DEVICE SPECIFICATIONS NI 5783 Four Channel Transceiver Adapter Module This document lists specifications for the NI 5783 adapter module. Pair these specifications with the specifications listed in your FlexRIO FPGA module specifications document or your Controller for FlexRIO specifications document. Caution The protection provided by the NI 5783 can be impaired if it is used in a manner not described in this document. Caution To avoid permanent damage to the NI 5783, disconnect all signals connected to the NI 5783 before powering down the module, and only connect signals after the module has been powered on by the FlexRIO FPGA module or the Controller for FlexRIO. Note All numeric specifications are typical unless otherwise noted. All graphs illustrate the performance of a representative module. Specifications are subject to change without notice. For the most recent device specifications, visit ni.com/manuals. Contents FlexRIO Documentation........................................................................................................... 2 Analog Input............................................................................................................................. 3 General Characteristics..................................................................................................... 3 Typical Specifications....................................................................................................... 4 Analog Output........................................................................................................................... 9 General Characteristics..................................................................................................... 9 Typical Specifications..................................................................................................... 10 CLK/REF IN........................................................................................................................... 13 TRIG General Characteristics................................................................................................. 14 AUX I/O (Port 0 DIO <0..3>, Port 1 DIO <0..3>, and PFI <0..3>......................................... 15 Power...................................................................................................................................... 15 Physical................................................................................................................................... 15 Environment............................................................................................................................ 16 Operating Environment................................................................................................... 16 Storage Environment.......................................................................................................16 Shock and Vibration................................................................................................................ 16 Compliance and Certifications................................................................................................ 17 Safety.............................................................................................................................. 17 Electromagnetic Compatibility....................................................................................... 17

Transcript of NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency...

Page 1: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

DEVICE SPECIFICATIONS

NI 5783Four Channel Transceiver Adapter Module

This document lists specifications for the NI 5783 adapter module. Pair these specificationswith the specifications listed in your FlexRIO FPGA module specifications document or yourController for FlexRIO specifications document.

Caution The protection provided by the NI 5783 can be impaired if it is used in amanner not described in this document.

Caution To avoid permanent damage to the NI 5783, disconnect all signalsconnected to the NI 5783 before powering down the module, and only connectsignals after the module has been powered on by the FlexRIO FPGA module or theController for FlexRIO.

Note All numeric specifications are typical unless otherwise noted. All graphsillustrate the performance of a representative module.

Specifications are subject to change without notice. For the most recent device specifications,visit ni.com/manuals.

ContentsFlexRIO Documentation........................................................................................................... 2Analog Input............................................................................................................................. 3

General Characteristics..................................................................................................... 3Typical Specifications....................................................................................................... 4

Analog Output...........................................................................................................................9General Characteristics..................................................................................................... 9Typical Specifications..................................................................................................... 10

CLK/REF IN........................................................................................................................... 13TRIG General Characteristics.................................................................................................14AUX I/O (Port 0 DIO <0..3>, Port 1 DIO <0..3>, and PFI <0..3>.........................................15Power...................................................................................................................................... 15Physical................................................................................................................................... 15Environment............................................................................................................................16

Operating Environment...................................................................................................16Storage Environment.......................................................................................................16

Shock and Vibration................................................................................................................16Compliance and Certifications................................................................................................17

Safety.............................................................................................................................. 17Electromagnetic Compatibility....................................................................................... 17

Page 2: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

CE Compliance .............................................................................................................. 17Online Product Certification........................................................................................... 18Environmental Management........................................................................................... 18

FlexRIO Documentation

Table 1. FlexRIO Documentation Locations and Descriptions

Document Location Description

Getting started guidefor your FlexRIOFPGA module orController forFlexRIO

Available from the Start menuand at ni.com/manuals.

Contains installation instructionsfor your FlexRIO system.

Specificationsdocument for yourFlexRIO FPGAmodule or Controllerfor FlexRIO

Available from the Start menuand at ni.com/manuals.

Contains specifications for yourFlexRIO FPGA module orController for FlexRIO.

Getting started guidefor your adaptermodule

Available from the Start menuand at ni.com/manuals.

Contains signal information,examples, and CLIP details foryour adapter module.

Specificationsdocument for youradapter module

Available from the Start menuand at ni.com/manuals.

Contains specifications for youradapter module.

LabVIEW FPGAModule Help

Embedded in LabVIEW Helpand at ni.com/manuals.

Contains information about thebasic functionality of theLabVIEW FPGA Module.

Real-Time ModuleHelp

Embedded in LabVIEW Helpand at ni.com/manuals.

