Om Pcr Eleva v1.1 en (4512 987 12111 Rev Af)
Transcript of Om Pcr Eleva v1.1 en (4512 987 12111 Rev Af)
-
English
PCR Eleva
4512
987
121
11 A
F/7
32 *
200
7-12
en
Instructions for Use
Version 1.1
-
PCR Eleva Version 1.1
4512
987
121
11
en
PCR Eleva Version 1.1
4512
987
121
11
en
PCR Eleva Version 1.145
12 9
87 1
2111
e
n
PCR Eleva Version 1.1
4512
987
121
11
en
PCR Eleva Version 1.1
4512
987
121
11
en
PCR Eleva Version 1.1
4512
987
121
11
en
PCR Eleva Version 1.1
4512
987
121
11
en
-
PPCR Eleva* n T U S V D U J P n T G P S 6 T e
Version 1.1
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
-
PIJlJQT .eEJDal 4ZTUeNT JT QaSU PG RPZalPIJlJQT EleDUSPOJDT
XXXNeEJDBMQIJMJQTDPNNeEJDBM!QIJMJQTDPN
Manufacturing address1IJMJQT .eEJDBM 4ZTUeNT %.$ (NC)3nUHenTUSBe 24225 )BNCVSH(eSNBnZ
,POJOLlJKLe PIJlJQT EleDUSPOJDT /7 "ll SJHIUT aSe SeTeSveE ReQSPEVDUJPO PS USaOTNJTTJPOJO XIPle PS JO QaSU JO aOZ GPSN PS CZ aOZ NeaOTeleDUSPOJD NeDIaOJDal PS PUIeSXJTe JT QSPIJCJUeEXJUIPVU UIe QSJPS XSJUUeO DPOTeOU PG UIe DPQZSJHIUPXOeS
5IJT .eEJDal %evJDe NeeUT UIe QSPvJTJPOT PG UIeUSaOTQPTJUJPO PG UIe .eEJDal %evJDe %JSeDUJve EEC XJUIJO UIe DPVOUSZ PG UIe E6 "VUIPSJ[eEReQSeTeOUaUJve GPS UIe EevJDe DPODeSOeE
CPQZSJHIUT aOE all PUIeS QSPQSJeUaSZ SJHIUT JO aOZTPGUXaSe aOE SelaUeE EPDVNeOUaUJPO i4PGUXaSew
NaEe avaJlaCle UP ZPV SeTU eYDlVTJvelZ XJUI PIJlJQT PSJUT lJDeOTPST /P UJUle PS PXOeSTIJQ JO UIe 4PGUXaSe JTDPOGeSSeE UP ZPV 6Te PG UIe 4PGUXaSe JT TVCKeDU UP UIeeOE VTeS lJDeOTe DPOEJUJPOT aT aSe avaJlaCle PO SeRVeTU
5P UIe NaYJNVN eYUeOU QeSNJUUeE CZ laX ZPV TIallOPU EeDPNQJle aOEPS SeveSTe eOHJOeeS UIe TPGUXaSe PSaOZ QaSU UIeSePG
PSJOUeE JO (eSNaOZ "' eO
-
5BCMe PG $PnUenUT
1 8IBUhT neX 1
2 *nUSPEVDUJPn 2
2.1 The PCR Eleva ................................................................. 2-3
2.2 About these Instructions for Use ....................................... 2-3
2.3 Normal use ....................................................................... 2-4
2.4 Prohibited use ................................................................... 2-4
2.5 Conformity ....................................................................... 2-52.5.1 Compliance with legal requirements ................... 2-52.5.2 Recycling ........................................................... 2-52.5.3 Mercury (USA only) .......................................... 2-52.5.4 Perchlorate ......................................................... 2-52.5.5 RoHS (China only) ............................................ 2-5
2.6 Training ............................................................................ 2-6
4BGeUZ JnTUSVDUJPnT
3.1 Safety ................................................................................ 3-3
3.2 Electrical safety .................................................................. 3-3
3.3 Mechanical safety .............................................................. 3-4
3.4 Fire safety .......................................................................... 3-4
3.5 Laser radiation warning ..................................................... 3-4
3.6 Special safety information ................................................. 3-43.6.1 Loss of Data Due to Power Outage .................... 3-43.6.2 Inferior Quality Due to Multiple Exposure
............................................................................ 3-43.6.3 Damage as a result of incorrect cassette insertion
............................................................................ 3-43.6.4 Data inconsistency ............................................. 3-53.6.5 Using Lead Letters ............................................. 3-5
3.7 Electromagnetic Compatibility .......................................... 3-5
3.8 Error messages ................................................................... 3-6
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 5
-
4 4ZTUeN EeTDSJQUJPn 4
4.1 System components ........................................................... 4-34.1.1 The basic system ................................................ 4-34.1.2 Possible configurations ....................................... 4-3
4.2 Standard workflow ............................................................ 4-5
4.3 The operators console ....................................................... 4-8
4.4 PCR plate reader ............................................................. 4-104.4.1 The PCR Compano plate reader ...................... 4-104.4.2 The PCR Corado, CosimaX and S Hi-res plate
readers .............................................................. 4-124.4.3 The PCR S and S Plus plate readers ................. 4-144.4.4 The PCR AC500 plate reader ......................... 4-184.4.5 The PCR AC5000 plate reader ....................... 4-224.4.6 The PCR Cosima plate reader .......................... 4-26
5 4XJUDIJnH UIe TZTUeN PnPGG 5
5.1 Switching on ..................................................................... 5-3
5.2 Switching off ..................................................................... 5-35.2.1 One-console system ............................................ 5-35.2.2 Multiple-console system ..................................... 5-4
1$3 QMBUe SeBEeS
6.1 The PCR Compano plate reader ....................................... 6-36.1.1 Reading out an image plate ................................ 6-36.1.2 Performing secondary erasure ............................. 6-76.1.3 Performing primary erasure ............................... 6-86.1.4 Switching on/off ................................................ 6-8
6.2 The PCR Corado, CosimaX and S Hi-res plate readers ......................................................................................... 6-116.2.1 Distinguishing features ..................................... 6-116.2.2 Switching on/off .............................................. 6-116.2.3 Reading out an image plate .............................. 6-136.2.4 Performing secondary erasure ........................... 6-176.2.5 Performing primary erasure ............................. 6-176.2.6 What to do if the cassette is not scanned .......... 6-18
6.3 The PCR S and S Plus plate readers ................................ 6-196.3.1 Distinguishing features ..................................... 6-196.3.2 Switching on/off .............................................. 6-196.3.3 Reading out an image plate .............................. 6-216.3.4 Performing secondary erasure ........................... 6-256.3.5 Performing primary erasure .............................. 6-26
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
6.4 The PCR AC500, AC5000 and Cosima plate readers ......................................................................................... 6-286.4.1 Distinguishing features ..................................... 6-286.4.2 Switching on/off .............................................. 6-286.4.3 Insert the cassette ............................................. 6-316.4.4 Removing the cassette ...................................... 6-346.4.5 Screen saver ...................................................... 6-346.4.6 Working with information windows ................ 6-346.4.7 Processing a cassette with priority ..................... 6-356.4.8 Performing secondary erasure ........................... 6-376.4.9 Performing primary erasure .............................. 6-386.4.10 Adjusting gray levels manually .......................... 6-406.4.11 Processing unregistered image plates ................. 6-416.4.12 Utilities ............................................................ 6-436.4.13 Error messages .................................................. 6-58
6.5 Making exposures ............................................................ 6-606.5.1 Exposure parameters ........................................ 6-606.5.2 Positioning the cassette ..................................... 6-60
6.6 Cassettes and image plates .............................................. 6-616.6.1 Handling cassettes and image plates ................. 6-61
6.7 Notes on radiography ...................................................... 6-626.7.1 UNIQUE, UM and DRR ................................ 6-626.7.2 Notes on radiography techniques with UNIQUE
.......................................................................... 6-636.8 The Exposure Index ........................................................ 6-63
7 5Ie PQeSBUPST DPnTPMe 7
7.1 About this Chapter ............................................................ 7-3
7.2 What the operators console can do ................................... 7-3
7.3 How to use the operators console ..................................... 7-57.3.1 The Touch Screen .............................................. 7-57.3.2 Selecting the Section .......................................... 7-57.3.3 Scrolling Through Lists ...................................... 7-57.3.4 Operating the Slide Controls .............................. 7-67.3.5 Operating Drop-Down Lists .............................. 7-67.3.6 Operating the Selection Fields ............................ 7-77.3.7 Operating Switches ............................................ 7-77.3.8 Setting the on-screen keyboard ........................... 7-8
7.4 In an emergency ................................................................ 7-9
8 1BUJenU BENJnJTUSBUJPn 8
8.1 The Patient list .................................................................. 8-3
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5BCMe PG $PnUenUT
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 7
-
8.2 Entering patient and examination data .............................. 8-58.2.1 Entering patient data .......................................... 8-58.2.2 Adding a Study .................................................. 8-78.2.3 Adding Examination Types ................................ 8-78.2.4 Deleting a study ................................................. 8-88.2.5 Deleting examination types ................................ 8-88.2.6 Customizing the Patient List .............................. 8-98.2.7 Finding a Patient in the Patient List ................ 8-108.2.8 Changing Patient and Examination Data ......... 8-108.2.9 Deleting a patient's data from the Patient List
.......................................................................... 8-118.3 Getting patient data from RIS ......................................... 8-13
8.3.1 Updating the patient list .................................. 8-138.3.2 Getting individual examinations from RIS (only via
DICOM RIS) .................................................. 8-13
9 1eSGPSNJnH UIe eYBNJnBUJPn 9
9.1 The Examination section ................................................... 9-3
9.2 Overview ........................................................................... 9-4
9.3 Examination procedure Overview .................................. 9-5
9.4 Examination steps in detail ................................................ 9-89.4.1 Assigning cassettes .............................................. 9-89.4.2 Reading out image plates on the plate reader
............................................................................ 9-99.4.3 Checking image quality on the operators console
............................................................................ 9-99.4.4 Exporting exposures, printing and completing the
examination ..................................................... 9-119.4.5 Making more than one exposure with the same view
.......................................................................... 9-129.5 Workflow control ............................................................ 9-14
1 3eWJeX JNBHe QSPDeTTJnH 1
10.1 The Review section ......................................................... 10-3
10.2 Exporting Images to an External DICOM Destination ......................................................................................... 10-5
10.3 Image processing tool functions ...................................... 10-5
10.4 Individual image processing tools .................................... 10-610.4.1 Calling up image processing tools ..................... 10-610.4.2 Setting Contrast and Brightness ....................... 10-710.4.3 Setting image processing protocols ................... 10-810.4.4 Setting image processing parameters ................. 10-910.4.5 Rotating and Flipping an Image ..................... 10-1110.4.6 Annotating an image ...................................... 10-11
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
8 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
10.4.7 Shutter Function ............................................ 10-1210.4.8 Scaling ........................................................... 10-1510.4.9 Moving images between examinations ............ 10-1610.4.10 Stitching ........................................................ 10-1710.4.11 Measurements ................................................ 10-2310.4.12 Resetting image .............................................. 10-2410.4.13 Customizing image processing tools ............... 10-24
11 1SJnU 11
