P375 Universal Output Probe - movingpixel.com · 2010. 1. 29. · P375 Universal Output Probe...

7
Squire Probe Adapter Data Sheet & Users Manual January 2009 - Rev 1.2

Transcript of P375 Universal Output Probe - movingpixel.com · 2010. 1. 29. · P375 Universal Output Probe...

  • Squire Probe Adapter

    Data Sheet&

    Users Manual

    January 2009 - Rev 1.2

  • Squire Probe Adapter

    1.0 General:The Squire Probe Adapter is used to connect a Tektronix P64341 logic analyzer probe to a Tektronix TLA logic analyzer module that uses high-density front panel connectors (eg TLA7AA4, TLA7BB4). It has been optimized to work well with high level logic signals such as TTL and LVCMOS. It will not work well with signals with low swings (like doubly-terminated SSTL-2 1.5 volt)

    The Squire amplifies and buffers the 34 channels of the Probe for use with the logic analyzer module. A proprietary autocal scheme ensures offset voltages are tightly controlled. The output of the Squire is analog; thresholding is performed in the logic analyzer as it would be in the case of a Probe without a Squire adapter.

    2.0 RequirementsP6434 probe or similar and TLA module (TLA7AAx or TLA7BBx)

    3.0 Usage

    CAUTION: once the Squire is attached to the Logic Analyzer, it forms a significant lever arm. Take great care to make sure no weight is placed on the Squire. Costly damage to the logic analyzer can occur if a user places any weight on the Squire. Do not lift or move the logic analyzer by using the Squire as a “handle” or grab point.

    For high-edge rate signals, we recommend setting the logic analyzer threshold to the center of the swing to avoid minor glitches undershoots and overshoots cause on the logic analyzer display. For example, for a signal rising at faster than 0.5V/nS and swinging from 0 to 3.3 volts, we recommend the user set the threshold on the logic analyzer to 1.70 volts. Settings more than 0.7 volts below or above this level may result in glitches being detected.

    1 and others, P6417, P6418, P6419

  • 3.1 Labels

    Locations are illustrated below.

    TLA Connector

    E0 E1 Q2

    PROBE CONNECTIONOBSERVE POLARITY

    BOTTOM VIEW

    Squire Adapter

    The Moving Pixel Companywww.movingpixel.com

    E0 E1 Q2 E2 E3 Q3

    TLA CONNECTIONOBSERVE POLARITY

    DO NOT OVERTIGHTEN SCREWS !!!

    TLA Connector

    E2 E3 Q3

    PROBE CONNECTIONOBSERVE POLARITY

    TOP VIEWwhen installed in TLA

    CAL

  • Apply labels as shown in the following table. The preferred labeling is show first, followed by the most likely legacy labeling. Note that some remapping of the logic analyzer is required for legacy labeling.

    For a P6434 probe:Probe high side refers to the TLA cable/connector from the pin 1 side of the Mictor.Probe low side refers to the TLA cable/connector from the pin 38 side of the Mictor.

    Mictor Pin 1 side (HIGH)

    Mictor Pin 38 side (LOW )

