NT® PRESSURE TRANSDUCERFOR VACUUM APPLICATIONS
User Guide
ENTEGRIS, INC.Corporate Headquarters3500 Lyman BoulevardChaska, Minnesota 55318 USACustomer Service Tel. 763-502 - 02 0 0Toll Free 877-503-02 0 0Customer Service Fax 763-502-0300w w w. e n t e g r i s . c o mw w w. e n t e g r i s f l u i d h a n d l i n g . c o m
Entegris® and Flaretek® are registered trademarks of Entegris, Inc.NT® is a registered trademark of NT International, an Entegris Company .Kalrez® and Viton® are registered trademarks of DuPont Dow Elastomers.
©2004 Entegris, Inc.Printed in USAP/N 1088 (Rev B 07/04)3960-1587MAX-0704
The materials integrity manage ment company
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 1
NT® PRESSURE TRANSDUCER USER GUIDE
Table of ContentsIntroduction .................................. 2
Ordering Information .................. 3
Identifying NonstandardProduct Configurations .......... 5
Installation .................................... 6
Mechanical Installation ............. 6
Electrical Installation ................ 7
Power Supply Requirements ..... 8
Load Resistance:Current Output ....................... 9
Unit Operation ............................ 10
Operating Environment ........... 10
Process Connections ................ 10
Pressure Transducer Cover Assembly ..................... 10
Temperature Requirements .... 10
Pressure Limits ......................... 10
Pressure ReferenceAccuracy ................................ 10
Linear Output Signal ............... 11
Troubleshooting .......................... 12
Reference ..................................... 13
Physical Specifications ............ 13
Electrical Specifications ......... 13
Performance Specifications .... 13
Certifications .............................. 14
CE Compliance ......................... 14
Dimensions .................................. 15
Repair and Warranty Service . . . . 2 1
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 3
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
2 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Introduction
This manual is for use with NT®
Pressure Transducers for vacuumapplications, single-port and flow-through models 4150 and 4250.These instruments have been designed for use in high purity applications in the semiconductorindustry and are compatible withhighly corrosive processes.
These products feature no movingparts and no filled cavities, whichreduces the possibility of a contami-nated process. The wetted parts ofthese nonmetallic transducers areconstructed with PTFE, sapphireor other similar high purity inertmaterials.
Ordering Information
CODE BASE MODEL
4150 Single-Port Vacuum Pressure Transducer
CODE PRESSURE RANGE
030V -12 to 30 PSIG (-83 to 207 kPa)060V -12 to 60 PSIG (-83 to 414 kPa)100V -12 to 100 PSIG (-83 to 690 kPa)
CODE INLET/OUTLET PORT CONNECTION TYPE
F02 1/4″ Flaretek®F03 3/8″ Flaretek®F04 1/2″ Flaretek®N02 1/4″ FNPT N04 1/2″ FNPT M02 1/4″ MNPT
CODE ELECTRICAL CONNECTOR TYPE
B06 FEP-jacketed 6′ pigtail electrical cableB12 FEP-jacketed 12′ pigtail electrical cableB30 FEP-jacketed 30′ pigtail electrical cableD00 3-pin polypropylene connectorD06 3-pin polypropylene connector and 6 ′ PVC-jacketed mating cableD12 3-pin polypropylene connector and 12 ′ PVC-jacketed mating cableD30 3-pin polypropylene connector and 30 ′ PVC-jacketed mating cable
CODE ELECTRICAL OUTPUTS
A 4-20 mA (12-28 VDC input)
CODE PRIMARY/SECONDARY SEAL
U1 Kalrez® 4079/Viton® (default)U2 Kalrez® 1050/Viton®U3 Kalrez® 6375 UP/Viton®S1 Kalrez® 4079/Kalrez® 4079S2 Kalrez® 1050/Kalrez® 1050S3 Kalrez® 6375 UP/Kalrez® 6375 UP
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 3
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
2 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Introduction
This manual is for use with NT®
Pressure Transducers for vacuumapplications, single-port and flow-through models 4150 and 4250.These instruments have been designed for use in high purity applications in the semiconductorindustry and are compatible withhighly corrosive processes.
These products feature no movingparts and no filled cavities, whichreduces the possibility of a contami-nated process. The wetted parts ofthese nonmetallic transducers areconstructed with PTFE, sapphireor other similar high purity inertmaterials.