11.1 Overview ......................................................................... 11-311.1.1 The Print Section ............................................. 11-311.1.2 The Patient Information Window .................... 11-511.1.3 The Image Memory ......................................... 11-511.1.4 The Preview Window ....................................... 11-511.1.5 The template editor .......................................... 11-811.1.6 The Text Box Editor ...................................... 11-1111.1.7 Activating Print Section ................................. 11-1311.1.8 Editing Images and Viewports ........................ 11-1411.1.9 The Tools for Free Composing ...................... 11-1511.1.10 Extras ............................................................. 11-1611.1.11 Using the template editor ............................... 11-17
12 J4JUe 12
12.1 What is iSite? .................................................................. 12-3
12.2 Log in ............................................................................. 12-4
1 4ZTUeN BENJnJTUSBUJPn BnE DVTUPNJ[BUJPn 1
13.1 The System section ......................................................... 13-3
13.2 Adjusting the touch screen .............................................. 13-813.2.1 Opening the setting screen ............................... 13-813.2.2 Brightness ........................................................ 13-813.2.3 Calibration ....................................................... 13-8
14 4eSWJDe DMeBnJnH EJTQPTBM 14
14.1 Service ............................................................................. 14-314.1.1 Safety Checks ................................................... 14-314.1.2 Repairs ............................................................. 14-414.1.3 Recording Results ............................................. 14-414.1.4 Obligations of the user ..................................... 14-4
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5BCMe PG $PnUenUT
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 9
-
14.2 Tests by the user .............................................................. 14-4
14.3 Cleaning .......................................................................... 14-514.3.1 Cleaning the System ......................................... 14-514.3.2 Cleaning the PCR plate reader ......................... 14-514.3.3 Cleaning the image plates ................................. 14-8
14.4 Disposal ........................................................................ 14-1014.4.1 Recycling ....................................................... 14-1014.4.2 RoHS requirements (China) .......................... 14-10
15 5eDInJDBM EBUB 15
15.1 General data .................................................................... 15-315.1.1 Ambient conditions and power supply ............. 15-315.1.2 Operators console ............................................ 15-315.1.3 Computer ........................................................ 15-415.1.4 The PCR Corado, CosimaX and S Hi-res plate
readers .............................................................. 15-415.1.5 The PCR S and S Plus plate readers ................. 15-615.1.6 The PCR AC500 plate reader .......................... 15-715.1.7 The PCR AC5000 plate reader ........................ 15-815.1.8 The PCR Cosima plate reader .......................... 15-915.1.9 PCR Compano plate reader ........................... 15-10
15.2 Electromagnetic compatibility (EMC) data ................... 15-1115.2.1 Electromagnetic emission ............................... 15-1115.2.2 Electromagnetic interference immunity
recommended safety distances ........................ 15-1215.3 Labels ............................................................................ 15-13
15.3.1 Operators console .......................................... 15-1315.3.2 PCR Compano plate reader ........................... 15-1415.3.3 The PCR Corado, CosimaX and S Hi-res plate
readers ............................................................ 15-1515.3.4 The PCR S and S Plus plate readers ............... 15-1515.3.5 The PCR AC500 plate reader ........................ 15-1615.3.6 The PCR AC5000 plate reader ...................... 15-1715.3.7 The PCR Cosima plate reader ........................ 15-18
1 "QQenEJY 1
16.1 Notes on Radiography Techniques with UM and DRR ......................................................................................... 16-3
16.2 If the barcode reader is not working .............................. 16-11
16.3 Error messages ............................................................... 16-1216.3.1 What you should know .................................. 16-1216.3.2 Type 1 messages ............................................. 16-1316.3.3 Type 2 messages ............................................. 16-2016.3.4 Type 3 messages ............................................. 16-2316.3.5 Type 4 messages ............................................. 16-25
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
1 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
17 (MPTTBSZ 17
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5BCMe PG $PnUenUT
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 11
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
12 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
5BCMe PG $PnUenUT
1 8IBUhT neX 1
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 11
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
12 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
1 8IBUhT neX
Software Rel. 1.1 offers a number of new functions over earlier versions.As a quick guide, the most important ones are explained in brief below;for further information please go the chapter you are referred to.
CContrast and brightness
The buttons for contrast and brightness are now on the monitor itself(chapter 10.4.2).
Turning a function on/off.
You can change contrast and brightness using the buttons or on the monitor itself, using your finger.
Moving images between examinations
You can move images between two examinations on the same patient between examinations on different patients.Exposure Index EI_s
Using the Exposure Index, you can compare similar exposures of thesame organ with the same radiation quality in terms of their dose, andanalyze deviations from the optimum dose (chapter 6.8).
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 8IaUhT OeX 1
-
The EI_s value appears beneath the image along with the exposure andexamination data in the Examination and Review sections.
In an emergency
The system has an operating mode for emergencies (chapter 7.4), if youneither have a valid user account nor know the password.
Length and angle measurement
You can measure lengths and angles including Cobbs angle (option)(chapter 10.4.11).
Magnifying an image and moving a section
The buttons for scaling and moving are now on the monitor itself(chapter 10.4.8).
Switching a function on/off.
You can move the image using the mouse or on the monitor itself, using your finger.Stitching
You can stitch several images together automatically or manually to formone large image (chapter 10.4.10.1).
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
1
14 8IaUhT OeX PCR Eleva Ver. 1.1
-
5BCMe PG $PnUenUT
2 *nUSPEVDUJPn 2
2.1 The PCR Eleva ................................................................. 2-3
2.2 About these Instructions for Use ....................................... 2-3
2.3 Normal use ....................................................................... 2-4
2.4 Prohibited use ................................................................... 2-4
2.5 Conformity ....................................................................... 2-52.5.1 Compliance with legal requirements ................... 2-52.5.2 Recycling ........................................................... 2-52.5.3 Mercury (USA only) .......................................... 2-52.5.4 Perchlorate ......................................................... 2-52.5.5 RoHS (China only) ............................................ 2-5
2.6 Training ............................................................................ 2-6
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 21
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
22 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
2 *nUSPEVDUJPn
21 5Ie 1$3 &MeWB
This system is designed for digitizing cassette exposures on conventionalexamination workstations in diagnostic radiology. The PCR system usesso-called image plates instead of screen-film systems.
With PCR systems you can make, digitize, process and transfer filmlesscassette exposures. They are fully automatic from the entering of thepatient data and processing parameters through to film composition.
You can print any required films of these exposures.
The PCR system automatically communicates with other DICOMpartners and archives via the hospital network. Two-way data transferwith an existing RIS (RRadiology IInformation SSystem) guarantees effi-cient data communication within the radiology department.
22 "CPVU UIeTe *nTUSVDUJPnT GPS 6Te
This manual is intended to make it possible for you to work with thedevice described safely.
You may only use this device in compliance with the safety instructionsin this manual and not for purposes other than those for which it wasintended.
Please pay particular attention to all the information given in the chapterSafety instructions.
Conventions
5IJTTZNCPMJEFOUJGJFTJOTUSVDUJPOTXIJDINVTUCFPCTFSWFEJOBOZDBTFUPBWPJEUIFEBOHFSPGJOKVSZUPUIFQBUJFOUBOEPSUPUIFTUBGG
%"/(&3
5IJTJEFOUJGJFTJOTUSVDUJPOTXIJDINVTUCFPCTFSWFEJOBOZDBTFUPBWPJEEBNBHJOHUIFEFWJDFEFTDSJCFEJOUIJTPQFSBUPShTNBOVBM
8"3/*/(
/05& 5IJTJTVTFEUPJEFOUJGZTQFDJBMBEWJDFFHUPBTTJTUUIFPQFSBUPSPSUPJNQSPWFBOPQFSBUJOHTFRVFODF
Conditions for operation are preceded by this icon. Single steps in an action are preceded by this icon.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 *OUSPEVDUJPO 2
-
The result produced by a step is preceded by this icon.SSystem configurations
This manual describes the system with its broadest range of features,including all possible functions, options and all accessories. Your systemwill not necessarily have every function described.
You can only use some of the functions if you have received training inthem by an application specialist, and have specially logged on to getthese privileges. Such functions are marked For the system manageronly.
Depending on the features you have, other manuals may be deliveredwith your system; these contain instructions on safety, calibration, sys-tem tests and service and maintenance.
These Instructions for Use are written in English and were created, au-thorized and marketed by Philips Medical Systems DMC GmbH underthe number given on the front cover.
2 /PSNBM VTe
PCR Eleva S, S Plus, S Hi-res, Corado and CosimaX readers may beinstalled in the patient environment of a medically used room, but onlywith an additional permanently connected protective grounding wire.All other equipment of PCR Eleva shall be installed either outside thepatient environment of a medically used room or in a non-medicallyused room.
Readers AC500, AC5000, Cosima and Compano may only be installedin a non-medically used room.
The PCR Eleva system is designed to process, display and output radio-logical images produced by means of cassette exposures and processedwithin the PCR system.
The PCR Eleva system is not suitable for processing and displayingdigital images produced on other manufacturers systems or compressedusing non-system software.
Diagnosis is only possible on suitable PACS monitors or on printedfilms.
This device may only be operated by suitably qualified persons who havebeen instructed on how to operate the device.
24 1SPIJCJUeE VTe
The system-own monitor is not approved for diagnosis.
PCR Eleva is not an archive and can save exposures only for a limitedperiod. You always should save exposures either to PACS or print themon films. PIJ
lJQT .
eEJDa
l 4ZT
UeN
T /
PveN
CeS
2 /PSNBM VTe
24 *OUSPEVDUJPO PCR Eleva Ver. 1.1
-
25 $PnGPSNJUZ
251 $PNQMJBnDe XJUI MeHBM SeRVJSeNenUT
This Medical Device meets the provisions of the Medical Device Di-rective MDD 93/42 EEC (93).If you have further questions regarding the applicable national or in-ternational standards, please address them to:
Philips Medical Systems DMC GmbH Quality Assurance Department Roentgenstrasse 24 22335 HamburgGermanyE-Mail: [email protected]
252 3eDZDMJnH
The return, proper disposal and recovery of the medical device take placein accordance with the European WEEE Directive (Waste Electricaland Electronic Equipment) and/or respective requirements of nationallegislation.
For further information see chapter Disposal.
For Product Recycling Passport see:www.medical.philips.com/main/company/sustainability/recycling
25 .eSDVSZ 64" PnMZ
/DPSVLQGLVSOD\DQGUHDGHUFRQWDLQPHUFXU\
GLVSRVHSURSHUO\
This product consists of devices that may contain mercury, which mustbe recycled or disposed of in accordance with local, state or federal laws.(Within this system, the backlight lamps in the monitor display and thefluorescent lamps in the reader contain mercury.)
