Oxygen Consumption in Two Portable Ventilators Using a ...Oxygen Consumption in Two Portable...

1
Oxygen Consumption in Two Portable Ventilators Using a High Pressure Gas Source JS Phillips 1,2 , LP Pangilinan 1 , EK Saliba 2 , MS Siobal 3,4 , EL Warnecke 2 University of California San Francisco, Department of Anesthesia and Perioperative Care at Zuckerberg San Francisco General Hospital, San Francisco, CA 1 ; Respiratory Therapist Program, Ohlone College, Newark, CA 2 . Background The VOCSN (Ventec Life Systems, Bothell, WA) and Trilogy (Philips Respironics, Murrysville, PA) ventilators can deliver oxygen via low or high pressure sources utilizing different circuits. We evaluated and compared oxygen consumption from a high pressure gas source using the two ventilators with a null hypothesis that there would be no difference. Methods The VOCSN and Trilogy ventilators were evaluated using a high pressure gas source (e-cylinders regulated to 50 psi) with both passive (constant leak) and active (exhalation valve) circuits connected to a TTL test lung (Michigan Instruments, Grand Rapids, MI). The three simulated TTL lung models were: normal - Cst 60 mL/cm H2O and Raw 5 cm H2O/L/s, restrictive - Cst 30 mL/cm H2O and Raw 5 cm H2O/L/s, and obstructive - Cst 60 mL/cm H2O, Raw 20 cm H2O/L/s. Using pressure ventilation modes, the peak pressure and rise time were titrated to achieve a tidal volume of 500 mL, as measured by a Certifier FA Plus (TSI Inc, Shoreview MN). Other settings include: frequency 12 breaths/min, PEEP 5 cm H2O and IT 1.0 second, set FIO2 of 0.40. Delivered FIO2 was measured by the Handi + oxygen analyzer (Maxtec, Salt Lake City, UT) at the lung inlet. In addition, the VOCSN pulse dose function was evaluated by titrating its oxygen flow to achieve an FIO2 of approximately 0.40. We measured the length of time to reduce e-cylinder pressure by 100 PSI for each lung model and test configuration, then calculated the liters of oxygen utilized per minute during each test run. A series of three tests were performed for all lung models and circuit configurations. Data for the three lung models were averaged for each test configuration and reported as the mean ± SD for both circuits. Results Tidal volume delivery and measured FIO2 remained relatively constant during all lung models, test configurations and circuit types (501 ± 7 mL and 0.397 ± 0.01 respectively). Oxygen consumption using VOCSN with pulse dose oxygen delivery was 1.7 ± 0.7 and 1.7 ± 0.1 L/min, using VOCSN with set FIO2 was 4.8 ± 0.8 and 2.4 ± 0.1 L/min, and using Trilogy with set FIO2 was 5.2 ± 1.0 and 4.8 ± 0.9 L/min with passive and active circuits respectively. Conclusions Oxygen utilization was lowest using the VOCSN ventilator with pulse dose oxygen delivery with both the passive and active circuits, and with VOCSN using a set FIO2 with the active circuit compared to the Trilogy 202 ventilator. Disclosures Study made possible through grant/research support from Ohlone College and Ventec Life Systems. Mr. Siobal serves as a consultant and discloses commercial relationships with Aerogen 3 and Origin Medical 4 . Acknowledgements This project would not have been possible without the vision of Mr. Siobal and Ms. Warnecke. We are extremely grateful for their guidance and thank them for their dedication to respiratory care research. Figure 1 Average liter flow (L/min) consumption required to maintain a target fraction of inspired oxygen (FIO2) equal to forty-percent (0.40). Figure 3 VOCSN multifunctional critical-care ventilator (Courtesy of Ventec Life Systems) Figure 2 Average liter flow (L/min) consumption required to maintain a target fraction of inspired oxygen (FIO2) equal to forty-percent (0.40), categorized by: ventilator, circuit type, and lung physiology

Transcript of Oxygen Consumption in Two Portable Ventilators Using a ...Oxygen Consumption in Two Portable...

