Options for Fast, Reliable Pesticide Residues Analysis in Food by Triple Quadrupole GC-MS/MS
Richard Fussell1
Vertical Marketing Manager, Food and Beverage Michael T. Hetmanski2, Cristian Cojocariu3, Victoria Bailey-Horne2, Dominic Roberts3 and Paul Silcock3 1Thermo Fisher Scientific, Hemel Hempstead, UK, 2 Food and Environment Research Agency (FERA), York, UK , 3Thermo Fisher Scientific, Runcorn, UK
PP71668-EN 0515S
![Page 2: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/2.jpg)
2
Outline
• The analytical challenge
• User requirements for GC-MS/MS analysis of pesticides • Collaboration between Fera and Thermo Fisher Scientific;
method optimisation, results and learning experiences
• Latest GC-MS/MS developments • Summary
![Page 3: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/3.jpg)
3
• Wide range of matrices • Food • Environment
• Wide analytical scope • Low limits of detection • High sample throughput • Fast turnaround • Low cost of analysis
Analytical Challenges
RT: 4.57 - 37.12
6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36Time (min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rela
tive
Abun
danc
e
6.32
6.59
4.63
7.00 11.13
10.12
11.9925.20
14.969.50 23.13 29.6117.388.89
12.207.315.77 19.1315.57 15.848.20
26.4712.87 17.9513.77 23.58
28.7125.3620.625.09 21.7023.8319.25
27.7724.12
35.78
34.8134.3932.9230.11 31.00
NL:3.94E8TIC MS 2july2104_011
![Page 4: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/4.jpg)
4
Requirements for GC-MS/MS
• Selectivity
• High sensitivity for 2nd transition to enable identification of pesticides at low target concentrations
• Robustness in operation 24/7 and ease of use
• Reproducibility
• Software: Effective and ease of use
• Cost, maintenance and service
![Page 5: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/5.jpg)
5
1. Sample Prep: March 24th 2. LC-MS Analysis: April 29th
3. GC-MS Analysis: June 17th 4. Data Processing/Analysis: July 15th
Typical Pesticides Workflow Register at www.chromatographyonline.com/LCGCwebseminars
![Page 6: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/6.jpg)
6
Pesticide Analysis: Triple Quad GC-MS
Highly Selective Reaction Monitoring (SRM) Improved detection limits
![Page 7: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/7.jpg)
7
Selectivity: SIM and SRM
DDE-p,p’, 0.05 mg/kg in green tea, 1.0 uL splitless injection
Ion ratio = 2.7 Ion ratio = 3.5
![Page 8: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/8.jpg)
Thermo Fisher Scientific and Fera GC-MS/MS Collaboration (1)
![Page 9: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/9.jpg)
9
Analysis of QuEChERS Acetonitrile by GC?
• High expansion coefficient-limited injection volume
• Offline solvent exchange - requires care to avoid losses of volatile or unstable pesticides
• LVI with higher capacity liners or liner packing materials, but potential for adsorption of pesticides?
• Will repeated injection of acetonitrile extracts lead to a decrease in column efficiency?
• How quickly will matrix extracts contaminate the GC-MS
system?