254 1eSDIMPSBUe
This product complies with the requirements of California. It containsPerchlorate Material. Special handling may apply.
For further information see:www.dtsc.ca.gov/hazardouswaste/perchlorate
255 3P)4 $IJnB PnMZ
This product complies with the Chinese requirements.
If necessary you can find further information in chapter Disposal.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
$PnGPSNJUZ 25
PCR Eleva Ver. 1.1 *OUSPEVDUJPO 25
-
2 5SBJnJnH
Operators of this system must have received adequate training on itssafe and effective use before attempting to operate the equipment de-scribed in these Instructions for use.
%POPUVTFUIJTTZTUFNJGZPVIBWFOPUSFDFJWFEQSPQFSBOEBEFRVBUFUSBJOJOHPOJUTTBGFBOEFGGFDUJWFVTF0QFSBUJOHUIFFRVJQNFOUXJUIPVUTVDIUSBJOJOHDPVMEMFBEUPGBUBMPSPUIFSTFSJPVTQFSTPOBMJOKVSZ*UDPVMEBMTPMFBEUPDMJOJDBMNJTEJBHOPTJT
%"/(&3
If you require further information about training in the use of theequipment, please contact your local Philips Medical Systems represen-tative.
Alternatively, contact:
Philips Medical Systems DMC GmbHRoentgenstrasse 2422335 Hamburg, GermanyE-Mail: [email protected]
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
2 5SBJnJnH
2 *OUSPEVDUJPO PCR Eleva Ver. 1.1
-
5BCMe PG $PnUenUT
4BGeUZ JnTUSVDUJPnT
3.1 Safety ................................................................................ 3-3
3.2 Electrical safety .................................................................. 3-3
3.3 Mechanical safety .............................................................. 3-4
3.4 Fire safety .......................................................................... 3-4
3.5 Laser radiation warning ..................................................... 3-4
3.6 Special safety information ................................................. 3-43.6.1 Loss of Data Due to Power Outage .................... 3-43.6.2 Inferior Quality Due to Multiple Exposure
............................................................................ 3-43.6.3 Damage as a result of incorrect cassette insertion
............................................................................ 3-43.6.4 Data inconsistency ............................................. 3-53.6.5 Using Lead Letters ............................................. 3-5
3.7 Electromagnetic Compatibility .......................................... 3-5
3.8 Error messages ................................................................... 3-6
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 1
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
2 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
4BGeUZ JnTUSVDUJPnT
1 4BGeUZ
You may only use this device in compliance with the safety instructionsin this manual and not for purposes other than those for which it wasintended.
It is always the user who is responsible for conforming to the regulationswhich apply to the setting up and operation of medical equipment.
:PVNVTUOPUVTFUIFFRVJQNFOUJGJUIBTBOZFMFDUSJDBMPSNFDIBOJDBMEFGFDUT5IJTQBSUJDVMBSMZBQQMJFTUPGBVMUTJOJOEJDBUPSTEJTQMBZTXBSOJOHTBOEBMBSNT
%"/(&3
If the user wishes to connect the equipment to other equipment,components or assemblies and if it is not apparent from the technicaldata whether it can be safely combined with such equipment, com-ponents or assemblies, the user must consult the manufacturersconcerned or an expert to ensure that the safety of the patient, op-erating staff and the environment is not affected by the intendedcombination.
Philips only accepts responsibility for the safety features of its prod-ucts if maintenance, repairs and modifications are performed byPhilips or persons explicitly authorized to do so by Philips.
As with any technical appliance this equipment also calls for properoperation and regular competent maintenance and care, which aredescribed in the section Service and cleaning.
If you operate the equipment incorrectly or if the user fails to main-tain it properly, Philips cannot be held liable for any resulting faults,damage or injuries.
SSafety circuits may be neither removed nor modified. You may only remove or open parts of the housing if you have
been instructed to do so by this manual.
2 &MeDUSJDBM TBGeUZ
This device conforms to class I according to IEC 60601-1/IEC 60950-1.This equipment may only be operated in medical rooms which meetIEC requirements.
To guarantee the reliable operation of the workstation and to preventthe device from overheating, the device openings must not be blockedoff or covered. The workstation must never be positioned near radiators.
Some of the workstation components are high-voltage. Only electriciansmay remove the devices covers.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 4aGeUZ JOTUSVDUJPOT
-
.eDIBnJDBM TBGeUZ
1MFBTFOPUF :PVNBZOPUUSBOTQPSUUIJTFRVJQNFOUXIJMFJUJTJOPQFSBUJPO4IVUEPXOUIFFRVJQNFOUCFGPSFUSBOTQPSUBUJPOBOEFOTVSFUIBUBMMQFSJQIFSBMQBSUTPGUIFTZTUFNNPOJUPSNPVTFLFZCPBSEDBCMFTFUDBSFEJTDPOOFDUFEBOETBGFMZUSBOTQPSUFE
#FGPSFVTJOHMPOHPSUIPDBTTFUUFTUIF1$3&MFWB441MVTQMBUFSFBEFSNVTUCFTFDVSFEUPUIFGMPPSVTJOHUIFTVQQMJFECSBDLFUTPUIFQMBUFSFBEFSEPFTOPUUPQQMFPWFS
8"3/*/(
4 'JSe TBGeUZ
0OFMFDUSJDBMPSDIFNJDBMGJSFTPOMZVTFFYUJOHVJTIFSTXIJDIBSFTQFDJGJDBMMZMBCFMFEGPSUIPTFQVSQPTFT"OZBUUFNQUUPFYUJOHVJTIFMFDUSJDBMGJSFTXJUIXBUFSPSPUIFSMJRVJETDBOSFTVMUJOTFSJPVTQFSTPOBMJOKVSZ
5IFBJSWFOUTNBZOPUCFDPWFSFEXIFOUIFEFWJDFJTTXJUDIFEPO%"/(&3
5 -BTeS SBEJBUJPn XBSnJnH
All plate readers use laser radiation (class IIIb, 50 mW).
6OEFSOPDJSDVNTUBODFTTIPVMEZPVPQFOUIFIPVTJOHPGUIFQMBUFSFBEFS$POUBDUFTQFDJBMMZFZFDPOUBDUXJUIMBTFSSBEJBUJPODBODBVTFEBNBHFUPIFBMUI
*OUIFFWFOUPGBNBMGVODUJPOQMFBTFOPUJGZTVJUBCMZRVBMJGJFETUBGGPOMZ%"/(&3
4QeDJBM TBGeUZ JnGPSNBUJPn
1 -PTT PG %BUB %Ve UP 1PXeS 0VUBHe
This device should be connected to an uninterruptible power system toprevent database damage and image loss in the event of a power outage.
2 *nGeSJPS 2VBMJUZ %Ve UP .VMUJQMe &YQPTVSe
When using UNIQUE image processing there must be only one imageon a cassette. Otherwise, the image processing may fail to correctly findthe region of interest.
%BNBHe BT B SeTVMU PG JnDPSSeDU DBTTeUUe JnTeSUJPn
Always slide the cassettes into the cassette compartment of the platereader with care. Incorrect insertion of a cassette, for instance if it is at
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
.eDIBnJDBM TBGeUZ
4 4aGeUZ JOTUSVDUJPOT PCR Eleva Ver. 1.1
-
an angle, can damage the plate reader. Please observe the instructionsin chapter 6.1.1.
4 %BUB JnDPnTJTUenDZ
Do not switch off the operators console at the power switch if the PCRapplication program is displayed. This can lead to database errors ordata inconsistencies. Always follow the sequence for switching off asdescribed in chapter 5.
5 6TJnH -eBE -eUUeST
To avoid false interpretation regarding body direction (right, left) andpatient orientation, Philips recommends the use of lead markers (let-ters), as traditionally used in conventional radiography.
7 &MeDUSPNBHneUJD $PNQBUJCJMJUZ
The system conforms to the classification Group 1, Class A underCISPR 11. It contains Class A IT components according to CISPR 22.Although it has been shown that this does not have a detrimental effecton the performance and safety of the system, electromagnetic incom-patibilities may arise when operated in residential areas. In such anevent, it may be necessary for the user to take appropriate measures.
For the intended use, this X-ray system complies with the EuropeanCouncil Directive concerning Medical Devices, as signified by the CEmark. This includes definitions on permitted levels on electromagneticemission and requirements on immunity against electromagnetic emis-sions from other equipment.
It is not possible, however, to exclude with absolute certainty the pos-sibility that other high frequency electronic equipment, which is fullycompliant to the EMC regulations, will not adversely affect the opera-tion of this system. If the other equipment has a comparatively highlevel of transmission power and is in close proximity, these EMC con-cerns (the risk of interference) may be more pronounced. It is thereforerecommended that the operation of equipment of this type such as mo-bile telephones, cordless microphones and other similar mobile radioequipment be restricted from the vicinity of this X-ray system.
Explanation:Electronic apparatus that satisfies the EMC requirements is designed sothat under normal conditions there is no risk of malfunction caused byelectromagnetic interference. However, in the case of radio signals fromhigh-frequency transmitters with a relatively high transmission power,the risk of electromagnetic incompatibility when operated in close prox-imity to electronic apparatus cannot be totally ruled out.
In unusual circumstances unintended functions of the apparatus couldbe initiated, possibly giving rise to undesirable risks for the patient oruser.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
&MeDUSPNBHneUJD $PNQBUJCJMJUZ 7
PCR Eleva Ver. 1.1 4aGeUZ JOTUSVDUJPOT 5
-
For this reason, all kinds of transmission with mobile radio equipmentshould be avoided. This also applies when the equipment is in standbymode.
Mobile telephones must be sswitched off in designated problem zones.
8 &SSPS NeTTBHeT
When an error has occurred in the system or a part thereof, an errormessage appears on the operators console monitor with instructions onhow to rectify it (example).
3ODWH,'IURPUHDGHULVQRWUHJLVWUHG3UHVVWKH&RQILUPEXWWRQWRUHPRYHWKHFDVVHWWH
2.
