVID-19 virus PCR test unable to be completed.
* * * * * * * * * * * * * * * * * * *
Test result is suggestive of inhibition of the Seegene
Allplex 2019-nCoV Assay PCR reaction. Please submit
another sample for testing if clinically indicated.
NOTE: The Seegene Allplex 2019-nCoV Assay has been approved
by Health Canada for Emergency Use Access (EUA) and has
been verified by the University Health Network/Sinai
Health Microbiology Laboratory.
![Page 35: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/35.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 35 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
FINAL RESULT for REPEAT of LOW LEVEL DETECTION:
Follow Seegene Reporting Process for Repeat Samples
CONFIRMED POSITIVE RESUTLS:
Follow FINAL RESULTS for POSITIVE reporting procedure.
INDETERMINATE RESUTLS:
Suppress COVID-19 virus isolate
Use the following Indeterminate test comment
Status as FINAL
Inform all updated confirmed results according to COVID RESULT
COMMUNICATION
Example:
Indeterminate for COVID-19 virus.
* * * * * * * * * * * * * * * * * * *
Results should be interpreted within the context of the
clinical signs, symptoms, and history of the patient.
INDETERMINATE result may indicate the presence of low levels
of virus, non-specific reactivity of the assay, or other unrecognized factors. Please submit a
follow-up sample if clinically indicated.
Testing performed using the BGI Real-Time Fluorescent
RT-PCR 2019-nCoV Assay.
NOTE: The BGI Real-Time Fluorescent RT-PCR 2019-nCoV Assay
has been approved by Health Canada for Emergency Use Access
(EUA) and has been verified by the University Health
Network/Sinai Health Microbiology Laboratory.
* * * * * * * * * * * * * * * * * *
DISCLAIMER:
NP swab, throat swab, sputum, BAL, saliva, EDTA plasma/serum, stool and breast milk
are verified samples for testing on Seegene.
Run all “non-verified” samples (tissue, placenta, CSF and other body fluids) as a batch at
the beginning or end of a run so that it’s easy to prevent these from being transmitted to
LIS.
![Page 36: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/36.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 36 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Follow General Seegene Reporting Process
Add the following disclaimer for non-verified sample types.
Its performance has NOT been validated on this specimen type.
Please take this into consideration when interpreting this result.
Cleaning
Routine Cleaning:
After each run:
After each Seegene run, wipe down with RNase Away and alcohol the Starlet platforms,
arms/probes that pick up the pipette tips to reduce potential cross-contamination.
Weekly:
Soak and disinfect sample racks in 10% bleach solution for 10 minutes and rinse thoroughly with
water. Let fully dry.
At shift end, perform cleaning protocol as outlined below:
Clean Room: Wipe down with RNAse Away or NucleoClean on paper towel, followed by distilled
water, and then 70% alcohol
Biological Safety Cabinet
Pipettes
Bench tops
Specimen Preparation Area: Wipe down with working 1% hypochloride (made daily), followed
by distilled water, and then 70% alcohol
Biological Safety Cabinet (BSC), pipettes, centrifuge, and bench top.
Seal and discard BSC waste.
Soak and wash sample loading racks weekly.
Amplification Area: Wipe down surfaces with RNAse Away or NucleoClean on KimWipe,
followed by UltraPure water, and then 70% alcohol Wipe.
Seal & discard reaction microtubes into biohazard waste after each run.
Perform the cleaning procedure according the daily maintenance sheet.
![Page 37: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/37.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 37 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Quality Control
Reagent QCs:
An External Control (external to Seegene reagent kits) is used to monitor the isolation,
amplification and detection procedures. The result must correspond to expected value supplied by
the manufacturer.
External control should run with each new shipment or with each new lot.
Daily QCs:
Every run:
Each patient specimen must have an Internal Control (IC) added to monitor both extraction
and PCR inhibition.
A Positive Control is included and shows amplified signal in each corresponding channel
A Negative Control is included and shows negative signal in each corresponding channel
Report all failed QCs to senior/charge technologist.
Failed QC:
Test is invalid without satisfactory QC results.
a. Do not release results pending resolution of QC failure.
b. Inform charge/senior technologist.
c. Record in Reagent Log Chart, Instrument Maintenance Log or Incident Report where
appropriate.
d. If the QC failure was due to a simple matter of position reversal or misplacement, the run
can be released (positive QC material yielded positive result, negative yielded negative
result) with approval from senior, charge or microbiologist.
e. If positive QC material yielded negative result, repeat the entire run.
f. If negative QC material yielded positive result, it may be due to cross-contamination from
adjacent positive sample within the run or carry-over contamination from previous runs via
equipment or the environment. Review procedure and equipment to establish and eliminate
potential sources of contamination.
g. The extent and nature of contamination can also be evaluated by comparing the positive rate
of the run with its expected positive rate.
h. If the contamination is extensive, it is necessary to quarantine/discard potentially
contaminated reagents and consumables and disinfect equipment and environment before
repeating the run.
i. If a carry-over contamination is suspected (e.g. two or more runs with negative QC being
positive or patient samples have higher than expected positive rate and these samples are
![Page 38: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/38.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 38 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
often non-repeatable positives), it is necessary to have a thorough environmental
disinfection followed by swabbing to monitor.
j. Successful ending to a carry-over contamination may be indicated by QC results and patient
positivity rate falling back to the expected normal range and three negative environmental
swabs.
