Oxy Megastation OXYGEN DERMABRASION VACUUM RF SYSTEM OPERATOR’S MANUAL Serial Number: OXY MEGASTATION Hermes Technology Co., Ltd www.starbelle.cn

Transcript of Oxy megastation - f01.s.alicdn.com

Page 1: Oxy megastation - f01.s.alicdn.com

Oxy MegastationOXYGEN DERMABRASION VACUUM RF SYSTEM

OPERATOR’S MANUAL

Serial Number: OXY MEGASTATION

Hermes Technology Co., Ltdwww.starbelle.cn

Page 2: Oxy megastation - f01.s.alicdn.com

Copyright © 2013, Guangzhou HEMS Technology Co., Ltd.

No part of this manual may be reproduced or copied in any form by any means -graphic, electronic or mechanical, including photocopying, typing, orinformation retrieval systems - without written permission of HEMS Company.

System specifications and the information contained in this document are subjectto change without notice.

All rights reserved. Hermes Technology Co., Ltd., its logo,

are the trademarks of Guangzhou HEMS Technology Co., Ltd

Guangzhou HEMS Technology Co., Ltd

Email: [email protected]

Office adress: Rm508A, Chuangzhan Business Building, Rongxi Street, Baiyun

Guangzhou, ChinaTel: +86-20-22847317, 22847318Fax: +86-20-22847315

Factory adress: Floor 1-4, No.38 Jiaoxin Village, Shijing Town ,Baiyun District,

Guangzhou, ChinaTel: +86-20-61163310

ESWT™, RSWT™, ShockWave™,

Cryo Cavitation™, iBelle™, PEMF™,

Dermolift RF™, Dermolipo RF™, Cyclone RF™,

Pixel RF™, Lipo Electromagnetic™, CrioWave™,

Oxygen Peel™, Acoustic Wave™, Cryolipo™

Page 3: Oxy megastation - f01.s.alicdn.com

Warning!!!

Before attempting to use or operate the system, operator should read this manual andbecome thoroughly familiar with all its safety requirements and operating procedures.

ContentsChapter 1 System Overview

Chapter 2 Installation

Chapter 3 General Precautions and Cautions

Chapter 4 Operating Instructions

Chapter 5 After Sale Service

Chapter 6 Engineering Drawing

Chapter 7 Technique Data

Chapter 8 Appendix A

Page 4: Oxy megastation - f01.s.alicdn.com

Chapter 1 System Overview

1. Face Oxygen Infusion connector

2. Oxygen Spray connector

3. Aqua Oxygen Peel connector

4. Body Oxygen Infusion connector

5. Dermabrasion connector

6. Breast Cup connector

7. Foot Switch connector for RF

8. Face RF connector

9. Body Tri Sculptor connector

10. Power Switch

11. Power Connector

12. Left filter for Dermabrasion

13. Right filter for Breast Cup

Page 5: Oxy megastation - f01.s.alicdn.com

Chapter 2 InstallationBefore the System leaving factory, the HEMS Company will do the best to protect the

equipment during transportation.

System transportation is carried out by third party, not by HEMS Technology Co., Ltd

NOTE

Any damage to the packaging or to the system found prior to opening the package should be

reported to HEMS Company by photos and to the insurance carrier.

Equipment List

Please carefully check the following illustration showing that the Pro Cryolipo System

should include:

1. Main Unit

2. Applicators ( Face Oxygen Infusion + Oxygen Spray + Aqua Oxygen Peel + Body

Oxygen Infusion + Dermabrasion + Breast Cup + Face RF + Body Tri Sculptor)

3. Operator’s Manual

4. Input Power Cable

5. Appplicator support6. Foot Switch

7. Extra Fuse

8. Plastic Tube

Page 6: Oxy megastation - f01.s.alicdn.com

Chapter 3 General Precautions and CautionsThe following precautions, cautions and warnings must be observed for the safe use of the

system.

Precautions

l Operator should read this manual thoroughly before attempting to operate thesystem.

l Probes must be kept clean at all times.l The system is heavy and may cause injury if proper care is not used when moving it.

The system is well balanced and is designed to be moved, but should always bemoved carefully and slowly. Never pull the system by the probes.

Cautions

l Only after Artemis’ authorization, then may service the system. This includesmaking internal adjustments to the power supply, cooling system, probes, etc.

l Maintenance performed by the operator must only take place when the system isshut down and disconnected from the electrical power source. Performingmaintenance procedures with the system powered-up can be hazardous to theoperator and/or destructive to the system.

l Always turn off the system when it is not in use.l Never leave the system in Pause Mode and Start Mode unattended.l Never press the Start Softkey unless the probe is safely oriented at a specific and

intended target.l The probes must always be visually inspected before treating the patient. If any

damage is apparent, do not use the probes.l Never leave the system turned on, open or unattended during system maintenance.

