NEW BIOSTAT Cplus internal presentation -...
Transcript of NEW BIOSTAT Cplus internal presentation -...
![Page 1: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/1.jpg)
BIOSTAT® Cplus
The SIP Stainless Steel Fermenter|Bioreactor for Your Lab
Internal Presentation
![Page 2: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/2.jpg)
Main Changes New BIOSTAT® CplusProduct Changes
- DFC controller replaced by DCU control platform
- up to 4 MFC for Additive Flow gassing strategy with up to 4 integrated pumps
- VP sensor connectors for pH and DO
- Optical DO sensor (Visiferm) available
- M12 connectors for external inputs, external feed pump connection etc.
- Substrate software controller not part of standard software function
- Software feature “unused integrated pump can be configured as substrate pump”not available
- Gas Flow Ration software controller not available -> configure “Advanced DO controller” instead
- More software features: Advanced DO Controller, Culture Vessel Pressure Hold Sequence and Gravimetric Feed Control available as option
Changes
![Page 3: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/3.jpg)
Main Changes New BIOSTAT® CplusProduct Changes
Discontinued:
– 2L working volume culture vessel
– “Exclusive Flow” gassing strategy
– Bubble-free aeration
Changes
![Page 4: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/4.jpg)
Main Changes New BIOSTAT® CplusConfigurator Changes
- Configurator structure harmonized with BIOSTAT® B and D-DCU
- Customer get what is configured in the configuratore.g. Substrate controller no longer part of standard software scope
- Steam and condensate tubing's are not listed as separate item in the quote -> Now included under “Sterilization” e.g. “manual sterilization sampling valve”
- Only one MO entry, selection in the “pO2 control” “Gassing strategy”-> O2-Enrichment or Gasflow Ratio
- New entry Dual Use. Additive Flow gassing strategy with free configurable gassing rates and vessel accessories
- Qualification now in the configurator
New Basic Configurations (BC) not include sampling valve, high foam kit, casters, tray for storage bottles, motor holder and foam measurement and control
Changes
![Page 5: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/5.jpg)
Page 2
Contents
1
2
A Profile of Sartorius
BIOSTAT® Cplus3
Bioprocess Solutions Division
Competition Overview4
![Page 6: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/6.jpg)
Sartorius Runs Three Divisions
Sartorius Group
Industrial WeighingLab Products & ServicesBioprocess Solutions
A Profile of Sartorius
![Page 7: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/7.jpg)
Global Presence
30.06.2012 Page 7Sales Production
A Profile of Sartorius
![Page 8: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/8.jpg)
Page 9
Contents
1
2
A Profile of Sartorius
BIOSTAT® Cplus3
Bioprocess Solutions Division
Competition Overview4
![Page 9: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/9.jpg)
Page 10
Our Portfolio
Bioprocess Solutions
Filtration
FluidManagement
Fermentation
Purification
IntegratedSolutions
Service
![Page 10: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/10.jpg)
The Widest Product Portfolio Along Biopharma’s Process Chain
Bioprocess Solutions
![Page 11: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/11.jpg)
Leading Worldwide Market Positions
Filtration
amongTOP 3Fermentation
#1 Fluid Management
#1
Bioprocess Solutions
![Page 12: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/12.jpg)
Our Solutions for Your Process...
... in Upstream Processing (fermentation, cell cultivation, media and buffer preparation)
Bioprocess Solutions
![Page 13: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/13.jpg)
Our Solutions for Your Process...
... in Downstream Processing (depth filtration, crossflow filtration, virus filtration, freeze & thaw)
Bioprocess Solutions
![Page 14: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/14.jpg)
Our Solutions for Your Process...
