BCM- 300 Cardio Potis User Manual ( Ver 1.1 ) Operation Manual Head Office/Factory 687-5, Sangoan-ri, Hongcheon-eup, Hongcheon-gun, Gangwon-do, Korea (zip.250-804) Tel : +82-33-434-9041 Fax : +82-33-434-9043 Seoul Office/R&D 1406, Masters Tower, 553, Dohwa-dong, Mapo-gu, Seoul, Korea (zip.121-748) Tel : +82-2-714-29602 Fax : +82-2-714-2963 Customer Service Dept. Tel : +82-2-714-2962 Fax : +82-2-714-2963 Address of Region Representative ET MEDICAL DEVICES SPA VIA DE ZINIS 6, 38011 CAVARENO (TN) ITALY Tel : +39 0463 85 01 25 Fax : +39 0463 85 00 88

Transcript of Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27....

Page 1: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

BCM-300Cardio Potis User Manual (Ver 1.1)

Operation Manual

Head Office/Factory 687-5, Sangoan-ri, Hongcheon-eup, Hongcheon-gun,Gangwon-do, Korea (zip.250-804)Tel : +82-33-434-9041 Fax : +82-33-434-9043

Seoul Office/R&D1406, Masters Tower, 553, Dohwa-dong, Mapo-gu,Seoul, Korea (zip.121-748)Tel : +82-2-714-2960∼2 Fax : +82-2-714-2963

Customer Service Dept.Tel : +82-2-714-2962 Fax : +82-2-714-2963

Address of Region Representative ET MEDICAL DEVICES SPAVIA DE ZINIS 6, 38011 CAVARENO (TN) ITALY Tel : +39 0463 85 01 25 Fax : +39 0463 85 00 88

Page 2: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

OPERATION MANUAL

BCM-300 Cardio Potis User Manual (Ver 1.1)

(MAR. 27. 2009)

Page 3: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

1. GENERAL INFORMATION 5

2. FEATURES 6

3. STANDARD SPECIFICATION 7

3.1. Features 7

3.2. Specification 7

3.2.1. Electrocardiography 7

3.2.2. General 8

3.2.3. Thermal Printer 8

3.2.4. Operation environments 8

3.2.5. Constitution 8

3.2.6. Standard accessories 8

4. PRODUCT DESCRIPTION 9

4.1. THE FRONT SIDE 9

4.2. THE LEFT SIDE 10

4.3. THE RIGHT SIDE 10

4.4. KEY BOARD PANEL 11

5. INSTALLATION 12

5.1. POWER CODE CONNECTION 12

5.2. PATIENT LEAD CONNECTION 12

5.3. EKG PAPER INSTALLATION 13

6. BCM-300 OPERATION 14

6.1. ATTACHMENT 14

6.2. PANEL KEY 15

6.2.1. LEAD Key 15

6.2.2. SENS Key 15

6.2.3. SPEED Key 15

6.2.4. FREEZE Key 15

6.2.5. PRINT Key 16

6.3. OPERATION OF THE MENU 16

6.3.1. Hospital Name 17

6.3.2. Patient Name 17

6.3.3. Patient Age 17

6.3.4. Patient Sex 17

6.3.5. EKG Lead 17

6.3.6. EKG Base 17

6.3.7. Notch Filter 18

6.3.8. LPF 18

6.3.9 BaseLine 18

6.3.10. Muscle 18

6.3.11. Date 19

6.3.12. Time 19

6.3.13. Exit 19

6.4. POWER ON/OFF 20

6.5. DEFAULT SET-UP VALUE 19

6.6. DIAGNOSIS 20

7. Power Supply 21

8. WARRANTY 23

9. AFTER SERVICE 24

Table of Contents

Page 4: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

GENERAL INFORMATION 5

1. GENERAL INFORMATION

Standard 12leads ( I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6).

