Ophthalmic USG
-
Upload
pabita-dhungel -
Category
Health & Medicine
-
view
311 -
download
6
Transcript of Ophthalmic USG
![Page 1: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/1.jpg)
Pabita DhungelB.Optometry3rd year
ULTRASONOGRAPHY PRINCIPLE, METHODS AND INTERPRETATION
![Page 2: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/2.jpg)
PRESENTATION LAYOUT
• Introduction• History• A- scan• B- scan• Methods for special purposes• Clinical pictures• Summary
![Page 3: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/3.jpg)
References
• Clinical Procedure in Optometry• Optometry: science, techniques and Clinical
managements• Ophthalmic ultrasonography• Borish Clinical Refraction
![Page 4: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/4.jpg)
Introduction
• Ultrasound is an acoustic wave (above audible frequency) that consists of oscillations of particles within a medium, frequencies greater than 20kHz(20,000 oscillations/s)
• Diagnostic ophthalmic Ultrasound– Frequency 8-10 MHz (Sandra Frazer et. Al)
• 8-25 MHz for Posterior segment & Orbit (Jager’s Duane Oph.)
• 50 MHz for imaging Anterior segment (Jager’s Duane Oph.)
– Produce short wavelength = 0.2 mm– Good resolution of ocular structures – Less
Penetration
![Page 5: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/5.jpg)
Contd…
• Examination of larger structures e.g abdominal or obstetric ultrasound requires frequencies in the range of 1 to 5 MHz
• Such wavelengths produced by these lower frequencies enable these instruments to penetrate deeper into the body decreasing their resolution capability
![Page 6: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/6.jpg)
Contd…
• Ultrasound is transmitted as a longitudinal wave so its speed is dependent upon the density of the medium it is passing through e.g air sound travels at 340m/s whereas in water its speed is much faster at approx 1480m/s.
• Fluid contact is essential betn the transducer and eye so normal saline is used in open eye or water soluble gel is used if reading taken through eyelids
![Page 7: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/7.jpg)
Sound Wave Velocities Through Various Media
Medium Velocity (m/sec) Water 1,480
Aqueous/ Vitreous 1,532
Silicon Lens 1,486
Crystalline Lens 1,641
PMMA Lens 2,718
Silicon Oil 986
Soft Tissue 1,550
Bone 3,500
![Page 8: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/8.jpg)
History
• In 1956, • First time: Mundt and Hughes, American Oph.
• A-scan (Time Amplitude ) to demonstrate various ocular disease
• Oksala et. Al in Finland• Ultrasound Basic Principle (Pulse-Echo Technique)• Studied reflective properties of globe
• In 1958, Baum and Greenwood • Developed the first two-dimensional(immersion) (B-scan)
ultrasound instrument for ophthalmology.• In the early 1960s, Jansson and associates, in Sweden,
• used ultrasound to measure the distances between structures in the eye.
![Page 9: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/9.jpg)
Contd…
• In the 1960s, Ossoinig, an Austrian ophthalmologist– first emphasized the importance of standardizing
instrumentation and technique.– developed standardized A-scan.
• In 1972, Coleman and associates made– first commercially available immersion B -scan instrument – Refined techniques for measuring Axial length, AC depth,
Lens thickness
• Bronson in 1974 made contact B scan machine.
![Page 10: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/10.jpg)
Advantages of USG
• Easy to use.• No ionizing radiation• Excellent tissue differentiation• Cost effectiveness
Primary uses in ophthalmology:– Posterior segment evaluation in Hazy media / Orbit
• Structural integrity of eye but no functional integrity – Detection and differentiation of intraocular and orbital
lesions – Tissue thickness measurements.– Location of Intra Ocular Foreign Body– Ocular Biometry for IOL power calculations
![Page 11: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/11.jpg)
Principle
Average velocity in Eye = 1500 m/sec– Takes about 33 microseconds to travel & return
back from the posterior part of eye•Principle
– Pulse- Echo System • Emission of multiple short pulses of ultrasound
waves with brief interval to –Detect, process and display the turning
Echoes
![Page 12: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/12.jpg)
• Piezoelectric crystal (Quartz & Ceramic Crystal) – Present in Probe tip
• Application of Pulse of Voltage electric energy – Mechanical Vibration Rapid vibration
» Generate short pulse Ultrasound energy (Transducer)
• Longitudinal ultrasound wave propagate through medium (Eye) – Echoes come back from different Acoustic Interfaces
– A pause of Several Milliseconds for receiving echoes • Creates mechanical vibration as it strikes the probe tip
and the piezoelectric crystals – Produce electric energy
– Transmitted to receiver and in a Cathode ray Tube to display Ultra sonogram
![Page 13: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/13.jpg)
Terms
PROBE – Consists of piezoelectric transducer. – A piezoelectric transducer consist of small ceramic plates
• which converts electricity into sound waves and sound waves into electric signals to be displayed on the screen.