NOTE

Please adjust the emitting energy step by step as per the communication with client.

Explosion and Fire Hazards

l The system is not suitable for use in the presence of flammable mixtures with air oroxygen.

l Do not operate in the presence of volatile solvents such as alcohol, gasoline or othersolvents.

l Do not use any flammable substances such as alcohol or acetone in the preparationof the skin for treatment. If necessary, use soap and water to clean skin beforetreatment.

l If alcohol is used to clean and disinfect any probe, allow it to dry thoroughly beforeoperating the system.

l Flammable materials must be kept at a safe distance from the system.

Page 7: Oxy megastation - f01.s.alicdn.com

Chapter 4 Operating Instructions

Prepare the System for Operation

To prepare the system for operation:

1. Check the connection of the system and the AC mains power.

2. Ensure that the applicators are connected correctly.

Operating the System:

Turn on the Probe Plug to enter into Main Interface

Touch each applicator icon and enter into each applicator treatment interface.

Page 8: Oxy megastation - f01.s.alicdn.com

Face Oxygen Infusion Operation Interface

1. Oxygen Power icon : Touch to adjust the

power.

2. Treatment time icon :Touch to set thetreatment time, every 1 minute a interval.

3. Oxygen switch icon :Touch to power onand off the Oxygen input and output

4. Start icon :Sart the treatment.

5. Pause icon :Pause the treatment.

6. Stop icon :Stop the treatment.

7. Back icon :Touch it and enter into the main interface.

Page 9: Oxy megastation - f01.s.alicdn.com

Oxygen Spray Operation Interface

Aqua Oxygen Peel Operation Interface

Inerface Operations of Oxygen Spray and Aqua Oxygen Peel are

similar with Face Oxygen Infusion

Page 10: Oxy megastation - f01.s.alicdn.com

Body Oxygen Infusion Operation Interface

1. Treatment time icon :Touch to set thetreatment time, every 1 minute a interval.

2. Oxygen Power icon : Touch to adjust the

power.

3. Oxygen Power icon : Touch to adjust the

power.

4. Electrophorasio Wave Form iron : Touch

to adjust the wave form, total 7 waves.

5. Electrophorasio Power iron : Touchto adjust E power from 0 to 100%

6. Start icon :Sart the treatment.

7. Pause icon :Pause the treatment.

8. Stop icon :Stop the treatment.

9. Back icon :Touch it and enter into the main interface.

Page 11: Oxy megastation - f01.s.alicdn.com

Dermabrasion Operation Interface

1. Vacuum Power icon : Touch to adjustthe Vacuum suction power from 0 to 100%.

2. Treatment time icon :Touch to set thetreatment time, every 1 minute a interval.

3. Start icon :Sart the treatment.

4. Pause icon :Pause the treatment.

5. Stop icon :Stop the treatment.

6. Back icon :Touch it and enter into the main interface.

Page 12: Oxy megastation - f01.s.alicdn.com

Breast Cup Operation Interface

1.Treatment time icon :Touch to set thetreatment time, every 1 minute a interval.

2. Vacuum Power icon : Touch toadjust the Vacuum suction power from 0 to 100%.

3. Vacuum mode icon : Touch to adjustthe Vacuum mode among M1, M2, M3, AUTO working mode.

4. Electrophorasio Wave Form iron : Touch

to adjust the wave form, total 7 waves.

5. Electrophorasio Power iron : Touchto adjust E power from 0 to 100%

6. Start icon :Sart the treatment.

7. Pause icon :Pause the treatment.

8. Stop icon :Stop the treatment.

9. Back icon :Touch it and enter into the main interface.

Page 13: Oxy megastation - f01.s.alicdn.com

Face RF Operation Interface

1. Treatment time icon :Touch to set thetreatment time, every 1 minute a interval.

2. RF Power icon : Touch to adjust the

Vacuum suction power from 0 to 100%.

3. RF mode icon : Touch to adjust theVacuum mode among M1, M2, M3, AUTO working mode.

4. Start icon :Sart the treatment.

5. Pause icon :Pause the treatment.

6. Stop icon :Stop the treatment.

7. Back icon :Touch it and enter into the main interface.

Page 14: Oxy megastation - f01.s.alicdn.com

Body Tri Sculptor Operation Interface

1.Treatment time icon :Touch to set thetreatment time, every 1 minute a interval.

2. RF Power icon : Touch to adjust the

Vacuum suction power from 0 to 100%.

