Ribbon Electron Beam Profile Monitor For Bunched Beam Tomography
Planmeca ProMax tomography. Four tomography modes Wide beam linear tomography (film) Wide beam true...
-
Upload
lyric-lockridge -
Category
Documents
-
view
220 -
download
0
Transcript of Planmeca ProMax tomography. Four tomography modes Wide beam linear tomography (film) Wide beam true...
Planmeca ProMax tomography
Planmeca ProMax tomographyFour tomography modes
•Wide beam linear tomography (film)
•Wide beam true linear tomography (film)
•Digital linear tomography•Digital Transtomography™
Film tomography
Wide beam linear tomography
• Cassette is perpendicular to the X-ray beam
Wide beam true linear tomography
• Cassette is kept parallel to the image layerConstant resolution over the whole
exposed area
Linear vs true linear tomoraphy
pin phantom
Linear:
• The thickness of the layer is not regular
True linear:
• The thickness of the layer is regular
Linear vs true linear tomoraphy
Tomography
Linear:
• The resolution is optimum only in the middle of the image
True linear:
• The resolution is constant all over the image
Digital tomography
Digital Transtomography™
Multiple sweep - the beam makes oscillating and transitional movement
Straight layers, adjustable thickness
Better layer separation
Digital linear tomography One sweep - the rotating point moves
along the layer during imaging Straight 6mm layer Fast
Tomography features• Straight cuttings• Layer thickness selectable (1, 3, 6,
9, 18, 36 mm)• Adjustable step between layers• Constant 1.50x magnification• Pre-programmed cutting sites
(based on jaw size)• Pre-programmed cutting angles
(based on jaw shape)• Automatic or manual operationsSpecialised imaging at the dental
practice High value for implantology
Cross-sectional tomograms• 1-4 cross-sectional images (manual)• 4 cross sectional images (automatic)• 3 cross sectional images with different angles (automatic)• 4 cross-sectional with anatomical angles• 3 cross sectional images from sinus and nasal cavity (automatic)
Tomography programs
Longitudinal tomograms•1-4 longitudinal images (manual)•3 longitudinal images (automatic)•3 longitudinal images with different angles (automatic)•3 longitudinal images from the sinus and nasal cavity
transtomogram
Tomography programs
Mixed tomograms• 1 cross-sectional and 1 longitudinal
images (automatic)• 3 cross-sectional images and 1
longitudinal image (automatic)• 3 cross-sectional images with
different angles and 1 longitudinal image (automatic)•Mixed tomograms give 3D-
information from the target site• Necessary for exact
measurementstranstomogram
Tomography programs
Taking tomogram
Step #1 Take impression
Step #2 Mark the target
Step #3 Impression in machine
Step #4 Make program selections
Step #5 Set layer to mark
Step #6 Patient in machine
Step #7 Touch Ready
Step #8 Exposure
Taking tomogram
Step #1 Take Impression
Taking tomogram
Step #1 Take impression Step #2 Mark the target
Taking tomogram• Step #3 Impression in machine
Taking tomogram
Step #4 Make program selections
Taking tomogram
• ProMax's concept positions the patient automatically so the area to be exposured is precisely within the focal layer.• The operator simply selects the
proper coordinates of the object, and ProMax goes to the right position.
Taking tomogram
Step #5 Set layer to mark with joystick Adjust angle with the angle buttons
Taking tomogramStep # 6 Patient in machine
Fine adjust the target with the positioning lights
Patient bites the impression When the target is upper jaw, adjust the maxillary alveolar ridge horizontal
When the target is lower jaw adjust the lower part of the mandible horizontal
Taking tomogram
Step # 7 touch Ready Step # 8 Press exposure button
3 x CRS 1xLNG Auto Angle
Taking tomogram
The diagnostic set must include the longitudinal and the one best cross-sectional images.
More information:
Erkki Hiltunen
Product Manager, X-rays
tel: +358 20 7795 456
Mark Niemi
Product Manager, X-rays
tel: +358 20 7795 743
www.planmeca.com
The End