![Page 10: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/10.jpg)
10
Evaluation of high sensitivity GC-MS/MS
• Thermo Scientific™ TSQ Quantum™ GC-MS/MS system
• Thermo Scientific™ Trace™ Ultra GC Chromatograph • Thermo Scientific™ TriPlus RSH ™ Autosampler • Column: TR-Pesticide (II), 30 m x 0.25 mm i.d., 0.25 µm film
• ‘test matrix’- fruit preserve
• 96 pesticides/metabolites:192 transitions
Evaluate direct analysis, of QuEChERS acetonitrile extracts
![Page 11: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/11.jpg)
11
SRM chromatograms – 1µL splitless
Splitless injection (1 µL) of acetonitrile extracts (d-PSA clean-up) of fruit preserve spiked at 0.01 mg/kg (= 10 pg on-column)
184.95>92.88 219.95>1
Dichlorvos
6.44
Pirimiphos- methyl
290.09>233.07 305.10>290.09
21.05
148.97>69.98 148.97>104.98
Captan (0.02 mg/kg)
24.9
226.03>206.03 163.02>206.03
Cyfluthrin (0.02 mg/kg)
47.03
180.99>151.99 252.99>93.00
Deltamethrin 51.6
Courtesy of Fera, UK
![Page 12: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/12.jpg)
12
Summary of validation data (1 µL splitless)
• QuEChERS extraction (citrate buffer/d-PSA) of fruit preserve
• Except chlorothalonil, recovery and % RSD data for the other 95 compounds met EU DG SANCO validation criteria
Results courtesy of Fera, UK
![Page 13: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/13.jpg)
13
Robustness (1 year) of instrument performance
• TSQ Quantum XLS • Analysed ~ 200 batches of samples
• ~ 6000 matrix injections (PTV-backflush)
• Kidney, liver, fish, meat, crustaceans, milk
• Response does vary but is always sufficient
• Analytical capillary column not replaced
• 2 breakdowns requiring service engineer
Information courtesy of Fera, UK
![Page 14: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/14.jpg)
Themo Fisher Scientific and Fera GC-MS/MS Collaboration (2)
![Page 15: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/15.jpg)
15
Thermo Scientific TSQ 8000 GC-MS/MS
The new Trace 1300 Series offers the most versatile GC platform in the market, with unique “instant connect” modularity for ground-breaking ease of use and performance, setting a new era in GC technology
![Page 16: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/16.jpg)
16
(No I/S)
Repeatability using the baffled liner
• [10ul (1/10 diln with EtAc] speed controlled solvent vent] • PTV Baffle Liner deactivated (Siltek),
Results courtesy of Fera, UK
![Page 17: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/17.jpg)
17
Repeatability Using Asymmetric Liners
• QuEChERS extracts of baby food (no IS) • High number of injector parameters to optimise
Results courtesy of Fera, UK
![Page 18: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/18.jpg)
18
Validation Results
Dichlorvos (m/z 185-93, 186.9-93)
Cyfluthrin (m/z 226.0-206.1, 163.0-127)
• Excellent results obtained • Cycle time approximately 40 minutes
(0.01 mg/kg, 10 pg on column) (0.001 mg/kg 2.5 pg on column)
Results courtesy of Fera, UK
![Page 19: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/19.jpg)
19
QuEChERS- GC-MS/MS Workflow Validation
Matrix Recovery (%) %RSD Average No out-with
criteria Average No outwith
criteria Apple 95 3 4.7 0 Grapes 100 4 4.1 2 lettuce 104 10 4,0 0 Oats 96 16 5.9 0
• According to SANCO/12571/2013 criteria • Recovery 70-120 % , RSDr (<20%)
• Splitless (1 µL) with cyclo-liner
Preliminary Data
Results courtesy of Fera, UK
![Page 20: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/20.jpg)
Thermo Scientific™ TSQ ™ 8000 Evo GC-MS/MS
![Page 21: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/21.jpg)
21
Increasing Laboratory Productivity
• Decrease analysis time by shortening the GC run times. • More samples in less time.
• Increase the number of pesticides in a run. • More SRMs to accommodate within an analytical run.
• Improve selectivity for various matrices.
• Increased number of SRMs per compound.
• See beyond the targets. • Full Scan and SRM data acquisition in the same experiment.
Expect More Performance
![Page 22: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/22.jpg)
22
Fast GC-MS Pesticide Residue Analysis
Challenges:
• Complexity of elution when using fast GC
• Large number of compounds (SRMs) in short time
• Many SRM transitions can result in sensitivity loss
Solution:
• High speed analyzer • Fast collision cell • Short SRM dwell times with very
short inter-scan delays
![Page 23: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/23.jpg)
23
TSQ 8000 Evo GC-MS/MS
Expect More Capacity
• Analytical instrumentation:
TSQ 8000 Evo GC-MS/MS Pesticide Analyzer TRACE 1310 GC Gas Chromatograph TriPlus RSH Autosampler (liquid injection set-up)
• EvoCell
• Rapid, innovative collision cell technology • Increased method capacity • More compounds • More SRM transitions • Up to 4x more transitions whilst maintaining method sensitivity low analyte concentrations
![Page 24: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/24.jpg)
24
Pesticide MRM Database
The problem:
• Growing list of target compounds require continuous adjustment to an existing SRM database.
• Some SRM transitions are not suitable for all matrices. Addition of new SRM transitions can be time consuming.