Portable, Battery operation

Real time printing (1Channel or 3Channel)

Auto Interpretation

Page 5: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

STANDARD SPECIFICATION 7

3. STANDARD SPECIFICATION

3.1. Features

LCD Graphic Display 320 x 240 dots, 1 or 3 Channel with base line movement possibility

Standard 3 channel

Real time heart rate display

Real time display and print out

10 seconds memory for all 12 derivations

Image Freeze with the possibility to scroll into the memory for review or hardcopy.

Auto interpretation

Real time printing

Various patient information (Patient Name, Sex, Age)

High Resolution Thermal Printer (8 dots/mm)

Defibrillator protection

Rechargeable Battery Operation

3.2. Specification

3.2.1. Electrocardiography

Input circuit : Floating input

Leads : Standard 12 leads

Input impedance : 20MΩ min

Input circuit current : 0.05 μA ma

Calibration voltage : 1mV with ±2%

CMRR : 100dB or more

Polarization voltage : ±350mV min

Time constancy : 3.2 seconds (0 ~ +20%)

Sensitivity : 5, 10, 20 mm/mV ± 5%

Frequency response : 0.05 Hz ~ 100Hz (13dB)

Filters : AC 50/60 Hz, -20dB or lower , Muscle tremor 35 to 45 Hz, -3db(-6db/oct)

Display : LCD Graphic 320 x 240 dots

Cardio Potis User Manual (Ver 1.1)

6 FEATURES

2. FEATURES

3 Channel Interpretation EKG

The Merits⇨Various patient information (Patient Name, Patient Age, Patient Sex)

⇨Easy administration of a patient’s history

Real time display of the heart rate.

Easy-to Operate

Rechargeable Battery Operation (Economical, mobile)

Page 6: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

PRODUCT DESCRIPTION 9

4. PRODUCT DESCRIPTION

Open the packing box and check if main body and all the standard accessories are put in the box.

Also, check any breakage or damage which might be taken place in transit.

If you find out any breakage or missing accessories, please inquire about the problem at the

service department.

4.1. THE FRONT SIDE

Connecting part with Patient Cable

LCD Screen

Potential equalization stud

Power Adapter

Power Switch

Key Board Panel

Print Feed Button

Print Error LED

Print Power LED

Cardio Potis User Manual (Ver 1.1)

8 STANDARD SPECIFICATION

A/D conversion : 12bits

Sampling rate : 2.5ms

3.2.2. General

Power requirements : 90 ~ 240V, 50/60Hz, DC 7.2V with built-in battery

Power consumption : less than 7.2 VA max

Continuous operating time with battery : Max. 2.5hours

Dimension : 248(W) x 179(L) x 70(H)mm

Weight : 1.5kg

3.2.3. Thermal Printer

Thermal array : 8 dots/mm

Chart speed : 12.5mm/sec, 25.0mm/sec, 50.0mm/sec

Chart paper : 57mm(W) Roll paper

Dimension : 80(W)x45(D)x75(H)mm

Weight : 160g(include roll paper)

3.2.4. Operation environments

Temperature : 10 ~ 40℃ (45 ~ 113℉)

Humidity : 95% RH Max. @ 40℃ (113℉)

Atmospheric pressure : 70 ~ 106 kPa

3.2.5. Constitution

BCM-300 EKG main body

3.2.6. Standard accessories

10 Lead EKG Cable 1ea

Power Code 1ea

Limb strap 4ea

Chest electrodes 6ea

EKG gel 1ea (250ml)

Roll paper 2ea(57mm)

Operating manual 1ea

Ground Cable 1ea

Page 7: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

PRODUCT DESCRIPTION 11

4.4. KEY BOARD PANEL

MENU : Information Setup Menu

LEAD : EKG Lead

SENS : Sensitivity

SPEED : Waveform & Printer Speed

FREEZE : Freeze

PRINT : Manual Print

UP, DOWN, LEFT, RIGHT , ENTER : Move the menu

Cardio Potis User Manual (Ver 1.1)

10 PRODUCT DESCRIPTION

4.2. THE LEFT SIDE

Connecting Part with Patient Cable

4.3. THE RIGHT SIDE

Power Switch

Power Adapter

Potential equalization stud

Page 8: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

INSTALLATION 13

NOTICE

Make sure the EKG derivations according to IEC

- N Black

- R Red

- C1 White/Red

- C2 White/Yellow

- C3 White/Green

- C4 White/Brown

- C5 White/Black

- C6 White/Violet

- L Yellow

- F Green

5.3. EKG PAPER INSTALLATION

Open printer door by pulling the door cover

Insert EKG paper into its compartment and have recording side turned up.

