Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS...

8
Brit. Heart J., 1967, 29, 608. Pacemaker 'Heart Sound' ALAN HARRIS From the Cardiac Department, St. George's Hospital, Hyde Park Corner, London S.W.1 Artificial pacing of the heart may alter the order of closure of the mitral and tricuspid valves and pul- monary and aortic valves depending on the site -of the myocardial electrode (Haber and Leatham, 1965). In some patients on artificial pacemakers we have also noticed an extra sound, always earlier, and sometimes louder, than the usual heart sounds, and it has been found to occur with both epicardial and endocardial systems (Nager et al., 1965). This extra sound can be heard, and more often recorded, 6 m.sec. after the pacemaker impulse is recorded on the electrocardiogram. In addition, an early out- ward systolic movement can be recorded in the apex cardiogram coinciding with the extra sound. Nager et al. (1965) concluded that the extra sound was pos- sibly of intracardiac origin related to premature con- traction of the heart muscle underlying the electrode. This implies that heart muscle could behave in an abnormal way in respect to its known electro- mechanical interval which is 21 m.sec. for conducted beats (Tafur, Cohen, and Levine, 1964) and contra- venes the "all or none law" (Bowditch, 1871). In view of the physiological importance of these find- ings, the pacemaker heart sound has been further investigated. SUBJECTS AND METHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel photographic oscilloscope* was used to record the extra sound, together with a second phonocardiogram, external carotid pulse, respira- tion, and electrocardiogram at a paper speed of 100 m.sec./cm. The apex cardiogram was recorded in the left lateral decubitus position, with the breath held at the end of expiration. One patient was also studied by intracardiac phonocardiography and pressure recordings. In order to measure the interval between the pacemaker Received November 30, 1966. * Cambridge Instrument Company of England. 608 impulse and extra sound with greater accuracy, high- speed phonocardiograms were also recorded on a measuring oscilloscope at a speed of 5 m.sec./cm. and 10 m.sec./cm. Five of the patients were being paced with a C.50 No. 5 unipolar electrodet positioned in the apex of the right ventricle: in four of them the pacemaker and positive electrode were buried in the axilla and in the fifth patient the C.50 No. 5 unipolar electrode was exteriorized in the left mammary region and attached to an external pace- maker unit, and the positive electrode was implanted in the subcutaneous tissue overlying the left pectoral muscle. The sixth patient was paced with an electrode placed on the epicardium of the left ventricle at thoraco- tomy and attached to an implanted abdominal pacemaker unit. The units were of the constant voltage type, the no-load voltage being 5 volts, and powered by 3 or 4 mercury batteries giving a square wave impulse of 0-7 to 1 m.sec. at a fixed rate of 75 a minute. RESULTS A typical recording of the extra sound (X) is shown in Fig. 1. The extra sound may be louder than any of the other heart sounds, and on inspira- tion it tends to diminish in intensity and is best heard in the mitral and tricuspid areas. Occasion- ally on auscultation the first sound may be quite soft and almost inaudible, so that the extra sound may be mistaken for the first heart sound, giving the impression of an abnormally long interval between the first and second heart sounds (Fig. 2). The apex cardiogram usually shows an early systolic out- ward movement, the onset of which coincides with the extra sound (Fig. 2), and less commonly is the only recorded movement (Fig. 3). High-speed phonocardiograms were recorded and the interval between the pacemaker stimulus and the onset of the extra sound was found to be 6 m.sec. (Fig. 4). t U.S. Catheter Corporation. on April 27, 2020 by guest. Protected by copyright. http://heart.bmj.com/ Br Heart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Downloaded from

Transcript of Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS...

Page 1: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

Brit. Heart J., 1967, 29, 608.

