OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks...

58
AUTOMATIC LEVEL OPERATOR'S MANUAL 31830 90073

Transcript of OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks...

Page 1: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

AUTOMATIC LEVELOPERATOR'S MANUAL

31830 90073

Page 2: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

This is the mark of the Japan Surveying Instruments Manufacturers Association.

Page 3: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

FOREWORD

Thank you for selecting the TOPCON instrument.• Please read this operator’s manual carefully

before using this instrument.• Verify that all equipment is included.

“ STANDARD SYSTEM COMPONENTS”(p. v)

• The specifications and general appearance ofthe instrument are subject to change withoutprior notice and without obligation by TopconCorporation and may differ from thoseappearing in this manual.

• Some of the diagrams in this manual may besimplified for easier understanding.

i

Page 4: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

HOW TO READ THIS MANUAL

Symbols

The following conventions are used in this manual.

Trademark

All other company and product names featured in this manual are trademarks or registered trademarks of each respective organization.

: Indicates precautions and important items which should be read before operations.

: Indicates the chapter title to refer to for additional information.

: Indicates supplementary explanation.

[SET] : Indicates operating keys.

ii

Page 5: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

CONTENTS

FOREWORD ......................................................... iHOW TO READ THIS MANUAL........................... iiCONTENTS ......................................................... iiiSTANDARD SYSTEM COMPONENTS ............... v1. PRECAUTIONS FOR SAFE OPERATION...... 12. PRECAUTIONS............................................... 53. NOMENCLATURE AND FUNCTIONS ............ 8

3.1 Names of Parts ........................................ 83.2 Display ................................................... 10

Battery level ....................................... 103.3 Operating Keys ...................................... 11

4. PREPARATION............................................. 124.1 Installing Batteries................................. 124.2 Setting Up the Instrument ...................... 134.3 Setting Up the Staff ................................ 14

Reading the staff ................................ 14Setting Up the Staff ............................ 15

4.4 Focusing and Sighting............................ 16Before using the instrument ............... 16Focusing when taking measurements .... 17

4.5 Power ON............................................... 184.6 Modes .................................................... 19

5. MEASUREMENT [MEASUREMENT MODE] .... 205.1 Height / Distance.................................... 20

Repeat measurement for height and distance.............................................. 20

5.2 Height Difference ................................... 21 Repeat measured distance of height

iii

Page 6: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

difference ........................................... 22Display of Height Difference............... 24Measurement from the ceiling............ 25

5.3 Using the Instrument as an Automatic Level (When not using digital reading) ............ 25

6. SETTING UP [SETTING MODE]................... 266.1 Setting Up Display Unit .......................... 266.2 Setting Up Automatic Power OFF.......... 276.3 Setting Up Single / Repeat Measurement..... 29

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE].................................. 317.1 Adjusting the Circular Level ................... 317.2 Adjusting the Cross-line ......................... 32

Soft adjustment .................................. 32Mechanical adjustment ...................... 37

8. ERROR.......................................................... 39What to do when an error message appears .............................................. 39

9. STORAGE IN THE CASE ............................. 40Diagram of storage case.................... 40

10.STAFF TYPES .............................................. 4111.REGULATION ............................................... 4212.SPECIFICATIONS......................................... 4413.ERROR CHART ............................................ 47

iv

Page 7: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

STANDARD SYSTEM COMPONENTS

1) ORION+ instrument........................................... 12) AA-size dry cell batteries .................................. 23) Barcode tape staff ............................................. 14) Hexagonal wrench ........................................... 15) Vinyl cover ........................................................ 16) Cleaning cloth ................................................... 17) Carrying case ................................................... 18) Operator’s manual ............................................ 19) Warranty card.................................................... 1

• Please make sure that all of above items are in thebox when you unpack.

v

Page 8: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

vi

Page 9: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

1. PRECAUTIONS FOR SAFE OPERATION

For the safe use of the product and prevention of injury to operators and other persons as well as prevention of property damage, items which should be observed are indicated in this operator’s manual by WARNING and CAUTION indications.The definitions of the indications are listed below. Be sure you understand them before reading the manual's main text.