Contains information about real-time programming concepts, step-by-step instructions for usingLabVIEW with the Real-TimeModule, reference informationabout Real-Time Module VIs andfunctions, and information aboutLabVIEW features on real-timeoperating systems.

2 | ni.com | NI 5783 Specifications

Page 3: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Table 1. FlexRIO Documentation Locations and Descriptions (Continued)

Document Location Description

FlexRIO Help Available from the Start menuand at ni.com/manuals.

Contains information about theFPGA module front panelconnectors and I/O, controller forFlexRIO front panel connectorsand I/O, programminginstructions, and adapter modulecomponent-level IP (CLIP).

FlexRIO AdapterModule DevelopmentKit User Manual

Available from the Start menuat Start»All Programs»National Instruments»NIFlexRIO»NI FlexRIOAdapter Module DevelopmentKit»Documentation.

Contains information about howto create custom adapter modulesfor use with FlexRIO FPGAmodules.

LabVIEW Examples Available in NI ExampleFinder. In LabVIEW, clickHelp»Find Examples»Hardware Input and Output»FlexRIO.

Contains examples of how to runFPGA VIs and Host VIs on yourdevice.

IPNet Located at ni.com/ipnet. Contains LabVIEW FPGAfunctions and intellectual propertyto share.

FlexRIO product page Located at ni.com/flexrio. Contains product information anddata sheets for FlexRIO devices.

Analog Input

General CharacteristicsNumber of channels Four, single-ended, simultaneously sampled

Connector Type HDBNC (high-density BNC)

Input type 50 Ω

Input coupling DC

NI 5783 Specifications | © National Instruments | 3

Page 4: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Sample rate

Internal Sample Clock 100 MHz

External Sample Clock 60 MHz to 100 MHz

Analog-to-digital converter (ADC)

Type Quad, 16-bit

Part number AD9653

Related InformationFor more information about the ADC, refer to the AD9653 datasheet at www.analog.com.

Typical SpecificationsFull-scale input range (normal operating conditions)

Elliptic 2.030 Vpk-pk

Butterworth 2.037 Vpk-pk

DC accuracy

Elliptic ±[(0.80% × reading) + 3.5 mV]

Butterworth ±[(1.00% × reading) + 3.75 mV]

Input impedance

Elliptic 50 Ω ± 0.5%

Butterworth 50 Ω ± 0.8%

Bandwidth (-3 dB)

Elliptic 39.4 MHz

Butterworth 39.5 MHz

Table 2. AI Spectral Performance

SNR1 SINAD1 ENOB (bits)2 SFDR3

74.5 dBFS 74.3 dBFS 12.05 -87 dBc

Note All AI spectral performance values apply to both the Elliptic and Butterworthvariants.

1 Measured at 10.1 MHz with a -1 dBFS signal adjusted to full-scale.2 Calculated from SINAD corrected to fullscale.3 Measured at 10.1 MHz with a -1 dBFS signal.

4 | ni.com | NI 5783 Specifications

Page 5: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Table 3. AI Noise Spectral Density

Filter Variant nV/√(Hz) dBm/Hz dBFS/Hz

Elliptic 17.5 -142.1 -152.2

Butterworth 17.8 -142.0 -152.1

Table 4. AI Channel Crosstalk (10 MHz)

Channel N±1 Channel N±2 Channel N±3

-79 dBc -87 dBc -91 dBc

Note All AI channel crosstalk values apply to both the Elliptic and Butterworthvariants.

Figure 1. AI Crosstalk

Am

plitu

de (

dBc)

–60

–100–98–96–94–92–90–88–86–84–82–80–78–76–74–72–70–68–66–64–62

Frequency (Hz)

40 M0 4 M 8 M 12 M 16 M 20 M 24 M 28 M 32 M 36 M

Elliptic N±1

Elliptic N±2

Elliptic N±3

Butterworth N±1

Butterworth N±2

Butterworth N±3

NI 5783 Specifications | © National Instruments | 5

Page 6: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Figure 2. AI Frequency Response (Zoomed Out)

Nor

mal

ized

Am

plitu

de to

100

kH

z (d

B)

0

–80

–75

–70

–65

–60

–55

–50

–45

–40

–35

–30

–25

–20

–15

–10

–5

Frequency (Hz)

120 M0 10 M 20 M 30 M 40 M 50 M 60 M 70 M 80 M 90 M 100 M 110 M

Butterworth Elliptic

Figure 3. AI Frequency Response (Zoomed In)

Nor

mal

ized

Am

plitu

de to

100

kH

z (d

B)

0

–10

–9

–8

–7

–6

–5

–4

–3

–2

–1

Frequency (Hz)