    note orientation of ground contact

    Probe LabelsPreferred

    Squire Labels TLA

    LOW HIGH Bottom Top TLA Connector

    Remapping

    E0 E1 Q2 E2 E3 Q3 E0 E1 Q2 E2 E3 Q3 E0 E1 Q2E2 E3 Q3

    none

    D2 D3 Q0 A2 A3 CK0 D2 D3 Q0 A2 A3 CK0 D2 D3 Q0A2 A3 CK0

    none

    D0 D1 CK2 A0 A1 CK1 D0 D1 CK2 A0 A1 CK1 D0 D1 CK2A0 A1 CK1

    none

    C0 C1 Q1 C2 C3 CK3 C0 C1 Q1 C2 C3 CK3 C0 C1 Q1C2 C3 CK3

    none

    Probe LabelsLegacy

    Squire Labels TLA

    LOW HIGH Bottom Top TLA Connector

    Remapping

    D0 D1 CK2 D2 D3 Q0 D0 D1 CK2 A0 A1 CK1 D0 D1 CK2A0 A1 CK1

    A0 = D2, A1 = D3, CK1 = Q0

    C0 C1 Q1 C2 C3 CK3 C0 C1 Q1 C2 C3 CK3 C0 C1 Q1C2 C3 CK3

    none

    A0 A1 CK1 A2 A3 CK0 D2 D3 Q0 A2 A3 CK0 D2 D3 Q0A2 A3 CK0

    D2 = A0, D3 = A1, q0 = CK1

    E0 E1 Q2 E2 E3 Q3 E0 E1 Q2 E2 E3 Q3 E0 E1 Q2E2 E3 Q3

    none

    Once labels are applied, connect the Squire adapter to the TLA connector indicated by the Squire label. Be careful not to over tighten the screws. We have found that the probe labels are better protected and last a bit longer if the user puts some clear tape over them.

    Connect the Probe connectors to the Squire, taking care to insert the Probe connector into the appropriate connector based on the labeling and watching the connector keying.

  • Connect the user end of the Probe to the system under test as shown below

    unit under test

    Probe Assembly

    Squireprobe

    adapter

    TLAModule

    Set the TLA logic thresholds based on the expected UUT signal voltages and the signal path gain of the combined Probe / Squire adapter.

    ProbeSquire Probe

    AdapterTLA ModuleUUT

    0

    1V

    0

    3.7mV

    0

    52 mV

    Gain = 1/267.7Offset = 0

    Gain = 14Offset = 0

    Signal Path Gains / Offsets

    0 – 1V input is an example.Allowed input range is +5 to -2.5 V

    End to end gain is nominally 0.052, End to end offset is nominally 0 volts

    For the 0 – 1V signal used as an example the TLA threshold should be set to 500 mV (internally the TLA makes 25 mV as the threshold).

  • The probe end of the Squire Adapter is shown below

    top when installed in TLA

    bottom LED2LED 1

    3.2 Controls and Indicators

    There are two LED indicators labeled led 1 and led 2 (see drawing above) and a CAL switch (see section 3.1).

    At power up, both leds blink a few times to let the user know they are operational.

    LED2 is a heartbeat LED to let you know that the unit is functioning. It blinks all the time except during calibration.

    The CAL switch is used to recalibrate DC offsets in the Squire adapter. The Squire is pre-calibrated at the factory. However, we recommend that the user recalibrate whenever the probe adapter is attached to a logic analyzer module or moved between logic analyzer connectors. Best results will be had if the logic analyzer and the probe adapter have both reached power-on thermal equilibrium, at least 10 minutes.

    The reason to calibrate the probe is so that the threshold between a one and zero does not drift around, changing timing and waveshape relationships at the logic analyzer. Calibration is more important with slow edges and low amplitude signals.

    To perform a calibration, use the following procedure.

    1. Allow the Squire to come to temperature equilibrium (10 minutes or more)2. Disconnect the Probe from the Squire.3. Push and hold the CAL button until LED1 lights (approx 1 second). 4. LED1 will go out when the calibration is complete (about 5 seconds)

  • 4.0 Specifications

    Characteristic Specification Notes

    DC Offset +/- 100mV typicalRisetime ~3nS or faster 20%-80%Data Rate/Bandwidth- Nominal frequency rolloff- -3dB (70.7% amplitude)- -6dB (half-amplitude)

    200Mbps/100 MHz400 Mbps/200 MHz600Mbps/300MHz

    Worst case, 85% amplitudeTypicalTypical

    Gain ~14 Nominal, inconsequential to the user.

    Specifications for the supported Tektronix logic analyzer probes can be found on Tek’s website: http://www2.tek.com/cmswpt/psdownload.lotr?ct=PS&cs=psu&ci=13438&lc=ENNote that Squire will support all these probes: P6434, P6410, P6419, and P6430 as well as the legacy P6417 and similar.

    http://www2.tek.com/cmswpt/psdownload.lotr?ct=PS&cs=psu&ci=13438&lc=ENhttp://www2.tek.com/cmswpt/psdownload.lotr?ct=PS&cs=psu&ci=13438&lc=EN

    Squire Probe Adapter