Pipe Clamp Temp Sensor

12
Product Data Sheet November 2014 00813-0100-4952, Rev FA Non-intrusive design for fast and easy temperature measurement in piping applications Platinum RTD Temperature Sensors with silver or nickel tip Integrated temperature assemblies provide time and cost savings Integral assembly with 248 or 648 Wireless Temperature (as shown above) Transmitter provides unparalleled mounting flexibility and accuracy for field-mounted pipeline tempera ture measurement s Rosemount 0085 Pipe Clamp Sensor

Transcript of Pipe Clamp Temp Sensor

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 1/12

Product Data SheetNovember 2014

00813-0100-4952, Rev FA

Non-intrusive design for fast and easy temperature measurement in piping applications

Platinum RTD Temperature Sensors with silver or nickel tip

Integrated temperature assemblies provide time and cost savings

Integral assembly with 248 or 648 Wireless Temperature (as shown above) Transmitter provides

unparalleled mounting flexibility and accuracy for field-mounted pipeline temperature measurements

Rosemount 0085 Pipe Clamp Sensor

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 2/12

2

Rosemount 0085 Pipe Clamp Sensor November 2014

www.rosemount.com

Rosemount 0085 Pipe Clamp Sensor

Proven pipe clamp sensors deliver excellent performance and reliability

Superior accuracy and stability

Improved response time with silver and nickel tip

Easy implementation and installation in existing application

Available in a wide variety of pipe sizes and material

Installation with only two bolts – no welding needed

Optimized surface contact by spring loaded sensor design

Minimized risk of sensor failure and unplanned shutdowns

Avoids stresses related to flow, pressure, chemical contact, abrasion, vibration, and bending

Maintenance of sensor without shutdown of process

Achieve optimal efficiency with Rosemount wireless transmitter offering

Measure your temperature anywhere

Explore the benefits of a Complete Point Solution from Rosemount Temperature Measurement

An “Assemble to Transmitter” option enables Emerson to provide a complete point

temperature solution, delivering an installation-ready transmitter and sensor

assembly

Emerson has a complete portfolio of Single Point, High Density and Wireless

Temperature Measurement solutions, allowing you to effectively measure and

control your processes with the reliability you trust from Rosemount products

Experience global consistency and local support from numerous worldwide Rosemount Temperature

sites

Experienced Instrumentation Consultants help select

the right product for any temperature application and

advise on best installation practices

An extensive global network of Emerson service and supportpersonnel can be on-site when and where they are needed

Contents

Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Pipe Clamp Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 3/12

3

Rosemount 0085 Pipe Clamp SensorNovember 2014

www.rosemount.com

Ordering Information

The Rosemount 0085 Pipe Clamp Sensor is designed for fast and easy non-intrusive

surface temperature measurements in piping applications.

Features include:

Temperature range of -50 to 300 °C (-58 to 572 °F) for silver tip, -200 to 300 °C

(-328 to 572 °F) for nickel tip

Suitable for pipe sizes 1/2-in. to 48-in. (22 mm to 1219 mm)

Single or Dual Element Class A Sensor

Assemble to Transmitter Option

Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See

page 7 for more information on Material Selection.

 Table 1. Rosemount Pipe Clamp Sensor Ordering TableH The Standard offering represents the most common options. The starred options (H) should be selected for best delivery.

