NGA 2000 Platform Hardware

72
www.EmersonProcess.com Instruction Manual Platform for NGA 2000 CLD, FID, HFID, MLT and WCLD Analyzer Modules 1/1 19“, 3 HU Housing 3 rd Edition 04/2008 Instruction Manual 90002496 04/2008

Transcript of NGA 2000 Platform Hardware

Page 1: NGA 2000 Platform Hardware

www.EmersonProcess.com

Instruction ManualPlatformfor NGA 2000 CLD, FID, HFID, MLTand WCLD Analyzer Modules1/1 19“, 3 HU Housing

3rd Edition 04/2008

Instruction Manual9000249604/2008

Page 2: NGA 2000 Platform Hardware

Platform for CLD, FID, HFID, MLT AND WCLD Instruction Manual90002496

04/2008

Emerson Process ManagementGmbH & Co. OHGIndustriestrasse 1D-63594 HasselrothGermanyT +49 (0) 6055 884-0F +49 (0) 6055 884-209Internet: www.EmersonProcess.com

ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!

Emerson Process Management (Rosemount Analytical) designs, manufactures and testsits products to meet many national and international standards. Because these instrumentsare sophisticated technical products, you MUST properly install, use, and maintainthem to ensure they continue to operate within their normal specifications. The followinginstructions MUST be adhered to and integrated into your safety program when installing,using and maintaining Emerson Process Management (Rosemount Analytical) products.Failure to follow the proper instructions may cause any one of the following situations tooccur: Loss of life; personal injury; property damage; damage to this instrument; and warrantyinvalidation.

• Read all instructions prior to installing, operating, and servicing the product.

• If you do not understand any of the instructions, contact your Emerson ProcessManagement (Rosemount Analytical) representative for clarification.

• Follow all warnings, cautions, and instructions marked on and supplied with the product.

• Inform and educate your personnel in the proper installation, operation, andmaintenance of the product.

• Install your equipment as specified in the Installation Instructions of the appropriateInstruction Manual and per applicable local and national codes. Connect all productsto the proper electrical and pressure sources.

• To ensure proper performance, use qualified personnel to install, operate, update, program,and maintain the product.

• When replacement parts are required, ensure that qualified people use replacement partsspecified by Emerson Process Management (Rosemount Analytical). Unauthorized partsand procedures can affect the product’s performance, place the safe operation of yourprocess at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire,electrical hazards, or improper operation.

• Ensure that all equipment doors are closed and protective covers are in place, exceptwhen maintenance is being performed by qualified persons, to prevent electricalshock and personal injury.

The information contained in this document is subject to change without notice. Misprintsreserved.

1st Edition 06/1995 2nd Edition 03/1999 3rd Edition 04/2008 © 2008 by Emerson Process Management

Page 3: NGA 2000 Platform Hardware

CONTENTS

i90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Table of Contents

SAFETY SUMMARY S - 1

General S - 2

Gases and Gas Conditioning (Sample Handling of analyzer modules) S - 3

Supply Voltage S - 4

Instrument specific notes for the user S - 5

Additional notes for service / maintenance S - 6

Electrostatic Discharge S - 7

INTRODUCTION E - 1

Platform E - 1

EMC-Housing E - 1

DESCRIPTION

1. Technical Description 1 - 1

1.1 Front Panel 1 - 3

1.2 Rear Panel 1 - 4

1.3 Internal Layout 1 - 5

1.3.1 Interconnection Board ICB 01 1 - 7

1.3.2 Power Supply Unit 1 - 9

1.3.3 Line - Fan - Module (LFM 01) 1- 10

Ventilation 1 - 11

1.3.4 Controller Board ACU 02 1 - 12

1.3.5 Network Module LEM 01 1 - 13

2.

- (open)

4.

Page 4: NGA 2000 Platform Hardware

CONTENTS

ii 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

OPERATION

5. Preparation 5 - 1

5.1 Installation Site 5 - 1

5.2 External Modules 5 - 2

5.3 Installation of an Analyzer Module into the Platform 5 - 3

5.4 Installation of PCBs / Rear Mounting Modules 5 - 4

5.4.1 Rear Mounting Modules 5 - 4

5.4.2 Internal Slots (with 5 A power supply only) 5 - 4

6. Switching On 6 - 1

6.1 24 Vdc - Operation 6 - 2

6.2 230 / 115 Vac - Operation 6 - 2

7. Switching Off 7 - 1

8.

- (open)

10.

11. TROUBLESHOOTING 11 - 1

11.1 Instrument has no Function (LCD display is dark) 11 - 1

11.2 No Measurement Screen 11 - 2

11.3 Line - Fan - Module (LFM 01) 11 - 3

11.4 Power Supply Unit 11 - 3

11.5 Fuses 11 - 4

11.5.1 AC - Fuses 11 - 4

11.5.2 DC - Fuses 11 - 4

12.

- (open)

14.

Page 5: NGA 2000 Platform Hardware

CONTENTS

iii90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

MAINTENANCE

15. Opening the Housing 15 - 1

15.1 Front Panel 15 - 1

15.2 Housing Cover 15 - 2

16. (open)

17. Removal / Replacement of Components 17 - 1

17.1 Removal of Analyzer Module 17 - 1

17.2 Removal / Replacement of PCBs / Rear Mounting Modules 17 - 2

17.2.1 Rear Mounting Modules 17 - 2

17.2.2 Internal Slots 17 - 2

17.3 Removal / Replacement of Front Panel 17 - 3

17.4 Replacement of EPROM / Buffer Battery on the ACU 02 17 - 4

17.4.1 Removal of ACU 02 17 - 4

17.4.2 Replacement of EPROM 17 - 5

17.4.3 Replacement of Buffer Battery 17 - 6

17.4.4 Installation of ACU 02 17 - 6

18. (open)

19. Cleaning of Housing Outside 19 - 1

Page 6: NGA 2000 Platform Hardware

CONTENTS

iv 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

20. TECHNICAL DATA 20 - 1

20.1 Platform 20 - 1

20.1.1 Housing 20 - 1

20.1.2 Environments 20 - 1

20.2 Line - Fan - Module (LFM 01) 20 - 4

20.2.1 Electrical Safety 20 - 4

20.3 Internal Power Supply Unit 20 - 5

20.3.1 5 A Power Supply 20 - 5

20.3.2 10 A Power Supply 20 - 5

SUPPLEMENT

21. Pin Assignments 21 - 1

21.1 24 Vdc Input (LFM 01) 21 - 1

21.2 230/115 Vac Input (LFM 01) 21 - 1

21.3 24 Vdc Output to analyzer module (internal) 21 - 1

INDEX R - 1

List of Figures R - 5

Page 7: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Safety Summary

Outside and/or inside platform or at operation manual resp. different symbols gives you a hint to

special sources of danger.

Source of danger !See Operation Manual!

Electrostatic Discharge (ESD) !

Explosives !

Hot components !

Toxic !

Risk to health !

Platform specific notes for the user !

In operation manual we will give partly additional informations to these symbols.

Strictly follow these instructions please !

GENERAL

Page 8: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalGENERAL

1. General

◆ The following general safety precautions must be observed during all phases of operation,

service and repair of this instrument !

Failure to comply with these precautions or with specific warnings elsewhere in this manual

violates safety standards of design, manufacture and intended use of this instrument !

Failure to comply with these precautions may lead to personal injury and damage to this

instrument !

◆ Fisher-Rosemount GmbH & Co. does not take responsibility (liability) for the customer´s

failure to comply with these requirements !

◆ Do not attempt internal service or adjustment unless other person, capable of rendering first

aid and resuscitation, is present !

◆ Because of the danger of introducing additional hazards, do not perform any unauthorized

modification to the instrument !

Return the instrument to a Fisher-Rosemount Sales and Service office for service or repair

to ensure that safety features are maintained !

◆ Instruments which appear damaged or defective should be made inoperative and secured

against unintended operation until they can be repaired by qualified service personnel.

Page 9: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 390002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalGENERAL / GASES AND GAS CONDITIONING (SAMPLE HANDLING)

Operating personnel must not remove instrument covers !Component replacement and internal adjustments must be made by qualifiedservice personnel only !

Read all operation manuals before attempting to operate with the instrument !Be sure to observe the additional notes, safety precautions and warnings givenin the individual operation manuals (e. g. platform, analyzer modules andI/O-modules) !

Do not operate the instrument in the presence of flammable gases or explosiveatmosphere without supplementary protective measures !

At photometer or heated components there could be exist hot components !

2. Gases and Gas Conditioning (Sample Handling of analyzer modules)

Be sure to observe the safety regulations for the respective gases(sample gas and test gases / span gases) and the gas bottles !

Inflammable or explosive gas mixtures must not be purged into the instrumentwithout supplementary protective measures !

To avoid a danger to the operators by explosive, toxic or unhealthy gascomponents, first purge the gas lines with ambient air or nitrogen (N2) beforecleaning or exchange parts of the gas paths.

Page 10: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 4 90002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalSUPPLY VOLTAGE

3. Supply Voltage

The socket outlet shall be installed near the equipment and shall be easilyaccessible to disconnect the device from the socket outlet.

