PD. Dr. Markus Ganzera - aoaceurope.com 12.pdf26 CE-MS Application NACE / CE/MS CE-conditions :...
Transcript of PD. Dr. Markus Ganzera - aoaceurope.com 12.pdf26 CE-MS Application NACE / CE/MS CE-conditions :...
1
Capillary Electrophoresis and Capillary Electro Chromatography –
Suitable for quality control and routine analysis of herbal drugs?
Symposium Erlangen October 2011
PD. Dr. Markus Ganzera
Institut of Pharmacy, Pharmacognosy
University of Innsbruck
Innsbruck, Austria
2
Analytical techniques
• trend to miniaturization (e.g. UPLC, CE, CEC, nano-LC, “lab on a chip”)
• established versus innovative
• novel approaches suitable for routine analysis (?)
Symposium Erlangen October 2011
Electrophoretic techniques
• gel electrophoresis
• capillary electrophoresis
• capillary electrochromatography
3
cation25 kV
CE
Capillary electrophoresis (CE, HPCE)
• separations in silica capillaries (50 to 75 µm ID)
• different modes (aqueous, NACE, MEKC)
• advantages: high separation efficiency, versatility, economic use
• disadvantages: sensitivity, reproducibility
Symposium Erlangen October 2011
anion
compoundanode cathode
High voltage unit
25 kV
silica capillary (negatively charged)
EOF EOF
Cationic analyte
Anionic analyte
Neutral analyte
4
Micellar electrokinetic chromatography (MEKC)
• for separation of neutral (and charged) analytes
• cationic / anionic tensides, CMC
• pseudostationary phase
• advantage: high efficiency, low solvent consumption
• disadvantage: reproducibility, sensitivity
http
://es
q-w
ww
.mit.
edu
CE
Symposium Erlangen October 2011
CMC of SDS:9 mM
anode cathode
ww
w.p
harm
chem
.tu-b
s.de
5
CE-MS
CE-ESI-MS Interface
Sheath liquidCE Capillary
Sheath liquid
Nebulizing gas
Nebulizing gas
CE inlet
Sheath liquidCE Capillary
Sheath liquid
Nebulizing gas
Nebulizing gas
CE inlet
CE
Symposium Erlangen October 2011
Drying gas
Sprayer
Nebulizing gas
Drying gas
MS entrancecapillary
HV
-4 kV
Drying gas
Sprayer
Nebulizing gas
Drying gas
MS entrancecapillary
HVHV
-4 kV
© A
gile
nt
6
Capillary electrochromatography
• capillaries (100 µm or smaller ID) filled with stationary phase
• hybrid technique (EOF and stationary phase)
• advantage: versatility, high efficiency, low solvent consumption
• disadvantage: reproducibility, sensitivity, few applications
CEC
Symposium Erlangen October 2011
365 µm OD
100 µm IDstationary phase
capillary
7
CEC separation materials
• packed capillaries (regular HPLC material, 2-5 µm; frits!)
• open tubular columns
• monolithic materials (silica or polymer based)
CEC
Symposium Erlangen October 2011
packed open tubular monolithw
ww
.bar
nett.
neu.
edu
© M
erck
© M
erck
8
air-cooled column cartridge
CE
Symposium Erlangen October 2011
HP 3D CE system, Agilent
9Application MEKC
Leontopodium alpinum
® In
stitu
te o
f Pha
rmac
y, U
nive
rsity
of I
nnsb
ruck
Symposium Erlangen October 2011
family: Asteraceae
origin: Alps, Switzerland, Austria
part used: aerial parts
indications: diarrhoea, bronchitis and fewer
constituents: sesquiterpenes, polyacetylenes,
flavonoids, phenolic acids
® In
stitu
te o
f Pha
rmac
y, U
nive
rsity
of I
nnsb
ruck
® In
stitu
te o
f Pha
rmac
y, U
nive
rsity
of I
nnsb
ruck
10Application MEKC
MEKC-conditions : capillary: barefused silica, 50 µm ID, 52 cm
O
OH O
R1
R2
R3
R4
O
OH
OH
O
OH OR1
OR2HOOC
O
HO
HO
O
O
O
OH
O
O
HO
O
O
OOH
OH
OH
OH
OH
O
Symposium Erlangen October 2011
fused silica, 50 µm ID, 52 cmeffective length; buffer: 60 mMsodium tetraborate solution, adjustedto pH 6.75 with 0.1 % H3PO4,containing 50 mM SDS and 35 %ACN; voltage: 30 kV; temperature: 40°C; detection: 254 nm; sampleinjection: hydrodynamic mode, 50mbar for 4 seconds
Ganzera et al., Electrophoresis, submitted August 2011.