The solution:
• Automated SRM development with AutoSRM.
![Page 25: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/25.jpg)
25
AutoSRM: Fast, Simple Route to Optimized SRM
1) Precursor ion selection
2) Product ion selection
3) Collision energy optimization
AutoSRM automates the development of SRM methodology
![Page 26: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/26.jpg)
26
Highlights of AutoSRM
• Automates the following: • Creation of full scan, product ion scan, and SRM methods • Creation of sample sequences • Creation of data layouts for analyzing results • Selection of precursor, product, and collision energies
End result showing optimized transition
![Page 27: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/27.jpg)
27
Timed-SRM: Using Dwell Times Efficiently
Classical segmented SRM
TSQ 8000 EVO timed SRM
Classical segmented SRM:
• Complex to set up • Wasted dwell time • Reduced sensitivity • Reduced tolerance to RT shifts
TSQ 8000 Evo timed-SRM:
• Automated set-up • Full optimized dwell time • Optimal sensitivity • Increased resistance to RT shifts
![Page 28: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/28.jpg)
28
Increasing Laboratory Productivity
• Decrease the analysis time by shortening the GC run times. • More samples in less time. • More SRMIncrease the number of pesticides in a run. • s to accommodate within an analytical run.
• Improved selectivity for various matrices
• Increase the number of SRMs per compound.
• Seeing beyond the targets • Full Scan and SRM data acquisition in the same experiment.
![Page 29: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/29.jpg)
29
Decrease the Analysis Time
RT: 4.57 - 37.12
6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36Time (min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Re
lativ
e A
bu
nd
an
ce
6.32
6.59
4.63
7.00 11.13
10.12
11.9925.20
14.969.50 23.13 29.6117.388.89
12.207.315.77 19.1315.57 15.848.20
26.4712.87 17.9513.77 23.58
28.7125.3620.625.09 21.7023.8319.25
27.7724.12
35.78
34.8134.3932.9230.11 31.00
NL:3.94E8TIC MS 2july2104_011
Full scan 144 pesticides in baby food @ 0.2 mg/kg TG-5 SILMS, 30m x 0.25 mm x 0.25 µm GC run time: ~37 min
![Page 30: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/30.jpg)
30
Fast GC: 144 Pesticides in Baby Food
As more compounds are added to a single run, more complexity arises with the management of acquisition windows and compound co-elution. RT: 3.68 - 10.75
4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5Time (min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rel
ativ
e Ab
unda
nce
8.68 8.819.31
8.96
6.789.10
4.66
8.258.007.78 8.49 9.61
7.927.61
9.687.00 7.31
6.56 8.236.83
7.096.08
7.46 9.958.045.956.36
5.25 6.28 10.359.58
7.165.22
4.90 5.43 10.055.894.10 5.68
4.46 10.574.44 5.113.74 3.99
NL:1.28E9TIC MS 2july2104_057_140704112840SRM acquisition of 144 pesticides in baby food @ 0.2 mg/kg
GC column: TG-5 SILMS, 20m x 0.18 mm x 0.18 µm, faster temperature ramp GC run time: <11 min
![Page 31: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/31.jpg)
31
Increase Laboratory Productivity
• Decrease the analysis time by shortening the GC run times. • More samples in less time.
• Increase the number of pesticides in a run. • More SRMs to accommodate within an analytical run.
• Improved selectivity for various matrices
• Increase the number of SRMs per compound.
• Seeing beyond the targets • Full Scan and SRM data acquisition in the same experiment.
