PCR quantitative en temps réel

24
PCR quantitative en temps réel Lydie Pradel

description

PCR quantitative en temps réel. Lydie Pradel. PCR. PCR semi-quantitative 25 cycles. Sybr Green. Fluorogenic 5’Nuclease Assay. Binds ds DNA. Use Taqman probe. Sybr Green fluoresces upon binding to double stranded PCR product - PowerPoint PPT Presentation

Transcript of PCR quantitative en temps réel

Page 1: PCR quantitative en temps réel

PCR quantitativeen temps réel

Lydie Pradel

Page 2: PCR quantitative en temps réel
Page 3: PCR quantitative en temps réel

PCR

Page 4: PCR quantitative en temps réel

PCR semi-quantitative25 cycles

Page 5: PCR quantitative en temps réel

Sybr Green Fluorogenic 5’Nuclease Assay

Binds dsDNA Use Taqman probe

Page 6: PCR quantitative en temps réel

Sybr Green fluoresces upon binding to double stranded PCR product

Emitted Fluorescence is proportional to amount of amplified product detected in every sample

Page 7: PCR quantitative en temps réel

Specificity check of Sybr Green

Gel or melting curve analysis (Real-time PCR system)

Sharp, single peak indicates specific amplification

Page 8: PCR quantitative en temps réel

Non specific amplification (genomic DNA and RT-qPCR)

Page 9: PCR quantitative en temps réel
Page 10: PCR quantitative en temps réel

Signal generation with TaqMan Probe

Uses2 principles:- FRET technology- 5’-Nuclease activity of the Taq polymerase

PCR specificity (primers)Hybridization specificity (probe)

Dyes: FAM, VIC, TAMRA

Page 11: PCR quantitative en temps réel

TaqMan Probe Sybr Green

Specificityprimer binding Primer bindingProbe hybridizationPCR conditions PCR conditions

FlexibilityMultiplex easy, only primers neededSNP detection

OptimizationCkeck primer dimer formation

Page 12: PCR quantitative en temps réel

Thermal Cycling Protocol (Applied Biosystem)

95°C 10’ Activation of AmpliTaq Gold Polymerase95°C 15’’ Denaturation60°C 1’ Annealing/Extension

Page 13: PCR quantitative en temps réel

This inert dye, whose fluorescence does not change during the reaction, may be added to quantitative, real-time PCR reactions to normalize the well-to-well differences that may occur due to artifacts such as pipetting errors or instrument limitations.

Passive reference ROX dye

ROX dye normalizes for non-PCR related fluorescence variation

FAM dye

ROX dye

RnFAM dye

ROX dye

Rn

Sample 1 Sample 2

Rn= Reporter/Passive reference

Fluo

recs

ence

Fluo

recs

ence

Page 14: PCR quantitative en temps réel

From fluorescence to results

105 104 103 104 103

Page 15: PCR quantitative en temps réel

Primer specificity: efficiency

If slope= -3,32 efficiency becomes 1

Page 16: PCR quantitative en temps réel

Quantification

Absolute quantificationStandard curve

Relative quantification

Relative increase or decrease

No standard curve

Calculation of results by comparison of Ct value« comparative Ct method »

Definition of - Endogenous Control- Calibrator

Page 17: PCR quantitative en temps réel

Endogenous Control (EC)

- Amount of cDNA per well

- Constant expression level in all samples

- EC normalizes for

- RNA input measurement errors

- RT efficiency variations

Ex: Actine, GAPDH …

Page 18: PCR quantitative en temps réel

Calibrator: an example using four samples

timet=0 t=12 t=24 t=48

Total RNA Total RNA Total RNA Total RNA

cDNA cDNA cDNA cDNA

Calibrator

Page 19: PCR quantitative en temps réel

Comparison of Target Gene and Endogenous Control

Ct=24-14=10

Ct=14 Ct=24

Rn

Cycles

Target gene

Endogenous control

What if we added the double amount of cDNA ?

Page 20: PCR quantitative en temps réel

What if we added the double amount of cDNA ?

Ct=14 Ct=24

Ct=23-13=10

Rn

Cycles

Target gene

Endogenous control

Ct=13 Ct=23

Page 21: PCR quantitative en temps réel

Ct=15 Ct=35

Rn

Cycles

EC

TG

Comparative Ct method: an example using the four samples

Ct=15 Ct=30

Rn

Cycles

EC

TG

Ct=9 Ct=24

Rn

Cycles

EC

TG

Ct=14 Ct=34

Rn

Cycles

EC

TG

t=0 t=12h

t=24h t=48h

Page 22: PCR quantitative en temps réel

Comparative Ct Method calculation Steps

Step 1: Normalisation to endogenous control

Ct target gene – Ct Endogenous gene =Ct

(do both for calibrator and sample)

Step 2: Normalization to calibrator sample

Ct Sample - Ct Calibrator = Ct

Step3: Use the formula

2-Ct

Page 23: PCR quantitative en temps réel

Ct=15 Ct=35

Rn

Cycles

EC

TG

Ct=15 Ct=30

Rn

Cycles

EC

TG

t= 0 t= 12h

t= 0 t= 12h

Threshold Threshold

Ct = Ct target gene – Ct Endogenous gene

Ctt=0 35-15= 20 Ctt=12 30-15= 15

Ct = Ct Sample - Ct Calibrator

Ct 15-20= -5

2-Ct

2^-(-5)= 32

Page 24: PCR quantitative en temps réel

Relative quantification of the 4 samples

50

40

30

20

0

10

Samples

X-f

old

exp

ress

ion

t = 0

t = 12 h

t = 24 h

t = 48 h

Calibratort=0

1

32 35

4