Passive Spectrometer Design
Transcript of Passive Spectrometer Design
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 1/30
DESIGN AND CONSTRUCTION OF A PHOTODIODE-BASED PASSIVE
SPECTROMETER
by
Daniel T.S. Cook
PRINCIPAL SUPERVISORS:
Dr. William Heidbrink
Department of Physics and Astronomy
University of California, Irvine
Dr. Thomas Askew
Physics Department
Kalamazoo College
A paper submitted in partial fulfillment
Of the requirements for the degree of Bachelor of Arts at
Kalamazoo College
Fall Quarter, 2005
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 2/30
Contents
Acknowledgments iii
Abstract iv
I. Introduction 1
A. Plasma 1
B. Nuclear Fusion 1
C. Plasma Confinement 2
D. The Field-Reversed Configuration 3
E. The University of California, Irvine FRC Experiment 3
F. Plasma Diagnostics 4
G. Passive Spectroscopy, Line-Emission Spectroscopy 4
II. Background 5
A. FRC Theory 5
B. Spectroscopy Theory 7
C. Electron Temperature Determination 8
III. Spectrometer Design and Construction 9
A. Experimental Arrangement and Light Collection 10
B. Signal Generation and Amplification 10
C. Spectrometer Calibration 12
D. Spectral Line Ratio Measurements 15
E. Noise Reduction 16
IV. Results and Discussion 18
V. Conclusion 20
A. Relevant Spectral Line Information 23
B. Photodiode Calibration 23
C. Photodiode Characteristics 23
i
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 3/30
D. Filter Characteristics 23
E. Operational Amplifier Characteristics 23
F. Symbols 24
References 24
ii
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 4/30
Acknowledgments
Thank you to William Heidbrink for mentoring, advising, and assisting me throughout
my time on this project; Eusebio Garate and Thomas Askew for also advising me on this
project; Erik Trask and Wayne Harris for advice and assistance inside and outside the
scope of the project; Alan Van Drie for assistance with data collection; Princeton Plasma
Physics Laboratory and the US Department of Energy for supporting this project through
the National Undergraduate Fellowship Program in Plasma Physics and Fusion Energy
Sciences.
iii
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 5/30
Abstract
The use of passive spectroscopy to determine electron temperatures is well documented
and has the important benefit of being noninvasive. The use of line intensities in particular is
important because other spectrographic techniques, such as those based on Stark broadening
or Thompson scattering, can be difficult to employ in certain situations. Here, we present a
model for the estimation of electron temperature utilizing the ratio of emitted line intensities
in a plasma. Details of design and construction for a photodiode based spectrometer taking
advantage of this model are also presented, as well as details regarding data acquisition and
analysis using this spectrometer.
iv
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 6/30
I. INTRODUCTION
A. Plasma
Plasmas, the name given to ionized gases by Nobel laureate Irving Langmuir in the
early 1900’s, are thought to comprise the vast majority of matter in the known universe.
They exist naturally as interstellar clouds and solar wind, on the surface of stars, as the
earth’s ionosphere, around lightning, and in many other places [1]. Indeed, previously in the
history of the universe, all matter is thought to have been plasma [2]. Ever since Langmuir
founded the field of plasma physics through his research of tungsten-filament light bulbs,
the discipline has proved vital to further our understanding of the world around us.
Of more immediate interest to most however, are the man-made plasmas that are usefulto us everyday. From fluorescent lighting and neon signs, to semiconductor chip etching, to
arc welding, and even new plasma televisions, plasmas are becoming a part of our daily life.
And the potential for further exploitation of the unique properties of plasmas is such that
they are likely to become only more prevalent. Perhaps the most interesting, and possibly
the most useful, potential application of plasmas is to generate electrical power through
nuclear fusion.
B. Nuclear Fusion
Nuclear fusion occurs when two nucleii are joined together to produce a heavier nucleus,
generally with some excess energy. The energy required to produce fusion is extreme –
found in nature, for example, in the center of stars. At an energy level this high, neutral
atoms can ionize to become plasma. The process through which this takes place is well
understood. If it can be controlled, and the excess energy produced can be harnessed, it has
the potential to produce nearly boundless amounts of safe, non-polluting electrical power
for the future [3]. The reason that we are not yet using nuclear fusion as a power source is
because of the myriad challenging problems that exist with its actual implementation. The
large amounts of energy required are in fact attainable using our technology and resources.
A problem arises however, when having gathered this amount of energy one must then find
a way to channel it into the ions which are to be fused. After having overcome this hurdle,
which we only recently have, one must then direct the energy in such a way as to cause the
1
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 7/30
ions to collide and fuse. This last problem, the biggest that we still face, is the problem of
confinement.