... in Engineering
Bioprocess Solutions
![Page 15: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/15.jpg)
Page 20
Contents
1
2
A Profile of Sartorius
BIOSTAT® Cplus3
Bioprocess Solutions Division
Competition Overview4
![Page 16: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/16.jpg)
BIOSTAT® Cplus - SIP stainless steel fermenter|bioreactor
Fits in your labEconomic integration in existing infrastructure. Choice of steam or electrical heating for operation and culture vessel sterilization
Advanced process control and intuitive operation DCU control platform for Advanced DO Control, Gravimetric feeding and touch screen operation
Compact and movable Minimum lab space requirement and casters for easy relocation
Excellent safety Automatic sequences for culture vessel sterilization and pressure hold test and a maintenance free agitation motor
Introduction
![Page 17: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/17.jpg)
The BIOSTAT® C Family Tree
1992-2008 2008-08/2013 from August 2013BIOSTAT® C BIOSTAT® Cplus New BIOSTAT® Cplus
- Largest installation base worldwide with reference applications for many processes
- Very experienced service and application specialists for local support
- Trainings and seminars for cell culture and fermentation
- NEW with DCU control platform and new additional features
BIOSTAT® Cplus
![Page 18: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/18.jpg)
Product feature BIOSTAT® Cplus New BIOSTAT® Cplus
Operation/ Display 10” Touch screen 10” Touch screen
Integrated amplifiers max. 7 Up to 12
Interface for optical DO sensor No Yes
Sensor connectors pH | DO DIN | Lemo V8 | V8
Gassing system Integrated, MO: O2-Enrichment, Gasflow Ratio CC: Exclusive Flow, Additive Flow
IntegratedMO: O2-Enrichment, Gasflow Ratio CC: Additive Flow
Mass Flow Controller - max. 2 MFC (dedicated position) - Up to 4 MFC per Vessel (freely selectable)
Pumps Integrated: up 4 (ACID, BASE, FOAM, LEVEL) 2 external substrate pumps
Integrated: up to 4, Total: up to 6 4 x digital (ACID, BASE, FOAM, LEVEL) or2 x digital + 2 x analogue (Feed) and 2 external substrate pumps (shared output w/ internal pumps)
Vessel design 5L – 30 L (2L discontinued)
System qualification yes yes
Gravimetric Feed Control Via BioPAT® MFCS/win only Up to 2
Culture vessel pressure hold sequence
No Yes (Option)
Advanced pO2 Control No Yes (Option)
BIOSTAT® Cplus vs. NEW BIOSTAT® Cplus at a glance...
BIOSTAT® Cplus
![Page 19: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/19.jpg)
Gassing system
BIOSTAT® Cplus
Intuitive Touch Screen
Motor holder
Casters
Resterilizable bottom seat valve
High foam sensor
Sampling bottle
Resterilizable sampling valve
Culture vessel
25 mm port for resterilizabale addition valve
Agitation motor
Up to 4 integrated pump
Fluid buffer system for double mechanical seal
Removable bottle tray
Temperature control systemChoice of electrical or steam heating
Exhaust filter
Supply unit
Control unit
![Page 20: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/20.jpg)
BIOSTAT® Cplus
Lift lifting crane
Motor holder
Casters
Accessories for easy handling
![Page 21: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/21.jpg)
Safety shut down of aeration and agitation OOHigh-Foam
Control of internal unused or additional optional peristaltic pump (internal or external)
O-Substrate A & B feeding
For integration of external device, i.e.BioPAT® Xgas-X3 x External Signal
Manual or automaticOOPressure
Maintenance free, low noise XXAgitation speed
OOVessel weight
Weight indication of feeding bottleOOGravimetric feed control
O
-
X
X
X
X
Control
For Foam or Level control; user selectionXFoam|Level
Free selectable OMFC 1 ...4
MO: Controlled via agitation speed; O2 supplementation; Aeration rate; Substrate CC: Controlled via O2/ N2 supplementation, agitation speed, Aeration rate, Substrate
XpO2
Controlled via CO2/ Acid + BaseXpH
O
X
Measurement
Redox and Turbidity
Closed loop pressurized water systemChoice of electrical or steam heating
Temperature
Comments
Overview Measurement and Control capabilities
BIOSTAT® Cplus
![Page 22: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/22.jpg)
DCU control software
BIOSTAT® Cplus
Process time
Process time
Weight measurement & Gravimetric Feed control
Pumps
Process parameter
Main menus Shut down & Status
![Page 23: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/23.jpg)
DO Control – CASCADE (standard)
− 4 stage, freely configurable
− Setting of min. / max. for each cascade
− PID parameters freely adjustable
− Next cascade is activated once prior one has reached its max.