– Damping material (metal powder with plastic or epoxy) • Help to produce shorter Pulse – Better for Axial
resolution• Axial resolution
– The minimum distance between two interfaces (Echo Sources) along the direction of the sound beam
![Page 14: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/14.jpg)
Contd…
• Shape of the Crystal: – Planer crystal
• Produce relatively parallel sound beam (A- Scan)
– Acoustic lens – Produce focused sound beam (B-scan)• Improves lateral resolution
–Minimum separation between two interfaces perpendicular to the direction of sound beam
![Page 15: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/15.jpg)
Contd…
• RECEIVER (computer unit) – Receives returning echoes
• Produces electrical signal that undergoes complex processing
–Amplification, Compensation, Compression, Demodulation and Rejection
![Page 16: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/16.jpg)
Contd…
• RESOLUTION – – Ability to discern two interfaces close to each
other.– Higher the frequency of ultrasound
• Shorter the wavelength & better the resolution.
• Echoes- – Reflected portion of the wave.
![Page 17: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/17.jpg)
• GAIN – – Relative units of Ultrasound intensity
• Expressed in Decibel (db)• Adjust of gain doesn't change the amount of energy
emitted by transducer – but chance in intensity of the returning echoes for
display – Electric Amplification of the echo signals received by the
transducer • Higher the gain – Greater the sensitivity of the
Instrument in displaying weaker echoes (i.e Vitreous opacities)
• Lower the gain – Weaker the depth of sound penetration
– only stronger echoes are displayed (i.e Retina / Sclera)
![Page 18: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/18.jpg)
Gain contd…
• Stronger echoes are located in the centre of the returning sound wave – Lowering gain effectively narrows the sound beam
• Improves both axial and lateral Resolution
• When the gain is Increased – A-scan gets taller and B-scan echoes gets brighter
conversely • When the gain is turned down the echoes get
shorter and dimmer.
![Page 19: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/19.jpg)
Displaying the Ultrasound
1. A-Mode Display
2. B-Mode Display
3. M-Mode Display
![Page 20: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/20.jpg)
A - scan
• ‘A’ for amplitude provides one dimensional display of returning echoes in the form of vertical spikes of various heights and distances from the initial signal
• Echoes from the structures deeper within the eye take longer to return to the transducer for conversion back to electric signal, so appear further along the time baseline
![Page 21: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/21.jpg)
Contd…
• Two fundamental data obtained are
i) distance of echo source from the probe face
- forms the basis of biometry
ii) amplitude of echo signal (spike) which partly depends on the nature of reflecting interface
- forms the basis of quantitative echography
![Page 22: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/22.jpg)
Standardization of A scan
• Credited to Dr. Karl Ossoinig
1.Unique sound Amplification :
S-shaped amplifier with flat upper and lower curves and a steep mid segment and a dynamic range of 36dB
This amplification enhances the difference between normal and abnormal signals
![Page 23: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/23.jpg)
A-Mode DisplayA-Mode Display• Sound velocity should be adjusted
– Time Dimension calculated according to the speed at which sound travels via a given medium
– Phakic Eye – Average Sound Velocity : 1,550 m/sec
Average Velocity Adjustment for Eye Length measurement
Ocular Media Velocity (m/sec)
Aphakia (Aqueous/ Vitreous) 1,532
Phakia 1,550
Pseudophakia PMMA Implant Silicone Implant
1,532 + 0.2 mm or 1,5501,486
Silicon oil 986
Note: Average Velocity : Average of Sound Velocities for the Aqueous + Vitreous + Lens
![Page 24: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/24.jpg)
Examination steps
• Patient is positioned with head near oscilloscope
• Topical anaesthetic drops are placed in the eye
• Probe is firmly placed on the globe without coupling jelly as tear acts as coupling agent
• Eight meridians are scanned, postero-anteriorly, by shifting and tilting the probe in a single, smooth arc movement from limbus to fornix
![Page 25: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/25.jpg)
Contd…
• In cases of traumatized or infected eyes, or soon after intraocular surgery, examination through the closed eyelids is safer