C. Plasma Confinement
Since we are unable to direct ions well enough to cause collisions with a high probability,
the only way to achieve fusion is to energize the plasma, contain it the best we can, and
wait for the ions to collide on their own. Even if we were able to energize plasma to a
sufficient level for nuclear fusion in some sort of standard container, say a glass tube, steel
box, or some sort of specialized ceramic, the atoms on the wall in contact with the plasma
would very quickly also be ionized and turned into plasma. We are not even able to do this
however, because the energy lost by the plasma colliding with the walls would prevent us
from ever reaching the necessary excitation. In stars, the plasma is confined through the use
of immense gravitational fields. Unfortunately, this is not an option for us. On the bright
side, the ionized nature of plasma - which is what allows us to channel the energy into it
in the first place - presents us with a solution. The plasma can be confined magnetically,
and in so doing kept from coming in contact with the walls. Although the ionized nature of
plasma also brings with it other properties which make this type of confinement inherently
unstable, we have been able to counter the instabilities to some degree.
There have been many approaches to confining high energy plasmas, all of which are
understood to varying degrees. Most of these confinement schemes, such as the tokamak,
spherical torus, stellerator, spheromak, and others employ toroidal magnetic field geome-
tries of differing complexity. They generally require two main magnetic fields, one that acts
in a toroidal direction, and one that acts in a poloidal direction. Perhaps the most well
understood system is that of the tokamak, as is evidenced by the beginning of a recent
multinational, multi-billion dollar experiment called ITER located in France. This exper-
iment is designed to study and demonstrate the self-sustainability of a large-scale fusion
reactor. While the tokamak is the best understood confinement scheme, it has the drawback
of requiring great external energy input. The potential exists with some other sources for
a plasma which is self-organized to a greater degree, however these schemes tend to be less
well understood [4]. One promising such scheme is that of the field-reversed configuration.
2
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 8/30
FIG. 1: Sketch of magnetic fields in typical field-reversed configuration.
D. The Field-Reversed Configuration
In a field-reversed configuration (FRC), plasma confinement is achieved without the use
of a toroidal magnetic field. This is accomplished by using the magnetic properties of theplasma to freeze an initial magnetic field in a portion of the plasma, while reversing the
magnetic field surrounding it. This has the effect of creating areas of closed magnetic field
loops, as shown in Figure 1. This method has the advantages of having a large ratio of
particle pressure to magnetic field pressure (high β ) plasmas, providing an efficient way of
energizing the plasma, as well as having more self-organization than in other schemes [4].
Both of these aspects of FRCs cause them to require less initial energy input, as well as
less continuous energy input. As of yet FRCs have only been achieved for short durations,although these durations have been shown to be significantly longer than theory would
predict [4]. Unfortunately, this merely demonstrates the lack of understanding that exists
surrounding them, and more basic research must be conducted. This is the goal of the
University of California, Irvine FRC experiment (IFRC).
E. The University of California, Irvine FRC Experiment
The aim of IFRC is not to research nuclear fusion directly, but instead to investigate FRC
creation and confinement. As a result, it does not implement the standard method for FRC
creation, that of the Field Reversed Theta Pinch (FRTP). Instead, it uses a method known
as Coaxial Slow Source (CSS) formation, as described in Section II A. This method was
originally proposed by Phillips [5, 6], and first implemented at the University of Washington
[7]. This method is beneficial because conventional FRC techniques may be difficult to
implement in relatively large sizes [7].
3
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 9/30
F. Plasma Diagnostics
When conducting magnetic fusion research, one of the challenges that must be met is
that of gathering information about the plasma. Many methods have been developed to this
end, each with their own advantages and disadvantages. Perhaps the most common class
of methods is that of material probes, where physical objects are placed inside the plasma.
These objects can then be used to monitor the plasma itself, as is done with Langmuir probes,
or to monitor characteristics inside the plasma, such as measuring interior magnetic fields
with magnetic pickup coils. These methods have the advantages of being simple and well
understood, but have the disadvantages of being invasive and sometimes non-implementable
[8, 9].
Another class of diagnostic methods is active spectroscopy, where the plasma is probed by
exciting it in a specific way and observing the effects. Some examples of this are heavy-ion
beam probes and laser-induced flourescence, which inject heavy ions or energetic photons
into plasma. Similar to these methods are those such as microwave interferometry and
microwave reflectometry, where microwaves are launched into the plasma and their phase
shift and reflectance, respectively, are observed. These methods have the benefits of being
less invasive than material probes, as well as being implementable in some cases where
material probes are not. However, they are often very complicated and/ or expensive to
perform. Finally, there is a class of plasma diagnostics known as passive spectroscopy,
which will be the subject of this paper.
G. Passive Spectroscopy, Line-Emission Spectroscopy
Passive spectroscopy consists essentially of observing plasma externally without interfer-
ing with it. Some of the information that can be gathered regarding a plasma using thismethod are electron temperature, ion temperature, and electrical field information [10]. This
type of diagnostic has the advantage of being completely non-invasive, although it often sac-
rifices some degree of precision and resolution. Of particular interest in IFRC, as in many
experiments, is the measurement of electron temperature.