Definition of Indication

WARNING

Ignoring this indication and making an operation error could possibly result in death or serious injury to the operator.

CAUTION

Ignoring this indication and making an operation error could possibly result in minor injury or property damage.

This symbol indicates items for which caution (hazard warnings inclusive) is urged. Specific details are printed in or near the symbol.

This symbol indicates items which are prohibited. Specific details are printed in or near the symbol.

This symbol indicates items which must always be performed. Specific details are printed in or near the symbol.

1

Page 10: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

1. PRECAUTIONS FOR SAFE OPERATION

GENERAL

WARNING

Do not perform disassembly or rebuilding. Fire, electric shock or burns could result.

Do not use the unit in areas exposed to high amounts of dust or ash, in areas where there is inadequate ventilation, or near combustible materials. An explosion could occur.

Never look at the sun through the telescope. Loss of eyesight could result.

Do not look at reflected sunlight from a prism or other reflecting object through the telescope. Loss of eyesight could result.

When securing the instrument in the carrying case make sure that all catches, including the side catches, are closed. Failure to do so could result in the instrument falling out while being carried, causing injury.

CAUTION

Do not use the carrying case as a footstool. The case is slippery and unstable so a person could slip and fall off it.

Do not place the instrument in a case with a damaged case or belt. The case or instrument could be dropped and cause injury.

2

Page 11: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

1. PRECAUTIONS FOR SAFE OPERATION

TRIPOD

CAUTION

When mounting the instrument to the tripod, tighten the centring screw securely. Failure to tighten the screw properly could result in the instrument falling off the tripod causing injury.

Tighten securely the leg fixing screws of the tripod on which the instrument is mounted. Failure to tighten the screws could result in the tripod collapsing, causing injury.

Do not carry the tripod with the tripod shoes pointed at other persons. A person could be injured if struck by the tripod shoes.

Keep hands and feet away from the tripod shoes when fixing the tripod in the ground. A hand or foot stab wound could result.

Tighten the leg fixing screws securely before carrying the tripod. Failure to tighten the screws could lead to the tripod legs extending, causing injury.

3

Page 12: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

1. PRECAUTIONS FOR SAFE OPERATION

STAFF

Power Supply

WARNING

Do not use under thunderous weather conditions. Staff is conductive and if struck by lightning, death or injury could result.

Handle with care when using near high voltage cables or transformers. Staff is conductive and contact could result in electric shock.

WARNING

Do not heat or throw batteries into fire. An explosion could occur, resulting in injury.

CAUTION

Do not touch liquid leaking from batteries. Harmful chemicals could cause burns or blisters.

4

Page 13: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

2. PRECAUTIONS

GENERAL

• The ORION+ is a precision instrument. Avoid severe shocks or vibration.

• Be careful when removing the instrument from its case.• Do not place the instrument directly on the ground.• Never carry the instrument on the tripod to another site.• Always turn the instrument off and remove the battery

before storing the instrument in its case.• When the instrument is placed in the carrying case, see

"Diagram of storage case". 9. STORAGE IN THE CASE (p. 40)

• Consult your local dealer before using the instrument under special conditions such as long periods of continuous use or high levels of humidity. In general, special conditions are treated as being outside the scope of the product warranty.

MAINTENANCE

• Wipe off moisture completely if the instrument gets wet during survey work.

• Always clean the instrument before returning it to the case. The lens requires special care. Dust it off with the lens brush first, to remove tiny particles. Then, after providing a little condensation by breathing on the lens, wipe it with the supplied cleaning cloth or lens tissue.

• If the display is dirty, carefully wipe it with a soft, dry cloth. To clean other parts of the instrument or the carrying case, lightly moisten a soft cloth in a mild detergent solution. Wring out excess water until the cloth is slightly damp, then carefully wipe the surface of the unit. Do not use any alkaline cleaning solutions,

5

Page 14: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

2. PRECAUTIONS

alcohol, or any other organic solvents on the instrument or display.

• Store the instrument and accessories in a dry room where the temperature remains fairly constant.

• If any trouble is found with the screws or optical parts (e.g. lens), contact your local dealer.