50 M2 M0 6 M 10 M 14 M 18 M 22 M 26 M 30 M 34 M 38 M 42 M 46 M

Butterworth Elliptic

6 | ni.com | NI 5783 Specifications

Page 7: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Figure 4. AI Step Response (Butterworth)

Am

plitu

de (

V)

1

–50 m0

50 m100 m150 m200 m250 m300 m350 m400 m450 m500 m550 m600 m650 m700 m750 m800 m850 m900 m950 m

Time (ns)

2000 20–10 40 60 80 100 120 140 160 180

Figure 5. AI Step Response (Elliptic)

Am

plitu

de (

V)

–50 m0

50 m100 m150 m200 m250 m300 m350 m400 m450 m500 m550 m600 m650 m700 m750 m800 m850 m900 m950 m

1

Time (ns)

2000 20–10 40 60 80 100 120 140 160 180

NI 5783 Specifications | © National Instruments | 7

Page 8: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Figure 6. AI Spectral (10.1 MHz at -1 dBFS input signal, 6.1 kHz RBW, 10 Averages)

Am

plitu

de (

dBF

S)

0

–120–115–110–105–100–95–90–85–80–75–70–65–60–55–50–45–40–35–30–25–20–15–10–5

Frequency (Hz)

50 M0 2 M 6 M 10 M 14 M 18 M 22 M 26 M 30 M 34 M 38 M 42 M 46 M

Note AI Spectral figure applies to both the Elliptic and Butterworth variants.

Figure 7. AI Return Loss

Am

plitu

de (

dB)

0

–42–40–38–36–34–32–30–28–26–24–22–20–18–16–14–12–10–8–6–4–2

Frequency (Hz)

60 M10 M 15 M 20 M 25 M 30 M 35 M 40 M 45 M 50 M 55 M

Butterworth Elliptic

8 | ni.com | NI 5783 Specifications

Page 9: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Analog Output

General CharacteristicsNumber of channels Four, single-ended, simultaneously updated

Connector type HDBNC (high-density BNC)

Output type 50 Ω

Output coupling DC

Digital-to-analog converter (DAC)

Type Quad, 16-bit

Part number DAC3484

Minimum analog input to analog output response time4

100 MS/s 1130 ns

200 MS/s 720 ns

400 MS/s 550 ns

Table 5. AO Sample Rates

Clocking Mode Data Rate (per channel) DAC Update Rate (perchannel)

Internal Clock, 2x interpolation5 400 MS/s 800 MS/s

Internal Clock, 4x interpolation(default clocking mode)

200 MS/s 800 MS/s

Internal Clock, 8x interpolation 100 MS/s 800 MS/s

External Clock, 2x interpolation5 240 MS/s to 400 MS/s 480 MS/s to 800 MS/s

External Clock, 4x interpolation 120 MS/s to 200 MS/s 480 MS/s to 800 MS/s

External Clock, 8x interpolation 60 MS/s to 100 MS/s 480 MS/s to 800 MS/s

Related InformationFor more information about the DAC, refer to the DAC3484 datasheet at www.ti.com.

4 Minimum time to digitize a signal (AI) and output a response (AO). Time measured from signalentering the AI connector, passing into and out of the LabVIEW FPGA diagram, and observed atthe AO connector.

5 400 MS/s with 2x interpolation is available only when operating in 2 channel analog output mode.

NI 5783 Specifications | © National Instruments | 9

Page 10: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Typical SpecificationsFull-scale output range (normal operation conditions)

50 Ω 1.001 Vpk-pk

High-Z 2.002 Vpk-pk

DC accuracy (into High-Z) ±[(2.0% × desired voltage) + 4.4 mV]

Output impedance 50 Ω ± 0.7%

SFDR6 -81 dBc

Table 6. AO Noise Spectral Density (into 50 Ω)

nV/√(Hz) dBm/Hz dBFS/Hz

5.8 -151.7 -155.7

Table 7. AO Channel Crosstalk (10 MHz)

Channel N±1 Channel N±2 Channel N±3

-90 dBc -98 dBc -99 dBc

Figure 8. AO Crosstalk (into 50 Ω Load)

Am

plitu

de (

dBc)

–106

–104

–102

–100

–98

–96

–94

–92

–90

–88

–86

–84

–82

–80

–78

–76

–74

Frequency (Hz)

100 M0 10 M 20 M 30 M 40 M 50 M 60 M 70 M 80 M 90 M

N±3N±2N±1

6 10.1 MHz tone at -1 dBFS.