Although electron temperature can be measured in many ways, the approach that we
chose to adopt was that of line-emission spectroscopy. In line-emission spectroscopy, the
4
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 10/30
intensities of different spectral lines emitted from the plasma are compared, and these ratios
are used to calculate the temperature. Like other passive spectrographic techniques, good
temporal resolution can be achieved with this method, giving information on the thermal
evolution of the plasma throughout the experiment. Unfortunately, also like other passive
spectrographic techniques, it provides poor spatial resolution. We chose this method in
particular because other measurements, such as Stark broadening and Thompson scattering,
can not be employed for low energy plasma [8, 10, 11].
II. BACKGROUND
A. FRC Theory
As mentioned in section I D, the theory behind FRCs is still not complete [4]. Although
there has been some degree of success in forming FRCs, their high sensitivity to formation
and confinement parameters has caused experiments to remain largely based around trial
and error. Although some more recent improvements have aided in creating more control
and better behaved plasmas, the basic steps in FRC formation have remained much the
same.
The FRTP method for creating FRCs involves five basic stages, as is illustrated in Fig-ure 2. First, an initial ”bias” magnetic field is applied to contained gas, which is then ionized
to create plasma. The bias field induces an overarching flux which will be frozen into the
plasma throughout the lifespan of the FRC. The next stage is known as field reversal, where
the direction of external magnetic field is quickly reversed. During field reversal, as the
magnetic field initially decreases, the plasma expands toward the walls of the container.
Then, as the magnetic field increases in magnitude in the opposite direction, the flux of the
plasma induced by the bias field causes the plasma to be pushed toward the center of thecontainer and to lift off the wall. In the third stage, which begins directly after lift-off, the
plasma continues to implode radially. This causes the plasma to begin to heat. Oppositely
directed magnetic field lines near the ends of the plasma then pinch together, tear, and
reconnect. This reconnection begins the next stage, in which the reconnection causes an
axial contraction of the now self-contained plasma configuration, causing the plasma to heat
further. This continues until the final stage, in which equilibrium between all of the fields
5
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 11/30
FIG. 2: Typical steps in FRC Formation.
is reached and the FRC is completely formed [12].
In CSS formation, the magnetic fields required are generally induced by pulsing large cur-
rents through many separate toroidal coils surrounding the plasma. This has the advantages
of requiring smaller capacitor banks, which makes them more cost effective. It also allows
more control over the formation of the FRC, as well as allowing slow FRC formation. This
helps to study FRCs with better temporal resolution. In IFRC, the coils that are used are a
bias coil, a mirror coil, and a flux coil. The bias coil applies the initial magnetic field for the
FRC, which lasts throughout the experiment. The mirror coil applies a magnetic field on
the axial boundaries of the experiment which replace the field line tearing and reconnection
in a FRTP by containing the plasma using a magnetic mirror. This field also lasts through-
out the experiment. Unlike in an FRTP, both the bias coil and the mirror coil are fired in
vacuum. After these fields are reasonably established, pre-ionized plasma is injected axially
into the system. Once the plasma has had time to acquire the bias flux, the third and final
coil, the flux coil, is fired. The flux coil is fired at a higher power than the other two coils,
which results in a dominating magnetic field. It is fired in the opposite direction of the bias
coil so as to provide the field reversal necessary for FRC formation. The approximate timing
for the firing of these coils can be seen in Figure 3.
6
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 12/30
FIG. 3: Plot of FRC driving coil currents vs time. (Shot # 00240: bias coil voltage 350 V, delay
0µs; mirror coil voltage 1250 V, delay 0µs; flux coil voltage 4500 V, delay 25µs; plasma gun voltage
9700 V, delay 0µs; chamber pressure 1.1E-6 Torr.
B. Spectroscopy Theory
The plasma in IFRC is relatively cool compared to high-energy plasma experiments, which
has the consequence that it is not in fact completely ionized. Although very hot plasmas
will be necessary to achieve nuclear fusion, it is the cool nature of IFRC and many other
plasma experiments that allows the use of spectroscopy as a diagnostic tool. When partially
ionized atoms in the plasma are heated, the electrons in those atoms are excited to higher
energy levels. When these electrons eventually decay back into lower energy levels, they
emit photons with an energy corresponding to the difference in those levels. The frequency
of the photon can then be measured and the initial energy of the excited electron calculated
by the equation ∆E = hν , where ∆E is the change in energy levels of the electron, h is
Plank’s constant, and ν is the frequency of the photon.
The different emitted photon frequencies are observed as spectral lines, and are the basis
for spectroscopy. Knowing the possible excitation levels for a particular atom, the spectral
line intensities can be observed and compared, giving a measurement of the abundances
of each electron excitation level. This intensity of the spectral line emitted by an electron
decaying from energy level k to i depends linearly on the number electrons excited to that
level in the line of observation, as given by
I ki =ωki
4πAki
10
N kdx, (1)
7
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 13/30
where I ki is the spectral line intensity, is Plank’s constant, ωki is the frequency of the
spectral line, Aki is the transition probability for spontaneous emission per unit time, N k is
the population density of electrons in state k, and the integral is along the viewing sight-line
[8, 9].
