OPERATING INSTRUCTIONS - Ideal Vac...Tbl. 12: Pirani gauges 33 Tbl. 13: Cold cathode gauges 34 Tbl....
Transcript of OPERATING INSTRUCTIONS - Ideal Vac...Tbl. 12: Pirani gauges 33 Tbl. 13: Cold cathode gauges 34 Tbl....
OPERATING INSTRUCTIONS EN
Translation of the Original
PLUG-IN BOARDSfor total pressure measuring and control units TPG 300/TPG 500
Dear Customer,Thank you for choosing a Pfeiffer Vacuum product. Your new total pressure measuring and control unit should support you in your individual application with full performance and without malfunctions. The name Pfeiffer Vacuum stands for high-quality vacuum technology, a comprehensive and complete range of top-quality products and first-class service. From this extensive, practical experience we have gained a large volume of information that can contribute to efficient deployment and to your personal safety.In the knowledge that our product must avoid consuming work output, we trust that our product can offer you a solution that supports you in the effective and trouble-free implementation of your individual application.Please read these operating instructions before putting your product into operation for the first time. If you have any questions or suggestions, please feel free to contact [email protected] operating instructions from Pfeiffer Vacuum can be found in the Download Center on our website.
Disclaimer of liabilityThese operating instructions describe all models and variants of your product. Note that your product may not be equipped with all features described in this document. Pfeiffer Vacuum constantly adapts its products to the latest state of the art without prior notice. Please take into account that online operating instructions can deviate from the printed operating instructions supplied with your product.Furthermore, Pfeiffer Vacuum assumes no responsibility or liability for damage resulting from the use of the product that contradicts its proper use or is explicitly defined as foreseeable misuse.
CopyrightThis document is the intellectual property of Pfeiffer Vacuum and all contents of this document are protected by copyright. They may not be copied, altered, reproduced or published without the prior written permission of Pfeiffer Vacuum.We reserve the right to make changes to the technical data and information in this document.
2/48
Table of contents
1 About this manual 71.1 Validity 7
1.1.1 Applicable documents 71.1.2 Variants 7
1.2 Target group 81.3 Conventions 8
1.3.1 Instructions in the text 81.3.2 Pictographs 81.3.3 Abbreviations 8
1.4 Trademark proof 9
2 Safety 102.1 General safety information 102.2 Safety instructions 102.3 Safety precautions 112.4 Proper use 112.5 Foreseeable improper use 112.6 Responsibilities and warranty 122.7 Owner requirements 122.8 Personnel qualification 12
2.8.1 Ensuring personnel qualification 122.8.2 Personnel qualification for maintenance and repair 132.8.3 Advanced training with Pfeiffer Vacuum 13
2.9 Operator requirements 13
3 Product description 143.1 Function 14
3.1.1 Pirani measuring system 143.1.2 Cold cathode measuring system 14
3.2 Identifying the product 143.3 Structure of the plug-in boards 143.4 Interfaces 15
3.4.1 "OUTPUT" connection on PI 300 and CP 300 153.4.2 "CONTROL" connection on PE 300DC9 153.4.3 "RELAY" connection on IF 300A/C/P 163.4.4 "RELAY" connection on IF 300B 163.4.5 "RS-232" connection on IF 300A 173.4.6 "RS-232" connection on IF 300B 173.4.7 "RS-422" connection on IF 300C 183.4.8 "Profibus" connection on IF 300P 18
3.5 Scope of delivery 19
4 Transport and storage 204.1 Transporting the product 204.2 Storing the product 20
5 Installation 215.1 Installing/removing plug-in boards 215.2 Connecting gauges 21
5.2.1 Connecting Pirani gauges 215.2.2 Connecting the cold cathode gauge 22
5.3 Connecting interfaces and connections 235.3.1 Connecting the analog signal to the measurement boards 235.3.2 Connecting relays to the interface boards 235.3.3 Connecting the "RS-232-C" interface of the IF 300A 245.3.4 Connecting the "RS-232-C" interface of the IF 300B 24
Table of contents
3/48
5.3.5 Connecting the "RS-422" interface of the IF 300C 245.3.6 Connecting the "Profibus-DP" interface on the IF 300P 25
6 Calibrating the Pirani measuring circuit 26
7 Malfunctions 27
8 Shipping 29
9 Recycling and disposal 309.1 General disposal information 309.2 Disposing of plug-in boards 30
10 Service solutions by Pfeiffer Vacuum 31
11 Accessories 3311.1 Accessory information 3311.2 Ordering accessories 33
12 Technical data and dimensions 3712.1 Technical data 3712.2 Dimensions 41
13 Appendix 4313.1 Relation: measuring signal and pressure 4313.2 Gas type dependence 4413.3 Units of pressure 4613.4 Gas throughputs 46
Table of contents
4/48
List of tablesTbl. 1: Applicable documents 7Tbl. 2: Variants 7Tbl. 3: Abbreviations used 9Tbl. 4: Switch functions on IF 300A/C/P 16Tbl. 5: Switching functions on IF 300B 17Tbl. 6: Disturbances during installation 27Tbl. 7: Interference during operation and calibration 27Tbl. 8: Disturbances as a result of defects 27Tbl. 9: Disturbances with the RS-232-C interface 28Tbl. 10: Disturbances with the RS-422 interface 28Tbl. 11: Disturbances with the Profibus-DP interface 28Tbl. 12: Pirani gauges 33Tbl. 13: Cold cathode gauges 34Tbl. 14: Measuring cable for TPR 010 34Tbl. 15: Measuring cable for TPR 017 and TPR 018 35Tbl. 16: Extension cable for TPR 010, TPR 017 and TPR 018 35Tbl. 17: Measuring cable for IKR 050 and IKR 060 35Tbl. 18: Measuring cable for IKR 070 36Tbl. 19: Plugs and interface cable 36Tbl. 20: Dummy plates 36Tbl. 21: Technical data PI 300D / PI 300DN 37Tbl. 22: Technical data PE 300DC9 37Tbl. 23: Technical data CP 300C9/C10/T11/T11L 39Tbl. 24: Technical data IF 300A 39Tbl. 25: Technical data IF 300B 40Tbl. 26: Technical data IF 300C 40Tbl. 27: Technical data IF 300P 41Tbl. 28: Units of pressure and their conversion 46Tbl. 29: Gas throughputs and their conversion 46
List of tables
5/48
List of figuresFig. 1: "OUTPUT" connection with 2 sockets 15Fig. 2: "CONTROL" connection on PE 300DC9 15Fig. 3: Switch and logic on the "CONTROL" connection 15Fig. 4: "RELAY" connection on IF 300A/C/P 16Fig. 5: "RELAY" connection on IF 300B 16Fig. 6: "RS-232" connection on IF 300A 17Fig. 7: "RS-232" connection on IF 300B 17Fig. 8: "RS-422" connection on IF 300C 18Fig. 9: "Profibus" connection on IF 300P 19Fig. 10: Coax cable and triax cable 22Fig. 11: Locking the triax plug 23Fig. 12: Folding ferrite for IF 300B 23Fig. 13: Connecting the interface cable to IF 300B 24Fig. 14: ATM and HIGH VAC trimming potentiometers 26Fig. 15: Pirani measuring cable connection diagram 34Fig. 16: Pirani extension cable connection diagram 34Fig. 17: Dimensions of measurement boards 41Fig. 18: Dimensions of interface and relay boards 42Fig. 19: Relation: Measuring signal and pressure with TPR gauges 43Fig. 20: Relation: Measuring signal and pressure with IKR gauges 44Fig. 21: Displayed pressure for TPR gauges 45Fig. 22: Displayed pressure for IKR gauges 45
List of figures
6/48
1 About this manualIMPORTANTRead carefully before use.Keep the manual for future consultation.
1.1 ValidityThis document describes the function of the products listed in the following and provides the most im-portant information for safe use. The description is written in accordance with the valid directives. The information in this document refers to the current development status of the products. The document retains its validity assuming that the customer does not make any changes to the product.