Pull the end of paper up by the black driving roller and then close the door cover.

Cardio Potis User Manual (Ver 1.1)

12 INSTALLATION

5. INSTALLATION

Please be sure following precautions for installing and use.

Operating temperature 10 ~ 40℃

Check the connecting status of power cord, patient cable for smooth using.

Don’t plug several cords in one electric outlet.

Please place main body on the plain spot.

Don’t use the outlet to which electric appliances causing noise and connected

Secure enough space for paper use.

5.1. POWER CODE CONNECTION

Connect main Body to AC power inlet by using power cable after plugging power cable to

power adaptor.

5.2. PATIENT LEAD CONNECTION

Connect patient cable to patient cable connector at the left side of main body.

Connect LIMB STRAP and CHEST ELECTRODES to patient cable

Page 9: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

BCM-300 OPERATION 15

6.2. PANEL KEY

6.2.1. LEAD Key

1 or 3 lead is displayed on the LCD screen.

Press the “LEAD” key to select lead on the screen.

1 Lead Mode

⇨ I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6

3 Leads Mode

⇨First Group : I, II, III

⇨Second Group : aVR, aVL, aVF

⇨Third Group : V1, V2, V3

⇨ Fourth Group : V4, V5, V6

WARNING

You can not change the Lead mode or Setup menu during the printing

6.2.2. SENS Key

Select the sensitivity value by using the “SENS” key.

Sensitivity : 5mm/mV, 10mm/mV, 20mm/mV

6.2.3. SPEED Key

Select the waveform and printing speed by using “SPEED” key.

Speed : 12.5mm/sec, 25.0mm/sce, 50.0mm/sec

6.2.4. FREEZE Key

Press the key “FREEZE” to stop the display. The data acquisition will continue for 10seconds.

Cardio Potis User Manual (Ver 1.1)

14 BCM-300 OPERATION

6. BCM-300 OPERATION

6.1. ATTACHMENT

When the patient is on stable condition, lay him down on the bed. Then spread EKG gel all

over the parts where the electrodes are to be attached.

Please confirm the EKG signal on LCD screen after attaching the patient electrodes to the

patient.

[Note]Please rub the skin with the gel, the alcohol or the water before measuring the patient.

Page 10: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

BCM-300 OPERATION 17

6.3.1. Hospital Name

Select the Hospital Name Menu, and input the Hospital Name by using the numeric key.

To come back to the previous stage, press the “ENTER”Key.

6.3.2. Patient Name

Select the Patient Name Menu, and input the Patient Name by using the numeric key.

To come back to the previous stage, press the “ENTER”Key.

6.3.3. Patient Age

Select the Patient Age Menu, and input the Patient Age by using the numeric key.

To come back to the previous stage, press the “ENTER”Key.

6.3.4. Patient Sex

Select the Patient Sex, and select the sex type by using direction key.

Sex Type : Male/Female/-------

To come back to the previous stage, press the “ENTER”Key.

6.3.5. EKG Lead

Select the EKG Lead, and select Ekg Lead type by using direction key.

EKG Lead type : 1Lead or 3Leads

To come back to the previous stage, press the “ENTER”Key.

6.3.6. Ekg Base

Base 1 : Enables the first trace base line shift by using the is ▼ key or ▲ key.

Base 2 : Enables the second trace base line shift by using the ▼ key or ▲ key.