Verify whether the line voltage stated on the instrument (LFM 01) or power supplyagrees with that of your mains line!

Be sure to observe the safety precautions and warnings given bymanufacturer of power supply !

◆ NGA 2000 platform is a Safety Class 1 instrument

The platform is provided with a protective earth terminal.To prevent shock hazard, the instrument chassis and cabinet must be connectedto an electrical ground. The instrument must be connected to the AC powersupply mains through a three-conductor power cable, with the third wire firmlyconnected to an electrical ground (safety ground) at the power outlet.If the instrument is to be energized via an external power supply, that goes for thepower supply too.Any interruption of the protective (grounding) conductor or disconnection of theprotective earth terminal will cause a potential shock hazard that could result inpersonal injury. Deliberate disconnection is inadmissible / prohibited !

Verify that the position of input voltage switch of the 10 A power supply agreeswith that of your mains line (Fig. 20-3 and 1-4) !

Verify correct polarity for 24 V dc - operation !

Page 11: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 590002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

4. Instrument specific notes for the user

The installation site for the instrument has to be dry and remain above freezingpoint at all times.The instrument must be exposed neither to direct sunlight nor to strong sourcesof heat. Be sure to observe the permissible ambient temperature !For outdoor sites, we recommend to install the instrument in a protective cabinet.At least, the instrument has to be protected against rain (e.g., shelter).

Free flow of air into and out of the platform (ventilation slits) must not be hinderedby nearby objects or walls !

Use only from our factory optional delivered cables or equivalent shielded cablesto be in agreement with the CE - conformity.The customer has to guarantee, that the shield is be connected bothsided.Shield and connectors housing have to be connected conductive.Sub.-min.-D-plugs/sockets have to be screwed to the analyzer.

By using of optional delivering terminal strip adapters the analyzer is not be inagreement with the CE - conformity. In this case CE - conformity is to be declaredby customer as “manufacturer of system”.

INSTRUMENT SPECIFIC NOTES FOR USER

Page 12: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 6 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

5. Additional notes for service / maintenance

Operating personnel must not remove instrument covers !Component replacement and internal adjustments must be made by qualifiedservice personnel only !

Always disconnect power, discharge circuits and remove external voltagesources before troubleshooting, repair or replacement of any component !

Any work inside the instrument without switching off the power must beperformed by a specialist, who is familiar with the related danger, only !

In case of exchanging fuses the customer has to be certain that fuses of specifiedtype and rated current are used. It is prohibited to use repaired fuses or defectivefuse holders or to short-circuit fuse carriers (fire hazard).

At component replacement or installation the RF shielding contacts must not bebended !

ADDITIONAL NOTES FOR SERVICE / MAINTENANCE

Page 13: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 790002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

5.1 Electrostatic Discharge

The electronic parts of the analyzer can be irreparably damaged if exposed to electrostatic

discharge (ESD).

The instrument is ESD protected when the covers have been secured and safety precautions

observed. When the housing is open, the internal components are not ESD protected anymore.

Although the electronic parts are reasonable safe to handle, you should be aware of the following

considerations:

Best ESD example is when you walked across a carpet and then touched an electrical grounded

metal doorknob. The tiny spark which has jumped is the result of electrostatic discharge (ESD).

You prevent ESD by doing the following:

Remove the charge from your body before opening the housing and maintain during work with

opened housing, that no electrostatic charge can be built up.

Ideally you are opening the housing and working at an ESD - protecting workstation.

Here you can wear a wrist trap.

However, if you do not have such a workstation, be sure to do the following procedure exactly:

Discharge the electric charge from your body. Do this by touching a device that is grounded

electrically (any device that has a three - prong plug is grounded electrically when it is plugged into

a power receptacle).

This should be done several times during the operation with opened housing (especially after

leaving the service site because the movement on a low conducting floors or in the air might cause

additional ESDs).

ELECTROSTATIC DISCHARGE

Page 14: NGA 2000 Platform Hardware

SAFETY SUMMARY

S - 8 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Page 15: NGA 2000 Platform Hardware

INTRODUCTION

E - 190002496(2) [NGA-Platform e] xx.03.99

Rosemount Analytical

Introduction

This operation manual informs you about the platform, which is used for accommodating

analyzers or analysis modules of the NGA 2000 series (19", 3 HU, IP 20-housing).

Platform

This housing, together with the operator interface and the built-in controller board ACU 02,

represents the platform for analyzers of the NGA 2000 series.

EMC-Housing

Should the housing be used as an EMC-assembly carrier for non-EMC analysis modules, please

be aware that the installation differs from the description given in this manual. The following

changes must be taken into consideration:

1. The 19" operator interface (chapter 1.1) will be substituted by two 1/2 19" blind front panels.

Each blind front panel can be dismounted separately by removing the respective screws

(6 for each panel).

2. The controller board ACU 02 (chapter 1.3.4) will be replaced by a bulkhead plate.

3. Only rear-panel modules using network can be built-in in this housing.

Modules which need the ICB-bus can not be connected, because the ACU 02 board is not

installed.

4. The operation of built-in analyzer module and any additional rear-panel modules must be

carried out via a "true-type" platform or a superior analyzer.

Page 16: NGA 2000 Platform Hardware

INTRODUCTION

E - 2 90002496(2) [NGA-Platform e] xx.03.99

Rosemount Analytical

Page 17: NGA 2000 Platform Hardware

1 - 1

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

1. Technical Description

The basis for analyzers of the NGA 2000 series is a 19", 3 HU housing of IP 20 protection class.

The instrument can be delivered as table top or rack mountable unit.

The platform contains all components necessary for operating the analyzer and can be used as

a "stand alone" analysis instrument when combined with an analyzer module and an Input/Output

(I/O) module, or it can be configured as a system when equipped with several analyzer and I/O

modules.

The conceptional design is depicted in the block diagram below (Fig. 1-1).

Fig. 1-1: Block diagram platform design (external)

Page 18: NGA 2000 Platform Hardware

1 - 2

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

NGA standard housing contains:

◆ the front panel (84 DU, see chapter 1.1)

◆ the interconnection board ICB 01 (see chapter 1.3.1)

◆ the line-fan module LFM 01 (12 DU, see chapter 1.3.3), the power supply (see chapter 1.3.2)

and the controller board ACU 02 (see chapter 1.3.4)

◆ the network module LEM 01 for connecting the platform with external analyzer modules

(see chapter 1.3.5)

◆ 5 free slots (PCBs, 5 DU) on the rear panel, which can be assigned optionally with

“Input / Output Modules” (I/O's)

e.g. programmable I/O's: one SIO and up to 4 DIO boards

◆ 4 internal free slots (Euro cards) which can be assigned optionally (for 5 A power supply only)

Fig. 1-1a: Analyzer system via platform (schematic diagram)

A n alyze r M od u le

A n alyze r M od u le

Po w e r s u p p lyNetwork Cable

24VD C Cable

24VD C Cable

AC DC

Plat fo r m

A n alyze r M od u le

Network Cable

Page 19: NGA 2000 Platform Hardware

1 - 3

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

F 3F 1 F 2 F 4 F 5

N G A 2 0 0 0

Softkeys

(Function depends on menu)

LCD displayCursor keys

Fastening screws

front panel

Footing

(table-top housing only)

ENTER key

Mounting

(with rack mount-

able housing only)

Fig. 1-2: Housing, Front View (Front panel)

1.1 Front Panel

The 1/1-19" front panel (84 DU) accommodates the LCD display, accompanying front panel circuit

board AFP 01 and the keys necessary for operating the instrument (see Fig. 1-2).

For assembling and disassembling the analyzer module and for maintenance and repair, the front

panel can be swivelled out after releasing the six fastening screws, while the instrument may

remain in operation.

FRONT PANEL

Page 20: NGA 2000 Platform Hardware

1 - 4

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

1.2 Rear Panel

The rear panel (84 DU) accommodates (see Fig. 1-3):

◆ the line-fan module LFM 01 (12 DU, see chapter 1.3.3)

◆ the network module LEM 01 for connecting the platform with external analyzer modules

(see chapter 1.3.5)

◆ 5 free slots (PCBs, 5 DU) on the rear panel, which can be assigned optionally with

“Input / Output Modules” (I/O's)

e.g. programmable I/O's: one SIO and up to 4 DIO boards

◆ the gas fittings and optional electrical connections of analyzer modules (see separately

operation manual)

Analyzer module

(optionally built-in)

Line - Fan - Module

LFM 01 (12 DU)5 free slots for optional

rear-panel modules

(PCB's, 5 DU)Mounting (at the front)

(with rack mountable

housing only)

Fig. 1-3: Housing, Rear View

Network Module

LEM 01 (5 DU)

Footing

(table-top housing only)

REAR PANEL

Page 21: NGA 2000 Platform Hardware

1 - 5

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

1.3 Internal Layout

The interior of the housing is divided into two 42 DU wide halves.When looking from the front side, on the right there is the electronics with interconnection boardand printed circuit board (PCB of Euro standard format).On the left side, there is spare room for locating an analyzer module.

