Out of Specification and Atypical Test Results. Contents Objective Definitions and terminology...

32
Out of Specification and Atypical Test Results

Transcript of Out of Specification and Atypical Test Results. Contents Objective Definitions and terminology...

Out of Specification and

Atypical Test Results

Contents

• Objective• Definitions and terminology• General investigation principles• Investigation stages

– Initial Laboratory Investigation– Full-Scale OOS Investigation– Additional Laboratory Testing– Review of Production

• Importance of corrective and preventive actions

Process overview

Evaluation of results

Analysis of sample

Sampling

Manufacturing of batch

Cause for an investigation

Evaluation of results

Analysis of sample

Sampling

Manufacturing of batchOut of

Specification arises:

Investigation

Analytical (lab) error?

Root cause assigned?

Corrective and

Preventive Action?

Production error?

References

FDA Guidance for Industry : Investigating Out-of-Specification (OOS)Test Results for Pharmaceutical Production, October 2006 : Contains Nonbinding Recommendations

USP– NF General Chapter <1010> Analytical Data – Interpretation and treatment

ICH Q7A Note for Guidance on Good Manufacturing Practice for Active Pharmaceutical Ingredients, November 2000

EU Guidelines to GMP Part I Chapter 6 Quality Control Section 6.32.

Definitions and Terminology (1)

• Out of Specification (OOS) Result– Reportable result that falls outside established specifications or

acceptance criteria.– Do not include situations where Control test is not continued due to

violations of system suitability tests (SST) limits– Identification of gross operator-errors during a run: e.g. wrong

instrument settings– Such deviation cases should be tracked in order to apply CAPA if

appropriate

• Reportable Result– Result that must be compared to a particular acceptance criterion.– May be “an average value, an individual measurement, or something

else”, as defined in the control test [USP]

Definitions and Terminology (2)

• Established Specifications or Acceptance criteria:– Approved specifications or acceptance criteria (e.g. limits) with respect to the

final result (reportable result) in order to evaluate the quality of a product or sample.

– May be official (part of a regulatory file) or internal (e.g. additional or alternative control tests)

These do not apply to:– Acceptance criteria during the course of analysis (before the reportable result)

such as system suitability tests (SST), warning limits, in-process control limits for process adjustments

• Atypical Test Result or Out of Trend / expectation, suspect, …– Result within specification– Result different from those usually obtained or expected– Same general investigation principles as OOS results– Site Quality Control Manager must determine extent and depth of any– investigation

Definitions and Terminology (3)

• Re-analysis– Analysis of retained sample preparations– Repeating some part of the testing, e.g.

re-measurement re-injection redilution of sample and standard preparations Further extraction

• Fresh analysis– Performing the testing as normal (New execution of the

control test)– As described in the control test– Only allowed if initial results could be invalidated

Definitions and Terminology (4)

• Retest :– New preparation and analysis of a portion of the original sample– Part of a full-scale OOS investigation– Using a predefined procedure/plan with a MAXIMUM number of

retests,– Used to verify/ reject the possibility of laboratory errors

• Control Sample :– A sample of material that has previously been tested and approved or

well characterised.

• Re-sampling :– Collecting and analyzing a new sample from the product– Must have a documented rationale for re-sampling, (e.g. original

sample not representative or compromised/contaminated)– Used for confirmation of product failure

Presentation Scope

• Applicable to QC laboratories, Manufacturing sites and Affiliates

• All GMP relevant laboratory testing including inprocess controls– Where approved specification or acceptance

limits are established.

Note: Compendial testsWhere interpretation and handling of test results are described, those procedures are to be followed.