Pacemaker 'Heart Sound'ALAN HARRIS

From the Cardiac Department, St. George's Hospital, Hyde Park Corner, London S.W.1

Artificial pacing of the heart may alter the order ofclosure of the mitral and tricuspid valves and pul-monary and aortic valves depending on the site-ofthe myocardial electrode (Haber and Leatham,1965). In some patients on artificial pacemakerswe have also noticed an extra sound, always earlier,and sometimes louder, than the usual heart sounds,and it has been found to occur with both epicardialand endocardial systems (Nager et al., 1965). Thisextra sound can be heard, and more often recorded,6 m.sec. after the pacemaker impulse is recorded onthe electrocardiogram. In addition, an early out-ward systolic movement can be recorded in the apexcardiogram coinciding with the extra sound. Nageret al. (1965) concluded that the extra sound was pos-sibly of intracardiac origin related to premature con-traction ofthe heart muscle underlying the electrode.This implies that heart muscle could behave in anabnormal way in respect to its known electro-mechanical interval which is 21 m.sec. for conductedbeats (Tafur, Cohen, and Levine, 1964) and contra-venes the "all or none law" (Bowditch, 1871). Inview of the physiological importance of these find-ings, the pacemaker heart sound has been furtherinvestigated.

SUBJECTS AND METHODSIn 6 patients with chronic heart block on artificial

pacemakers a four-channel photographic oscilloscope*was used to record the extra sound, together with asecond phonocardiogram, external carotid pulse, respira-tion, and electrocardiogram at a paper speed of 100m.sec./cm. The apex cardiogram was recorded in theleft lateral decubitus position, with the breath held at theend of expiration. One patient was also studied byintracardiac phonocardiography and pressure recordings.In order to measure the interval between the pacemaker

Received November 30, 1966.* Cambridge Instrument Company of England.

608

impulse and extra sound with greater accuracy, high-speed phonocardiograms were also recorded on ameasuring oscilloscope at a speed of 5 m.sec./cm. and10 m.sec./cm.Five of the patients were being paced with a C.50 No.

5 unipolar electrodet positioned in the apex of the rightventricle: in four of them the pacemaker and positiveelectrode were buried in the axilla and in the fifth patientthe C.50 No. 5 unipolar electrode was exteriorized in theleft mammary region and attached to an external pace-maker unit, and the positive electrode was implanted inthe subcutaneous tissue overlying the left pectoralmuscle. The sixth patient was paced with an electrodeplaced on the epicardium of the left ventricle at thoraco-tomy and attached to an implanted abdominal pacemakerunit. The units were of the constant voltage type, theno-load voltage being 5 volts, and powered by 3 or 4mercury batteries giving a square wave impulse of 0-7 to1 m.sec. at a fixed rate of 75 a minute.

RESULTSA typical recording of the extra sound (X) is

shown in Fig. 1. The extra sound may be louderthan any of the other heart sounds, and on inspira-tion it tends to diminish in intensity and is bestheard in the mitral and tricuspid areas. Occasion-ally on auscultation the first sound may be quitesoft and almost inaudible, so that the extra soundmay be mistaken for the first heart sound, giving theimpression of an abnormally long interval betweenthe first and second heart sounds (Fig. 2). Theapex cardiogram usually shows an early systolic out-ward movement, the onset of which coincides withthe extra sound (Fig. 2), and less commonly is theonly recorded movement (Fig. 3). High-speedphonocardiograms were recorded and the intervalbetween the pacemaker stimulus and the onset ofthe extra sound was found to be 6 m.sec. (Fig. 4).

t U.S. Catheter Corporation.

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from

Page 2: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

Pacemaker 'Heart Sound'

x I

l l l l l 11 111111111111111112 X I

1111111111111112~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~-

PA

S LSE

iS

N.i~ /

FIG. 1.-Endocardial pacing from the right ventricular cavity with an implanted axillary pacemaker, demon-stratingthe extra sound (X) which decreases in intensity on inspiration. The time intervals in this and subse-quent records, unless otherwise stated, are 0.20 and 0.04 sec. External phonocardiograms: PA, pulmonaryarea; LSE, left sternal edge; 1, first heart sound; 2, second heart sound; Insp., inspiration; II, lead II of

electrocardiogram; S, pacemaker stimulus; HF, high frequency; Car., external carotid pulse.

lil llll lHF

PA

ACG

S

1111111111111111111111111 11111111111111

FIG. 2.-An early outward movement of the apex cardiogram(ACG) coincides with the extra sound (X) which is relativelyloud compared with the first heart sound. LF, low fre-

quency; TA, tricuspid area.FIG. 3.-The outward movement of the apex cardiogram is

the only movement recorded. Exp., expiration.