• At the end of procedure, the eye is irrigated with sterile saline, and the probe tip is cleaned with alcohol wipe or other suitable disinfectant
![Page 26: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/26.jpg)
Orientation and labelling of scan
• The labelling of sections is determined by the projection of the beam and not the probe location
• E.g a section labelled 12 equator (12E) is produced by placing the probe at 6 0’clock and mid distance betn limbus and fornix, a section labelled 6 anterior (6 A) is produced by placing the probe at 12 0’clock fornix
![Page 27: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/27.jpg)
Macular screening
1) Axial section• easier of the two approaches•Probe is placed on cornea and directed axially•Suitable for measurement but not sensitive in detecting early macular thickening or in differentiation of its lesions because of strong sound attenuation by the lens
![Page 28: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/28.jpg)
Contd…
• Posterior section• In the RE this is 9P position and in the LE the
3P position• Produced by directing the patient’s gaze
temporally and placing the probe at the nasal limbus and aiming it posteriorly thus avoiding the lensand achieving better resolution
![Page 29: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/29.jpg)
Examination of fundus periphery
• Patient’s gaze is directed maximally towards the meridian to be scanned and probe is placed at the opposite fornix the beam being aimed across the globe towards the opposite periphery
• Useful for detecting peripheral retinal cysts/retino-schisis, choroidal detachments and ciliary body lesions
![Page 30: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/30.jpg)
B - Scan
• B for brightness produces two dimensional slice of tissue images, composed of coalescing dots of varying degrees of brightness , depending on the reflectivity of the echo source
• Probe emits a focused sound beam at the frequency of 10 MHz , eye dedicated scanners produce a sound beam whose focal zone coincides with the posterior globe wall and anterior orbit
![Page 31: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/31.jpg)
Contd…
• Marker at the probe tip indicates the beam orientation and the top of the echogram as it displays on the screen
• Coupling jelly is applied to the probe tip to ensure adequate sound transmission
• Probe sterilized with alcohol, impregnating B-probe in sterilizing soln isn’t recommended as it damages the transducer
![Page 32: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/32.jpg)
Fundamental objectives for high quality B- scan
• Lesions must be placed in the center of the scanning beam
• Beam must be directed perpendicularly to interfaces at the area of interest
• Lowest possible decibel gain that is consistent with the maintenance of adequate intensity should be used to optimize resolution of the images
![Page 33: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/33.jpg)
B- scan sections: orientation and labelling
• 1) Axial section• Easiest to perform and interpret• Probe is placed on the cornea and directed
axially• Posterior lens surface and optic nerve head
are placed in the center of the echogram• Optic nerve head is used as an echographic
centre
![Page 34: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/34.jpg)
Contd…
• Depending on the clock hour location of the probe marker, a vertical axial section (marker at 12 0’clock) horizontal axial scan (marker at 3 0’clock RE and 9 0’clock LE) and sections of all other clock hours can be performed
• Easy orientation and demonstration of posterior pole lesions and attachments of membranes to optic nerve head
![Page 35: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/35.jpg)
Contd…
• Because of scatter and strong sound attenuation created by the lens higher decibel gain levels are needed to show structures at the posterior segment
• In pseudophakia strong artefacts created by the lens implant will hamper the adequate visualization
![Page 36: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/36.jpg)
Contd…
2) Transverse section•Probe is placed on the limbus and directed posteriorly •Echograms are labelled according to clock hour at the center of the beam and also to the beam’s anteroposterior location•E.g section labelled 12 posterior (12P) is produced by a probe located at 6 0’clock limbus, a 6 0’clock (6E) by probe located at 12 0’clock
![Page 37: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/37.jpg)
Contd…