-eHenE 'VnDUJPn .eBnJnH 8IBU ZPV NVTU EP
ESSPS NeTTaHe eYaNQle
4ZTUeN TUaUVT EJTQlaZ#lVe EveSZUIJOH 0,:ellPX ESSPSReE 6OSeDPveSaCleeSSPS
(SeeO DJSDleEveSZUIJOH 0,
:ellPX USJaOHleESSPSeH SeaEeS OPU SeaEZGPS VTe
ClJDL PO TZNCPl NPSe JOGPSNaUJPO aQQeaST
ReE DSPTT6OSeDPveSaCle eSSPS
ClJDL PO TZNCPl NPSe JOGPSNaUJPO aQQeaST
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
8 &SSPS NeTTBHeT
4aGeUZ JOTUSVDUJPOT PCR Eleva Ver. 1.1
-
&WFOJGOPFSSPSNFTTBHFBQQFBSTCVUUIFFRVJQNFOUEPFTOPUGVODUJPOBTVTVBMGJSTUTJHOTPGBEFGFDU1IJMJQT$VTUPNFS4FSWJDFNVTUCFJOGPSNFE
%"/(&3
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
&SSPS NeTTBHeT 8
PCR Eleva Ver. 1.1 4aGeUZ JOTUSVDUJPOT 7
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
8 &SSPS NeTTBHeT
8 4aGeUZ JOTUSVDUJPOT PCR Eleva Ver. 1.1
-
5BCMe PG $PnUenUT
4 4ZTUeN EeTDSJQUJPn 4
4.1 System components ........................................................... 4-34.1.1 The basic system ................................................ 4-34.1.2 Possible configurations ....................................... 4-3
4.2 Standard workflow ............................................................ 4-5
4.3 The operators console ....................................................... 4-8
4.4 PCR plate reader ............................................................. 4-104.4.1 The PCR Compano plate reader ...................... 4-104.4.2 The PCR Corado, CosimaX and S Hi-res plate
readers .............................................................. 4-124.4.3 The PCR S and S Plus plate readers ................. 4-144.4.4 The PCR AC500 plate reader ......................... 4-184.4.5 The PCR AC5000 plate reader ....................... 4-224.4.6 The PCR Cosima plate reader .......................... 4-26
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 41
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
42 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
4 4ZTUeN EeTDSJQUJPn
41 4ZTUeN DPNQPnenUT
411 5Ie CBTJD TZTUeN
It consists of an operators console with monitor or touch screen, barcode reader,
keyboard and mouse a plate reader.When you work with a touch screen you require neither a mouse norkeyboard for operation.
You can combine more than one plate reader and operators console toform a system.
During exposure there is no mechanical connection between plates andsystem, so that in the event of incorrect configuration or cassette use,the system cannot, for instance, accurately show the status of a flippedimage or the assignment to a patient.
412 1PTTJCMe DPnGJHVSBUJPnT
OOne-console system with one plate reader
This configuration is the basic configuration for a PCR system. You canwork with or without being connected up to the hospital network.
3ODWHUHDGHU
3ULQWHU
3&5RSHUDWRUVFRQVROH
+RVSLWDOQHWZRUN
Multiple-console system with one plate reader
This configuration is the extended basic configuration. To achieve amore efficient workflow, you can set up more than one operators con-sole, e.g. in the various examination rooms and in patient admissions.At the same time, the entire system is connected up to the hospitalnetwork. The assignment of image plate to patient and vice versa isalways unique.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 4
-
Plate reader
Printer
PCR operators console in examination room
PCR operators console at plate reader
Hospital network
Multiple-console system with more than one plate reader (option)
With configurations of this type you can read out any plate cassette onany plate reader, regardless of the operators console on which you as-signed the image plate, or with which console the patient data ismanaged. After readout the image always comes to the operators con-sole to which the image plate is assigned.
3ODWHUHDGHU
3ULQWHU
3&5RSHUDWRUVFRQVROHV
+RVSLWDOQHWZRUN
3ODWHUHDGHU
3&5RSHUDWRUVFRQVROHV
The following applies to all configurations All operators consoles have independent worklists and are
synchronized with the RIS separately. On any one operators console you can view, process, print and ex-
port only the images assigned to it. You should therefore performan examination to the fullest possible extent on one operators con-sole only.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
41 4ZTUeN DPNQPnenUT
44 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
SServer and client In a multiple-console system, one console is always the server, which
assigns the image plates and distributes the images. If this breaksdown or is not switched on, no image plates can be assigned or im-ages read out in the entire system. Always switch the server on firstand off last.
If, in a multiple-console system with more than one plate reader, theserver is down, the system manager can convert the operators con-soles which operate another plate reader into servers so that it ispossible to continue working with part of the system.
42 4UBnEBSE XPSLGMPX
The following diagram shows the standard workflow with the PCR sys-tem, by way of example. You will find alternatives to the describedassignment with the barcode reader in the rest of the manual.
The differences to conventional radiography are as follows: You work with plate cassettes instead of film cassettes. Before radiography, you must assign the plate cassette to a view using
the barcode reader. After radiography, you must read out the plate cassette in the plate
reader. The X-ray exposure is automatically transferred from the plate reader
to the operators console. For filing purposes, you can print out the X-ray exposure or save it
on computer.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
4UBnEBSE XPSLGMPX 42
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 45
-
2QWKHGLDJQRVWLFXQLW
2QWKHRSHUDWRUVFRQVROH6WDQGDUGZRUNIORZRQWKH3&5V\VWHPH[DPSOH
0DNHH[SRVXUHV
6HOHFWSDWLHQW
$VVLJQLPDJHSODWHVZLWKEDUFRGHUHDGHU
,QVHUWFDVVHWWHV
2QWKHRSHUDWRUVFRQVROH
7DNHRXWFDVVHWWHV
,PDJHSODWHLVUHDGRXWDQGHUDVHG
3ULQWILOP
2QWKHSODWHUHDGHU
6HOHFWYLHZV
6HOHFWH[DPLQDWLRQVHFWLRQ
&KHFNLPDJHTXDOLW\
&RPSOHWHH[DPLQDWLRQ
([DPLQDWLRQVHFWLRQRSHQV
,PDJHSODWHLVDVVLJQHGWRWKHVHOHFWHGYLHZ
(GLWH[SRVXUHLIQHFHVVDU\
5,6TXHU\RUSDWLHQWHQWU\
$UFKLYHH[SRVXUHVVDYHWRLPDJHDUFKLYHRQFRPSXWHUDQGRU
([SRVXUHVDSSHDURQWKHRSHUDWRUVFRQVROH
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
42 4UBnEBSE XPSLGMPX
4 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
4UBnEBSE XPSLGMPX 42
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 47
-
4 5Ie PQeSBUPST DPnTPMe
You can select the sections using the main selector buttons (11-6). Whena section is selected the button for it turns yellow.
You can do the following in the sections:
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
4 5Ie PQeSBUPST DPnTPMe
48 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P 'VnDUJPn .eBnJnH
1BUJenU MJTU EOUeS QaUJeOU aOE eYaNJOaUJPO EaUa aOE TeleDUQaUJeOUT
&YBNJnBUJPn CPOUSPl UIe eYaNJOaUJPO aTTJHO aOE JOTeSU JNaHe QlaUe
3eWJeX CIeDL aOE JG OeDeTTaSZ JNQSPve UIe JNaHeRValJUZ
1SJnU :PV XJll GJOE UIe QSJOUJOH UPPlT IeSe
J4JUe ETUaClJTI DPOOeDUJPO UP UIe PIJlJQT J4JUe 8eC4eSveS PQUJPO
4ZTUeN EYJU UIe aQQlJDaUJPO QSPHSaN GVODUJPOT GPS UIeTZTUeN NaOaHeS
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie PQeSBUPST DPnTPMe 4
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 49
-
44 1$3 QMBUe SeBEeS
441 5Ie 1$3 $PNQBnP QMBUe SeBEeS
8
1 2 3 4 5 6 7
1011
9
The plate reader reads out the exposed image plates. The image platesare contained in the plate cassettes.
/P .eBnJnH
CPOUSPl QaOel XJUI PQeSaUJOH JOEJDaUPST
-PaE laNQ-JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe aGUeS TXJUDIJOH PO IaT CeeOTVDDeTTGVllZ DPNQleUeE aOE UIe QlaUe SeaEeS JT SeaEZ UP QSPDeTT DaTTeUUeT
4UaUVT EJTQlaZT GPS DaTTeUUe QSPDeTTJOH5IeZ GlaTI EVSJOH DaTTeUUe QSPDeTTJOH
6OlPaE laNQ-JHIUT VQ HSeeO PODe DaTTeUUe QSPDeTTJOH JT GJOJTIeE 5Ie DaTTeUUe DaO CeUaLeO PVU P
IJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
41 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
.eTTaHe JOEJDaUPS8IeO JU lJHIUT VQ ZellPX
a XJOEPX XJUI TQeDJal JOTUSVDUJPOT aQQeaSTPO UIe NPOJUPS PG UIe PQeSaUPShT DPOTPle 'PllPX UIeTe JOTUSVDUJPOT *GaO eSSPS NeTTaHe aQQeaST aO aVEJCle XaSOJOH TJHOal JT HJveO
ESaTe CVUUPO GPS QSJNaSZ aOE TeDPOEaSZ eSaTVSePSeTTJOH UIe CVUUPO PODe aDUJvaUeT UIe TeDPOEaSZ eSaTVSe NPEe aOE
aQQeaST OeYU UP UIe CVUUPO PSeTT JU aHaJO UP aDUJvaUe UIe QSJNaSZeSaTVSe NPEe aOE aQQeaST 5P SeUVSO UP UIe SPVUJOe NPEe QSeTTUIe CVUUPO aHaJO
PJlPU laNQ-JHIUT VQ HSeeO XIeO UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
CaTTeUUe TlPU" DaTTeUUe XJUI UIe JNaHe QlaUe JT JOTeSUeE IeSe
ReTeU TXJUDI0OlZ GPS SeTUaSUJOH JO UIe eveOU PG EJTSVQUJPOT EP OPU QSeTT JU VOEeSOPSNal PQeSaUJOH DPOEJUJPOT
.aJO TXJUDI*T alXaZT leGU JO QPTJUJPO i*w 0O 4XJUDIJOH JU PGG JT POlZ aEvJTaCle JO UIeeveOU PG QSPlPOHeE TUPQQaHeT
%VTU GJlUeSveOUJlaUPS" EVTU GJlUeS JT GJUUeE PveS UIe veOUJlaUPS
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 411
-
442 5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST
910
1
8 2
3
7
4
5
6
The plate reader reads out the exposed image plates. The image platesare contained in the plate cassettes.
Distinguishing features
All three plate readers perform the same functions.The CCorado and CCosimaX plate readers are identical externally and areshown in the following picture. The SS Hi-res has only one cassette compartment.