Base 3 : Enables the third trace base line shift by using the ▼ key or ▲ key.

To come back to the previous stage, press the “ENTER”Key.

6.3.7. Notch Filter

Select the Notch Filter, and select Notch Filter type by using direction key.

Notch Filter Type : 50Hz / 60Hz / Off

To come back to the previous stage, press the “ENTER”Key.

Cardio Potis User Manual (Ver 1.1)

16 BCM-300 OPERATION

When the acquisition is finished(MEMORY), press the <- key or -> key to review the traces

memorized and you can print them by pressing the “PRINT KEY”.

In Freezing status, you can change the LEAD, SPEED and SENSITIVITY.

To restart, just push “FREEZE” key.

6.2.5. PRINT Key

Press the key “PRINT” to print out the EKG data on the screen. The printed data will continue

for 5seconds.

WARNING

You can not change the LEAD, SPEED or SENSITIVITY during the manual Printing.

▣ CAUTION

During the printing, Keep the patient on a stable condition.

6.3. OPERATION OF THE MENU

If you press the “MENU” key, the Information menu can be displayed on the screen.

Setup Information

▶Hospital Name : PITALPatient Name : TIENTPatient Age : 35Patient Sex : MaleEkg Lead : 3 leadsEkg Base : BASE 1Notch Filter : 50HzLPF : 40HzBaseLine : ONMuscle : ONDate (D/M/Y) : 01/01/05Time (H/M) : 10:05

EXIT

Page 11: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

BCM-300 OPERATION 19

6.4. POWER ON/OFF

To turn on/off the equipment, press the main switch which is placed on its right side.

If the battery LED lights on, you must immediately plug AC power cord connected the

equipment.

6.5. DEFAULT SET-UP VALUE

Traces display/print number : 3

Paper Speed : 25.0mm/sec

Sensitivity of recording : 10mm/mV

Traces Group : I, II, III

EKG Lead Mode : 3 Leads

Filters : 50Hz Notch Filter

LPF(LowPass Filter) : 40Hz

BaseLine Filter : ON

Muscle Filter : ON

Hardcopy frame length : 10 seconds

6.6. DIAGNOSIS

If you want DIAGNOSIS,

Press [▼] key

After displaying “DIAGNOSIS”on the screen, press “ PRINT”key.

After acquiring the EKG data for 10 seconds and processing for 20seconds, the results can be

printed out.

Cardio Potis User Manual (Ver 1.1)

18 BCM-300 OPERATION

6.3.8. LPF(Lowpass Filter)

Select the LPF, and select Lowpass Filter type by using direction key.

Lowpass Filter can be selected to optimize ECG signal if ECG signal is still interfered.

Lowpass Filter Type : 40Hz / 100Hz /150Hz/ Off

To come back to the previous stage, press the “ENTER”Key.

6.3.9. BaseLine(Drift Filter)

Select the BaseLine, and select BaseLine Drift Filter type by using direction key.

BaseLine Drift is caused by patient breath.

BaseLine Filter Type : On/ Off

To come back to the previous stage, press the “ENTER”Key.

6.3.10. Muscle(EMG Filter)

Select the Muscle, and select Muscle Filter type by using direction key.

Muscle(EMG Signal) is induced from muscle and internal organs of the patient.

Muslce Filter Type : On/ Off

To come back to the previous stage, press the “ENTER”Key..

6.3.11. Date

Select the Date Menu, and input the Date by using the numeric key.

Display format : DD / MM / YY

Year : 2000 ~ 2099

Month : 01 ~ 12

Day : 01 ~ 31

To come back to the previous stage, press the “ENTER”Key.

6.3.12. Time

Select the Time Menu, and input the Time by using the numeric key.

Display format : HH / MM / SS

Hour : 00 ~ 23

Minute : 00 ~ 59

To come back to the previous stage, press the “ENTER”Key.

6.3.13. Exit

Used to go out from the Information Setup Menu.