Principles of Investigation (1)

• All OOS or atypical test results must be thoroughly investigated and documented– Includes identification of root cause

• OOS or Atypical test results are only allowed to be invalidated if a laboratory error can be assigned– Demonstrated or indicated with high probability and/or a product failure definitely

excluded– Supporting information and/or data

• Timely initiation (e.g. 2 business days) and completion (e.g. 20 business days)– Otherwise written rationale– Sterile products: a parallel investigation in the manufacturing area must be

initiated immediately– A risk assessment and notification in case of confirmed OOS for product already on

the market or for a risk that could affect other batches or products on the market

Principles of Investigation (2)

• Consider:– Problem has occurred previously (historical data)– Possibility that other batches or products are affected

• Corrective actions must be defined as a conclusion of the investigation and followed through to prevent reoccurence

• A system must be in place for the tracking of OOS results

• Investigation phases– Initial Laboratory Investigation– Full-Scale OOS Investigation– Additional Laboratory Testing– Review of Production

Responsibilities (1)

• Analyst– Achieves accurate analytical results– Should not use systems that do not meet system suitability

requirements (SST); or in such case during analysis identifies data collected during suspect time period (e.g. reference standard injection runs in a chromatographic system)

– Should not continue test in case of obvious error (spilling of a sample solution, incorrect dilution etc.)

– Immediately reports the OOS to his/her supervisor– Retains the original test solutions, materials and source data– Participates in the investigation

Responsibilities (2)

• Supervisor / QC Manager– Conducts an objective and timely investigation– Informs Site Quality Manager (immediately in the case of a sterile

product)– Decides upon investigation of atypical results

• Site Quality Manager– Ensures specific site procedures and systems exist– Ensures appropriate training given– Makes batch related decisions when an OOS or atypical result is

confirmed– Informs all relevant departments– Assuring that corrective actions are taken to prevent reoccurrence– For sterile products, promptly initiates parallel investigation in

laboratory AND manufacturing area.

Initial Laboratory Investigation

OOS/OOT results identified by Analyst. Supervisor informed immediately

INITIAL LABORATORY INVESTIGATION

Assess whether the test procedure was followed correctly (calculation; sampling; sample preparation; equipment

functionality& qualification; training etc)

Laboratory Error Identified?

YES NO

See slide 18(Lab error IS identified)

Full scale OOS investigation required(See slide 20 )

Initial Laboratory Investigation:examples (1)

• Method discussed between analyst and supervisor : confirm analyst knowledge of and performance of the correct procedure

• Verification of calculations• Review of samples for correct labelling and identification• Verification of correct laboratory test methods• Examination for proper documentation• Review of chromatograms, spectra, data and calculations• Review of reagents, media, diluents, controls and standards• Examination of the instruments and laboratory systems

(qualification, calibration, maintenance)• Review other samples run concomitantly (if any)

Initial Laboratory Investigation:examples (2)

• Verification of analyst’s training history• Inspection of prepared sample• Reanalysis/Redilution of test samples (if stable)• Supervisor fully document and preserve records of the

Initial Assessment Everything except new sample preparation

Laboratory Error is Identified

Invalidate original test results

(Initial Lab investigation- slide 15)

Laboratory error IS identified

Consider Implication on other analytical tests/products CAPA plan

Fresh analysis

New Evaluation

Conclusion

Evaluation if positive

Implement CAPA action plan and follow through

Categorising Errors

Error type: Description:

Transcription Error Wrong interpretation or calculation

Identifiable Analytical Error

• Non-adherence to control test instructions (wrong materials, dilutions, instrument settings etc.)• Equipment malfunctions

Anticipated OOS E.g. stability, accelerated conditions or beyond shelf-life

Consider Actions for improvement of processes/ prevention of errors – e.g.

training analysts; instrument re-qualification, update to procedures.

Full scale OOS Investigation

Additional Laboratory testing

Laboratory error is NOT identified (from slide 15):

Full Scale OOS investigation

Review of Production (slide 22)

Full Laboratory investigation

Follow Company protocols on:• Re-testing• Re-sampling

Areas to review will include:• Batch Dossier Evaluation• Equipment: Validation,

Qualification, Calibration• Training of Personnel• Risk Management Evaluation

OOS Confirmed?(See slide 21)

Deviation explaining OOS confirmed?

NO

YES

Batch Failure Investigation(See slide 21)

OOS Confirmed?

Batch Failure Investigation

(Additional Laboratory testing, slide 20)

OOS Confirmed?

Lab error identified?