− e.g.
C1: Nitrogen Flow
C2: Air Flow
C3: O2 Flow
C4: Stirrer Speed
BIOSTAT® Cplus
![Page 24: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/24.jpg)
Advanced DO Control (option)
− Parallel activation of multiple controllers
− Max. flexibility for advanced process development
− Based on DO controller output up to five parameters can be activated at the same time
− e.g.
Increase O2 flow & decrease AIR flow while total flow is kept constant
Increase flow rate & stirrer speed at the same time
BIOSTAT® Cplus
![Page 25: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/25.jpg)
Culture vessel pressure hold sequence (option)
− Automatic pressure hold test for the culture vessel
− Adjustable time, test and process pressure
− Monitoring of the pressure characteristics on the Trend display or SCADA system
− Requires pressure measurement and automatic pressure control valve
BIOSTAT® Cplus
![Page 26: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/26.jpg)
“O2-Enrichment”
– Air flow gassing system with O2 Supplementation capability via 3/2-way solenoid valve
– Rotameter for Total Sparger flow
– O2 Supplementation automaticallycontrolled via pO2 controller
– Optional Mass Flow Controller for Total Sparger Flow
Control unitGassing system “O2-Enrichment”
BIOSTAT® Cplus
![Page 27: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/27.jpg)
“Gas Flow Ratio”
- Air flow gassing system with O2 Supplementation capability via mass flow controller for Air and O2
- Rotameter for Total Sparger flow
- O2 Supplementation automaticallycontrolled via pO2 controller
Control unit Gassing system “Gas Flow Ratio”
BIOSTAT® Cplus
![Page 28: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/28.jpg)
“Additive Flow”
- 4-Gasmixing system gasmixing of Air, N2, O2 and CO2
- Individual Rotameter
- Overlay flow and Sparger aerationOverlay: Air only Sparger: Air, O2, N2 and CO2 automatically controlled via pH/ pO2 controller
- up to 4 optional Mass Flow Controller for
Control unit Gassing System “Additive Flow”
BIOSTAT® Cplus
![Page 29: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/29.jpg)
Agitation motor
Agitation motor
• Brushless DC Servomotor
• Gear and maintenance free
• Motor performance : 2L & 5L 500 W 20-1500 rpm : 10 & 15L 800 W 20-1000 rpm : 20 & 30 L 1200 W 20-600 rpm
BIOSTAT® Cplus
![Page 30: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/30.jpg)
Supply Unit 2L & 5L, Benchtop
Open frame design with...• Frame for culture vessel• Console for control unit • Thermostat system
: Closed-loop pressurized water system with circulation pumpand heat exchanger for cooling
: Heating Alternatively via stainless steel heat exchanger (cupper soldered) orelectrical heating cartridge
• Temperature control range : 8 °C above cooling water up to 90 °C
• Sterilization temperature : up to 130 °C
Supply Unit
BIOSTAT® Cplus
![Page 31: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/31.jpg)
Supply Unit 10-30L, Floor standing
Open frame design with...