Ordering Information
CODE BASE MODEL
4150 Single-Port Vacuum Pressure Transducer
CODE PRESSURE RANGE
030V -12 to 30 PSIG (-83 to 207 kPa)060V -12 to 60 PSIG (-83 to 414 kPa)100V -12 to 100 PSIG (-83 to 690 kPa)
CODE INLET/OUTLET PORT CONNECTION TYPE
F02 1/4″ Flaretek®F03 3/8″ Flaretek®F04 1/2″ Flaretek®N02 1/4″ FNPT N04 1/2″ FNPT M02 1/4″ MNPT
CODE ELECTRICAL CONNECTOR TYPE
B06 FEP-jacketed 6′ pigtail electrical cableB12 FEP-jacketed 12′ pigtail electrical cableB30 FEP-jacketed 30′ pigtail electrical cableD00 3-pin polypropylene connectorD06 3-pin polypropylene connector and 6 ′ PVC-jacketed mating cableD12 3-pin polypropylene connector and 12 ′ PVC-jacketed mating cableD30 3-pin polypropylene connector and 30 ′ PVC-jacketed mating cable
CODE ELECTRICAL OUTPUTS
A 4-20 mA (12-28 VDC input)
CODE PRIMARY/SECONDARY SEAL
U1 Kalrez® 4079/Viton® (default)U2 Kalrez® 1050/Viton®U3 Kalrez® 6375 UP/Viton®S1 Kalrez® 4079/Kalrez® 4079S2 Kalrez® 1050/Kalrez® 1050S3 Kalrez® 6375 UP/Kalrez® 6375 UP
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 5
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
4 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Identifying NonstandardProduct Configurations
CODE BASE MODEL
4250 Flow-Through Vacuum Pressure Transducer
CODE PRESSURE RANGE
030V -12 to 30 PSIG (-83 to 207 kPa)060V -12 to 60 PSIG (-83 to 414 kPa)100V -12 to 100 PSIG (-83 to 650 kPa)
CODE INLET/OUTLET PORT CONNECTION TYPE
F02 1/4″ Flaretek®F03 3/8″ Flaretek®F04 1/2″ Flaretek®F06 3/4″ Flaretek®F08 1″ Flaretek® (030G, 060G types only)
CODE ELECTRICAL CONNECTOR TYPE
B06 FEP-jacketed 6′ pigtail electrical cableB12 FEP-jacketed 12′ pigtail electrical cableB30 FEP-jacketed 30′ pigtail electrical cableD00 3-pin polypropylene connectorD06 3-pin polypropylene connector and 6 ′ PVC-jacketed mating cableD12 3-pin polypropylene connector and 12 ′ PVC-jacketed mating cableD30 3-pin polypropylene connector and 30 ′ PVC-jacketed mating cable
CODE ELECTRICAL OUTPUTS
A 4-20 mA (12-28 VDC input)
CODE PRIMARY/SECONDARY SEAL
U1 Kalrez® 4079/Viton® (default)U2 Kalrez® 1050/Viton®U3 Kalrez® 6375 UP/Viton®S1 Kalrez® 4079/Kalrez® 4079S2 Kalrez® 1050/Kalrez® 1050S3 Kalrez® 6375 UP/Kalrez® 6375 UP
This User Guide applies to productmanufactured as standard NT®
Pressure Transducers single-port and flow-through models 4150 and4250. Entegris also manufacturesnonstandard product to meet theneeds of specific applications.Nonstandard product may havedifferent materials of construction,accuracy specifications, performanceand other specifications that differ-entiate the nonstandard productfrom the standard offering.
NOTE: Nonstandard pressuretransducers may be identified bythe model number found on theproduct label.
Nonstandard single-port pressuretransducers, model 4150 productline, are identified with a “P” fol-lowed by a number code.
For example, part number 4150-030V-F02-B06-A-P04
the “P04” designates the product as anonstandard product manufacturedto certain specifications designatedunder the “P04”code.
Nonstandard flow-through pressuretransducers, model 4250 productline, are identified with a “T” fol-lowed by a number code.
For example, part number 4250-030V-F02-B06-A-T02
the “T02” designates the product as anonstandard product manufacturedto certain specifications designatedunder the “T02” code.
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 5
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
4 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Identifying NonstandardProduct Configurations
CODE BASE MODEL
4250 Flow-Through Vacuum Pressure Transducer
CODE PRESSURE RANGE
030V -12 to 30 PSIG (-83 to 207 kPa)060V -12 to 60 PSIG (-83 to 414 kPa)100V -12 to 100 PSIG (-83 to 650 kPa)
CODE INLET/OUTLET PORT CONNECTION TYPE
F02 1/4″ Flaretek®F03 3/8″ Flaretek®F04 1/2″ Flaretek®F06 3/4″ Flaretek®F08 1″ Flaretek® (030G, 060G types only)
CODE ELECTRICAL CONNECTOR TYPE
B06 FEP-jacketed 6′ pigtail electrical cableB12 FEP-jacketed 12′ pigtail electrical cableB30 FEP-jacketed 30′ pigtail electrical cableD00 3-pin polypropylene connectorD06 3-pin polypropylene connector and 6 ′ PVC-jacketed mating cableD12 3-pin polypropylene connector and 12 ′ PVC-jacketed mating cableD30 3-pin polypropylene connector and 30 ′ PVC-jacketed mating cable
CODE ELECTRICAL OUTPUTS
A 4-20 mA (12-28 VDC input)
CODE PRIMARY/SECONDARY SEAL
U1 Kalrez® 4079/Viton® (default)U2 Kalrez® 1050/Viton®U3 Kalrez® 6375 UP/Viton®S1 Kalrez® 4079/Kalrez® 4079S2 Kalrez® 1050/Kalrez® 1050S3 Kalrez® 6375 UP/Kalrez® 6375 UP
This User Guide applies to productmanufactured as standard NT®
Pressure Transducers single-port and flow-through models 4150 and4250. Entegris also manufacturesnonstandard product to meet theneeds of specific applications.Nonstandard product may havedifferent materials of construction,accuracy specifications, performanceand other specifications that differ-entiate the nonstandard productfrom the standard offering.
