NW6S_6SF-UK

104
ENGLISH INSTRUCTION MANUAL NON-MYDRIATIC RETINAL CAMERA TRC-NW6SF/ TRC-NW6S

Transcript of NW6S_6SF-UK

ENGLISH

INSTRUCTION MANUAL

NON-MYDRIATIC RETINAL CAMERA

TRC-NW6SF/ TRC-NW6S

Copyrights and trademarksNon-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S is a trademark of the TOPCON Corporation. TOPCON 2000

INTRODUCTION

Thank you for purchasing the TOPCON Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S. This manual contains instructions for the TOPCON Non-Mydriatic Retinal Camera. This includes the main features, basic operation, troubleshooting, and the maintenance and cleaning of this instrument. Please read the section Warning labels on page 5 and the section Safety precautions on page 6 carefully, before operating the instrument. To ensure the best performance of this instrument, read all instructions carefully and place this manual in a convenient location for future reference.

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

This machine offers the following innovative featuresObserve the fundus of the eye and take images without using mydriatics. Improved picture quality and ease of use. Allows images of the periphery, using the alignment point and the external fixation target for peripheral photography. Allows Fluoro-photography by changing the photography mode (TRC-NW6SF only). Allows Fluoro-observation by using the built-in color video monitor (TRC-NW6SF only).

4

Precautions

This machine is a precision instrument. Be sure to install and use it in a controlled environment under normal temperature, humidity and atmospheric pressure conditions (10~40C, 30-85%, 70~106kPa) and avoid direct exposure to sunlight. Never install the machine on a slope or an unstable surface. Before using the machine, ensure that all cables are correctly connected. Use a power supply within the range of 10% of the rated voltage (50/60Hz). Ensure the surrounding area is always clean. When the instrument is not in use, turn off the Power switch, cap the objective lens, and install the dust cover. To ensure the best imaging results, handle the objective lens with particular care and keep it free of dust, fingerprints, or other stains.

Selecting externally connected equipmentThe TOPCON TRC-NW6SF/ TRC-NW6S complies with the CE standard. Before connecting a personal computer, image recorder, image printer, TV camera, or monitor to the TOPCON product, ensure that such external equipment is in compliance with the CE marking.Note: The illustrations in this manual pertain to the model TRC-NW6SF only. Please note that the TRC-NW6S differs from the SF model.

Warning labelsImportant warnings are placed on the products and inserted in the instruction manuals, in order to encourage the safe use of products and prevent any danger to the operator and others or damage to existing facilities. We recommend that the following information about labels and icons be read before going on to read "Safety precautions" and the rest of the manual.

Meaning of labels

Label

Meaning

Ignoring or disregarding this label may lead to death or serious injury. WARNING

Ignoring or disregarding this label may result in personal injury or severe damage to the instrument or other facilities. PRECAUTIONS Possible injuries include wounds, burns, electric shocks, etc. Damage to other facilities means extensive damage to buildings, equipment and furniture.

INSTRUCTION MANUAL

5

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

Meaning of icons

Icons

Meaning This icon indicates Caution. A further explanation or picture is placed close to the icon. This icon indicates Prohibition (DO NOT!). A further explanation or picture is placed close to the icon. This indicates Mandatory Action (DO!). A further explanation or picture is placed close to the icon.

Safety precautionsThis instruction manual details the safety precautions that need to be observed to prevent accidents. Always observe these precautions and use the instrument correctly. WARNING Icons Precautions To avoid the risk of electric shock, disconnect the cable from the power supply before assembling the instrument. Always ensure the power supply is disconnected before assembly. Ensure the power supply is equipped with a 3-plug AC receptacle to safeguard against power leakage. Power leakage could cause fire or electric shock. To avoid the risk of electric shock, do not attempt disassembling, rebuilding and/or repairs. Always consult your dealer in case of repairs. Page

6

29

36

95

WARNING Icons Precautions To avoid the risk of electric shock do not remove any instrument components, TV relay lens covers, chinrest or power supply components, other than the lamp house cover. To avoid the risk of electric shock, always disconnect the power cable from the instrument before removing the fuse cover. Always replace the fuse cover before reconnecting the instrument to the power supply. Install the instrument in a clean, dry area.-

Page

INSTRUCTION MANUAL

95

89

To avoid the risk of fire or electric shock, do not place any liquids near the instrument.

-

To avoid the risk of electric shock, do not insert metal items into any vents and/or slots.

-

To avoid the risk of fire, always use the correctly rated fuse as specified on the fuse holder.

-

To avoid the risk of fire, always use the correctly rated fuse as specified on the fuse holder.

89

7

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

CAUTION Icons Precautions Ensure that the monitor lamp is no brighter than necessary, to avoid any discomfort or injury to the patient. Ensure that the photography light is no brighter than necessary, to avoid any discomfort or injury to the patient. Do not continue visible fluoro-observation any longer than necessary (TRC-NW6SF only), to avoid any discomfort or injury to the patient. Always install the instrument on a fixed, stable surface, to prevent damage or injury. Page

8

-

65, 66

77

29, 34, 85

To avoid potential injury, do not place your hand or fingers into the area between the instrument body and the power supply unit. To avoid potential burns, do not touch the lamp during, or immediately after use.

68

86

To avoid the risk of electric shock, never handle a plug with wet hands.

36

To avoid the risk of electric shock, never touch the xenon lamp immediately after a flash or burn out.

87

To avoid potential injury to the patient, always watch the patient carefully while adjusting the chinrest.

61

CAUTION Icons Precautions To avoid potential injury to the patient while moving the instrument, always keep the instrument at a safe distance from the patient. Always ensure the TV relay lens and TV camera are securely mounted, to avoid potential injury during adjustment and operation. Use an approved automatic instrument table to avoid potential injury while moving the instrument. Page

INSTRUCTION MANUAL

68

32, 85

34

Ensure the fixing knob at the bottom of the instrument is secure, before moving the instrument.

34

The instrument should always be carried by two people, both holding the bottom of the instrument. Attempting to carry the instrument alone could result in backache or other injury. Attempting to carry the instrument without holding it at the bottom could result in injury or damage. Ensure that the camera mount lever is firmly tightened to prevent the TV camera from falling and causing injury or damage. Ensure that the lens mount lever is firmly tightened to prevent the TV relay lens from falling and causing injury or damage. To avoid the risk of electric shock, always disconnect the power supply and unplug the power cable before replacing the lamp. To prevent the instrument from falling and to avoid injury during movement, use the base locking knob to secure the power supply unit to the instrument.

29, 34

85

85

86, 87

-

9

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

Operation and maintenancePurposeThis Non-Mydriatic Retinal Camera is a precision electronic medical instrument, and should only be used under the direction of a medical doctor.

User maintenanceTo ensure the safe and proper operation of the instrument, unless otherwise specified in this manual, all maintenance should be carried out by a trained service technician. The following maintenance tasks, however, can be performed by the user. Please refer to the relevant text in this manual for the maintenance method. Replacing lamps The illumination lamp and xenon lamp may be replaced by the user. For details, see Replacing the illumination lamp on page 86 and Replacing the xenon lamp on page 87. Fuses used with the instrument may be replaced by the user. For details, see Changing the fuses on page 89. The objective lens may be cleaned by the user. For details, see Cleaning the objective lens on page 91.

Replacing fuses

Cleaning the objective lens

10

TOPCON Instruments general statements

TOPCON shall not bear any responsibility for damage due to fire, earthquakes, actions by third persons and other accidents, or the negligence and misuse by the user and use under unusual conditions. TOPCON shall not bear any responsibility for damage resulting from the inability to use this equipment, such as a loss of business profit and suspension of business. TOPCON shall not bear any responsibility for damage caused by operations other than those described in this Instruction Manual. The diagnoses that are made are the responsibility of relevant doctors, and TOPCON shall not take any responsibility for the results of such diagnoses.

Warning labels and positionsTo ensure safe operation of this equipment, warning labels are provided. Please adhere to the following warning instructions.

INSTRUCTION MANUAL

11

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

CAUTION To avoid potential injury to the patient, always watch the patient carefully while adjusting the chinrest.

CAUTION Never touch the patient's eyes or nose with the instrument during operation as this may result in personal injury.

WARNING Electrical shock may cause burns or fire. Turn the main power switch OFF and UNPLUG the power cable before replacing fuses. Only replace with fuses of the correct rating.

CAUTION To avoid the risk of electric shock, always disconnect the power supply and unplug the power cable before replacing the lamp. To avoid potential burns, do not touch the lamp during, or immediately after use.

WARNING To avoid the risk of electric shock, do not attempt to remove the instrument cover.

No user serviceable parts inside, refer servicing to qualified personnel.