Network Module

LEM 01 (5 DU)

Line- Fan - Module

LFM 01 (12 DU)

Interconnection Board

ICB 01

Controller Board

ACU 02

Analysis module

(optionally built-in)

Front panel with LCD display

and AFP 01 - PCB

5 A

Power Supply(for MLT, TFID

analyzer modules)

Mounting

rack mountable housing

(option)Carrying handles

Fig. 1-4a: Internal view of housing (with 5 A power supply)

Retention pin of analysis module(dependent on analysis module version)

hinge

front panel

5 free slots for optional I/O's

(PCBs, 5 DU)

4 optional internal PCBs

INTERNAL LAYOUT

Page 22: NGA 2000 Platform Hardware

1 - 6

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalINTERNAL LAYOUT

Line- Fan - Module

LFM 01 (12 DU)

Interconnection Board

ICB 01

Fan

Analyzer module

(optionally built-in)

Front panel with LCD display

and AFP 01 - PCB

10 A

Power Supply(for CLD, WCLD,

HFID, McFidanalyzer modules)

Carrying handles

Fig. 1-4b: Internal view of housing (with 10 A power supply)

hinge

front panel

5 free slots for optional I/O's

(PCBs, 5 DU)

Controller Board ACU 02

Network Module

LEM 01 (5 DU)

24 Vdcto

analyzer module

Page 23: NGA 2000 Platform Hardware

1 - 7

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

1.3.1 Interconnection Board ICB 01

The board ICB 01 (see Figs. 1-4 and 1-5) interconnects the line-fan module (LFM 01), the power

supply (PSM) and all inserted PCBs [Euro cards with parallel ICB-bus system] and built-in analysis

module (RJ 45 network interface).

PCBs and the required voltage supply can be connected to the ICB with the so-called ICB-bus

[(a 64-pin plug-in connection DIN 41612 C 64) see Fig. 1-7].

ICB ensures the internal network connection which is necessary for operating the analyzer /

analyzer module.

With 5 A power supply it supplies the analyzer module with 24 Vdc required.

INTERCONNECTION BOARD ICB 01

Slot for

power supply

Slot for

Controller board ACU 02

(with 5 A power supply)

24 Vdc to

analyzer module

(with 5 A power supply)

internal network - connection

to analyzer module

(RJ 45 socket)

Fig. 1-5: Printed Circuit Board ICB 01(viewed from front panel)

Slots for PCBs

with 64-pin ICB-Bus

(double-sided)

Slot for network module

LEM 01

(Rear side)

Slot for

Controller board ACU 02

(with 10 A power supply)

Page 24: NGA 2000 Platform Hardware

1 - 8

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Standard Inputs/OutputsSIO

(Power Supply)24 Vdc / 5 A

Main-filter Fuse

PlatformFront panel

Analyzer Module (AM)

Network - Module

Optional Inputs/Outputs (4 x)Optional PCBs (4 x)

Control Unit ACU 01

Autorange230 / 115 Vac

Line-Fan-ModuleLFM 01

Mains

Fuse

24 VdcFilterLogic24 Vdc

Keys

AFP - PCB

LCD - Dusplay

24 V

dc

Ner

twor

k

Inte

r-C

onne

ctio

n-B

oard

ICB

01

Analyzer Module

24 Vdc

Network

Network

24 Vdc

24 Vdc

24 Vdc

230/115 Vac

24 Vdc out

FanPower Supply

L1, N, PE

ACU AFP LCDPCB

PCB PCB

ICB 01

Keyboard

L1, N, PE

24 Vdc

NEB 01

LFM 01

Fan

ext. 24 Vdc in

EMI filter

or

Fig. 1-6b: Block diagram internal platform design (with 10 A power supply)

Fig. 1-6a: Block diagram internal platform design (with 5 A power supply)

INTERCONNECTION BOARD ICB 01

Page 25: NGA 2000 Platform Hardware

1 - 9

TECHNICAL DESCRIPTION

90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Fig. 1-7: Pin assignment, 64-pin ICB-bus connector

1.3.2 Power Supply Unit

The power supply unit converts the voltage supply from 230 / 115 Vac to 24 Vdc required for

operating the instrument.

Alternatively two different power supplys can be built in (see Fig. 1-4a and 1-4b too):

One power supply can be loaded with up to 5 A and is provided with autoranging between 230

Vac and 115 Vac (rear panel of LFM 01 shows “230/115 VAC”).

The other power supply can be loaded with up to 10 A and is provided with a manual switch

between 230 Vac and 115 Vac (rear panel of LFM 01 shows “230 VAC” or “115 VAC”). To prevent

excessively high temperature inside the housing, a fan is installed inside (see Figs. 1-4b). The air

enters through openings in the front of the side walls.

The necessary connection of line-fan module (see 1.4) is provided by the ICB 01 board via a

matching plug (DIN 41612 / H 15).

The slot is located in front on the outer right side of ICB 01 (when looking from the front panel),

see Figs. 1-5 and 1-4.

To exchange 5 A power supply (standard) to 10 A power supply via retrofit kit, contact service

support center please.

INTERCONNECTION BOARD ICB 01 / POWER SUPLLY UNIT

Page 26: NGA 2000 Platform Hardware

TECHNICAL DESCRIPTION

1 - 10 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

13

2

-DC 24 V IN-

LFM 01

F1

/F2

T3

,5A

/ 2

30

V

POWER

230 V

115 V

50/60 Hz

~~

LINE-FAN MODULE (LFM 01)

1.3.3 Line - Fan - Module (LFM 01)

The NGA basic design has been provided for an operating voltage of 230 Vac (196 - 264 Vac), 47

- 63 Hz and 115 Vac (93 - 132 Vac) resp., 47 - 63 Hz and / or 24 Vdc (± 10 %).

The line-fan-module (LFM 01) connects the NGA unit to the AC mains line or to a 24 Vdc low

voltage supply.

The unit (12 DU) is put in the housing from the rear side (see Figs. 1-8, 1-3 and 1-4).

Fig. 1-8: Rear panel LFM 01 (sectional view of the housing)

Fan

Fuses AC

(F1/F2: T 3,5 A)

AC - connector

(230 / 115 Vac;

50 / 60 Hz)

24 Vdc - connector

Page 27: NGA 2000 Platform Hardware

TECHNICAL DESCRIPTION

1 - 1190002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalLINE-FAN MODULE (LFM 01)

AC - voltage cable is connected via an IEC power lead. Power failure (RF noise) or internal

interference is suppressed by an RF filter with two integrated fuses (5 x 20 mm).

If rear panel of LFM 01 shows “230/115 VAC”, there is built-in the 5 A power supplywith autoranging of input AC voltage.If rear panel of LFM 01 shows “230 VAC” or “115 VAC”, there is built-in the 10 Apower supply with manual switch of input AC voltage..

Verify beforehand that the line voltage stated on the LFM 01agrees with that of your power supply line !Verify that the position of input voltage switch of the 10 A power supply agreeswith that of your power supply line (Fig. 20-3 and 1-4) !

The 24 Vdc voltage is connected via a 3-pin socket (XLR), whereby filtered DC voltage has to be

available. The remaining interferences will be prevented by internal filters.

Verify correct polarity for 24 V dc - operation !

The instrument is protected against reversed polarity.

If both 24 Vdc and 230/115 Vac voltage are supplied simultaneously, the AC voltage will be

preferred. The 24 Vdc then is automatically disconnected by a relay. The 24 Vdc will be used only

while cessation of the AC voltage.

Ventilation

To prevent excessively high temperature inside the housing, a ventilator is installed on the rear side

of the LFM 01 (see Figs. 1-8 and 1-3).

The ventilator sucks air from the housing.

The air enters through openings in the front of the side walls.

Page 28: NGA 2000 Platform Hardware

TECHNICAL DESCRIPTION

1 - 1290002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

1.3.4 Controller Board ACU 02

The ACU 02 controller board is the “heart” of the NGA platform. It is a one board CPU card. When

the voltage supply is disconnected, all user data will be safe loaded via a battery buffer.

This board functions as a networks manager, allows the access to each built-in component, and

controls the analyzer modules, the LCD display, and the operation procedures.

The slot for ACU 02 is located on the ICB 01 (when viewed from the front panel) on the right side

of the 64-pin ICB bus with internal5 A power supply and on the left side with internal 10 A power

supply (left from power supplys), see Figs. 1-5 and 1-4.

The board contains the following function blocks:

◆ 32 / 16-bit micro controller

◆ Floating point co-processor

◆ dependent on platform version different RAM / Flash-EEPROM

(Standard: 1 MB RAM extension for up to 15 software channels*); Option: additional 0.5 MB RAM extension

if more than 15 software channels*);contact factory if more than 20 software channels are requested)

◆ Real-time Clock with

Calendar function

Alarm function

◆ Watchdog function

◆ Serial interface RS 232 C

◆ A network interface with ECHELON chip

◆ System bus:

Parallel bus A6 : D8

12 TPU wiring

network bus

Synchron serial bus

◆ Buffered parallel interface for LCD control

◆ Local bus interface (e.g. storage extension)

◆ Switch-mode power supply for 5 V supply (on board)*) Calculate software channels (SW) as follows: MLT = 1 SW ch. for each measuring channel; all other NGA analyzer modules like

CLD, FID, etc. = 1 SW ch.; SIO/DIO = zero SW ch.; network I/O's = 1/2 SW ch.

