Online Solid Phase Extraction and LC/MS/MS Detection of Thyroid Hormones in Biological Fluids

Xiaoning Lu and David S. BellSupelco, Div. of Sigma-AldrichBellefonte, PA 16823 USA

T414067

© 2014 Sigma-Aldrich Co. All rights reserved.

2

Introduction

Thyroid hormones play critical roles in the regulation of biological processes such as growth, metabolism, protein synthesis, and brain development. Specifically, thyroid hormones, 3,3’,5,5’-tetraiodo-L-thyronine (thyroxine or T4) and 3,3’,5-triiodo-L-thyronine (T3), are essential for development and maintenance of normal physiological functions. For a clinical laboratory, measurements of total T4 and total T3, along with estimates of free T4 (FT4) and free T3 (FT3), are important for the diagnosis and monitoring of thyroid diseases. Most clinical laboratories measure thyroid hormones using immunoassays. The immunoassay-based methods offer a relatively rapid, high patient sample throughput that lends itself to automation but are significantly compromised by problems with assay interference and are perturbed by changes in protein levels that alter the free hormone availability (1).

© 2014 Sigma-Aldrich Co. All rights reserved.

3

Introduction (contd.)

Liquid chromatography mass spectrometry (LC/MS) has been reported to offer superior specificity and speed over the immunoassays for determination of thyroid hormones in biological matrices such as serum and tissues. Nevertheless, the reported sample preparation procedures, typically by liquid-liquid extraction followed by solid phase extraction (SPE), involve multiple time consuming steps, and are less compatible with automation (2-3). The present work exploited online SPE with LC/MS for rapid determination of T4, T3 and 3,3′,5′-triiodo-L-thyronine (rT3) from biological matrices.

© 2014 Sigma-Aldrich Co. All rights reserved.

4

Experiment

Figure 1. Chemical Structures of the Thyroids Analytes in this Study

Thyroid compoundsOH

II

NH2 COOHH

OII

Thyroxine (T4), C15H11I4NO4, Monoisotopic

mass: 776.686 Da

OH

I

NH2 COOHH

OII

3, 3', 5-Triiodothyronine (T3), C15H12I3NO4,

Monoisotopic mass: 650.790 Da

OH

I

NH2 COOHH

OI

I

3, 3', 5'-Triiodothyronine (Reverse T3), C15H12I3NO4,

Monoisotopic mass: 650.790 Da

© 2014 Sigma-Aldrich Co. All rights reserved.

5

Sample Processing Procedure

A 100 µL of rabbit plasma spiked with a desired amount of T4, T3 and rT3, were protein precipitated by addition of 25 µL zinc chloride and 200 µL methanol, agitated by vortex for 1 min, and then centrifuged at 9000 g x 3 min. The resulting supernatant was collected and directly injected for LC/MS analysis.

Online SPE-LC/MSinstrument: Shimadzu® LCMS-8030

trapping column: Ascentis® Express RP-Amide guard cartridge, 5 x 2.1 mm, (53514-U) or Ascentis Express C8 guard cartridge, 5 x 2.1 mm, (53509-U)

HPLC column: Ascentis Express Phenyl Hexyl, 5 cm x 2.1 mm (53334-U)mobile phase: (A) water; (B) methanol, both with 0.1% acetic acid

column pressure: 1800 psi (88 bar)det.: MS, ESI(+), MRM mode (777.7/731.8 for T4; 651.8/605.5 for T3 or rT3).

injection: 5 µL

© 2014 Sigma-Aldrich Co. All rights reserved.

6

Figure 2a. Configuration of the LC/MS System and Valve Positions for Online SPE-LC/MS

© 2014 Sigma-Aldrich Co. All rights reserved.

7

Figure 2b. A Representative Online SPE-LC/MS Program

© 2014 Sigma-Aldrich Co. All rights reserved.

8

Column Screening for Separation of ThyroidsResults and Discussion

Table 1. Ascentis Express Columns (5 cm x 2.1 mm I.D.,2.7 µm) Screened in this Study

Column Phase Sigma-Aldrich Part No.C18 C18 53822-U F5 pentafluorophenylpropyl 53567-U

Phenyl Phenyl-Hexyl 53334-U

LC-UV conditions:instrument: Agilent® 1200 HPLC-DAD

HPLC column: Ascentis Express Columnsmobile phase: (A) water; (B) methanol, both with 0.1% acetic acid

isocratic: 70% Bflow rate: 0.2 mL/min

temp.: 35 Cdet.: UV 210 nm with DAD scan

column pressure: 1600-1750 psi injection: 1 µL

© 2014 Sigma-Aldrich Co. All rights reserved.

9

Figure 3a. Comparison of Separation of Thyroids by Different Phases

• All the tested phases have adequate selectivity to resolve all of the 3 thyroids.

• The C18 gives least retention. • F5 and Phenyl offer better

selectivity than C18.• Phenyl-Hexyl phase tend to

give better peak shapes than C18 and F5, and best selectivity. Therefore, it was selected for further method development.

C18

F5

Phenyl

© 2014 Sigma-Aldrich Co. All rights reserved.

10

Figure 3b. Separation of Thyroids by Phenyl-Hexyl Phase under Optimized LC Conditions

T3rT3 T4

mobile phase: (A) water; (B) methanol, both with 0.1% acetic acid

isocratic: 65% B

© 2014 Sigma-Aldrich Co. All rights reserved.