10 | ni.com | NI 5783 Specifications

Page 11: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Figure 9. AO Frequency Response Across Data Rate

Nor

mal

ized

Am

plitu

de to

100

kH

z (

dB)

0

–10

–9.5–9.0–8.5–8.0–7.5–7.0–6.5–6.0–5.5–5.0–4.5–4.0–3.5–3.0–2.5–2.0–1.5–1.0–0.5

Frequency (Hz)

200 M0 20 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M

100 MS/s 200 MS/s 400 MS/s

Figure 10. AO Phase Noise (Signal at 12.1 MHz)

Am

plitu

de (

dBc/

Hz)

–95

–155

–150

–145

–140

–135

–130

–125

–120

–115

–110

–105

–100

Offset from Carrier (Hz)

1 M10 100 1 k 10 k 100 k

–160

NI 5783 Specifications | © National Instruments | 11

Page 12: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Figure 11. AO Return Loss

Am

plitu

de (

dB)

0

–32

–30

–28

–26

–24

–22

–20

–18

–16

–14

–12

–10

–8

–6

–4

–2

Frequency (Hz)

300 M10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280 M

Figure 12. AO Two Tone, Each Tone at -7 dBFS, 69.95 MHz, and 70.05 MHz, 500 HzRBW

0

–95–90–85–80–75–70–65–60–55–50–45–40–35–30–25–20–15–10–5

Frequency (Hz)

70.5 M69.5 M 69.6 M 69.7 M 69.8 M 69.9 M 70.0 M 70.1 M 70.2 M 70.3 M 70.4 M

Am

plitu

de (

dBm

)

Note The noise floor in the above figure is limited by the noise floor of themeasurement device. Refer to the AO Noise Spectral Density table for moreinformation.

12 | ni.com | NI 5783 Specifications

Page 13: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

CLK/REF INConnector type HDBNC (high-density BNC)

Input impedance 50 Ω

Input coupling AC

Reference input voltage range 0.75 Vpk-pk to 5.2 Vpk-pk

Sample Clock input voltage range 0.4 Vpk-pk to 5.2 Vpk-pk

Absolute maximum voltage ±8.0 VDC, 8.0 Vpk-pk AC

Duty cycle 45% - 55%

Clock Configuration ExternalClock Type

ExternalClock

Frequency

Description

Internal Clock PLL Off7 The internal VCXO acts as a free-running Sample Clock.

Internal Clock PLL On(TbRef)

10 MHz The internal VCXO locks toTbRefClk, which is providedthrough the backplane.

Internal Clock PLL On(CLK/REF IN)

ReferenceClock

10 MHz The internal VCXO locks to anexternal Reference Clock, whichis provided through the CLK/REFIN front panel connector.

External Clock PLL Off(CLK/REF IN)

SampleClock

60 MHz to100 MHz

An external Sample Clock can beprovided through the CLK/REFIN front panel connector.

Internal VCXO phase noise

10 Hz -80 dBc/Hz

100 Hz -110 dBc/Hz

1 kHz -140 dBc/Hz

10 kHz -150 dBc/Hz

100 kHz -155 dBc/Hz

1 MHz -160 dBc/Hz

10 MHz -162 dBc/Hz

7 Default clocking configuration.

NI 5783 Specifications | © National Instruments | 13

Page 14: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Figure 13. Internal Sample Clock Phase Noise

Am

plitu

de (

dBc/

Hz)

–85

–175–170–165–160–155–150–145–140–135–130–125–120–115–110–105–100–95–90

Offset Frequency (from carrier)

10 M10 100 1 k 10 k 100 k 1 M

TRIG General CharacteristicsNumber of channels 1, single-ended

Connector type HDBNC

Coupling DC

Impedance

Input 10 kΩ

Output 50 Ω

Logic level 3.3 V LVCMOS

Voltage

VIH_MIN 2 V

VIL_MAX 0.8 V

VOH_MIN (unloaded) 3.1 V

VOL_MAX (unloaded) 0.2 V

Absolute maximum voltage ±20 VDC, +21 dBm (7.1 Vpk-pk)

14 | ni.com | NI 5783 Specifications

Page 15: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

AUX I/O (Port 0 DIO <0..3>, Port 1 DIO <0..3>,and PFI <0..3>Number of channels 12 bidirectional (8 DIO and 4 PFI)

Connector type HDMI

Interface standard 3.3 V LVCMOS

Interface logic

Maximum VIL 0.8 V

Minimum VIH 2.0 V

Maximum VOL 0.4 V

Minimum VOH 2.7 V

Maximum VOH 3.6 V

Zout 50 Ω ± 20%

Iout (DC) ±2 mA

Pull-down resistor 150 kΩ

Recommended operating voltage -0.3 V to 3.6 V

Overvoltage protection ±10 V

Maximum toggle frequency 100 MHz

+5 V maximum current 10 mA

+5 V voltage tolerance 4.2 V to 5 V

PowerTotal power, typical operation 4.6 W

PhysicalDimensions 12.9 x 2.0 x 12.1 cm (5.1 x 0.8 x 4.7 in.)