NOTE: Nonstandard pressuretransducers may be identified bythe model number found on theproduct label.
Nonstandard single-port pressuretransducers, model 4150 productline, are identified with a “P” fol-lowed by a number code.
For example, part number 4150-030V-F02-B06-A-P04
the “P04” designates the product as anonstandard product manufacturedto certain specifications designatedunder the “P04”code.
Nonstandard flow-through pressuretransducers, model 4250 productline, are identified with a “T” fol-lowed by a number code.
For example, part number 4250-030V-F02-B06-A-T02
the “T02” designates the product as anonstandard product manufacturedto certain specifications designatedunder the “T02” code.
NT® PRESSURE TRANSDUCER USER GUIDE
6 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC. ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 7
NT® PRESSURE TRANSDUCER USER GUIDE
CAUTION!The NT® Pressure Transducer has been factory sealed. Do notattempt to remove the coverof the pressure transducer. Anyattempt at removal of the pressure transducer cover willvoid the warranty and damagethe unit.
CAUTION!Do not tighten the nuts thatprotect the process connectionsduring shipment. Do not tightenthe nuts unless the propertubing has been installed.Tightening these nuts may resultin damage to the pressure trans-ducer process connections.
Installation Electrical Installation The pressure transducer provides an analog (4-20 mA) electricaloutput proportional to the pressuremeasured.
NOTE: The white wire is not required, please refer to the wiring diagram below.
Pressure transducer4-20 mA output version
+24 VDC GND
RED
WHITE
BLACK4-20 mA pressure signal
no connection
display or PLC
Mechanical InstallationNT® Pressure Transducers utilizeFlaretek® flared tube connectionsand must be used with the propertubing size and coupling nut.
When installing flared tubing to thepressure transducer, the flared tubeis pushed over the transducer’sfitting until the fitting reaches thesmaller tube diameter. The amountof torque required to tighten thecoupling nut is dependent upon thesize of the fitting.
Care should be taken when installingflared tubing onto the pressuretransducer. Do not use excessivetorque or subject the transducer tohigh heat during installation. Thetransducer should be firmly mountedto a solid surface to ensure stability.Verify that the pressure transducerand the signal cable are free frommechanical stress or excessive bend-ing from the surrounding equipment.
FITTING SIZE 1/4˝ 3/8˝ 1/2˝ 3/4˝ 1˝
Torque (in•lbs) 5 8 11 14 30
NT® PRESSURE TRANSDUCER USER GUIDE
6 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC. ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 7
NT® PRESSURE TRANSDUCER USER GUIDE
CAUTION!The NT® Pressure Transducer has been factory sealed. Do notattempt to remove the coverof the pressure transducer. Anyattempt at removal of the pressure transducer cover willvoid the warranty and damagethe unit.
CAUTION!Do not tighten the nuts thatprotect the process connectionsduring shipment. Do not tightenthe nuts unless the propertubing has been installed.Tightening these nuts may resultin damage to the pressure trans-ducer process connections.
Installation Electrical Installation The pressure transducer provides an analog (4-20 mA) electricaloutput proportional to the pressuremeasured.
NOTE: The white wire is not required, please refer to the wiring diagram below.
Pressure transducer4-20 mA output version
+24 VDC GND
RED
WHITE
BLACK4-20 mA pressure signal
no connection
display or PLC
Mechanical InstallationNT® Pressure Transducers utilizeFlaretek® flared tube connectionsand must be used with the propertubing size and coupling nut.
When installing flared tubing to thepressure transducer, the flared tubeis pushed over the transducer’sfitting until the fitting reaches thesmaller tube diameter. The amountof torque required to tighten thecoupling nut is dependent upon thesize of the fitting.
Care should be taken when installingflared tubing onto the pressuretransducer. Do not use excessivetorque or subject the transducer tohigh heat during installation. Thetransducer should be firmly mountedto a solid surface to ensure stability.Verify that the pressure transducerand the signal cable are free frommechanical stress or excessive bend-ing from the surrounding equipment.
FITTING SIZE 1/4˝ 3/8˝ 1/2˝ 3/4˝ 1˝
Torque (in•lbs) 5 8 11 14 30
NT® PRESSURE TRANSDUCER USER GUIDE
8 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC. ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 9
NT® PRESSURE TRANSDUCER USER GUIDE
Power Supply Requirements
Figure 1. Power Required for a 4-20 mA Loop
Load Resistance: Current Output (4-20 mA Configuration)
If a load resistor, RLoad, is used inseries with the current output, thevalue of RLoad is dependent on thesupply voltage and the meter resis-tance and is calculated from the following formula.
where:
RLoad = maximum load resistance (ohms)
VPS = power supply voltage (volts)
Rmeter = meter resistance (ohms)(typically not used, therefore Rmeter = 0)
Rwire = resistance of interconnecting wire
RLoad = - (Rmeter + Rwire)VPS - 12 V
0.020A
Operating region
Power Supply Voltage(VPS)
12 24 28
800
0
The pressure transducer requires a12-28 volt DC power supply with lessthan 2% ripple at 100 or 120 Hz. Therequired power supply voltage varieswith the load resistance (RLoad),please refer to the diagrams belowand the formulas on page 9.