LF I -,X. 1, I I 12k I I

PALF _TA ITA

II1111111111111111x 1 2

_ '. I,

-

IW-n-

-~~~~~~uo

609

I

n s-0

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from

Page 3: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

6Alan Harris

FIG. 4.-High-speed phonocardiogram recorded on a measuring oscilloscope showing the time interval betweenthe pacemaker stimulus and the onset of the extra sound to be 6 m.sec. Time intervals are 5 and 1 m.sec.

An electro-mechanical interval of 6 m.sec. is ofthe same order as that of skeletal muscle and thepossibility that diaphragmatic contraction was thesource of the extra sound was considered. How-ever, x-ray screening of the patients did not revealany abnormal diaphragmatic contraction, and thiswas also the finding of Nager et al. (1965). It waspossible to reproduce the extra sound and earlyapical movement in one patient by implanting thepositive electrode of the pacing system in the leftpectoral-muscle overlying the apex of the heart.

This patient had visible and palpable contractions ofthe pectoral muscle with every pacemaker stimulus,and recordings were obtained with (Fig. 5A) andwithout (Fig. 5B) pectoral muscle stimulation.Pectoral muscle stimulation resulted in an extrasound and early systolic movement (Fig. 5, 6, and 7),which were similar in timing (8 m.sec.) to the extrasound and apical movement recorded in the other 5patients. Phonocardiograms were recorded overthe implanted pacemaker units and positive elec-trodes in these 5 patients to see if local muscle con-

A B111111 'xi 111111 I'''11111111 111111 111111'1 111 11111 111111LF LF

TA TA

FIG. 5(A).-Unipolar endocardial pacing with a positive electrode implanted in the left pectoral muscle. Theextra sound (X) is produced by pectoral muscle contraction and coincides with an early outward movement of

the apex cardiogram.FIG. 5(B).-Bipolar endocardial pacing abolished the extra sound and early movement of the apex cardiogram.

610

sG s

ACGn /%.....0000"

I I I I HM I-i 1 1 l-1

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from

Page 4: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

Pacemaker 'Heart Sound'

A B

1111 I Iixi III t 1 1 1111111111,111iiiII 1111111LF , _ LF _

_Iw-'~ 1PTA

LF' LA I& ..

MA

Car. p.A

bom oS ooms

TA

S

111111111 111111111111111111111111111111111111111111FIG. 6(A).-Unipolar endocardial pacing with a positive electrode implanted in the left pectoral muscle. The

extra sound (X) is produced by pectoral muscle contraction.FIG. 6(B).-Bipolar endocardial pacing abolished the extra sound. MA, mitral area.

traction was causing the extra sound, but no soundwas present (Fig. 8).

Since the extra sound coincided with an early out-ward systolic movement of the apex cardiogram itwould presumably be associated with an early rise inventricular pressure if arising from the ventricle.During the routine change of an implanted axillarypacemaker the opportunity was taken to obtain intra-cardiac phonocardiograms and pressure recordingsin one patient with an obvious extra sound, and itwas found that the right ventricular pressure trace

did not show any evidence of an early rise in pres-sure (Fig. 9), nor was the extra sound recorded withinthe cavity of the right ventricle. The external caro-tid pulse recording was also normal, excluding thepossibility that the left ventricle was producing anearly rise in pressure. In the refractory phase ofectopic beats the extra sound still followed the pace-maker impulse, despite the absence of a right ven-tricular pressure pulse (Fig. 10). Hence it seemedalmost certain that the extra sound was not cardiacin origin and must be related to pacemaker-induced

FIG. 7.-High-speed phonocardiogram recorded on a measuring oscilloscope, showing the time intervalbetween the pacemaker stimulus and the onset of the sound produced by pectoral muscle contraction to be

8 m.sec. Time intervals 10 and 2 m.sec.

61

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from

Page 5: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

Alan Harris

l 1 | l 1i I I I 1 I 1 21 1 1 1

S LSE rv r.-

rHFUn

Unit

Car. 5 / /

l i l IIII

FIG. 8.-Phonocardiogram recorded over the implanted axillary pacemaker unit and positive electrode, show-ing absence of any sound, though present at the fifth left intercostal space.

contraction of skeletal muscle overlying the myo-cardial electrode. Thus, during the course of thegeneral anesthetic, the patient was given a neuro-muscular blocking agent (suxamethonium chloride),and this was found to abolish the extra sound (Fig.9).

