Neurosurgery
Image of valves
Strata™ ValveThe Adjustable Delta® Valve
Neurosurgery
RegularSmall
• 體外調節控制閥的壓力
• 無需反覆更換閥門,減少手術費用
• 避免二次手術創傷
• 抗虹吸閥一體化設計,同時解決虹吸現象和壓力分級
Strata Valve
42866 42856
Neurosurgery
Strata ValveCutaway of Regular Valve
Delta Chamber
Reservoir
Valve Mechanism
Proximal Occluder
Inlet Connector
Outlet Connector
Distal Occluder
Polypropylene Base
Neurosurgery
紅寶石圓錐
流體壓力彈簧
磁性轉軸
性能级别
轉軸定位彈簧
轉軸鎖定彈簧設計,
降低閥門受外界磁場干擾機會
不對稱齒和特殊底盤設計將外磁場影響減小到最小程度
Strata Valve MechanismHow it works...
Neurosurgery
Strata Valve Performance CharacteristicsFive Performance Levels
Neurosurgery
Strata ValveAdjustment Tools
Locator Tool Adjustment ToolIndicator Tool
Neurosurgery
輕鬆四步體外調壓 :
1. Locator Tool 底座定位2. Indicator Tool 指示盤讀數3. Adjustment Tool 磁性轉盤調壓4. Indicator Tool 指示盤確認結果
體外調壓工具,多重確認壓力,確保安全
Strata ValveAdjustment Tools
Neurosurgery
Strata ValveValve Adjustment Reliability
X-ray Hand tool
0.5 1.0
1.0 0.5
1.5 0.5
2.0 2.5
•All initial valve settings were confirmed by X-ray
•All post-operative adjustments were confirmed by X-ray
•238 valve adjustments
•4 instances where the x-ray did not match patient chart
98% ACCURACY RATE
Neurosurgery
Codman Programmable Hakim Valve FDA Study
•377 Patients
–235 First Insertion
–142 Revisions
•540 total valve adjustments.
•118 discrepancies between x-rays and patient charts.
Error Rate of 22%
Neurosurgery
Image of X-Ray
0.5 1.0 1.5 2.0 2.5
Dot Code
Magnet
根據磁性轉軸相對於編碼的方向用 X-Ray確認壓力級別
Strata ValveX-Ray
Neurosurgery
Strata NSC ValvePressure/Flow Ranges
Performance Level Old Description
Performance Level 0.5 Low-Low Pressure
Performance Level 1.0 Low Pressure
Performance Level 1.5 Medium Pressure
Performance Level 2.0 High Pressure
Performance Level 2.5 New higher pressure to match the upper Hakim valve range.
Neurosurgery
競爭對手
Neurosurgery
Integra NeuroCareNovus Valve
Hydrodynamic Leverage 8:1 槓桿14 cm H20 > 10 cm H20 Under drainage 引流不足
Neurosurgery
Impact of Hydrodynamic Leverage
• The hydrodynamic leverage ratio determines the amount of added resistance the anti-siphon mechanism will contribute in the upright position.
• To determine this added resistance, the negative hydrostatic column is divided by the leverage ratio to determine the added resistance.
Example: Hydrodynamic Leverage: 8:1
Negative Hydrostatic Column: 48cm H2O
Additional Shunt Resistance: 48/8 = 6 cm H2O
Neurosurgery
Codman SiphonGuard DeviceSiphonGuard(Flow Restrictor)
Secondary Flow
Primary Flow
Neurosurgery
Codman SiphonGuardFacts
• The siphon guard is merely a flow restrictor. A ball in the center of the device diverts fluid through a tortuous circumferential flow path, when the patient is upright.
• 當病人站立時其圓球會阻塞住主通道 ,CSF由側細支流流通 ,其機械原理是限制其流速。
• The small flow path is highly susceptible to occlusion.• 因其側細支流通道極小 ,阻塞機率極高。• The product was recalled in 2001, reportedly due to
performance issues.• 產品曾於 2001年被原廠招回 ,主要是因 CSF引流問題。