PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System with High Caco-2 and Human Absorption Predictability PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System with High Caco-2 and Human Absorption Predictability Xiaoxi (Kevin) Chen August 20, 2008 Xiaoxi (Kevin) Chen August 20, 2008

Transcript of PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or...

Page 1: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

PAMPA: Novel BD Gentest™ Pre-coatedPAMPA Plate System with High Caco-2and Human Absorption Predictability

PAMPA: Novel BD Gentest™ Pre-coatedPAMPA Plate System with High Caco-2and Human Absorption Predictability

Xiaoxi (Kevin) ChenAugust 20, 2008

Xiaoxi (Kevin) ChenAugust 20, 2008

Page 2: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

High Throughput Permeability Assay

• Drug candidates are screened for their oral-absorption potential early in the discovery and development phase, as a filter to remove poor performers and identify candidates that need to be modified.

• Most drugs are absorbed through the intestines without using cellular pumps. Therefore, passive permeability assays are useful for screening oral-absorption potential of drug candidates.

• Two permeability assays have become prevalent in recent years:– Cell-based (especially Caco-2) permeability assay– Parallel artificial membrane permeability assay (PAMPA)

Page 3: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Caco-2 vs. PAMPA

Easy, fast preparationRequires cell culture (up to 21 days)

Assay preparation

LowHighCost

Type of Permeability

Type of model

Measures passive permeability in absence of transporters or efflux systems

Measures the sum of passive and active permeability

Artificial membraneCell monolayer

PAMPACaco-2

PAMPA and Caco-2 tests can be complementary tests to determine both passive and active permeability. Caco-2 tests alone measure the sum of passive and active permeability which can not be decoupled without the information obtained from PAMPA tests.

Page 4: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

PAMPA Experimental Setup

Buffer Compound solutionMultiwell filter plate

Multiwell receiver plate

(Incubation for several hours)

Compounds permeate through membrane barrier

Compound concentration in acceptor well: CA(t)

Compound concentration in donor well: CD(t)

Permeability is calculated from CA(t) and CD(t)

Page 5: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Existing PAMPA Methods• Original PAMPA

– The lipid solution consists of 10% lecithin in dodecane.– Reference: Kansy, M. et al. (1998) J. Med. Chem. 41,1007-1010.

• DOPC-PAMPA– The lipid solution consists of 2% DOPC in dodecane.– Reference: Avdeef, A. et al. (2001) Eur. J. Pharm. Sci. 14, 271-280.

• HDM-PAMPA– The lipid solution is 100% hexadecane.– Wohnsland, F. and Faller, B. (2001) J. Med. Chem. 44, 923-930.

• Bio-mimetic PAMPA (BM-PAMPA)– The lipid solution consists of a mixture of PC, PE, PS, PI and cholesterol in an organic solvent– Reference: Sugano, K. et al. (2001) Int. J. Pharm. 228,181-188.

• Double-Sink™ PAMPA (DS-PAMPA), available from pION Inc– The lipid solution consists of 20% dodecane solution of a phospholipid mixture– The acceptor solution contains a surfactant mixture– Reference: Avdeef, A. et al. (2005) Eur. J. Pharm. Sci. 24:333-349 (2005).

• BD Gentest™ Pre-coated PAMPA Plate, available from BD Biosciences– A lipid-oil-lipid tri-layer structure is constructed in the pores of the porous filter. – Solvent contents are reduced by using volatile solvents in membrane preparation. – Improves correlation with human absorption and Caco-2, and reduces mass retention– Reference: Chen, X. et al., (2008), Pharm. Res., In Press

Page 6: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Traditional PAMPA Membrane

• A porous filter is soaked with a solution of lipids• The lipids are likely dispersed in the solvent• The PAMPA membrane is dominated by solvent, instead of lipids• The excess amount of solvent may become an extra barrier for the

compounds to permeate through

Lipids solution:Red: lipidsGreen: solvent

Filter plate

Pores

Page 7: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Issues of the Traditional PAMPA Membrane that We Would Like to Address

1. Incorrect prediction of a group of commercial compounds that are classified by the biopharmaceutical classification system (BCS) as high permeability compounds.

• These compounds have high human absorption values and high Caco-2 permeability values, but are predicted by traditional PAMPA as “low permeability”.

2. Some “sticky” compounds are trapped in the artificial membrane, resulting in high mass retention values.

3. The artificial membrane needs to be prepared immediately before the assay, due to the poor stabilityof the artificial membrane.