__The Expanded offering is subject to additional delivery lead time.

Model Product description

0085 Non Intrusive Pipe Clamp Sensor H

Code Connection head IP rating Conduit entry

C Connection Head Rosemount, Aluminum 68 M20 x 1.5 H

D Connection Head Rosemount, Aluminum 68 1/2-in. NPT H

G Connection Head Rosemount, Stainless Steel 68 M20 x 1.5 H

H Connection Head Rosemount, Stainless Steel 68 1/2-in. NPT H

N No connection head H

1 Connection Head Rosemount, Aluminum with LCD display cover 68 M20 x 1.5 H

2 Connection Head Rosemount, Aluminum with LCD display cover 68 1/2 in. NPT H

3 Connection Head Rosemount, Stainless Steel with LCD display cover 68 M20 x 1.5 H

4 Connection Head Rosemount, Stainless Steel with LCD display cover 68 1/2 in.NPT H

Code Sensor connection

3 Spring Loaded Adapter H

5 Spring Loaded Adapter with Terminal Block H

Code Sensor type Temperature range

P1 RTD, Single Element, 4-Wire, Silver tip -50 to 300 °C (-58 to 572 °F) H

P2 RTD, Dual Element, 3-Wire, Silver tip -50 to 300 °C (-58 to 572 °F) H

P3 RTD, Single Element, 4 Wire, Nickel tip -200 to 300 °C (-328 to 572 °F) H

P4 RTD, Dual Element, 3 Wire, Nickel tip -200 to 300 °C (-328 to 572 °F) H

Code Extension type Head connectionInstrumentconnection

Material

 J Nipple -Union None 1/2 in. NPT Stainless steel H

N No extension (sensor only option) H

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 4/12

4

Rosemount 0085 Pipe Clamp Sensor November 2014

www.rosemount.com

Code Extension length (N)

0080 80 mm H

0150 150 mm H

XXXX Non Standard lengths 200 - 500 mm - Available in 50 mm increments

Code Pipe clamp material

N No Clamp (sensor only option) H

P ASTM 304 SST (1.4301) H

B Duplex F51 (1.4462)

C Carbon steel (1.0037)

Code Inner diameter (D) Suitable pipe size inchSuitable pipesize DIN

Clamp/boltdimensions

0022 22 mm1

/2-in. DN15 30 x 5 mm, M10 H

0034 34 mm 1-in. DN25 30 x 5 mm, M10 H

0061 61 mm 2-in. DN50 40 x 6 mm, M12 H

0089 89 mm 3-in. DN80 40 x 6 mm, M12 H

0115 115 mm 4-in. DN100 50 x 8 mm, M16 H

0140 140 mm 5-in. DN125 50 x 8 mm, M16 H

0169 169 mm 6-in. DN150 50 x 8 mm, M16 H

0220 220 mm 8-in. DN200 50 x 8 mm, M16 H

0273 273 mm 10-in. DN250 60 x 8 mm, M20 H

0027 27 mm 3/4-in. DN20 30 x 5 mm, M10

0030 30 mm N/A DN25 30 x 5 mm, M10

0043 43 mm 11/4-in. DN32 30 x 5 mm, M10

0049 49 mm 11/2-in. DN40 30 x 5 mm, M10

0077 77 mm 21/2-in. DN65 40 x 6 mm, M12

0159 159 mm N/A DN150 50 x 8 mm, M16

0306 306 mm N/A N/A 60 x 8 mm, M20

0324 324 mm 12-in. DN300 60 x 8 mm, M20

0356 356 mm 14-in. DN350 60 x 8 mm, M20

0368 368 mm N/A DN350 60 x 8 mm, M20

0407 407 mm 16-in. DN400 70 x 10mm, M24

0458 458 mm 18-in. DN450 70 x 10 mm, M24

0508 508 mm 20-in. DN500 70 x 10 mm, M24

0521 521 mm N/A DN500 70 x 10 mm, M24

0610 610 mm 24-in. DN600 70 x 10 mm, M24

0660 660 mm 26-in. N/A 70 x 10 mm, M24

0720 720 mm N/A N/A 70 x 10 mm, M24

0762 762 mm 30-in. DN790 70 x 10 mm, M24

0813 813 mm 32-in. DN900 70 x 10 mm, M24

 Table 1. Rosemount Pipe Clamp Sensor Ordering Table

H The Standard offering represents the most common options. The starred options (H) should be selected for best delivery.

__The Expanded offering is subject to additional delivery lead time.

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 5/12

5

Rosemount 0085 Pipe Clamp SensorNovember 2014

www.rosemount.com

Code Inner diameter (D) Suitable pipe size inchSuitable pipesize DIN

Clamp/boltdimensions

0915 915 mm 36-in. DN1000 70 x 10 mm, M24

1016 1016 mm 40-in. N/A 70 x 10 mm, M24

1070 1070 mm 42-in. N/A 70 x 10 mm, M24

1219 1219 mm 48-in. N/A 70 x 10 mm, M24

Code Corrosion protection inlay

N None H

A Material NBR

Options (include with selected model number)