C. Electron Temperature Determination
To calculate electron temperature from the intensities, we must make an assumption
about the distribution of energy in the system. We assume that the plasma is in partial
local thermodynamic equilibrium (PLTE), by which it is meant that the Saha and Boltzmann
thermal equilibrium relations provide good approximations. This assumption is possible
when free electron distributions are close to Maxwellian or Fermi distributions, and electron
densities are high enough that radiative rates are at least one order of magnitude smaller
than collisional rates [8, 13]. Applying PLTE to electron excitation populations, the ratio
in which the electrons decay to the same lower energy level is described by the equation
N kN i
=gkgi
exp
−E k −E ikT e
, (2)
where gk is the statistical weight – the total number of associated states – of the upper
energy level, gi is the statistical weight of the lower energy level, E k is the energy of the
upper level, E i is the energy of the lower level, k is Boltzmann’s constant, T e is the average
thermal energy of an electron in Kelvin, and so kT e is the plasma electron temperature
(average thermal energy) in electron volts.
If two different spectral lines are similarly observed, the ratio of the line intensities can
then be described by
I k1i1I k2i2
= Ak1i1gk1λk2i2Ak2i2gk2λk1i1
exp−E k1 −E k2
kT e
, (3)
[8, 9, 13, 14].
Now the electron temperature can be calculated by
8
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 14/30
kT e =− (E k1 −E k2)
ln
I k1i1I k2i2
Ak2i2gk2λk1i1Ak1i1gk1λk2i2
(4)
=−
(E k1−E k2)
ln
I k1i1I k2i2
/α
, (5)
where α is the line ratio coefficient:
α =Ak1i1gk1λk2i2Ak2i2gk2λk1i1
. (6)
The use of ratios is crucial because it only requires measurements of relative line intensi-
ties, as opposed to absolute line intensities. This has the result that if the two wavelengths
are observed in a similar fashion, factors such as population density and observed solid angle
can be ignored. This is beneficial because these factors may be unknown and/ or difficult
to measure.
Differentiating equation (3), we get
∆T eT e
=kT e
E k1 −E k2
∆(I k1i1/I k2i2)
(I k1i1/I k2i2). (7)
From this is is seen that if kT eE k1−E k2
becomes much larger than 1, errors in relative line inten-
sity measurements are amplified, which could result in large inaccuracies in the calculated
temperature [8, 14]. This is an effect that must be considered before deciding to implement
this method of plasma diagnostic for a particular experiment.
III. SPECTROMETER DESIGN AND CONSTRUCTION
In IFRC, two spectral line ratios were measured. The first ratio was of lines from neutral
oxygen impurities in the system. The lines compared were one centered at a wavelength
of 777nm and one at 615nm. The second ratio was that of hydrogen Hα and Hβ lines,
which are centered at 656nm and 486nm. The relevant information for calculating electron
temperature from these lines is shown in Appendix A [CITE]. These particular lines were
chosen because they have high relative intensity, and are isolated from other spectral lines
that could be emitted from the plasma [8].
9
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 15/30
FIG. 4: A schematic of the experimental set up for measuring electron temperature in IFRC using
spectrometer.
A. Experimental Arrangement and Light Collection
A schematic representation of the experimental set up for measuring electron temperature
in IFRC is shown in Figure 4.
The light emitted from the plasma was collected radially from the FRC through two
adjacent parallel aluminum pipes of diameter 0.25in and length 4in. These pipes were
placed directly against the viewing window of the FRC chamber. The pipe diameter was
chosen out of availability, and the length was chosen to balance the cross sectional viewed
area with the solid angle. The cross sectional viewed area needed to be small to eliminate
inaccuracies that could be caused by varying plasma density radially in the FRC. For this
reason also, the two pipes were aligned side-by-side axially along the FRC. The solid angle,
on the other hand, had to be as large as possible to allow enough light to be easily observed.
The collected light was then filtered through two 10nm full-width half-maximum narrow-
bandpass filters (TFI Technologies, Inc). The filters used were TFI 780-10 (780nm) and
610-10 (610nm) for the oxygen ratios, and TFI 660-10 (660nm) and 490-10 (490 nm) for the
hydrogen lines. The spectral characteristics of the specific filters is shown in Appendix D.
The collection pipes were placed inside an aluminum filter holder fabricated specially for the
spectrometer, and the filtered light was then directed through the same holder to a pair of
photodiodes.
B. Signal Generation and Amplification
The light through each filter was collected by a silicon photodiode. The particular photo-
diodes used (Hamamatsu Corp., S1227-33BR) were chosen for their high spectral response
10
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 16/30
FIG. 5: A schematic of the circuit used to provide a regulated reverse bias voltage to the photodi-
odes.
in the relevant range, as well as their fast rise time. Complete characteristics of these pho-
todiodes is shown in Appendix C. High spectral response was necessary to increase the
signal-to-noise ratio (SNR). The fast rise time was important to allow good temporal resolu-
tion throughout the FRC discharge, which in IFRC lasts between 50 and 100 microseconds.