WARNING Electrical shock may cause burns or fire. Turn the main power switch OFF and UNPLUG the power cable before replacing fuses. Only replace with fuses of the correct rating.

12

Contents

INTRODUCTIONThis machine offers the following innovative features Precautions Warning labelsMeaning of labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Meaning of icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4 4 55 6

Safety precautions Operation and maintenancePurpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . User maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6 1010 10

TOPCON Instruments general statements Warning labels and positions

11 11

11.1 1.2 1.3 1.4 1.5

COMPONENTSComponent names Control panel components Monitor screen Standard accessories Assembly components 17 21 22 24 26

Non-Mydriatic Retinal Camera TRC-NW6SF/TRC-NW6S

22.1 2.2 2.3 2.4 2.5 2.62.6.1 2.6.2 2.6.3 2.6.4

ASSEMBLY PROCEDUREAssembling the instrument body Connecting the tv relay lens with the instrument body Confirmation after assembly Installing the instrument Connecting the power cable Connecting the external device 29 32 33 34 36 36

Connecting the TV camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Connecting the external recording device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Outputting images to an external monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 VIDEO OUT terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

2.72.7.1 2.7.2 2.7.3 2.7.4 2.7.5 2.7.6 2.7.7 2.7.8

Menu setting

39

Preparation for menu setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Displaying the Menu screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Returning to the Monitor screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Setting the Record/Playback mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Switching of fixation target on/flicker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Switching of internal/external fixation targets . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Fixation target pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

2.8

Reset from power save state

43

33.1 3.2 3.33.3.1 3.3.2

INITIAL SETTINGPreparation for initial set-up Setting the record/playback mode Setting the fixation target 45 46 47

Switching the fixation target on/flicker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Switching of internal/external fixation targets . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

3.43.4.1 3.4.2 3.4.3

Setting peripheral photography

48

Fixation target pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 R/L reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Display mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

3.53.5.1 3.5.2 3.5.3 3.5.4

Setting the screen display

51

Flash level compensation display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Flash level display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Illumination level display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Picture angle display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

xiv

3.5.5 3.5.6

Illumination aperture display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Peripheral fixation position display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

3.6 3.73.7.1 3.7.2 3.7.3 3.7.4 3.7.5 3.7.6 3.7.7

System setting Setting the initial state

55 55

Flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Commercial power supply frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 LED display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Timer sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Operation sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Power save time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Quick mirror up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

3.8

Setting the language

60

44.14.1.1 4.1.2

BASIC OPERATIONPreparation for photography 61Connecting to the power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Positioning the patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

4.24.2.1 4.2.2 4.2.3 4.2.4 4.2.5 4.2.6 4.2.7

Color photography (center)

64

Setting the photography mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Setting the picture angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Setting illumination level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Setting the flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Changing the diopter compensation lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

4.34.3.1 4.3.2 4.3.3 4.3.4 4.3.5 4.3.6 4.3.7 4.3.8 4.3.9

Fluoro-photography (center)

75

Setting the photography mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Setting the picture angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Setting illumination level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Setting the flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Changing the diopter compensation lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Preparing the patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Fluoro-observation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

4.44.4.1 4.4.2 4.4.3

Peripheral photography

79

Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 Other settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

INSTRUCTION MANUAL

xv

Non-Mydriatic Retinal Camera TRC-NW6SF/TRC-NW6S

4.54.5.1 4.5.2 4.5.3 4.5.4 4.5.5 4.5.6

Anterior segment photography

82

Setting the photography mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Setting the picture angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Setting the illumination level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Changing the diopter compensation lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

4.6

Finishing

84

55.1 5.25.2.1 5.2.2

MAINTENANCE AND CHECKSDaily care Changing the lamp 85 86

Replacing the illumination lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 Replacing the xenon lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87

5.3 5.4 5.5 5.65.6.1 5.6.2

Changing the fuses Supplying the chinrest tissue Adjusting the color video monitor Cleaning the instrument

89 90 90 91

Cleaning the dust cover, control panel and monitor screen . . . . . . . . . . . . . . . . 91 Cleaning the objective lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

5.7 5.85.8.1

List of expendables and spare parts Optional accessories Contacting TOPCON

92 93 94

Adjustable instrument table AIT-11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

5.9

66.1 6.2 6.3 6.4

TROUBLESHOOTINGBefore requesting service Troubleshooting table Error code list Emergency chinrest operation 95 96 99 99

7

SPECIFICATIONS

xvi

1

COMPONENTS

1.1 Component names

(1)

(2) (9) (3) (10) (11) (12) (13) (4) (14) (15) (16) (5) (6) (18) (7) (19) (8) (20)

(17)

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

(1) (2) (3) (4) (5) (6) (7) (8) (9) (10) (11) (12) (13) (14) (15) (16) (17) (18) (19) (20)

TV camera connection terminal Diopter compensation lens selector Focusing knob IR filter selector Image quality adjustment knob Base brake knob Vertical position mark Fuse holder TV relay lens attaching lever TV relay lens Angle selector Mode selector * Fluoro-observation knob * Name plate slot Color video monitor Photography switch Omni-directional joystick Power lamp Control panel Fixing knob (for use when carrying the instrument)

* TRC-NW6SF only

18

(21)

(22)

(26)

(27) (23) (28) (24)

(25)

(21) (22) (23) (24) (25) (26) (27) (28)

Forehead rest Canthus marker Chin-rest tissue pin Lamp house cover Chinrest Anterior segment fixation target Objective lens Lamp house cover screw

INSTRUCTION MANUAL

19

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

(34)

(29)

(30)

(31) (32) (33)

(29) (30) (31) (32) (33) (34)

Base Power switch Fuse holder Power supply unit External output terminal: External fixation target

20

1.2 Control panel components

(40) (35) (41)

(36)

(37)

(42)

(38)

(39)

(35)

Menu switch Displays the Menu screen.

(36)

Split switch Turns split lines on/off.

(37)

Flash Level switch ([-], [reset], [+]) Adjusts the flash level for the patients eye condition.

(38)

Illumination Level switch ([-], [+]) Adjusts the illumination level for the patients eye condition.

(39)

Chinrest switch ([down], [up]) Positions the patients head by electronically moving the chinrest up/down.

(40)

Illumination Aperture switch * Changes the flash level for color photography. Also, changes the illumination aperture in fluoro-photography with 45/ picture angle.

(41)

Timer switch * Displays the elapsed time on the monitor/photography image.

* TRC-NW6SF only

INSTRUCTION MANUAL

21

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

(42)

Periphery Fixation switch ([left turn], [reset], [right turn]) Switches the position of the internal fixation target, therefore guiding the patients eye to the periphery fixation point.

1.3 Monitor screenMonitor screen(43) (44) (45) (46) (47) (48) (49) (50) (54) (51) (52)

(53)

(43) (44) (45) (46) (47) (48) (49) (50) (51) (52) (53) (54) * TRC-NW6SF only

Nameplate/timer image Xenon charging display Flash level compensation display Flash level display IR warning display Picture angle display Illumination aperture display * Fixation target position display Scale Alignment bright spots Split lines Illumination level

22

Menu screen

(55)

(56)

Preview display (color)(57)

(58)

(55) (56) (57) (58)

Cursor Input guide display Nameplate/timer image Photography image

INSTRUCTION MANUAL

23

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

1.4 Standard accessoriesUpon unpacking, make sure that all the following standard accessories are delivered. Figures in parentheses, ( ), indicate the quantity.

(59)

(60)

(61)

(62)

(63)

(64)

(65) (66)

24

(67)

(68)

(69)

(70)

(71)

(72)

(59) (60) (61) (62) (63) (64) (65) (66) (67) (68) (69) (70)

Power cable (1) Nameplate (10) Illumination lamp (1) Fuse (9) Chinrest tissue (1) Chinrest tissue pin (2) Emergency chinrest knob (1) Spare parts case (1) Instruction Manual Warranty (1) Dust cover (1) External fixation target (1)

INSTRUCTION MANUAL

25

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

(71) (72)

Power cable (1) TV camera signal cable (1)

1.5 Assembly components

(6) (3)

(7) (4) (1) + (2)

(9)

(5)

(8)

(10)

(11)

(12)

(13)

(15)

(16)

(17)

(14)

(20) (18) (19)

26

1.5.1

ComponentsDescription Quantity Container

(1) Instrument body (2) TV relay lens (3) Rail cover (4) TV camera power cable (5) TV camera signal cable (6) Power cable (7) External fixation target (8) Star screwdriver (9) Hexagon rod spanner (10) Focusing tool (11) Dust cover (12) Nameplate (13) Chin rest tissue pin (spare) (14) Illumination lamp (15) Chinrest tissue (16) Fuse (spare) (17) Emergency chinrest knob (18) Accessories case (19) TRC-NW6SF Instruction Manual (20) Warranty

1 1 2 1 1 1 1 1 1 1 1 10 2 1 1 9 1 1 1 1

O O O O O O O O O Accessories case O Accessories case Accessories case Accessories case Accessories case Accessories case Accessories case O O O

INSTRUCTION MANUAL

27

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

28

2

ASSEMBLY PROCEDURE

2.1 Assembling the instrument bodyTo avoid the risk of electric shock, disconnect the cable from the power supply before assembling the instrument. Never plug in the power cable before assembly.