3) Longitudinal section•Section produces an antero-posterior slice of the ocular wall along one meridian only; from the optic nerve (lower portion of echogram) to the ciliary body (upper portion of the echogram)•Probe is located on sclera with marker at its limbal side
![Page 38: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/38.jpg)
Contd…
• Echograms are labelled after the clock-hour location of the beam
• E.g L-3 is a section produced by a probe at 9 0’clock limbus and L-12 by a probe at 6 0’ clock limbus
![Page 39: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/39.jpg)
Steps for ocular screening
1) Transverse 12:
Patient directs gaze superiorly
Probe is placed at the 6 0’clock limbus with its marker nasally
Shifting and rotating the probe from limbus to fornix scans the superior retina postero-anteriorly
![Page 40: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/40.jpg)
Contd…
2) Transverse 3:•The patient gazes at 3 0’ clock and the probe is placed at 9 0’clock with its marker up•Probe is manoeuvred from the limbus to fornix, scanning the nasal retina in the RE and temporal retina in the LE
![Page 41: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/41.jpg)
Contd…
3) Transverse 6:•The patient looks downwards and the probe is positioned at 12 0’clock with the marker nasally and swept from limbus to fornix to scan inferior retina
4) Transverse 9:•The patient gazes at 9 0’clock and the probe is manoeuvred at 3 0’clock with its marker kept up, section scans the temporal retina in the RE and nasal retina in the LE
![Page 42: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/42.jpg)
Contd…
• Four additional transverse scans of the oblique quadrants i.e 1:30 , 4:30, 7:30 and 10:30 0’ clock hours may be performed in a similar fashion if abnormalities in the oblique periphery are suspected
![Page 43: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/43.jpg)
Macular screening
1) horizontal axial • an axial section with the marker nasally will display the macular area and adjcent optic nerve head
2) Vertical macula•Probe is tilted temporally without losing the posterior lens echoes •Vertical beam is shifted from optic nerve to the macular area, lens acts as a reference point for accurate placement and future reference
![Page 44: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/44.jpg)
Contd…
3) Transverse 9:00 RE and 3:00 LE•The probe is placed on the corresponding nasal limbus with its marker up, and the patient gazes temporally•Avoiding the lens allows better visualization of the vertical extent of macular masses
![Page 45: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/45.jpg)
Contd…
• 4) Longitudinal 9:00 RE and 3:00 LE• Patient’s gaze is directed temporally• Probe is placed on the nasal side of the globe
with the marker at the limbus• Macular area will appear at the centre of the
echogram with the optic nerve at the bottom and ciliary body at the top
• Lateral extension of lesion is studied in this section
![Page 46: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/46.jpg)
M- Mode Display
• Motion Mode & Time-Motion Mode • Used to monitor Integrity of Blood Vessels/ Lens
Accommodation Dynamic • Aka Color Doppler Imaging (CDI), the technique
provides information about the ocular vascular supply superimposed on a B-scan image
![Page 47: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/47.jpg)
Contd…
• This modality gives approx flow velocity of the principal vessels in the eye and orbit, by providing a color doppler display for pulse doppler examination (Lieb et al 1991; Giovagnorio et al 1993)
• CDI scans of the eye and orbit are performed with patient lying in supine position with their eye closed using an ultrasound frequency > 7 MHz and a Doppler frequency >5 MHz (Tranquart et al 2003)
![Page 48: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/48.jpg)
Contd…
• The transducer or probe is applied to closed eyelid using a thick layer of conducting gel
• Care must be taken not to exert any pressure on the globe as this might result in a decrease in blood flow velocity, thus increasing the RI with no diagnostic significance
![Page 49: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/49.jpg)
FACTOR INFLUENCING HEIGHT IN A-SCAN AND BRIGHTNESS IN B-SCANFACTOR INFLUENCING HEIGHT IN A-SCAN AND BRIGHTNESS IN B-SCAN
1. Angle of the sound beam
1. Interface
2. Size and shape of interfaces
![Page 50: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/50.jpg)
• When the sound beam is directed perpendicularly to a structure
– Maximum amount of sound will be reflected back to the probe.