Weight 420 g (14.8 oz)

Front panel connectors Ten HDBNC and one HDMI

NI 5783 Specifications | © National Instruments | 15

Page 16: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

EnvironmentMaximum altitude 2,000 m (800 mbar) (at 25 °C ambient

temperature)

Pollution Degree 2

Indoor use only.

Operating EnvironmentAmbient temperature range 0 °C to 55 °C (Tested in accordance with

IEC 60068-2-1 and IEC 60068-2-2. MeetsMIL-PRF-28800F Class 3 low temperaturelimit and MIL-PRF-28800F Class 2 hightemperature limit.)

Relative humidity range 10% to 90%, noncondensing (Tested inaccordance with IEC 60068-2-56.)

Storage EnvironmentAmbient temperature range -20 °C to 70 °C (Tested in accordance

with IEC 60068-2-1 and IEC 60068-2-2. MeetsMIL-PRF-28800F Class 3 limits.)

Relative humidity range 5% to 95%, noncondensing (Tested inaccordance with IEC 60068-2-56.)

Shock and VibrationOperating shock 30 g peak, half-sine, 11 ms pulse (Tested in

accordance with IEC 60068-2-27. MeetsMIL-PRF-28800F Class 2 limits.)

Random vibration

Operating 5 Hz to 500 Hz, 0.3 grms

Nonoperating 5 Hz to 500 Hz, 2.4 grms (Tested in accordancewith IEC 60068-2-64. Nonoperating testprofile exceeds the requirements ofMIL-PRF-28800F, Class 3.)

16 | ni.com | NI 5783 Specifications

Page 17: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Compliance and Certifications

SafetyThis product is designed to meet the requirements of the following electrical equipment safetystandards for measurement, control, and laboratory use:• IEC 61010-1, EN 61010-1• UL 61010-1, CSA 61010-1

Note For UL and other safety certifications, refer to the product label or the OnlineProduct Certification section.

Electromagnetic CompatibilityThis product meets the requirements of the following EMC standards for electrical equipmentfor measurement, control, and laboratory use:• EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity• EN 55011 (CISPR 11): Group 1, Class A emissions• EN 55022 (CISPR 22): Class A emissions• EN 55024 (CISPR 24): Immunity• AS/NZS CISPR 11: Group 1, Class A emissions• AS/NZS CISPR 22: Class A emissions• FCC 47 CFR Part 15B: Class A emissions• ICES-001: Class A emissions

Note In the United States (per FCC 47 CFR), Class A equipment is intended foruse in commercial, light-industrial, and heavy-industrial locations. In Europe,Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intendedfor use only in heavy-industrial locations.

Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medicalequipment that does not intentionally generate radio frequency energy for thetreatment of material or inspection/analysis purposes.

Note For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

CE Compliance This product meets the essential requirements of applicable European Directives, as follows:• 2014/35/EU; Low-Voltage Directive (safety)• 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

NI 5783 Specifications | © National Instruments | 17

Page 18: NI 5783 Specifications - National Instruments · –14 –12 –10 –8 –6 –4 –2 Frequency (Hz) 10 M 40 M 60 M 80 M 100 M 120 M 140 M 160 M 180 M 200 M 220 M 240 M 260 M 280

Online Product CertificationRefer to the product Declaration of Conformity (DoC) for additional regulatory complianceinformation. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in theCertification column.

Environmental ManagementNI is committed to designing and manufacturing products in an environmentally responsiblemanner. NI recognizes that eliminating certain hazardous substances from our products isbeneficial to the environment and to NI customers.

For additional environmental information, refer to the Minimize Our Environmental Impactweb page at ni.com/environment. This page contains the environmental regulations anddirectives with which NI complies, as well as other environmental information not included inthis document.

Waste Electrical and Electronic Equipment (WEEE)EU Customers At the end of the product life cycle, all NI products must bedisposed of according to local laws and regulations. For more information abouthow to recycle NI products in your region, visit ni.com/environment/weee.

电子信息产品污染控制管理办法(中国 RoHS)中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物

质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录

ni.com/environment/rohs_china。(For information about China RoHScompliance, go to ni.com/environment/rohs_china.)

Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks.Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patentscovering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, thepatents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information aboutend-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the ExportCompliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy andhow to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIESAS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to theapplicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2016 National Instruments. All rights reserved.

375445A-01 Apr16