CONTROLLER BOARD ACU 02

Page 29: NGA 2000 Platform Hardware

TECHNICAL DESCRIPTION

1 - 1390002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

NET1

NET2

1.3.5 Network Module LEM 01

The network module LEM 01 (see Fig. 1-9) connects the platform (controller board) with external

modules via network. The slot on the interconnection board is located on the outer right (viewed

from the rear side), left from the optionally built-in analyzer module.

Note !

The external modules cannot be supplied with 24 Vdc from the platform and, therefore, have to

be externally powered (see Fig. 1-1a, too).

NETWORK MODULE LEM 01

Analyzer module

(optionally built-in)Network - twin - connector

to external modules

(two RJ 45-sockets)

Fig. 1-9: Rear panel network module LEM 01 (sectional view of the housing)

RJ 45 sockets are serving for network interconnection.

It is necessary to terminate the ends of a twisted pair bus to minimize reflections (see Fig. 1-11).

Failure to terminate the bus will degrade network performance.

Termination will be done via RJ 45 connectors (see Fig. 1-10).

Fig. 1-10: RJ 45 network termination connector

Page 30: NGA 2000 Platform Hardware

TECHNICAL DESCRIPTION

1 - 14 90002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalNETWORK MODULE LEM 01

AM AM AM AM

Platform

AM AM

Term

inat

ion

Term

inat

ion

AM AM AM AM AM

Term

inat

ion

Term

inat

ion

AM

Platform

AM

AM AM

Term

inat

ion

Platform

Term

inat

ion

AM

Platform

Fig. 1-11: Network termination (examples)

Page 31: NGA 2000 Platform Hardware

PREPARATION

5 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

5. Preparation

Please check the packing and its contents immediately upon arrival.

If any damage is observed or items are missing, then we request that you notify the forwarder to

undertake a damage survey and report the loss or damage to us immediately.

5.1 Installation Site

Be sure to observe the additional notes, safety precautions and warnings givenin the individual manuals (platform, modules, I/O boards etc.) !

The platform must not operate in explosive atmosphere without supplementaryprotective measures !

Free flow of air into and out of the platfrom (ventilation slits) must not be hinderedby nearby objects or walls !

The installation site for the Platform has to be dry and remain above freezing pointat all times. The platform must be exposed neither to direct sunlight nor to strongsources of heat.Be sure to observe the permissible ambient temperatures (c.f. Item 20: TechnicalData). For outdoor installation, we recommend to install the platform in aprotective cabinet. At least, the platorm has to be protected against rain (e.g.,shelter).

INSTALLATION SITE

Page 32: NGA 2000 Platform Hardware

PREPARATION

5 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

NET1

NET2

5.2 External Modules

Observe the safety precautions and warnings for the modules !Consider also the procedure steps with regard to handling, configuration,and operation if given in the individual manual(s) !

The platform (controller board) can be connected with external analyzer modules via the network

module LEM 01 (see Fig. 5-1). RJ 45 sockets are serving for network interconnection. Be sure

to observe the network termination (see section 1.3.5).

Note !

The external modules cannot be supplied with 24 Vdc from the platform and, therefore, have to

be externally powered (see Fig. 1-1a).

Fig. 5-1: Rear panel network module LEM 01 (sectional view of the housing)

Analyzer module

(optionally built-in)network - twin - connector

to external modules

(two RJ 45-sockets)

EXTERNAL MODULES

Page 33: NGA 2000 Platform Hardware

PREPARATION

5 - 390002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

5.3 Installation of an Analyzer Module into the Platform

Be shure to observe the safety precautions and warnings !Installation must be made by qualified service personnel only !Consider also the procedure steps with regard to handling, configuration,and operation if given in the individual manual(s) !

For the installation of analyzer module into the platform consider the following steps:

❍ Loosen the six fastening screws for the front panel (Fig. 1-2), hold the handles, and swing

the front panel to the farest right (see also section 15.1).

❍ Lift the retention pin(s), push the analyzer module in the left half of the housing until the

retension pin(s) lock(s) into the pin seats in the lower analyzer module guides.

❍ Connect the required network wiring between the analyzer module (see the individual

operation manual) and the interconnection board ICB 01 (Fig. 5-2 and 1-5).

❍ Connect the required 24 Vdc wiring between the analyzer module (see the individual

operation manual) and the interconnection board ICB 01 (platform with 5 A power supply,

Fig. 5-2 and 1-5) or the PCB of power supply (platform with 10 A power supply, Fig. 1-4b)

resp..

❍ Close the instrument (i.e., re-attach the front panel, see section 15.1).

Fig. 5-2: Printed Circuit Board ICB 01(viewed from front panel)

24 V DC to

internal analyzer mod-

ule

internal network - connection

to analyzer module

(RJ 45 socket)

INSTALLATION OF AN ANALYZER MODULE

Page 34: NGA 2000 Platform Hardware

PREPARATION

5 - 4 90002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalINSTALLATION OF PCBS / REAR MOUNTING MODULES

5.4 Installation of PCBs / Rear Mounting Modules

Insert or remove PCBs when the instrument is electrically disconnected only !Be shure to observe the safety precautions and warnings !Installation must be made by qualified service personnel only !

Consider also the procedure steps with regard to handling, configuration,and operation if given in the individual manual(s) !

5.4.1 Rear Mounting Modules

Follow the procedure steps to insert cards:

❍ Unscrew the shielding blind-plate of the chosen slot and remove it.

❍ Put the backplane module in the chosen slot, and fasten the back panel with screws.

5.4.2 Internal Slots (with 5 A power supply only)

Follow the procedure steps to insert cards:

❍ Loosen the six fastening screws of the front panel (Fig. 1-2) and swing the front panel to the

farest left (see also section 15.1).

❍ Tilt the card ejector upwards and push the card in the chosen slot until the card catch locks

into place.

❍ Close the instrument (i.e., re-attach the front panel, see section 15.1).

Page 35: NGA 2000 Platform Hardware

SWITCHING ON

6 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

13

2

-DC 24 V IN-

LFM 01

F1

/F2

T3

,5A

/ 2

30

V

POWER

230 V

115 V

50/60 Hz

~~

6. Switching On

Be shure to observe the safety precautions and warnings !Observe the additional notes, safety precautions and warnings given in theindividual manuals (modules, I/O boards) !

Once the instrument has been correctly assembled and installed in accordance with the general

instructions given in section 5., the equipment is ready for operation.

The instrument is switched on by providing the required voltage.

The instrument is specified for an operating voltage of 230 V AC (196 - 264 V AC), 47 - 63 Hz or

115 V AC (93 - 132 V AC) resp., 47 - 63 Hz and / or 24 V DC (± 10 %).

The platform must be switched on only after switching on all modules connectedto the network. Be sure to observe the network termination (see section 1.3.5)!

Fig. 6-1: Voltage Supply of NGA 2000 - Platform

AC Fuses

F1/F2: T 3,5 A)

AC - connector

(230 / 115 Vac;

50 / 60 Hz)

24 Vdc - connector

Page 36: NGA 2000 Platform Hardware

SWITCHING ON

6 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

6.1 24 Vdc - Operation

The 24 Vdc voltage is connected via a 3-pin socket (XLR), whereby filtered DC voltage has to be

available. The remaining interferences will be prevented by internal filters.

When both 24 Vdc and Vac voltage are supplied simultaneously, the AC voltage will be preferred.

The 24 Vdc then is automatically disconnected by a relay. The 24 Vdc will be used only while

cessation of the AC voltage.

❍ Connect power supply and Line-Fan-Module (LFM 01), (see Fig. 5-1, DC - connector).

Verify whether the line voltage stated on the instrument or power supply agreeswith that of your mains line!

Verify correct polarity for 24 V dc - operation !

Be sure to observe the safety precautions and warnings given bymanufacturer of power supply !

❍ Connect mains line and power supply.

6.2 230 / 115 Vac - Operation

❍ Connect mains line and Line-Fan Module (LFM 01), (see Fig. 5-1, AC - connector).

Verify beforehand that the line voltage stated on the LFM 01agrees with that of your mains line !Verify that the position of input voltage switch of the 10 A power supply agreeswith that of your mains line (Fig. 20-3 and 1-4) !

If rear panel of LFM 01 shows “230/115 VAC”, there is built-in the 5 A power supplywith autoranging of input AC voltage.If rear panel of LFM 01 shows “230 VAC” or “115 VAC”, there is built-in the 10 Apower supply with manual switch of input AC voltage..

Upon switching on, the analyzer will perform a self-diagnostic test routine.

After switching on the analyzer, the software intergrates automatically all the individual modules,

which are incorporated in the analyzer via network and have been previously switched on, in the

system.

For former informations to the displays while switching on look at the respective software manual.

Page 37: NGA 2000 Platform Hardware

SWITCHING OFF

7 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

7. Switching Off

The platform is switched off by disconnecting the voltage supply.