It was estimated that an amplifier voltage gain near 100 was necessary to provide good
signal. A desired response frequency of 1MHz then necessitated the use of an amplifier with
a gain bandwidth product of approximately 100MHz. The operational amplifier, or op-amp,
chosen for its low noise and fast response time used was an LM6172, produced by National
Semiconductor. The LM6172 is a dual voltage feedback amplifier, the characteristics for
which are shown in Appendix E.
Each photodiode was then connected to an amplifying circuit, which was designed with
the desired characteristics (high amplification, low noise, and fast response time). The
photodiodes were reversed biased to allow operation in photo-conductive mode, which has
the advantage of faster and more linear response. The reverse biasing was accomplished
through the use of an Analog Devices ADP667 voltage regulator. The voltage regulator
was powered by one of two 9V batteries in the spectrometer, and the regulating circuit wasconstructed to provide a constant reverse bias voltage of approximately 2mV. This voltage
was chosen to balance the high linearity, high speed advantages of high reverse bias with
the low noise, high signal advantages of low reverse bias. Information regarding the voltage
regulating circuit is shown in Figure 5 and Table I.
The schematic for the amplifying circuit of one photodiode is shown in Figure 6 (note that
although the majority of the circuit components for each amplifying circuit were separate,
11
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 17/30
TABLE I: Voltage Regulating Circuit Characteristics
Set Resistor Output Resistor Output Capacitor Measured Output Voltage
Rset[Ω] Rout[Ω] C out[µF ] V out
473000 277400 18 1.925
the same op-amp could be used for both as it was dual channel). The photo-current from
each photodiode was passed through a load resistor of approximately 1000 ohms, which was
terminated to ground at a minimum distance. This provided current to voltage conversion,
and the voltage was passed directly into the non-inverting input of the op-amp, which was
powered by supply voltages of ±9V by the batteries. The inverting input of the op-amp
was also terminated to ground through approximately 1000 ohms at a minimum distance.
The output of the op-amp was then terminated to ground at 50 ohms and measured by
the oscilloscope, as well as being fed back into the inverting input through a resistance of
approximately 105 ohms. This circuit provided a theoretical V/A gain of approximately 105,
as is shown by
V o
I photo= R
l1 +
Rf
Rt (8)
= 1000
1 +
105
1000
≈ 105 (9)
where V o is the final output voltage, I photo is the photo-current from the photodiode, Rl is
the load resistor from the non-inverting input to ground, Rf is the feedback resistor to the
inverting input of the op-amp, and Rt is the terminating resistor from the inverting input
to ground. The actual gains were measured by driving the circuits with a known voltage
and recording the output voltage. The measured component characteristics, as well as themeasured gains are shown in Table II.
C. Spectrometer Calibration
To accurately calculate electron temperature, it is crucial that the relative line intensities
are accurately recorded. To insure this, all aspects of collection for each line intensity
measurement have to be calibrated against each other and taken into account. The different
12
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 18/30
FIG. 6: A schematic of the circuit used to amplify the photocurrent output from photodiodes.
TABLE II: Measured Spectrometer Circuit Characteristics
Circuit A Circuit B
Rl [Ω] 1041 1062
Rf [Ω] 100600 100600
Rt [Ω] 1043 1105
Rb [Ω] 9880 9860
Gainmeasured [V/A] 91439.18919 88978.37838
C 1 [µ F] 18
C 2 [nF] 0.3
C 3 [µ F] 18
C 4 [nF] 0.3
factors that had to be incorporated into the calibration were the photodiode efficiencies, the
different levels of circuit amplification, spectral response of the filters, and any geometric
differences resulting from gathering light from two separate pipes.
A schematic diagram of the experimental setup used to gather most of the calibration
information is shown in figure 7. In this setup the spectrometer was placed 10cm from
a tungsten bulb, which was used because it was adequate to provide light power over the
spectral regions necessary. The output voltages were then recorded for each photodiode with
no light reaching the diodes, unfiltered light reaching the diodes, and light passed through
each filter reaching the diodes. These values were compared to the light intensity at the
13
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 19/30
FIG. 7: A schematic of the experimental setup used to measure the relative photodiode efficiencies.
central wavelength as measured by a Newport Corp. Model 840 Optical Power Meter at 10cm
in the same scenarios. This setup was used because it provided a single calibration coefficient
(c1) that incorporated the photodiode efficiencies, circuit amplification, filter transmissions,
which was beneficial because the calculated electron temperature could depend sensitively
on all of these calibration factors. Although this method did not yield an accurate absolute
efficiency, it was sufficient because only a relative measurement was necessary. The results of
the calibration tests for both the oxygen and hydrogen filter sets are shown in Appendix B.Next, the geometric differences resulting from gathering light from two separate pipes
were measured (c2). This was found by first firing the FRC with the spectrometer attached
and in the standard orientation. The spectrometer was then inverted, and the FRC was
fired again. The geometric coefficient was then taken to be the quotients of the voltages
measured in these different orientations. A sample of the data from which this coefficient
was calculated is shown in Figure 8.