The farther away from the ideal angle , the lower the amplitude.
1. ANGLE OF INCIDENCE
![Page 51: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/51.jpg)
• Depends upon the difference between Acoustic Impedance
– Greater the diff. AI stronger the Reflected Echoes
• EG:
– Anterior lens surface produce strong echo when bordered by Aqueous than by Blood
– Interface between vitreous and fresh blood is very slight resulting in small echo.
– The difference between a detached retina and the vitreous is great producing a large echo
2. INTERFACE
![Page 52: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/52.jpg)
Smooth surface like retina will give strong reflection.
Smooth and rounded surface scatter the beam.
Coarse surface like ciliary body or membrane with folds tend to scatter the beam without any single strong reflection.
Small interface produces scattering of reflection.
3. TEXTURE AND SIZE OF INTERFACE
![Page 53: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/53.jpg)
Indications for Intraocular Examination
Opaque Ocular MediaAnterior Segment
Corneal opacificationHyphema /hypopyonCataractPupillary or
retrolenticular membrane
Posterior Segment Vitreous hemorrhage or
inflammation
Clear Ocular Media
Anterior segment
Iris lesions
Ciliary body lesionsPosterior segment
Tumors
Choroidal detachment
Retinal detachment
Optic disc abnormalities-Intraocular foreign bodies
•
BiometryBiometry
![Page 54: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/54.jpg)
� Indications-Signs & Symptoms-• Unilateral or Bilateral
Exophthalmos• Enophthalmos• Globe Displacement• Lid Abnormalities - Ptosis,
Retraction, Swelling, Ecchymosis• Palpable or Visible Masses• Chemosis• Motility Disturbances;
Diplopia• Pain
Additional Indications-
• Tissue Differentiation of Mass Lesions
• Clarification of CT or MRI Findings
• Assessment of blood flow within lesions
• Follow-up studies
Intraorbital ExaminationIntraorbital Examination
![Page 55: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/55.jpg)
Examination Techniques For The Globe:
• Positioning the patient
• Topical anesthesia
• Techniques
– Contact Technique• Probe - placed directly on the globe
– Immersion Technique
• Methylcellulose - a coupling medium (B-Scan)
![Page 56: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/56.jpg)
Contact Technique• The Probe Placed
– Directly on the cornea (A-Scan) or over closed lids (B- Scan)
– To evaluate the posterior segment only (B Scan)
Small solid probe fits in the Tonometer holder for A-scan biometry.
Small solid probe fits in the Tonometer holder for A-scan biometry.
![Page 57: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/57.jpg)
Applanation Technique
![Page 58: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/58.jpg)
Contact Technique A Scan – Phakic Eye
![Page 59: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/59.jpg)
Contact Technique A- Scan - Aphakia
![Page 60: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/60.jpg)
Contact Technique
• Sources of Error – Corneal compression
(Shorter Axial length) • 1mm error in Axial length –
2.5 to 3.0 Ds error in IOL Power
– Misalignment of sound beam
– Fluid Miniscus Trapped• Erroneously Long AL
![Page 61: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/61.jpg)
• Correct– ( no compression )
No Fluid meniscus trapped between probe tip and cornea.
A- Scan
![Page 62: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/62.jpg)
Immersion Technique
• Can use in the same instrument – Requires Scleral Cup
• Coupling Agent – Methylcellulose • Probe is not directly placed on the
cornea – immersed into the fluid
• Error – Small air bubbles in the fluid
gives falsely long AL measurement
Prager Scleral Shell
![Page 63: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/63.jpg)
Scleral Shells
The Hansen scleral shells available in 16, 18, 20, 22, and 24 mm
The Hansen scleral shells available in 16, 18, 20, 22, and 24 mm
The Prager Shell The Prager Shell
The Kohn shell The Kohn shell
![Page 64: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/64.jpg)
Immersion Technique
The cup is placed between the lids and methylcellulose 1% is poured into the cup.
The ultrasound probe is immersed in the solution, keeping it 5 to 10 mm away from the cornea.
![Page 65: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/65.jpg)
A-Scan (Immersion Biometry)
Initial spike (IS), the anterior (C1) and posterior (C2) corneal surfaces, the anterior (L1) and posterior (L2) lens surfaces, the retina (R), sclera (S), and orbital tissues (O).
