New Devices & Techniques In Carotid Artery Stenting: Clinically Meaningful ? Dr Sumaira Macdonald, MBChB (Comm.), FRCP, FRCR, PhD, EBIR Consultant Vascular Radiologist & Honorary Clinical Senior Lecturer, Freeman Hospital, Newcastle, UK TCT 2013

Transcript of New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS...

Page 1: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

New Devices & Techniques In Carotid Artery Stenting:

Clinically Meaningful ?

Dr Sumaira Macdonald, MBChB (Comm.), FRCP, FRCR, PhD, EBIR

Consultant Vascular Radiologist

& Honorary Clinical Senior Lecturer,

Freeman Hospital,

Newcastle,

UK

TCT 2013

Page 2: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Disclosures: Research / Educational Grants & / or consultancy:

Abbott Vascular

CR Bard

Ev3/Covidien

Pyramed

WL Gore

Medtronic / Invatec

Biotronik

Cordis (J & J)

St Jude/AGA

Spectranetics

Tryton Medical

Silk Road Medical

Terumo

Merit Medical

Volcano

COOK

Vascular Perspectives

Bridgepoint / EPS vascular

Page 3: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Prevailing Limitations Of CAS:

• Excess minor stroke risk cf. CEA (may be age dependent)

• Excess DWMRI lesions of brain cf. CEA (may be age dependent)

• Anatomic constraints (largely “ access ” related & ? influenced by operator experience)

• “ Day Zero ” Strokes for CAS cf. CEA

• Real world outcomes

• “ Off table ” to 30-day strokes

Page 4: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

“ Real World ” CAS Outcomes:

Page 5: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Nallamothu BK et al JAMA 2011;28:1338-1343

N = 24,701

Page 6: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Excess Minor Stroke:

Page 7: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

CREST: Transfemoral

Filter-Protected CAS

Major Unmet Needs CEA CAS p

CREST Peri-procedural Stroke1 2.3% 4.1% 0.01

N Engl J Med 2010;363:11-23;

Page 8: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Carotid Stenting Trialists’ Collaboration

Lancet 2010; 376; 1062-1073

Page 9: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Day Zero Strokes:

The Timing of Strokes &

Their Proposed Aetiology:

Page 10: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Fairman R et al Annals of Surgery 2007; 246: 551-558

N = 3,500

Page 11: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

“ Day Zero ” Strokes In CREST:

** p value for CREST only

0

5

10

15

20

25

30

35

Day 0 Day 1-7 Day 8-30

CAS

CEA

P = 0.0008**

P = 0.44 P = 0.26

11 Brooks et al. Circulation 2012

Page 12: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

ICSS: Timing of first stroke after allocated procedure

~ 40% of “ Day 0 ” events

occurred during prediltation / stenting / post-dilatation

Courtesy of Martin Brown & Roland Featherstone, ICSS

Page 13: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid
Page 14: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

The Arch Is A Hostile Territory:

N = 42

Procedure N Incidence MES Procedural Stage

CEA 15 15.3 (+/- 22) Post procedure

Filter protected CAS 20 319.3 (+/- 110.3) During protection

Flow reversal CAS 7 184.2 (+/- 110.5) Pre protection

CEA vs filter p = 0.001 CEA vs flow reversal p = 0.007 Flow reversal vs filter p = 0.053

Gupta N et al. JVS. 2011;53:316-322

Page 15: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Non-target territory embolization implies embolic burden of the arch & great vessel origins

from a transfemoral approach with a 9F sheath

JVIR; 2013;24:528-533

Page 16: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Anatomic Constraints:

Page 17: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Complex CAS Anatomy: Dictated By Arch Type:

Green Amber Red

Left ICA score < 4.9 Left ICA score 5.0-5.9 Right ICA score > 7.0

Macdonald S et al. Stroke 2009; 40: 1698-703

Page 18: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

P<0.0001 P<0.0001

P<0.0001

Mann-Whintey U

Kruskall-Wallis P<0.005

P<0.0001

P<0.0001

P=0.0020

Mann-Whitney U

Kruskall-Wallis P<0.005

Total Procedure Time (mins)

Fluoroscopy Time (mins)

Red case: 10/20 >15 minutes to cannulate CCA Amber case: 2/20 Green case: 0/20

Willaert W et al. J Vasc Surg 2012; 56(6): 1763-70

Page 19: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

≤ 30-Day Strokes:

The Timing of Strokes &

Their Proposed Aetiology:

Page 20: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Delayed Stroke & Death At 1-30 Days Especially with Open Cell Stents

Bosiers et al. Eur J Vasc Endovasc Surg 2007;33:135

2/3 of events delayed

Page 21: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Increased Neurologic Events With Open Cell Stents