A final coefficient, c3, was required to account for the multiplicity of spectral lines withinthe transmission interval of each filter. This coefficient was calculated to allow the use of
values from one specific spectral line using a product of the atomic transition probability
and upper-energy-level statistical weight (see Appendix A):
14
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 20/30
FIG. 8: Plot of geometric coefficients (normal orientation signal / inverted orientation signal)
calculated from intensities of Hα and Hβ lines. (Shots # 00242, 00243)
I tot = I 1 + I 2 + · · · + I n (10)
≈ I 1
1 +
g2A2
g1A1
+ · · · +gnAn
g1A1
(11)
I 1 ≈ I totc3 (12)
where
c3 =
1 +
g2A2
g1A1
+ · · · +gnAn
g1A1
−1
(13)
c3 accounts for the multiplicity of spectral lines within the transmission interval of each
filter, and allows the use of spectral line characteristics from specific spectral lines. The
lines chosen were those with the greatest product of transition probability and statistical
weight. The final relative coefficients for both the oxygen and hydrogen ratios are shown in
Table III. The relevant spectral line characteristics are shown in Table IV
D. Spectral Line Ratio Measurements
Using this information, the electron temperature can be calculated by multiplying equa-
tion (4) by the calibration coefficients:
15
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 21/30
TABLE III: Relative Oxygen and Hydrogen Line Coefficients
Spectral Line Calibration Coefficient Geometric Coefficient Multiplicity Coefficient
Oxygen
780nm 1.118 1 0.467
615nm 1 1.414 0.525
Hydrogen
660nm 1.720 1 0.812
490nm 1 1.414 0.900
TABLE IV: Relevant Oxygen and Hydrogen Spectral Line Characterisitcs
Spectral Line λ [A] Aki[s−1] gk E k [eV]
Oxygen
780nm 777.194 36900000 7 10.741
615nm 615.818 7620000 9 12.754
Hydrogen
660nm 656.467 64650000 6 12.087
490nm 486.269 20620000 6 12.749
kT e =− (E k1 −E k2)
ln
I k1i1I k2i2
C α
, (14)
where C is the ratio of calibration coefficients:
C =c11c12c13
c21c22c23
(15)
E. Noise Reduction
To obtain usable data from the spectrometer, its SNR must be as high as possible. The
first aspect to be considered is photodiode noise. There are two main types of noise inherent
in the operation of a photodiode: thermal noise (or Johnson noise) and shot noise. The
thermal noise is associated with the shunt resistance of the photodiode, and is the same as
16
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 22/30
that present in all resistors. The shot noise is more specific to photodiodes, and is the result
of statistical fluctuations in photocurrent and darkcurrent. These are reduced by limiting
the operating temperature and reverse voltage bias on the photodiode, while maintaining
the maximum possible shunt resistance [15].
The other main types of noise inherent in the spectrometer result from the signal amplifi-
cation circuit. First, there is thermal noise associated with the op-amp and circuit resistors.
This is again reduced by limiting the operating temperature. Next, noise in op-amp supply
voltage can affect the degree of amplification of the signal. This noise is first reduced by
the use of batteries to provide DC power. This noise is then also rejected to some degree by
the op-amp itself. It can be further reduced by the use of voltage regulators on the input
voltage, and capacitors grounded at a minimum distance from the op-amp supply pins. All
of these methods are implemented in the IFRC spectrometer. Finally, external electromag-
netic (EM) noise can be coupled into the amplifying circuit. When coupled to the input or
feedback loops of the op-amp, this noise becomes amplified to the same degree as the signal.
For this reason, it is of the utmost importance to reduce this last noise as much as possible.
The first step in reducing external EM coupling is to reduce the size of all closed loops
inside the circuit as much as possible. This is accomplished by running any necessary wires
in twisted pairs, using the smallest components available, and connecting and terminating all
components at a minimum distance. This was also done with the IDE spectrometer. Next,
the entire spectrometer was enclosed in an aluminum box to provide a degree of shielding
from external EM noise. Upon firing IFRC however, large amounts of coupled noise were still
seen as ringing in the signal. An example of this noise is shown in Figure 9, in which all FRC
driving coils were fired, but no plasma was injected into the system. This was found to be
mainly the result of EM noise coupling to the output of the spectrometer from the crowbar
circuits used to limit the output voltage of the driving coils. This was reduced by moving the
oscilloscope to the maximum achievable distance from the crowbars, approximately 20 feet.
Further EM noise from the driving coils was reduced by using CAT-5 cable, which consists
wholly of twisted pairs, as the output cable from the spectrometer. Finally, large aluminum
plates were placed axially around the spectrometer for further shielding. Implementing these
measures reduced the noise considerably, as can be seen in Figure 10
17
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 23/30
FIG. 9: Plots of measured voltage vs time from spectrometer with no plasma, showing coupled
electromagnetic noise. (Shot # 00198)
FIG. 10: Plots of measured voltage vs time from spectrometer after noise reduction methods
implemented, showing final coupled electromagnetic noise. (Shot # 00216)
IV. RESULTS AND DISCUSSION
Voltages from both photodiode circuits were transmitted along the CAT-5 output cable
to an oscilloscope, which was set to run in single-shot mode to avoid accidental triggering.