Page 1: Oxygen Consumption in Two Portable Ventilators Using a ...Oxygen Consumption in Two Portable Ventilators Using a High PressureGas Source JS Phillips1,2, LP Pangilinan1, EK Saliba2,

Oxygen Consumption in Two Portable Ventilators Using a High Pressure Gas SourceJS Phillips1,2, LP Pangilinan1, EK Saliba2, MS Siobal3,4, EL Warnecke2

University of California San Francisco, Department of Anesthesia and Perioperative Care at Zuckerberg San Francisco General Hospital, San Francisco, CA1; Respiratory Therapist Program, Ohlone College, Newark, CA2.

BackgroundThe VOCSN (Ventec Life Systems, Bothell, WA) andTrilogy (Philips Respironics, Murrysville, PA) ventilatorscan deliver oxygen via low or high pressure sourcesutilizing different circuits. We evaluated and comparedoxygen consumption from a high pressure gas sourceusing the two ventilators with a null hypothesis that therewould be no difference.

MethodsThe VOCSN and Trilogy ventilators were evaluated using ahigh pressure gas source (e-cylinders regulated to 50 psi)with both passive (constant leak) and active (exhalationvalve) circuits connected to a TTL test lung (MichiganInstruments, Grand Rapids, MI).

The three simulated TTL lung models were: normal - Cst60 mL/cm H2O and Raw 5 cm H2O/L/s, restrictive - Cst 30mL/cm H2O and Raw 5 cm H2O/L/s, and obstructive - Cst60 mL/cm H2O, Raw 20 cm H2O/L/s.

Using pressure ventilation modes, the peak pressure andrise time were titrated to achieve a tidal volume of 500 mL,as measured by a Certifier FA Plus (TSI Inc, ShoreviewMN). Other settings include: frequency 12 breaths/min,PEEP 5 cm H2O and IT 1.0 second, set FIO2 of 0.40.Delivered FIO2 was measured by the Handi + oxygenanalyzer (Maxtec, Salt Lake City, UT) at the lung inlet. Inaddition, the VOCSN pulse dose function was evaluated bytitrating its oxygen flow to achieve an FIO2 ofapproximately 0.40.

We measured the length of time to reduce e-cylinderpressure by 100 PSI for each lung model and testconfiguration, then calculated the liters of oxygen utilizedper minute during each test run. A series of three testswere performed for all lung models and circuitconfigurations. Data for the three lung models wereaveraged for each test configuration and reported as themean ± SD for both circuits.

ResultsTidal volume delivery and measured FIO2 remainedrelatively constant during all lung models, testconfigurations and circuit types (501 ± 7 mL and 0.397 ±0.01 respectively). Oxygen consumption using VOCSNwith pulse dose oxygen delivery was 1.7 ± 0.7 and 1.7 ±0.1 L/min, using VOCSN with set FIO2 was 4.8 ± 0.8 and2.4 ± 0.1 L/min, and using Trilogy with set FIO2 was 5.2 ±1.0 and 4.8 ± 0.9 L/min with passive and active circuitsrespectively.

ConclusionsOxygen utilization was lowest using the VOCSN ventilatorwith pulse dose oxygen delivery with both the passive andactive circuits, and with VOCSN using a set FIO2 with theactive circuit compared to the Trilogy 202 ventilator.

Disclosures

Study made possible through grant/research support from Ohlone College and Ventec Life Systems.

Mr. Siobal serves as a consultant and discloses commercial relationships with Aerogen3 and Origin

Medical4.

Acknowledgements

This project would not have been possible without the vision of Mr. Siobal and Ms. Warnecke. We are extremely

grateful for their guidance and thank them for their dedication to respiratory care research.

Figure 1Average liter flow (L/min) consumption required to maintain a target

fraction of inspired oxygen (FIO2) equal to forty-percent (0.40).

Figure 3VOCSN multifunctional critical-care ventilator

(Courtesy of Ventec Life Systems)

Figure 2Average liter flow (L/min) consumption required to maintain a target

fraction of inspired oxygen (FIO2) equal to forty-percent (0.40), categorized by: ventilator, circuit type, and lung physiology