11Application MEKC
3
4
5
6
perc
ent
1
2
3
4
5
6
Flavonoids and phenolic acids in different L. alpinum samples
flavonoids
Symposium Erlangen October 2011
0
1
2
3
perc
ent
6
7
8
9
10
phenolic acids
leontopodic acid
12
Eleutherine americana
ww
w.p
harm
.su.
ac.th
Application CE / CEC
Symposium Erlangen October 2011
ww
w.p
harm
.su.
ac.th
family: Iridaceae
origin: southeast Asia, Thailand
part used: bulb
indications: diuretic, coronary diseases, cold
constituents: naphthoquinone derivatives
13
Established HPLC method
mAU
200
400
eleuthoside B
isoeleutherin
eleutherin
eleutherol
eleutherinoside A
O
O
O
CH3O CH3
CH3
H
O
O
CH3CH3O
OOH
OHHOO
OH
OHHOHO O
O
O
O
CH3
CH3
OH
OOH
OHHOHO
O
O
O
O
CH3O CH3
H
CH3
OHCH3O
O
O
CH3
Application CE / CEC
Symposium Erlangen October 2011
0
200
min0 5 10 15 20 250
40
80
standard
sample
LC-conditions : column: Synergi MAX-RP, 150 x 4.6 mm, 4 µm; mobile phase: water (A), acetonitrile (B); gradient: 15 B in 10 min to 30 B, in 10 min to 50 B, in 10 min to 80 B; flow: 1.0 ml/min; sample: 10 µl; temperature: 40 °C; 254 nm
Paramapojn et al., J. Pharm. Biomed. Anal. 47 (2008) 990-993
14
Development of MEKC method
Challenges
• most analytes are neutral
• solubility of compounds in aqueous systems
• eleutherin and isoeleutherin are isomers
MEKC
solvent in buffer
cyclodextrins
Application CE / CEC
Symposium Erlangen October 2011
0
5
10
15
20
25
30
35
0 25 50 75 100
sodium cholate (mM)
reso
lutio
n
1-2
2-3
3-a
a-4
4.5
15
0
2
4
6
8
10
12
14
16
18
no solvent THF 20 % MeOH 20 % EtOH 20 % ProOH 20% ACN 20 %
reso
lutio
n
1-2 2-3 3-a a-4 4-5
7,00
8,00 1-2 2-3 3-a 4-5
influence of solvent
Application CE / CEC
Symposium Erlangen October 2011
0,00
1,00
2,00
3,00
4,00
5,00
6,00
7,00
8,50 9,00 9,25 9,50 9,75 10,00 10,50
pH-value
reso
lutio
n
influence of pH
Addition of cyclodextrins did not enable separation of 2 and 3 !!