WARNING

CAUTION

CAUTION

Note:

Since the upper part and lower part of the instrument are connected only by the power cable, take them out together so they do not separate from each other.

Always install the instrument on a fixed, stable surface, to prevent damage or injury.

The instrument should always be carried by two people, both holding the bottom of the instrument. Attempting to carry the instrument alone could result in backache or other injury. Attempting to carry the instrument without holding it at the bottom could result in injury or damage.

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

1 Remove the instrument body (1) from the container and place it on the appropriate table. 2 Slightly raise the omni-directional joystick, and pull out the cushion from the lower part of the base in the direction of the arrow.

Omni-directional joystick

Styrofoam

Sponge Sponge Cushion

3 Wipe the sliding board with a cloth to remove any dirt. 4 Remove the styrofoam from the transportation bracket (A), on the left hand side of the patient, and unscrew the bracket (B).

transportation bracket (A)

transportation bracket (B)

5 Slide the base to the right and unscrew the transportation bracket (A).

30

6 Fasten the rail covers (3), using the small screws attached.

7 Set the external fixation target (7). Align the connector groove of the external fixation target with the connector of the instrument body and insert it. Secure it firmly and ensure that it cannot be moved.

8 Fix the chinrest tissue (15) with the chinrest tissue pin.Chinrest tissue pin

Chinrest tissue

INSTRUCTION MANUAL

31

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

2.2 Connecting the tv relay lens with the instrument bodyAlways ensure the TV relay lens and TV camera are securely mounted, to avoid potential injury during adjustment and operation.

CAUTION

This machine can be connected to an optional TV camera. If your TV camera is SONY DXC-970MD, the attached TV camera power cable (4) and TV camera signal cable (5) can be connected to the instrument body (1). 1 Turn the TV camera fixing ring in the direction of the arrow until it reaches the stopper, and remove the protective cap.

2 Mount the TV camera so the lens positioning hole is aligned to the pin of the TV relay lens (2).

32

Lens fixing ring

Protective cap

3 Turn the lens fixing ring in the direction of the arrow and fasten it.

2.3 Confirmation after assembly1 Slide the base to the left of the operator, and ensure that the primary voltage is within the correct range of voltage as set by the voltage selector.

2 Ensure that the input voltage is within the range 10% of the rated voltage. If the input voltage exceeds the range, use a constant-voltage power supply (marketed: 400VA up). 3 Loosen the base locking knob, and move the omni-directional joystick to confirm that it moves smoothly. - Right-left movement - Back-forth movement - Up-down movement

INSTRUCTION MANUAL

33

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

Note:

The right-left movement of the instrument may be uneven when it is first unpacked. If so, move the joystick to its limit in each direction.

2.4 Installing the instrumentEnsure the fixing knob at the bottom of the instrument is secure, before moving the instrument.

CAUTION

CAUTION

CAUTION

CAUTION

34

The instrument should always be carried by two people, both holding the bottom of the instrument. Attempting to carry the instrument alone could result in backache or other injury. Attempting to carry the instrument without holding it at the bottom could result in injury or damage.

Use an approved automatic instrument table to avoid potential injury while moving the instrument.

Always install the instrument on a fixed, stable surface, to prevent damage or injury.

1 Fasten the fixing knob. 2 Hold the instrument firmly at the specified positions and place it on the adjustable instrument table. For details about the adjustable instrument table, see Optional accessories on page 93.

Holding positions

Fixing knob

Holding the instrument body

3 After installing the instrument, loosen the fixing knob fully. The instrument body can then be moved. 4 If the instrument body is slightly off level, fine-adjust the height using the four adjusters. Do not extend the adjuster past 1 cm.

Adjuster

INSTRUCTION MANUAL

35

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

2.5 Connecting the power cableEnsure the power supply is equipped with a 3-plug AC receptacle to safeguard against power leakage. Power leakage could cause fire or electric shock.

WARNING

CAUTION

1 Make sure that the power switch (30) of the instrument body is OFF. 2 Connect the power cable to the instrument body.

3 Plug the power cable into an 3-p AC outlet that is properly grounded.

2.6 Connecting the external device2.6.1 Connecting the TV cameraThis machine can be connected and used with an optional TV camera. If your TV camera is SONY DXC-970MD, the attached TV camera power cable and TV camera signal cable can be connected to the instrument body.

36

To avoid the risk of electric shock, never handle a plug with wet hands.

1 Connect the TV camera connection terminal of the instrument body to the DC IN/REMOTE terminal of the TV camera with the TV camera power cable. 2 Connect the TV camera connection terminal of the instrument body to the RGB/SYNC terminal of the TV camera with the TV camera signal cable.DC IN/REMOTE terminal

RGB/SYNC terminal

TV camera (SONY DXC-970MD)

Note:

If your TV camera is not a SONY DXC-970MD, contact your dealer or TOPCON (see the back cover) for details about connection.

2.6.2

Connecting the external recording deviceThis machine can be connected to an external recording device via the external connection terminal. 1 Connect the external recording device to the image connection terminal of the instrument with an image cable (optional). 2 Connect the external recording device to the control terminal of the instrument with a control cable (optional).

INSTRUCTION MANUAL

37

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

3 When using a TOPCON IMAGEnet system, connect the photography data cable (optional) to the photography data output terminal of the instrument.

Image connection terminal

Control terminal

Photography data output terminal

Note:

Various types of connection cables are available for connectiing the instrument with external recording devices. For details about connecting external recording devices, contact your dealer or TOPCON (see the back cover).

2.6.3

Outputting images to an external monitorThe images displayed on the color video monitor may be outputted to an external monitor via the VIDEO OUT terminal of the external connection terminal. 1 Connect the BNC cable (optional) to the VIDEO OUT terminal of the instrument. 2 Connect the other end of the BNC cable to the input terminal of the external monitor.

Control terminal

38

2.6.4Note:

VIDEO OUT terminalFor details about video output to external recording devices, contact your dealer or TOPCON (see the back cover).

2.7 Menu settingThe following items can be set within Menu setting: record/playback, fixation target on/flicker, internal/ external fixation targets, fixation target patterns and flash level.

2.7.1

Preparation for menu setting1 Check the power cable connection. For details about the connection, see Connecting the power cable on page 36. 2 Turn the power switch (30) on.

2.7.2

Displaying the Menu screen1 Check the monitor screen. 2 Press the Menu switch (35) on the control panel. Check the MAIN MENU screen.

Cursor

3 Press the flash level switch (37) (+) or (-). The cursor then moves.

INSTRUCTION MANUAL

39

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

2.7.3

Returning to the Monitor screen1 Press the menu switch (35).

2.7.4

Setting the Record/Playback modeIt is possible to alternate between Record and Playback modes. In the Record mode, ordinary photography is available. Use the Playback mode to display the recorded images of the recording device to the color video monitor. The record mode RECORD is set automatically once the power supply is turned on. 1 On the MAIN MENU screen, make sure that the cursor is at RECORD/PLAYBACK, and press the flash level switch (37) (reset). The monitor screen changes to the RECORD/PLAYBACK screen.

2 Press the flash level switch (37) (+) or (-) and select RECORD (record mode) or PLAYBACK (playback mode). 3 Press the flash level switch (37) (reset); the setting is complete and the MAIN MENU screen returns.Note: When the playback mode is on, photography is prohibited.

2.7.5

Switching of fixation target on/flickerThe fixation target can be switched between on and flicker states. When the instrument is shipped, FLICKERING is set.

40

1 Move the cursor to FIXATION TYPE on the MAIN MENU screen, and press the flash level switch (37) (reset) to display the FIXATION TYPE screen.

2 Press the flash level switch (37) (+) or (-) and select CONSTANT (on) or FLICKERING (flicker). 3 Press the flash level switch (37) (reset); the setting is complete and the MAIN MENU screen returns.