Initial spike (IS), the anterior (C1) and posterior (C2) corneal surfaces, the anterior (L1) and posterior (L2) lens surfaces, the retina (R), sclera (S), and orbital tissues (O).
![Page 66: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/66.jpg)
Immersion Technique: Advantages
• Important in patients with shorter axial length < 22.00 mm – no corneal compression
• Where small error in measurement cause significant error in Biometry
• Can examine the cornea, anterior chamber, iris, lens & retrolental space
• Ultrasound Biomicroscopy:– developed by Pavlin & colleagues– uses sound wave of 50 to 100 MHz
![Page 67: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/67.jpg)
Immersion Technique• Should be cautious in eyes with recent intraocular
surgery or penetrating trauma.
• No corneal compression
• No problem of fluid meniscus
• The display of a separate corneal spike makes it easier to determine when the sound beam is properly aligned along the optical axis.
![Page 68: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/68.jpg)
Examination Techniques For The Globe-A scan
• The probe should remain perpendicular to the structure studied– The probe is first adjusted to the Tissue Sensitivity gain setting.
• One dimensional acoustic display-echoes-vertical spikes from baseline
Uses•For IOL calculations (biometry)•For differentiating and monitoring certain ocular conditions • Congenital glaucoma• Progressive Myopia
Uses•For IOL calculations (biometry)•For differentiating and monitoring certain ocular conditions • Congenital glaucoma• Progressive Myopia
![Page 69: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/69.jpg)
Axial Eye Length Measurements
• Two techniques: – The contact method– The immersion method
• In both techniques, the sound beam must be directed along the optical axis of the eye.
![Page 70: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/70.jpg)
B scan
• Different types of Probe • The probe face is Oval
Shaped with the transducer movement in the direction of longest diameter
• The probe contains transducer that moves back and forth near the tip
• Fig of probe face
![Page 71: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/71.jpg)
B scan
• Each probe has a marker – A dot / Line – Indicates the side of
the probe that is represented on the upper portion of B-Scan Display
![Page 72: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/72.jpg)
• Basic Screening Examination– transverse scans of the four major
quadrants at a high gain setting, from limbus to fornix
– first superior portion → nasal portion → inferior portion → temporal portion
Examination Techniques For The Globe:..
![Page 73: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/73.jpg)
Examination Techniques For The Globe-B scan
• After transverse scan – Eye should be scanned with longitudinal scan at
least in four major quadrants– Helpful in detecting lesions in peripapillary region and macula
• Globe should also be evaluated with vertical and horizontal axial scan
![Page 74: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/74.jpg)
A-scan basic screening examination
![Page 75: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/75.jpg)
Special Examination Techniques
• Topographic– Location– Extension– Shape
• Quantitative– Reflectivity estimate: spike height– Internal structure: histologic architecture– Sound attenuation: absorption, shadowing
• Kinetic– Mobility: after movement– Vascularity: blood flow
![Page 76: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/76.jpg)
Topographic Echography
A. point-like
e.g. fresh V.H
B. membrane-like
e.g. R.D
C. mass-like
e.g. choroidal melanoma
![Page 77: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/77.jpg)
Quantitative Echography
1-Reflectivity estimate:
PVD Vs RD– PVD produces 50%
high spike– RD produces 100%
high spike
![Page 78: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/78.jpg)
Quantitative ..internal structure
• Choroidal melanoma-homogenous histologically-regular internal stucture,low reflectivity
• Choroidal hemangioma-multiple vascular spaces-regular internal structure high reflectivity
• Metastatic carcinoma-irregular arrangement of tumor cells,variable reflectivity
![Page 79: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/79.jpg)
Kinetic Echography• Lesion mobility (after
movement)– Non-solid lesion (e.g.
vitreous membrane, RD) displays after movement.
– Solid lesion (e.g. tumor) does not.