• Frame for culture vessel
• Console for control unit
• Thermostat system: Closed-loop pressurized water system with circulation pump
and heat exchanger for cooling: HeatingAlternatively via stainless steel heat exchanger (cupper soldered) orelectrical heating cartridge
: OptionalStainless steel heat exchanger (stainless steel welded) for use with clean steam
• Temperature control range : 8 °C above cooling water up to 90 °C
• Sterilization temperature : up to 130 °C
Supply Unit
BIOSTAT® Cplus
![Page 32: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/32.jpg)
- Storage bottle support (Floor stand version only)
- Storage bottles 500 ml or 1000 ml
- Motor holder
- Installation kit
- NTP connection kit
- Casters (Floor stand version only)
- Recirculation cooler FRIGOMIX® 2000 (230 VAC only)
and installation kit
- Top plate lifting device (manual operated) (Floor stand version only)
Supply unit accessories
BIOSTAT® Cplus
![Page 33: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/33.jpg)
Culture vessel 5L
- Vessel with top agitation
- Working volume 5L
- Aspect ratio H : D (total): 2:1
- Jacketed stainless steel vessel with torospherical bottom and upper glass cylinder
- Vessel fabrication: PED
- Material: Product side: stainless steel 1.4435 (AISI 316L) ,
EPDM, borosilicate glass
others: 1.4301 (AISI 304)
: Surface: internal: Ra <= 0,5, electropolished
external: glass beads blasted, electropolished
Culture vessel
BIOSTAT® Cplus
![Page 34: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/34.jpg)
Culture vessel 10L - 30 L
- Vessel with top agitation
- Working volume 10, 15, 20, 30 L
- Aspect ratio H : D (total): 2:1 or 3:1
- Jacketed stainless steel vessel with torospherical bottom
- Longitudinal viewing window
- Vessel fabrication: PED or ASME
- Material: Product side: stainless steel 1.4435 (AISI 316L) ,
EPDM, borosilicate glass
others: 1.4301 (AISI 304)
: Surface: internal: Ra <= 0,5, electropolished
external: glass beads blasted, electropolished
Culture vessel
BIOSTAT® Cplus
![Page 35: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/35.jpg)
Culture vessel Volume overview
429 – 30 | 6 – 3030-2 | 30-3
307,5 – 20 | 5,5 – 2020-2 | 20-3
225,5 – 15 | 5.0 - 1515-2 | 15-3
154,5 – 10 | 3,5 - 1010-2 | 10-3
6.81.6 - 55-2
Total Volume [L]
Working Volume[L]
Vessel description
BIOSTAT® Cplus
![Page 36: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/36.jpg)
6554 | 54Top plate ports 19 mm
11111Bottom
55555Lower side ports (25 mm port) incl. 1 x 12 mm port for Pt100
3333-Upper side ports (25 mm port)ASME vessel addition port for bursting disc
Sparger inlet, Exhaust cooler, agitation, sight glass for illumination (10-30 L), safety valve (PED vessel only)
Top plate port (fixed installations)
30 L20 L15 L10 L
2:1 | 3:15 LDescription Vessel
Culture vessel Port overview
BIOSTAT® Cplus
![Page 37: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/37.jpg)
Culture vessel equipment
• 4 x baffle, removable
• Exhaust cooler
• Light glass Ø 25 mm in lid, with removable lamp (10l – 30 L only)
• Safety valve, ASME vessel with bursting disc
• Temperature sensor Pt 100
• Combined harvesting/sampling valve
BIOSTAT® Cplus
![Page 38: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/38.jpg)
Agitation shaft sealing
Single mechanical seal
• Lubricated by condensate/aerosol in head space
• Sealing material: ceramic/carbon
Double mechanical seal
• Lubricated by pressurized condensate
• Sealing material: ceramic/carbon
• Condensate vessel with viewing window
• High containment class
Culture vessel accessories
BIOSTAT® Cplus
![Page 39: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/39.jpg)
Vessel accessories: Impeller for microbial application
6-blade disk impeller
The 6-blade disk impeller produces a flow which moves radially away from the shaft and has a high energy dissipation density. This impeller breaks up air bubbles effectively and thereby improves oxygen transport in the liquid phase. At the same time, high local shear forces are also generated which can have a negative influence on the growth of shear-stress-sensitive organisms (cell cultures, filament-producing fungi).