Evaluate rejection of batch• Risk Assessment• Quality Alert Notification

Impact on Other Products?

Take all results into consideration

YES NO

Investigation report

Document. If yes, further

evaluation required

NO

YESSee slide

18, above.

Review of Production

• Involvement required of all departments that could be implicated: Manufacturing, Process Development, Maintenance, Engineering, Sub-contractor

• Timely, thorough, well-documented review• Review of: batch dossier, equipment (qualification,

calibration), personnel training and risk management evaluation, etc.

• Once Root cause is identified and CAPA plan implemented: OOS investigation may be terminated

Additional Laboratory testing; retesting; Maximum number;

design

• Variables : number specified in advance, analyst, control samples …

• Must be defined before testing

• Case-by case decision, e.g. depending on– Possible cause of the OOS (analyst, equipment, sample …)– Level of confidence required– Analytical variability– “Extent” of OOS

Additional Laboratory testing: Resampling

• Resampling to be carried out: – in accordance with predetermined procedures and sampling strategies – where it is identified that the original sample was improperly prepared

and not representative of batch quality

• Use same sampling method unless it is demonstrated as inadequate

Reporting Testing results: Averaging General Considerations

• Mean Estimate of the true value Reliability increases with the number of determinations However: may hide (unacceptable) variability

• Reportable results and acceptance limits Correlated Evaluation of single results requires wider limits due to the broader

distribution If the mean is defined as reportable result, the variability should be

checked (standard deviation or range)

Reporting Testing results: Appropriate and Inappropriate Uses of

Averaging

Appropriate • When sample assumed homogeneous and if the written and

approved test method specify that the average of multiple replicates (assays) is considered one test and represents one reportable result

Inappropriate • When testing is intended to measure variability of a product

(blend uniformity, content uniformity)• In case of additional testing during OOS investigation because

variability is hidden– Specifically when some results are OOS and other within specifications

• Provide all individual results to quality management responsible for approving or rejecting the product

Reporting Testing results: Outlier Tests

• Outlier result, Definition: A value obtained, on rare occasions, markedly different from the others in a series, with a validated method

• Outlier testing: statistical procedure for identifying from an array those data that are extreme: – SOP is mandatory (includes minimum number of results required, to obtain a

statistically significant assessment)

• Used when it is not possible to reveal the root cause of OOS/ deviation

• May be used as investigative tool

• Should not be used as sole criteria to invalidate data

Concluding the Investigation

• Results evaluated, batch quality determined, release/rejection: decision by Quality Management

• OOS cause revealed: – Suspect result invalidated and not used– OOS confirmed and is caused by factor affecting batch quality– OOS result used to evaluate quality of lot

• OOS Cause not revealed: – Inconclusive investigation– OOS not confirmed– OOS result given full consideration for batch decision

Reporting and Documentation

• OOS report should include the following: Chronology of the investigation Full description of the initial laboratory investigation including

an analysis of all data derived from all testing Batches or products involved Justification for invalidating any data Conclusion including conformity of final results and batch disposition Corrective and preventive actions taken to prevent the OOS from occurring again

• Uniquely numbered to permit tracking• Maintained in a controlled and retrievable

manner

OOS Investigations

• Clear definition of the reportable result• Thorough, timely, well documented, scientifically

justified with no preconceived assumptions– Clear and methodical escalation

• Objective to assign a cause– Product failure– Laboratory (analytical) error

• Corrective actions to prevent reoccurrence• Atypical results: chance to increase data quality

– E.g. in stability studies

Examples: Identification of Analytical Errors

Examples of Errors Solutions

One injection OOS, reanalysis/reinjection ok

Instrument failure (consider maintenance, repair, qualification)

All injections of one sample preparation OOS, redilution ok

Dilution error (consider training, clarification of control test)

One sample preparation OOS, retesting 2nd analyst ok

Analyst error (consider training, clarification of control test)

Results near limit, one result just OOS; 1 out of 4 retests also OOS

Reportable result? May indicate unsuitable specification limits (too tight for the analytical variability)

All sample preparations OOS, retestingOOS, control samples ok

Sampling error or batch failure