1.1.1 Applicable documents
Designation DocumentOperating instructions "Total pressure measuring and control unit"TPG 300
BG 5970
Operating instructions "Total pressure measuring and control unit"TPG 500
BG 6008
Communication protocol "Total pressure measuring and control unit"TPG 300
BG 5973
Communication protocol "Total pressure measuring and control unit"TPG 500
BG 6009
“Gauges” operating instructionsModulLine gauges
(depending on the gauge used)
Tbl. 1: Applicable documents
You can find these documents in the Pfeiffer Vacuum Download Center.
1.1.2 VariantsThis document applies to the products with the following part numbers:
Part number Type DesignationPT 546 920-T Measurement board (Pirani) PI 300DPT 549 214-T PI 300DNPT 441 375-T Measurement board (cold cathode) PE 300DC9PT 441 000-T Measurement board (Pirani/cold cathode) CP 300C9PT 441 114-T CP 300C10PT 441 080-T CP 300T11PT 441 120-T CP 300T11LPT 441 130-T Interface and relay board (RS-232-C) IF 300APT 441 250-T IF 300BPT 441 390-T Interface and relay board (RS-422) IF 300CPT 441 395-T Interface and relay board (Profibus) IF 300P
Tbl. 2: Variants
You can find the part number on the rating plate of the product.Pfeiffer Vacuum reserves the right to make technical changes without prior notification.The figures in this document are not to scale (dimensions in mm).
About this manual
7/48
1.2 Target groupThese operating instructions are aimed at all persons performing the following activities on the product:
● Transportation● Setup (Installation)● Usage and operation● Decommissioning● Maintenance and cleaning● Storage or disposal
The work described in this document is only permitted to be performed by persons with the appropriate technical qualifications (expert personnel) or who have received the relevant training from Pfeiffer Vac-uum.
1.3 Conventions
1.3.1 Instructions in the textUsage instructions in the document follow a general structure that is complete in itself. The required ac-tion is indicated by an individual step or multi-part action steps.Individual action stepA horizontal, solid triangle indicates the only step in an action.
► This is an individual action step.Sequence of multi-part action stepsThe numerical list indicates an action with multiple necessary steps.
1. Step 12. Step 23. ...
1.3.2 PictographsPictographs used in the document indicate useful information.
Note
Tip
1.3.3 Abbreviations
Abbreviation ExplanationASCII American Standard Code for Information Interchange
ATM AtmosphereCNTR-P Repeater direction control signal (Profibus)
ESD Electrostatic DischargeIF InterfaceGND GroundHV High vacuumMHV Miniature High Voltage
(coax high voltage connector in miniature design)NN Mean sea levelPE PenningProfibus DP Profibus distributed periphery
About this manual
8/48
Abbreviation ExplanationRS-232 Recommended Standard 232
(Standard for a serial interface often available on computers)RS-422 Recommended Standard 422
(Interface standard for a wired differential serial data transfer)RxD Received eXchange Data (incoming data line)SHV Safe High Voltage (connector for coax cable)TPG Total Pressure Gauge (total pressure measuring and control unit)TxD Transmitted eXchange Data (outgoing data line)
Tbl. 3: Abbreviations used
1.4 Trademark proof● Profibus® is a registered trading name of Profibus Nutzerorganisation e.V.
About this manual
9/48
2 Safety
2.1 General safety informationThe following 4 risk levels and 1 information level are taken into account in this document.
DANGERImmediately pending dangerIndicates an immediately pending danger that will result in death or serious injury if not observed.
► Instructions to avoid the danger situation
WARNINGPotential pending dangerIndicates a pending danger that could result in death or serious injury if not observed.
► Instructions to avoid the danger situation
CAUTIONPotential pending dangerIndicates a pending danger that could result in minor injuries if not observed.
► Instructions to avoid the danger situation
NOTICEDanger of damage to propertyIs used to highlight actions that are not associated with personal injury.
► Instructions to avoid damage to property
Notes, tips or examples indicate important information about the product or about this docu-ment.
2.2 Safety instructionsSafety instructions according to product’s life stagesAll safety instructions in this document are based on the results of a risk assessment. Pfeiffer Vacuum has taken into account all the relevant life stages of the product.
Risks during installation
DANGERDanger to life due to dangerous contact voltageVoltages above 30 V (AC) or 60 V (DC) are considered dangerous in accordance with EN 61010. If you come into contact with dangerous contact voltage, this can result in injury through electric shocks or even death.
► Only connect the gauge to professionally grounded devices.► Take additional safety measures on the system-side if processes in the vacuum system (e.g.
flashovers) can result in dangerous voltages at the gauge connections.► Secure the line to the gauge.
Safety
10/48
Risks during disposal
WARNINGHealth hazard through poisoning from toxic contaminated components or devicesToxic process media result in contamination of devices or parts of them. During maintenance work, there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub-stances causes environmental damage.
► Take suitable safety precautions and prevent health hazards or environmental pollution by toxic process media.
► Decontaminate affected parts before carrying out maintenance work.► Wear protective equipment.
2.3 Safety precautionsThe product is designed according to the latest technology and recognized safety engineering rules. Nevertheless, improper use can result in danger to operator all third party life and limb, and product damage and additional property damage.
Duty to provide information on potential dangersThe product holder or user is obliged to make all operating personnel aware of dangers posed by this product.Every person who is involved in the installation, operation or maintenance of the product must read, understand and adhere to the safety-related parts of this document.
Infringement of conformity due to modifications to the productThe Declaration of Conformity from the manufacturer is no longer valid if the operator changes the original product or installs additional equipment.
● Following the installation into a system, the operator is required to check and re-evalu-ate the conformity of the overall system in the context of the relevant European Direc-tives, before commissioning that system.
General safety precautions when handling the product► When handling the gases and contaminated parts used, observe the applicable guidelines.► Observe the protective measures.► Observe the safety guidelines specified in this document.
– All work is only permissible when observing the relevant guidelines and adhering to the pro-tective measures.
► Inform yourself about any contamination before starting work.► Pass on safety instructions to all other users.
2.4 Proper useThe plug-in boards are analog measuring, interface and relay boards for the ModulLine total pressure measuring and control units TPG 300 and TPG 500.
► Install, operate and maintain the product only in accordance with these operating instructions.► Comply with the application limits.► Observe the technical data.
2.5 Foreseeable improper useImproper use of the product invalidates all warranty and liability claims. Any use that is counter to the purpose of the product, whether intentional or unintentional, is regarded as misuse, in particular:
● Use outside the mechanical and electrical application limits in accordance with the technical data● Use with corrosive or explosive media, if this is not explicitly permitted● Use outdoors● Use after technical changes (on the inside or the outside of the product)● Use with replacement or accessory parts that are unsuitable or are not approved
Safety
11/48
2.6 Responsibilities and warrantyPfeiffer Vacuum shall assume no responsibilities and warranty if the operating company or a third party:
● disregards this document.● does not use the product for its intended purpose.● carries out any modifications to the product (conversions, changes, maintenance work, etc.) that
are not listed in the corresponding operating instructions.● operates the product with accessories that are not listed in the corresponding operating instruc-
tions.The operator is responsible for the process media used.
2.7 Owner requirementsSafety-conscious working
1. Only operate the product in a technically flawless state.2. Operate the product in line with its intended purpose, safety and hazard-conscious and only in
compliance with these operating instructions.3. Fulfill the following instructions and monitor the observation of the following instructions:
– Proper use– Generally applicable safety instructions and accident prevention regulations– International, national and locally applicable standards and guidelines– Additional product-related guidelines and regulations
4. Only use original parts or parts approved by Pfeiffer Vacuum.5. Keep the operating instructions available at the place of installation.6. Ensure personnel qualification.
2.8 Personnel qualificationThe work described in this document may only be carried out by persons who have appropriate profes-sional qualifications and the necessary experience or who have completed the necessary training as provided by Pfeiffer Vacuum.Training people
1. Train the technical personnel on the product.2. Only let personnel to be trained work with and on the product when under the supervision of
trained personnel.3. Only allow trained technical personnel to work with the product.4. Before starting work, make sure that the commissioned personnel have read and understood
these operating instructions and all applicable documents, in particular the safety, maintenance and repair information.