2.7.6

Switching of internal/external fixation targetsIt is possible to alternate between Internal and External fixation targets. When the instrument is shipped, INTERNAL (internal fixation) is set.. 1 Move the cursor to INTERNAL/EXTERNAL FIX on the MAIN MENU screen, and press the flash level switch (37) (reset) to display the INTERNAL/EXTERNAL FIX screen.

2 Press the flash level switch (37) (+) or (-) and select INTERNAL (internal fixation) or EXTERNAL (external fixation). 3 Press the flash level switch (37) (reset); setting is complete and the MAIN MENU screen returns.

2.7.7

Fixation target patternIt is possible to select a fixation target pattern.

INSTRUCTION MANUAL

41

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

When the instrument is shipped, 8 is set. 1 Move the cursor to FIXATION PATTERNS on the MAIN MENU screen, and press the flash level switch (37) (reset) to display the FIXATION PATTERNS screen.

2 Press the flash level switch (37) (+) or (-) and select the desired pattern from the following:4A 4B 5A 5B 8 9 (diagonal 4 positions), (vertical-horizontal 4 positions), (diagonal 4 positions + center), (vertical-horizontal 4 positions + center), (8 positions), or (8 positions + center).

Note:

The center position is assigned to the optical axis center and is used for photographing around the macular segment.

3 Press the flash level switch (37) (reset); setting is complete and the MAIN MENU screen returns.

2.7.8

Flash levelIt is possible to change the reference value of the flash level. When the instrument is shipped, 0 (no change) is set.

42

1 Select FLASH LEVEL on the MAIN MENU screen, and press the flash level switch (37) (+) or (-) to display the FLASH LEVEL screen.

2 Press the flash level switch (37) (+) or (-) and select the desired level from the following:+2 +1 0 -1 -2 (2 steps up), (1 step up), (no change), (1 step down), or (2 steps down).

3 Press the flash level switch (37) (+) or (-); setting is complete and the MAIN MENU screen returns.Note: A step up/down changes the flash level reference value by about 20%.

2.8 Reset from power save stateThis machine features a power save function. When the instrument body has not been operated for a set time, the power save function stops power supply to the monitor, CCD camera, illumination light source and photography light source. During power saving, the power lamp of the control panel flickers and the monitor screen switches off. 1 Press the photography switch (16). In a few seconds, the video monitor displays and is ready for photography.Note: When the instrument is shipped, the power save set time is 10 minutes. To change the set time, contact your dealer or TOPCON (see the back cover).

INSTRUCTION MANUAL

43

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

44

3

INITIAL SETTING

On the INITIAL MENU screen, the following items can be programmed; record/playback, internal/external fixation targets, peripheral photography, system setting, initial setting and language.

3.1 Preparation for initial set-up1 Ensure that the power cable is connected. 2 Press the menu switch (35) and photography switch (16) of the control panel and turn the power switch (30) on. Press the menu switch (35) and photography switch (16) until the buzzer sounds. The Title screen is displayed, and a few seconds later, the Monitor screen is displayed.

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

3 Press the menu switch (35) of the control panel to display the SETTING MENU screen.

4 To finish with the INITIAL MENU screen, press the menu switch (35).Note: Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

3.2 Setting the record/playback modeIt is possible to alternate between Record and Playback modes. Use the Playback mode to display the recorded images of the recording device to the color video monitor. The record mode RECORD is set automatically once the power supply is turned on. 1 Move the cursor to the RECORD/PLAYBACK of the INITIAL MENU screen, and press the flash level switch (37) (reset). The monitor screen changes to the CSTM SCREEN.

2 Press the flash level switch (37) (+) or (-) and select the RECORD or PLAYBACK mode. 3 Press the flash level switch (37) (reset) and the INITIAL MENU screen displays.

46

Note:

Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

3.3 Setting the fixation targetIt is possible to set the fixation target on/flicker and internal/external fixation target switching on the FIXATION screen. 1 On the INITIAL MENU screen, move the cursor to FIXATION and press the flash level switch (37) (reset). The monitor screen changes to the FIXATION screen.

2 To finish with the FIXATION screen, press the menu switch (35).Note: Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

3.3.1

Switching the fixation target on/flickerThe fixation target can be switched between on and flicker states. When the instrument is shipped, FLICKERING (flicker) is set. 1 Select FIXATION TYPE on the FIXATION screen to display the FIXATION TYPE screen.

INSTRUCTION MANUAL

47

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

2 Press the flash level switch (37) (+) or (-) and select CONSTANT (on) or FLICKERING (flicker). 3 Press the flash level switch (37) (reset); setting is complete and the FIXATION screen returns.

3.3.2

Switching of internal/external fixation targetsInternal/external fixation targets can be switched. When the instrument is shipped, INTERNAL (internal fixation) is set. 1 Select INTERNAL/EXTERNAL FIX on the FIXATION screen, to display the INTERNAL/EXTERNAL FIX screen.

2 Press the flash level switch (37) (+) or (-) and select INTERNAL (internal fixation) or EXTERNAL (external fixation). 3 Press the flash level switch (37) (reset); setting is complete and the FIXATION screen returns.

3.4 Setting peripheral photographyThe following items can be programmed on the PERIPHERAL FIXATION screen: fixation target pattern, R/L reset and display mode.

48

1 Move the cursor to the PERIPHERAL FIXATION of the INITIAL MENU screen, and press the flash level switch (37) (reset). The monitor screen changes to the PERIPHERAL FIXATION screen.

2 To finish with the PERIPHERAL FIXATION screen, press the menu switch (35).Note: Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

3.4.1

Fixation target patternIt is possible to select a fixation target pattern. When the instrument is shipped, 8 is set. 1 Move the cursor to FIXATION PATTERNS on the PERIPHERAL FIXATION screen, to display the FIXATION PATTERNS screen.

2 Press the flash level switch (37) (+) or (-) and select the desired pattern from the following:4A 4B 5A 5B 8 9 (diagonal 4 positions), (vertical-horizontal 4 positions), (diagonal 4 positions + center), (vertical-horizontal 4 positions + center), (8 positions), or (8 positions + center).

3 Press the flash level switch (37) (reset); the setting is complete and the PERIPHERAL FIXATION screen returns.

INSTRUCTION MANUAL

49

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

Note:

The center position is assigned to the optical axis center and is used for photographing around the macular segment.

3.4.2

R/L resetIf the base is moved during peripheral photography, right/left eye reset can be set. (When OFF is set, peripheral photography cannot be reset by right/left eye switching.) When the instrument is shipped, OFF (no reset) is set. 1 Select R/L RESET on the PERIPHERAL FIXATION screen, to display the R/L RESET screen.

2 Press the flash level switch (37) (+) or (-) and select ON (reset) or OFF (no reset). 3 Press the flash level switch (37) (reset); the setting is complete and the PERIPHERAL FIXATION screen returns.

3.4.3

Display modeIt is possible to set the display mode of the peripheral fixation target. When FIXATION POSITION (fixation position display) is set, the monitor screen display responds to the fixation target, and when PHOTOGRAPHIC POSITION (picture position display) is set the monitor screen display responds to the photography result. When the instrument is shipped, FIXATION POSITION (fixation position display) is set. 1 Select FIXATION INDICATOR on the PERIPHERAL FIXATION screen, to display the FIXATION INDICATOR screen.

50

2 Press the flash level switch (37) (+) or (-) and select FIXATION POSITION (fixation position display) or PHOTOGRAPHIC POSITION (picture position display). 3 Press the flash level switch (37) (reset); the setting is complete and the PERIPHERAL FIXATION screen returns.

3.5 Setting the screen displayOn the MONITOR DISPLAY screen, the following display settings can be set: flash level compensation, flash level, illumination level, picture angle, illumination aperture* and fixation position. 1 Ensure that the cursor is on the MONITOR DISPLAY of the INITIAL MENU screen, and press the flash level switch (37) (reset). The monitor screen changes to the MONITOR DISPLAY screen.

TRC-NW6SF

TRC-NW6S

2 To finish with the MONITOR DISPLAY screen, press the menu switch (35).Note: Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

3.5.1

Flash level compensation displayThe flash level compensation display can be set. When the instrument is shipped, ON (display) is set.

* TRC-NW6SF only

INSTRUCTION MANUAL

51

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

1 Select FLASH LEVEL IN 5STEPS on the MONITOR DISPLAY screen, to display the FLASH LEVEL IN 5STEPS screen.

2 Press the flash level switch (37) (+) or (-) and select ON (display) or OFF (no display). 3 Press the flash level switch (37) (reset); the setting is complete and the MONITOR DISPLAY screen returns.

3.5.2

Flash level displayThe flash level display can be set. When the instrument is shipped, OFF (no display) is set. 1 Select FLASH LEVELS IN WS on the MONITOR DISPLAY screen, to display the FLASH LEVELS IN WS screen.

