Plates and Screws Supervised by Dr. Marwan Abuhashem
By Dr. Ismael Al-jabiri
Al-Bashir Orthopedic Department Ministry Of Health Jordan
![Page 2: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/2.jpg)
1. Anatomical Reduction.2. Stable internal fixation.3. Preservation of Blood supply4. Early mobilization.
Principles of Fixation :
![Page 3: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/3.jpg)
3
• “Surfaces of the fracture do not displace under functional load”
• Can only be achieved by interfragmentary compression
Absolute stability
• A plate by itself rarely provides absolute stability
• The key tool of absolute stability is the lag screw• Compression must sufficiently neutralize all forces[bending, tension, shear and rotation]
![Page 4: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/4.jpg)
Relative Stability:
A fixation device that allows small amounts of motion in proportion to the load applied. The deformation or displacement is inversely proportional to the stiffness of the implant. Examples: Intramedullary rod, bridge plating, external fixation.
![Page 5: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/5.jpg)
PLATES
Introduction : Bone plates are like internal splints holding
together the fractured ends of a bone. A bone plate has two mechanical functions.
• It transmits forces from one end of a bone to the other, bypassing and thus protecting the area of fractures.
• It also holds the fracture ends together while maintaining the proper alignment of the fragments throughout the healing process.
![Page 6: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/6.jpg)
History
Hansman’s Bone Plate (1886)
Hansmann’s plates were:Bent at the end to protude through the skinAttched to bone by screw with long shanks that projected outside the soft tissues.
![Page 7: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/7.jpg)
Bone Suture Stabilization Inserts (Koenig, 1905)
![Page 8: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/8.jpg)
History
Since 1958, AO has devised a family of plates for long bone fractures, starting with a round holed plate.
In 1969 the Dynamic Compression Plate was developed.
In 1994 LC DCP was created. In 2011 LCP with combination
holes has come into use.
![Page 9: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/9.jpg)
Names of plates.
1. Shape (Semitubular, 1/3rd tubular)2. Width of plate (Small, Narrow,
Broad)3. Shape of screw holes. (Round, Oval)4. Surface contact characteristics. (LC,
PC)5. Intended site of application
(Condylar Plate)6. According to the function
![Page 10: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/10.jpg)
Standard Plates
Narrow DCP-4.5 mm
Broad DCP – 4.5 mm
3.5 mm DCP
![Page 11: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/11.jpg)
LC-DCP 3.5 & 4.5mm
Reconstruction plate 3.5 & 4.5mm
![Page 12: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/12.jpg)
1/3 tubular plate 2.7, 3.5 & 4.5 mm
![Page 13: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/13.jpg)
Special Plates
T Plates
T&L Buttress plates
![Page 14: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/14.jpg)
Lateral proximal Tibial buttress plates
Condylar buttress plate
Narrow lengthening plates
![Page 15: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/15.jpg)
Broad Lengthening plate
Spoon plate
Clover leaf plate
![Page 16: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/16.jpg)
DCP - 3.5 and 4.5
First introduced in 1969 by Danis
Revolutionary concept of compression plating
Featured a new hole designed for axial
compression
Broad 4.5 for Femur & Narrow 4.5 for
Humerus & Tibia
DCP 3.5 for Forearm, Fibula, Pelvis & Clavicle
![Page 17: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/17.jpg)
Advantage of DCP :
1. Inclined insertion 25°longitudinal and 7° sideways
2. Placement of a screw in neutral position without the danger of distraction of fragments
3. Insertion of a lag screw for the compression4. Usage of two lag screws in the main
fragments for axial compression5. Compression of several fragments individually
in comminuted fractures6. Application as a buttress plate in articular area
![Page 18: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/18.jpg)
Problems with DCP
Unstable fixation leads to fatigue &
failure
Strict adherence to principles of
compression
Compromised blood supply due to
intimate contact with underlying cortex
“Refractures” after plate removal
![Page 19: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/19.jpg)
LC-DCP Represents a design change Overcome problems with DCP Plate footprint reduced Minimized kinking at screw holes,
more countourable, reduced plate fatigue at
hole Allows more inclination of screw
in longitudinal plane and transverse
plane.
![Page 20: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/20.jpg)
In the DCP (A), the area at the plate holes is less stiff than the area between them so while bending, the plate
tends to bend only in the areas of the hole.
The LC-DCP(B) has an even stiffness without the risk of buckling at the screw holes.