2.8.1 Ensuring personnel qualificationSpecialist for mechanical workOnly a trained specialist may carry out mechanical work. Within the meaning of this document, special-ists are people responsible for construction, mechanical installation, troubleshooting and maintenance of the product, and who have the following qualifications:
● Qualification in the mechanical field in accordance with nationally applicable regulations● Knowledge of this documentation
Specialist for electrotechnical workOnly a trained electrician may carry out electrical engineering work. Within the meaning of this docu-ment, electricians are people responsible for electrical installation, commissioning, troubleshooting, and maintenance of the product, and who have the following qualifications:
● Qualification in the electrical engineering field in accordance with nationally applicable regulations● Knowledge of this documentation
In addition, these individuals must be familiar with applicable safety regulations and laws, as well as the other standards, guidelines, and laws referred to in this documentation. The above individuals must have an explicitly granted operational authorization to commission, program, configure, mark, and earth devices, systems, and circuits in accordance with safety technology standards.
Safety
12/48
Trained individualsOnly adequately trained individuals may carry out all works in other transport, storage, operation and disposal fields. Such training must ensure that individuals are capable of carrying out the required activi-ties and work steps safely and properly.
2.8.2 Personnel qualification for maintenance and repair
Advanced training coursesPfeiffer Vacuum offers advanced training courses to maintenance levels 2 and 3.
Adequately trained individuals are:● Maintenance level 1
─ Customer (trained specialist)● Maintenance level 2
─ Customer with technical education─ Pfeiffer Vacuum service technician
● Maintenance level 3─ Customer with Pfeiffer Vacuum service training─ Pfeiffer Vacuum service technician
2.8.3 Advanced training with Pfeiffer VacuumFor optimal and trouble-free use of this product, Pfeiffer Vacuum offers a comprehensive range of courses and technical trainings.For more information, please contact Pfeiffer Vacuum technical training.
2.9 Operator requirementsObserving relevant documents and data
1. Read, observe and follow this operating instruction and the work instructions prepared by the op-erating company, in particular the safety and warning instructions.
2. Install, operate and maintain the product only in accordance with these operating instructions.3. Carry out all work only on the basis of the complete operating instructions and applicable docu-
ments.4. Comply with the application limits.5. Observe the technical data.6. Please contact the Pfeiffer Vacuum Service Center if your questions on operation or maintenance
of the product are not answered by these operating instructions.– You can find information in the Pfeiffer Vacuum service area.
Safety
13/48
3 Product description
3.1 Function
3.1.1 Pirani measuring systemWithin certain limits, the heat conductivity of gases depends on the pressure. This physical occurrence is utilized in the Pirani thermal conductivity vacuum gauge to measure the pressure. A thin wolfram wire with high temperature coefficients is used as the measuring element. With a suitable control circuit, the resistance of the wire and thus its temperature are kept constant. The electrical power that is fed to the wire is a measure for the heat power dissipated in the gas through heat conduction and therefore a measure of the pressure.
3.1.2 Cold cathode measuring systemThe current that flows into a self-sustaining gas discharge with cold cathode (similar to Penning) de-pends on the applied voltage, the composition of the gas and the pressure. A magnetic field penetrates into the measuring room that causes the electrons on the spiral trajectories to move from the cathode towards the anode, and as a result perform a sufficient number of ionizing impacts to maintain the dis-charge, even with low gas density. If the anode voltage and magnetic field are now kept constant (with known gas type), the discharge current is a measure for the pressure.
3.2 Identifying the productYou will need all the data from the rating plate to safely identify the product when communicating with Pfeiffer Vacuum.
1. Read the data on the product rating plate.2. Record this data.3. Always have all rating plate specifications to hand.
3.3 Structure of the plug-in boardsPirani measurement boardsWhen the total pressure measuring and control unit is switched on, the Pirani measuring circuits are al-ways in operation. The analog signals are always available, irrespective of the pressure display.Pirani measurement boards have:
● 2 independent fine vacuum measuring circuits, each with─ 1 gauge power supply plug─ 2 calibration potentiometers─ 1 analog signal output
Cold cathode measurement boardsWhen the gauge is switched on, the analog signals are always available, irrespective of the pressure display. This measurement board has special electronics that limit the measurement current to 100 μA. This results in a considerably longer service life for the gauges.Cold cathode measurement boards for high vacuum have:
● 2 measuring circuits, each with─ 1 gauge power supply plug─ 1 analog signal output
Pirani/cold cathode measurement boardsThe properties of these measurement boards correspond with the Pirani and cold cathode measure-ment boards. The measurement boards for the range 10-10 hPa (CP 300C10) and 10-11 hPa (CP 300T11 and CP 300T11L) have special electronics that limit the measurement current to 100 μA. This results in a considerably longer service life for the gauges.Pirani/cold cathode measurement boards have:
● 1 Pirani measuring circuit● 1 cold cathode measuring circuit
Product description
14/48
Interfaces and relay boardsThere are 4 types of interface and relay boards:
● 2 with RS-232-C interfaces● 1 with RS-422 interface● 1 with Profibus-DP interface
All 4 types each have 5 relays with potential-free switching contacts. The types each have a different switching voltage for the relays and different interface connections.
3.4 Interfaces
3.4.1 "OUTPUT" connection on PI 300 and CP 300Each measuring circuit has an analog signal output. Suitable plugs are included with each measure-ment board.
OUTPUT
0 ... 10 V
– +
Fig. 1: "OUTPUT" connection with 2 sockets
3.4.2 "CONTROL" connection on PE 300DC9Each measuring circuit has a control input as well as an analog signal output. A suitable plug is included with each measurement board.
4
5
1
2
3
6
7
Screen
Fig. 2: "CONTROL" connection on PE 300DC9
1 Analog GND IKR 1 5 IKR ON 22 Digital GND 6 Analog Output3 Analog GND 7 Analog Output4 IKR ON 1 Gehäuse Screen
2
4, 5
2
4, 5
Fig. 3: Switch and logic on the "CONTROL" connection
Switching the measuring circuit on or off externally (PE 300DC9)1. First switch the gauge on manually or automatically if you want to control the gauge via the exter-
nal input.2. Switch the measuring circuit on or off using one of the three options.
● manual● automatic● external, via a contact on the "CONTROL" connection
Product description
15/48
3.4.3 "RELAY" connection on IF 300A/C/P
19
15 8
Screen
Pin
1
2
9
3
11
10
12
5
4
6
14
13
8
15
7
Fig. 4: "RELAY" connection on IF 300A/C/P
The figure shows the contacts in rest position.
Pins Designation Switching function15 8 7 Relay 1 114 6 13 Relay 2 25 12 4 Relay 3 311 3 10 Relay 4 42 1 9 Relay 5 Error status
Tbl. 4: Switch functions on IF 300A/C/P
3.4.4 "RELAY" connection on IF 300B
32
28
30
26
22
24
20
16
18
14
10
12
8
4
6
IF 300B
3032
26
22
18
14
10
6
28
24
20
16
12
8
4
Fig. 5: "RELAY" connection on IF 300B
The figure shows the contacts in rest position.
Pins Designation Switching function4 8 6 Relay 1 110 14 12 Relay 2 216 20 18 Relay 3 3
Product description
16/48
Pins Designation Switching function22 26 24 Relay 4 428 32 30 Relay 5 Error status
Tbl. 5: Switching functions on IF 300B
3.4.5 "RS-232" connection on IF 300A
RXD
TXD
6 1
59
2
3
7
IF 300A
3
2
7
2
3
7
CTS
RTS
TXD
S GND
RXD
Screen
S GND
DSR
DTR
CCD
5
4
6
20
8
Computer
Fig. 6: "RS-232" connection on IF 300A
2 Receive data (RxD)(relating to the measuring device)
7 Signal ground (S GND)
3 Transmit data (TxD)(relating to the measuring device)
3.4.6 "RS-232" connection on IF 300B
RXD
TXD2
3
7
IF 300B
3
2
7
2
3
7
CTS
RTS
TXD
S GND
RXD
Screen
S GND
DSR
DTR
CCD
5
4
6
20
8
Computer
14
1 13
25
Fig. 7: "RS-232" connection on IF 300B
2 Receive data (RxD)(relating to the measuring device)
7 Signal ground (S GND)
3 Transmit data (TxD)(relating to the measuring device)
Product description
17/48
3.4.7 "RS-422" connection on IF 300C
5
1
6
IF 300C
8
7
9
9 5
16
TX+
RX–
8
7
9
Screen
TX–
S GND
P GND
6
5
1
RX+
Fig. 8: "RS-422" connection on IF 300C
1 Protective ground (P GND) 7 Reception data (Rx-)(relating to the measuring device)
2, 3, 4 unassigned 8 Transmission data (Tx+)(relating to the measuring device)
5 Signal ground (S GND) 9 Transmission data (Tx)-(relating to the measuring device)
6 Reception data (Rx+)(relating to the measuring device)
Screen Screen
3.4.8 "Profibus" connection on IF 300P
Details on the Profibus-DP interfaceYou can find details on the Profibus-DP interface in the corresponding communication pro-tocol.