16
mAU
0
4
8
12
1
2+3
4 5
MEKC
Optimized MEKC method
1 eleuthoside B2 isoeleutherin3 eleutherin4 eleutherol5 eleutherinoside A
Application CE / CEC
Symposium Erlangen October 2011
min4 6 8 10 12 14
5 eleutherinoside A
MEKC-conditions : capillary: bare fused silica, 75 µm ID, 52 cm effective length; buffer: 25 mM sodium tetraborate solution, adjusted to pH 9.75 with 0.1 N NaOH, containing 50 mM sodium cholate and 20 % THF; voltage: 25 kV; temperature : 35 °C; detection: 254 nm; sample injection: hydrodynamic mode, 50 mbar for 2 seconds
17
Development of CEC method
1) Selection of monolithic phase (methacrylates, ac rylates)
• hydrophilic (glycidyl methacrylate, ethylene dimethacrylate)
• hydrophobic (butyl methacrylate, ethylene dimethacrylate)
• cationic ((acryloxy)ethyltrimethyl ammonium chloride, butanediol
diacrylate)
Application CE / CEC
Symposium Erlangen October 2011
diacrylate)
2) Selection of mobile phase
• type of electrolyte
• pH value
• molarity
• organic solvent
0
0,2
0,4
0,6
0,8
1
1,2
1,4
4,5 6,6 8,1
pH
reso
lutio
n
1-5 2-3
18
mAU
05
10
mAU
050
100
mAU
0204060
mAU
60 % ACN
70 % ACN
80 % ACN
Influence of acetonitrile concentration
Application CE / CEC
Symposium Erlangen October 2011
min2 4 6 8 10 12 140
102030
90 % ACN
3) Other settings
• voltage and temperature with impact on separation time, not resolution
19
Optimized CEC method
1 eleuthoside B2 isoeleutherin3 eleutherin4 eleutherol
mAU
60
100
140
15
2 34
CEC
Application CE / CEC
Symposium Erlangen October 2011
4 eleutherol5 eleutherinoside A
CEC-conditions : stationary phase: acrylate monolith with cationic properties; capillary: bare fused silica, 100 µm ID, 35 cm effective length;buffer: 3 mM ammonium formate solution (pH 6.6) containing 70 % ACN; voltage: - 25 kV; temperature: 25 °C; detection: 254 nm; sample injection: electrokinetic mode, - 10 kV for 6 seconds; both vials pressurized with 8 bar during analysis
min2 4 6 8 10 12
20
Ganzera et al., Electrophoresis 30 (2009) 3757-3763
20
min2 4 6 8 10 12 14
mAU
0
2
4
6
1 2+3a
4 5
mAU
20 15
MEKC
CEC
Analysis of samples
1 eleuthoside B2 isoeleutherin
Application CE / CEC
Symposium Erlangen October 2011
min2 4 6 8 10 12 14
0
5
10
15
2 3
4
min0 5 10 15 20 25
0
40
80mAU
5 1
23
4HPLC
2 isoeleutherin3 eleutherin4 eleutherol5 eleutherinoside A
Sample: methanolic extract of E. americana bulbs
21
method compound correlation coefficient linear range (µg/ml) LOD (µg/ml) LOQ (µg/ml) recovery
MEKC 1 0.999 20 - 540 5 16 96.5 – 100.5 %
2 0.998 20 - 540 4 13 n.d.
3* - - - - n.d.
4 0.999 20 - 540 5 16 99.4 – 102.7 %
5 0.999 20 - 540 5 15 n.d.
CEC 1 0.998 15 - 480 2 8 97.1 – 103.5 %
2 0.999 15 - 480 4 15 n.d.
Validation data
Application CE / CEC
Symposium Erlangen October 2011
2 0.999 15 - 480 4 15 n.d.
3 0.997 15 - 480 8 25 n.d.
4 0.998 15 - 480 5 16 98.6 – 103.4 %
5 0.999 15 - 480 2 7 n.d.
HPLC 1 0.999 7 - 227 0.4 1.5 98.4 – 102.6 %
2 0.999 6 - 203 0.4 1.3 96.3 – 100.5 %
3 1.000 7 - 214 0.8 2.8 97.1 – 102.5 %
4 0.999 7 - 233 0.5 1.5 96.8 – 101.9 %
5 1.000 6 - 201 0.4 1.3 96.4 – 103.5 %
* co-eluted with compound 2 n.d. not determined
22
Sample Method eleuthoside B isoeleutherin eleutherin eleut herol eleutherinoside A Total
EA-1 MEKC 0.12 (1.02) 0.08 (1.45) -* 0.10 (1.35) 0.03 (0.42) 0.33
CEC 0.13 (0.48) 0.03 (0.24) 0.03 (0.01) 0.09 (0.28) 0.04 (0.24) 0.32
EA-2 MEKC 0.09 (1.04) 0.11 (0.56) -* 0.11 (0.61) 0.03 (1.99) 0.34
CEC 0.10 (1.27) 0.03 (2.08) 0.05 (1.68) 0.10 (0.22) 0.03 (0.71) 0.32
EA-3 MEKC 0.13 (1.51) 0.14 (0.66) -* 0.14 (0.87) 0.05 (0.72) 0.46
CEC 0.13 (0.52) 0.05 (0.88) 0.07 (0.08) 0.11 (0.50) 0.05 (0.33) 0.41
EA-4 MEKC 0.06 (1.36) 0.12 (2.43) -* 0.16 (1.28) 0.04 (0.69) 0.38
Quantitative results
Application CE / CEC
Symposium Erlangen October 2011
0
0,05
0,1
0,15
0,2
0,25
0,3
0,35
MEKC CEC HPLC
perc
ent
5
4
3
2
1
EA-4 MEKC 0.06 (1.36) 0.12 (2.43) -* 0.16 (1.28) 0.04 (0.69) 0.38
CEC 0.09 (2.02) 0.04 (0.54) 0.06 (0.74) 0.18 (2.21) 0.04 (1.06) 0.41
*co-eluted with 2, thus values for 2 reflect sum of both compounds
sample EA-1
2323
Lupinus species
family: Fabaceae
Application NACE / CE/MS
Com
mon
s.w
ikim
edia
.org
Symposium Erlangen October 2011
family: Fabaceae
origin: Mediterranean region, Latin America
part used: seeds (some as food)
constituents: proteins, quinolizidine alkaloids
method of analysis: GC
ww
w.s
cien
ceph
oto.