The power supply must provideclean power and must be used onlyto power similar measurement-typedevices. The power supply must notbe used to power inductive loads,such as motors, relays or solenoids.These devices may produce tran-sients that may affect the pressuretransducer measurements when theinductive device is powered-up orpowered down.
NOTE: Be sure to ground theshield of the cable to local ground.
NT® PRESSURE TRANSDUCER USER GUIDE
8 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC. ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 9
NT® PRESSURE TRANSDUCER USER GUIDE
Power Supply Requirements
Figure 1. Power Required for a 4-20 mA Loop
Load Resistance: Current Output (4-20 mA Configuration)
If a load resistor, RLoad, is used inseries with the current output, thevalue of RLoad is dependent on thesupply voltage and the meter resis-tance and is calculated from the following formula.
where:
RLoad = maximum load resistance (ohms)
VPS = power supply voltage (volts)
Rmeter = meter resistance (ohms)(typically not used, therefore Rmeter = 0)
Rwire = resistance of interconnecting wire
RLoad = - (Rmeter + Rwire)VPS - 12 V
0.020A
Operating region
Power Supply Voltage(VPS)
12 24 28
800
0
The pressure transducer requires a12-28 volt DC power supply with lessthan 2% ripple at 100 or 120 Hz. Therequired power supply voltage varieswith the load resistance (RLoad),please refer to the diagrams belowand the formulas on page 9.
The power supply must provideclean power and must be used onlyto power similar measurement-typedevices. The power supply must notbe used to power inductive loads,such as motors, relays or solenoids.These devices may produce tran-sients that may affect the pressuretransducer measurements when theinductive device is powered-up orpowered down.
NOTE: Be sure to ground theshield of the cable to local ground.
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 11
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
10 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Operating EnvironmentThe pressure transducer is to bemounted in a well vented and con-trolled environment. Refer to theReference section on page 13 foradditional specifications.
Process ConnectionsTo avoid possible pressure leaks,make sure all process connectionshave been performed in accordancewith the Mechanical Installationguidelines on page 6.
Pressure Transducer Cover AssemblyNT® Pressure Transducer covers are factory sealed and should not betampered with or opened. Openingthe cover shall void the product warranty.
NOTE: Any attempt to remove ortamper with the transducer coverwill void the warranty and damagethe unit.
Temperature RequirementsThe pressure transducer is rated for use with fluids at processtemperatures between 50° to 95°F(10° to 35°C) under normal operating conditions.
Pressure LimitsTRANSDUCER MAX. PRESSURE LIMITRANGE @73°F (23°C)
-12 to 30 PSIG 45 PSIG(-83 to 207 kPa) (310 kPa)
-12 to 60 PSIG 90 PSIG(-83 to 414 kPa) (621 kPa)
-12 to 100 PSIG 150 PSIG(-83 to 690 kPa) (1034 kPa)
WARNING!The pressure limits for NT®Pressure Transducers models4150 and 4250 decrease sig-nificantly for temperaturesabove 95°F (35°C) Please consult the factory.
Pressure Reference AccuracyThe accuracy of the pressure trans-ducer output is ±1% of span. Forexample, a -12 to 30 PSIG rangedpressure transducer has a span of 42 PSIG and an accuracy of ±0.42PSIG. This accuracy includes theeffects of linearity, hysteresis and repeatability, measured at room temperature.
Unit Operation Linear Output SignalThe output signal of the NT® PressureTransducer is a linear function pro-portional to the applied pressure.
Pressure measurement-12 PSIG Full scale
20 mA
4 mA
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 11
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
10 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Operating EnvironmentThe pressure transducer is to bemounted in a well vented and con-trolled environment. Refer to theReference section on page 13 foradditional specifications.
Process ConnectionsTo avoid possible pressure leaks,make sure all process connectionshave been performed in accordancewith the Mechanical Installationguidelines on page 6.
Pressure Transducer Cover AssemblyNT® Pressure Transducer covers are factory sealed and should not betampered with or opened. Openingthe cover shall void the product warranty.
NOTE: Any attempt to remove ortamper with the transducer coverwill void the warranty and damagethe unit.
Temperature RequirementsThe pressure transducer is rated for use with fluids at processtemperatures between 50° to 95°F(10° to 35°C) under normal operating conditions.