Code Sensor options

A1 Single Element Class A Sensor from -50 to 300 °C (-58 to 572 °F) H

A2 Dual Element Class A Sensor from -50 to 300 °C (-58 to 572 °F) H

Code Assemble to option

XA Assemble Sensor to Specific Temperature Transmitter H

Code Cable gland options

G2 Cable Gland, Ex d, Brass, 7.5-11.9 mm H

G7 Cable Gland, M20x1.5, Ex e, Blue, Polyamide, Diam 5-9 mm H

Code Product certifications

E1 ATEX Flameproof   H

E7 IECEx Flameproof and Dust H

E5 FM Explosion-Proof   H

E6 CSA Explosion-Proof   H

EM Technical Regulations Customs Union (EAC) Flameproof H

Code Cover chain option

G3 Cover Chain - only available with Rosemount Connection Head Material Codes C, D, G, and H H

Code Product certifications

LT Special Material to meet extended Temperature range of -51 °C H

Typical pipe clamp model number: 0085 C 3 P 1 J 0080 P 0061 NReplacement sensor only model number: 0085 N3 P 1 N 0080 N 0061 N

 Table 1. Rosemount Pipe Clamp Sensor Ordering Table

H The Standard offering represents the most common options. The starred options (H) should be selected for best delivery.

__The Expanded offering is subject to additional delivery lead time.

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 6/12

6

Rosemount 0085 Pipe Clamp Sensor November 2014

www.rosemount.com

Overview

Rosemount Pipe Clamp overview

Emerson offers a range of RTDs alone, or as integrated

temperature assemblies including Rosemount TemperatureTransmitters and connection heads.

Rosemount Pipe Clamp Platinum RTD Sensors are highly linear

and have a stable resistance versus temperature relationship.

They are used primarily in industrial environments where high

accuracy, durability, and long-term stability are required, and

are designed to meet the most critical parameters of

international standards: DIN EN 60751/IEC 751 1983

incorporating Amendments 1 and 2.(1)

Rosemount Pipe Clamp Sensors are available in single and dual

element types.

Selecting the extension length for a

pipe clamp sensor

A direct mounting configuration allows heat from the process,

aside from ambient temperature variations, to transfer from the

pipe clamp to the transmitter housing. If the expected pipe

surface temperature is near or above the transmitter

specification limits, consider using additional extension length

or a remote mounting configuration to isolate the transmitter.

Figure 1 provides an example of the relationship between

transmitter housing temperature rise and distance from the

process.

Figure 1. Transmitter Housing Temperature Rise vs.

Uninsulated Distance from the Process

Example

The rated ambient temperature specification for the transmitter

is 85 °C. If the maximum ambient temperature is 40 °C and the

temperature to be measured is 540 °C, the maximum allowable

housing temperature rise is the rated temperature specification

limit minus the existing ambient temperature (85 – 40), or

45 °C.

As shown in Figure 1, an uninsulated distance from the process

of 90 mm will result in a housing temperature rise of 22 °C.

Therefore, 100 mm would be the minimum recommended

distance from the process providing a safety factor of about

25 °C. A longer length, such as 150 mm, is desired to reduce

errors caused by transmitter temperature effect, although in

that case the transmitter may require extra support.

Sensor tip material configuration

The pipe clamp sensor tip is constructed from silver or nickel for

better thermal conductivity and to reduce the thermal response

time. The silver tip has a slightly faster response time while the

nickel tip has a larger temperature range, which allows for

cryogenic applications. The silver tip temperature range is -50 to

300 °C (-58 to 572 °F), and the nickel tip temperature range is

-200 to 300 °C (-328 to 572 °F).

(1) 100   at 0°C,   = 0.00385   x °C/  .