/P .eBnJnH
CPOUSPl QaOel5Ie QlaUe SeaEeST UPVDITeOTJUJve DPOUSPl QaOel UPVDI TDSeeO
PJlPU laNQ-JHIUT VQ HSeeO XIeO UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
PPXeS TXJUDI4XJUDIeT PO UIe QlaUe SeaEeS JG UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
CaTTeUUe DPNQaSUNeOU OVNCeS
6OlPaE laNQ-JHIUT VQ HSeeO PODe DaTTeUUe QSPDeTTJOH JT GJOJTIeE 5Ie DaTTeUUe DaO CeUaLeO PVU
PSPDeTTJOH laNQ'laTIeT XIJle UIe DaTTeUUe JT CeJOH QSPDeTTeE CZ UIe QlaUe SeaEeS %VSJOHUIJT UJNe UIe DaTTeUUe JT lPDLeE JO UIe DaTTeUUe DPNQaSUNeOU
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
412 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
-PaE laNQ-JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe aGUeS TXJUDIJOH PO IaT CeeOTVDDeTTGVllZ DPNQleUeE aOE UIe QlaUe SeaEeS JT SeaEZ UP QSPDeTT DaTTeUUeT
'PVS DaTTeUUe TlPUT 4 )JSeT POe DaTTeUUe TlPU
'PS UIe TJNVlUaOePVT QSPDeTTJOH PG VQ UP GPVS DaTTeUUeT
%VTU GJlUeSveOUJlaUPS" EVTU GJlUeS JT GJUUeE PveS UIe veOUJlaUPS
.aJO TXJUDI*T alXaZT leGU JO QPTJUJPO i*w 0O 4XJUDIJOH JU PGG JT POlZ aEvJTaCle JO UIeeveOU PG QSPlPOHeE TUPQQaHeT
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 41
-
44 5Ie 1$3 4 BnE 4 1MVT QMBUe SeBEeST
The plate reader reads out the exposed image plates. The image platesare contained in the plate cassettes.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
414 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
.POJUPS Tee CelPX
#VUUPOT GPS TeleDUJOH NPEe8JUI UIeTe CVUUPOT ZPV DaO TeleDU UIe NPEe aTTJHOeE PO TDSeeO
PJlPU laNQ-JHIUT VQ HSeeO XIeO UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
EOeSHZTavJOH JOEJDaUPS*G UIe lJHIU JT GlaTIJOH UIe NPOJUPS JT JO eOeSHZTavJOH NPEe
ReaEZ JOEJDaUPS*G JU JT lJU HSeeO
UIe QlaUe SeaEeS JT SeaEZ GPS PQeSaUJPO
4UaUVT EJTQlaZ GPS DaTTeUUe QSPDeTTJOH-JHIUT VQ HSeeO UIe DaTTeUUe XaT JOTeSUeE DPSSeDUlZ'laTIeT HSeeO QSPDeTTJOH DPNQleUe 0GG UIe JNaHe QlaUe JT JO UIe DaTTeUUe aHaJO
6OlPaE laNQ'laTIeT ClVe XIeO UIe DaTTeUUe DaO Ce SeNPveE "GUeS UIe DaTTeUUe IaTCeeO SeNPveE UIe laNQ HPeT PVU
.eTTaHe JOEJDaUPS8IeO JU GlaTIeT ZellPX
a XJOEPX XJUI TQeDJal JOTUSVDUJPOT aQQeaST POUIe NPOJUPS PG UIe PQeSaUPShT DPOTPle 'PllPX UIeTe JOTUSVDUJPOT *G aOeSSPS NeTTaHe aQQeaST aO aVEJCle XaSOJOH TJHOal JT HJveO
PPXeS TXJUDI4XJUDIeT PO UIe QlaUe SeaEeS JG UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
CaTTeUUe TlPU XJUI EVTU GJlUeS" DaTTeUUe XJUI JNaHe QlaUe JT JOTeSUeE IeSe
.aJO TXJUDI*T alXaZT leGU JO QPTJUJPO i*w 0O 4XJUDIJOH JU PGG JT POlZ aEvJTaCle JO UIeeveOU PG QSPlPOHeE TUPQQaHeT
.aJOT DPOOeDUJPO
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 415
-
SStart screen
/P .eBnJnH
5eYU CPY5Ie OaNe PG UIe TeleDUeE NeOV aQQeaST IeSe
'JelE GPS TZTUeN NeTTaHeT
6TeS JOTUSVDUJPOT 4UaUVT EJTQlaZ
PSPHSeTT JOEJDaUPS*U EJTQlaZT UIe TUaUVT PG DaTTeUUe QSPDeTTJOH Tee CelPX
'VODUJPO EaUa aOOPUaUJPO
/eUXPSL DPOOeDUJPO TUaUVT
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
41 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
TThe progress indicator
OLWXS1RFDVVHWWHLQVHUWHG
&DVVHWWHLQVHUWHGVWDUWRIUHDGRXW
5HDGRXWLQSURJUHVV
(UDVXUH
(UDVXUHDQGUHDGRXWILQLVKHG
5HDGRXWILQLVKHGUHPRYLQJFDVVHWWH
OLWXS
OLWXS
OLWXS
IODVKHV
IODVKHV
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 417
-
444 5Ie 1$3 "$5 QMBUe SeBEeS
Fig. 4.1 Hardware, front view
/P .eBnJnH
4DSeeODPOUSPl QaOel ReaEPVU eYQPTVSeT aSe EJTQlaZeE JO UIe leGU TJEe PG UIe TDSeeO GPSDIeDLJOH UIe SJHIU TJEe PG UIe TDSeeO JT a UPVDI TDSeeO GPS TeleDUJOHGVODUJPOT
'lPQQZ EJTL ESJve'PS PIJlJQT CVTUPNeS 4eSvJDe POlZ
CPOUSaTU DPOUSPl 'PS TeUUJOH UIe TDSeeO EJTQlaZ DPOUSaTU 5VSO VQXaSE SeEVDeT DPOUSaTUUVSO EPXOXaSE JODSeaTeT DPOUSaTU PQeO UIe DPveS QaOel CZ QSeTTJOH JUJO
-PaE laNQ -JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe IaT CeeO TVDDeTTGVllZ DPNQleUeE aOE UIe DaTTeUUe TlPU JT SeaEZ GPS JOTeSUJPO PG a DaTTeUUe *G JU EPeTOPU lJHIU VQ a DaTTeUUe NVTU OPU Ce JOTeSUeE
6OlPaE laNQ-JHIUT VQ HSeeO PODe DaTTeUUe QSPDeTTJOH JT GJOJTIeE 5Ie DaTTeUUe XJUIUIe eSaTeE JNaHe QlaUe DaO Ce UaLeO PVU
4UPSaHe DPNQaSUNeOU 5eNQPSaSZ TUPSaHe GPS VQ UP GPVS DaTTeUUeT CeGPSe PS aGUeS QSPDeTTJOH
CaTTeUUe TlPU5aLeT a DaTTeUUe XJUI eYQPTeE JNaHe QlaUe
PPXeS TXJUDI4XJUDIeT PO UIe QlaUe SeaEeS JG UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
PPXeS JOEJDaUPS -JHIUT VQ HSeeO XIeO UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O aOE QPXeSJT TVQQlJeE 4JODe UIe NaJO TXJUDI JT OPSNallZ leGU JO UIe 0/ QPTJUJPO UIeQPXeS JOEJDaUPS JT alTP DPOTUaOUlZ lJU
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
418 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
ESSPS JOEJDaUPS -JHIUT VQ PSaOHe XIeO a DSJUJDal eSSPS IaT PDDVSSeE EVSJOH PQeSaUJPO"O alaSN TJHOal JT IeaSE aOE a XJOEPX aQQeaST PO UIe TDSeeO 'PllPX UIeJOTUSVDUJPOT HJveO JO UIJT XJOEPX
.eTTaHe JOEJDaUPS 8IeO UIJT JOEJDaUPS lJHIUT VQ ZellPX
aO alaSN TJHOal JT IeaSE " XJOEPXXJUI TQeDJal JOGPSNaUJPO aQQeaST PO UIe TDSeeO 'PllPX UIe JOTUSVDUJPOTHJveO IeSe 8IeO UIe XJOEPX EJTaQQeaST UIe NeTTaHe JOEJDaUPS alTPHPeT PVU
ReaEZ JOEJDaUPS -JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe aGUeS TXJUDIJOH PO IaT CeeOTVDDeTTGVllZ DPNQleUeE aOE UIe QlaUe SeaEeS JT SeaEZ UP QSPDeTT DaTTeUUeT
5(6(7
Fig. 4.2 Hardware, side view
/P .eBnJnH
.aJO TXJUDI*T alXaZT leGU JO UIe 0/ QPTJUJPO i*w
%VTU GJlUeS" EVTU GJlUeS JT GJUUeE PveS UIe veOUJlaUPS
ReTeU TXJUDI 0OlZ GPS SeTUaSUJOH JO UIe eveOU PG PQeSaUJOH eSSPST EP OPU QSeTT JU EVSJOH OPSNal PQeSaUJPO
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 419
-
5($',1*
6
$,'6PLWK %RE&+(673RWDEOH&KHVW
Fig. 4.3 User interface
The diagram shows the main window on the AC500 which appears afterswitching on. It is the central window for all displays and the startingpoint for the selection of other functions.