![Page 21: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/21.jpg)
The LC-DCP offers additional advantage
Improve blood circulation by minimizing plate-bone contact
More evenly distribution of stiffness through the plate
Allows small bone bridge beneath the plate
![Page 22: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/22.jpg)
Tubular plates
3.5 system - 1/3rd Tubular 4.5 system - Semitubular Limited stability-Oval holes – Axial
compression can be achieved.
-Low rigidity (1mm thick). Lateral malleolus Distal ulna / Olecranon Distal humerus
![Page 23: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/23.jpg)
limited stability. The thin design allows for easy shaping and is primarily used on the lateral malleolus and distal
ulna. The oval holes allow for limited fracture compression with eccentric screw placement.
![Page 24: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/24.jpg)
Reconstruction plates
Deep notches between holes
Accurate contouring in any plane
Pelvis Acetabulum Distal humerus Clavicle Olecranon
![Page 25: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/25.jpg)
Reconstruction plates are thicker than third tubular plates but not quite as thick as dynamic compression plates. Designed with deep notches between the holes, they can be contoured in 3 planes to fit
complex surfaces, as around the pelvis and acetabulum. Reconstruction plates are provided in straight and slightly thicker and stiffer precurved lengths. As with tubular plates, they have oval screw
holes, allowing potential for limited compression.
![Page 26: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/26.jpg)
LCP – Locking Compression Plate
![Page 27: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/27.jpg)
LCP
Latest in the evolution
“ Internal fixator ” Combination of
locking screw with conventional screw
Extraperiosteal location of plate
![Page 28: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/28.jpg)
LCP: internal external fixator
![Page 29: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/29.jpg)
LCP
Combines advantages of DCP principle and locking head principle.
Flexibility of choice within a single implant. Screw hole have been specially designed to accept either:
cortical screw and locking screw
![Page 30: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/30.jpg)
The locking screws, by achieving angular stability within the plate holes are able to produce a similar hoop with just two unicortical screws.
LCP
![Page 31: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/31.jpg)
LOCKING COMPRESSION PLATE (LCP) Principle :
Angular-stability whereas stability of conventional plates is friction between the plate and bone
Screw locking principle
Provides the relative stability Healing by callus formation (Secondary Healing)
![Page 32: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/32.jpg)
Stability under load
• By locking the screws to the plate, the axial force is transmitted over the length of the plate
• secondary loss of the intraoperative reduction is reduced
Blood supply to the bone
• No additional compression after locking
• Periosteal blood supply will be preserved
![Page 33: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/33.jpg)
LCP used as internal fixator to bridge multifragmentary diaphyseal fracture zone.
Locking compression plate is used.
Standard cortical and cancellous screws are used as a traditional plate.
LCP
![Page 34: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/34.jpg)
Principle of internal fixation using LCP :
1. 1st reduced the # as anatomical as possible2. Cortical screw should be used 1st in a fracture
fragment3. If the locking screw have been put, use of the
cortical screw in the same fragment without loosening and retightening of the locking screw is not recommended
4. If locking screw is used first avoid spinning of plate
5. Unicortical screws causes no loss of stability
![Page 35: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/35.jpg)
6. In Osteoporotic bone bicortical screws should be used.
7. In comminuted # screw holes close to the fracture should be used to reduce strain.
8. In the fracture with small or no gap the immediate screw holes should be left unfilled to reduced the strain.
Principle of internal fixation using LCP :
![Page 36: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/36.jpg)
Indications :
1. Osteoporotic #
2. Periprosthetic #
3. Multifragmentry #
4. Delayed change from external fixation to internal fixation.
![Page 37: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/37.jpg)
Advantages :
1. Angular stability 2. Axial stability3. Plate contouring not required4. Less damage to the blood supply
of bone5. Decrease infection because of
submuscular technique6. Less soft tissue damage
![Page 38: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/38.jpg)
Sizes of DCPName of plate Small Narrow Broad
Width 11 mm 13.5 mm 17.5mm
Profile 4 mm 5.4 mm 5.4 mm
Screw 2.7 , 3.5 cortex screw and 4 mm cancellous screw
4.5 mm cortex screw & 6.5mm canellous screw
4.5 mm cortex screw & 6.5mm canellous screw
Sizes of LCDCPName of plate Small Narrow Broad
Width 11 mm 13.5 mm 17.5mm
Profile 4 mm 5.4 mm 5.4 mm
Screw 2.7 , 3.5 and 4 mm cancellous screw
4.5 mm & 6.5mm canellous screw
4.5 mm & 6.5mm canellous screw
Name of plate Small Narrow Broad
Width 11 mm 13.5 mm 17.5mm
Profile 4 mm 5.0 mm 5.0 mm
Screw 4 mm locking screw 5 mm locking screw 5 mm locking screw
Sizes of LCP
![Page 39: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/39.jpg)
![Page 40: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/40.jpg)
LISS System
Preshaped plates with self drilling self tapping screws with threaded heads.