com
24
NACE
non aqueous CE
buffer in organic solvent
sparteine (1)N
N
HH
H
Application NACE / CE/MS
Symposium Erlangen October 2011
lupanine (2)
angustifoline (3)
13-α hydroxylupanine (4)
N
H
N
N
HH
H
H
O
HN
N
HH
H
H
O
N
N
HH
H
H
O
OHGanzera et al., J. Pharm. Biomed. Anal. 53 (2010) 1231-1235.
25Application NACE / CE/MS
Symposium Erlangen October 2011
26
CE-MS
Application NACE / CE/MS
CE-conditions : fused-silica capillary 60 cm x 50 µm ID; (MS: effective length 90 cm); 100 mM ammonium acetate in MeOH/ACN/water (72/18/10) with 1 % HOAc; hydrodynamic injection at 50 mbar for 2 sec; separation voltage: 25 kV; temperature: 30 °C; detection wavelength: 210 nm
Symposium Erlangen October 2011
MS-conditions : ESI positive; sheath gas: 4 psi; dry gas: 4 l/min; probe temperature: 250 °C; spray voltage: 4.5 kV; sheath liquid: methanol/water = 1/1, with 0.1 % acetic acid; sheath liquid flow rate: 0.20 ml/h
wavelength: 210 nm
27
Species Sparteine Lupanine Angustifol. 13-OH Lup. Others* Σ of alkaloids
L. albus - 1.47 (0.46) 0.59 (0.20) 0.21 (1.59) albine: 0.08 (0.49) 2.35
L. arboreus 2.28 (1.26) 0.17 (0.41) - - 2.45
L. densiflorus - 0.02 (1.39) 0.01 (2.55) 0.02 (2.31) cytisine: 0.11 (1.56)
anagyrine: 0.02 (0.08)
methylcytisine: 0.51 (3.93)
0.69
Quantitative results (in percent; relative standard deviation in parenth esis)
Application NACE / CE/MS
Symposium Erlangen October 2011
L. exaltatus - 0.89 (0.56) 1.25 (0.10) 0.03 (1.23) 2.17
L. hartwegii 0.18 (0.10) 0.79 (2.91) 1.17 (1.29) 0.04 (0.33) 2.18
L. microcarpus 0.10 (0.37) 0.02 (1.84) - - anagyrine: 0.08 (0.96)
methylcytisine: 0.22 (1.90)
0.42
L. polyphyllus - 1.08 (0.62) 3.28 (0.93) 0.77 (0.78) 5.13
* tentatively identified based on MS data, quantified like lupanine
28Conclusions (1)
Conclusions
CE / CEC HPLC
Separation standards
samples
Precision standards
sample
+++ +++
++ +++
+++ +++
++ +++
Symposium Erlangen October 2011
Separation time
Sensitivity
Durability of stationary phase
Cost equipment
analysis
General handling
+++ ++
+ +++
++ +++
+ ++
+++ +
++ +++
29Conclusions (2)
… and the winner is ??
versatilityspeed
sensitivityreproducibility (column to column)
CE / CEC
Symposium Erlangen October 2011
speedease of preparation (monoliths)high efficiencyenormous potentialcost of analysis
reproducibility (column to column)handlingfew CEC materials availablefew applicationstedious sample preparation
Type of analytes and separation requirements will be crucial factors for decision.
30
Symposium Erlangen October 2011
Thank you for your attention.