The data recorded by the oscilloscope was then sent to a computer which compiled it along
with other data taken according to a computer program written by Alan Van Drie. The data
was recorded as a voltage, and then analyzed automatically at each time point according to
equation (14). The data was also analyzed manually using Microsoft Excel and Matlab. A
18
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 24/30
sample calculation using Hydrogen ratios for one specific time point is shown below (Shot
# 00245, 70 µsec):
C = c11c12c13c21c22c23
=1.720 ∗ 1 ∗ 1.232
1 ∗ 1.414 ∗ 1.111(16)
= 1.097 (17)
α =Ak1i1gk1λk2i2Ak2i2gk2λk1i1
= 64650000 ∗ 6 ∗ 486.26920620000 ∗ 6 ∗ 656.467
(18)
= 2.322 (19)
kT e =− (E k1 −E k2)
ln
I k1i1I k2i2
C α
=− (12.087 − 12.749)
ln−0.504−0.034
1.097
2.322 (20)
= 0.340 (21)
Unfortunately, the SNR of emitted oxygen lines in IFRC was initially too low to provide
provide good data regarding the evolution of the plasma electron temperature, and the
610nm filter became damaged before further testing could be done to remedy the problem.
The low SNR is likely the result of the relatively low temperature of the plasma, as well as
a large reduction of the line intensity as a result of the filtering. It is also possible that the
level of oxygen impurities in the plasma was less than expected, which could again lead to
low line intensities.
Hydrogen atoms and ions made up the bulk of the plasma, and so contributed the bulk of
the total emitted light. As a result, the SNR of the emitted hydrogen lines was much higher
than that of the oxygen lines, giving better data. There is a possibility that the observed
intensities in either of the hydrogen wavelengths could have been artificially inflated, and
hence the temperature made inaccurate, by the presence of carbon line in the same spectral
19
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 25/30
range. Estimates indicate however, that these carbon lines are lower in intensity by at least
an order of magnitude.
Figure 11 shows the results of shot # 00246. Plasma was injected into the system at
approximately10 µsec, at which point it was already slightly energized, and becomes further
energized due to the bias and mirror fields. Emitted light reached the photodiodes almost
immediately, however the SNR is too low to determine the electron temperature. Upon the
firing of the flux coil at approximately 35 µsec, the overarching magnetic field begins to
reverse. FRC was achieved at approximately 37.5 µsec and lasted until approximately130
µsec. After a slight delay, the electron temperature became easily discernible at approxi-
mately 50 µsec. This delay is likely the result of the specific spectral lines being observed:
Hα and Hβ are radiated from neutral Hydrogen atoms, so they must be heated secondarily
through collisional processes with free ions and electrons as opposed to being compressed or
heated directly by field reversal. During FRC, the electron temperature was observed to rise
from approximately 0.28 eV at approximately 50 µsec to a maximum of approximately 0.38
eV at approximately 90 µsec, and then decay to approximately 0.30 eV at approximately
110 µsec before the SNR becomes to great. The peak and subsequent decay are likely the
result of the plasma expanding as rotational or tilt instabilities began to break down the
FRC [6].
V. CONCLUSION
In this paper, we showed that the electron temperature of a low temperature FRC such as
IFRC can be estimated through spectral line emissions as measured by silicon photodiodes.
Although we were unable to use emissions from oxygen impurities in the system in my limited
time, we demonstrated the use of the Hα and Hβ line ratio to calculated this temperature.
Examples of the data that can be acquired were shown and analyzed. A passive spectrometer
constructed from silicon photodiodes along with amplifying circuits and narrow bandpass
filters is a feasible, low-cost diagnostic tool, and we presented one possible electrical and
mechanical design as well as the theoretical background. The main obstacle in constructing
and implementing this spectrometer was increasing the SNR in order to obtain useful data.
With a properly designed and constructed spectrometer as described above, the temporal
evolution of this temperature can be easily observed.
20
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 26/30
(a)
(b)
(c)
FIG. 11: Plots of (a) magnetic fields at radial distance from center R = 14, 44cm and axial distance
from chamber end Z = 10.7cm, (b) Hα and Hβ line intensity, and (c) corresponding calculated
plasma electron temperature vs time. (Shot # 00246: bias coil voltage 400 V, delay 0µs; mirror
coil voltage 1250 V, delay 0µs; flux coil voltage 4900 V, delay 25µs; plasma gun voltage 9500 V,
delay 0µs; chamber pressure 1.2E-6 Torr.)