Product description
18/48
IF 300P
59
6 18
9
6
7
4
5
2
3
1
Pin
Fig. 9: "Profibus" connection on IF 300P
1 Screen 6 VP (only required with bus end devices)2 Do not connect 7 Do not connect3 RxD/TxD-P (Line B) 8 RxD/TxD-N (Line A)4 CNTR-P (optional) 9 Do not connect5 DGND
3.5 Scope of deliveryThe shipment includes the following parts:
● Plug-in board● Plugs for analog signals● Folding ferrite (for IF 300B)● Operating instructions
Unpacking the product and checking completeness of the shipment1. Unpack the product.2. Remove the transport fasteners, transport protection etc.3. Store the transport fasteners, transport protection etc. in a safe place.4. Check that the shipment is complete.5. Ensure that no parts are damaged.
Product description
19/48
4 Transport and storage
4.1 Transporting the product
NOTICEDamage caused by incorrect transportationTransportation in unsuitable packaging, or failure to install all transport locks, can damage the prod-uct.
► Comply with the instructions for safe transportation.
PackingWe recommend keeping the transport packaging and original protective cover.
General information regarding safe transport► Observe the weight specified on the transport packaging.► Where possible, always transport or ship the product in the original transport packaging.► Always use dense and impact-proof transport packaging for the product.► Remove the existing protective cover and transport protections only immediately prior to installa-
tion.► Reattach transport locks and transport protections prior to each transport.
4.2 Storing the product
NOTICEDamage caused by improper storageImproper storage will lead to damage to the product.Static charging, moisture, etc. lead to defects on the electronic components.
► Comply with the instructions for safe storage.
PackingWe recommend storing the product in its original packaging.
General information for safe storage► Store the product in a cool, dry, dust-free place, where it is protected against impacts and me-
chanical vibration.► Always use dense and impact-proof packaging for the product.► Where possible, store the product in its original packaging.► Store electronic components in antistatic packaging.► Maintain the permissible storage temperature.► Avoid extreme fluctuations of the ambient temperature.► Avoid high air humidity.► Seal connections with the original protective caps.► Protect the product with the original transport protection (where available).
Transport and storage
20/48
5 Installation
5.1 Installing/removing plug-in boards
NOTICEDamage to unit from electrostatic dischargeElectrostatic discharge damages electronic components. Defects resulting from failure to comply with this warning result in loss of any warranty claim entitlement.
► Carry out all work on ESD-protected workstations only, in compliance with the appropriate work-ing methods.
► When the unit is open, take suitable precautions against electrostatic charges.► Always store electronic units and components in anti-static packaging (see Storage).► Observe EN 61340, Parts 5-1 and 5-2 – Protection of electronic devices from electrostatic phe-
nomena.
Dummy plates for free slotsAlways cover the free slots on the total pressure measuring and control unit with dummy plates for safety reasons.Suitable dummy plates are available as accessories.
Required tools● Flat-tip screwdriver
Accessories required● Dummy plates, if free slots are available on the TPG.
Procedure1. Switch the TPG off.2. Wait at least 1 minute.3. Disconnect all cable connections, except the mains cable of the TPG.4. Afterwards, connect the mains cable of the TPG.5. Unscrew the dummy plate/plug-in board.6. Remove the dummy plate/plug-in board from the TPG.7. Carefully insert the plug-in board into the TPG.8. Carefully tighten the plug-in board.
– If plug-in board screws are loose, trouble-free operation cannot be guaranteed.9. Connect the mains cable back onto the TPG.
10. Re-establish all other cable connections to the TPG.11. Switch the TPG back on again.
5.2 Connecting gauges
5.2.1 Connecting Pirani gauges
DANGERDanger to life due to dangerous contact voltageVoltages above 30 V (AC) or 60 V (DC) are considered dangerous in accordance with EN 61010. If you come into contact with dangerous contact voltage, this can result in injury through electric shocks or even death.
► Only connect the gauge to professionally grounded devices.► Take additional safety measures on the system-side if processes in the vacuum system (e.g.
flashovers) can result in dangerous voltages at the gauge connections.► Secure the line to the gauge.
Installation
21/48
Procedure1. Make sure that the TPG is switched off.2. Despite being partially screened, do not lay the gauge cable parallel with cables with too much
interference.3. Connect the gauge with the measuring cable to the "IKR" connection on the measurement board.4. Lock the plug to prevent unintentional disconnection.
5.2.2 Connecting the cold cathode gauge
DANGERDanger to life due to dangerous contact voltageVoltages above 30 V (AC) or 60 V (DC) are considered dangerous in accordance with EN 61010. If you come into contact with dangerous contact voltage, this can result in injury through electric shocks or even death.
► Only connect the gauge to professionally grounded devices.► Take additional safety measures on the system-side if processes in the vacuum system (e.g.
flashovers) can result in dangerous voltages at the gauge connections.► Secure the line to the gauge.
Maximum length of coax cablesThe maximum length of 100 m for coax cables applies for compliance with EN 61010. Lon-ger lengths are not permitted without additional safety measures.
2 4 6 10 2 4 6 2 4 500 100
l [m]
10-7
8 6
4
2
P [hPa]
10-8
86
4
2
10-9
86
4
2
10-10
86
4
2
10-11
5×10 -9
Coax cable
Triax cable
Coax cablee
Triax cable
Fig. 10: Coax cable and triax cable
Installation
22/48
Required tool● With triax cable: Open-end wrench (17 mm WAF)
17 mm
unlock
lock
Fig. 11: Locking the triax plug
Procedure1. Make sure that the TPG is switched off.2. Despite being partially screened, do not lay the gauge cable parallel with cables with too much
interference.3. Choose the suitable measuring cable type in accordance with Figure "Coax cable and triax cable".4. Connect the gauge with the measuring cable to the "IKR" connection on the measurement board.5. With triax cable: Lock the triax plug to prevent unintentional disconnection, loosening and touching
of the inner conductor.6. Additionally, ground the gauge if the gauge is not grounded via the recipients.
5.3 Connecting interfaces and connections
5.3.1 Connecting the analog signal to the measurement boardsProcedure
1. Only use screened cable.2. Connect the screen to the plug sleeve.3. If you ground both ends of the screen you prevent the occurrence of compensating currents (e.g.
when connecting all devices involved to the same mains distributor).4. Make sure that components to be connected externally contain the technical data of the plug-in
board.5. Connect the analog signal to the plugs included with the plug-in board.
5.3.2 Connecting relays to the interface boardsConnecting relays (IF 300A/C/P)
1. Only carry out work on the plug when it is de-energized.2. Make sure that components to be connected externally contain the technical data of the plug-in
board.3. Connect the relay with the relay plug.
Insulated flat pin terminals for IF 300BWith voltages < 50 V, you may use 4.8 × 0.8 mm insulated flat pin terminals. However, Pfeiffer Vacuum recommends using the relay plug, as the relay plug enables fast discon-nection of the connection as well as strain relief.
Fig. 12: Folding ferrite for IF 300B
Installation
23/48
Connecting relays (IF 300B)Screened cable is not required for the connection.
1. Only carry out work on the plug when it is de-energized.2. In general, use the relay plug for voltages ≥ 50 V for safety reasons.3. Make sure that components to be connected externally contain the technical data of the plug-in
board.4. Mount the folding ferrite in order to prevent electromagnetic interference.
– The folding ferrite is included in the scope of delivery.5. Maintain a distance of max. 5 cm between the folding ferrite and the relay plug.6. Connect the relay to the relay plug or insulated flat pin terminals.