2 Press the flash level switch (37) (+) or (-) and select ON (display) or OFF (no display). 3 Press the flash level switch (37) (reset); the setting is complete and the MONITOR DISPLAY screen returns.

3.5.3

Illumination level displayThe illumination level display can be set. shipped, ON (display) is set. When the instrument is

52

1 Select ILLUMINATION LEVELS on the MONITOR DISPLAY screen, to display the ILLUMINATION LEVELS screen.

2 Press the flash level switch (37) (+) or (-) and select ON (display) or OFF (no display). 3 Press the flash level switch (37) (reset); the setting is complete and the MONITOR DISPLAY screen returns.

3.5.4

Picture angle displayThe picture angle display can be set. When the instrument is shipped, ON (display) is set. 1 Select ANGLE INDICATION on the MONITOR DISPLAY screen, to display the ANGLE INDICATION screen.

2 Press the flash level switch (37) (+) or (-) and select ON (display) or OFF (no display). 3 Press the flash level switch (37) (reset); the setting is complete and the MONITOR DISPLAY screen returns.

3.5.5

Illumination aperture display *The illumination aperture display can be set. When the instrument is shipped, ON (display) is set.

* TRC-NW6SF only

INSTRUCTION MANUAL

53

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

1 Select ILLUMINATION APERTURE on the MONITOR DISPLAY screen, to display the ILLUMINATION APERTURE screen.

2 Press the flash level switch (37) (+) or (-) and select ON (display) or OFF (no display). 3 Press the flash level switch (37) (reset); the setting is complete and the MONITOR DISPLAY screen returns.

3.5.6

Peripheral fixation position displayThe peripheral fixation position display can be set. When the instrument is shipped, ON (display) is set. 1 Select PERIPHERAL PATTERN on the MONITOR DISPLAY screen, to display the PERIPHERAL PATTERN screen.

2 Press the flash level switch (37) (+) or (-) and select ON (display) or OFF (no display). 3 Press the flash level switch (37) (reset); the setting is complete and the MONITOR DISPLAY screen returns.

54

3.6 System settingSettings of the connected external recording device can be programmed on the SYSTEM SETTINGS screen. When the instrument is shipped, IMAGENET MODE (image mode connection) is set. 1 Ensure that the cursor is on the SYSTEM SETTINGS of the INITIAL MENU screen, and press the flash level switch (37) (reset). The monitor screen changes to the SYSTEM SETTINGS screen.

2 Press the flash level switch (37) (+) or (-) and select:IMAGE MODE TYPE1 MODE TYPE2 MODE (IMAGEnet connection); (single-action external device connection); or (double-action external device connection).

3 Press the flash level switch (37) (reset) and the INITIAL MENU screen displays.Note: Note: For details about single/double-action, refer to the instruction manual of the external recording device. Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

3.7 Setting the initial stateOn the INITIAL SETTINGS screen, the following items can be set: flash level, commercial power supply frequency, LED display, timer sound*, operation sound and quick mirror-up of power save time.

* TRC-NW6SF only

INSTRUCTION MANUAL

55

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

1 Ensure that the cursor is on the INITIAL SETTINGS of the INITIAL MENU screen, and press the flash level switch (37) (reset).

TRC-NW6SF

TRC-NW6S

2 To finish with the INITIAL SETTINGS screen, press the menu switch (35).Note: Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

3.7.1

Flash levelYou can change the flash level of the instrument. When the instrument is shipped, 0 (no change) is set. 1 On the INITIAL MENU screen, choose INITIAL SET to display the INITIAL SET screen.

2 Press the flash level switch (37) (+) or (-) and select the desired level from the following:+2 +1 0 -1 -2 (2 steps up), (1 step up), (no change), (1 step down), or (2 steps down).

3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL SETTINGS screen returns.Note: A step up/down changes the flash level reference value by about 20%.

56

3.7.2

Commercial power supply frequencyThe frequency of commercial power supply can be set. When the instrument is shipped, 50HZ (50Hz) is set. 1 Select FREQUENCY on the INITIAL SETTINGS screen, to display the FREQUENCY screen.

2 Press the flash level switch (37) (+) or (-) and select 50Hz or 60Hz. 3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL SETTINGS screen returns.

3.7.3

LED displayThe LED display can be set. For TRC-NW6SF, when shipped, TIMER (timer) is set. For TRC-NW6S, when shipped, PERIPHERAL POINT (fixation target position display) is set. 1 Select SEGMENT LED on the INITIAL SETTINGS screen, to display the SEGMENT LED screen.

TRC-NW6SF

TRC-NW6S

2 Press the flash level switch (37) (+) or (-) and select:TIMER PERIPHERAL POINT OFF (timer display); (fixation target position display); or (no display).

INSTRUCTION MANUAL

57

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL SETTINGS screen returns.

3.7.4

Timer sound *The timer sound can be set. When the instrument is shipped, ON (buzzer sound) is set. 1 Select TIMER SOUND on the INITIAL SETTINGS screen, to display the TIMER SOUND screen.

2 Press the flash level switch (37) (+) or (-) and select ON (buzzer sound) or OFF (no buzzer sound). 3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL SETTINGS screen returns.

3.7.5

Operation soundThe operation sound can be set. When the instrument is shipped, ON (operation sound) is set. 1 Select OPERATION SOUND on the INITIAL SETTINGS screen, to display the OPERATION SOUND screen.

2 Press the flash level switch (37) (+) or (-) and select ON (operation sound) or OFF (no operation sound).

* TRC-NW6SF only

58

3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL SETTINGS screen returns.

3.7.6

Power save timeThe power save time can be set. When the instrument is shipped, 10 MINUTES (10 minutes) is set. 1 Select POWER SAVE TIME on the INITIAL SETTINGS screen, to display the POWER SAVE TIME screen.

2 Press the flash level switch (37) (+) or (-) and select between 5 MINUTES up to 60 MINUTES in 5 minutes steps. 3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL SETTINGS screen returns.

3.7.7

Quick mirror upThis function is used to fix the quick mirror in the up position during assembly adjustment of the TV relay lens. Unfix OFF is set automatically once the power supply is turned on. 1 Select QUICK MIRROR UP on the INITIAL SETTINGS screen, to display the QUICK MIRROR UP screen.

2 Press the flash level switch (37) (+) or (-) and select ON (fix) or OFF (unfix).

INSTRUCTION MANUAL

59

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL SETTINGS screen returns.

3.8 Setting the languageThe language to be used on the menu screen can be set on the LANGUAGE screen. When the instrument is shipped, ENGLISH is set. 1 Ensure that the cursor is on the LANGUAGE of the INITIAL MENU screen, and press the flash level switch (37) (reset). The monitor screen changes to the LANGUAGE screen.

2 Press the flash level switch (37) (+) or (-) and select:ENGLISH; GERMAN; FRENCH; SPANISH or ITALIAN.

3 Press the flash level switch (37) (reset); the setting is complete and the INITIAL MENU screen returns.Note: Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.

60

4

BASIC OPERATION

4.1 Preparation for photography4.1.1 Connecting to the power supply1 Check the power cable connection. For details about connection, see Connecting the power cable on page 36. 2 Turn on the power switch (30) of both the instrument and the external recording device. 3 Confirm that the title screen is displayed first, followed by the Monitor screen.

4.1.2

Positioning the patientTo avoid potential injury to the patient, always watch the patient carefully while adjusting the chinrest.

CAUTION

Note:

If the patient wears glasses or contact lens, remove them beforehand.

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

1 Ensure the main monitor screen is displayed. 2 When imaging the patients data, write the date on the nameplate and insert it into the nameplate slot.

Nameplate insert

Monitor screen

Note:

Use a felt-tip marker to write the patient's name on the nameplate. Do not use an oil felt-tip as this can appear unclear on the monitor screen.

3 Press in the IR filter selector.

Pressing the IR filter selector

62

Note:

If the IR filter selector has not been pressed, the IR warning display (IR) flickers on the color video monitor, to indicate that the filter is inactive.

4 Seat the patient comfortably in the examination chair. 5 Adjust the table or chair height, so that the patient is comfortable when placing his/her chin on the chin rest (14). 6 Let the patient rest his/her chin on the chinrest, with the forehead against the forehead rest.

7 Use the chinrest switch to adjust the height of the chinrest (39) so the tail of the eye comes level with the Canthus marker of the chinrest post.

Canthus marker

Note:

The chinrest can be moved up or down as required when the chinrest switch (39) is pressed. If you are unable to move the chinrest even when the chinrest switch (39) is pressed, discontinue the examination. Turn the power switch (30) off and remove the power cable, and contact your dealer or TOPCON (see the back cover).

