OPTICAL QUALITY OF DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

12
1 OPTICAL QUALITY OF DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS Andrea Galvis, MD 1,3 a Ivan Ossma, MD MPH MSc 1,2, a Sanjeev Kasthurirangan, PhD 4, b 1 Fundacion Valle del Lili, Cali-Colombia 2 Universidad Industrial de Santander, Bucaramanga-Colombia 3 Universidad Autonoma de Bucaramanga, Bucaramanga-Colombia 4 Abbott Medical Optics Inc, Irvine, USA Financial Disclosure a Consultant Abbott Medical Optics b Scientist at Abbott Medical Optics Inc. The Synchrony IOL is not currently FDA-approved

description

The Synchrony IOL is not currently FDA-approved. OPTICAL QUALITY OF DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS. 1 Fundacion Valle del Lili, Cali-Colombia 2 Universidad Industrial de Santander, Bucaramanga-Colombia 3 Universidad Autonoma de Bucaramanga, Bucaramanga-Colombia - PowerPoint PPT Presentation

Transcript of OPTICAL QUALITY OF DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

Page 1: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

1

OPTICAL QUALITY OF DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

Andrea Galvis, MD 1,3 a

Ivan Ossma, MD MPH MSc 1,2, a

Sanjeev Kasthurirangan, PhD 4, b

1 Fundacion Valle del Lili, Cali-Colombia

2 Universidad Industrial de Santander, Bucaramanga-Colombia

3 Universidad Autonoma de Bucaramanga, Bucaramanga-Colombia

4 Abbott Medical Optics Inc, Irvine, USA

Financial Disclosure

a Consultant Abbott Medical Optics

b Scientist at Abbott Medical Optics Inc.

The Synchrony IOL is not currently FDA-approved

Page 2: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

2

Synchrony Design

Anterior Optic 32 Diopters

Posterior Optic Divergent (negative)

2

Page 3: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

3

+ Compare Optical Quality Pseudophakic Eyes

+ Monofocal IOL

+ Synchrony Dual Optic Accommodating IOL

+ Parameters

+ Objective iTrace aberrometry

+ Custom Matlab Software to calculate MTF

+ Subjective Contrast Sensitivity Testing

+ Optec 6500

Study Purpose

Page 4: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

4

Study Population

iTrace aberrometry

Cycloplegia

3 mm zone

Matlab Software to calculate MTF

Monofocal IOL : 15 subjects (30 eyes)

Synchrony IOL : 16 subjects (32 eyes)

Page 5: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

5

iTrace - Matlab conversion

Page 6: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

6

Results

Page 7: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

7

Objective Measurements3 mm pupil

Difference 0.01 um

Page 8: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

8

Best-focus MTF 3 mm pupil

Differences are less than 10%

Page 9: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

9

Mesopic Contrast Sensitivity

p=0.61 Wilcoxon Rank Sum Test

9

Page 10: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

10

Dual Optic Design DOES NOT affect optical or retinal image quality of the eye

Objective evidence (iTrace, MTF)

Subjective evidence (CSF)

Contrast sensitivity of Synchrony is equivalent to that of monofocal IOLs

Discussion

Page 11: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

11Discussion

Effect of the slight spherical aberration on Synchrony would only yield 0.42 D of increased depth of focus

Calculations based on Rocha’s work *

The improved near vision of Synchrony patients MOST LIKELY due to true accommodation

* Rocha KM. Using Adaptive Optics Simultion to Explore the Effect of Higher Order Aberrations on Depth of Focus for the Correction of Presbyopia. Wavefront Congress 2008

Page 12: OPTICAL QUALITY OF  DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS

12Conclusion

Quality of Vision of Synchrony IOL similar to standard monofocal IOLs

Measured Subjectively

Measured Objectively

Improved Near Vision with Synchrony

Not related to increased depth of focus