5.3.3 Connecting the "RS-232-C" interface of the IF 300AProcedure
1. Only use screened cable.2. Connect the screen to the plug sleeve.3. If you ground both ends of the screen you prevent the occurrence of compensating currents (e.g.
when connecting all devices involved to the same mains distributor).4. Make sure that components to be connected externally contain the technical data of the plug-in
board.5. Use a suitable interface cable to connect the interface.
5.3.4 Connecting the "RS-232-C" interface of the IF 300BAccessories required
● Interface cable, for IF 300 B, RS-232-C, 0.4 m
Fig. 13: Connecting the interface cable to IF 300B
Procedure1. Guide the interface cable to connection "X2" on the interface board.2. Fasten the plug of the interface cable to connection "X2" on the circuit board.3. Connect the interface cable in the front plate of the interface board.4. Tighten the interface cable using the fastening parts.5. Pull the interface cable through the folding ferrite.
– The folding ferrite is included in the scope of delivery.6. Secure the interface cable to the circuit board using the cable tie.7. Mount the folding ferrite in order to prevent electromagnetic interference.8. Maintain a distance of max. 5 cm between the folding ferrite and the front plate.
5.3.5 Connecting the "RS-422" interface of the IF 300CProcedure
1. Only use screened cable.2. Connect the screen to the plug sleeve.3. If you ground both ends of the screen you prevent the occurrence of compensating currents (e.g.
when connecting all devices involved to the same mains distributor).
Installation
24/48
4. Make sure that components to be connected externally contain the technical data of the plug-in board.
5. Use a suitable interface cable to connect the interface.
5.3.6 Connecting the "Profibus-DP" interface on the IF 300P
Details on the Profibus-DP interfaceYou can find details on the Profibus-DP interface in the corresponding communication pro-tocol.
Procedure1. Only use the special cable provided.2. Make sure that components to be connected externally contain the technical data of the plug-in
board.3. Connect the interface using the special cable.
Installation
25/48
6 Calibrating the Pirani measuring circuitMaintenance in the Pfeiffer Vacuum Service CenterPfeiffer Vacuum offers a complete maintenance service for all products.Pfeiffer Vacuum recommends: Contact your Pfeiffer Vacuum Service Center to arrange the maintenance of defective products and components.
Utilization in other climatic conditions, long-term operation, extreme temperatures, a different mounting orientation and aging or contamination can lead to a zero point shift in the Pirani measuring circuit. This will in turn necessitate recalibration or cleaning.Pfeiffer Vacuum has calibrated the gauge to standard values at the factory. However, as a result of indi-vidual scattering, gauge tolerances, gauge contamination or different cable lengths, you must accept deviations. You can compensate these factors with 2 trimming potentiometers in certain limits. You do not need to calibrate cold cathode measuring circuits.
Required tool● Flat-tip screwdriver
OUTPUT
0 ... 10 V
– +
ATM
HIGH VAC
Fig. 14: ATM and HIGH VAC trimming potentiometers
Preparing for calibration1. Perform a calibration at regular intervals.2. Perform the calibration under the same, constant ambient conditions and with the same mounting
orientation as normally used for the gauge.3. Start up the measuring tube.
You may now perform a calibration.Performing an HV calibration
1. Operate the gauge (for at least 10 minutes) until the gauge has reached the operating tempera-ture.
2. Evacuate the vacuum system to pressure < 10-4 hPa.3. Select the measuring circuit you wish to calibrate on the TPG.
– You can find further information on the "Sensor" mode in the TPG operating instructions.4. Rotate the "HIGH VAC" trimming potentiometer until the pressure 8.0 × 10-4 hPa appears on the
TPG display.5. Rotate the “HIGH VAC” trimming potentiometer around a 1/4 of a turn in an anticlockwise direc-
tion.– An underrange "ur 10-4" appears on the TPG display.
Performing an ATM calibration1. Operate the gauges for at least 10 minutes at atmospheric pressure.2. Select the measuring circuit you wish to calibrate on the TPG.
– You can find further information on the "Sensor" mode in the TPG operating instructions.3. Rotate the "ATM" trimming potentiometer until the pressure 1.0 × 103 hPa appears on the TPG
display.4. Evacuate the vacuum system to pressure < 10-4 hPa.5. Check the high vacuum value.6. Adjust the high vacuum value if necessary.
Calibrating the Pirani measuring circuit
26/48
7 MalfunctionsWarrantyMalfunctioning of the equipment as a direct result of contamination or wear, as well as wear parts, is not covered by the warranty.
Disturbance Possible cause RemedyGauge cable will not connect to IKR gauge. Old IKR gauge with MHV
plugUse a correct cable.Change the plug.
Gauge cable will not connect with measure-ment board.
Gauge cable has old plug Use a correct cable.1. Disconnect the plug.2. Replace the plug.
Tbl. 6: Disturbances during installation
Disturbance Possible cause RemedyPirani display too high Pirani gauge contaminated Calibrate the Pirani measuring circuit
(see chapter “Calibrating the Pirani measuring circuit”, page 26).Clean the gauge.Replace the gauge.
Cold cathode display too high
Plug insulation is contaminat-ed or damp
● Clean the insulation.● Replace the plug.
Air humidity (leakage current) Keep the air humidity level low.Keep the gauge in continuous operation.
Cold cathode display too low
Cold cathode gauge contami-nated
Clean the gauge.
Pirani will not calibrate. Incorrect combination: Meas-urement board - gauge - cable
Choose the correct combination (see chapter “Connecting Pirani gauges”, page 21).
Gauge too contaminated Clean the gauge.Replace the gauge.
Tbl. 7: Interference during operation and calibration
Disturbance Possible cause RemedyCold cathode permanently displays OR (overrange), although the pressure is within the measurement range.
Short circuit cold cath-ode cable/gauge
Repair the cable/gauge.Replace the cable/gauge.
Cold cathode meas-urement board defec-tive
Replace the cold cathode measure-ment board.
Cold cathode permanently displays U (underrange), al-though the pressure is within the measurement range.
No IKR gauge con-nected
Connect a gauge.
Interruption cold cath-ode cable
Repair the cable.Replace the cable.
Cold cathode gauge failed
Replace the gauge.
Cold cathode meas-urement board defec-tive
Replace the cold cathode measure-ment board.
Tbl. 8: Disturbances as a result of defects
Malfunctions
27/48
Disturbance Possible cause RemedyNo communication Pins 2 and 3 of interface
cable not crossedUse a correct cable.(see chapter “"RS-232" connection on IF 300A”, page 17)(see chapter “"RS-232" connection on IF 300B”, page 17)
Incorrect Baud rate Adjust the Baud rate.Incorrect data format Maintain the data format in accordance with the da-
ta in the TPG basic unit.
Tbl. 9: Disturbances with the RS-232-C interface
Disturbance Possible cause RemedyNo communication Incorrect Baud rate Adjust the Baud rate.
Incorrect data format Maintain the data format in accordance with the data in the TPG basic unit.
Tbl. 10: Disturbances with the RS-422 interface
Disturbance Possible cause RemedyNo communication Incorrect Baud rate set 1) Set the correct Baud rate (19200 Baud).
Incorrect data format Observe the Profibus-DP communication instruc-tions.
Cycle time > 100 ms Incorrect firmware Check the firmware:TPG 300 firmware: 302-654Profibus firmware: V1.5
Incorrect Baud rate set 2) Set the correct Baud rate (19200 Baud).
Tbl. 11: Disturbances with the Profibus-DP interface
1) On TPG 300 basic unit, from firmware 302-654.2) On TPG 300 basic unit, from firmware 302-654.
Malfunctions
28/48
8 Shipping WARNING
Risk of poisoning from contaminated productsWhere products that contain harmful substances are shipped for maintenance or repair purposes, the safety of service personnel is at risk.
► Comply with the instructions for safe shipping.
Decontamination subject to chargePfeiffer Vacuum decontaminates products not clearly declared "Free of contamination" at your expense.
Instructions for safe shipping► Do not ship microbiological, explosive or radioactively contaminated products.► Observe the shipping guidelines for the participating countries and transport companies.► Highlight any potential dangers on the outside of the packaging.► Download the explanation for contamination at Pfeiffer Vacuum Service.► Always enclose a completed declaration of contamination.