Blurred appearance of spikes
![Page 80: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/80.jpg)
Kinetic…• Vascularity (spontaneous motion)
– Vascularity is a characteristic that is assessed in tumors spontaneous motion (low-amplitude flickering of the internal lesion spikes)
![Page 81: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/81.jpg)
• Three major portions:– orbital soft tissue assessment
– extraocular muscle evaluation
– retrobulbar optic nerve
examination
• Two approaches:– Transocular (through the globe)
• for lesions located within the posterior & mid-aspects of the orbital cavity
– Paraocular (next to the globe)• for lesions located within the lids or anterior orbit
Examination Techniques For The Orbit:
![Page 82: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/82.jpg)
• Basic Screening Examination– Mainly Transocular approach
– Transverse scans of the four major meridians using a medium-high gain setting• From limbus to fornix
– First superior portion → nasal portion → inferior portion → temporal portion
Examination Techniques For The Orbit:…
![Page 83: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/83.jpg)
Clinical Examples:
Order of disappearance with gain reduction : Vitreous Opacities –Blood- Retinal Tissue – Ca++Order of disappearance with gain reduction : Vitreous Opacities –Blood- Retinal Tissue – Ca++
![Page 84: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/84.jpg)
Vitreoretinal Disease
• Mild and extremely dense Asteroid hyalosis• Dense Vitreous hemorrhage
![Page 85: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/85.jpg)
Vitreous Haemorrhage
• US showing hemorrhagic track through vitreous cavity along path of penetrating nail injury
![Page 86: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/86.jpg)
Posterior Scleritis
• High reflective thickening of retinochoroid layer and sclera
![Page 87: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/87.jpg)
RD Vs PVDRD Vs PVD
Choroidal Detachment Choroidal Detachment
![Page 88: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/88.jpg)
Posterior Vitreous Detachment
• Low reflective vitreous opacities and a posterior vitreous detachment as seen with normal aging of the eye.
![Page 89: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/89.jpg)
Tractional RD
• US showing tent shaped tractional RD (Large arrow). Small arrows show vitreous membranes of low reflecting attached to apex of detached retina.
![Page 90: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/90.jpg)
Dislocated Lens:
![Page 91: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/91.jpg)
Choroidal Detachment
B-scan of eye with “kissing” Choroidal detachment & Haemorrhage
![Page 92: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/92.jpg)
Subarachnoid Fluid Around the Optic Nerve
Positive crescent sign.
![Page 93: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/93.jpg)
Ciliary Body Detachment (Immersion Technique)
Ciliary body detachment as seen on high-resolution scan. Note the large cleft in the subciliary space
![Page 94: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/94.jpg)
Choroidal Melanoma:
![Page 95: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/95.jpg)
Thyroid Ophthalmopathy Vs Myositis
• Thickened muscles due to thyroid ophthalmopathy.
• Orbital myositis
Thickened muscle
to be irregular and
medium-high reflective
in thyroid but regular and
low reflective in myositis.
![Page 96: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/96.jpg)
Posterior Staphylomas
![Page 97: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/97.jpg)
![Page 98: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/98.jpg)
.
Positive T-sign at the insertion of the optic nervePositive T-sign at the insertion of the optic nerve
Nodular posterior scleritis with fluid in the Tenon capsuleNodular posterior scleritis with fluid in the Tenon capsule
ScleritisScleritis
![Page 99: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/99.jpg)
AC – Hyphema (Immersion Technique)
US showing anterior chamber hyphema.
C represents cornea, white arrow - hyphema and L - Lens.
![Page 100: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/100.jpg)
Metallic Foreign Body
• US of eye demonstrated a metal F.B. (arrow) which is highly reflective. This can produce a region that may look like a nerve shadow or even resemble a dehiscence in sclera.
![Page 101: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/101.jpg)
Metallic IOFB
• Thick, very highly reflective spike from foreign body(F) and decreased spike height of sclera and orbital tissue as a result of sound attenuation.
.
![Page 102: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/102.jpg)
Retinoblastoma
• Non-calcified RB• Calcified RB
![Page 103: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/103.jpg)
Extrusion of Lens Material
• B-scan of eye with severe blunt trauma showing extrusion of lens material through small posterior lens rupture (arrow).
![Page 104: Ophthalmic USG](https://reader030.fdocuments.in/reader030/viewer/2022020123/55a7290d1a28ab885e8b4719/html5/thumbnails/104.jpg)
S
THANK YOU