• Always close the case when empty to protect the interior from humidity and dust.

• Regular checking and adjustment is recommended to maintain instrument precision.

EXCEPTIONS FROM RESPONSIBILITY

• The user of this product is expected to follow all operating instructions and make periodic checks (hardware only) of the product’s performance.

• The manufacturer, or its representatives, assumes no responsibility for results of faulty or intentional usage or misuse including any direct, indirect, consequential damage, or loss of profits.

• The manufacturer, or its representatives, assumes no responsibility for consequential damage, or loss of profits due to any natural disaster, (earthquake, storms, floods etc.), fire, accident, or an act of a third party and/or usage under unusual conditions.

• The manufacturer, or its representatives, assumes no responsibility for any damage (change of data, loss of data, loss of profits, an interruption of business etc.) caused by use of the product or an unusable product.

• The manufacturer, or its representatives, assumes no responsibility for any damage, and loss of profits caused by usage different to that explained in the operator’s manual.

6

Page 15: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

2. PRECAUTIONS

• The manufacturer, or its representatives, assumes no responsibility for damage caused by incorrect operation, or action resulting from connecting to other products.

7

Page 16: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

3. NOMENCLATURE AND FUNCTIONS

1 Circular level2 Objective lens3 Notch sight

Use for coarse adjustment of the orientation of the instrument.

4 Battery cover5 Horizontal circle6 Horizontal circle positioning ring

You can rotate the horizontal scale while the instrument is fixed in position.Use to align benchmarks with 0°, 90, 180, 270,etc.

7 Reticle adjusting screw and screw coverUse this screw to mechanically adjust the reticle.

3.1 Names of Parts

78

9

10

11

12

13

14

15

1

2

3

4

56

8

Page 17: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

3. NOMENCLATURE AND FUNCTIONS

8 Keyboard3.3 Operating Keys (p. 11)

9 EyepieceAdjust the reticle focus to suit your eyesight.

10 DisplayDisplay the menu screen or measured values.

3.1 Names of Parts (p. 8)11 Focusing knob

Use this knob to focus on the staff.12 Measure key

(Appears as [MEAS] in this manual.)3.3 Operating Keys (p. 11)

13 Horizontal fine motion knobsUse these knobs to finely adjust the instrument's horizontal position.

14 Leveling foot screw15 Base plate

9

Page 18: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

3. NOMENCLATURE AND FUNCTIONS

Battery level

Battery level display shows the remaining in the battery.Short push the key to display battery level while the power is ON.

3.2 Display

Height / Distance measurement

Height difference measurementSetting Mode

Check and adjustment modeBattery level

Staff level display (To raise)

Staff level display (To lower)

With display: Repeat measurementWithout display: Single measurement

Height:1.968mDistance: 11.13 m

Unit display6.1 Setting Up

Display Unit (p. 26)

Ample power for operation

Ample power for operation

Power remaining for operation

Near power depletion. Please replace with new batteries.

10

Page 19: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

3. NOMENCLATURE AND FUNCTIONS

3.3 Operating Keys

Key Display or Name FunctionLong push: Turns the power ON /

OFF.Short push: Displays battery level

Switches the display backlight ON/OFF

[MENU] • Selects operations• Recovers from an error screen

[H/Dif]

In the measurement mode:Toggles distance/height measurement and height difference measurement

In the setting mode:Returns to the previous screen

[SET]• Toggles the measurement mode

and setting mode• Confirms set items

[MEAS] Starts or stops measurement (Side of the unit)

11

Page 20: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

*1 Always replace both batteries at the same time.Do not mix old and new batteries.

*2 Use alkali batteries. (Batteries for operation check are included in the package.) It is possible to use nickel cadmium or nickel hydride batteries; however, service life will differ.

*3 Generally, battery performance will weaken at low temperature. However, this is only temporary and the performance will recover when the temperature rises to the normal operational range.

4.1 Installing Batteries

1 Push down the section of the battery cover specified in the diagram at the left.

2 Open the cover.

3 Remove the cover.

4 Install batteries.*1), *2), *3)

5 Close the battery cover. Push the battery cover until you hear a click.

12

Page 21: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

1 Set up the tripod.Spread the tripod legs about the same distance apart so that the tripod head is approximately level. Tread the tripod shoes firmly into the ground.