Pressure LimitsTRANSDUCER MAX. PRESSURE LIMITRANGE @73°F (23°C)
-12 to 30 PSIG 45 PSIG(-83 to 207 kPa) (310 kPa)
-12 to 60 PSIG 90 PSIG(-83 to 414 kPa) (621 kPa)
-12 to 100 PSIG 150 PSIG(-83 to 690 kPa) (1034 kPa)
WARNING!The pressure limits for NT®Pressure Transducers models4150 and 4250 decrease sig-nificantly for temperaturesabove 95°F (35°C) Please consult the factory.
Pressure Reference AccuracyThe accuracy of the pressure trans-ducer output is ±1% of span. Forexample, a -12 to 30 PSIG rangedpressure transducer has a span of 42 PSIG and an accuracy of ±0.42PSIG. This accuracy includes theeffects of linearity, hysteresis and repeatability, measured at room temperature.
Unit Operation Linear Output SignalThe output signal of the NT® PressureTransducer is a linear function pro-portional to the applied pressure.
Pressure measurement-12 PSIG Full scale
20 mA
4 mA
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 13
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
12 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Troubleshooting the NT® PressureTransducers may be accomplishedby measuring the output signal ofthe pressure transducer with abattery powered current/voltagemeter. The meter may be placed inseries with the pressure transducerto measure the current output or itmay be used to directly measure thevoltage output.
Using the battery powered current/volt meter is an effective method todetermine whether the device or theon-site data acquisition system is notfunctioning properly.
Troubleshooting
The following lists the specificationsfor NT® Pressure Transducers,models 4150 and 4250. Consult thefactory for product specificationsmanufactured for nonstandard applications.
NOTE: Specifications are subject to change without notice. Pleaseconsult the factory for the mostcurrent information.
Reference
CONSTRUCTION MATERIALS
Wetted parts:
Body: PTFE
Sensor Interface: Sapphire
O-Ring: Kalrez® 4079, Kalrez® 1050 or Kalrez® 6375 UP
Nonwetted parts: Polypropylene, polyethylene and PVDF (In addition to materials listed above)
Connection type: Flaretek®
Physical Specifications
Electrical Specifications
Input power: 24 VDC (12-28 VDC)
Pressure signal output ranges: 4-20 mA
Electrical connection: 6′, 12′, 30′ FEP-coated pigtail or polypropylene 3-pin connector with PVC cable
Reference accuracy: ±1% of span (includes linearity, hysteresisand repeatability) at 73°F (23°C)
Process temperature: 50° to 95°F (10° to 35°C)
Performance Specifications
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 13
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
12 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
Troubleshooting the NT® PressureTransducers may be accomplishedby measuring the output signal ofthe pressure transducer with abattery powered current/voltagemeter. The meter may be placed inseries with the pressure transducerto measure the current output or itmay be used to directly measure thevoltage output.
Using the battery powered current/volt meter is an effective method todetermine whether the device or theon-site data acquisition system is notfunctioning properly.
Troubleshooting
The following lists the specificationsfor NT® Pressure Transducers,models 4150 and 4250. Consult thefactory for product specificationsmanufactured for nonstandard applications.
NOTE: Specifications are subject to change without notice. Pleaseconsult the factory for the mostcurrent information.
Reference
CONSTRUCTION MATERIALS
Wetted parts:
Body: PTFE
Sensor Interface: Sapphire
O-Ring: Kalrez® 4079, Kalrez® 1050 or Kalrez® 6375 UP
Nonwetted parts: Polypropylene, polyethylene and PVDF (In addition to materials listed above)
Connection type: Flaretek®
Physical Specifications
Electrical Specifications
Input power: 24 VDC (12-28 VDC)
Pressure signal output ranges: 4-20 mA
Electrical connection: 6′, 12′, 30′ FEP-coated pigtail or polypropylene 3-pin connector with PVC cable
Reference accuracy: ±1% of span (includes linearity, hysteresisand repeatability) at 73°F (23°C)
Process temperature: 50° to 95°F (10° to 35°C)
Performance Specifications
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 15
NT® PRESSURE TRANSDUCER USER GUIDE
Installation Drawings
Dim AØ 1.88″ (47.8 mm)
1.00″ (25.4 mm)
Dim B
3 conductor FEP-jacketed cable, pigtail connection (BXX) (Cable length up to 30 feet)
4150 FEP-jacketed Electrical Cable Connection, Single-port
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ Flaretek® 2.90″ (73.7 mm) 0.99″ (25.2 mm)
3/8″ Flaretek® 2.90″ (73.7 mm) 1.06″ (26.9 mm)
1/2″ Flaretek® 2.90″ (73.7 mm) 1.14″ (29.0 mm)
ELECTRICAL CONNECTOR INFORMATION
NT® PRESSURE TRANSDUCER USER GUIDE
14 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
CE Compliance
Entegris products have been testedto various test standards required bythe EMC 89/336/EEC directive. Theresults of this testing are on file atEntegris and are available uponrequest.
Please contact the factory for thelatest information. The most current specifications may be found on theInternet at:http://www.nt-intl.com/ce_etl-info.cfm.