11

Online SPE-LC/MS Method

As shown in Figure 2a, the online SPE-LC/MS system comprises a trapping column, a switching valve, and an LC/MS instrument. Figure 2b shows a typical online SPE-LC/MS program, which consists of three distinct steps. The valve switches between positions 1 for loading/washing (Step 1) and equilibration (Step 3), and position 2 for LC/MS analysis (Step 2), respectively. It should be noted the online SPE and LC/MS analysis use the same pumping system. In the Step 1, samples are loaded and washed with the mobile phase containing low organic (5% methanol in this case) and at higher flow rate where the salts or other interferences in the samples are washed off and directed to the waste. Subsequently (In Step 2), the samples are eluted, separated and detected by the LC/MS at optimum flow (0.25 mL/min) for both chromatographic separation and MS signals. Lastly, the system returns back to valve position 1 for re-equilibration under the sample loading/washing conditions.

© 2014 Sigma-Aldrich Co. All rights reserved.

12

Choice of Online SPE Trapps

The choice of the right trapping column is critical for the successful online SPE-LC/MS method development as it affects both trapping efficiency (e.g. the recovery of the analytes) and the downstream LC separation. As a rule of thumb, the trapping column should be less retentive than the analytical column to avoid possible peak broadening. Two guard cartridges (traps), RP-Amide and C8, of the same dimensions, were evaluated and compared. As can be seen in Figure 3, both trapping columns led to comparable sharp peak shape (separation efficiency). However, the RP-Amide delivered slightly (15%) higher peak responses than the C8 trap under the trapping conditions with 5% methanol as the washing solvent. Further increasing methanol content to 20% in the washing solvent resulted in significant (about 50%) signal decrease with C8 traps, but almost no change in signals with RP-Amide traps (See Figures 4a and 4b). These results indicate better washing can be achieved with minimal sample loss with RP-Amide traps.

© 2014 Sigma-Aldrich Co. All rights reserved.

13

Figure 4. Comparison of the LC/MS Chromatograms with Online Trapping with C8 (4a) and RP-Amide (4b) Guard Cartridges (Under the Conditions with 5% Methanol as the Washing Solvent)

Online SPE-LC/MS conditions as in Figure 2.

(4b) RP-Amide Trap

Note the higher peaks with RP-Amide trap.

4.0 5.0 6.0 7.0 8.0Time (min)

010

020

030

0

Inte

nsity

/ cp

s

T3

T4rT3

4.0 5.0 6.0 7.0 8.0Time (min)

010

020

030

0

Inte

nsity

/ cp

s

T3

T4

rT3

(4a) C8 Trap

© 2014 Sigma-Aldrich Co. All rights reserved.

144.0 5.0 6.0 7.0 8.0Time (min)

010

020

030

0

Inte

nsity

/ cp

s

4.0 5.0 6.0 7.0 8.0Time (min)

010

020

030

0

Inte

nsity

/ cp

s

T3

T4

rT3

Note the much lower response with C8 trap under 20% MeOH washing conditions.

T3

T4

rT3

(5a) C8 Trap

(5b) RP-Amide Trap

Figure 5. Comparison of the LC/MS Chromatograms with Online Trapping with C8 (5a) and RP-Amide (5b) Guard Cartridges (Under the Conditions with 20% Methanol as the Washing Solvent, 5c)

© 2014 Sigma-Aldrich Co. All rights reserved.

15

(5c)

Figure 5. (contd.)

© 2014 Sigma-Aldrich Co. All rights reserved.

16

Conclusion

An online SPE-LC/MS method was developed for the determination of thyroid hormones in biological fluids. The method exploits RP-Amide or C8 as the trapping column and Phenyl phase as the separation column, respectively. The preliminary experiments demonstrated that, under the optimized conditions, both RP-Amide and C8 effectively trapped the thyroid hormones extracted from spiked rabbit plasma sample which had been protein precipitated. ith both traps, sharp peak shapes were observed. However, the RP-Amide traps are advantageous over the commonly used C8 traps with higher signals and recoveries. Additionally, RP-Amide traps are more flexible with washing solvent as it is compatible with 100% aqueous mobile phases and can be used with up to 20% methanol with minimal sample loss.

© 2014 Sigma-Aldrich Co. All rights reserved.

17

References

1. Kahric-Janicic N, Soldin SJ, Soldin OP, West T, Gu J, Jonklaas J, Tandem mass spectrometry improves the accuracy of free thyroxine measurements during pregnancy. Thyroid. 2007;17(4): 303-11.

2. Susan S-C. Taia, Lorna T. Sniegoski and Michael J. Welch, Candidate Reference Method for Total Thyroxine in Human Serum Use of Isotope-Dilution Liquid Chromatography-Mass Spectrometry with Electrospray Ionization. Clinical Chemistry, 2002; 48(4): 637-642.

3. Dongli Wang and Heather M. Stapleton, Analysis of thyroid hormones in serum by liquid chromatography-tandem mass spectrometry. Anal BioanalChem. 2010; 397(5): 1831–1839.

Ascentis is a registered trademark of Sigma-Aldrich Co. LLC. Agilent is a registered trademark of Agilent Technologies, Inc. Shimadzu is a registered trademark of Shimadzu Corporation.