Be shure to observe the safety precautions and warnings !Consider also the procedure steps with regard to handling, configuration,and operation if given in the individual manual for analysis module(s)and additional PCBs !

Page 38: NGA 2000 Platform Hardware

SWITCHING OFF

7 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Page 39: NGA 2000 Platform Hardware

TROUBLESHOOTING

11 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

11. Troubleshooting

The instrument cover can be removed when necessary (see 15.).

Be sure to observe the safety precautions and warnings !Trouble-shooting, component replacement and internal adjustments must bemade by qualified service personnel only !

11.1 Instrument has no Function (LCD display is dark)

Possible Causes Check / Correct

a) External supply voltage is absent: Check electrical supply

- DC polarity is reversed Check electrical supply

b) Possible error at line-fan module: LFM 01 is not correctly connected,

for further causes see 11.3

c) Possible error at internal power supply: Causes see 11.4

d) Internal connections incorrect or absent: Check internal connections:

Check whether ACU 02 is in correct place

Check ACU 02 - AFP 01 connection cable

e) AFP 01 board or LCD display defective Exchange front panel (see 17.3)

g) EPROM / ACU 02 defective: Replace EPROM / ACU 02

(see 17.4 and 17.2)

NO FUNCTION (LCD DISPLAY)

Page 40: NGA 2000 Platform Hardware

TROUBLESHOOTING

11 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

11.2 No Measurement Screen

Possible Causes Check / Correct

Internal analyzer module

a) Internal connection incorrect: Check internal connection (24 Vdc/Network)

between ICB and analyzer module, and

replace cable if necessary.

b) Defective analyzer module: See the operating and servicing manual of

the analyzer module.

External modules

a) Switched off module: Switch on the module

b) No external power: Check external power supply

c) Defective network connection: Check network termination (see section 1.3.5).

Check network connection between platform

and external module, or replace the

connection cable or network module (platform)

if necessary.

d) Analyzer module has not been integrated Integrate the analyzer module in the

in the system software (network): system software (see software manual).

e) Defective analyzer module: See the operation manual of the analyzer

module.

NO MEASUREMENT SCREEN

Page 41: NGA 2000 Platform Hardware

TROUBLESHOOTING

11 - 390002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

11.3 Line - Fan - Module (LFM 01)

Possible Causes Check / Correct

a) Fan does not function: For DC operation:

Pull out LFM, check fuse F 3 and replace

if necessary (see section 11.5).

In case the fuse function is properly:

LFM 01 defective, exchange module

For AC operation: see section 11.3 b)

b) For AC operation: Fuses defective Check fuses F 1 and F 2 and replace

if necessary (see section 11.5).

In case the fuses function is properly:

LFM 01 defective, exchange module

11.4 Power Supply Unit

Possible Causes Check / Correct

a) External supply voltage is absent: Check electrical supply

b) Green LED does not lighten: Instrument is DC operated

- Possible error at line-fan module: LFM 01 is not correctly connected,

for further causes see 11.3

- Power supply is not correctly connected Open the housing (see section 15.)

and check correct fit

- Power supply is defective Exchange the module (see 17.2.2)

LINE - FAN - MODULE / POWER SUPPLY UNIT

Page 42: NGA 2000 Platform Hardware

TROUBLESHOOTING

11 - 4 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

11.5 Fuses

Be sure to observe the safety precautions and warnings !Always disconnect power, discharge circuits and remove external voltage beforechecking the fuses !

In case of exchanging fuses the customer has to be certain that fuses of specifiedtype and rated current are used. It is prohibited to use repaired fuses or defectivefuse holders or to short-circuit fuse carriers (fire hazard).

After visual checking, check the fuses with an ohmmeter.

If low impedance has been measured, the fuse is in order.

High impedance means, the fuse is out of order and must be replaced.

11.5.1 AC - Fuses

❍ Hinge down the fuse cover of the AC socket (see Fig. 11-1).

❍ Take out and check the fuses. Replace the fuse(s) if necessary [T 3.5 A / 250 V (5 x 20 mm)].

❍ Reclose the cover.

11.5.2 DC - Fuses

❍ Loosen the four fastening screws for the LFM 01 (see Fig. 11-1) and take out the LFM 01.

❍ unscrew the fuse located on the left rear side of the LFM (see Fig. 11-2), and check the fuse.

Replace the fuse if necessary [T 6.25 A / 250 V (5 x 20 mm)].

❍ Reinsert and secure the LFM with the four screws.

FUSES

Page 43: NGA 2000 Platform Hardware

TROUBLESHOOTING

11 - 590002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

13

2

-DC 24 V IN-

LFM 01

F1

/F2

T3

,5A

/ 23

0V

POWER

230 V

115 V

50/60 Hz

~~

AC - connector

Fastening Screws

Fig. 11-2: DC - Fuse of LFM 01

Fig. 11-1: AC - Fuses of LFM 01

DC - Fuse

F3: T 6,25 A / 250 V

FUSES

AC - Fuses

F1/F2:

T 3,5 A / 230 V

fuse cover

24 Vdc - connector

Page 44: NGA 2000 Platform Hardware

TROUBLESHOOTING

11 - 6 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Page 45: NGA 2000 Platform Hardware

OPENING THE HOUSING

15 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

15. Opening the Housing

Open the housing in case of troubleshooting, repair and service.

Be sure to observe the safety precautions and warnings !Trouble-shooting, component replacement and internal adjustments must bemade by qualified service personnel only !

15.1 Front Panel

Use the following steps to remove the front panel:

❍ Loosen the six fastening screws of the front panel (Fig. 1-2.).

❍ Swing the front panel outwards.

Due to the built-in hinge mechanism, it is possible to swing the front panel at a certain

angle.

Thus, the inserted PCBs or the analysis module are rendered accessible.

The instrument can be operated even with opened front panel which is advantageous

during servicing.

❏ Reassembling is carried out in reverse sequence.

Page 46: NGA 2000 Platform Hardware

OPENING THE HOUSING

15 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

15.2 Housing Cover

If required, the cover can be removed.

❍ Loosen the eight lateral recessed head screws on top.

❍ Pull the cover straight up.

❏ Reassembling is carried out in reverse sequence.

The cellular plastic insulation sticked on the bottom side of the cover mustalways be placed above the power supply unit !

Page 47: NGA 2000 Platform Hardware

REMOVAL / REPLACEMENT OF COMPONENTS

17 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

17. Removal / Replacement of Components

Exchange components if requested by repair or servicing

Be sure to observe the safety precautions and warnings !Component replacement and internal adjustments must be made by qualifiedservice personnel only !Consider also the procedure steps with regard to handling, configuration,and operation if given in the individual manual(s) !

17.1 Removal of Analyzer Module

Use the following steps to remove the analyzer module:

❍ Disconnect power.

❍ Loosen the six fastening screws of the front panel (Fig. 1-2) and swing the front panel to the

farest right (see also section 15.1).

❍ Loosen the gas connections if necessary.

To avoid a danger to the operators by explosive, toxic or unhealthy gascomponents, first purge the gas lines with ambient air or nitrogen (N2) beforecleaning or exchange parts of the gas paths.

In case it is necessary to open the gas paths, close the analyzersgas connections with PVC caps immediatly !

❍ Disconnect analyzer module wiring (24 Vdc and network) if necessary (see section 5.).

❍ Unlock the retention pin(s) (Fig. 1-4a) and pull out the analysis module carefully.

❏ For installation see section 5.2.

REMOVAL OF ANALYSIS MODULE

Page 48: NGA 2000 Platform Hardware

REMOVAL / REPLACEMENT OF COMPONENTS

17 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

17.2 Removal / Replacement of PCBs / Rear Mounting Modules

Always disconnect power, discharge circuits and remove external voltagesources before troubleshooting, repair or replacement of any component !

Be sure to observe the safety precautions and warnings !Component replacement and internal adjustments must be made by qualifiedservice personnel only !

17.2.1 Rear Mounting Modules

❍ Loosen fastening screws of the module of interest, press down the lever to unlock module

and remove module.

❍ Replace the module und fasten the screws.

17.2.2 Internal Slots

❍ Loosen the six fastening screws of the front panel (Fig. 1-2) and swing the front panel to the

farest left (see also section 15.1).

PCBs:

❍ Push the card ejector downwards to unlock the card of intereset and remove the card.

❍ Tilt the card ejector of the new card upwards and put in the card until the catch locks

into place.

Power Supply Unit:

❍ Push the catch downwards and remove the power supply unit by pulling at its handel

(Fig. 1-4) .

❍ Put in the new unit until the catch locks into place.

❍ Close the instrument (i.e., re-attach the front panel, see section 15.1).

REMOVAL / REPLACEMENT OF PCBS / REAR MOUNTING MODULES

Page 49: NGA 2000 Platform Hardware

REMOVAL / REPLACEMENT OF COMPONENTS

17 - 390002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

17.3 Removal / Replacement of Front Panel

Front panel plate together with LCD display, circuit board AFP 01, and hinge mechanism

constitutes an unit. Therefore, the front panel has to be disassembled completely when one

component of the unit is out of order.

Be sure to observe the safety precautions and warnings !Component replacement and internal adjustments must be made by qualifiedservice personnel only !