BIOSTAT® Cplus
![Page 40: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/40.jpg)
Vessel accessories: Impeller for cell culture application
3-blade segment impeller
3-blade segment impeller
3-blade segment impellers are specifically designed for use in culturing of cells sensitive to shear stresses and especially for cell cultures immobilized in suspension or on microcarriers. The impeller creates an axial direction of flow and thereby achieves homogenous and thorough mixing of the culture medium while minimizing shearing forces.
BIOSTAT® Cplus
![Page 41: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/41.jpg)
Vessel accessories: Aeration devices direct aeration
Microbial applications
• Ring sparger for 19 mm top plate port
Cell culture applications
• Micro sparger for 19 mm top plate port
• Sintered stainless steel frit 20 μmRing sparger μ-sparger
BIOSTAT® Cplus
![Page 42: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/42.jpg)
30.06.2012 Page 51
DO Sensor Hamilton (standard) or Mettler (optional)Clark or Optical electrode
pH SensorHamilton (standard) or Mettler (optional)
Foam & Level sensorConductive probes
Temperature sensorPt100 sensor with sleeve
Sensors & Connections
BIOSTAT® Cplus
![Page 43: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/43.jpg)
30.06.2012 Page 52
Redox sensor Combined pH & Redox sensor Hamilton (standard) or Mettler (optional)
Turbidity (cell density) sensor On-line measurement of cell growth and biomassReduction of sampling frequency and bench analysesAbsorption based
Sensors & Connections
BIOSTAT® Cplus
![Page 44: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/44.jpg)
Vessel accessories: Addition devices
Needle assembly for 19 mm portProduct description
• For installation 19 mm top plate port
• Available as : 1-Channel assembly with 4 x 1 mm needle for tubing 1.6 x 1.6 mm with sterile sleeve
: 1-Channel assembly with 6 x 0.5 mm needlefor tubing 3.2 x 1.6 mm with sterile sleeve
: 3-Channel assembly with 8 x 1 mm needlefor tubing 1.6 x 1.6 mm with sterile sleeve
Operation:
1. Sterilization in an autoclave together with the addition bottles
2. Sterilization of the culture vessel with septum/ blind plug assembly
3. Remove septum bind plug and tne sterile sleeve of the needle and pierce the septum with the needle
BIOSTAT® Cplus
![Page 45: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/45.jpg)
Vessel accessories: Addition devices
Safety Containment Valve (Sacova)
Product description
• Sterile valve for a 19 mm top plate port
• 1-Channel or 3-Channel assembly available
• For tubing 3,2 x 1,6 mm
• High containment class
Operation:
1. Sterilization in an autoclave together with the addition bottle or connected C-flex tubing using a BioWelder for sterile connection
2. After autoclaving installation in the culture vessel and sterilization together with the culture vessel
Sacova valve open
Sacova valve closed
BIOSTAT® Cplus
![Page 46: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/46.jpg)
Vessel accessories: Addition devices
Steam line APC 19
Steam line
Condensate Line
APC 25
Condensateline
Resterilizable Addition Valve (AP19 & APC25)
Product description
• Resterilizable valve for 19 or 25 mm port
• For multiple sterile connections during the process
• For tubing 3,2 x 1,6 mm
• High containment class
Operation:
1. APC19 & APC25 are installed in culture vessel
2. Sterlization together with culture vessel
3. Sterilization by autoclave of addition bottles with valve assembly
4. After autoclaving connection of valve assembly to culture vessel and sterlization by steaming of connection line
autoclave valveassembly
autoclave valveassembly
BIOSTAT® Cplus
![Page 47: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/47.jpg)
Vessel accessories: Sampling
Resterilizable Sampling Valve (SVC 25)
Product description