2 Mount the instrument on the tripod.Hold the instrument on the tripod head and tighten the centering screw.

3 Level the instrument.

The compensator's operational range is ±12' however, an error message will not be displayed when the range is exceeded. Please make sure that the leveling operation is performed accurately.

At a correctly adjusted level, horizontal level is maintained by the automatic correction mechanism provided that the bubble is within the circle.

4.2 Setting Up the Instrument

Spherical head tripod:Loosen the centering screw and slide the instrument across the tripod head until the bubble is centered in the circular level. Tighten the centering screw.

Flat head tripod: Extend or push in the tripod legs until the bubble is centered in the circular level. When the bubble is more or less centered, turn the leveling foot screws until the bubble is exactly centered in the circle.When you turn a screw clockwise, raising the instrument, the bubble will move toward the position of that screw.

13

Page 22: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

Simply focus on the RAB code and press the [MEAS] key for an automatic reading of the staff. The following explains how to read the RAB code of the staff.

• RAB code (Random Bi-directional code) is dedicatedly used for our own automatic levels.

• The barcode tape staff is a coded staff dedicatedly used with ORION+.

Reading the staff

• Set up the staff in an area free of obstacles.• Avoid placing the staff next to mirror-like surfaces. The

effect of strong light could make measurement impossible.

• Support the staff so that it is perpendicular, checking the circular bubble scope on the staff. If the staff is tilting, height and distance measurements will be incorrect.

• If the surface is catching the light, turn it just enough to stop the reflection.

• Make sure shadows are not cast on the staff as this could make measurement impossible.

4.3 Setting Up the Staff

Barcode tape staff

Turn a few degrees to the left or right.

RAB code side of the staff

14

Page 23: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

• When holding the staff during measurement, make sure your hand does not cover the RAB code.

• If dark locations make measurement difficult, shine a flashlight on the staff. Stand at a distance so that the beam shines evenly over the length of the staff.

• If waterdrops or dirt adhere to the barcode, measurement may be impossible.Wipe the staff clean with a soft cloth.

• Clean the staff if grit or dirt gets stuck between the sections. If the staff is dirty, measurements will not be very accurate.

• Avoid scratching or soiling the barcode surface as this could make measurement impossible. Store and carry the staff inside its case.

• If direct sunlight or strong light enters the eyepiece during operation, measurement may not be performed. ([ 1 ] (Measurement error) or [ 4 ] (Too bright) will be displayed) 8. ERROR (p. 39)Shield the eyepiece from the light source using your body or cover the eyepiece with your hand. Measurement can be performed when the eyepiece is shielded.

• If the equipment is subjected to shock or vibration during use, measurement may not be possible. Resume measurement under more stable conditions.

Setting Up the Staff

1 Connect the staffs correctly by making sure the numbers on the numeric scale side of the staffs run in unbroken order.

2 Set the foot plate on the ground so that the staff will not sink.

3 Keeping an eye on the circular bubble scope on the staff, set the staff in the foot plate so that it stands up straight.

4 Turn the barcode scale toward the instrument.

15

Page 24: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

Before using the instrument

Turn the eyepiece to adjust the eyepiece image.

1 Using the gunsight, aim the objective lens toward the staff.

2 Gradually turn the eyepiece outward, stopping just before the reticle cross-lines become blurred.

3 Turn the horizontal fine motion knobs until the staff is nearly centered in the field of view, then turn the focusing knob to focus on the staff.

4 Looking through the telescope, shift your eyes slightly up and down and to each side.

5 If the staff and reticle show no deviation, the instrument is ready for use.If the staff and reticle deviate, return to step 2.

4.4 Focusing and Sighting

16

Page 25: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

If the cross-line blurs, or deviation occurs at step 5, errors will occur when taking measurements. Always make sure the instrument is correctly focused.

Focusing when taking measurements

1 Using the gunsight, aim the objective lens toward the staff.

2 Turn the horizontal fine motion knobs until the staff is nearly centered in the field of view, then turn the focusing knob to focus on the staff.