/P .eBnJnH
EYQPTVSe OVNCeS *OUeSOal aENJOJTUSaUJPO OVNCeS PO UIe QlaUe SeaEeS *U JT DPVOUeE DPOTeDVUJvelZ
PaUJeOU OVNCeS *%
EYaNJOaUJPO7JeX
PaUJeOU OaNe 'PS UIe eYQPTVSe DVSSeOUlZ CeJOH QSPDeTTeE
0QeSaUJOH NPEeT JO UIe DaTTeUUe TlPU
3 PSJNaSZ eSaTVSe
6 4eDPOEaSZ eSaTVSe
PSJPSJUZ DaTTeUUe QSPDeTTJOH
8aSOJOH JOEJDaUPS 'laTIeT XIeO a XaSOJOH PDDVST OPUJGZ CVTUPNeS 4eSvJDe JG OeDeTTaSZ
0QeSaUJOH NPEe TXJUDI 'PS TeleDUJOH NPEeT lJLe QSJNaSZTeDPOEaSZ eSaTVSe aOE QSJPSJUZ DaTTeUUeQSPDeTTJOH
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
42 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
PSPHSeTT JOEJDaUPS
/P DaTTeUUe ZeU JOTeSUeE aGUeS SeNPval
CaTTeUUe aDDeQUeE QSeQaSJOH UP SeaE
ReaEJOH CeHVO o ESaTVSe CeHVO
ESaTVSe JO QSPHSeTT o ESaTVSe DPODlVEeE
PSPDeTTJOH DPNQleUeE aXaJUJOH DaTTeUUe SeNPval
PSPUeDUeEVOQSPDeTTeE eYQPTVSeT /VNCeS PG eYQPTVSeT UIaU XeSe QSPUeDUeE GSPN aVUPNaUJD eSaTVSe PSOVNCeS PG eYQPTVSeT OPU ZeU QSPDeTTeE
"ENJOJTUSaUJPO CallT VQ UIe lJTU PG JNaHeT ZeU UP Ce QSPDeTTeE
6UJlJUJeTCallT VQ VUJlJUJeT
%JTQlaZ aSea %JTQlaZT UIe eYQPTVSe DVSSeOUlZ CeJOH SeaE PVU
0QeSaUJOH NPEe %JTQlaZT UIe DVSSeOU QlaUe SeaEeS NPEe
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 421
-
445 5Ie 1$3 "$5 QMBUe SeBEeS
Fig. 4.4 Hardware, front view
/P .eBnJnH
4DSeeODPOUSPl QaOel ReaEPVU eYQPTVSeT aSe EJTQlaZeE JO UIe leGU TJEe PG UIe TDSeeO GPSDIeDLJOH UIe SJHIU TJEe PG UIe TDSeeO JT a UPVDI TDSeeO GPS TeleDUJOHGVODUJPOT
'lPQQZ EJTL ESJve'PS PIJlJQT CVTUPNeS 4eSvJDe POlZ
CPOUSaTU DPOUSPl 'PS TeUUJOH UIe TDSeeO EJTQlaZ DPOUSaTU 5VSO VQXaSE SeEVDeT DPOUSaTUUVSO EPXOXaSE JODSeaTeT DPOUSaTU PQeO UIe DPveS QaOel CZ QSeTTJOH JUJO
CaTTeUUe TlPU OVNCeS CaTTeUUe TlPU OVNCeS
-PaE laNQ -JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe IaT CeeO TVDDeTTGVllZ DPNQleUeE aOE UIe DaTTeUUe TlPU JT SeaEZ GPS JOTeSUJPO PG a DaTTeUUe *G JU EPeTOPU lJHIU VQ a DaTTeUUe NVTU OPU Ce JOTeSUeE
6OlPaE laNQ-JHIUT VQ HSeeO PODe DaTTeUUe QSPDeTTJOH JT GJOJTIeE 5Ie DaTTeUUe XJUIUIe eSaTeE JNaHe QlaUe DaO Ce UaLeO PVU
4UPSaHe DPNQaSUNeOU 5eNQPSaSZ TUPSaHe GPS VQ UP GPVS DaTTeUUeT CeGPSe PS aGUeS QSPDeTTJOH
CaTTeUUe TlPU5aLeT a DaTTeUUe XJUI eYQPTeE JNaHe QlaUe
PPXeS TXJUDI4XJUDIeT PO UIe QlaUe SeaEeS JG UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
PPXeS JOEJDaUPS -JHIUT VQ HSeeO XIeO UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O aOE QPXeSJT TVQQlJeE 4JODe UIe NaJO TXJUDI JT OPSNallZ leGU JO UIe 0/ QPTJUJPO UIeQPXeS JOEJDaUPS JT alTP DPOTUaOUlZ lJU
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
422 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
ESSPS JOEJDaUPS -JHIUT VQ PSaOHe XIeO a DSJUJDal eSSPS IaT PDDVSSeE EVSJOH PQeSaUJPO"O alaSN TJHOal JT IeaSE aOE a XJOEPX aQQeaST PO UIe TDSeeO 'PllPX UIeJOTUSVDUJPOT HJveO JO UIJT XJOEPX
.eTTaHe JOEJDaUPS 8IeO UIJT JOEJDaUPS lJHIUT VQ ZellPX
aO alaSN TJHOal JT IeaSE " XJOEPXXJUI TQeDJal JOGPSNaUJPO aQQeaST PO UIe TDSeeO 'PllPX UIe JOTUSVDUJPOTHJveO IeSe 8IeO UIe XJOEPX EJTaQQeaST UIe NeTTaHe JOEJDaUPS alTPHPeT PVU
ReaEZ JOEJDaUPS -JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe aGUeS TXJUDIJOH PO IaT CeeOTVDDeTTGVllZ DPNQleUeE aOE UIe QlaUe SeaEeS JT SeaEZ UP QSPDeTT DaTTeUUeT
5(6(7
Fig. 4.5 Side view
/P .eBnJnH
.aJO TXJUDI*T alXaZT leGU JO UIe 0/ QPTJUJPO i*w
%VTU GJlUeS" EVTU GJlUeS JT GJUUeE PveS UIe veOUJlaUPS
ReTeU TXJUDI 0OlZ GPS SeTUaSUJOH JO UIe eveOU PG PQeSaUJOH eSSPST EP OPU QSeTT JU EVSJOH OPSNal PQeSaUJPO
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 42
-
36
5($',1*
$,'6PL WK %RE&+(673RWDEOH&KHVW
Fig. 4.6 User interface
The diagram shows the main window on the AC5000 which appearsafter switching on. It is the central window for all displays and the start-ing point for the selection of other functions.
/P .eBnJnH
EYQPTVSe OVNCeS *OUeSOal aENJOJTUSaUJPO OVNCeS PO UIe QlaUe SeaEeS *U JT DPVOUeE DPOTeDVUJvelZ
PaUJeOU OVNCeS *%
EYaNJOaUJPO7JeX
PaUJeOU OaNe 'PS UIe eYQPTVSe DVSSeOUlZ CeJOH QSPDeTTeE
0QeSaUJOH NPEeT JO UIe DaTTeUUe TlPU
3 PSJNaSZ eSaTVSe
6 4eDPOEaSZ eSaTVSe
PSJPSJUZ DaTTeUUe QSPDeTTJOH
8aSOJOH JOEJDaUPS 'laTIeT XIeO a XaSOJOH PDDVST OPUJGZ CVTUPNeS 4eSvJDe JG OeDeTTaSZ
0QeSaUJOH NPEe TXJUDI 'PS TeleDUJOH NPEeT lJLe QSJNaSZTeDPOEaSZ eSaTVSe aOE QSJPSJUZ DaTTeUUeQSPDeTTJOH
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
424 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
PSPHSeTT JOEJDaUPS
/P DaTTeUUe ZeU JOTeSUeE aGUeS SeNPval
CaTTeUUe aDDeQUeE QSeQaSJOH UP SeaE
ReaEJOH CeHVO o ESaTVSe CeHVO
ESaTVSe JO QSPHSeTT o ESaTVSe DPODlVEeE
PSPDeTTJOH DPNQleUeE aXaJUJOH DaTTeUUe SeNPval
PSPUeDUeEVOQSPDeTTeE eYQPTVSeT /VNCeS PG eYQPTVSeT UIaU XeSe QSPUeDUeE GSPN aVUPNaUJD eSaTVSe PSOVNCeS PG eYQPTVSeT OPU ZeU QSPDeTTeE
"ENJOJTUSaUJPO CallT VQ UIe lJTU PG JNaHeT ZeU UP Ce QSPDeTTeE
6UJlJUJeTCallT VQ VUJlJUJeT
%JTQlaZ aSea %JTQlaZT UIe eYQPTVSe DVSSeOUlZ CeJOH SeaE PVU
CaTTeUUe TlPU OVNCeS
'VODUJPOJOH QSPQeSlZ CaTTeUUe TlPU EeGeDUJve JNQPTTJCle UP JOTeSU PS SeNPve UIe DaT
TeUUe PleaTe OPUJGZ PIJlJQT CVTUPNeS 4eSvJDe
0QeSaUJOH NPEe %JTQlaZT UIe DVSSeOU QlaUe SeaEeS NPEe
CaTTeUUe TlPU OVNCeS CaTTeUUe TlPU OVNCeS PG UIe eYQPTVSe DVSSeOUlZ CeJOH QSPDeTTeE
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 425
-
44 5Ie 1$3 $PTJNB QMBUe SeBEeS
Fig. 4.7 Hardware, front view
/P .eBnJnH
4DSeeODPOUSPl QaOel ReaEPVU eYQPTVSeT aSe EJTQlaZeE JO UIe leGU TJEe PG UIe TDSeeO GPSDIeDLJOH UIe SJHIU TJEe PG UIe TDSeeO JT a UPVDI TDSeeO GPS TeleDUJOHGVODUJPOT
'lPQQZ EJTL ESJve'PS PIJlJQT CVTUPNeS 4eSvJDe POlZ
CPOUSaTU DPOUSPl 'PS TeUUJOH UIe TDSeeO EJTQlaZ DPOUSaTU 5VSO VQXaSE SeEVDeT DPOUSaTUUVSO EPXOXaSE JODSeaTeT DPOUSaTU PQeO UIe DPveS QaOel CZ QSeTTJOH JUJO
-PaE laNQ -JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe IaT CeeO TVDDeTTGVllZ DPNQleUeE aOE UIe DaTTeUUe TlPU JT SeaEZ GPS JOTeSUJPO PG a DaTTeUUe *G JU EPeTOPU lJHIU VQ a DaTTeUUe NVTU OPU Ce JOTeSUeE
6OlPaE laNQ-JHIUT VQ HSeeO PODe DaTTeUUe QSPDeTTJOH JT GJOJTIeE 5Ie DaTTeUUe XJUIUIe eSaTeE JNaHe QlaUe DaO Ce UaLeO PVU
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
42 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
4UPSaHe DPNQaSUNeOU 5eNQPSaSZ TUPSaHe GPS VQ UP GPVS DaTTeUUeT CeGPSe PS aGUeS QSPDeTTJOH
CaTTeUUe TlPU5aLeT a DaTTeUUe XJUI eYQPTeE JNaHe QlaUe
PPXeS TXJUDI4XJUDIeT PO UIe QlaUe SeaEeS JG UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O
PPXeS JOEJDaUPS -JHIUT VQ HSeeO XIeO UIe NaJO TXJUDI JT JO QPTJUJPO i*w 0O aOE QPXeSJT TVQQlJeE 4JODe UIe NaJO TXJUDI JT OPSNallZ leGU JO UIe 0/ QPTJUJPO UIeQPXeS JOEJDaUPS JT alTP DPOTUaOUlZ lJU
ESSPS JOEJDaUPS -JHIUT VQ PSaOHe XIeO a DSJUJDal eSSPS IaT PDDVSSeE EVSJOH PQeSaUJPO"O alaSN TJHOal JT IeaSE aOE a XJOEPX aQQeaST PO UIe TDSeeO 'PllPX UIeJOTUSVDUJPOT HJveO JO UIJT XJOEPX
.eTTaHe JOEJDaUPS 8IeO UIJT JOEJDaUPS lJHIUT VQ ZellPX
aO alaSN TJHOal JT IeaSE " XJOEPXXJUI TQeDJal JOGPSNaUJPO aQQeaST PO UIe TDSeeO 'PllPX UIe JOTUSVDUJPOTHJveO IeSe 8IeO UIe XJOEPX EJTaQQeaST UIe NeTTaHe JOEJDaUPS alTPHPeT PVU
ReaEZ JOEJDaUPS -JHIUT VQ HSeeO PODe UIe TUaSUVQ QSPDeEVSe aGUeS TXJUDIJOH PO IaT CeeOTVDDeTTGVllZ DPNQleUeE aOE UIe QlaUe SeaEeS JT SeaEZ UP QSPDeTT DaTTeUUeT
%VTU GJlUeS" EVTU GJlUeS JT GJUUeE PveS UIe veOUJlaUPS
.aJO TXJUDI*T alXaZT leGU JO UIe 0/ QPTJUJPO i*w
ReTeU TXJUDI 0OlZ GPS SeTUaSUJOH JO UIe eveOU PG PQeSaUJOH eSSPST EP OPU QSeTT JU EVSJOH OPSNal PQeSaUJPO
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 427
-
5($',1*
6
$,'6PLWK %RE&+(673RWDEOH&KHVW
Fig. 4.8 User interface
The diagram shows the main window on the Cosima which appearsafter switching on. It is the central window for all displays and the start-ing point for the selection of other functions.