Shipping
29/48
9 Recycling and disposal WARNING
Health hazard through poisoning from toxic contaminated components or devicesToxic process media result in contamination of devices or parts of them. During maintenance work, there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub-stances causes environmental damage.
► Take suitable safety precautions and prevent health hazards or environmental pollution by toxic process media.
► Decontaminate affected parts before carrying out maintenance work.► Wear protective equipment.
Environmental protectionYou must dispose of the product and its components in accordance with all applicable reg-ulations for protecting people, the environment and nature.
● Help to reduce the wastage of natural resources.● Prevent contamination.
Environmental protectionThe product and its components must be disposed of in accordance with the applica-ble regulations relating to environmental protection and human health, with a view to reducing natural resource wastage and preventing pollution.
9.1 General disposal informationPfeiffer Vacuum products contain materials that you must recycle.
► Dispose of our products according to the following:– Iron– Aluminium– Copper– Synthetic– Electronic components– Oil and fat, solvent-free
► Observe the special precautionary measures when disposing of:– Fluoroelastomers (FKM)– Potentially contaminated components that come into contact with media
9.2 Disposing of plug-in boardsPfeiffer Vacuum plug-in boards contain materials that you must recycle.
1. Separate the components into recyclable materials.2. Recycle the components.3. Dispose of the product or components in a safe manner according to locally applicable regula-
tions.
Recycling and disposal
30/48
10 Service solutions by Pfeiffer VacuumWe offer first-class serviceHigh vacuum component service life, in combination with low downtime, are clear expectations that you place on us. We meet your needs with efficient products and outstanding service.We are always focused on perfecting our core competence – servicing of vacuum components. Once you have purchased a product from Pfeiffer Vacuum, our service is far from over. This is often exactly where service begins. Obviously, in proven Pfeiffer Vacuum quality.Our professional sales and service employees are available to provide you with reliable assistance, worldwide. Pfeiffer Vacuum offers an entire range of services, from original replacement parts to service contracts.
Make use of Pfeiffer Vacuum serviceWhether preventive, on-site service carried out by our field service, fast replacement with mint condition replacement products, or repair carried out in a Service Center near you – you have various options for maintaining your equipment availability. You can find more detailed information and addresses on our homepage, in the Pfeiffer Vacuum Service section.
You can obtain advice on the optimal solution for you, from your Pfeiffer Vacuum representa-tive.For fast and smooth service process handling, we recommend the following:
1. Download the up-to-date form templates.─ Explanations of service requests─ Service requests─ Contamination declaration
a) Remove and store all accessories (all external parts, such as valves, pro-tective screens, etc.).
b) If necessary, drain operating fluid/lubricant.c) If necessary, drain coolant.
ANFORDERUNG SERVICE
ERKLÄRUNG KONTAMINIERUNG
2. Complete the service request and contamination declaration.
3. Send the forms by email, fax, or post to your local Service Center.
4. You will receive an acknowledgment from Pfeiffer Vacuum.
Submission of contaminated productsNo microbiological, explosive, or radiologically contaminated products will be accepted. Where products are contaminated, or the contamination declaration is missing, Pfeiffer Vacuum will contact you before starting service work. Depending on the product and degree of pollution, additional decontamination costs may be incurred.
Service solutions by Pfeiffer Vacuum
31/48
5. Prepare the product for transport in accordance with the provisions in the contamination declaration.
a) Neutralize the product with nitrogen or dry air.b) Seal all openings with blind flanges, so that they are airtight.c) Shrink-wrap the product in suitable protective foil.d) Package the product in suitable, stable transport containers only.e) Maintain applicable transport conditions.
ERKLÄRUNG KONTAMINIERUNG
6. Attach the contamination declaration to the outside of the packag-ing.
7. Now send your product to your local Service Center.
8. You will receive an acknowledgment/quotation, from Pfeiffer Vac-uum.
Our sales and delivery conditions and repair and maintenance conditions for vacuum devices and com-ponents apply to all service orders.
Service solutions by Pfeiffer Vacuum
32/48
11 AccessoriesView the range of accessories for ModulLine on our website.
11.1 Accessory informationMeasuring boardsFor the connection of Pirani gauges (TPR) and cold cathode gauges (IKR) for measuring atmospheric pressure (TPR) up to UHV (IKR)
Interface boardsSerial and fieldbus interfaces with switchpoints with contacts of varying loading capacity
Measurement cable and extension measurement cableMeasurement cable in standard and high temperature versions in different lengths, for connecting to a Pfeiffer Vacuum total pressure measuring and control unit or an ionization measuring device
Dummy platesCover plate for free slots on Pfeiffer Vacuum total pressure measuring and control units that are not fully configured
11.2 Ordering accessoriesDescription Order number suitable for measurement
boardsPirani gauge head TPR 010, DN 10 ISO-KF PT R02 270 PI 300D, CP 300C9, CP
300C10Pirani gauge head TPR 017, corrosion resistant, DN 16 ISO-KF
PT R13 270 PI 300DN
Pirani gauge head TPR 017, corrosion resistant, DN 16 CF-F
PT R13 271 PI 300DN
Pirani gauge head TPR 018, DN 16 ISO-KF PT R15 010 PI 300D, CP 300C9, CP 300C10
Pirani gauge head TPR 018, DN 16 CF-F PT R15 011 PI 300D, CP 300C9, CP 300C10
Pirani gauge head TPR 018, DN 40 CF-F PT R15 014 PI 300D, CP 300C9, CP 300C10
Tbl. 12: Pirani gauges
Description Order number suitable for measure-ment boards
Cold cathode gauge head IKR 050, coaxial, FPM seal DN 25 ISO-KF
PT R18 500 PE 300DC9, CP 300C9, CP 300C10
Cold cathode gauge head IKR 050, coaxial, FPM seal, DN 40 ISO-KF
PT R18 501 PE 300DC9, CP 300C9, CP 300C10
Cold cathode gauge head IKR 050, coaxial, FPM-Seal, DN 40 CF-F
PT R18 502 PE 300DC9, CP 300C9, CP 300C10
Cold cathode gauge head IKR 060, coaxial, metal seal, DN 40 ISO-KF
PT R18 753 PE 300DC9, CP 300C9, CP 300C10
Cold cathode gauge head IKR 060, coaxial, metal seal, DN 40 CF-F
PT R18 751 PE 300DC9, CP 300C9, CP 300C10
Accessories
33/48
Description Order number suitable for measure-ment boards
Cold cathode gauge head IKR 070, triaxial, metal seal, DN 40 ISO-KF
PT R20 501 CP 300T11, CP 300T11L
Cold cathode gauge head IKR 070, triaxial, metal seal, DN 40 CF-F
PT R20 502 CP 300T11, CP 300T11L
Tbl. 13: Cold cathode gauges
Measuring and extension cableThe cable is available in a standard version for temperatures up to max. 80 °C and as a high temperature table for temperatures up to 250 °C.