If the staff barcodes are out of focus, they will not be readable and measurements cannot be taken. Make sure they are correctly focused.

17

Page 26: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

1 Press the power switch.

Height / Distance Measurement screen will be displayed.The previously measured height and distance will be displayed.

At the time of initial use, height and distance will not be displayed.

4.5 Power ON

18

Page 27: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

4. PREPARATION

Procedures to change modes are as follows:

Method for Version Display:Press the [SET] key while in the measurement mode.

4.6 Modes

Power ON

Measurement Mode Setting Mode

Height / Distance measurement Display unit setting

Height difference measurementAutomatic power OFF setting

Single / Repeat measurement setting

Version display

Software VersionPrinted Circuit Board Number

19

Page 28: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

5. MEASUREMENT [MEASUREMENT MODE]

In the status mode you can sight point A, take the staff reading (h), and measure the horizontal distance (d) to the staff.

There are two measurement methods: single and repeat. Below is the method for repeat measurement, as an example.

Repeat measurement for height and distance

Setting up Single/Repeat Measurement6.3 Setting Up Single / Repeat Measurement

(p. 29)This setting will also apply to the height/distance and height difference.

1 Select Height / Distance measurement.4.5 Power ON (p. 18)

2 Collimate the staff A.

5.1 Height / Distance

h

d

20

Page 29: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

5. MEASUREMENT [MEASUREMENT MODE]

3 Press the [MEAS] key. Repeat measurement will begin.

Height (h) and distance (d) will be displayed with each measurement.

4 To end measurement, press the [MEAS] key again.

Measurement will end and the last values will be displayed.

This function measures the difference in height between the standard point (Point A) and measured point (Point B).At this time, the up and down directions to set the standard point are shown with directional marks. Move the staff to measure the Point B at the same height as the standard point A.

5.2 Height Difference

21

Page 30: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

5. MEASUREMENT [MEASUREMENT MODE]

There are two methods of measurements: single and repeat. Below is the method for repeat measurement, as an example.

Repeat measured distance of height difference

Setting up Single/Repeat Measurement4.5 Power ON (p. 18)

• Measure height / distance for standard point A.1 Select Height / Distance measurement.

4.5 Power ON (p. 18)2 Set up the staff at the standard point A.3 Collimate the staff and press the [MEAS] key.

Repeat measurement will begin.

4 Press the [MEAS] key.The measured value of the standard point A is determined.

22

Page 31: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

5. MEASUREMENT [MEASUREMENT MODE]

• Continue height difference measurement5 Press the [H/Dif] key.

At this time, the measured value in the procedure 4 (EX: 1.312m) will be stored in the unit as the standard value for the height difference.

6 Set up the staff at the standard point A.7 Collimate the staff.8 Press the [MEAS] key.

Repeat measurement of height difference will begin.

The results are as follows:H: Standard point: 1.312 md: Difference between the standard and measured points: 0.026 mStaff position: It is higher than the standard point. Lower the staff.

Height difference being measured

23

Page 32: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

5. MEASUREMENT [MEASUREMENT MODE]

9 Using the directional marks set the staff so the height difference (d) to the standard point is 0.

10 To end measurement, press the [MEAS] key again.

Display of Height Difference

The directional marks on the right screen show the direction of movement of the staff to the standard point.

Height of standard point(1.312 m)

26 mm higher than the standard point, so raise26 so raise

38 mm higher than the standard point, so raise38 so lower

24

Page 33: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

5. MEASUREMENT [MEASUREMENT MODE]

Measurement from the ceiling

To measure the height from the ceiling or from under the joint, turn the staff upside down vertically.

By using the numeric scale side of the staff, you can use the ORION+ as an automatic level (When not using digital reading). Simply focus on the staff and read the scale.

5.3 Using the Instrument as an Automatic Level (When not using digital reading)

Height of standard point (1.312 m)

26 mm higher than the standard point, so raise26 so lower

38 mm higher than the standard point, so raise38 so raise

25

Page 34: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

6. SETTING UP [SETTING MODE]

1 Turn ON the power.The measurement mode is displayed.

2 Press the [MENU] key.The screen will switch to the setting mode and the display unit setting screen (dISPLAy UnIT) will be displayed.