21
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 27/30
TABLE V: Relevant Oxygen Spectral Lines
Wavelength Transition Probability Statistical Weights Relative Weight
(Lower, Upper)
λ [A] Aki[s−1] gi, gk
6046.23 1050000 3, 3 0.024091778
6046.44 1750000 5, 3 0.040152964
6046.49 350000 1, 3 0.008030593
6155.98 5720000 3, 3 0.13124283
6156.77 5080000 5, 7 0.271969407
6158.18 7620000 7, 9 0.524512428
7771.94 36900000 5, 7 0.466666667
7774.17 36900000 5, 5 0.333333333
7775.39 36900000 5, 3 0.2
TABLE VI: Oxygen Line Energies
Wavelength Lower Energy Level Upper Energy Level
λ [A] E i [eV] E k [eV]6046.23 10.988792 13.0388262
6046.44 10.988861 13.0388262
6046.49 10.98888 13.0388262
6155.98 10.740224 12.753715
6156.77 10.740475 12.753715
6158.18 10.740931 12.753715
7771.94 9.1460906 10.740931
7774.17 9.1460906 10.740475
7775.39 9.1460906 10.740224
22
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 28/30
TABLE VII: Relevant Hydrogen Spectral Lines
Wavelength Transition Probability Statistical Weights Relative Weight
(Lower, Upper)
λ [A] Aki[s−1] gi, gk
4862.69 20620000 6, 4 0.899991271
4862.69 3437000 4, 4 0.100008729
6564.5 22470000 4, 2 0.188120055
6564.6 64650000 6, 4 0.811879945
TABLE VIII: Hydrogen Line Energies
Wavelength Lower Energy Level Upper Energy Level
λ [A] E i [eV] E k [eV]
4862.69 10.1988511 12.7485394
4862.69 10.1988511 12.7485375
6564.5 10.1988101 12.08705066
6564.6 10.1988511 12.0870511
APPENDIX A: RELEVANT SPECTRAL LINE INFORMATION
APPENDIX B: PHOTODIODE CALIBRATION
APPENDIX C: PHOTODIODE CHARACTERISTICS
APPENDIX D: FILTER CHARACTERISTICS
APPENDIX E: OPERATIONAL AMPLIFIER CHARACTERISTICS
23
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 29/30
APPENDIX F: SYMBOLS
c1 Relative spectrometer calibration coefficient 14
c2 Relative spectrometer geometric coefficient 14
c3 Spectral line multiplicity coefficient 14
CSS Coaxial Slow Source 3
E i Energy of an electron in energy level i 8
EM Electromagnetic 17
FRC Field-Reversed Configuration 3
FRTP Field Reversed Theta Pinch 3
gi Statistical weight of energy level i 8
h Planck’s Constant = 6.6262 × 10−
34 J s 7 Planck’s Constant = 1.0546 × 10−34 J s 8
i, k Specific electron energy levels 7
I ik Spectral line intensity emitted by an electron decaying from level i to k 8
I photo Photo-current from photodiode 12
IFRC University of California, Irvine FRC experiment 3
kT e Electron temperature (in electron volts) 8
N i Population density of electrons in state i 8Op-amp Operational Amplifier 11
Rf Feedback resistor 12
Rl Load resistor 12
Rt Terminating resistor 12
SNR Signal-To-Noise Ration 11
T e Electron temperature (in Kelvin) 8
V o Output voltage from Operational Amplifier 12α Line Ratio Coefficient 9
β Ratio of plasma particle pressure to magnetic field pressure 3
∆E Change in energy 7
ν Frequency 7
ω the frequency of an emitted spectral line 8
[1] R. Goldston and P. Rutherford, Introduction to plasma physics (Institute of Physics Publish-
ing, 1995). 24
8/7/2019 Passive Spectrometer Design
http://slidepdf.com/reader/full/passive-spectrometer-design 30/30
[2] R. Fitzpatrick, Introduction to plasma physics (1998), course PHY380L, University of Texas
at Austin.
[3] The surprising benefits of creating a star (2001), for U.S. Department of Energy, Offices of
Fusion Energy Sciences.[4] M. Brown, Alternate concepts, Lecture (2005), for National Undergraduate Fellowship Pro-
gram in Plasma Physics and Fusion Engineering.
[5] J. Phillips, Tech. Rep. LA-8711-P, Los Alamos National Laboratory (1981).
[6] M. Tuszewski, Nuclear Fusion 28, 2033 (1988).
[7] Z. Pietrzyk, G. Vlases, R. Brooks, K. Hahn, and R. Raman, Nuclear Fusion 27, 1478 (1987).
[8] O. Garate, for TriAlpha Energy Corporation, Irvine CA.
[9] R. McWhirter, Spectral intensities (Academic Press, 1965), chap. 5.
[10] J. Foley, Plasma diagnostics, Lecture (2005), for National Undergraduate Fellowship Program
in Plasma Physics and Fusion Engineering.
[11] I. Hutchinson, Principles of plasma diagnostics (Cambridge University Press, 1987).
[12] A. Hoffman, R. Milroy, J. Slough, and L. Steinhauer, Fusion Technology 9, 48 (1986).
[13] H. Griem, Principles of plasma spectroscopy (Cambridge University Press, 1997).
[14] Z. Andreic, Ph.D. thesis, University of Zagreb (1993).
[15] Photodiode Technical Information , Hamamatsu Corporation (????).
25