❍ Loosen the six fastening screws for the front panel (Fig. 1-2) and swing the front panel

outwards preferably to the right position (see also section 15.1).

❍ Remove the cable (ACU 02 - AFP 01 connection) from the circuit board ACU 02.

❍ Unscrew both fastening screws (top and bottom screw) by using a short screwdriver

for the hinge mechanism to complete remove the front panel.

❍ Re-insert swivel mechanism and complete front panel in the housing and tighten both

fastening screws (top and bottom screws) by using a short screwdriver (see Fig. 17-1).

❍ Re-attach the cable (ACU 02 - AFP 01 connection) to the circuit board ACU 02.

❍ Close the instrument (i.e., re-attach the front panel, see section 15.1).

REMOVAL / REPLACEMENT OF FRONT PANEL

Fig. 17-1: Fastening screws for the hinge mechanism (partial view)

lower fastening screw

hinge

mechanism

upper fastening screw

metal washer

lower housing strut

hinge

mechanism

metal washer

NYLON washer

upper housing strut

Page 50: NGA 2000 Platform Hardware

REMOVAL / REPLACEMENT OF COMPONENTS

17 - 4 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

17.4 Replacement of EPROM / Buffer Battery on the ACU 02

Always disconnect power, discharge circuits and remove external voltagesources before troubleshooting, repair or replacement of any component !

17.4.1 Removal of ACU 02

Use the following steps to remove the ACU 02 board:

❍ Loosen the six fastening screws of the front panel (Fig. 1-2) and swing the front panel to the

farest left (see also section 15.1).

❍ Remove the cable (ACU 02 - AFP 01 connection) from the circuit board ACU 02.

❍ Push the card ejector for the ACU 02 downwards and remove the board.

REPLACEMENT OF EPROM / BUFFER BATTERY OF ACU 02

Fig. 17-2: Controller Board ACU 02 (partial view, component side)

Battery Jumper (P 23)

EP

RO

M

D32

Connection

to LCD dispaly

+

Buffer battery

on soldering pins-

(P20

)

(P23

)

(P1)

Page 51: NGA 2000 Platform Hardware

REMOVAL / REPLACEMENT OF COMPONENTS

17 - 590002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalREPLACEMENT OF EPROM / BUFFER BATTERY OF ACU 02

17.4.2 Replacement of EPROM

The EPROM located on the ACU 02 may easily be replaced (e.g. in case of a defective board or

a new program version). Use the following procedure to replace the EPROM:

❍ Remove the ACU 02 (see 17.4.1).

❍ Remove the jumper “P 23” for battery buffer (Fig. 17-2).

All data and compensation values entered by the user will be deleted (RAM-fail) !

❍ Remove the EPROM from its socket (see Fig. 17-2).

❍ Insert the new EPROM.

Verify polarity when inserting the new EPROM !

A semicircular notch can be found both on the EPROM and the socket.

For the correct insertion of the EPROM, both notches have to be superimposed.

After the replacement:

❍ Re-insert the jumper for battery buffer (Fig. 17-2).

❍ Install the ACU 02 (see 17.4.4).

Page 52: NGA 2000 Platform Hardware

REMOVAL / REPLACEMENT OF COMPONENTS

17 - 6 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

17.4.3 Replacement of Buffer Battery

Use the following procedure to replace the battery:

❍ Remove the ACU 02 (see 17.4.1).

❍ Remove the jumper “P 23” for battery buffer (Fig. 17-2).

All data and compensation values entered by the user will be deleted (RAM-fail) !

❍ Unsolder the battery from the soldering pins (see Fig. 17-2).

❍ Solder the new battery (Ordering No. 03 765 180) to the soldering pins (see Fig. 17-2).

Verify polarity when soldering the new battery (Fig. 17-2) !

After the replacement:

❍ Re-insert the jumper for battery buffer (Fig. 17-2).

❍ Install the ACU 02 (see 17.4.4).

17.4.4 Installation of ACU 02

❍ Tilt the card ejector of the ACU 02 upwards and put in the card until the catch locks into

place.

❍ Re-attach the cable (ACU 02 - AFP 01 connection) to the circuit board ACU 02.

❍ Close the instrument (i.e., re-attach the front panel, see section 15.1).

❍ Switch on instrument.

Now all the data required by the user can be entered again, e.g. system parameters.

REPLACEMENT OF EPROM / BUFFER BATTERY OF ACU 02

Page 53: NGA 2000 Platform Hardware

CLEANING OF HOUSING OUTSIDE

19 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

19. Cleaning of Housing Outside

For cleaning housing outside, you need a soft, fluff free cloth and all purpose detergent.

❍ Disconnect all voltage supplies.

To avoid a danger to the operators by explosive, toxic or unhealthy gas

components, first purge the gas lines with ambient air or nitrogen (N2) before

cleaning or exchange parts of the gas paths.

If it is necessary to disconnect the gas connections of analysis modules, the gas

line fittings of the module have to be closed with PVC-caps before cleaning !

❍ Moisten of the soft, fluff free cloth with the cleaning solution(mixture of 3 parts water, 1 part all purpose detergent max.).

Be sure to use a moisted, but not wet, cloth only !

Be sure, that non liquid can drop into the housing inside !

❍ Cleaning of the platform housing outside with the moisted cloth.

❍ If required, rub off the housing with a dry cloth afterwards.

Page 54: NGA 2000 Platform Hardware

CLEANING OF HOUSING OUTSIDE

19 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Page 55: NGA 2000 Platform Hardware

TECHNICAL DATA

20 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

20. Technical Data

20.1 Platform

20.1.1 Housing

Certifications 1) EN 50081-1, EN 50082-2, EN 61010-1

NAMUR, CSA NRTL/C, C-Tick, PAC, IS,

GOSSTANDART

Weight approx. 10 kg (including all standard modules)

Type of housing 19" - 3 HU, rack or table top housing

Materials Aluminum

Protection Class IP 20 (according to DIN standard 40050)

Dimensions see Figs. 20-1 and 20-2

20.1.2 Environments

Permissible ambient temperature + 5 °C to + 45 °C

Solar radiation (sunshine) The housing must not directly be irradiated by

sun light or fluorescence lamps.

Altitude 0 - 1500 m (above sea level)

Humidity (not- condensing) < 90 % rel. humidity at + 20 °C

< 50 % rel. humidity at + 40 °C

Rain / Drop- and splash water The housing must not be exposed to rain or

drop-/splash water.

Explosive atmosphere The housing must not be operated in explosive

atmosphere

1) partly in preparation

PLATFORM

Page 56: NGA 2000 Platform Hardware

TECHNICAL DATA

20 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

488

537

Fig. 20-1: Housing dimensions (in [mm]) (with standard analyzer module)

PLATFORM

Page 57: NGA 2000 Platform Hardware

TECHNICAL DATA

20 - 390002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

488

615

664

537

Fig. 20-2: Housing dimensions (in [mm]) (with extended analyzer module)

PLATFORM

Page 58: NGA 2000 Platform Hardware

TECHNICAL DATA

20 - 4 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

20.2 Line - Fan - Module (LFM 01)

Electrical connections

external IEC connector with 2 fuses and RFI filter.

Audio connector (XLR) / 24 V DC;

pre-connecting shield, 3-pin

internal 230 Vac connector, pre-connecting shield

24 Vdc connector for low voltage

Nominal voltage 230 / 115 Vac, 50 / 60 Hz

Input voltage 93 - 132 Vac, 47 - 63 Hz

or

196 - 264 Vdc, 47 - 63 Hz

and / or

24 V DC ± 10 %

Ripple and noise < 100 mV pp

Fuses AC F1 / F2: T 3,5 A / 230 V (5 x 20 mm)

DC F3: T 6,25 A / 230 V (5 x 20 mm)

Power consumption maxi. 240 VA / 240 W

Outline dimensions 3 HU, 12 DU, 160 mm (length)

Weight 280 g, approx.

Housing material Aluminum, light chromated

20.2.1 Electrical Safety

Over-voltage category II

Pollution degree 2

Safety Class 1

LINE - FAN - MODULE (LFM 01)

Page 59: NGA 2000 Platform Hardware

TECHNICAL DATA

20 - 590002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

20.3 Internal Power Supply Unit

20.3.1 5 A Power Supply

Outline dimensions 8 DU - Euro card slot with handle

Weight 550 g, approx.

Nominal voltage 230 / 115 Vac, 50 / 60 Hz

Input voltages 93 - 132 Vac, 47 - 63 Hz

or

196 - 264 Vac, 47 - 63 Hz

Output voltages 24 Vdc ± 5 % / 5 A

Electrical connections Connector H 15 / DIN 41612

20.3.2 10 A Power Supply

Input terminal strips

Electrical connections to ICB 01 Connector H 15 / DIN 41612 via PCB PPA 01

Nominal voltage 230 / 115 V ac, 50 / 60 Hz

Input voltage 196–264 V ac or 93–132 V ac, 47-63 Hz

with manual switch

Input power max. 700 VA

Output terminal strips

Electrical connections to analyzer module 24 Vdc connector for low voltage via PCB PPA 01

Output voltage 24 V dc

max. 10.0 A

Output power max. 240 VA

Dimensions 125 x 122 x 103 mm (HxWxD), see Fig. 20-3

POWER SUPPLY UNIT

Page 60: NGA 2000 Platform Hardware

TECHNICAL DATA

20 - 6 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Fig. 20-3: Dimensional Sketch power Supply SL10 [mm]

Front View Side View

Manual Switch 230 / 115 Vac

Page 61: NGA 2000 Platform Hardware

PIN ASSIGNMENTS

21 - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

21. Pin Assignments

Use only from our factory optional delivered cables or equivalent shielded cables

to be in agreement with the CE - conformity.