Page 8: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Traditional PAMPA Issue 1A: Correlation with Human Absorption

0

20

40

60

80

100

-8 -7 -6 -5 -4

Under-predicted Compounds:phenytointimololpindololmetoprololtheophylinenaproxenacetaminophenacetylsalicylic acidantipyrineindomethacinketoprofenHigh PermeabilityLow Permeability

Traditional PAMPA PermeabilityLog Pe (cm/s)

Traditional PAMPA vs. Human Absorption Data

Hum

an A

bsor

ptio

n (%

FA)

• Both traditional PAMPA (DOPC-PAMPA) permeability data and human absorption data are cited from:Ruell, J.A.; Avdeef, A.; Du, C.; Tsinman, K. A Simple PAMPA Filter for Passively Absorbed Compounds, Poster presented at 2002 ACS National Meeting, Boston, MA, USA. (available at http://www.pion-inc.com).

Page 9: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

0

50

100

150

200

250

300

350

0 100 200 300 400 500 600 700

Traditional PAMPA Issue 1B: Correlation with Caco-2 Data

Traditional PAMPA vs. Caco-2

Cac

o-2

Perm

eabi

lity

(nm

/s)

Traditional PAMPA Permeability(nm/s)

Under-predicted Compounds:naproxenantipyrinecaffeineketoprofen

• Both traditional PAMPA (Double-Sink™ PAMPA) permeability data and Caco-2 permeability data are cited from:Balimane, P.V., Han, Y-H., Chong, S. Current Industrial Practices of Assessing Permeability and P-Glycoprotein Interaction. AAPS Journal 8(1):E1-E13 (2006); and other publications from the same laboratory.

Page 10: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Traditional PAMPA Issue 2: Mass Retention by the Membrane

• Mass retention: the percentage of the total mass of the compound lost during the permeability measurement as a result of binding to the plastic surface and being retained in the filter membrane

• Mass retention of some “sticky” compounds by the traditional PAMPA membrane (data cited from: Avdeef, A., Strafford, M., Block, E, Balogh, M., Chambliss, W., and Khan, I. Eur. J. Pharm. Sci. 14:271 (2001)).– Amitriptyline: 53%– Ketoconazole: 64%– Phenazopyridine: 69%

• “Sticky” compounds are very common among pharmaceutical research compounds

Page 11: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Traditional PAMPA Issue 3: Stability of the Membrane

Because of this instability, traditional PAMPA membrane needs to be prepared by the user immediately before the assay.

Perm

eabi

lity

(nm

/s)

Traditional PAMPA Membrane: Fresh vs. 1-day old

020406080

100

Propran

olol

Verapa

milCarb

amaz

epine

Atenolo

lRan

itidine

Hydroc

hlorot

hiazid

e

Results obtainedwith freshmembrane

Results obtainedwith 1-day oldmembrane

Page 12: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Design of the New PAMPA Membrane

• Ideally, the PAMPA artificial membrane should be a true lipid bilayer. However, a true lipid bilayer is too fragile for use in high throughput assays.

• To create a model closer to the true lipid bilayer and robust enough for high throughput screening, a lipid-oil-lipid tri-layer structure is constructed in the pores of the porous filter.

Page 13: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

New PAMPA Membrane

• A lipid-oil-lipid tri-layer structure is constructed in the pores of the porous filterusing three consecutive coating steps

• The oil layer mimics the hydrophobic interior of the biological membrane; and the amphiphilic lipids anchoring on the oil / water interface mimic the exterior of the biological membrane.

• The oil layer is crucial for maintaining a robust and stable PAMPA membrane• The oil layer is ultra thin to minimize compound retention and interference with

the compound permeation

Filter plate

PoresRed: lipid layer

Green: oil layer

Page 14: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Comparison of traditionaland new PAMPA membrane

• Excessive solvents• No structured layers• Long permeation pathway

Traditional PAMPA membrane construction

New PAMPA membrane construction

Porous filter

• No excessive solvents• Structured layers• Short permeation pathway

(closer to biological membrane)

Page 15: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Traditional PAMPA membrane construction

New PAMPA membrane construction

Porous filter

Excessive solvents

No excessive solvents

Comparison of traditionaland new PAMPA membrane

Page 16: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

New PAMPA Membrane Improves Correlation with Human Absorption

Traditional PAMPA

Correlation plots using the same group of compounds

0

20

40

60

80

100

-8 -7 -6 -5 -4

False negativeFalse positiveHigh permeability

Low permeability

Hum

an A

bsor

ptio

n (%

FA)

PAMPA PermeabilityLog Pe (cm/s)

BD Gentest™ Pre-coated PAMPA

0

20

40

60

80

100

-8 -7 -6 -5 -4

False negativeFalse positiveHigh permeability

Low permeability

Hum

an A

bsor

ptio

n (%

FA)

PAMPA PermeabilityLog Pe (cm/s)

Page 17: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

0

50

100

150

200

250

300

350

0 100 200 300 400 500 600 700

Cac

o-2

Perm

eabi

lity

(nm

/s)

Traditional PAMPA Permeability(nm/s)