Troubleshooting
Refer to
For technical support Email description of errors and include files below as attachments to Seegene technician support staff as per Equipment Maintenance Procedure for the Starlet.
1. Trace logs (always) 2. Screenshot or exact error or error message (if applicable) 3. Abort logs (if applicable) 4. .pcrd file (result-based rather than mechanical issue)
Related Documents
Virology Accessioning Manual MI_MD_ACC
Training Checklist
References:
Seegene Sars-CoV-2 Allplex Assay package insert.
![Page 39: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/39.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 39 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Record of Edited Revisions
Manual Section Name: SARS-CoV-2 PCR by AllplexTM Seegene
Page Number / Item Date of Revision Signature of
Approval
Update pre-lysis procedure to heat fix procedure
Remove email communication to microbiologists,
director, manager
Updated reporting section
March 16, 2020 Dr. T. Mazzulli
Add manually scan sample IDs
Add extraction reagent barcode process
Add PCR reagent barcode process
Update reporting and send-out section
March 18, 2020 Dr. T. Mazzulli
Updated reporting to remove any indeterminate or PHOL
send out reporting.
March 20, 2020 Dr. T. Mazzulli
Updated Collection and Transport & Storage:
Changed ‘4°C for up to 7 days’ to ‘4°C for up to 72
hours’ due to initial typo
March 27, 2020 Dr. T. Mazzulli
Updated acceptable sample types
Added invalid reporting
April 14, 2020 Dr. T. Mazzulli
Updated acceptable sample types:
Added Placenta/ conception swabs Stool samples
Added the sample preparation for stool and viscous
sputum
Updated the reporting:
No need to report the individual gene
April 17, 2020 Dr. T. Mazzulli
Updated Report:
No need to repeat samples with initial invalid results;
report final results of “INVALID”
April 20, 2020 Dr. T. Mazzulli
Updated phrases for invalid reporting
Added back “No gene targets were detecte…..
Please submit another sample for testing if
clinically indicated.”
Updated to repeat all invalid non-NP/non-Throat/non-
sputum samples in neat
Clarified NP/Throat/Sputum/BAL/EDTA plasma and
serum are verified sample types for this assay.
April 23, 2020 Dr. T. Mazzulli
![Page 40: Policy # MI MD COVID Page 1 of 40 Department of ......Department of Microbiology Quality Manual Policy # MI_MD_COVID Page 2 of 40 Version: 5.0 CURRENT Section: Molecular Diagnostics](https://reader036.fdocuments.in/reader036/viewer/2022071413/610c6a68f39ef764b7685c64/html5/thumbnails/40.jpg)
Department of Microbiology
Quality Manual
Policy # MI_MD_COVID
Page 40 of
40
Version: 5.4 CURRENT
Section: Molecular Diagnostics Procedures Subject Title: SARS-CoV-2 PCR by AllplexTM
Seegene
UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY
NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that do not state "CONTROLLED COPY” in red print are not controlled and should be checked against the
document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Standard Operating Procedures\Molecular Diagnostics Procedures\
Page Number / Item Date of Revision Signature of
Approval
Added the disclaimer for reporting non-verified samples
Addition of C and L kit procedure.
Addition of new interpretation and report testing
algorithm for N-Gene only testing.
May 20, 2020 Dr. T. Mazzulli
Addition of Seegene troubleshooting guide link August 20, 2020 Dr. T. Mazzulli
Addition of Low positive algorithm for single gene
positives. Addition of low positive confirmed, or
indeterminate repeats from BGI. Addition of auto-
resulting instructions
August 24, 2020 Dr. T. Mazzulli
Full document review included in all updates. Bi-annual review conducted when no revision had
been made within 2 years.
Page Number / Item Date of Revision Edited by:
Addition of General Seegene Reporting Process,
Seegene Reporting Process for Repeat Samples;
Updated Seegene report interpretation table, reporting
rules and preliminary and final report statements in
test window and isolate window
Updated disclaimer for non-verified samples
Addition of routine cleaning after each Seegene run
December 31, 2020 Dorna Zareianjahromi
Updated Saliva testing info March 22, 2021 Dorna Zareianjahromi
Updated Tissue processing info April 9, 2021 Dorna Zareianjahromi
Updated the list of verified sample types with stool and
break milk included
April 9, 2021 Dorna Zareianjahromi
Minor formatting change April 11, 2021 Jessica Bourke
Update on a typo in the reporting part April 20, 2021 Oliver Li
Updated files required for troubleshooting with technician
support
April 24th
, 2021 Jessica Bourke