4
2
1
5
3
4
2
1
5
Amphenol C 091 Emale
TPR 010: Amphenol C 091 E
femaleTPR 017/018: LEMO
Fig. 15: Pirani measuring cable connection diagram
4
2
1
5
3
4
2
1
5
Amphenol C 091 Emale
Amphenol C 091 Efemale
Fig. 16: Pirani extension cable connection diagram
Description Order numberMeasurement Cable, TPR 010, 2.0 m, 80 °C PT 548 401-TMeasurement Cable, TPR 010, 3.0 m, 80 °C PT 548 402-TMeasurement Cable, TPR 010, 6.0 m, 80 °C PT 548 403-TMeasurement Cable, TPR 010, 10 m, 80 °C PT 548 450-TMeasurement Cable, TPR 010, 15 m, 80 °C PT 548 451-TMeasurement Cable, TPR 010, 20 m, 80 °C PT 548 452-TMeasurement Cable, TPR 010, 25 m, 80 °C PT 548 453-TMeasurement Cable, TPR 010, 30 m, 80 °C PT 548 415-TMeasurement Cable, TPR 010, 35 m, 80 °C PT 548 454-TMeasurement Cable, TPR 010, 40 m, 80 °C PT 548 416-TMeasurement Cable, TPR 010, 45 m, 80 °C PT 548 455-TMeasurement Cable, TPR 010, 50 m, 80 °C PT 548 417-T
Tbl. 14: Measuring cable for TPR 010
Description Order numberMeasurement Cable, TPR 017/018, 3.0 m, 80 °C PT 548 308-THigh-Temperature Measurement Cable, TPR 017/018, 3.0 m, 250 °C PT 548 414-TMeasurement Cable, TPR 017/018, 6.0 m, 80 °C PT 548 309-THigh-Temperature Measurement Cable, TPR 017/018, 6.0 m, 250 °C PT 548 465-TMeasurement Cable, TPR 017/018, 10 m, 80 °C PT 548 456-THigh-Temperature Measurement Cable, TPR 017/018, 10 m, 250 °C PT 448 047-TMeasurement Cable, TPR 017/018, 15 m, 80 °C PT 548 457-THigh-Temperature Measurement Cable, TPR 017/018, 15 m, 250 °C PT 448 043-TMeasurement Cable, TPR 017/018, 20 m, 80 °C PT 548 458-THigh-Temperature Measurement Cable, TPR 017/018, 20 m, 250 °C PT 448 044-TMeasurement Cable, TPR 017/018, 25 m, 80 °C PT 548 459-T
Accessories
34/48
Description Order numberMeasurement Cable, TPR 017/018, 30 m, 80 °C PT 548 460-TMeasurement Cable, TPR 017/018, 35 m, 80 °C PT 548 461-TMeasurement Cable, TPR 017/018, 40 m, 80 °C PT 548 462-TMeasurement Cable, TPR 017/018, 45 m, 80 °C PT 548 463-TMeasurement Cable, TPR 017/018, 50 m, 80 °C PT 548 464-T
Tbl. 15: Measuring cable for TPR 017 and TPR 018
Description Order numberExtension for measurement cable, TPR 010/017/018, 10 m, 80 °C PT 548 466-TExtension for measurement cable, TPR 010/017/018, 20 m, 80 °C PT 548 468-TExtension for measurement cable, TPR 010/017/018, 30 m, 80 °C PT 548 470-TExtension for measurement cable, TPR 010/017/018, 40 m, 80 °C PT 548 472-TExtension for measurement cable, TPR 010/017/018, 50 m, 80 °C PT 548 474-T
Tbl. 16: Extension cable for TPR 010, TPR 017 and TPR 018
Description Order numberMeasurement Cable, IKR 050/060, 3.0 m, 80 °C PT 548 406-THigh-Temperature Measurement Cable, IKR 050/060, 3.0 m, 250 °C PT 548 542-TMeasurement Cable, IKR 050/060, 6.0 m, 80 °C PT 548 407-THigh-Temperature Measurement Cable, IKR 050/060, 6.0 m, 250 °C PT 548 543-TMeasurement Cable, IKR 050/060, 10 m, 80 °C PT 548 419-THigh-Temperature Measurement Cable, IKR 050/060, 10 m, 250 °C PT 448 045-TMeasurement Cable, IKR 050/060, 15 m, 80 °C PT 548 483-THigh-Temperature Measurement Cable, IKR 050/060, 15 m, 250 °C PT 548 989-TMeasurement Cable, IKR 050/060, 20 m, 80 °C PT 548 484-THigh-Temperature Measurement Cable, IKR 050/060, 20 m, 250 °C PT 448 046-TMeasurement Cable, IKR 050/060, 25 m, 80 °C PT 548 485-TMeasurement Cable, IKR 050/060, 30 m, 80 °C PT 548 422-TMeasurement Cable, IKR 050/060, 35 m, 80 °C PT 548 486-TMeasurement Cable, IKR 050/060, 40 m, 80 °C PT 548 487-TMeasurement Cable, IKR 050/060, 45 m, 80 °C PT 548 488-TMeasurement Cable, IKR 050/060, 50 m, 80 °C PT 548 489-TCoax cable with SHV plugs, test voltage: 6 kV DC
Tbl. 17: Measuring cable for IKR 050 and IKR 060
Description Order numberMeasurement Cable, IKR 070, 3.0 m, 80 °C PT 548 306-TMeasurement Cable, IKR 070, 6.0 m, 80 °C PT 548 317-TMeasurement Cable, IKR 070, 10 m, 80 °C PT 548 490-TMeasurement Cable, IKR 070, 15 m, 80 °C PT 548 491-TMeasurement Cable, IKR 070, 20 m, 80 °C PT 548 492-TMeasurement Cable, IKR 070, 30 m, 80 °C PT 548 493-TMeasurement Cable, IKR 070, 40 m, 80 °C PT 548 494-TTriax cable with triax plugs, test voltage: 6 kV DC
Accessories
35/48
Description Order numberMeasurement Cable, IKR 070, 45 m, 80 °C PT 548 495-TMeasurement Cable, IKR 070, 50 m, 80 °C BP 229 748-TTriax cable with triax plugs, test voltage: 6 kV DC
Tbl. 18: Measuring cable for IKR 070
Description Order numberD-sub plug, female, 15-pin, for IF 300 A/C, relay output PT 441 129-TD-sub plug, female, for IF 300 A, RS-232-C PT 441 128-TRelay connector, complete, for IF 300 B PT 546 999-TInterface cable, for IF 300 B, RS-232-C, 0.4 m PT 548 932-T
Tbl. 19: Plugs and interface cable
Description Order numberBlanking Panel, for IF 300 PT 441 017Blanking Panel, for Measurement Board PT 441 259
Tbl. 20: Dummy plates
Accessories
36/48
12 Technical data and dimensions
12.1 Technical dataParameter Value/designationMeasuring board PI 300D for Pirani gauges with wolfram fila-
mentPI 300DN for Pirani gauges with nickel fila-
mentNumber of measuring circuits 2Measuring range (N2 equivalent) 1000 hPa to 8 × 10-4 hPaMeasuring cable length, max. 100 mSignal output Measurement value, analog 0 to +10 V
Error message (error) > 11.5 VCurrent, max. 2 mAOutput resistance 400 Ω
Response time (10 to 90 %) with rapid pressure change
< 10-3 to 103 hPa < 50 ms103 to < 10-3 hPa < 600 ms
Connection, device side TPR gauge 6-pin Amphenol C 091 B socketSignal output Ø 2 mm sockets
Weight approx. 0.14 kg
Tbl. 21: Technical data PI 300D / PI 300DN
Parameter Value/designationNumber of measuring circuits 2Measuring range (N2 equivalent) 5 × 10-3 hPa to 1 × 10-9 hPaMeasuring cable length, max. 60 3) up to 100 mPower supply for IKR gauges Operating voltage 3.3 kV
Measurement current, max. 100 μASignal output Measurement value, analog 0 to +10 V
Error message (error) > 11.5 VCurrent, max. 2 mAOutput resistance 400 Ω
Response time (10 to 90 %) with rapid pressure change
< 10-9 to 103 hPa < 20 ms103 to < 10-9 hPa < 20 ms
Connection, device side IKR gauge SHV, coax, socketSignal output 7-pin Amphenol C 091 B sock-
etWeight approx. 0.26 kg
Tbl. 22: Technical data PE 300DC9
3) with use of the lower measurement range limit
Technical data and dimensions
37/48
Parameter Value/designationNumber of measuring circuits 1 each ofMeasuring range (N2 equivalent) Pirani 1000 hPa to 8 × 10-4 hPa
Cold cathode CP 300C95 × 10-3 hPa to 5 × 10-9 hPaCP 300C105 × 10-3 hPa to 1 × 10-10 hPaCP 300T11/T11L5 × 10-3 hPa to 1 × 10-11 hPa
Measuring cable length, max. Pirani CP 300C9/C10/T11100 mCP 300T11L500 m
Cold cathode CP 300C9/C1060 4) up to 100 mCP 300T11/T11L500 m
Power supply for IKR gauge Operating voltage 3.3 kVMeasurement current, max.