INSTRUCTION MANUAL

63

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

In the case of an emergency, you can manually operate the chinrest. For details about the manual operation, see Troubleshooting table on page 96.

4.2 Color photography (center)Note: To ensure correct imaging, adjust the table height so the patient can rest his/her chin comfortably on the chinrest.

4.2.1

Setting the photography mode*Turn the mode selector counterclockwise and set it for color photography.

4.2.2

Setting the picture positionSet the picture position using the periphery fixation switch (42). Set the fixation target to the centre by pressing the periphery fixation switch (42) (reset).

* TRC-NW6SF only

64

The center part of the fixation position display flickers on the color video monitor, indicating that the fixation target is set in the center.

Fixation position display (with fixation target pattern "8")

Note:

The picture position is set automatically in the center when the power supply is first turned on.

4.2.3

Setting the picture angleSet the picture angle to 45 or 30 by operating the angle selector.

Note:

The picture angle of the image on the color video monitor does not change, even when the picture angle has been changed. To confirm the current picture angle, check the picture angle display on the color video monitor (displayed only when 30 is set).

4.2.4

Setting illumination levelEnsure that the photography light is no brighter than necessary, to avoid any discomfort or injury to the patient.

CAUTION

INSTRUCTION MANUAL

65

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

Set the illumination level using the illumination level switch (38). You can confirm the current level by refering to the illumination level display on the color video monitor.

Illumination level

Note:

Normally the illumination level has five levels - during a visible fluoro-observation it has seven illumination levels to choose from. The illumination level is automatically set to level 3, when the instrument is first turned on.

4.2.5

Setting the flash levelEnsure that the photography light is no brighter than necessary, to avoid any discomfort or injury to the patient.

CAUTION

Set the flash level using the flash level switch (37). Check the compensation value by refering to the flash level compensation display on the color video monitor.

66

Flash level compensation display

Note:

You can compensate for the flash level in either direction, (+) and (-), to a maximum of two steps from the currently set position. When the flash level is identical to the reference value, no compensation value is displayed. The flash level is automatically set to the reference value, when the instrument is first turned on. The illumination level is automatically set to the reference value, when the power supply is turned on. You can adjust the reference value of the flash level in either direction, (+) or (-), to a maximum of two steps. See the Flash level on page 56. The flash level display can also display the light intensity level (unit: WS), as well as the the compensation value. For details about the flash level setting, contact your dealer or TOPCON (see the back cover).

Note:

Note:

4.2.6

Changing the diopter compensation lensPull out the diopter compensation lens selector and change the diopter compensation lens for the patients eye.

Note:

If the patient has strong myopia, pull out the diopter compensation lens selector one step and set it to (-) myopia. If the patient has strong hyperopia, pull out the diopter compensation lens selector two steps and set it to (+) hyperopia.

Compensation range:

0 +

:-13~+12D :-12~-33D :+9~+40D

INSTRUCTION MANUAL

67

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

4.2.7

Collimation and photographyTo avoid potential injury, do not place your hand or fingers into the area between the instrument body and the power supply unit.

CAUTION

CAUTION

Use the omni-directional joystick to perform a collimating operation. Using the omni-directional joystick to move the instrument body. Fine movements of the base, back and forth and right and left, are achieved by tilting the joystick. First free the base by turning the base brake knob left. To fix the base, turn the base brake knob right. Operating the omni-directional joystick (in all directions)Base brake knob

68

To avoid potential injury to the patient while moving the instrument, always keep the objective lens at a safe distance from the patient..

To move the instrument body up/down, turn the omni-directional joystick right for the upward movement, and left for the downward movement. The vertical position of the instrument body can be checked by the vertical position mark.

Operating the omni-directional joystick (in vertical directions)

Vertical position mark

1 Hold the omni-directional joystick and pull the instrument back towards the operator. When the internal fixation target has turned on, instruct the patient to look at the fixation target in the center. Observe the anterior segment image on the color video monitor. 2 Move the instrument body as required until the patients eye can be seen in the center of the color video monitor.

INSTRUCTION MANUAL

69

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

3 Now hold the joystick upright, in order to facilitate the eyeground collimation.

4 On the color video monitor, bring the ( ) scale to the patients pupil, ensuring that the patient's eye is larger than the ( ) scale.Note: Comparison of the ( ) scale and the patient's eye will indicate whether the pupil is large enough for retinal photography.

Well dilated.

Narrowly dilated for photography.

Pupil diameter is too small: darken the room in order to dilate the pupil further.

Note:

If dilation is insufficient, you can change the flash level to twice the set value with the illumination aperture switch (40)*. When the illumination aperture switch (40) is pressed, the flash level compensation display appears on the color video monitor, indicating that the flash level is now twice the set value. Pressing the illumination aperture switch (40) again, returns the flash level to the set value.

* TRC-NW6SF only

70

Flash level compensation display

5 Bring the instrument slowly closer to the patient; the retina image appears on the video monitor.

6 Instruct the patient to look at the green light (internal fixation target). 7 While watching the image on the video monitor, adjust the brightness of the image using the illumination level switch (38).Note: For details about mapping data printing, see Setting illumination level on page 65. If the illumination level is too bright, the split lines will be difficult to see.

8 Bring the instrument still closer to the patient; and two bright spots for the working distance alignment will become visible.

Bright spots

INSTRUCTION MANUAL

71

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

9 Using the joystick, bring the two bright spots together on the video monitor.

10 Using the joystick, move the instrument body until the bright spot of the video monitor moves into the ( ) scale.

72

11 Operating the focusing knob, align the focusing split lines into one line on the video monitor.

Split lines

Note:

If the split lines cannot be aligned using the focusing knob, change the diopter compensation lens. For details, see Changing the diopter compensation lens on page 67. Because split lines are inactive when the diopter compensation lens is other than (0), turn the focusing knob so the eyeground image is clearly visible on the color video monitor. If the split lines are not clearly visible, lower the illumination level, or lower the brightness of the color video monitor. For details about adjusting the brightness of the video monitor, see Adjusting the color video monitor on page 90. If one of the split lines cannot be seen, check if dilation is insufficient, or if the eyelashes or eyelid are shading the eye. You can remove split lines from the Monitor screen. Pressing the split switch (36) removes the split lines from the monitor screen. Pressing the split switch (36) again, displays the split lines on the monitor screen.

Note:

Note:

INSTRUCTION MANUAL

73

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

12 Ensure that the split line is aligned with the bright spot on the color video monitor. Press the photography switch (16).

Photography switch

Xenon charging display

Note:

It is not possible to take an image while the xenon charging display is flickering on the color video monitor. Charging takes about 2 seconds. Wait until the xenon charging lamp turns on, for the next image.

13 The captured image is sent to the external recording device and displayed on the color video monitor.

Note:

If the image is insufficiently bright, you can compensate by pressing the flash level switch (37), and repeating the collimation and photography procedure.

74

Note:

The display of images on the color video monitor depends on the settings of the external recording device as well as those of the instrument.

14 To return to the capture screen on the color video monitor, press the photography switch (16).

4.3 Fluoro-photography (center)*Note: To ensure correct photography, adjust the table height so the patient can rest his/her chin comfortably on the chinrest.

4.3.1

Setting the photography modeSet the fluoro-photography by switching the mode selector.

4.3.2

Setting the picture positionSet the picture position using the periphery fixation switch (42). See Setting the picture position on page 64.

4.3.3

Setting the picture angleSet the picture angle to 45 or 30 by operating the angle selector. See Setting the picture angle on page 65.

* TRC-NW6SF only

INSTRUCTION MANUAL

75

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

4.3.4

Setting illumination levelSet the illumination level using the illumination level switch (38). See Setting illumination level on page 65.

4.3.5

Setting the flash levelSet the flash level using the flash level switch (37). See Setting the flash level on page 66.

4.3.6

Changing the diopter compensation lensPull out the diopter compensation lens selector and change the diopter compensation lens for the patients eye. See Changing the diopter compensation lens on page 67.

4.3.7

Preparing the patient1 Apply mydriatics to the patients eyes, wait for thorough dilation, and then guide the patient to the photography room.

Note:

If picturing at a 45 picture angle or if there is insufficient dilation, photography is possible up to 3.7 mm pupil diameter by changing the stop using the illumination aperture switch (40).

The illumination aperture display appears on the color video monitor. 2 Seat the patient comfortably in the examination chair. See Positioning the patient on page 61. 3 Prepare for an intravenous injection of fluorescent contrast medium.

4.3.8

Collimation and photographyUse the omni-directional joystick to perform a collimating operation. For details about movement or adjustment of the instrument body using the omni-directional joystick, see Collimation and photography on page 68. 1 As all operations are identical to color photography up to the point of focusing, please refer to collimation of color photography. 2 Upon injecting fluorescent contrast medium, press the timer switch (41).