3 Press the [SET] key.The current display unit, which is meters, will be displayed. Press the [MENU] key to change to feet.

4 Press the [MENU] key.Feet /inch units will be displayed.

5 Press the [SET] key to complete.

6 Press the [H/Dif] key to return to the original screen.

6.1 Setting Up Display Unit

26

Page 35: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

6. SETTING UP [SETTING MODE]

Please refer to the following table for display unit and accuracy.

1 Turn ON the power.The measurement mode is displayed.

Display unit

Display value

Accuracy in measurement/ height difference mode

Accuracy in check/adjustment mode

m(meter)

Reading/Height difference

1mm 0.1mm

Horizontal distance

1cm 1cm

ft(feet)

Reading/Height difference

0.01ft 0.001ft

Horizontal distance

0.02ft 0.02ft

f.i(feet/inch)

Reading/Height difference

1/8inch 1/8inch

Horizontal distance

1/8inch 1/8inch

6.2 Setting Up Automatic Power OFF

27

Page 36: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

6. SETTING UP [SETTING MODE]

2 Press the [MENU] key.3 Press the [MENU] key.

The automatic power OFF setting screen (AUTO P-OFF) will be displayed.

4 Press the [SET] key.The current setting will be displayed.On: The power will turn off automatically

approximately 30 minutes after the last key operation.

OFF: Stops the automatic power OFF function.

5 Press the [MENU] key to select the setting.

6 Press the [SET] key to complete.This setting will remain after the power is turned OFF.

7 Press the [H/Dif] key to return to the original screen.

[MENU]

28

Page 37: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

6. SETTING UP [SETTING MODE]

1 Turn ON the power.The measurement mode is displayed.

2 Press the [MENU] key.3 Press the [MENU] key again.

Single/Repeat setting screen (SELECT rEPEAT) will be displayed.

4 Press the [SET] key.The current setting will be displayed.*The default setting is single measurement.

rEPEAT OFF: single measurementrEPEAT ON: repeat measurement

5 Press the [MENU] key to select the setting.

6.3 Setting Up Single / Repeat Measurement

[MENU]

29

Page 38: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

6. SETTING UP [SETTING MODE]

6 Press the [SET] key to complete.

7 Press the [H/Dif] key to return to the original screen.

30

Page 39: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

Always check and adjust before use to ensure accurate measurements.

Make sure the instrument is securely set up and stable before performing checks and adjustments.

Check that the bubble remains centered in the circular level. Adjust if the bubble shifts position.1 Adjust the leveling foot screws until the bubble is

centered in the circle.2 Turn the instrument 180°.

The bubble should not shift from the center. If the bubble does move, adjust as follows:

3 Compensate for one-half of the shift by adjusting the leveling foot screws.

4 Eliminate the remaining shift by turning the circular level adjusting screws with the hexagonal wrench until the bubble is centered.

7.1 Adjusting the Circular Level

Adjust with the leveling foot screw

Adjust with the level adjusting screw

Adjusting screws

31

Page 40: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

5 Turn the instrument 180°.If the bubble stays in the center, adjustment is completed. If the bubble moves, repeat steps 3 and 4.

Soft adjustment

1 Turn ON the power while simultaneously pressing the [MENU] key.The check / adjustment mode is displayed.

2 Set the equipment at the center (Point A) of the staff a and b.

3 Collimate the staff a and press the [MEAS] key.The measured distance of Aa will be displayed.

7.2 Adjusting the Cross-line

32

Page 41: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

Pressing the [MENU] key will switch the screen over to the distance display for Aa.

Pressing the [MEAS] key will repeat measurement at the same collimation.

4 Press the [SET] key.The measurement value for Aa (1.5064m) will be stored.

5 Collimate the staff b and press the [MEAS] key.

The measurement value for Ab will be displayed.

33

Page 42: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

Pressing the [MENU] key will switch the screen over to the distance display for Ab.

Pressing the [MEAS] key will repeat measurement at the same collimation.