• Resterilizable sampling valve 25 mm port
• For multiple sterile samplings during the process
• Incl. Sterile sleeve
• Optional: Sampling bottle with valve assemblyStainless steel kill tank for contaminated condensate
Steam line
Sampling valveSVC 25
Condensate line
Sampling bottlewith valve assembly(optional)
BIOSTAT® Cplus
![Page 48: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/48.jpg)
Vessel accessories: Sampling Schematic overview SVC 25 with kill tank
BIOSTAT® Cplus
![Page 49: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/49.jpg)
The perfusion system with internal spin filter is needed for the continuous medium exchange at long-term cultivation. Internal Spinfilter are mounted on the agiation shaft. Our recommendation for mesh sizes are as follow:
Mesh size 20 μm: Suspension cells
Mesh size 40 μm: aggregate-forming cells
Mesh size 75 μm: Micro carrier cultivations
Vessel accessoriesPerfusion via Spinfilter
BIOSTAT® Cplus
![Page 50: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/50.jpg)
Vessel accessories: Perfusion via SpinfilterSchematic Drawing of a Cell Culture Bioreactor for Perfusion of Medium
BIOSTAT® Cplus
![Page 51: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/51.jpg)
Vessel accessories: Miscellaneous
- Vessel pressure gauge; -1/+3 bar
- Empty vessel sterilization
- Blind plug for 19 mm Blind plug 25 mm port
- Kill tank
- Sampling bottle for SVC 25
- Autoclavable storage bottles 0.5 – 50 L
BIOSTAT® Cplus
![Page 52: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/52.jpg)
30.06.2012 Page 37
BioPAT® MFCS | win – SCADA Software
Besides the core functionality of a real SCADA system BioPAT® MFCS/win offers a number of additional modules for adaptation to individual requirements:
- Bioprocess optimization with Design of Experiments module (DoE)
– S88 recipe control
– Batch control compliant with ANSI | ISA-88.01
– 21 CFR, Part 11 compatibility
– Secure integration into company networks
– Flexible device connection facility through OPC
– Expanded data logging and data archiving
– Logging of alarms and transmission to multimedia devices
SCADA Software
![Page 53: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/53.jpg)
30.06.2012 Page 61
Our Services
Installation
FAT & SAT
IQ | OQ support & documentation
Calibration services
Spare part sales
Preventive maintenance contracts
Customer-specific maintenance contracts
Instrument repairs
Technical & application support
Practical & technical operator training
Service
![Page 54: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/54.jpg)
Sartorius bioreactors are built to last for a decade (& more)-> You will enjoy many years of high performance
We set high standards to sell excellent - quality products-> You will receive only quality and innovative technologies with your Sartorius product
We are represented worldwide -> You‘ll always have a Sartorius contact to talk to directly in your country
Application Specialists–> Are available globally to help you with your processes and eliminate any obstacles you encounter
Value Products for Value Customers
Values
![Page 55: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/55.jpg)
Page 20
Contents
1
2
A Profile of Sartorius
BIOSTAT® Cplus
4
Bioprocess Solutions Division
Competition Overview
3
![Page 56: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/56.jpg)
Price
System performance
BioFlo 510BioBench
Techfors-S
BIOSTAT® Cplus
Competition overview
BIOSTAT® Cplus
![Page 57: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/57.jpg)
BIOSTAT® Cplus Applikon BioBench
Competition overview
BIOSTAT® Cplus
![Page 58: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/58.jpg)
Electrical heating for operation Steam required for vessel sterilization
Choice of steam heating or electrical heating operation and sterilization