Certifications
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 15
NT® PRESSURE TRANSDUCER USER GUIDE
Installation Drawings
Dim AØ 1.88″ (47.8 mm)
1.00″ (25.4 mm)
Dim B
3 conductor FEP-jacketed cable, pigtail connection (BXX) (Cable length up to 30 feet)
4150 FEP-jacketed Electrical Cable Connection, Single-port
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ Flaretek® 2.90″ (73.7 mm) 0.99″ (25.2 mm)
3/8″ Flaretek® 2.90″ (73.7 mm) 1.06″ (26.9 mm)
1/2″ Flaretek® 2.90″ (73.7 mm) 1.14″ (29.0 mm)
ELECTRICAL CONNECTOR INFORMATION
NT® PRESSURE TRANSDUCER USER GUIDE
14 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
CE Compliance
Entegris products have been testedto various test standards required bythe EMC 89/336/EEC directive. Theresults of this testing are on file atEntegris and are available uponrequest.
Please contact the factory for thelatest information. The most current specifications may be found on theInternet at:http://www.nt-intl.com/ce_etl-info.cfm.
Certifications
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 17
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
16 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
4150 FEP-jacketed Electrical Cable Connection, Single-port (NPT version)
1.00″ (25.4 mm)
Dim A
Dim B
Ø 1.88″ (47.8 mm)
3 conductor FEP-jacketed cable, pigtail connection (BXX) (Cable length up to 30 feet)
4150 Polypropylene Connector Connection, Single-port
3.00″ (76.2 mm)
Dim A
Dim B
Ø 1.88″ (47.8 mm)
3 conductor connector/cable (DXX)(Cable length up to 30 feet)
WIRE 4-20 MA OUTPUT VERSION
Black Ground (-) 4-20 mA
Red 24 VDC (+) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ Flaretek® 2.90″ (73.7 mm) 0.99″ (25.2 mm)
3/8″ Flaretek® 2.90″ (73.7 mm) 1.06″ (26.9 mm)
1/2″ Flaretek® 2.90″ (73.7 mm) 1.14″ (29.0 mm)
ELECTRICAL CONNECTOR INFORMATION
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ MNPTMALE 2.90″ (73.7 mm) 0.59″ (15.0 mm)
1/4″ FNPTFEMALE 3.24″ (82.3 mm) –
1/2″ FNPTFEMALE 3.50″ (88.9 mm) –
ELECTRICAL CONNECTOR INFORMATION
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 17
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
16 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
4150 FEP-jacketed Electrical Cable Connection, Single-port (NPT version)
1.00″ (25.4 mm)
Dim A
Dim B
Ø 1.88″ (47.8 mm)
3 conductor FEP-jacketed cable, pigtail connection (BXX) (Cable length up to 30 feet)
4150 Polypropylene Connector Connection, Single-port
3.00″ (76.2 mm)
Dim A
Dim B
Ø 1.88″ (47.8 mm)
3 conductor connector/cable (DXX)(Cable length up to 30 feet)
WIRE 4-20 MA OUTPUT VERSION
Black Ground (-) 4-20 mA
Red 24 VDC (+) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ Flaretek® 2.90″ (73.7 mm) 0.99″ (25.2 mm)
3/8″ Flaretek® 2.90″ (73.7 mm) 1.06″ (26.9 mm)
1/2″ Flaretek® 2.90″ (73.7 mm) 1.14″ (29.0 mm)
ELECTRICAL CONNECTOR INFORMATION
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ MNPTMALE 2.90″ (73.7 mm) 0.59″ (15.0 mm)
1/4″ FNPTFEMALE 3.24″ (82.3 mm) –
1/2″ FNPTFEMALE 3.50″ (88.9 mm) –
ELECTRICAL CONNECTOR INFORMATION
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 19
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
18 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
1.86″ (47.2 mm)
1.75″ (44.5 mm)
1.75″ (44.5 mm)
2.36″ (59.9 mm)
0.20″ (5.1 mm) (2)
Dim A
Dim B
0.25″ (6.4 mm)Dim C
1.00″ (25.4 mm)
3 conductor FEP-jacketed cable, pigtail connection (BXX)(Cable length up to 30 feet)
4250 FEP-jacketed Electrical Cable Connection, Flow-through
INLET/OUTLETPORT CONNECTION DIM A DIM B DIM C
1/4″ Flaretek® 3.53″ (89.7 mm) 3 . 7 2″ (94.5 mm) 0.73″ (18.5 mm)
3/8″ Flaretek® 3.53″ (89.7 mm) 3.87″ (98.3 mm) 0.67″ (17.0 mm)
1/2″ Flaretek® 3.78″ (96.0 mm) 4.03″ (102.4 mm) 0.85″ (21.6 mm)
3/4″ Flaretek® 4.11″ (104.4 mm) 4.27″ (108.5 mm) 1.00″ (25.4 mm)
1″ Flaretek® 4.43″ (112.5 mm) 4.75″ (120.7 mm) 1.20″ (30.5 mm)
3.00″ (76.2 mm)
Dim A
Dim B
Ø 1.88″ (47.8 mm)
3 conductor connector/cable (DXX)(Cable length up to 30 feet)
4150 FEP-jacketed Electrical Cable Connection, Single-port (NPT Version)
WIRE 4-20 MA OUTPUT VERSION
Black Ground (-) 4-20 mA
Red 24 VDC (+) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ MNPTMALE 2.90″ (73.7 mm) 0.59″ (15.0 mm)
1/4″ FNPTFEMALE 3.07″ (78.0 mm) –
1/2″ FNPTFEMALE 3.33″ (84.6 mm) –
ELECTRICAL CONNECTOR INFORMATION
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
ELECTRICAL CONNECTOR INFORMATION
ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 19
NT® PRESSURE TRANSDUCER USER GUIDENT® PRESSURE TRANSDUCER USER GUIDE
18 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC.