It also seemed likely that the position of the elec-trode tip might be critical, since current was pre-sumably spreading to the adjacent intercostal nerves.In one patient the position of the endocardial elec-trode could be altered without loss of right ventricu-lar stimulation, as indicated by the heart sounds(Fig. 11), and yet the extra sound could be abolished.

DISCUSSIONThe results of this investigation show that the

extra sound produced by electrical pacing of theheart is due to spread of current to the nearby inter-costal nerves producing intercostal muscle contrac-tion. Nager et al. (1965) considered that theabolition of the extra sound, when the pacemakerimpulse fell in the refractory period of a normalconducted beat and therefore failed to stimulate theheart, must indicate that the sound was due to prema-ture heart contraction, but this was not confirmed inthis present study (Fig. 10). One possible explana-tion for this difference in findings is the interpreta-

A

x 2HF

TA t ~ ---I-PCG

RV

RVP

11.I111II11\ 1111111111111 111111

B11111111111I1111111111111111111111

1 2HF i

TAII-PCMG I

100 Mo.Scoline_0g

11111111IAJ'I 1111111111111111 III

FIG. 9.-Intracardiac phonocardiogram (I-PCG) recorded in the right ventricle (RV) simultaneously with theright ventricular pressure (RVP) failed to record the extra sound which was shown on the external phono-cardiogram. The extra sound shown on the external phonocardiogram in (A) is abolished by intravenous

suxamethonium chloride (B).

612

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from

Page 6: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

Pacemaker 'Heart Sound'

HF

TA

L

I[ .1

RVP

' _ ~ ~~~~ES

FIG. 10.-The external phonocardiogram shows the extra sound coinciding with a pacenmaker impulse duringthe second cycle which is due to a ventricular ectopic beat (E). The first cycle is a normal paced beat.

tion of the phonocardiograms which Nager et al.(1965) used to illustrate this point. It is possiblethat the extra sound was coinciding with the firstheart sound in the section showing competition be-tween sinus rhythm and pacemaker rhythm. Inanother patient, Nager et al. (1965) were unable

satisfactorily to explain, on the basis of the prema-ture heart contraction, the continuation of the extrasound when pacing was intermittent. The probableexplanation of their findings in this patient was thatthe threshold for pacing had risen, resulting in inter-mittent pacing of the heart, but the threshold for

A B

11111§ | Ix 1,1 111 2iil l lllHF

PAAP

HF

TAI

HF

PA~~~~~~~~1 w - wTv*PAPA

HF A ,X M T. _

I.1TA

Coar.

Resp.- -M

s

II S II

liI li l ii I 11111 I 1111 11111111111111FIG. 1 1.-External phonocardiograms before (A) and after (B) repositioning the right ventricular endocardialunipolar electrode, showing abolition of the extra sound. In (A), pacing is causing physiological closure of thesecond sound indicating that the left ventricle is being stimulated before the right. In (B), closure of the secondsound is reversed, which is the normal sequence of events in right ventricular endocardial pacing. A, aortic

component; P, pulmonary component.

v-I I Is .1. -

Now" if -V r 11 -,-- I

613

I1 A.

i

Resp.

I I I I I ' l lx I 'X

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from

Page 7: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

Alan Harris

FIG. 12.-Lateral chest x-ray film showing the proximity ofthe endocardial electrode tip to the anterior chest wall.

stimulation of the intercostal nerves adjacent to themyocardial electrode may not have altered and thusthe extra sound continued to be recorded.Nager et al. (1965) considered the possibility that

skeletal muscle stimulation might be. responsible forthe extra sound and that the diaphragm would be itslikely source. Abnormal diaphragmatic contractionwas not observed on x-ray screening, which wasconfirmed in this present study, and they concludedthat skeletal muscle was unlikely to be the source ofthe extra sound. Intercostal muscle contractioncausing the extra sound is in fact too localized to beapparent on x-ray screening. From time to time,however, endocardial pacing does result in diaphrag-matic stimulation (Harris et al., 1965), and patientsare always aware of this, and it is of interest that noneof the patients with extra sound studied were awareof the localized intercostal muscle contraction.The position of the myocardial electrode is pre-

sumably fairly critical for stimulation of the inter-costal nerves, since by repositioning the endocardialelectrode the extra sound could be abolished (Fig.11). Lateral chest x-ray films of two patients, one

,.;a ............-9io_:c ~~~~~~~~~~~~~~~~~~~~~~~~.. ........