Through a small incision (using this jig ) plate is slid along the bone surface. position of plate and wire are checked radiologically before insertion of metaphyseal screw .
![Page 41: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/41.jpg)
LISS-Less Invasive Stabilization System
![Page 42: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/42.jpg)
LISS
![Page 43: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/43.jpg)
![Page 44: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/44.jpg)
Type of plate – Functional Regardless of their length, thickness,
geometry, configuration and types of hole, all plates may be classified in to 4 groups according to their function.
1. Neutralization plate.2. Compression plates.3. Buttress plate.4. Tension band plates.
![Page 45: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/45.jpg)
NEUTRALIZATION PLATE
• Acts as a ""bridge”” protection
• No compression at the fracture site
• neutralization plate is to protect the screw fixation of • a short oblique fracture
• a butterfly fragment
• a mildly comminuted fracture of a long bone
• fixation of a segmental bone defect in combination with bone grafting.
![Page 46: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/46.jpg)
The Neutralization Plate
Lag screws: compression and
initial stability
Plate: protects the
screws from bending and torsional loads
![Page 47: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/47.jpg)
NEUTRALIZATION PLATE
![Page 48: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/48.jpg)
COMPRESSION PLATE
• produces a locking force across a fracture site
• plate is attached to a bone fragment then pulled across the fracture site , producing tension in the plate
• Direction of the compression force is parallel to the plate
![Page 49: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/49.jpg)
Compression Plate
Principle :
- a self compression plate due to the special geometry of screw holes which allow the axial compression.
![Page 50: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/50.jpg)
Dynamic compression principle:a The holes of the plate are shaped like an inclined and transverse cylinder.b–c Like a ball, the screw head slides down the inclined cylinder.d–e Due to the shape of the plate hole, the plate is being moved horizontally when the screw is driven home.f The horizontal movement of the head, as it impacts against the angled side of the hole, results in movement of the plate and the fracture fragment already attached to the plate by the first screw (1).
This leads to compression of the fracture.
![Page 51: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/51.jpg)
51
Compression plate:eccentric DC (dynamic compression) hole
Removable device:compression device
Interfragmentary compression by plate
![Page 52: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/52.jpg)
52
External compression device
![Page 53: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/53.jpg)
METHODS OF ACHIEVING COMPRESSION
With tension devise By overbending With dynamic compression principle
(DCP/LC-DCP) By contouring plate Additional lag screw thro plate
![Page 54: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/54.jpg)
BUTTRESS PLATE• is to strengthen
(buttress) a weakened area of cortex
• The plate prevents the bone from collapsing during the healing process.
• A buttress plate applied a force to the bone which is perpendicular (normal) to the flat surface of the plate.
![Page 55: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/55.jpg)
• The fixation to the bone should begin in the middle of the plate, closest to the fracture site on the shaft. The screws should then be applied in an orderly fashion, one after the other, towards both ends of the plate.
example : the T-plate used for the fixation of fractures of the distal radius and the tibial plateau.
BUTTRESS PLATE
![Page 56: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/56.jpg)
Bridge Plating
Bridge Plating for comminuted fracture
-instead of individually fixing each fragment
-minimal disruption to blood supply
-reduction is performed indirectly
- compression is only sometimes possible
![Page 57: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/57.jpg)
TENSION BAND PRINCIPLE
Tension-band principle.
![Page 58: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/58.jpg)
Tension Band Principle :- Its describes how the tensile forces are
converted into compressive forces by applying a devise eccentrically or to the convex side of a curved tube or bone.
Indications :- Fracture Patella, olecranon, medial
malleolus, greater trochanter of the femur.