/P .eBnJnH
EYQPTVSe OVNCeS *OUeSOal aENJOJTUSaUJPO OVNCeS PO UIe QlaUe SeaEeS *U JT DPVOUeE DPOTeDVUJvelZ
PaUJeOU OVNCeS *%
EYaNJOaUJPO7JeX
PaUJeOU OaNe 'PS UIe eYQPTVSe DVSSeOUlZ CeJOH QSPDeTTeE
0QeSaUJOH NPEeT JO UIe DaTTeUUe TlPU
3 PSJNaSZ eSaTVSe
6 4eDPOEaSZ eSaTVSe
PSJPSJUZ DaTTeUUe QSPDeTTJOH
8aSOJOH JOEJDaUPS 'laTIeT XIeO a XaSOJOH PDDVST OPUJGZ CVTUPNeS 4eSvJDe JG OeDeTTaSZ
0QeSaUJOH NPEe TXJUDI 'PS TeleDUJOH NPEeT lJLe QSJNaSZTeDPOEaSZ eSaTVSe aOE QSJPSJUZ DaTTeUUeQSPDeTTJOH
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
428 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
/P .eBnJnH
PSPHSeTT JOEJDaUPS
/P DaTTeUUe ZeU JOTeSUeE aGUeS SeNPval
CaTTeUUe aDDeQUeE QSeQaSJOH UP SeaE
ReaEJOH CeHVO o ESaTVSe CeHVO
ESaTVSe JO QSPHSeTT o ESaTVSe DPODlVEeE
PSPDeTTJOH DPNQleUeE aXaJUJOH DaTTeUUe SeNPval
PSPUeDUeEVOQSPDeTTeE eYQPTVSeT /VNCeS PG eYQPTVSeT UIaU XeSe QSPUeDUeE GSPN aVUPNaUJD eSaTVSe PSOVNCeS PG eYQPTVSeT OPU ZeU QSPDeTTeE
"ENJOJTUSaUJPO CallT VQ UIe lJTU PG JNaHeT ZeU UP Ce QSPDeTTeE
6UJlJUJeTCallT VQ VUJlJUJeT
%JTQlaZ aSea %JTQlaZT UIe eYQPTVSe DVSSeOUlZ CeJOH SeaE PVU
0QeSaUJOH NPEe %JTQlaZT UIe DVSSeOU QlaUe SeaEeS NPEe
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
1$3 QMBUe SeBEeS 44
PCR Eleva Ver. 1.1 4ZTUeN EeTDSJQUJPO 429
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
44 1$3 QMBUe SeBEeS
4 4ZTUeN EeTDSJQUJPO PCR Eleva Ver. 1.1
-
5BCMe PG $PnUenUT
5 4XJUDIJnH UIe TZTUeN PnPGG 5
5.1 Switching on ..................................................................... 5-3
5.2 Switching off ..................................................................... 5-35.2.1 One-console system ............................................ 5-35.2.2 Multiple-console system ..................................... 5-4
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 51
-
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
52 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
5 4XJUDIJnH UIe TZTUeN PnPGG
51 4XJUDIJnH Pn
Philips recommends the following sequence:
Switch on PC. Log on to program.
Enter user name Enter password
Switch on the plate reader.
You will find information about the emergency mode in chapter 7.4.
52 4XJUDIJnH PGG
521 0neDPnTPMe TZTUeN
/05& 5IFTZTUFNJTEFTJHOFEGPSDPOUJOVPVTPQFSBUJPO*UJTUIFSFGPSFPOMZOFDFTTBSZUPTXJUDIPGGBMMDPNQPOFOUTJOUIFFWFOUPGQSPMPOHFETUPQQBHFT
Switch off plate reader.System Press SystemPIJlJ
QT .
eEJDa
l 4ZT
UeN
T /
PveN
CeS
"
'
eO
PCR Eleva Ver. 1.1 4XJUDIJOH UIe TZTUeN POPGG 5
-
The System section appears
([LW Press Exit Switch off PC
522 .VMUJQMeDPnTPMe TZTUeN
/05& 5IFTZTUFNJTEFTJHOFEGPSDPOUJOVPVTPQFSBUJPO*UJTUIFSFGPSFPOMZOFDFTTBSZUPTXJUDIPGGBMMDPNQPOFOUTJOUIFFWFOUPGQSPMPOHFETUPQQBHFT
Proceed as under chapter 5.2.1, but switch the server on first and offlast.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
52 4XJUDIJnH PGG
54 4XJUDIJOH UIe TZTUeN POPGG PCR Eleva Ver. 1.1
-
5BCMe PG $PnUenUT
1$3 QMBUe SeBEeS
6.1 The PCR Compano plate reader ....................................... 6-36.1.1 Reading out an image plate ................................ 6-36.1.2 Performing secondary erasure ............................. 6-76.1.3 Performing primary erasure ............................... 6-86.1.4 Switching on/off ................................................ 6-8
6.2 The PCR Corado, CosimaX and S Hi-res plate readers ......................................................................................... 6-116.2.1 Distinguishing features ..................................... 6-116.2.2 Switching on/off .............................................. 6-116.2.3 Reading out an image plate .............................. 6-136.2.4 Performing secondary erasure ........................... 6-176.2.5 Performing primary erasure ............................. 6-176.2.6 What to do if the cassette is not scanned .......... 6-18
6.3 The PCR S and S Plus plate readers ................................ 6-196.3.1 Distinguishing features ..................................... 6-196.3.2 Switching on/off .............................................. 6-196.3.3 Reading out an image plate .............................. 6-216.3.4 Performing secondary erasure ........................... 6-256.3.5 Performing primary erasure .............................. 6-26
6.4 The PCR AC500, AC5000 and Cosima plate readers ......................................................................................... 6-286.4.1 Distinguishing features ..................................... 6-286.4.2 Switching on/off .............................................. 6-286.4.3 Insert the cassette ............................................. 6-316.4.4 Removing the cassette ...................................... 6-346.4.5 Screen saver ...................................................... 6-346.4.6 Working with information windows ................ 6-346.4.7 Processing a cassette with priority ..................... 6-356.4.8 Performing secondary erasure ........................... 6-376.4.9 Performing primary erasure .............................. 6-386.4.10 Adjusting gray levels manually .......................... 6-406.4.11 Processing unregistered image plates ................. 6-416.4.12 Utilities ............................................................ 6-436.4.13 Error messages .................................................. 6-58
6.5 Making exposures ............................................................ 6-606.5.1 Exposure parameters ........................................ 6-606.5.2 Positioning the cassette ..................................... 6-60
6.6 Cassettes and image plates .............................................. 6-616.6.1 Handling cassettes and image plates ................. 6-61
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 5aCle PG CPOUeOUT 1
-
6.7 Notes on radiography ...................................................... 6-626.7.1 UNIQUE, UM and DRR ................................ 6-626.7.2 Notes on radiography techniques with UNIQUE
.......................................................................... 6-636.8 The Exposure Index ........................................................ 6-63
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
5BCMe PG $PnUenUT
2 5aCle PG CPOUeOUT PCR Eleva Ver. 1.1
-
1$3 QMBUe SeBEeS
1 5Ie 1$3 $PNQBnP QMBUe SeBEeS
11 3eBEJnH PVU Bn JNBHe QMBUe
To read out an image plate you must slide the plate cassette into the plate reader, have the image plate read out and remove the plate cassette again.IInsert the cassette
*ODPSSFDUMZJOTFSUJOHUIFDBTTFUUFDBOEBNBHFUIFQMBUFSFBEFS&OTVSFUIBUUIFDBTTFUUFJTOPUJOTFSUFEVQTJEFEPXOBUBOBOHMFPSJODPSSFDUMZBMJHOFE*GZPVVTFEFYQPTVSFNBSLTXIFONBLJOHUIFFYQPTVSFSFNPWFUIFTFGSPNUIFDBTTFUUFCFGPSFJOTFSUJOHJU
8"3/*/(
/05& *GZPVBSFXPSLJOHXJUIPVUBCBSDPEFSFBEFSUIFMPBEMBNQPOMZBQQFBSTJOUIFi&YBNJOBUJPOwTFDUJPO*GZPVBSFXPSLJOHXJUIBCBSDPEFSFBEFSZPVNVTUGJSTUPGBMMBTTJHOUIFQMBUFDBTTFUUFUPBWJFXVTJOHUIFPQFSBUPSTDPOTPMF0UIFSXJTFQSPDFTTJOHPGUIJTQMBUFDBTTFUUFXJMMCFEFOJFE3FNPWFUIFDBTTFUUFBHBJOBOEBTTJHOUIFQMBUFDBTTFUUFUPBWJFXVTJOHUIFCBSDPEFSFBEFSPOUIFPQFSBUPSTDPOTPMF
&OTVSFUIBUZPVJOTFSUUIFDBTTFUUFTUSBJHIUBOESJHIUBMPOHUIFSJHIUGFFEFEHF
8"3/*/(
The patient has been X-rayed. The exposures can be found onone or more plate cassettes.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS
-
Check that the load lamp is lit on the control panel of the platereader.
Turn the cassette so that the side to be exposed faces down, andthe barcode window goes in first and faces up.
;
Slide the cassette into the cassette compartment straight along theright feed edge.
Never insert the cassette like this:
Fig. 6.1 Wrong! Gap between cassette and feed edge
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
1 5Ie 1$3 $PNQBnP QMBUe SeBEeS
4 PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
Fig. 6.2 Wrong! Cassette is at an angle
The cassette slots into place. You will feel slight resistance wheninserting the cassette, which you must overcome.
The load lamp on the control panel of the plate reader goes out.Image plate readout begins automatically.
The status displays flash on the control panel:The first arrow flashes during readout. The second arrow flasheswhen erasure begins until the image plate is reloaded in the cassette.
Remove cassette
/05& 8IJMFUIFJNBHFQMBUFJTCFJOHSFBEPVUUIFDBTTFUUFJTMPDLFEJOUIFQMBUFSFBEFSUIFTUBUVTEJTQMBZTGMBTI
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie 1$3 $PNQBnP QMBUe SeBEeS 1
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS 5
-
Check that the unload lamp is lit on the control panel.
Pull the cassette in a straight line and slowly out of the cassettecompartment.
The unload lamp goes out; the load lamp lights up.
The cassette is ready for use again. The exposure which has been read out is added to the patients
exposures. Unless you have selected another patient or an-other exposure for processing, the exposure which has beenread out appears on the monitor.