Under-predicted Compounds

Traditional PAMPA

0

50

100

150

200

250

300

350

0 20 40 60 80 100 120

Cac

o-2

Per

mea

bilit

y (n

m/s

)

New PAMPA Permeability(nm/s)

BD Gentest™ Pre-coated PAMPA

Correlation plots using the same group of compounds

New PAMPA Membrane Improves Correlation with Caco-2 Data

Page 18: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Reproducibility of the New PAMPA Membrane across 2 sites*

73.3

60.3

87.5

43.4

86.4

19.2

95.8

41.3

4.4

1.5

1.1

1.6

2.0

BDPe (nm/sec)

84.2 ± 3.2

67.6 ± 4.9

82.5 ± 6.6

42.7 ± 1.2

80.8 ± 10.2

27.1 ± 4.1

95.8 ± 6.3

46 ± 2.9

1.1 ± 1.9

1.8 ± 0.8

5.2 ± 4.7

0.1

2.4 ± 1.5

BMSPe (nm/sec)

100Naproxen

95Dexamethasone

33Nadolol

97Antipyrine

95Verapamil

95Metoprolol

90Propranolol

100Caffeine

90Ketoprofen

50Etoposide

55Acebutolol

35Norfloxacin

12Sulfasalazine

% Human absorptionCompound

Data Correlation Between 2 Labs

y = 0.9751xR2 = 0.98

0

20

40

60

80

100

120

0 20 40 60 80 100 120

Permeability data obtained at BMS (nm/s)

Perm

eabi

lity

data

obt

aine

d at

BD

(nm

/s)

* Poster presented by Bristol-Myers Squibb (BMS) at 2006 AAPS Annual Meeting, San Antonio, TX, USA.

Page 19: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

New PAMPA Membrane Reduces Mass Retention of “Sticky” Compounds

* Avdeef, A., Strafford, M., Block, E., Balogh, M., Chambliss, W., Khan, I. Eur. J. Pharm. Sci. 14:271 (2001)

Mas

s R

eten

tion

(R) )/(])()([1 0 DAADD VCVtCVtCR ∗∗+∗−=

0%

50%

100%

amitriptyline ketoconazole phenazopyridine

Traditional PAMPAMembrane*

New PAMPA Membrane

Traditional DOPC-PAMPABD Gentest™ Pre-coated PAMPA

Page 20: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Pre-coated PAMPA plate is stableand highly reproducible

0

2

4

6

8

10

12

famoti

dine

amilo

ride

furos

emide

hydro

chlor

othiaz

idean

tipyri

neca

ffeine

metopro

lolna

proxe

n

Plate 1 (1-month old)Plate 2 (6-month old)Plate 3 (12-month old)

Perm

eabi

lity

(nm

/s)

BD Gentest™ Pre-coated PAMPA Plate System is stable when stored at -20 ˚C

Page 21: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

02468

1012141618

famoti

dine

amilo

ride

furos

emide

hydro

chlor

othiaz

idean

tipyri

neca

ffeine

metopro

lolna

proxe

n

PBSPBS+10%MethanolPBS+20%Methanol

BD Gentest™ Pre-coated PAMPA Plate is Compatible with Buffers Containing Organic Solvents

Perm

eabi

lity

(nm

/s) Using buffers

containing organic solvents helps increase the solubility of low solubility compounds.

Page 22: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

BD Gentest™ Pre-coated PAMPA Plate is Compatible with Automation Systems

Page 23: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

BD Gentest™ Pre-coated PAMPA Plate has been Validated using both UV Plate Reader and LC/MS Analysis

20

30

40

50

60

70

80

90

100

-8 -7 -6 -5 -420

30

40

50

60

70

80

90

100

-8 -7 -6 -5 -4

False negativeFalse positiveHigh permeability

Low permeability

Hum

an A

bsor

ptio

n (%

FA)

PAMPA PermeabilityLog Pe (cm/s)

Results obtained using UV Plate Reader

Correlation plots using the same group of compounds

Results obtained using LC/MS

False negativeFalse positiveHigh permeability

Low permeability

Hum

an A

bsor

ptio

n (%

FA)

PAMPA PermeabilityLog Pe (cm/s)

Both methods produced correct predictions for high and low permeability compounds.

Page 24: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

Conclusions

BD Gentest™ Pre-coated PAMPA Plate System• Improves correlation with Caco-2• Improves correlation with human absorption• Reduces mass retention• Is stable and highly reproducible• Is compatible with buffers containing organic

solvents (to help low solubility compounds)• Is compatible with automation systems• Has been validated using both UV Plate Reader

analysis and LC/MS analysis

Page 25: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer

BD Gentest™ Pre-coatedPAMPA Plate System

Page 26: PAMPA: Novel BD Gentest Pre-coated PAMPA Plate System …permeability in absence of transporters or efflux systems Measures the sum of passive and active permeability Cell monolayer