The customer has to guarantee, that the shield is be connected bothsided.

Shield and connectors housing have to be connected conductive.

Sub.-min.-D-plugs/sockets have to be screwed to the analyzer.

21.1 24 Vdc Input (LFM 01)

Pin 1: MEPin 2: + 24 VdcPin 3: 0 V DC (⊥⊥⊥⊥⊥)shield: housing flange

21.2 230/115 Vac Input (LFM 01)

Fig. 21-1: Pin assignments 24 Vdc Input (LFM 01)

Fig. 21-3: Pin assignments 24 Vdc output to analyzer moduel (internal)(see Item 5.3, too)

VOLTAGE INPUT

Pin 1: MEPin 2: + 24 VdcPin 3: 0 V DC (⊥⊥⊥⊥⊥)shield: housing flange

21.3 24 Vdc Output to analyzer module (internal)

Fig. 21-2: Pin assignments 230/115 Vac Input (LFM 01)

21

3

(sh

ield

)

PE

2 3 1

Supply from front(see Fig. 6-1, too)

Page 62: NGA 2000 Platform Hardware

PIN ASSIGNMENTS

21 - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Page 63: NGA 2000 Platform Hardware

INDEX

R - 190002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Index

A

AC - Operation 6 - 2

10 A Power Supply Unit 6 - 2

5 A Power Supply Unit 6 - 2

ACU 02 (Controller Board) 1 - 12

Electrostatic Discharge S - 7

Installation of ACU 02 17 - 6

Removal of ACU 02 17 - 4

Replacement of Buffer Battery 17 - 6

Replacement of EPROM 17 - 5

Safety Measures S - 1

Electrostatic Discharge S - 7

Additional notes for service / maintenance S - 6

Ambient Temperature, permissible 20 - 1

Analyzer Module 1 - 1, 1 - 4, 1 - 5, 1 - 6

Installation of an Analyzer Module 5 - 3

Removal of Analyzer Module 17 - 1

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Gases and Gas Conditioning S - 3

B

Buffer Battery 17 - 4

Replacement 17 - 6

Safety Measures S - 1

C

Cleaning of platform housing outside 19 - 1

Controller Board ACU 02 1 - 12

Electrostatic Discharge S - 7

Installation of ACU 02 17 - 6

Removal of ACU 02 17 - 4

Replacement of Buffer Battery 17 - 6

Replacement of EPROM 17 - 5

Safety Measures S - 1

Electrostatic Discharge S - 7

D

DC - Operation 6 - 2

Dimensions 20 - 1

E

Electrical Safety 20 - 4

Safety Class 20 - 4

Electrostatic Discharge S - 7

Environments 20 - 1

External Modules 5 - 2

F

Front Panel 1 - 3

Removal / Replacement of 17 - 3

Safety Measures S - 1

Fuses 11 - 4. See Technical Data, too.

AC - Fuses 11 - 4

DC - Fuse 11 - 4

Troubleshooting 11 - 4

H

Housing

Controller Board ACU 02 1 - 12

Dimensions 20 - 2, 20 - 3

Front Panel 1 - 3

Inside View 1 - 5, 1 - 6

Line - Fan - Module (LFM 01) 1 - 10

Network Module (LEM 01) 1 - 13

Network Termination 1 - 13, 1 - 14

Power Supply Unit 1 - 9

10 A loading 1 - 6, 1 - 9, 1 - 11, 6 - 2

5 A loading 1 - 5, 1 - 9, 1 - 11, 6 - 2

Printed Circuit Boards (PCB)

Controller Board ACU 02 1 - 12

Interconnection Board ICB 01 1 - 7

Rear Mounting Modules

Line - Fan - Module (LFM 01) 1 - 10

Network - Module (LEM 01) 1 - 13

Rear View 1 - 4

I

Installation of an Analyzer Module 5 - 3

Safety Measures S - 1

Installation of PCBs / Rear Mounting Modules

Internal Slots 5 - 4

Rear Mounting Modules 5 - 4

Safety Measures S - 1

Installation Site 5 - 1

Environments 20 - 1

Permissible Ambient Temperature 20 - 1

Safety Measures S - 1

Instrument specific notes for the user S - 5

Interconnection Board ICB 01 1 - 7

Internal Slots 5 - 4

Installation 5 - 4

Removal / Replacement 17 - 2

Safety Measures S - 1

Page 64: NGA 2000 Platform Hardware

INDEX

R - 2 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

L

LEM 01 (Network Module) 1 - 13, 5 - 2

Network Termination 1 - 13, 1 - 14

LFM 01 (Line - Fan - Module) 1 - 10

Technical Data 20 - 4

Line - Fan - Module (LFM 01) 1 - 10

Technical Data 20 - 4

M

Maintenance

Cleaning of platform housing outside 19 - 1

Installation of an Analyzer Module 5 - 3

Internal Slots

Installation 5 - 4

Removal / Replacement 17 - 2

Opening the Housing 15 - 1

Front Panel 15 - 1

Housing Cover 15 - 2

Rear Mounting Modules

Installation 5 - 4

Removal / Replacement 17 - 2

Removal / Replacement of Components 17 - 1

Removal / Replacement of Front Panel 17 - 3

Removal of Analyzer Module 17 - 1

Removal/Replacement of PCBs/Rear Mounting Modules

Internal Slots 17 - 2

Rear Mounting Modules 17 - 2

Replacement of EPROM / Buffer Battery of ACU 02 17 - 4

Installation of ACU 02 17 - 6

Removal of ACU 02 17 - 4

Replacement of Buffer Battery 17 - 6

Replacement of EPROM 17 - 5, 17 - 6

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Electrostatic Discharge S - 7

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Instrument specific notes for the user S - 5

Supply Voltage S - 4

Switching Off 7 - 1

N

Network Module (LEM 01) 1 - 13, 5 - 2

Network Termination 1 - 13, 1 - 14

Network Termination 1 - 13

O

Opening the Housing 15 - 1

Front Panel 15 - 1

Housing Cover 15 - 2

Safety Measures S - 1

Operating Voltage 6 - 1

10 A Power Supply Unit 6 - 2

5 A Power Supply Unit 6 - 2

Operation

Preparation 5 - 1

Installation of an Analyzer Module 5 - 3

Installation Site 5 - 1

Network Termination 1 - 13, 1 - 14

Safety Measures S - 1

Switching Off 7 - 1

Switching On 6 - 1

230 / 115 Vac - Operation 6 - 2

24 Vdc - Operation 6 - 2

Operating Voltage 6 - 1

Safety Measures S - 1

P

Permissible Ambient Temperature 5 - 1, 20 - 1

Pin Assignments 21 - 1

230/115 Vac Input (LFM 01) 21 - 1

24 Vdc Input (LFM 01) 21 - 1

24 Vdc Output to analyzer module (internal) 21 - 1

Power consumption 20 - 4

Power Supply Unit 1 - 9

10 A loading 1 - 6, 1 - 9, 6 - 2

Retrofit kit 1 - 9

Technical Data 20 - 5

5 A loading 1 - 5, 1 - 9

Technical Data 20 - 5

Technical Data 20 - 5

Troubleshooting 11 - 3

Preparation 5 - 1

External Modules 5 - 2

Network Termination 1 - 13, 1 - 14

Installation of an Analyzer Module 5 - 3

Installation of PCBs / Rear Mounting Modules

Internal Slots 5 - 4

Rear Mounting Modules 5 - 4

Installation Site 5 - 1

Ambient Temperature, permissible 20 - 1

Environments 20 - 1

Network Termination 1 - 13, 1 - 14

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Instrument specific notes for the user S - 5

Supply Voltage S - 4

Printed Circuit Boards (PCB)