1.86″ (47.2 mm)
1.75″ (44.5 mm)
1.75″ (44.5 mm)
2.36″ (59.9 mm)
0.20″ (5.1 mm) (2)
Dim A
Dim B
0.25″ (6.4 mm)Dim C
1.00″ (25.4 mm)
3 conductor FEP-jacketed cable, pigtail connection (BXX)(Cable length up to 30 feet)
4250 FEP-jacketed Electrical Cable Connection, Flow-through
INLET/OUTLETPORT CONNECTION DIM A DIM B DIM C
1/4″ Flaretek® 3.53″ (89.7 mm) 3 . 7 2″ (94.5 mm) 0.73″ (18.5 mm)
3/8″ Flaretek® 3.53″ (89.7 mm) 3.87″ (98.3 mm) 0.67″ (17.0 mm)
1/2″ Flaretek® 3.78″ (96.0 mm) 4.03″ (102.4 mm) 0.85″ (21.6 mm)
3/4″ Flaretek® 4.11″ (104.4 mm) 4.27″ (108.5 mm) 1.00″ (25.4 mm)
1″ Flaretek® 4.43″ (112.5 mm) 4.75″ (120.7 mm) 1.20″ (30.5 mm)
3.00″ (76.2 mm)
Dim A
Dim B
Ø 1.88″ (47.8 mm)
3 conductor connector/cable (DXX)(Cable length up to 30 feet)
4150 FEP-jacketed Electrical Cable Connection, Single-port (NPT Version)
WIRE 4-20 MA OUTPUT VERSION
Black Ground (-) 4-20 mA
Red 24 VDC (+) 4-20 mA
White Not used
INLET/OUTLETPORT CONNECTION DIM A DIM B
1/4″ MNPTMALE 2.90″ (73.7 mm) 0.59″ (15.0 mm)
1/4″ FNPTFEMALE 3.07″ (78.0 mm) –
1/2″ FNPTFEMALE 3.33″ (84.6 mm) –
ELECTRICAL CONNECTOR INFORMATION
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
ELECTRICAL CONNECTOR INFORMATION
NT® PRESSURE TRANSDUCER USER GUIDE
20 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC. ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 21
NT® PRESSURE TRANSDUCER USER GUIDE
Repair and WarrantyServiceE n t e g r i s, Inc. warrants each new productagainst defects in materials and work-manship for 18 months of time fromshipment. Entegris makes no warranty,express or implied, with respect to anyc o m p o n e n t s, products, information orservices provided by any third party. With respect to products specificallymanufactured to Buyer’s specifications,Entegris warrants only that such productshave been manufactured, without testing,to Buyer’s specifications but makes noother warranty, express or implied. UponEntegris’ confirmation of defects, Entegrisa g r e e s, as its E X C L U S I V E R E M E DY t orepair or, in its discretion, replace defec-tive product without charge. Entegris isnot responsible for product damage r e-sulting from accident, misuse or lack of reasonable care. IM P O R TA N T: NOOT H E R WA R R A N T Y, W R I T T E NO RV E R B A L ,I SA U T H O R I Z E D BY E N T E G R I S .E N T E G R I SI SN OT L I A B L EF O RA N YS P E C I A L ,I N C I D E N TA LO RC O N S E Q U E N-T I A LD A M A G E S .I M P L I E D WA R R A N T I E SO FM E R C H A N TA B I L I T YA N DF I T N E S SF O R A PA R T I C U L A RP U R P O S EA R ES P E C I F I C A L LY D I S C L A I M E D .
Repair and warranty service is availableat the Entegris factory. To expedite thereturn and repair of the product, contactEntegris at 877-503-0200. A ReturnMaterials Authorization (RMA) numberand a product packaging and return procedure will be provided at that time.
WA R N I N G !All products returned to Entegrismust be accompanied by a list ofall chemicals that have been incontact with the unit and thecorresponding Material SafetyData Sheet (MSDS) for eachchemical. Returned products willnot be accepted without an MSDS.