/05& *GZPVXPSLXJUIUIFNVMUJQMFDPOTPMFTZTUFNUIFFYQPTVSFXIJDIIBTCFFOSFBEPVUJTTUPSFEBOEEJTQMBZFEPOUIFDPOTPMFPOXIJDIZPVIBWFBTTJHOFEUIFQMBUFDBTTFUUFUPBWJFX
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
1 5Ie 1$3 $PNQBnP QMBUe SeBEeS
PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
12 1eSGPSNJnH TeDPnEBSZ eSBTVSe
Normally, the image plates are automatically erased after readout. If,however, an image plate has not been used in the past 8 hours, so-calledsecondary erasure is necessary. Here, the cassette is merely erased, andnot read out.
With time, the image plates absorb natural radiation, with the resultthat images with low exposure levels have a higher noise component. Inapplications with a low radiation dose you should therefore performsecondary erasure on the image plate after prolonged storage.
Proceed as follows:
Press the erase button. The field beside it displays the secondary erasure mode for
the next cassette to be inserted.
Insert cassette.During erasure, the status display for cassette processing flashes.On the monitor, a message concerning the plate reader modeappears above the progress indicator. After erasure, the platereader automatically switches back into normal read mode, mean-ing the erasure procedure only ever applies to one cassette.P
IJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie 1$3 $PNQBnP QMBUe SeBEeS 1
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS 7
-
When the unload lamp lights up, erasure is complete. The cassettecan be taken out and used for a new exposure.
1 1eSGPSNJnH QSJNBSZ eSBTVSe
Normally, the image plates are automatically erased after readout. Afterincorrect exposure, however, special, primary erasure is necessary. Here,the cassette is merely erased, and not read out.
/05& *GUIFQMBUFSFBEFSEFUFDUTBOJODPSSFDUMZFYQPTFEJNBHFQMBUFBOFSSPSNFTTBHFUPUIBUFGGFDUBQQFBST:PVNBZOPUVTFJODPSSFDUMZFYQPTFEJNBHFQMBUFTGPSBSPVOEIPVSTBGUFSQSJNBSZFSBTVSF
Proceed as follows:
Press the erase button twice. The field beside it displays the primary erasure mode for the
next cassette to be inserted.
Insert cassette.During erasure, the status display for cassette processing flashes.After erasure, the plate reader automatically switches back intonormal read mode, meaning the erasure procedure only ever ap-plies to one cassette.
When the unload lamp lights up, erasure is complete. The cassettecan be taken out and used for a new exposure.
14 4XJUDIJnH PnPGG
/05& 5IFQMBUFSFBEFSJTEFTJHOFEGPSDPOUJOVPVTPQFSBUJPO*UJTUIFSFGPSFPOMZOFDFTTBSZUPTXJUDIJUPGGJOUIFFWFOUPGQSPMPOHFETUPQQBHFT
Power ON
The PC and the application program on the operator's consolehave been started up. PIJ
lJQT .
eEJDa
l 4ZT
UeN
T /
PveN
CeS
1 5Ie 1$3 $PNQBnP QMBUe SeBEeS
8 PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
Press main switch to the I position.
The pilot lamp on the control panel of the plate reader lightsup.
The plate reader begins the start-up procedure, which takes ashort time. Then the load lamp lights up.
The plate reader is ready when the green load lamp lightsup.
Power OFF
Processing of the last cassette is complete.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie 1$3 $PNQBnP QMBUe SeBEeS 1
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS 9
-
The unload lamp lights up.
Remove the cassette from the cassette compartment.
Press main switch to the 0 position.
The PCR Compano is now properly switched off.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
1 5Ie 1$3 $PNQBnP QMBUe SeBEeS
1 PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
2 5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUeSeBEeST
21 %JTUJnHVJTIJnH GeBUVSeT
All three plate readers perform the same functions.The CCorado and CCosimaX plate readers are identical externally.The SS Hi-res plate reader has only one cassette slot.The ooperation of all three plate readers is identical; the operation of aplate reader with four cassette slots is described below.
22 4XJUDIJnH PnPGG
/05& 5IFQMBUFSFBEFSJTEFTJHOFEGPSDPOUJOVPVTPQFSBUJPO*UJTPOMZOFDFTTBSZUPTXJUDIJUPGGJOUIFFWFOUPGQSPMPOHFETUPQQBHFT%POPUTXJUDIPOUIFQMBUFSFBEFSBHBJOJNNFEJBUFMZBGUFSTXJUDIJOHJUPGGXBJUBUMFBTUT
Power ON
The pilot lamp lights up.
If the pilot lamp does not light up, the main switch may not beswitched on. If so, press the main switch to the I position (on).
Switch on the plate reader.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST 2
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS 11
-
The operating system loads the programs required and per-forms a selftest. Various different displays appear.
After the start-up procedure the main window appears: (Onlyone cassette compartment is displayed in the main window ofthe S Hi-res)
The plate reader is ready when the load lamps of the cassettecompartments light up.
Power OFF
Make sure that processing of all cassettes is complete (unloadlamps flashing).
Remove all cassettes from the cassette compartments. Touch the Shutdown button in the main window.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
2 5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST
12 PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
The following confirmation window appears:
Touch the Power OFF button and then OK.
After a short time the plate reader switches itself off automatically.The main switch can be left turned on.
2 3eBEJnH PVU Bn JNBHe QMBUe
The image does not appear on-screen until read-out is finished.
To read out an image plate you must slide the plate cassette into the plate reader, have the image plate read out and remove the plate cassette again.Insert the cassette
*ODPSSFDUMZJOTFSUJOHUIFDBTTFUUFDBOEBNBHFUIFQMBUFSFBEFS&OTVSFUIBUUIFDBTTFUUFJTOPUJOTFSUFEVQTJEFEPXOBUBOBOHMFPSJODPSSFDUMZBMJHOFE*GZPVVTFEFYQPTVSFNBSLTXIFONBLJOHUIFFYQPTVSFSFNPWFUIFTFGSPNUIFDBTTFUUFCFGPSFJOTFSUJOHJU
8"3/*/(
The patient has been X-rayed. The exposures can be found onone or more plate cassettes.
Turn the cassette so that the front faces up (barcode window goesin first and faces up).
Check that the load lamp of the cassette compartment is lit.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST 2
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS 1
-
Place the cassette into the cassette compartment with the frontfacing up.
/RDGODPSLVOLW
Guide the cassette along the rright feed edge, until it slots intothe cassette compartment.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
2 5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST
14 PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
NNever insert the cassette like this:
Fig. 6.3 Wrong! Gap between cassette and feed edge
Fig. 6.4 Wrong! Cassette is at an angle
When the cassette correctly slots into the cassette compartment, the loadlamp of the cassette compartment goes out and the processing lamplights up. Image plate readout begins. During this, the cassette is lockedand cannot be taken out. The progress of processing is shown on themonitor.
Cassette processing is complete when the unload lamp of the cassettecompartment lights up and the processed image is shown in full.
/05& *GZPVBSFXPSLJOHXJUIPVUBCBSDPEFSFBEFSUIFMPBEMBNQEPFTOPUBQQFBSVOUJMZPVIBWFTFMFDUFEBWJFXJOi&YBNJOBUJPOw
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST 2
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS 15
-
RRemove cassette
Check that the blue unload lamp of the cassette compartment isflashing.
8QORDGODPSIODVKLQJ
Pull the cassette in a straight line and slowly out of the cassettecompartment. The unload lamp goes out; the load lamp lights up. The cassette is ready for use again. The exposure which has been read out is added to the patients
exposures. Unless you have selected another patient or an-other exposure for processing, the exposure which has beenread out appears on the monitor.
/05& *GZPVXPSLXJUIUIFNVMUJQMFDPOTPMFTZTUFNUIFFYQPTVSFXIJDIIBTCFFOSFBEPVUJTTUPSFEBOEEJTQMBZFEPOUIFDPOTPMFPOXIJDIZPVIBWFBTTJHOFEUIFQMBUFDBTTFUUFUPBWJFX
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
2 5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST
1 PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
24 1eSGPSNJnH TeDPnEBSZ eSBTVSe
Normally, the image plates are automatically erased after readout. If,however, an image plate has not been used in the past 8 hours, so-calledsecondary erasure is necessary. Here, the cassette is merely erased, andnot read out.
With time, the image plates absorb natural radiation, with the resultthat images with low exposure levels have a higher noise component. Inapplications with a low radiation dose you should therefore performsecondary erasure on the image plate after prolonged storage.
Proceed as follows:
Select the mode in the main window. Select Secondary erasure.
1
2
The main window appears once again.
The Secondary erasure mode is now activated for the next cas-sette to be inserted. After erasure, the plate reader automaticallyswitches back into normal read mode, meaning the erasure pro-cedure only ever applies to one cassette.
Insert cassette. During erasure the indicator for the cassette compartment in
use flashes.
The status of erasure is displayed. When the unload lamp lights up, erasure is complete. The
cassette can be taken out and used for a new exposure.
25 1eSGPSNJnH QSJNBSZ eSBTVSe
Normally, the image plates are automatically erased after readout. Afterincorrect exposure, however, special, primary erasure is necessary. Here,the cassette is merely erased, and not read out.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
"'
eO
5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST 2
PCR Eleva Ver. 1.1 PCR QlaUe SeaEeS 17
-
/05& *GUIFQMBUFSFBEFSEFUFDUTBOJODPSSFDUMZFYQPTFEJNBHFQMBUFBOFSSPSNFTTBHFUPUIBUFGGFDUBQQFBST:PVNBZOPUVTFJODPSSFDUMZFYQPTFEJNBHFQMBUFTGPSBSPVOEIPVSTBGUFSQSJNBSZFSBTVSF
Proceed as follows:
Select the mode in the main window. Select Primary erasure.
1
2
The main window appears once again.
The Primary erasure mode is now activated for the next cassetteto be inserted. After erasure, the plate reader automaticallyswitches back into normal read mode, meaning the erasure pro-cedure only ever applies to one cassette.
Insert cassette. During erasure the indicator for the cassette compartment in
use flashes.
The status of erasure is displayed. When the unload lamp lights up, erasure is complete. The
cassette can be taken out and used for a new exposure.
2 8IBU UP EP JG UIe DBTTeUUe JT nPU TDBnneE
If you have forgotten to scan a cassettes barcode before loading it intothe plate reader, an error message, stating that the image plate cannotbe read out, appears on the monitor, as the plate reader does not havethe data for reading out the cassette. At the same time, the following,general error message appears:
Touch Stop alarm on the monitor and confirm unloading ofthe cassette by pressing Take out cassette.
PIJlJQ
T .eE
JDal 4
ZTUe
NT
/Pv
eNCe
S
2 5Ie 1$3 $PSBEP $PTJNB9 BnE 4 )JSeT QMBUe SeBEeST
18 PCR QlaUe SeaEeS PCR Eleva Ver. 1.1
-
The error message disappears; the cassette is ejected. Remove the cassette and enter the barcode with the correspond-
ing patient and examination data on the operator's console.
Reload the cassette in a free cassette compartment.
5Ie 1$3 4 BnE 4 1MVT QMBUe SeBEeST
1 %JTUJnHVJTIJnH GeBUVSeT
The S and S Plus plate readers a