60

50

40

30

20

10

0

135 160 185 210

8  1  5  C   

 p r  o c  e s  s  

 p r  o c  e s  s  

p r o c e s s 

t  e m  p e r  a t  u r  e t e m p e r a t u r e 

t emper at ur e

235 260 285

Uninsulated distance from the process (mm)

    H   o   u   s    i   n   g   r    i   s   e   a    b   o   v   e   a   m    b    i   e   n   t    (    C    )

5    4    0    

C    

2  5  0  C  

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 7/12

7

Rosemount 0085 Pipe Clamp SensorNovember 2014

www.rosemount.com

Pipe Clamp Specifications

Material selection

Emerson provides a variety of Rosemount product with variousproduct options and configurations including materials of

construction that can be expected to perform well in a wide

range of applications. The Rosemount product information

presented is intended as a guide for the purchaser to make an

appropriate selection for the application. It is the purchaser’s

sole responsibility to make a careful analysis of all process

parameters (such as all chemical components, temperature,

pressure, flow rate, abrasives, contaminants, etc.), when

specifying product, materials, options and components for the

particular application. Emerson Process Management is not in a

position to evaluate or guarantee the compatibility of the

process fluid or other process parameters with the product,options, configuration or materials of construction selected.

Rosemount Pipe Clamp Platinum RTD

Nominal resistance

In accordance with DIN EN 60751/IEC 751 1983 incorporating

Amendments 1 and 2, the nominal resistance is defined:

100 RTD at 0 °C

 = 0.00385 x °C/., averaged between 0 and 100 °C

Limit deviations

Tolerance Class B, as standard t = ± (0.3 + 0.005 x [t]);

temperature range -200 to 300 °C (-328 to 572 °F)

Tolerance Class A, as option t = ± (0.15 + 0.002 x [t]);

temperature range -50 to 300 °C (-58 to 572 °F)

Process temperature range

-200 to 300 °C (-328 to 572 °F)

Ambient temperature range

-40 to 85 °C (-40 to 185 °F)

Self-heating

0.15 K/mW when measured as defined in DIN EN 60751; 1996.

Insulation resistance (RTD)

1,000 M minimum insulation resistance when measured at

500 V dc at room temperature

Sheath material

321 SST with mineral insulated cable construction and silver or

nickel tip

Lead wires

PTFE insulated, silver-coated copper wire (Figure 2)

Identification data

The model and serial numbers are engraved directly on the

spring loaded adapter.

Ingress Protection (IP) rating for connection head

IP68 and NEMA 4X

Figure 2. Sensor Lead Wire Termination

Pipe Clamp RTD spring loaded

Single element 4-wire Dual element 3-wire

Pipe Clamp RTD spring loaded

Single element 4-wire Dual element 3-wire

Red

Red

White

White

Black 

Red

Red

Green

Blue

Blue

White

White

Red

Red

Black 

Red

Red

Blue

Blue

Green

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 8/12

8

Rosemount 0085 Pipe Clamp Sensor November 2014

www.rosemount.com

Dimensional Drawings

Figure 3. 1/2-in. ANPT Spring Loaded Adapter

Figure 4. Pipe Clamp Sensor Assembly with Rosemount 3144P

Dimensions are in millimeters.

L

11 2+

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 9/12

9

Rosemount 0085 Pipe Clamp SensorNovember 2014

www.rosemount.com

Figure 5. Pipe Clamp Sensor Assembly with Rosemount Connection Head

Figure 6. Pipe Clamp Sensor Assembly with 648 Wireless Transmitter

Dimensions are in millimeters.

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 10/12

10

Rosemount 0085 Pipe Clamp Sensor November 2014

www.rosemount.com

Accessories

 Table 2. Connection Head

Figure 7. Connection Head Dimensional Drawing

Part number Model/material IP ratingConduit

connection

Process

connection00644-4410-0011 Rosemount, Aluminum 68 1/2 in. NPT 1/2 in. NPT

00644-4410-0021 Rosemount, Aluminum 68 M20 x 1.5 1/2 in. NPT

00644-4410-0111 Rosemount, Aluminum with LCD Display Cover 68 1/2 in. NPT 1/2 in. NPT

00644-4410-0121 Rosemount, Aluminum with LCD Display Cover 68 M20 x 1.5 1/2 in. NPT

00644-4411-0011 Rosemount, Stainless Steel 68 1/2 in. NPT 1/2 in. NPT

00644-4411-0021 Rosemount, Stainless Steel 68 M20 x 1.5 1/2 in. NPT

00644-4411-0111 Rosemount, Stainless Steel with LCD Display Cover 68 1/2 in. NPT 1/2 in. NPT

00644-4411-0121 Rosemount, Stainless Steel with LCD Display Cover 68 M20 x 1.5 1/2 in. NPT

With LCD display cover With standard cover

A. LCD Display

Dimensions are in millimeters.