Dim A
Dim B
0.25″ (6.4 mm)Dim C
3.00″ (76.2 mm)
3 conductor connector/cable (DXX)(Cable length up to 30 feet)
4250 Polypropylene Connector Connection, Flow-through
1.86″ (47.2 mm)
1.75″ (44.5 mm)
1.75″ (44.5 mm)
2.36″ (59.9 mm)
0.20″ (5.1 mm) (2)
INLET/OUTLETPORT CONNECTION DIM A DIM B DIM C
1/4″ Flaretek® 3.53″ (89.7 mm) 3 . 7 2″ (94.5 mm) 0.73″ (18.5 mm)
3/8″ Flaretek® 3.53″ (89.7 mm) 3.87″ (98.3 mm) 0.67″ (17.0 mm)
1/2″ Flaretek® 3.78″ (96.0 mm) 4.03″ (102.4 mm) 0.85″ (21.6 mm)
3/4″ Flaretek® 4.11″ (104.4 mm) 4.27″ (108.5 mm) 1.00″ (25.4 mm)
1″ Flaretek® 4.43″ (112.5 mm) 4.75″ (120.7 mm) 1.20″ (30.5 mm)
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
ELECTRICAL CONNECTOR INFORMATION
NT® PRESSURE TRANSDUCER USER GUIDE
20 Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 ENTEGRIS, INC. ENTEGRIS, INC. Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300 21
NT® PRESSURE TRANSDUCER USER GUIDE
Repair and WarrantyServiceE n t e g r i s, Inc. warrants each new productagainst defects in materials and work-manship for 18 months of time fromshipment. Entegris makes no warranty,express or implied, with respect to anyc o m p o n e n t s, products, information orservices provided by any third party. With respect to products specificallymanufactured to Buyer’s specifications,Entegris warrants only that such productshave been manufactured, without testing,to Buyer’s specifications but makes noother warranty, express or implied. UponEntegris’ confirmation of defects, Entegrisa g r e e s, as its E X C L U S I V E R E M E DY t orepair or, in its discretion, replace defec-tive product without charge. Entegris isnot responsible for product damage r e-sulting from accident, misuse or lack of reasonable care. IM P O R TA N T: NOOT H E R WA R R A N T Y, W R I T T E NO RV E R B A L ,I SA U T H O R I Z E D BY E N T E G R I S .E N T E G R I SI SN OT L I A B L EF O RA N YS P E C I A L ,I N C I D E N TA LO RC O N S E Q U E N-T I A LD A M A G E S .I M P L I E D WA R R A N T I E SO FM E R C H A N TA B I L I T YA N DF I T N E S SF O R A PA R T I C U L A RP U R P O S EA R ES P E C I F I C A L LY D I S C L A I M E D .
Repair and warranty service is availableat the Entegris factory. To expedite thereturn and repair of the product, contactEntegris at 877-503-0200. A ReturnMaterials Authorization (RMA) numberand a product packaging and return procedure will be provided at that time.
WA R N I N G !All products returned to Entegrismust be accompanied by a list ofall chemicals that have been incontact with the unit and thecorresponding Material SafetyData Sheet (MSDS) for eachchemical. Returned products willnot be accepted without an MSDS.
Dim A
Dim B
0.25″ (6.4 mm)Dim C
3.00″ (76.2 mm)
3 conductor connector/cable (DXX)(Cable length up to 30 feet)
4250 Polypropylene Connector Connection, Flow-through
1.86″ (47.2 mm)
1.75″ (44.5 mm)
1.75″ (44.5 mm)
2.36″ (59.9 mm)
0.20″ (5.1 mm) (2)
INLET/OUTLETPORT CONNECTION DIM A DIM B DIM C
1/4″ Flaretek® 3.53″ (89.7 mm) 3 . 7 2″ (94.5 mm) 0.73″ (18.5 mm)
3/8″ Flaretek® 3.53″ (89.7 mm) 3.87″ (98.3 mm) 0.67″ (17.0 mm)
1/2″ Flaretek® 3.78″ (96.0 mm) 4.03″ (102.4 mm) 0.85″ (21.6 mm)
3/4″ Flaretek® 4.11″ (104.4 mm) 4.27″ (108.5 mm) 1.00″ (25.4 mm)
1″ Flaretek® 4.43″ (112.5 mm) 4.75″ (120.7 mm) 1.20″ (30.5 mm)
WIRE 4-20 MA OUTPUT VERSION
Red 24 VDC (+) 4-20 mA
Black Ground (-) 4-20 mA
White Not used
ELECTRICAL CONNECTOR INFORMATION
NT® PRESSURE TRANSDUCERFOR VACUUM APPLICATIONS
User Guide
ENTEGRIS, INC.Corporate Headquarters3500 Lyman BoulevardChaska, Minnesota 55318 USACustomer Service Tel. 763-502 - 02 0 0Toll Free 877-503-02 0 0Customer Service Fax 763-502-0300w w w. e n t e g r i s . c o mw w w. e n t e g r i s f l u i d h a n d l i n g . c o m
Entegris® and Flaretek® are registered trademarks of Entegris, Inc.NT® is a registered trademark of NT International, an Entegris Company .Kalrez® and Viton® are registered trademarks of DuPont Dow Elastomers.
©2004 Entegris, Inc.Printed in USAP/N 1088 (Rev B 07/04)3960-1587MAX-0704
The materials integrity manage ment company