6 Press the [SET] key.The measurement value for Ab (1.3887m) will be stored.

The display below indicates a setup error with differences in the distance between Aa and Ab exceeding 1 m. Turn OFF the power and restart the check / adjustment from the beginning.

7 Set the instrument at approximately 3 meters away from the staff a (Point B).

Approximately 3 meters

34

Page 43: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

8 Collimate the staff a after turning it 180° and press the [MEAS] key.The measured distance of Ba will be displayed.

Pressing the [MENU] key will switch the screen over to the distance display for Ba.

Pressing the [MEAS] key will repeat measurement at the same collimation.

9 Press the [SET] key.The measurement value for Ba (1.4005 m) will be stored.

10 Collimate the staff a and press the [MEAS] key.

35

Page 44: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

11 Collimate the staff b and press the [MEAS] key.The measured distance of Bb will be displayed.

Pressing the [MENU] key will switch the screen over to the distance display for Bb.

Pressing the [MEAS] key will repeat measurement at the same collimation.

12 Press the [SET] key.The difference (dIFFEr) obtained from the 4 measurements will appear.

The display below indicates an adjustment error. Turn OFF the power and restart the check / adjustment from the beginning.

The display below indicates a collimation error with differences in the distance of Ba exceeding the distance of Bb.Press the [H/Dif] key and start over from the step 8.

36

Page 45: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

13 Press the [SET] key.

The adjusted reading (1.2828 m) for the staff b is displayed.

14 Turn over the staff b.15 Align the reticle cross-lines of the instrument to the

reading (1.2828 m).16 Turn OFF the power. The adjustment has been

completed.

Mechanical adjustment

1 Collimate the barcode face of the staff b from the position B and measure with the instrument.

2 Turn the staff 180° from the position B, collimate the scale face of the staff b, and take a visual reading.

The adjusted measurement is set in the instrument.

37

Page 46: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

7. CHECKS AND ADJUSTMENTS [CHECK AND ADJUSTMENT MODE]

3 If the difference between the measurements taken at steps 1 and 2 is 2 mm or more, adjust as described below.If the difference is less than 2mm, the following steps are not necessary.

4 Remove the adjusting screw cover and insert a hexagonal wrench in the adjusting screw.

5 Turn the adjusting screw, then repeat steps 1 and 2. Adjust so that the difference between the two measurements is less than 2 mm.If the measurement at step 2 is larger than the measurement at step 1, lower the cross-line by slightly loosening the adjusting screw.If the measurement at step 2 is smaller than the measurement at step 1, raise the cross-line by slightly tightening the adjusting screw.

6 Close the adjusting screw cover.

Hexagonal wrench

Adjusting screw cover

38

Page 47: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

8. ERROR

The table below shows the error messages displayed by the instrument.

What to do when an error message appears

After the above state has been resolved,• During repeat measurement: automatically resumes

measuring operation• During single measurement: Press the [MEAS] key.

Error Number Type of error[ 1 ] Measurement[ 2 ] Vibration[ 3 ] Too dark[ 4 ] Too bright[ 5 ] Too close[ 6 ] Too far

Error number [1]: Measurement error

Example:

39

Page 48: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

9. STORAGE IN THE CASE

Diagram of storage case

• When storing the instrument in the storage case, align the positions of the leveling foot screw and notch on the case.

• Batteries included in the package are to confirm the initial operation.Their quality (including service life) is not guaranteed.

AA size dry cell battery Barcode tape staff

Hexagonal wrench

Operator’s manual Cleaning clothVinyl cover

ORION+Leveling foot screw

Notch

40

Page 49: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

10. STAFF TYPES

Staff Material Length / Surface treatment

Note

BAS55 Aluminum

5.0 m (5 positions)

RetractableFront surface: RAB cord

Back surface: Scale marks

BCS25 Stainless steel

6.01yd

TapeFront surface: RAB cord (2 m)

Back surface: Scale marks

41

Page 50: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

11. REGULATION

Region/Country: U.S.A.Directives/Regulations: FCC Class A

Region/Country: CanadaDirectives/Regulations: ICES Class A

FCC ComplianceWARNING:Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.NOTE:This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful inter-ference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Declaration of ConformityModel Number: AT-100D/AT-124DTrade Name: TOPCON CORPORATION