Page 12: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

Power Supply 21

Cardio Potis User Manual (Ver 1.1)

20 Power Supply

7. Power Supply

7.1 Power Supply

BCM-300 can be operated by both AC and DC power. It can be operated while on the move by

getting the power from the internal battery which can be used again by recharging with the

AC/DC adapter.

7.1.1. AC/DC adapter power

In case battery power is running out, it is automatically recharged if you connect AC/DC adapter

to BCM-300 and to the outlet.

▣ CAUTION

AC/DC Adapter Aultkorea MW160-AKOP03F01 use only.

7.1.2. Battery power

The internal battery supplies the sources of electricity, if you separate the AC/DC adapter from

BCM-300.

Recharging period : Longer than 4 hours

Continuous Operation Time per full recharge : 2.5 hours

[Note]

In case battery power is running out, [ Low Bat.] will display at the bottom portion of the LCD

screen before about 10 minutes of the power shortage. And the power can be immediately

turned off according to the using parameter. Therefore, AC/DC power has to be supplied at

once.

7.1.3. Battery Specifications

BCM-300 uses a Li-ion 7.2V 4800mAh battery pack.

In case the battery gets damaged, it should be replaced.

Do not use the defective battery in the equipment.

It is automatically recharged, if the equipment is connected to the outlet through the AC/DC

adapter. It can not be recharged to be separated with the adapter.

[Note]

For preservation of environment, do not discard the batteries in an inappropriate places and

dispose in assigned place by the appropriate process after query at the Medical Electronics

Dept. of the hospital.

Page 13: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

AFTER SERVICE 23

9. AFTER SERVICE

If there is any problem with the equipment, please follow the steps below :

Contact the below BIONIS Co.,Ltd. Service Department immediately, after gathering the

model name, serial Number, date of purchase and description of the problem.

First, try to solve the problem over the phone with the service department personnel.

If the problem can not be solved over the phone, the service personnel will visit and fix the

problem directly

Cardio Potis User Manual (Ver 1.1)

22 WARRANTY

8. WARRANTY

BIONIS Co.,Ltd. guarantees all new equipment about faults in materials or workmanship for

a period of one year from date of shipment to purchaser. BIONIS Co.,Ltd. will, as sole and

exclusive remedy, at no charge, replace any such defective unit returned to its service

department within the designated warranty period.

This warranty applies only to failures from operating the equipment under conditions for which

it was designed. Warranted equipment is to be used only for intended and labeled indications

presented in the literature accompanying the equipment.

This warranty dose not cover loss or damage resulting from external cause, such as, but not

limited to fire, flood, windstorm, hail, lightening, earthquake, theft, misuse, or abuse. Any

damage caused by equipment relocation is not covered.

This warranty is void in cases where the product has been damaged as a result of an accident,

misuse, abuse, dropping, abnormal operating conditions, or attempts to modify or alter any part

or assembly.

BIONIS Co.,Ltd. will not be responsible for any loss, damage, or injure resulting from delay

in rendering service under this warranty.

BIONIS Co.,Ltd. will not be responsible to you for incidental or consequential damage of any

kind arising from connected with the use of its equipment.

Defective equipment shipped from you to BIONIS Co.,Ltd. must be packed in the

replacement cartons. Shipping and insurance costs for the return of the defective equipment

must be prepaid by you.

Page 14: Operation Manual - BIONICS OPERATION MANUAL BCM-300 Cardio Potis User Manual (Ver 1.1) (MAR. 27. 2009)

Cardio Potis User Manual (Ver 1.1)

Product Warranty

Product Name

Model Code

Manufacture Approval No.

Manufacture Approval Date

Manufacture Serial No.

Warranty Period

Purchase Date

Name ofHospital

Address

Name

Tel

Thank you for using BCM-300

This product is a medical device.

Only the products that have Passed our extensive and thorough quality tests are

offered to our customers

Our policies on repairing, exchange and refund complies with “Regulation on

Consumer Protection & Compensation Standards” issued by Economic

Planning Board of Korea