SPACE Trial

Jansen O et al. Stroke 2009;40:841-846

Page 22: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Excess DWMRI Lesions:

Page 23: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Meta-Analysis Comparing DWMRI Lesions After CEA & CAS

Page 24: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

ICSS Substudy: N = 231

62 of 124 (50%) transfemoral distal filter CAS

18 of 107 (17%) CEA

New white lesions on DWI

(OR 5.21, 2.78-9.79; p < 0.0001)

ICSS Primary Analysis CEA Vs. CAS in 1713 symptomatic

patients

Lancet Neurol. 2010 Apr;9(4):353-62

Page 25: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

ICSS Substudy: N = 231

Recurrent stroke OR TIA (5 year cumulative)

CAS:

DWMRI +ve: 12/62

DWMRI -ve: 6/62

22.8% vs. 8.8% (p=0.04) HR 2.85 (1.05-7.720

Bonati L et al. European Stroke Congress May 2013

Page 26: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Association of DWMRI Lesion Volumes & Neurological Events

Page 27: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Schnaudigel et al. Stroke 2008;39:911

Closed

Open

Page 28: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

The potential of new technology to

solve the remaining

limitations of CAS

Page 29: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

MICHI™ Neuroprotection System

Flow Controller

Large Bore Arteriovenous Shunt Circuit

Transcervical Arterial Sheath

(8F) Venous Return Sheath (8F)

Page 30: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Baseline scan within 72 hours

Post-procedure scan within 12-48 hours

Submitted to core laboratory for blinded evaluation by two independent neuroradiologists

Parameter Value (n=56)

Subjects with new DW-MRI lesion(s) 11 (19.6%)

PROOF: FIRST IN MAN DWI SUBSTUDY

Page 31: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Prospective DWMRI outcomes

for various carotid interventional

regimes:

Page 32: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Study Procedure Embolic Protection

# subjects % w/ New DWI Lesions

ICSS1 Transfemoral CAS Distal filter (various)

51 73

ICSS1 CEA Clamp,

backbleed

107 17

PROFI2 Transfemoral

CAS

Distal filter

(Embosheild)

31 87

Leal5 Transfemoral Distal Filter

(FilterWire)

33 33

PROFI2 Transfemoral

CAS

Proximal occlusion

(MoMA)

31 45

DESERVE4 Transfemoral

CAS

Proximal

Occlusion

(MoMa)

127 30

PROOF3 Transervical

CAS

High flow rate flow reversal

48 16.7

Leal5 Transervical

CAS

Flow Reversal 31 12.9

1 Lancet Neurol. 2010 Apr;9(4):353-62 2. J Am Coll Cardiol. 2012;59:1383-1389 3. JVS 2011;54:1317-1323

4. Rubino P, EuroPCR 2011 5. JVS 2012;56:1585-1590

Page 33: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Prevailing Limitations Of CAS:

• Excess minor stroke risk cf. CEA (may be age dependent)

• Excess DWMRI lesions of brain cf. CEA (may be age dependent)

• Anatomic constraints (largely “ access ” related & ? influenced by operator experience)

• “ Day Zero ” Strokes for CAS cf. CEA

• Real world outcomes

• “ Off table ” to 30-day strokes

Page 34: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

GORE® Carotid Stent The Next Generation

Open cell nitinol frame

Closed cell 500 µ lattice on outside of frame

Permanently bound CBAS heparin on all device surfaces

*CAUTION: Investigational Device. Limited by United States Law

to Investigational Use only.

Page 35: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Hypotube (2.4 Fr)

Handle

Hub

Stent/Sheath (5Fr or 6Fr)

RO Markers

Carotid Stent Delivery System Attributes •Single handed delivery

•5Fr Introducer Sheath Compatible (White Tip)

•6Fr Introducer Sheath Compatible (Gray Tip)

•Hypotube Design

– Allows for complete closure of hemostatic valve

•135 cm Working Length

•30 cm Rx

*CAUTION: Investigational Device. Limited by United States Law to Investigational

use only.

Page 36: New Devices & Techniques In Carotid Artery Stenting...Filter-Protected CAS Major Unmet Needs CEA CAS p CREST Peri-procedural Stroke1 2.3% 4.1% 0.01 N Engl J Med 2010;363:11-23; Carotid

Prevailing Limitations Of CAS:

• Excess minor stroke risk cf. CEA (may be age dependent)

• Excess DWMRI lesions of brain cf. CEA (may be age dependent)

• Anatomic constraints (largely “ access ” related & ? influenced by operator experience)

• “ Day Zero ” Strokes for CAS cf. CEA

• Real world outcomes

• “ Off table ” to 30-day strokes