Note:

When the timer switch (41) is pressed, the timer begins to operate, and the buzzer sounds every 1 second for the first 20 seconds. Pressing the timer switch (41) again, will turn off the timer. The result of fluoro-photography depends on the injection time and the quantity of the fluorescent contrast medium used.

Note:

76

3 For details about visible fluoro-observation, see Fluoro-observation on page 77. 4 Ensure that the aligned bright spot is in the ( ) scale and the split lines are aligned to a straight line on the color video monitor, and then press the photography switch (16).

5 The photography image is sent to the external recording device and displayed on the color video monitor.

6 To return to the capture screen on the color video monitor, press the photography switch (16).

4.3.9

Fluoro-observationDo not continue visible fluoro-observation any longer than necessary to avoid any discomfort or injury to the patient.

CAUTION

Normally, observation is performed using infrared rays, but fluoro-observation of dynamic images can be performed using a color video monitor at the instant the fluorescent contrast medium begins to flow through the fundus artery.

INSTRUCTION MANUAL

77

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

1 Turn the fluoro-observation knob to the visible side. The IR warning display flickers on the color video monitor.

IR warning display

Note:

The illumination level is automatically set to the maximum light intensity and the 6th and 7th steps of the illumination level display flicker on the color video monitor, indicating that the maximum light intensity is set.

2 Pull out the IR filter selector. The IR warning display disappears from the color video monitor.

3 It is possible to perform visible fluoro-observation on the video monitor.

78

Note:

Because the internal fixation target cannot be used when the fluoro-observation knob is on the visible side, instead use the external fixation target to guide the patient's eye. You can remove split lines from the Monitor screen. Pressing the split switch (36) removes the split lines from the Monitor screen. Pressing the split switch (36) again displays the split lines on the Monitor screen. During the visible fluoro-observation, observation TV camera settings are changed for visible fluoro-observation. In this case, as the TV camera gain is increased, images may become grainy.

Note:

Note:

4.4 Peripheral photography4.4.1 Setting the picture positionEach time the periphery fixation switch (42) (right) is pressed the fixation target moves clockwise, from the patient's point of view, and the set point display flickers on the color video monitor.

Fixation position display (with fixation target pattern "8")

INSTRUCTION MANUAL

79

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

Each time the periphery fixation switch (42) (left) is pressed the fixation target moves counterclockwise, from the patient's point of view, and the set point display flickers on the color video monitor.

Fixation position display (with fixation target pattern "8")

4.4.2

Other settingsFor other settings, see Color photography (center) on page 64 and Fluoro-photography (center) on page 75.

4.4.3

Collimation and photographyUse the omni-directional joystick to perform a collimating operation. For details about movement or adjustment of the instrument body using the omni-directional joystick, see Collimation and photography on page 68.

80

1 Hold the omni-directional joystick and pull the instrument back towards the operator. When the anterior segment fixation target has turned on, instruct the patient to look at the fixation target. Monitor the anterior segment image on the color video monitor.

Anterior segment fixation target (flickering points corresponding to periphery setting)

2 Using the omni-directional joystick, move the instrument body as required until the patient's eye can be seen in the center of the color video monitor.

3 On the color video monitor, bring the ( ) scale to the patients pupil, and ensuring that the patient's eye is larger than the ( ) scale.Note: For details about dilation, see the note on page 70.

4 Bring the instrument slowly closer to the patient; the eyeground image appears on the color video monitor.

INSTRUCTION MANUAL

81

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

5 In this case, the ( ) scale on the color video monitor moves into alignment with the picture position. At the same time the anterior segment fixation targets changes to the internal fixation target.

6 Using the illumination level switch (38), adjust the brightness of the image on the color video monitor.Note: For details about illumination level setting, see the Setting illumination level on page 65.

7 For the steps to follow, see Collimation and photography on page 68 for color photography and Collimation and photography on page 76 for fluoro-photography (TRC-NW6SF only).

4.5 Anterior segment photography4.5.1 Setting the photography mode*Set the color photography mode by switching the mode selector.

* TRC-NW6SF only

82

4.5.2

Setting the picture positionSet the fixation target to the center using the periphery fixation switch (42). See Setting the picture position on page 64.

4.5.3

Setting the picture angleSet the picture angle by operating the angle selector. See Setting the picture angle on page 65.

4.5.4

Setting the illumination levelSet the flash level using the illumination level switch (38). See Setting the flash level on page 66.

4.5.5

Changing the diopter compensation lensPress in the diopter compensation lens selector and change the diopter compensation lens (0). See Changing the diopter compensation lens on page 67.

4.5.6

Collimation and photographyUse the omni-directional joystick to perform a collimating operation. For details about movement/adjustment of the instrument body using the omni-directional joystick, see Collimation and photography on page 68. 1 Hold the omni-directional joystick and pull the instrument back towards the operator. Observe the anterior segment image on the color video monitor.

INSTRUCTION MANUAL

83

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

2 Move the instrument body as required until the patient's eye can be seen in the center of the color video monitor.

3 Turn the focusing knob so the anterior segment image is clearly visible on the color video monitor, and press the photography switch (16).Note: In the anterior segment photography, the flash level is changed automatically to an optimal set value.

4.6 Finishing1 Turn off the power switch (30) of the instrument body and of the external recording device. 2 Use the omni-directional joystick to move the instrument body so that it comes to just above the base. 3 Turn the base brake knob to the right to brake the base and prevent it from moving suddenly.Note: To prepare the instrument body for the next imaging session, turn the omni-directional joystick and move the body into the center position. The vertical center position of the instrument body is indicated by the vertical posi-tion mark. If the instrument will not be in use for a long period, remove the power cables of the instrument body and external recording device from the power outlets, and remove all cables connecting the instrument to the external capture device to avoid accidental cable damage.

Note:

84

5

MAINTENANCE AND CHECKS

5.1 Daily careAlways install the instrument on a fixed, stable surface, to prevent damage or injury.

CAUTION

CAUTION

CAUTION

Ensure that the camera mount lever is firmly tightened to prevent the TV camera from falling and causing injury or damage.

Ensure that the lens mount lever is firmly tightened to prevent the TV relay lens from falling and causing injury or damage.

Keep the instrument protected from dust at all times. Keep the objective lens protected from fingerprints and dirt to ensure the best possible images.

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

When not in use, be sure to cap the objective lens and cover the instrument with the dust cover. If the objective lens becomes stained, follow the instructions in Cleaning the objective lens on page 91. When not in use, always turn the power switch (30) off.

5.2 Changing the lamp5.2.1 Replacing the illumination lampTo avoid the risk of electric shock, always disconnect the power supply and unplug the power cable before replacing the lamp.

CAUTION

CAUTION

Note: Note:

Do not touch the lamp with bare fingers: fingerprints can damage (whiten) the lamp. The lamp can be easily damaged: handle it with particular care.

1 Turn the power switch (30) off and unplug the power cable. 2 Turn the omni-directional joystick to raise the instrument body completely. 3 Unscrew the lamp house cover.

86

To avoid potential burns, do not touch the lamp during, or immediately after use.

The service life of illumination lamp is approx. 2,000 hours. Replacing the illumination lamp if it is burned or becomes whitened.

4 Loosen the two set screws and remove the lamp terminal.

Lamp holder

Lamp terminal Lamp terminal set screw

5 Hold the lamp at the root and pull it out of the lamp holder. 6 Hold the new lamp so the convex part faces the operator, and slide it into the lamp holder until it comes to a stop. Ensure that the lamp is firmly fixed in the lamp holder.

Lamp holder

7 Fasten the lamp terminal securely using the two set screws. 8 Align the projection at the bottom part of the lamp house cover with the groove of the body cover. Screw in the lamp house cover.Note: To avoid the risk of electric shock, an error is displayed on the monitor if the lamp house cover is left unfixed, and all operations are prohibited.

5.2.2

Replacing the xenon lampTo avoid the risk of electric shock, always disconnect the power supply and unplug the power cable before replacing the lamp.

CAUTION

INSTRUCTION MANUAL

87

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

To avoid the risk of electric shock, never touch the xenon lamp immediately after a flash or burn out.