![Page 32: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/32.jpg)
32
More SRM Transitions/Compound for More Confidence
Tecnazene in baby food at 0.01 mg/kg level
![Page 33: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/33.jpg)
33
RT: 14.95 - 16.71
15.0 15.2 15.4 15.6 15.8 16.0 16.2 16.4 16.6Time (min)
0
10
20
30
40
50
60
70
80
90
100
0
10
20
30
40
50
60
70
80
90
100
Rel
ativ
e A
bund
ance
RT: 15.79AA: 4093580SN: INF
RT: 15.79AA: 1901556SN: INF
NL: 2.60E6TIC F: + c EI SRM ms2 [email protected] [176.095-176.105] MS Genesis 19May2014_03
NL: 1.21E6TIC F: + c EI SRM ms2 [email protected] [245.995-246.005] MS Genesis 19May2014_03
RT: 14.84 - 16.62
15.0 15.2 15.4 15.6 15.8 16.0 16.2 16.4 16.6Time (min)
0
10
20
30
40
50
60
70
80
90
100
0
10
20
30
40
50
60
70
80
90
100
Rel
ativ
e A
bund
ance
RT: 15.79AA: 3837740
RT: 15.79AA: 1690756
NL: 2.35E6TIC F: + c EI SRM ms2 [email protected] [176.095-176.105] MS Genesis 19May2014_05
NL: 1.03E6TIC F: + c EI SRM ms2 [email protected] [245.995-246.005] MS Genesis 19May2014_05
More Speed Maintaining the Sensitivity
DDE-p,p’ in green tea, 1917 SRMs Inj. 0.01 mg/kg on column
DDE-p,p’ in green tea, 44 SRMs Inj. 0.01 mg/kg on column
0.7 ms 27 ms
![Page 34: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/34.jpg)
34
Linearity: Dichlorvos
• Dichlorvos peak area response over 0.5–10 ppb (mg/kg), matrix-matched standard (baby food).
• Chromatograms (quantification and confirmation ions) at 10 ppb level.
8 SRMs/compound
2 SRMs/compound R2= 0.998
R2= 0.997
![Page 35: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/35.jpg)
35
Examples of Peak Area Repeatability: Fast GC method
Data analysis method maintenance
GC Performance Maintenance
MS Performance Maintenance
Acquisition Method Maintenance
Routine results
%RSD pesticides (0.001 mg/kg on column) baby food matrix ~144 compounds in <10 minutes
Compound % RSD (n = 10) BHC, Alpha 7.0 BHC, Beta 8.8 BHC, gamma 9.2 Chlorobenzilate 12.5 Chlorothalonil 12.6 Clomazone 6.3 Cyfluthrin peaks 1-4 9.3 DDE p, p 8.2 Dichlobenil 5.3 Dichlorvos 8.1 EPTC 5.3 Etridiazole (Terrazole) 4.6 Hexachlorobenzene 8.9 Methacrifos 7.6 Propachlor 11.0 Propham 11.5 Simazine 9.1 Tecnazene 5.6 Tefluthrin 7.0 Triallate 11.0
![Page 36: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/36.jpg)
36
Comprehensive Detection of Pesticides in a Sample
• Targeted screening and quantification of a number of pesticides is important.
• Increased interest in screening the samples for additional compounds beyond a target list.
• Less common or new pesticides or unknown
metabolic/transformation products could be detected in addition to targeted compounds.
• This requires fast analytical instrumentation able to acquire both full scan and SRM/SIM data simultaneously.
![Page 37: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/37.jpg)
37
Summing up
• Compound database management: • Auto-SRM automates SRM transitions development.
• Using the available dwell time wisely: • Timed-SRM ensures minimal loss of time spent to acquire data.
• Enhancing the number of compounds and increasing the number of SRM transitions per compound:
• EvoCell fast collision cell technology.
• Comprehensive detection of target pesticides and of additional contaminants potentially hazardous to human health:
• Simultaneous full scan and SRM/SIM data acquisition.
• A full validation of this fast method is ongoing and will be available soon.
Fast, easy, and robust pesticide analysis can be achieved using:
![Page 38: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/38.jpg)
38
Thermo Scientific Food and Beverage Community
• View application notes, on-demand webinars, product information, and many more resources on our Pesticides and Food Communities Libraries:
www.thermoscientific.com/pesticides www.thermoscientific.com/foodandbeverage
![Page 39: Options for Fast, Reliable Pesticide Residues Analysis in ...tools.thermofisher.com/content/sfs/brochures/13-PP... · Options for Fast, Reliable Pesticide Residues Analysis in Food](https://reader035.fdocuments.in/reader035/viewer/2022081606/5e79130037d83d66206fdc3b/html5/thumbnails/39.jpg)
39
Thank You for Listening
Questions?
Stay connected with us Twitter @ChromSolutions Chromatography Solutions Blog http://chromblog.thermoscientific.com/blog YouTube http://www.youtube.com/ChromSolutions Facebook http://www.facebook.com/ChromatographySolutions Pinterest http://pinterest.com/chromsolutions/