ManufacturerName: TOPCON CORPORATIONAddress: 75-1, Hasunuma-cho, Itabashi-ku, Tokyo, 174-8580

JAPANCountry: JAPAN

U.S.A. RepresentativeResponsible party: TOPCON POSITIONING SYSTEMS,INC.Address: 7400 National Drive Livermore, CA94551, U.S.ATelephone number: 925-245-8300

This Class A digital apparatus meets all requirements of Canadian Interference-Causing Equipment Regulations.Cet appareil numérique de la Class A respecte toutes les exigences du Réglement sur le matériel brouilleur du Canada.

This class A digital apparatus complies with Canadian ICES-003.

Cet appareil numerique de la classe A est conforme a la norme NMB-003 duCanada.

42

Page 51: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

11. REGULATION

Region/Country: EUDirectives/Regulations: EMC-Class B

Directives/Regulations: WEEE Directive

Directives/Regulations: EU Battery Directive

Region/Country: AustraliaDirectives/Regulations: C-Tick

Region/Country: Republic of KoreaDirectives/Regulations: KC-ClassA

43

Page 52: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

12. SPECIFICATIONS

TelescopeLength 206mm

Objective aperture Ø30mm

Magnification AT-100D 20x

AT-124D 24x

Image Erect

Resolving Power 4.5"

Field of view 1°20' or more

Minimum collimation distance 0.9m

Stadia ratio 1:100

Stadia additive constant (mm) 100

MeasurementHorizontal circle Arc-minute 90°

Measuring range Height 0 to 5m(With BAS55)0 to 2m(With barcode tape staff)

Distance 2.0 to 50m(With BAS55)0.9 to 30m(With barcode tape staff)

Minimum display Height 1mm

Distance 1cm

Accuracy(With BAS55) Height

Electronic Measurement

2.0mm at 30m away

44

Page 53: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

12. SPECIFICATIONS

Visual Measurement

2.0mm at 30m away

Distance

Electronic Measurement

Less than 10m: ±10mm or less10m to 50m:±0.30% x D and under (D: measured distance)

Visual Measurement

Distance: not prescribed

Measurement Measurement Mode

(Height / Distance) / (Height difference)

Procedure Single / Repeat (Accuracy)(Selectable)

Measurement time Single / Repeat (Accuracy) Within about 2 sec. (At favorable condition)

Automatic compensator

Magnetic damping and pendulum mechanism

Compensation range

±12'

PowerPower 2 x AA size dry cell batteries

(alkaline)

Battery state indicator

4 levels

Working duration Approx. 120 hours (at 20°C)(One measurement per 5 sec.)

45

Page 54: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

12. SPECIFICATIONS

GeneralLCD 7 digits X 2 lines segment display

(With illuminator)

Keyboard 6 keys(5 keys on front pane; 1 key on side panel)

Auto Power OFFON : power OFF if

not used for 30 min.

OFF : power not turn OFF automatically

Circular level sensitivity

8'/2mm

Operating temperature range

-20~50°C

Storage temperature range

-30~70°C

Water resistance IEC 60529IP55 (Category 2)

Dimensions Width (W) 123mm

Length (L) 206mm

Height (H) 137mm

Weight 1.3kg (with battery)

46

Page 55: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do

13. ERROR CHART

This is the Error Number Chart.Cut it off and paste it on the upper area of the instrument.

ErrorNumber

Type of error

[ 1 ] Measurement[ 2 ] Vibration[ 3 ] Too dark[ 4 ] Too bright[ 5 ] Too close[ 6 ] Too far

ErrorNumber

Type of error

[ 1 ] Measurement[ 2 ] Vibration[ 3 ] Too dark[ 4 ] Too bright[ 5 ] Too close[ 6 ] Too far

47

Page 56: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do
Page 57: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do
Page 58: OPERATOR'S MANUAL AUTOMATIC LEVEL · • The ORION+ is a precision instrument. Avoid severe shocks or vibration. • Be careful when removing the instrument from its case. • Do