Static Dynamic
![Page 59: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/59.jpg)
Dynamic and static tension band
In dynamic tension band the tensile forces are converted to compression on the convex side of an eccentrically loaded bone
Examples : Patella Olecranon Greater tuberosity
Tension band principle to the medial malleolus example of static tension band
![Page 60: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/60.jpg)
HOW MANY SCREWS ?
Bones No. of Cortices
No. of Holes
Type of Plate
Forearm 5 to 6 Cortex 6 holes Small 3.5
Humerus
7 to 8 Cortex 8 holes Narrow 4.5
Tibia 7 to 8 Cortex 7 holes Narrow 4.5
Femur 7 to 8 Cortex 8 holes Narrow 4.5
Clavicle 5 to 6 Cortex 6 holes` Small 3.5
![Page 61: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/61.jpg)
Timing of Plate Removal
Malleolar fractures. 8-12mo The tibial pilon. 12-18mo The tibial shaft. 12-18mo The tibial plateau. 12-18mo The femoral condyles. 12-24mo The femoral shaft. 24-36mo Upper extremity. 12-18mo Shaft of radius / ulna. 24-28mo Distal radius. 8-12mo Metacarpals. 4-6mo
![Page 62: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/62.jpg)
SCREW: INTRODUCTION
An elementary machine to change the small applied rotational force into a large compression force
Function Holds the plate or other prosthesis to the
bone Fixes the # fragments ( Position screw) Achieves compression between the #
fragments (Lag screw)
![Page 63: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/63.jpg)
SCREWS
4 functional parts Head Shaft\ Shank \Core Thread Tip
![Page 64: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/64.jpg)
Head: Recess Types
1. Slotted 2. Cruciate 3. Philips 4. Hex/ Allen 5. Torx (eg Stardrive of Synthes)
![Page 65: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/65.jpg)
Screw: Shaft/ Shank/ Core Smooth link Almost not present in standard
cortex screw Present in cortical SHAFT SCREW or
cancellous screw
![Page 66: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/66.jpg)
Screw: Run out
Transition between shaft and thread Site of most stress riser Screw break
Incorrectly centered hole Hole not perpendicular to the plate
![Page 67: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/67.jpg)
Screw: Thread
Inclined plane encircling the root Single thread May have two or more sets of
threads
V-thread profile: more stress at sharp corner
Buttress thread profile: less stress at the rounded corner
![Page 68: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/68.jpg)
Core
Solid section from which the threads project out wards. The size of core determines the strength of screw and its fatigue resistance. The size of drill bit used is equal to the core diameter.
![Page 69: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/69.jpg)
Screw: Core Diameter
Narrowest diameter across the base of threads
Also the weakest part Smaller root shear
off Torsional strength
varies with the cube of its root diameter
![Page 70: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/70.jpg)
Screw: Thread Diameter
Diameter across the maximum thread width
Affects the pull out strength
Cancellous have larger thread diameter
![Page 71: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/71.jpg)
Screw: Tip Designs
1. Self-tapping tip: Flute Cuts threads in the bone over which
screw advances Cutting flutes chisel into the bone
and direct the cut chips away from the root
![Page 72: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/72.jpg)
Screw: 2.Non self tapping
Lacks flutes Rounded tip Must be pre-cut in the pilot hole by
tap Pre-tapped threads help to achieve
greater effective torque and thus higher inter-fragmental compression
Better purchase
![Page 73: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/73.jpg)
Screw: 3.Corkscrew tip
Thread forming tips In Cancellous screws
which form own threads by compressing the thin walled trabecular bone
Inadequate for cortical bone
![Page 74: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/74.jpg)
Screw: 4.Trochar Tip
Like self tapping Displaces the bone as it
advances Malleolar screw Schanz screws Locking bolts for IMIL
![Page 75: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/75.jpg)
Screw: 5.Self drilling self tapping
Like a drill bit In locked internal
fixator plate hole Pre-drilling not
required Flute Good purchase in
osteoporotic and metaphyseal area
![Page 76: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/76.jpg)
Locking Screws vs Cortical Screws
Creates Fixed Angle Generates Friction/Compression
4.4mm Core Dia. 3.5mm Core Dia.