128

78

A

104   104

100

12878

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 11/12

11

Rosemount 0085 Pipe Clamp SensorNovember 2014

www.rosemount.com

Product Certifications

European Directive Information

A copy of the EC Declaration of Conformity can be found at theend of the Quick Start Guide.The most recent revision of the EC

Declaration of Conformity can be found at

www.rosemount.com.

Ordinary Location Certification fromFM Approvals

As standard, the transmitter has been examined and tested to

determine that the design meets the basic electrical,

mechanical, and fire protection requirements by FM Approvals,

a nationally recognized test laboratory (NRTL) as accredited by

the Federal Occupational Safety and Health Administration(OSHA).

North America

E5 FM Explosionproof and Dust-Ignitionproof Certificate: 0R7A2.AEStandards: FM Class 3600- 2011, FM Class3615-2006, FM Class 3810-2005, ANSI/NEMA250-1991

Markings: XP CL I, DIV 1, GP B, C, D, T6; DIP CL II/III,DIV 1, GP E, F, G, T6; Type 4X; Installed per00068-0013;

E6  CSA Explosionproof, Dust-Ignitionproof and

Division 2Certificate: 1063635

Standards: CAN/CSA C22.2 No. 0-M91, CSA Std.C22.2 No. 25-1966, CSA Std. C22.2 No. 30-M1986,CSA Std. C22.2 No.94-M91, CSA Std. C22.2 No.142-M1987, CSA Std. C22.2 No. 213-M1987

Markings: XP Class I Groups B, C, and D; DIP Class IIGroups E, F, G; Class III; Class I Div. 2 Groups A, B, C,D; Class I Zone 1 Group IIB+H2; Class I Zone 2 GroupIIC; Installed per 00068-0033;

Europe

E1  ATEX FlameproofCertificate: FM12ATEX0065XStandards: EN60079-0:2012, EN60079-1:2007Markings: II 2 G Ex d IIC T6…T1 Gb

 Special Conditions for Safe Use (X):

1. See certificate for ambient temperature range.

2. The non-metallic label may store an electrostatic chargeand become a source of ignition in Group III environments.

3. Guard the LCD display cover against impact energiesgreater than 4 joules.

4. Consult the manufacturer if dimensional information onthe flameproof joints is necessary.

Technical Regulations Customs Union (EAC)

EM Contact an Emerson Process Management representative

for additional information.

International

E7  IECEx FlameproofCertificate: IECEx FMG 12.0022XStandards: IEC60079-0:2011, IEC60079-1:2007Markings: Ex d IIC T6…T1 Gb

 Special Conditions for Safe Use (X):

1. See certificate for ambient temperature range.2. The non-metallic label may store an electrostatic charge

and become a source of ignition in Group III environments.

3. Guard the LCD display cover against impact energiesgreater than 4 joules.

4. Consult the manufacturer if dimensional information onthe flameproof joints is necessary.

7/23/2019 Pipe Clamp Temp Sensor

http://slidepdf.com/reader/full/pipe-clamp-temp-sensor 12/12

Standard Terms and Conditions of Sale can be found at www.rosemount.com\terms_of_sale.The Emerson logo is a trade mark and service mark of Emerson Electric Co.Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc.All other marks are the property of their respective owners.© 2014 Rosemount Inc. All rights reserved.

Emerson Process ManagementRosemount Inc.8200 Market BoulevardChanhassen, MN 55317 USAT (U.S.) 1-800-999-9307

T (International) (952) 906-8888F (952) 906-8889www.rosemount.com

Emerson Process ManagementBlegistrasse 23P.O. Box 1046CH 6341 BaarSwitzerland

T +41 (0) 41 768 6111F +41 (0) 41 768 6300www.rosemount.com

Emerson Process ManagementAsia Pacific Pte Ltd1 Pandan CrescentSingapore 128461T +65 6777 8211F +65 6777 0947Service Support Hotline: +65 6770 8711Email: [email protected]

Rosemount 0085 Pipe Clamp Sensor00813-0100-4952, Rev FA

Product Data SheetNovember 2014

Emerson Process ManagementLatin America1300 Concord Terrace, Suite 400Sunrise, Florida 33323 USAT + 1 954 846 5030www.rosemount.com