CAUTION

1 Turn the power switch (30) off and unplug the power cable, then wait for at least 5 minutes to allow natural electrical discharge. 2 Unscrew the lamp house cover. 3 Loosen the three xenon set screws.Xenon set screw

4 Hold the xenon board at the top and bottom, slightly slide it to the right and pull it out toward the operator side. 5 Insert the new xenon board, so the xenon lamp does not touch surrounding metal components, and upon reaching the stopper, slightly move it to the left and slide it into the xenon set screws. 6 Fasten the three xenon set screws securely. 7 Align the projection at the bottom part of the lamp house cover with the groove of the body cover. Screw in the lamp house cover.Note: To avoid the risk of electric shock, an error is displayed on the monitor if the lamp house cover is left unfixed, and all operations are prohibited.

88

The service life of xenon lamp is approx. 10,000 cycles. Replacing the xenon lamp if it is burned or becomes whitened.

Xenon board

5.3 Changing the fusesTo avoid the risk of electric shock, always disconnect the power cable from the instrument before removing the fuse cover. Always replace the fuse cover before reconnecting the instrument to the power supply.

WARNING

WARNING

1 Turn the power switch (30) off and unplug the power cable. 2 Using a screwdriver, press and turn the fuse holder counterclockwise and remove.

3 Replace the old fuse with a new one of the same capacity.

4 Using a screwdriver, lightly press the fuse holder and turn it clockwise to fix it.

INSTRUCTION MANUAL

To avoid the risk of fire, always use the correctly rated fuse as specified on the fuse holder.

Changing the fuse

89

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

The usage of each fuse and the result after it has burned are shown below: Fuse No.F1, F2 (5A) F3 (6A) F4 (0.5A) F5 (5A)

UsePower input Illumination light Xenon lamp Control circuit

When the fuse blowsInstrument will not work. Illumination lamp will not work. ERR3: Xenon lamp has not been charged. Instrument will not work.

5.4 Supplying the chinrest tissueTo change the chinrest tissue, remove the chinrest tissue pin and replace the tissue.

5.5 Adjusting the color video monitorThis machine is adjusted for the best image quality before shipment, however, readjustment may be required after transportation. The image quality adjustment knob is located on the left hand side of the color video monitor.

90

Chinrest tissue pin

Please note that the ideal monitor settings for fundus viewing may be different than those for anterior segment viewing. Adjust the brightness, contrast and hue adjustment knobs to attain the desired image quality.

5.6 Cleaning the instrument5.6.1Note:

Cleaning the dust cover, control panel and monitor screenTo protect the instrument body from discoloration and deterioration, do not use volatile solvents to clean the instrument, including benzine, thinner, ether, gasoline, etc.

5.6.2

Cleaning the objective lensTo check the objective lens, turn the illumination lamp on. Darken the room, pull out the IR filter selector and set it to visible fluoro-observation. Press the illumination level switch (38) (+) four times and maximise the level of light. Examine the objective lens from the front and at a angle. The lens condition can be seen clearly. Dust and dirt adhered to the surface

Brightness adjustment knob

Contrast adjustment knob

Hue adjustment knob

1 If the dust cover, control panel and monitor screen become stained, clean them with dry cloth. 2 If the dust cover is badly stained, prepare a tepid solution of neutral detergent, suitable for kitchenware. Moisten the cloth with the solution and rinse it thoroughly.

INSTRUCTION MANUAL

91

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

Blow off dust with a blower. Fingerprints and oil spots on the surface:

Note:

If the stains are persistent and difficult to remove, contact your dealer or TOPCON (see the back cover). Use the ethanol-ether mix at temperatures (10~40C) and moisture (30~85%) under normal conditions.

5.7 List of expendables and spare partsWhen ordering expendables and spare parts, please specify the code number and the quantity required, in addition to the description.

92

1 Blow off any large particles. 2 Prepare a mix of ethanol (2) and ether (8). 3 Moisten a clean, dried gauze with the mix, and lightly wipe the lens spirally from the center outward. Continue until the stains have been completely removed. To prevent the objective lens from being scratched, always use clean gauze and wipe the surface of the lens gently.

DescriptionXenon lamp Illumination lamp Expendables Chin rest paper Data plate Dust cover Fuses (F1)(F2) 5A/125V Spare parts Fuse (F3) 6A/125V Fuse (F4) 0.5A/250V Fuse (F5) 5A/125V

Code Number40531 1810 40531 1880 40310 4082 40547 1132 40488 1007 44691 5006 T2400 0001A 40547 5302 44691 5006

5.8 Optional accessories5.8.1 Adjustable instrument table AIT-11This allows the height of the instrument to be adjusted as required, ensuring greater ease of measurement.

Size: Table height: Table top size: Weight: Power consumption

554 (W) 428 (D) mm 566 to 766 mm 450 (W) 500 (D) mm approx. 21 kg 270VA (100V)

5.8.2

Adjustable instrument table AIT-20

INSTRUCTION MANUAL

93

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

5.9 Contacting TOPCONWhen you contact us, please mention the following: Model name: Serial number: Period of use: Type of problem: TRC-NW6SF/TRC-NW6S This is located on the rating nameplate on the right side of the base. Please inform us of the purchase date of this instrument. Please give us a brief description of the problem.

94

6

TROUBLESHOOTING

6.1 Before requesting serviceIf a problem should occur, first consult the following troubleshooting table, and follow the instructions suggested. If you cannot solve the problem, or the instrument is defective, contact your authorized dealer or TOPCON. Be sure to disconnect the power input plug from the outlet before replacing a lamp or fuse.

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

To avoid the risk of electric shock, do not attempt disassembling, rebuilding or repairs. Consult your dealer for repairs.

WARNING

WARNING

6.2 Troubleshooting tableTrouble Check points

96

Periphery of photography image is dark.

Color video monitor does not display clearly.

Color video monitor does not work.

To avoid the risk of electric shock, do not remove any body components, covers of the TV relay lens, the chinrest or power supply. - other than the lamp house cover.

Power cable is not connected Connect power cable. Fuse has burned. Change fuse. Power save function is on (flickering power lamp). Cancel power save function (Photography switch). Image contrast is needs adjustment. Adjust contrast (contrast control). Image is dark. Adjust brightness (brightness control). Adjust light intensity (Illumination Level switch). Darken room and thoroughly dilate patients pupil. Image color is abnormal. Adjust hue (hue control). Collimation is incorrect. Adjust collimation.

Focusing is incorrect. Adjust focus. Patients pupil is not large enough. Darken room and thoroughly dilate patients pupil. Press illumination aperture switch (40).*

Illumination lamp cannot be turned on.

Photography image is completely dark.

Photography image contains a dim white spot.

Photography image is whitened.

Photography image is completely flared.

Collimation is incorrect. Adjust collimation Focusing is incorrect. Adjust focus. Opacity in patients eye. Flare cannot be removed. Patient blinked during photography. Take another picture. Objective lens is stained. Clean lens. Eyelashes obscured patients eye during photography. (Dim light at the bottom of the screen during alignment.) Instruct the patient to open the eye wider and retake the picture. If wide enough, open eyelid (i.e., Take the picture while holding the eyelid open). Flash level is insufficient. Adjust flash level (Flash level switch). Xenon set screws are loose. Refasten screws to terminal. Xenon lamp has exceeded the recommended service life. Change xenon lamp. Power save function is on (flickering power lamp). Press Photography switch and cancel power save function. Lamp terminal is loose. Refasten lamp terminal. Fuse is burnt. Change fuse. The lamp is burnt. Change lamp.

* TRC-NW6SF only

INSTRUCTION MANUAL

97

Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S

98

Nothing is recorded in external recording device.

Cannot see patients pupil center.

Xenon lamp cannot be turned on.

Split lines cannot be seen.

( ) scale is off centre of the monitor.

Internal fixation target cannot be seen.

Collimation is incorrect. Adjust collimation. Instrument is set to external fixation target. Reset to internal fixation target. Fluoro-observation knob is set to visible.* Change to infrared-observation (fluoro-observation knob (13)).* Fuse is burnt. Change fuse. Internal fixation target is set to periphery. Change fixation position to center (periphery fixation switch (42) (reset)). Split line is set to off. Turn split line on (split switch (36)). Diopter compensation lens selector is other than (0). Return diopter compensation lens selector to (0). Patients pupil is not large enough. Darken room and thoroughly dilate patients eye. Power save function is on (flickering power lamp). Cancel power save function (photography switch (16)). Xenon lamp has exceeded the recommended service life. Change xenon lamp. Anomaly in external recording device. Check power supply, settings, ink ribbon, paper, etc. Incorrect cable connection. Check and correct cable connection. Position of patients face (incl. chin, forehead) is incorrect. Allow patient to correct position of face. Patients face height is incorrect. Adjust face height (chinrest switch). Anomaly in external recording device. Check power supply, settings, ink ribbon, paper, etc. Cable connection is incorrect. Check and correct cable connections.

* TRC-NW6SF only

6.3 Error code listIn the event that "Err#" (# ind