Controller Board ACU 02 1 - 12

Interconnection Board ICB 01 1 - 7

Protection Class 20 - 1

Page 65: NGA 2000 Platform Hardware

INDEX

R - 390002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

R

Rear Mounting Modules 1 - 4

Installation 5 - 4

Line - Fan - Module (LFM 01) 1 - 10

Network Module (LEM 01) 1 - 13

Removal / Replacement 17 - 2

Removal / Replacement of Components 17 - 1

Installation of an Analyzer Module 5 - 3

Removal / Replacement of Front Panel 17 - 3

Removal of Analyzer Module 17 - 1

Removal/Replacement of PCBs/Rear Mounting Modules

Internal Slots 17 - 2

Rear Mounting Modules 17 - 2

Replacement of EPROM / Buffer Battery of ACU 02 17 - 4

Safety Measures S - 1

Removal of Analyzer Module 17 - 1

Safety Measures S - 1

Removal/Replacement of PCBs/Rear Mounting Modules

Internal Slots

Installation 5 - 4

Removal 17 - 2

Rear Mounting Modules

Installation 5 - 4

Removal 17 - 2

Safety Measures S - 1

Replacement of Buffer Battery

Installation of ACU 02 17 - 6

Removal of ACU 02 17 - 4

Replacement 17 - 6

Safety Measures S - 1

Replacement of EPROM

Installation of ACU 02 17 - 6

Removal of ACU 02 17 - 4

Replacement 17 - 5

Safety Measures S - 1

Replacement of EPROM / Buffer Battery of ACU 02 17 - 4

Installation of ACU 02 17 - 6

Removal of ACU 02 17 - 4

Replacement of Buffer Battery 17 - 6

Replacement of EPROM 17 - 5

Safety Measures S - 1

Replacement of Front Panel 17 - 3

Safety Measures S - 1

S

Safety Class 20 - 4

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Electrostatic Discharge S - 7

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Instrument specific notes for the user S - 5

Supply Voltage S - 4

Switching Off 7 - 1

Switching On 6 - 1

230 / 115 Vac - Operation 6 - 2

24 Vdc - Operation 6 - 2

Operating Voltage 6 - 1

10 A Power Supply Unit 6 - 2

5 A Power Supply Unit 6 - 2

Pin Assignments 21 - 1

230/115 Vac Input (LFM 01) 21 - 1

24 Vdc Input (LFM 01) 21 - 1

24 Vdc Output to analyzer module (internal) 21 - 1

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Instrument specific notes for the user S - 5

Supply Voltage S - 4

T

Technical Data 20 - 1

Ambient Temperature, permissible 20 - 1

Dimensions 20 - 1

Electrical Safety 20 - 4

Safety Class 20 - 4

Environments 20 - 1

Line - Fan - Module (LFM 01) 20 - 4

Operating Voltage 6 - 1

Permissible Ambient Temperature 20 - 1

Power Supply Unit 20 - 5

10 A loading 20 - 5

5 A loading 20 - 5

Protection Class 20 - 1

Weight 20 - 1

Technical Description 1 - 1

Housing

Front Panel 1 - 3

Inside View 1 - 5, 1 - 6

Line - Fan - Modul (LFM 01) 1 - 10

Network Module (LEM 01) 1 - 13

Rear View 1 - 4

Power Supply Unit 1 - 9

10 A loading 1 - 6, 1 - 9

5 A loading 1 - 5, 1 - 9

Printed Circuit Boards (PCB)

Controller Board ACU 02 1 - 12

Interconnection Board ICB 01 1 - 7

Rear Mounting Modules

Line - Fan - Modul (LFM 01) 1 - 10

Network Module (LEM 01) 1 - 13

Termination (Network) 1 - 13, 1 - 14, 11 - 2

Page 66: NGA 2000 Platform Hardware

INDEX

R - 4 90002496(2) [NGA-Platform e] 30.03.99

Rosemount Analytical

Troubleshooting 11 - 1

Front Panel 11 - 1

Fuses 11 - 4

AC - Fuses 11 - 4

DC - Fuse 11 - 4

Internal Slots

Installation 5 - 4

Power Supply Unit 11 - 3

Removal / Replacement 17 - 2

Safety Measures S - 1

No instrument function (LCD display) 11 - 1

No measurement screen 11 - 2

Network Termination 1 - 13, 1 - 14, 11 - 2

Opening the Housing 15 - 1

Front Panel 15 - 1

Housing Cover 15 - 2

Safety Measures S - 1

Power Supply Unit 11 - 3

Safety Measures S - 1

Rear Mounting Modules

Installation 5 - 4

Removal / Replacement 17 - 2

Safety Measures S - 1

Removal / Replacement of Components 17 - 1

Safety Measures S - 1

Removal / Replacement of Front Panel 17 - 3

Removal of Analyzer Module 17 - 1

Safety Measures S - 1

Removal/Replacement of PCBs/Rear Mounting Modules

Internal Slots 17 - 2

Rear Mounting Modules 17 - 2

Safety Measures S - 1

Replacement of EPROM / Buffer Battery of ACU 02 17 - 4

Installation of ACU 02 17 - 6

Removal of ACU 02 17 - 4

Replacement of Buffer Battery 17 - 6

Replacement of EPROM 17 - 5

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Electrostatic Discharge S - 7

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Instrument specific notes for the user S - 5

Supply Voltage S - 4

Switching Off 7 - 1

V

Voltage Supply

Operating Voltage 6 - 1

10 A Power Supply Unit 6 - 2

5 A Power Supply Unit 6 - 2

Pin Assignments

230/115 Vac Input (LFM 01) 21 - 1

24 Vdc Input (LFM 01) 21 - 1

24 Vdc Output to analyzer module (internal) 21 - 1

W

Weight 20 - 1

Page 67: NGA 2000 Platform Hardware

INDEX

R - 590002496(2) [NGA-Platform e] 30.03.99

Rosemount AnalyticalLIST OF FIGURES

List of Figures

Fig. Title Page

Fig. 1-1: Block diagram platform design (external) 1 - 1Fig. 1-1a: Analyzer system via platform (schematic diagram) 1 - 2

Fig. 1-2: Housing, Front View (Front panel) 1 - 3Fig. 1-3: Housing, Rear View 1 - 4

Fig. 1-4a: Internal view of housing (with 5 A power supply) 1 - 5Fig. 1-4b: Internal view of housing (with 10 A power supply) 1 - 6

Fig. 1-5: Printed Circuit Board ICB 01(viewed from front panel) 1 - 7Fig. 1-6a: Block diagram internal platform design (with 5 A power supply) 1 - 8Fig. 1-6b: Block diagram internal platform design (with 10 A power supply) 1 - 8

Fig. 1-7: Pin assignment, 64-pin ICB-bus connector 1 - 9Fig. 1-8: Rear panel LFM 01 (sectional view of the housing) 1 - 10Fig. 1-9: Rear panel network module LEM 01 (sectional view of the housing) 1 - 13

Fig. 1-10: RJ 45 network termination connector 1 - 13Fig. 1-11: Network termination (examples) 1 - 14

Fig. 5-1: Rear panel network module LEM 01 (sectional view of the housing) 5 - 2Fig. 5-2: Printed Circuit Board ICB 01 (viewed from front panel) 5 - 3

Fig. 6-1: Voltage Supply of NGA 2000 - Platform 6 - 1

Fig. 11-1: AC - Fuses of LFM 01 11 - 5Fig. 11-2: DC - Fuse of LFM 01 11 - 5

Fig. 17-1: Fastening screws for the hinge mechanism (partial view )17 - 3Fig. 17-2: Controller Board ACU 02 (partial view, component side) 17 - 4

Fig. 20-1: Housing dimensions (in [mm]) (with standard analyzer module) 20 - 2Fig. 20-2: Housing dimensions (in [mm]) (with extended analyzer module) 20 - 3Fig. 20-3: Dimensional Sketch power Supply SL10 [mm] 20 - 6

Fig. 21-1: Pin assignments 24 Vdc Input (LFM 01) 21 - 1Fig. 21-2: Pin assignments 230/115 Vac Input (LFM 01) 21 - 1Fig. 21-3: Pin assignments 24 Vdc output to analyzer moduel (internal) (see Item 5.3, too) 21 - 1

Page 68: NGA 2000 Platform Hardware

NGA 2000 PlatformInstruction Manual

9000249604/2008

APPENDIX Declarations of Conformity

A-1 NGA 2000 P

A-2 NGA 2000 PE

Page 69: NGA 2000 Platform Hardware
Page 70: NGA 2000 Platform Hardware
Page 71: NGA 2000 Platform Hardware

Instruction Manual9000249604/2008 NGA 2000 Platform

Page 72: NGA 2000 Platform Hardware

Instruction Manual9000249604/2008 NGA 2000 Platform

© Emerson Process Management GmbH & Co. OHG 2007

WORLD HEADQUARTERSROSEMOUNT ANALYTICAL EUROPEEmerson Process ManagementGmbH & Co. OHGIndustriestrasse 163594 HasselrothGermanyT 49 6055 884 0F 49 6055 884209

Emerson Process ManagementRosemount Analytical Inc.6565 P Davis Industrial ParkwaySolon, OH 44139 USAT 440.914.1261Toll Free in US and Canada 800.433.6076F 440.914.1271e-mail: [email protected]

www.raihome.com

GAS CHROMATOGRAPHY CENTERAND LATIN AMERICAEmerson Process ManagementRosemount Analytical Inc.11100 Brittmoore Park Drive

Houston, TX 77041

T 713 467 6000

F 713 827 3329

EUROPE, MIDDLE EAST AND AFRICAEmerson Process ManagementShared Services LimitedHeath PlaceBognor RegisWest Sussex PO22 9SHEnglandT 44 1243 863121F 44 1243 845354

ASIA-PACIFICEmerson Process ManagementAsia Pacific Private Limited1 Pandan CrescentSingapore 128461Republic of SingaporeT 65 6 777 8211F 65 6 777 0947e-mail: [email protected]