FIG. 13.-Lateral chest x-ray film showing the proximityof the epicardial electrode to the anterior chest wall.

with an endocardial electrode and the other with anepicardial electrode, are shown in Fig. 12 and 13and demonstrate the proximity of the electrode tothe chest wall. Chest wall movement duringinspiration away from the electrode could accountfor the decrease in intensity of the extra sound,since under these circumstances spread of currentwould be expected to stimulate fewer intercostalnerve fibres owing to the greater separation of theelectrode from the chest wall. Respiratory varia-tion in the intensity of the extra sound should onlyoccur with low power units, and the more powerfulunits would be expected to produce an extra soundof constant intensity during inspiration. The earlysystolic movement of the apex cardiogram is local-ized over the apex beat of the heart, which is theregion where the endocardial and epicardial elec-trodes are usually positioned. The localizedintercostal muscle contraction gives a false impres-sion of an early outward movement of the heart.The failure to record any premature rise in ven-

tricular pressure or intracardiac sound coincidingwith the extra sound, and the abolition of the extrasound by a neuromuscular blocking agent, formsconclusive evidence of an extracardiac source forthis sound and of its origin from contraction ofskeletal muscle.

614

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from

Page 8: Pacemaker 'Heart Sound' · ings, the pacemaker heart sound has been further investigated. SUBJECTS ANDMETHODS In 6 patients with chronic heart block on artificial pacemakers a four-channel

Pacemaker 'Heart Sound'

Many patients on artificial pacemakers have an

extra heart sound which is always earlier, and oftenlouder, than the usual heart sounds. Five patientswith this extra sound have been studied by externalphonocardiograms and apex cardiograms, and one

by intracardiac phonocardiography and pressure

recordings.The extra sound occurred 6 m.sec. after the pace-

maker impulse was recorded on the phonocardio-gram, and was accompanied by an early outwardmovement of the apex cardiogram, but was not asso-ciated with premature ventricular pressure rise, norcould it be recorded within the right ventricle dur-ing right ventricular endocardial pacing.The extra sound continued to occur when the

pacemaker impulse fell in the refractory period ofventricular ectopic beats, and could be abolished bya neuromuscular blocking agent or by repositioningthe endocardial electrode. It was concluded thatthe extra sound was due to stimulation of intercostalnerves adjacent to the epicardial or endocardial

electrode and was not due to premature contractionof cardiac muscle.

I wish to thank Dr. Aubrey Leatham for his helpfuladvice in the preparation of this paper.

REFERENCESBowditch, H. P. (1871). tYber die EigenthUimlichkeiten der

Reizbarkeit, welche die Muskelfasern des Herzenszeigen. Ber. sachs. Ges. Wiss. (Math.-phys. classe), 23,652.

Haber, E., and Leatham, A. (1965). Splitting of heart soundsfrom ventricular asynchrony in bundle-branch block,ventricular ectopic beats, and artificial pacing. Brit.HeartJ., 27, 691.

Harris, A., Bluestone, R., Busby, E., Davies, G., Leatham, A.,Siddons, H., and Sowton, E. (1965). The managementof heart block. Brit. Heart_J., 27, 469.

Nager, F., Biihlmann, A., Schaub, F., Schwarz, H., andSenning, A. (1965). Auskultatorische und kardio-graphische Befunde bei Patienten mit implantiertemelektrischem Schrittmacher. Klin. Wschr., 43, 1232.

Tafur, E., Cohen, L. S., and Levine, H. D. (1964). Thenormal apex cardiogram. Its temporal relationship toelectrical, acoustic, and mechanical cardiac events.Circulation, 30, 381.

615

on April 27, 2020 by guest. P

rotected by copyright.http://heart.bm

j.com/

Br H

eart J: first published as 10.1136/hrt.29.4.608 on 1 July 1967. Dow

nloaded from