5.0 mm Locking Screw4.5 mm Cortical Screw
![Page 77: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/77.jpg)
Bending stiffness proportional to the core diameter
Pull out strength is proportional to the size of the thread
Cannulated screws have less bending stiffness
![Page 78: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/78.jpg)
AO/ASIF Screws: Types
Cortical Fully threaded Shaft screw 1.5:phalanx *drill bit 1.1 mm 2.7: mc and phalanx *bit:2.0 3.5: Radius/ Ulna/ Fibula/
Clavicle*bit:2.5 4.5: Humerus/ Tibia/ femur *bit:3.2
![Page 79: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/79.jpg)
AO/ASIF Screws
• Cortical screws:– a machine type– Smaller threads– Lower pitch– Large core diameter– Smaller pitch higher
holding power– greater surface area of
exposed thread for any given length
– better hold in cortical bone
![Page 80: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/80.jpg)
AO/ASIF Screws
• Cancellous screws:– a wood type– core diameter is less– the large threads– Higher pitch– Greater surface are for purchase– Untaped pilot hole– Pilot hole equals the core diameter– lag effect option with partially
threaded screws – theoretically allows better fixation in
soft cancellous bone.
![Page 81: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/81.jpg)
Cancellous Fully threaded
Cannulated or Non- cannulated Partially threaded
16mm or 32 mm Cannulated or Non-cannulated
4.0, drill bit 2.5mm humeral condyle
6.5 drill bit 3.2mm tibial and femoral condyle
![Page 82: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/82.jpg)
MALLEOLAR SCREW: - smooth shaft - partially threaded - trephine tip : no tapping
needed - was designed as lag screw for
malleoli fixation NOW small cancellous screws preffered
- distal humerus and lesser trochanter - size : 25mm – 75 mm
![Page 83: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/83.jpg)
Cannulated screws 3.0 4.0 4.5 6.5 7.0 7.3
![Page 84: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/84.jpg)
Special Screws
Locking bolt Herbert Screw Dynamic Hip
Screw Malleolar
Screw Interference
screw Acutrak screw
![Page 85: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/85.jpg)
Pedicle screw
Suture anchor
![Page 86: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/86.jpg)
Headless Screws
Herbert screw bridging a scaphoid fracture
Acutrak screw
![Page 87: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/87.jpg)
Bioabsorbable Screws
The most common materials used are polylactic acid (PLA), poly-L-lactic acid (PLLA), and polyglycolic acid.
![Page 88: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/88.jpg)
Advantages of bioabsorbable screws
• Does not interfere with MRI.• Does not interfere with future revision surgery if needed.• Decreased incidence of graft laceration.• Does not need implant removal
![Page 89: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/89.jpg)
Disadvantages of bioabsorbable screws
• Major disadvantage is screw failure during insertion. Special screw drivers that span the entire length of screw reduce incidence of screw breakage.• Foreign body reaction may be seen in some.
![Page 90: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/90.jpg)
Function or mechanism.
Neutralization screws – neutralizes forces on the plate in plate fixation.
Lag screws – For inter-fragmentary compression.
Reduction screw – To reduce displaced fracture by pushing or pulling.
Position screw – Holds two fragments in position without compression. Eq. Syndesmotic screw
Anchor screw – Acts as an anchor for wire or suture. In tension band wiring
Locking head screw – In locking plates Locking screw – In interlocking nails Poller screw – To guide the nail path in
interlocking nailing of fractures close to the bone ends.
![Page 91: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/91.jpg)
TAP To cut threads in bone of same size as
the screw to facilitate insertion Flutes : to clear the bone debris Two turns forward and half turn
backward recommended to clear debris Used with sleeve Done manually Power tapping NOT recommended For cancellous bone : short and wide
thread , slightly smaller dia than screw
![Page 92: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/92.jpg)
For cortical screws : - as fixation screw : both cortices - as lag screw : only far cortex For cancellous screw: - only near cortex - sometimes in young patients
tapping entire screw length needed
![Page 93: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/93.jpg)
LAG SCREW TECHNIQUE :
- to achieve interfragmentary compression
- this technique is used if a screw is to be inserted across a # , even through a plate.
- screw has no purchase in near fragment, thread grips the far fragment only
-achieved either with screw with shaft or fully threaded screw
![Page 94: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/94.jpg)
Positioning of screws:
-max. interfragmentary compression :
placed in middle of fragment, right angle to fracture plane
- max. axial stability: right angle to long
axis of bone
![Page 95: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/95.jpg)
![Page 96: Plates and screws 11](https://reader035.fdocuments.in/reader035/viewer/2022062216/55c6d403bb61ebfc428b45e9/html5/thumbnails/96.jpg)
Thank You