Open loop standard: Closed loop option Closed loop pressurized water temp. control system
Touch screen displayTouch screen display
2:1 onlyVessel aspect H:D ratio 3:1 and 2:1 (2&5L 2:1 only)
Central jacket onlyFully jacketed vessel
Local controller and PC based SCADA systemLocal controller and PC based SCADA system
Vessel volume WV 5L, 10L, 20LVessel volume WV 2L, 5L, 10L, 15L, 20L, 30L
Top driven agitation system Top driven agitation system
Process control for agitation speed, Temp., pH, pO2, feed controllerOptional: Pressure, Vessel weight
Process control for Agitation speed, Temp., pH, pO2; Foam|Level; 2 x Substrate feed controller; High Foam:Optional up to 2 MFC, Vessel weight, Pressure Redox (M), Turbidity (M), 2 x feeding bottle weight
Open frame supply unit integrated controller designOpen-frame with integrated Control Tower design
Applikon BioBenchBIOSTAT® Cplus
Competition overview
BIOSTAT® Cplus
![Page 59: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/59.jpg)
BIOSTAT® Cplus Infors Techfors-S
Competition overview
BIOSTAT® Cplus
![Page 60: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/60.jpg)
Open loopClosed loop pressurized water temp. control system
Electrical heating for operation Steam required for vessel sterilization
Choice of steam heating or electrical heating operation and sterilization
Jacketed vessel, no upper side portsFully jacketed vessel, 3 upper side ports
Top driven agitation systemTop driven agitation system
Vessel volume WV 5L, 10L, 20LVessel volume WV 2L, 5L, 10L, 15L, 20L, 30L
PC based integrated control and SCADA software Local controller and PC based SCADA system
2:1 onlyVessel aspect H:D ratio 3:1 and 2:1 (2&5L 2:1 only)
Touch screen displayTouch screen display
Process control for Agitation speed, Temp., pH, pO2; Optional: Foam|Level; Substrate feed controller, Vessel weight, Pressure, MFC
Process control for Agitation speed, Temp., pH, pO2; Foam|Level; 2 x Substrate feed controller; High Foam:Optional up to 2 MFC, Vessel weight, Pressure Redox (M), Turbidity (M), 2 x feeding bottle weight
Open-frame with integrated Control Tower designOpen-frame with integrated Control Tower design
Infors Techfors-SBIOSTAT® C plus
Competition overview
BIOSTAT® Cplus
![Page 61: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/61.jpg)
BIOSTAT® Cplus NBS BioFlo® 510
Competition overview
BIOSTAT® Cplus
![Page 62: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/62.jpg)
Steam required for vessel sterilization and operationChoice of steam heating or electrical heating operation and sterilization
Open loop standard: Closed loop option Closed loop pressurized water temp. control system
Touch screen displayTouch screen display
2:1 onlyVessel aspect H:D ratio 3:1 and 2:1 (2&5L 2:1 only)
full jacket Fully jacketed vessel
Local controller and PC based SCADA systemLocal controller and PC based SCADA system
Vessel volume WV 16L, 32LVessel volume WV 2L, 5L, 10L, 15L, 20L, 30L
Top driven agitation system Top driven agitation system
Process control for agitation speed, Temp., pH, pO2, Gas Flow Control, feed controller, vessel weightOption: Redox, Pressure,
Process control for Agitation speed, Temp., pH, pO2; Foam|Level; 2 x Substrate feed controller; High Foam:Optional up to 2 MFC, Vessel weight, Pressure, Redox (M), Turbidity (M), 2 x feeding bottle weight
Open frame supply unit integrated controller designOpen-frame with integrated Control Tower design
NBS BioFlo® 510BIOSTAT® C plus
Competition overview
BIOSTAT® Cplus
![Page 63: NEW BIOSTAT Cplus internal presentation - Sartoriusmicrosite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · - up to 4 MFC for Additive Flow gassing strategy with](https://reader031.fdocuments.in/reader031/viewer/2022020108/5af0f3897f8b9ad0618ec39d/html5/thumbnails/63.jpg)
Page 64
Thank You very much foryour Attention
turning science into solutions