3. RF mode icon : Touch to adjust theVacuum mode among M1, M2, M3, AUTO working mode.

4. Ultrasonic mode icon : Touch toadjust the Vacuum mode among M1, M2, M3, AUTO working mode.

5. Ultrasonic power iron : Touch to

adjust E power from 0 to 100%

6. Start icon :Sart the treatment.

7. Pause icon :Pause the treatment.

8. Stop icon :Stop the treatment.

9. Back icon :Touch it and enter into the main interface.

Page 15: Oxy megastation - f01.s.alicdn.com

Remark:

Electrophoresis working modes:

1. Mode : Galvanic Current,

Leading ion to permeate into the skinBreaking down the contamination in the skinAccelerating cell updated speedStimulation the speed of fluid flow

2. Mode : Faradic Current Square Wave

Stimulating muscle contractions & passivemovement

Lifting, tightening and Contouring Sagging Skin

3. Mode : Faradic Current Spike Wave

Stimulating Muscle intensivelyTraining muscle to build a perfect & firm bodyline

4. Mode : Faradic Current Twin Spike Wave

Powerful Electrophoresis penetrationTransforming fat into energy to consume

5. Mode : Faradic Current Step Wave

Contracting the muscles rhythmicallyAiming at swollen leg exercise

6. Mode : Faradic Current Trapezoid WaveContracting Muscles quicklyImproving the volume of blood circulationDraining away the internal waste of our body

7. Mode : Faradic Current Triangle WaveStimulating lightly to relax exhausted musclesAiming at Lymphatic System Care

Page 16: Oxy megastation - f01.s.alicdn.com

Vacuum Working Modes:

M1 working mode: Continuous mode

M2 working mode: Pulse mode, suck 2 seconds and then releaseM3 working mode: Pulse mode, suck 3 seconds and then releaseM4 working mode: Pulse mode, suck 4 seconds and then release

RFWorking Modes:

Auto working mode: Continuous-(2/3)-(1/2) repeats automaticallyM1 working mode: Continuous pulsation (0.5s/0.5s)M2 working mode: 1/2 pulsation (1s/2s)M3 working mode: 2/3 pulsation (2s/3s)

Total Time Interface

If need to clearance the total treated time for each treatment, please touch each

ten times, and then touch the treatment time icon once.

Page 17: Oxy megastation - f01.s.alicdn.com

Chapter 5 After Sale Service

Warranty Coverage

l HEMS warrants defects in materials and workmanship for a period of 1 year from the

date of original purchase.

l HEMS will repair these products at no charge or refurnish the parts if it has any

abnormal condition.

Exclusions

This warranty is void and invalid in case of ;l Damage caused by accident, abuse, misuse or misapplication, it will be repaired with

repair fee

l Damage caused by service performed by anyone who is not technician or qualified by

manufacturer.

l A product or part that has been modified without any permission of manufacturer.

l AProduct that has no serial number.

Maintenance

HEMS provides lifetime maintenance. Only charge the cost price for replacing the

component.

NOTE

1. After treatment, keep the power on about 3 minutes to make inner board cooling.

2. Please keep equipment in the shade, avoiding direct lay of light.

3. Don’t drop the equipment & keep it out of any shock.

4. Clean probes after use.

5. Moving the equipment, be careful not to drop applicators

Page 18: Oxy megastation - f01.s.alicdn.com

Chapter 6 Engineering DrawingThe Engineering Drawing guides to check and repair the system.

When the system meets problem, please take the photos and give the feedback to Hermes

Technology by email, Artemis will handle the problems as soon as possible.

Burned Fuse Replacement

1. Turn off the system and disconnect the power cable from the outlet.

2. Unplug the power cable from the power cable connection port on the service panel.

3. Remove the fuse housing as shown in Figure .

4. Remove the burnt fuses from the housing and replace them with the new ones.

5. Make sure that the fuses fit snugly within the fuse housing.

6. Insert the fuse housing into its receptacle and push it in until it snaps into place.

7. Connect the power cable to the system and to the mains outlet.

8. Turn on the system and verify that it is functioning properly.

Caution

l Make sure that the type and rating of the new and the replaced fuses match.

l Always replace both fuses at the same time.

Page 19: Oxy megastation - f01.s.alicdn.com

Chapter 7 Technique Data

Display 8” Color touch screen

Electro Frequency 1000Hz

Dermabrasion tips 9 pieces of different sizes

Oxyge purity Concentration

Oxygen source Without Cylinder

RF Frequency 6MHz

RF power 50W

Vacuum range 650mmHg

Ultrasonic Frequency 3MHz

Input Voltage 100V-110V, 220V240V/50-60Hz

Time Setting 5 to 30 minutes

Mark CE

Chapter 8 Appendix A

Please connect with your sales manager.