CP 300C9600 μACP 300C10/T11/T11L100 μA
Signal output Measurement value, an-alog
0 to +10 V
Error message (error) > 11.5 VCurrent, max. 2 mAOutput resistance 400 Ω
Response time (10 to 90 %) with rapid pressure change
Pirani< 10-3 to 103 hPa < 50 ms103 to < 10-3 hPa < 600 msCold cathode< 10-9 to 103 hPa CP 300C9
< 10 msCP 300C10/T11/T11L< 50 ms
103 to < 10-9 hPa CP 300C9< 10 msCP 300C10/T11/T11L< 50 ms
4) with use of the lower measurement range limit
Technical data and dimensions
38/48
Parameter Value/designationConnection, device side TPR gauge 6-pin Amphenol C 091 B socket
IKR gauge CP 300C9/C10SHV, coax, socketCP 300T11/T11Ltriax, socket
Signal output Ø 2 mm socketsWeight CP 300C9
approx. 0.21 kgCP 300C10approx. 0.23 kgCP 300T11/T11Lapprox. 0.25 kg
Tbl. 23: Technical data CP 300C9/C10/T11/T11L
Parameter Value/designationRelay Quantity, number 5
Contacts 1 x switching contact, potential-free, max. 50 V AC against other contacts and against ground
AC characteristics Switching voltage, max. 50 V ACSwitching current, max. 1.5 ASwitching output, max. 75 VA
DC characteristics Switching current, max. Switching voltages > 50 V DC are not permitted for safety reasons.0.6 A with 50 V DC0.8 A with 40 V DC1.5 A with 30 V DC
Switching output, max. -Connection, device side D-Sub plug, 15-pin, pinsElectrical resistance with plug 125 mΩInterface Type RS-232-C, asynchronous
Baud rates 300, 1200, 2400, 4800, 9600Data format ASCII, 1 start bit, 8 data bits, 1 stop bit, no parity bitConnection, device side D-Sub plug, 9-pin, pinsCable length, max. 30 m
Weight approx. 0.14 kg
Tbl. 24: Technical data IF 300A
Parameter Value/designationRelay Quantity, number 5
Contacts 1 x switching contact, potential-free, max. 250 V AC against other contacts and against ground
AC characteristics Switching voltage, max. 250 V ACSwitching current, max. 4 ASwitching output, max. 1000 VA
Technical data and dimensions
39/48
Parameter Value/designationDC characteristics Switching current, max. 0.25 A with 200 V DC
0.3 A with 140 V DC0.4 A with 100 V DC0.5 A with 60 V DC0.6 A with 50 V DC0.8 A with 40 V DC4.0 A with 30 V DC
Switching output, max. 120 WConnection, device side GdsA-H, DIN 41612, 15-pin, pinsElectrical resistance with plug 70 mΩInterface Type RS-232-C, asynchronous
Baud rates 300, 1200, 2400, 4800, 9600Data format ASCII, 1 start bit, 8 data bits, 1 stop bit, no parity bitConnection, device side 0.4 m cable with D-Sub plug 25-pin, pinsCable length, max. 30 m
Weight approx. 0.15 kg
Tbl. 25: Technical data IF 300B
Parameter Value/designationRelay Quantity, number 5
Contacts 1 x switching contact, potential-free, max. 50 V AC against other contacts and against ground
AC characteristics Switching voltage, max. 50 V ACSwitching current, max. 1.5 ASwitching output, max. 75 VA
DC characteristics Switching current, max. Switching voltages > 50 V DC are not permitted for safety reasons.0.6 A with 50 V DC0.8 A with 40 V DC1.5 A with 30 V DC
Switching output, max. -Connection, device side D-Sub plug, 15-pin, pinsElectrical resistance with plug 125 mΩInterface Type RS-422, asynchronous
Baud rates 300, 1200, 2400, 4800, 9600Data format ASCII, 1 start bit, 8 data bits, 1 stop bit, no parity bitConnection, device side D-Sub plug 9-pin, socketsCable length, max. 1200 m
Weight approx. 0.14 kg
Tbl. 26: Technical data IF 300C
Details on the Profibus-DP interfaceYou can find details on the Profibus-DP interface in the corresponding communication pro-tocol.
Parameter Value/designationRelay Quantity, number 5
Contacts 1 x switching contact, potential-free, max. 50 V AC against other contacts and against ground
Technical data and dimensions
40/48
Parameter Value/designationAC characteristics Switching voltage, max. 50 V AC
Switching current, max. 1.5 ASwitching output, max. 75 VA
DC characteristics Switching current, max. Switching voltages > 50 V DC are not permitted for safety reasons.0.6 A with 50 V DC0.8 A with 40 V DC1.5 A with 30 V DC
Switching output, max. -Connection, device side D-Sub plug, 15-pin, pinsElectrical resistance with plug 125 mΩInterface Type Profibus DP
Baud rates ≤ 12 MBaudData format (see communication protocol)Connection, device side D-Sub plug 9-pin, socketsCable length, max. (see communication protocol)
Weight approx. 0.16 kg
Tbl. 27: Technical data IF 300P
12.2 DimensionsDimensions in mm.
11
2 (
4.4
1“)
30 (1.18“)
Fig. 17: Dimensions of measurement boards
Technical data and dimensions
41/48
11
2 (
4.4
1“)
20 (0.79“)
300PIF IF 300B
Fig. 18: Dimensions of interface and relay boards
Technical data and dimensions
42/48
13 Appendix
13.1 Relation: measuring signal and pressurep [hPa]
UA [V]
105
0 1 2 3 4 5 6 7 8 9 10 11
104
103
102
101
100
10–1
10–2
10–3
10–4
TPR 010 / TPR 018 with PI 300D
TPR 017 with PI 300DN
Fig. 19: Relation: Measuring signal and pressure with TPR gauges
p Pressure UA Measuring signal [V] (output voltage)
Appendix
43/48
IKR 050/060 mit CP 300C9IKR 050/060 mit CP 300C10
IKR 070 mit PE 300T11IKR 050/060 mit PE 300DC9
p [hPa]
UA [V]
10-2
0 1 2 3 4 5 6 7 8 9 10 11
10-3
10-4
10-5
10-6
10-7
10–8
10–9
10–10
10–11
p =
10 U
–10.5
Fig. 20: Relation: Measuring signal and pressure with IKR gauges
p Pressure UA Measuring signal [V] (output voltage)
13.2 Gas type dependenceThe measuring signal is gas type-dependent. The characteristics apply for nitrogen (N2), oxygen (O2), dry air and carbon monoxide (CO).
Appendix
44/48
102
86
4
2
101
86
4
2
100
86
4
2
10–1
86
4
2
10–2
86
4
2
10–3
10–3 2 4 6 10–2 2 4 6 10–1 2 4 6 100 2 4 6 101 2 4 6 102
H2 He Ne
Xe
Kr
Ar
CO2
Freon 12
AirO2
CON2
p [hPa]
peff [hPa]
Water vapor
Fig. 21: Displayed pressure for TPR gauges
4
2
10–3
86
4
2
10–4
86
4
2
10–5
86
4
2
10–6
86
4
2
10–7
p [hPa]
peff [hPa]
10–7 2 4 6 10–6 2 4 6 10–5 2 4 6 10–4 2 4 6 10–3 2 4 6 10–2
Xe Kr Ar
AirO2
CO
N2 H2 Ne He
Fig. 22: Displayed pressure for IKR gauges
Appendix
45/48
13.3 Units of pressureUnit mbar bar Pa hPa kPa Torr / mm Hgmbar 1 1 · 10-3 100 1 0.1 0.75bar 1000 1 1 · 105 1000 100 750Pa 0.01 1 · 10-5 1 0.01 1 · 10-3 7.5 · 10-3
hPa 1 1 · 10-3 100 1 0.1 0.75kPa 10 0.01 1000 10 1 7.5Torr / mm Hg 1.33 1.33 · 10-3 133.32 1.33 0.133 11 Pa = 1 N/m2
Tbl. 28: Units of pressure and their conversion
13.4 Gas throughputsUnit mbar l/s Pa m3/s sccm Torr l/s atm cm3/smbar l/s 1 0.1 59.2 0.75 0.987Pa m3/s 10 1 592 7.5 9.87sccm 1.69 · 10-2 1.69 · 10-3 1 1.27 · 10-2 1.67 · 10-2
Torr l/s 1.33 0.133 78.9 1 1.32atm cm3/s 1.01 0.101 59.8 0.76 1
Tbl. 29: Gas throughputs and their conversion
Appendix
46/48
Appendix
47/48
*BG5972*
ed. C
- D
ate
1911
- P/
N:B
G59
72BE
N